METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY
|
|
|
- Maximillian McCormick
- 9 years ago
- Views:
Transcription
1 METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8020 is used to determine the concentration of various aromatic volatile organic compounds. The following compounds can be determined by this method: Appropriate Technique Direct a Compound Name CAS No. Purge-and-Trap Injection Benzene b b Chlorobenzene b b 1,2-Dichlorobenzene b b 1,3-Dichlorobenzene b b 1,4-Dichlorobenzene b b Ethylbenzene b b Toluene b b Xylenes b b a Chemical Abstract Services Registry Number. b adequate response by this technique. 1.2 Table 1 lists the method detection limit for each target analyte in organic-free reagent water. Table 2 lists the estimated quantitation limit (EQL) for other matrices. 2.0 SUMMARY OF METHOD 2.1 Method 8020 provides chromatographic conditions for the detection of aromatic volatile compounds. Samples can be introduced into the GC using direct injection or purge-and-trap (Method 5030). Ground water samples must be determined using Method A temperature program is used in the gas chromatograph to separate the organic compounds. Detection is achieved by a photo-ionization detector (PID). 2.2 If interferences are encountered, the method provides an optional gas chromatographic column that may be helpful in resolving the analytes from the interferences and for analyte confirmation. 3.0 INTERFERENCES 3.1 Refer to Method 5030 and CD-ROM 8020A - 1 Revision 1
2 3.2 Samples can be contaminated by diffusion of volatile organics (particularly chlorofluorocarbons and methylene chloride) through the sample container septum during shipment and storage. A field sample blank prepared from organic-free reagent water and carried through sampling and subsequent storage and handling can serve as a check on such contamination. 4.0 APPARATUS AND MATERIALS 4.1 Gas chromatograph Gas Chromatograph - Analytical system complete with gas chromatograph suitable for on-column injections or purge-and-trap sample introduction and all required accessories, including detectors, column supplies, recorder, gases, and syringes. A data system for measuring peak heights and/or peak areas is recommended Columns Column 1: 6 ft x in ID #304 stainless steel or glass column packed with 5% SP-1200 and 1.75% Bentone-34 on 100/120 mesh Supelcoport, or equivalent Column 2: 8 ft x 0.1 in ID stainless steel or glass column packed with 5% 1,2,3-Tris(2-cyanoethoxy)propane on 60/80 mesh Chromosorb W-AW, or equivalent Detector - Photoionization (PID) (h-nu Systems, Inc. Model PI or equivalent). 4.2 Sample introduction apparatus - Refer to Method 5030 for the appropriate equipment for sample introduction purposes. 4.3 Syringes - A 5 ml Luerlok glass hypodermic and a 5 ml, gas-tight with shutoff valve. 4.4 Volumetric flask, Class A - Appropriate sizes with ground glass stoppers. 4.5 Microsyringe - 10 and 25 µl with a in ID needle (Hamilton 702N or equivalent) and a 100 µl. 4.6 Analytical balance g. 5.0 REAGENTS 5.1 Organic-free reagent water. All references to water in this method refer to organic-free reagent water, as defined in Chapter One. 5.2 Methanol (CH OH) - pesticide quality or equivalent. Store away from 3 other solvents. CD-ROM 8020A - 2 Revision 1
3 5.3 Stock standards - Stock solutions may be prepared from pure standard materials or purchased as certified solutions. Prepare stock standards in methanol using assayed liquids. Because of the toxicity of benzene and 1,4-dichlorobenzene, primary dilutions of these materials should be prepared in a hood Place about 9.8 ml of methanol in a 10 ml tared ground glass stoppered volumetric flask. Allow the flask to stand, unstoppered, for about 10 min or until all alcohol wetted surfaces have dried. Weigh the flask to the nearest g Using a 100 µl syringe, immediately add two or more drops of assayed reference material to the flask; then reweigh. The liquid must fall directly into the alcohol without contacting the neck of the flask Reweigh, dilute to volume, stopper, and then mix by inverting the flask several times. Calculate the concentration in milligrams per liter (mg/l) from the net gain in weight. When compound purity is assayed to be 96% or greater, the weight may be used without correction to calculate the concentration of the stock standard. Commercially prepared stock standards may be used at any concentration if they are certified by the manufacturer or by an independent source Transfer the stock standard solution into a Teflon-sealed o screw-cap bottle. Store, with minimal headspace, at 4 C and protect from light All standards must be replaced after 6 months, or sooner if comparison with check standards indicates a problem. 5.4 Secondary dilution standards: Using stock standard solutions, prepare in methanol secondary dilution standards, as needed, that contain the compounds of interest, either singly or mixed together. The secondary dilution standards should be prepared at concentrations such that the aqueous calibration standards prepared in Section 5.5 will bracket the working range of the analytical system. Secondary dilution standards should be stored with minimal headspace for volatiles and should be checked frequently for signs of degradation or evaporation, especially just prior to preparing calibration standards from them. 5.5 Calibration standards: Calibration standards at a minimum of five concentrations are prepared in organic-free reagent water from the secondary dilution of the stock standards. One of the concentrations should be at a concentration near, but above, the method detection limit. The remaining concentrations should correspond to the expected range of concentrations found in real samples or should define the working range of the GC. Each standard should contain each analyte for detection by this method (e.g., some or all of the compounds listed in the target analyte list may be included). In order to prepare accurate aqueous standard solutions, the following precautions must be observed Do not inject more than 20 µl of alcoholic standards into CD-ROM 8020A - 3 Revision 1
4 100 ml of organic-free reagent water Use a 25 µl Hamilton 702N microsyringe or equivalent (variations in needle geometry will adversely affect the ability to deliver reproducible volumes of methanolic standards into water) Rapidly inject the alcoholic standard into the filled volumetric flask. Remove the needle as fast as possible after injection Mix aqueous standards by inverting the flask three times only Fill the sample syringe from the standard solution contained in the expanded area of the flask (do not use any solution contained in the neck of the flask) Never use pipets to dilute or transfer samples or aqueous standards Aqueous standards are not stable and should be discarded after 1 hr, unless properly sealed and stored. The aqueous standards can be stored up to 24 hr, if held in sealed vials with zero headspace. 5.6 Internal standards (if internal standard calibration is used): To use this approach, the analyst must select one or more internal standards that are similar in analytical behavior to the compounds of interest. The analyst must further demonstrate that the measurement of the internal standard is not affected by method or matrix interferences. Because of these limitations, no internal standard can be suggested that is applicable to all samples. Alpha,alpha,alpha-trifluorotoluene has been used successfully as an internal standard Prepare calibration standards at a minimum of five concentrations for each parameter of interest as described in Section Prepare a spiking solution containing each of the internal standards using the procedures described in Sections 5.3 and 5.4. It is recommended that the secondary dilution standard be prepared at a concentration of 15 mg/l of each internal standard compound. The addition of 10 µl of this standard to 5.0 ml of sample or calibration standard would be equivalent to 30 µg/l Analyze each calibration standard according to Section 7.0, adding 10 µl of internal standard spiking solution directly to the syringe. 5.7 Surrogate standards: The analyst should monitor both the performance of the analytical system and the effectiveness of the method in dealing with each sample matrix by spiking each sample, standard, and organic-free reagent water blank with surrogate compounds (bromochlorobenzene, bromofluorobenzene, 1,1,1- trifluorotoluene, fluorobenzene, and difluorobenzene are recommended) which encompass the range of the temperature program used in this method. From stock standard solutions prepared as in Section 5.3, add a volume to give 750 µg of CD-ROM 8020A - 4 Revision 1
5 each surrogate to 45 ml of organic-free reagent water contained in a 50 ml volumetric flask, mix, and dilute to volume for a concentration of 15 ng/µl. Add 10 µl of this surrogate spiking solution directly into the 5 ml syringe with every sample and reference standard analyzed. If the internal standard calibration procedure is used, the surrogate compounds may be added directly to the internal standard spiking solution (Section 5.6.2). 6.0 SAMPLE COLLECTION, PRESERVATION, AND HANDLING 6.1 See the introductory material to this chapter, Organic Analytes, Section PROCEDURE 7.1 Volatile compounds are introduced into the gas chromatograph either by direct injection or purge-and-trap (Method 5030). Method 5030 may be used directly on ground water samples or low-concentration contaminated soils and sediments. For medium-concentration soils or sediments, methanolic extraction, as described in Method 5030, may be necessary prior to purge-and-trap analysis. Method 5030 also provides guidance on the analysis of aqueous miscible and nonaqueous miscible liquid wastes (see Section below). 7.2 Gas chromatography conditions (Recommended): Column 1: Carrier gas (He) flow rate: 36 ml/min For lower boiling compounds: o Initial temperature: 50 C, hold for 2 min; o o o Temperature program: 50 C to 90 C at 6 C/min, hold until all compounds have eluted. For higher boiling range of compounds: o Initial temperature: 50 C, hold for 2 min; o o o Temperature program: 50 C to 110 C at 3 C/min, hold until all compounds have eluted. Column 1 provides outstanding separations for a wide variety of aromatic hydrocarbons. Column 1 should be used as the primary analytical column because of its unique ability to resolve para-, meta-, and orthoaromatic isomers Column 2: Carrier gas (He) flow rate: 30 ml/min o Initial temperature: 40 C, hold for 2 min; o o o Temperature program: 40 C to 100 C at 2 C/min, hold until all compounds have eluted. Column 2, an extremely high polarity column, has been used for a number of years to resolve aromatic hydrocarbons from alkanes in complex CD-ROM 8020A - 5 Revision 1
6 samples. However, because resolution between some of the aromatics is not as efficient as with Column 1, Column 2 should be used as a confirmatory column. 7.3 Calibration: Refer to Method 8000 for proper calibration techniques. Use Table 1 and especially Table 2 for guidance on selecting the lowest point on the calibration curve Calibration must take place using the same sample introduction method that will be used to analyze actual samples (see Section 7.4.1) The procedure for internal or external calibration may be used. Refer to Method 8000 for a description of each of these procedures. 7.4 Gas chromatographic analysis: Introduce volatile compounds into the gas chromatograph using either Method 5030 (purge-and-trap method) or the direct injection method. If the internal standard calibration technique is used, add 10 µl of internal standard to the sample prior to purging Direct injection: In very limited applications (e.g., aqueous process wastes), direct injection of the sample into the GC system with a 10 µl syringe may be appropriate. The detection limit is very high (approximately 10,000 µg/l); therefore, it is only permitted when concentrations in excess of 10,000 µg/l are expected or for water soluble compounds that do not purge. The system must be calibrated by direct injection (bypassing the purgeand-trap device). Non-aqueous miscible wastes may also be analyzed by direct injection if the concentration of target analytes in the sample falls within the calibration range. If dilution of the sample is necessary, follow the guidance for High Concentration samples in Method 5030, Section Method 8000 provides instructions on the analysis sequence, appropriate dilutions, establishing daily retention time windows, and identification criteria. Include a mid-concentration standard after each group of 10 samples in the analysis sequence Table 1 summarizes the estimated retention times and detection limits for a number of organic compounds analyzable using this method. An example of the separation achieved by Column 1 is shown in Figure 1. Figure 2 shows an example of the separation achieved using Column Record the sample volume purged or injected and the resulting peak sizes (in area units or peak heights) Calculation of concentration is covered in Method CD-ROM 8020A - 6 Revision 1
7 7.4.6 If analytical interferences are suspected, or for the purpose of confirmation, analysis using the second GC column is recommended If the response for a peak is off scale, i.e., beyond the calibration range of the standards, prepare a dilution of the sample with organic-free reagent water. The dilution must be performed on a second aliquot of the sample which has been properly sealed and stored prior to use. 8.0 QUALITY CONTROL 8.1 Refer to Chapter One for specific quality control procedures and Method 8000 for gas chromatographic procedures. Quality control to ensure the proper operation of the purge-and-trap device is covered in Method Quality control required to validate the GC system operation is found in Method The quality control check sample concentrate (Method 8000) should contain each parameter of interest at a concentration of 10 mg/l in methanol Table 3 indicates the calibration and QC acceptance criteria for this method. Table 4 gives method accuracy and precision as functions of concentration for the analytes of interest. The contents of both tables should be used to evaluate a laboratory's ability to perform and generate acceptable data by this method. 8.3 Calculate surrogate standard recovery on all samples, blanks, and spikes. Determine if recovery is within limits (limits established by performing QC procedure outlined in Method 8000) If recovery is not within limits, the following is required. Check to be sure that there are no errors in calculations, surrogate solutions and internal standards. Also, check instrument performance. Recalculate the data and/or reanalyze the extract if any of the above checks reveal a problem. Reextract and reanalyze the sample if none of the above are a problem or flag the data as "estimated concentration". 9.0 METHOD PERFORMANCE 9.1 This method was tested by 20 laboratories using organic-free reagent water, drinking water, surface water, and three industrial wastewaters spiked at six concentrations over the range µg/l. Single operator precision, CD-ROM 8020A - 7 Revision 1
8 overall precision, and method accuracy were found to be directly related to the concentration of the parameter and essentially independent of the sample matrix. Linear equations to describe these relationships are presented in Table The accuracy and precision obtained will be determined by the sample matrix, sample introduction technique, and by the calibration procedure used. 9.3 The method detection limits reported in Table 1 were generated under optimum analytical conditions by an Agency contractor (Ref. 7) as guidance, and may not be readily achievable by all laboratories at all times REFERENCES 1. Bellar, T.A., and J.J. Lichtenberg, J. Amer. Water Works Assoc., 66(12), pp , Bellar, T.A., and J.J. Lichtenberg, "Semi-Automated Headspace Analysis of Drinking Waters and Industrial Waters for Purgeable Volatile Organic Compounds", in Van Hall (ed.), Measurement of Organic Pollutants in Water and Wastewater, ASTM STP 686, pp , Dowty, B.J., S.R. Antoine, and J.L. Laseter, "Quantitative and Qualitative Analysis of Purgeable Organics by High Resolution Gas Chromatography and Flame Ionization Detection", in Van Hall, ed., Measurement of Organic Pollutants in Water and Wastewater. ASTM STP 686, pp , Development and Application of Test Procedures for Specific Organic Toxic Substances in Wastewaters. Category 11 - Purgeables and Category 12 - Acrolein, Acrylonitrile, and Dichlorodifluoromethane. Report for EPA Contract "EPA Method Validation Study 24, Method 602 (Purgeable Aromatics)", report for EPA Contract U.S. EPA 40 CFR Part 136, "Guidelines Establishing Test Procedures for the Analysis of Pollutants Under the Clean Water Act; Final Rule and Interim Final Rule and Proposed Rule", October 26, Gebhart, J.E., S.V. Lucas, S.J. Naber, A.M. Berry, T.H. Danison and H.M. Burkholder, "Validation of SW-846 Methods 8010, 8015, and 8020"; Report for EPA Contract , Work Assignment 2-15; US EPA, EMSL- Cincinnati, 1987." CD-ROM 8020A - 8 Revision 1
9 TABLE 1. CHROMATOGRAPHIC CONDITIONS AND METHOD DETECTION LIMITS FOR AROMATIC VOLATILE ORGANICS Retention time Method (min) detection limit Compound Col. 1 Col. 2 (µg/l) Benzene b Chlorobenzene ,4-Dichlorobenzene ,3-Dichlorobenzene ,2-Dichlorobenzene Ethyl Benzene Toluene Xylenes a Using purge-and-trap method (Method 5030). See Sec b Chlorobenzene and m-xylene may co-elute on some columns. TABLE 2. DETERMINATION OF ESTIMATED QUANTITATION LIMITS (EQLs) FOR VARIOUS MATRICES a Matrix Factor Ground water 10 Low-concentration soil 10 Water miscible liquid waste 500 High-concentration soil and sludge 1250 Non-water miscible waste 1250 a EQL = [Method detection limit (see Table 1)] X [Factor found in this table]. For non-aqueous samples, the factor is on a wet-weight basis. Sample EQLs are highly matrix-dependent. The EQLs determined herein are provided for guidance and may not always be achievable. CD-ROM 8020A - 9 Revision 1
10 TABLE 3. QC ACCEPTANCE CRITERIA a Range Limit Range_ Range for Q for s for x P, Ps Parameter (µg/l) (µg/l) (µg/l) (%) Benzene Chlorobenzene ,2-Dichlorobenzene ,3-Dichlorobenzene ,4-Dichlorobenzene Ethylbenzene Toluene Q = Concentration measured in QC check sample, in µg/l. s = Standard deviation of four recovery measurements, in µg/l. _ x = Average recovery for four recovery measurements, in µg/l. P, P = Percent recovery measured. s a Criteria from 40 CFR Part 136 for Method 602, using packed columns, and were calculated assuming a check sample concentration of 20 µg/l. These criteria are based directly upon the method performance data in Table 4. Where necessary, the limits for recovery have been broadened to assure applicability of the limits to concentrations below those used to develop Table 1. When capillary columns are used, see Method 8021 for performance data. CD-ROM 8020A - 10 Revision 1
11 TABLE 4. METHOD ACCURACY AND PRECISION AS FUNCTIONS OF CONCENTRATION Accuracy, as Single analyst Overall recovery, x' precision, s r' precision, Parameter (µg/l) (µg/l) S' (µg/l) Benzene 0.92C x+0.59 _ 0.21x+0.56 _ Chlorobenzene 0.95C x+0.23 _ 0.17x+0.10 _ 1,2-Dichlorobenzene 0.93C x-0.04 _ 0.22x+0.53 _ 1,3-Dichlorobenzene 0.96C x-0.10 _ 0.19x+0.09 _ 1,4-Dichlorobenzene 0.93C x+0.28 _ 0.20x+0.41 _ Ethylbenzene 0.94C x+0.46 _ 0.26x+0.23 _ Toluene 0.94C x x+0.71 x' = Expected recovery for one or more measurements of a sample containing concentration C, in µg/l. s r' = Expected single analyst standard _ deviation of measurements at an average concentration of x, in µg/l. S' = Expected interlaboratory standard _ deviation of measurements at an average concentration found of x, in µg/l. C = True value for the concentration, in µg/l. _ x = Average recovery found for measurements of samples containing a concentration of C, in µg/l. CD-ROM 8020A - 11 Revision 1
12 Figure 1 Chromatogram of Aromatic Volatile Organics (column 1 conditions) CD-ROM 8020A - 12 Revision 1
13 Figure 2 Chromatogram of Aromatic Volatile Organics (column 2 conditions) CD-ROM 8020A - 13 Revision 1
14 METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY CD-ROM 8020A - 14 Revision 1
1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.
Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative
METHOD 8310 POLYNUCLEAR AROMATIC HYDROCARBONS
METHOD 8310 POLYNUCLEAR AROMATIC HYDROCARBONS 1.0 SCOPE AND APPLICATION 1.1 Method 8310 is used to determine the concentration of certain polynuclear aromatic hydrocarbons (PAH) in ground water and wastes.
METHOD FOR THE DETERMINATION OF VOLATILE PETROLEUM HYDROCARBONS (VPH)
METHOD FOR THE DETERMINATION OF VOLATILE PETROLEUM HYDROCARBONS (VPH) Massachusetts Department of Environmental Protection Division of Environmental Analysis (Senator William X. Wall Experiment Station)
With proper calibration, the upper limit may be extended as needed.
While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the
HEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0.
HEXANES Prepared at the 51st JECFA (1998), published in FNP 52 Add 6 (1998) superseding specifications prepared at the 14th JECFA (1970), published in NMRS 48B (1971) and in FNP 52 (1992). ADI "limited
METHOD 3510C SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION
METHOD 3510C SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION 1.0 SCOPE AND APPLICATION 1.1 This method describes a procedure for isolating organic compounds from aqueous samples. The method also describes concentration
ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
NWTPH-HCID. Hydrocarbon Identification Method for Soil and Water
NWTPH-HCID Hydrocarbon Identification Method for Soil and Water Summary This method is a qualitative and semi-quantitative procedure. It is used for groundwater or surface water, and soil/sediment from
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography
Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine
METHOD 8082A. POLYCHLORINATED BIPHENYLS (PCBs) BY GAS CHROMATOGRAPHY
METHOD 8082A POLYCHLORINATED BIPHENYLS (PCBs) BY GAS CHROMATOGRAPHY SW-846 is not intended to be an analytical training manual. Therefore, method procedures are written based on the assumption that they
Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*
Method 552.1 Revision 1.0 Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD* UCT Products: EUQAX156 (quaternary amine with Cl - counter ion, 6 ml cartridge)** CLTTP050 (CLEAN-THRU
GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE
ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas
METHOD 550. July 1990. J.W. Hodgeson. W.J. Bashe (Technology Applications Inc.) T.V. Baker (Technology Applications Inc.)
METHOD 550 DETERMINATION OF POLYCYCLIC AROMATIC HYDROCARBONS IN DRINKING WATER BY LIQUID-LIQUID EXTRACTION AND HPLC WITH COUPLED ULTRAVIOLET AND FLUORESCENCE DETECTION July 1990 J.W. Hodgeson W.J. Bashe
ß-CYCLODEXTRIN SYNONYMS
ß-CYCLODEXTRIN Prepared at the 44th JECFA (1995), published in FNP 52 Add 3 (1995) superseding specifications prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications
Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection
52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis
METHOD 8031 ACRYLONITRILE BY GAS CHROMATOGRAPHY
METHOD 8031 ACRYLONITRILE BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Methd 8031 is used t determine the cncentratin f acrylnitrile in water. This methd may als be applicable t ther matrices. The
Version 1.0 September, 2008. Jean W. Munch and Paul E. Grimmett
EPA/600/R-08/101 METHOD 522 DETERMINATION OF 1,4-DIOXANE IN DRINKING WATER BY SOLID PHASE EXTRACTION (SPE) AND GAS CHROMATOGRAPHY/ MASS SPECTROMETRY (GC/MS) WITH SELECTED ION MONITORING (SIM) Version 1.0
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
METHOD 8082. POLYCHLORINATED BIPHENYLS (PCBs) BY GAS CHROMATOGRAPHY
METHOD 8082 POLYCHLORINATED BIPHENYLS (PCBs) BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8082 is used to determine the concentrations of polychlorinated biphenyls (PCBs) as Aroclors or as
SUCRALOSE. White to off-white, practically odourless crystalline powder
SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA
Gas Chromatography Liner Selection Guide
Gas Chromatography Liner Selection Guide Peter Morgan, Thermo Fisher Scientific, Runcorn, Cheshire, UK Technical Note 20551 Key Words Liner, focus Abstract The liner serves an important function in allowing
EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.
EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer HT3 Application Note Introduction The analysis of dissolved gases in ground water is important in
TANNIC ACID. SYNONYMS Tannins (food grade), gallotannic acid, INS No. 181 DEFINITION DESCRIPTION
TANNIC ACID Prepared at the 39th JECFA (1992), published in FNP Add 1 (1992) superseding specifications prepared at the 35th JECFA (1989), published in FNP 49 (1990) and in FNP 52 (1992). Metals and arsenic
Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2
Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall
Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)
Method OIV-MA-AS313-22 Type II method Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) 1. Introduction
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Introduction. The following definitions may help you better understand the components of the data report.
Introduction This document is designed to help our clients understand the quality control requirements and limitations of data reporting. There are three sections to this document. The first section will
A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water
A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water Sjaak de Koning 1, Mark van Lieshout 2, Hans-Gerd Janssen 2 and Wil van Egmond 1 1 ATAS, De Run 4334, PO Box
THERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles
THERMAL DESORPTION Introduction and Principles Thermal Desorption Thermal desorption is a simple extension of the technique of gas chromatography (GC) It involves the use of heat and a flow of inert gas
ORGANIC SAMPLE PREPARATION
ORGANIC SAMPLE PREPARATION W W W.LA BT E C H S R L.CO M WSPE MANUAL VACUUM MANIFOLD SPE Process control of the flow rate is critical to guarantee reproducible extractions. Differently then any other systems,
This study was performed by Patrick Sullivan and staff at TestAmerica Analytical Testing Corp., Westfield, MA, USA
An Automated Method for the Fractionation of Extractable Petroleum Hydrocarbons (EPH) from Water and Soil Extracts Using the Gilson GX-274 ASPEC System Keywords: Aliphatic Hydrocarbons, Aromatic Hydrocarbons,
METHOD 8330A NITROAROMATICS AND NITRAMINES BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)
METHOD 8330A NITROAROMATICS AND NITRAMINES BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) SW-846 is not intended to be an analytical training manual. Therefore, method procedures are written based on
Pesticide Analysis by Mass Spectrometry
Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes Brussels, 18th May 2005 CED No 494 rev 2 Final TAXUD/3711/2004 POETRY:
ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS
1 ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS Buck Scientific Atomic Absorption Spectrophotometer, Model 200 Atomic absorption spectroscopy (AAS) has for many years
The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port
APPLICATION NOTE Gas Chromatography/ Mass Spectrometry Authors: Andrew Tipler Sheila Eletto PerkinElmer, Inc. Shelton, CT The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory
GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits
Issue Implementation Page 1 of 7 GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits Scope and Application This method will determine the level of vanillin and ethyl vanillin in distilled
Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity
For the benefit of VaporFi, Inc. Samples RMN070714 Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity GC/MS Analysis (QG02) Anabasine, Anatabine, Myosmine, Nornicotine Enthalpy Analytical,
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection
Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection Application Note 1097 Liu Qian, Zheng Hongguo and
STANDARD OPERATING PROCEDURES
PAGE: 1 of 33 CONTENTS 1.0 SCOPE AND APPLICATION 2.0 METHOD SUMMARY 3.0 SAMPLE PRESERVATION, CONTAINERS, HANDLING, AND STORAGE 4.0 INTERFERENCES AND POTENTIAL PROBLEMS 5.0 EQUIPMENT/APPARATUS 6.0 REAGENTS
A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas
A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas Paper #46 Heidi C. Hayes Air Toxics, Ltd., 180 Blue Ravine Rd. Ste. B, Folsom, CA 95630 Diane J.
New Jersey Department of Environmental Protection Division of Remediation Management Response Site Remediation Program
New Jersey Department of Environmental Protection Division of Remediation Management Response Site Remediation Program NJDEP- SRP Low Level USEPA TO-15 Method (NJDEP-LLTO-15-3/2007) March 2007 March 2009
Aliphatic/Aromatic Separation of Extractable Petroleum Hydrocarbons in Solids or Water by Silica gel Fractionation
ANALYTICAL METHOD 7 FOR CONTAMINATED SITES Aliphatic/Aromatic Separation of Extractable Petroleum Hydrocarbons in Solids or Water by Silica gel Fractionation Prepared pursuant to Section 64 of the Contaminated
Dissolved and precipitated oxalate
Accepted 2005 Process liquors from bleach plants Dissolved and precipitated oxalate Using Ion Chromatography 0 Introduction In bleach plants of pulp mills with a high degree of system closure, there is
METHOD 3545 PRESSURIZED FLUID EXTRACTION (PFE)
METHOD 3545 PRESSURIZED FLUID EXTRACTION (PFE) 1.0 SCOPE AND APPLICATION 1.1 Method 3545 is a procedure for extracting water insoluble or slightly water soluble semivolatile organic compounds from soils,
CCME. Reference Method for the Canada- Wide Standard for Petroleum Hydrocarbons in Soil - Tier 1 Method
CCME Canadian Council of Ministers of the Environment Le Conseil canadien des ministres de l environnement Reference Method for the Canada- Wide Standard for Petroleum Hydrocarbons in Soil - Tier 1 Method
ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE
ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of carbohydrates in foodstuff (LC/RID method) Exercise guarantor: Assoc. Prof. Karel Cejpek, Ph.D. CONTENT Required knowledge...
METHOD 353.2 DETERMINATION OF NITRATE-NITRITE NITROGEN BY AUTOMATED COLORIMETRY
METHOD 353.2 DETERMINATION OF NITRATE-NITRITE NITROGEN BY AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 2.0 August 1993 ENVIRONMENTAL MONITORING
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.
Halogen Free: What, Why and How Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc. We will discuss What are Halogens and how we interact with them. Why Halogens, within consumer
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
GAS CHROMATOGRAPHY. DETERMINING THE PERCENT COMPOSITION OF A MIXTURE OF VOLATILE LIQUIDS USING A GAS CHROMATOGRAPH. (Revised: 1-12-93)
DETERMINING THE PERCENT COMPOSITION OF A MIXTURE OF VOLATILE LIQUIDS USING A GAS CHROMATOGRAPH. (Revised: 1-12-93) INTRODUCTION Gas chromatography is a technique used to analyze mixtures. The instrument
Determination of Caffeine in Beverages HPLC-1
Determination of in s HPLC-1 Introduction This experiment provides an introduction to the application of High Performance Liquid Chromatography (HPLC) to the solution of complex analytical problems. Cola
Automation of Solid Phase Extraction and Column Chromatographic Cleanup
Automation of Solid Phase Extraction and Column Chromatographic Cleanup Haibin Wan Ltd. Problems with Manual Operation 1. Unstable flow rate affects reproducibility and throughput. 2. Not convenient for
Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan. Interim Final SECTION 10
Technical Guidance Manual for the Implementation of the Hawai`i State Contingency Plan SECTION 10 DATA QUALITY ASSURANCE AND QUALITY CONTROL (QA/QC) Hawai`i HEER TGM 10-1 December 31, 2008 CONTENTS Acronyms
Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids
Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C
[ Care and Use Manual ]
The Waters GPC Cleanup System uses two Envirogel GPC Cleanup Columns to perform routine cleanup of environmental samples as specified by EPA Method 3640A, GPC Cleanup. The columns, 19 x 150 mm and 19 x
Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.
Gas Chromatography Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC. Samples in 10mL sealed glass vials were placed in the MPS-2
Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database
C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting
Lecture Chromo-3: Gas Chromatography. CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography. Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder
Lecture Chromo-3: Gas Chromatography CHEM 5181 Fall 2004 Mass Spectrometry & Chromatography Jessica Gilman and Prof. Jose-Luis Jimenez CU-Boulder Outline Introduction Instrument overview Carrier gas Sample
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
Identification of Arson Accelerants by Gas Chromatography
Identification of Arson Accelerants by Gas Chromatography Purpose Arson is an insidious crime that annually claims the lives of hundreds of Americans and costs billions of dollars. 1 Scientists and criminal
Fractional Distillation and Gas Chromatography
Fractional Distillation and Gas Chromatography Background Distillation The previous lab used distillation to separate a mixture of hexane and toluene based on a difference in boiling points. Hexane boils
23 The Thermal Conductivity Detector
23 The Thermal Conductivity Detector General Information TCD pneumatics Conditions that prevent the detector from operating Filament passivation Carrier, reference, and makeup gas Negative polarity Analyzing
Purification of reaction mixtures using flash chromatography.
Purification of reaction mixtures using flash chromatography. This technical note details the use of ISOLUTE Flash chromatography columns for the purification of reaction mixtures. What is flash chromatography?
In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..
Identification of a Substance by Physical Properties 2009 by David A. Katz. All rights reserved. Permission for academic use provided the original copyright is included Every substance has a unique set
APPENDIX C. DATA VALIDATION REPORT
APPENDIX C. DATA VALIDATION REPORT T-117 Upland Investigation Upland Investigation Data Report Appendices PROJECT NARRATIVE Basis for the Data Validation This report summarizes the results of the validation
U.S. Department of Labor Occupational Safety & Health Administration
PHOSGENE - (Organic Method #61) Page 1 of 20 U.S. Department of Labor Occupational Safety & Health Administration www.osha.gov Technical Links > Sampling & Analytical Methods > Index Search Advanced Search
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS
Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS UCT Part Numbers: ECQUEU750CT-MP - QuEChERS, Mylar packs containing 4 g magnesium sulfate, 1 g sodium chloride, 0.5 g sodium citrate dibasic
Expectations for GC-MS Lab
Expectations for GC-MS Lab Since this is the first year for GC-MS to be used in Dr. Lamp s CHEM 322, the lab experiment is somewhat unstructured. As you move through the two weeks, I expect that you will
LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)
October 2006 RESTRICTED LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) DRAFT FOR DISCUSSION World Health Organization 2006 All rights reserved. This draft is intended for
1.2 This TAP is applicable to concrete, charcoal, coral, debris, wipes and soil.
March 2012 8 1 of 17 1.0 SCOPE 1.1 This procedure describes the preparation and analysis of solid matrices and wipe samples. 1.2 This TAP is applicable to concrete, charcoal, coral, debris, wipes and soil.
α-cyclodextrin SYNONYMS α-schardinger dextrin, α-dextrin, cyclohexaamylose, cyclomaltohexaose, α- cycloamylase
α-cyclodextrin New specifications prepared at the 57th JECFA (2001) and published in FNP 52 Add 9 (2001). An ADI not specified was established at the 57th JECFA (2001). SYNONYMS α-schardinger dextrin,
Paper Chromatography: Separation and Identification of Five Metal Cations
Paper Chromatography: Separation and Identification of Five Metal Cations Objectives Known and unknown solutions of the metal ions Ag +, Fe 3+, Co 2+, Cu 2+ and Hg 2+ will be analyzed using paper chromatography.
Determination of Tricyclazole in Water Using Solid Phase Extraction and Liquid Chromatography
Journal of Liquid Chromatography & Related Technologies 1, 32: 2712 2720, 2009 Copyright # Taylor & Francis Group, LLC ISSN: 1082-6076 print/1520-572x online DOI: 10.1080/10826070903245839 Determination
Quality Assurance Manual
Clinical Laboratory of San Bernardino, Inc. Quality Assurance Manual Clinical Laboratory of San Bernardino, Inc. 21881 Barton Road, Grand Terrace, CA 92313 Telephone (909) 825-7693 Facsimile (909) 825-7696
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
CHAPTER ONE TABLE OF CONTENTS 1.0 INTRODUCTION... 1
CHAPTER ONE TABLE OF CONTENTS Section Page 1.0 INTRODUCTION... 1 2.0 QA PROJECT PLAN... 1 2.1 DATA QUALITY OBJECTIVES... 2 2.2 PROJECT OBJECTIVES... 2 2.3 SAMPLE COLLECTION... 3 2.4 ANALYSIS AND TESTING...
MOLECULAR WEIGHT BY BOILING POINT ELEVATION
MOLECULAR WEIGHT BY BOILING POINT ELEVATION BACKGROUND This experiment demonstrates the use of colligative properties. The goal is to measure the molecular weight of a non-volatile solute by determining
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
Tamsulosin Hydrochloride Capsules
. nal Revision Bulletin Official October 1, 2011 Tamsulosin 1 standard solution, and shake well. Centrifuge at 1500 rpm for 10 min, and use the supernatant, passing it if Tamsulosin Hydrochloride Capsules
Ecology Quality Assurance Glossary
Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
S olid Ph ase Extraction (SP E) Method:
Solid Phase Extraction of US EPA 535 for Chloroacetanilide and Acetamide Degradates in Drinking Water Samples Prior to Liquid Chromatography/ Tandem Mass Spectrometry Toni Hofhine, Horizon Technology,
Appendix D lists the Field Services Standard Operating Procedures. Appendix E lists the Biological Monitoring Standard Operating Procedures.
Page 16 of 87 3.3 Sample Collection, Storage and Preservation Figure 3 details required containers, sample volumes, preservation techniques, and holding times for proper sample collection. A discussion
Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation
Application Note Determination of Nitrite and Nitrate in Fruit Juices by UV Detection Category Food Matrix Fruit Juice Method HPLC Keywords Ion pair chromatography, fruit juice, inorganic anions AZURA
Split / Splitless Injection for Capillary GC
Split / Splitless Injection for Capillary GC Speakers John V Hinshaw CHROMacademy GC Dean GC Connections editor Moderator Tony Taylor CHROMacademy Trainer Technical Director, Crawford Scientific Dave Walsh
Selecting Analytical Methods. in Environmental Samples and Gasoline. Alcohols and ethers are added to gasoline to. for the Determination of Oxygenates
Selecting Analytical Methods for the Determination of Oxygenates in Environmental Samples and Gasoline Demand for analyses is higher, but are the methods reliable? By I.A.L. Rhodes and A.W. Verstuyft Alcohols
AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY. Andrew Tipler
AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY Andrew Tipler Chromatography Research and Technology Manager PerkinElmer, Inc. Table of Contents Introduction 3 Fundamental
INSTRUCTIONS 56-1190-98. Edition AC
Sephacryl S-100 High Resolution Sephacryl S-200 High Resolution Sephacryl S-300 High Resolution Sephacryl S-400 High Resolution Sephacryl S-500 High Resolution INSTRUCTIONS Sephacryl High Resolution chromatography
CORESTA RECOMMENDED METHOD Nº 80
CORESTA RECOMMENDED METHOD Nº 80 USE OF THE PART-FILTER METHOD FOR THE ESTIMATION OF SMOKERS EXPOSURE TO NICOTINE AND NICOTINE-FREE DRY PARTICULATE MATTER (January 2016) 0. INTRODUCTION The CORESTA Smoking