Hydras 3 LT Quick Start
|
|
|
- Bruce Toby Henderson
- 10 years ago
- Views:
Transcription
1 Catalog Number Hydras 3 LT Quick Start SOFTWARE MANUAL December 2005, Edition 2
2 Hach Company, All rights reserved. Printed in the U.S.A.
3 Catalog Number Hydras 3 LT SOFTWARE MANUAL December 2005, Edition 2 Hach Company, All rights reserved. Printed in the U.S.A. jk/dk
4 Visit us at
5 Table of Contents Section 1 General Information... 5 Section 2 Startup Downloading Data Operating a Sonde System Tab Online Monitoring Tab Time Series Vertical Profile Manual Profile Stability Check Real-time Graph Functions Log Files Tab Creating a Log File Storing a Template File Recalling Template Files Parameter Setup Tab Calibration Tab Settings Tab Define SDI Parameter Order Software Tab Hydras 3 Options (F2) Graphics Tab DataSonde Tab Section 3 Contact Information Appendix A Troubleshooting
6 Visit us at
7 Section 1 General Information Hydras 3LT is software used to connect Hydrolab sondes to a PC. It automatically scans for connected sondes and recognizes any log files held in memory. Up to 32 sondes can be connected at once, and all log files can be downloaded simultaneously. When operating the sonde, Hydras 3LT provides seven function tabs to help view real-time data, or to set up a sonde for operation: The System tab allows the user to setup basic sonde identification information. Also, passwords can be established to allow up to four different access levels, which can help to protect the data. The Online Monitoring tab allows viewing real-time data as either a time series graph, or as a vertical profile graph. Up to six parameters can be viewed at a time on either graph. Data can also be downloaded into an Excel or text file. The Log Files tab allows quick and easy setup of logging runs with a GUI interface. Choose the start and stop time, the logging interval, as well as the circulator and sensor warm-up time. Add or remove parameters and arrange them in order. When the logging run is complete, download all the files at once. The Parameter Setup tab allows the user to configure available parameters. The Calibration tab sets the calibration parameters for an individual sensor. Choose the sensor to calibrate and enter the calibration standards. The current value and temperature readings are shown for temperature sensitive calibrations. The Settings tab sets the communication settings between the Hydras 3LT and the sonde. The Software tab uploads software to the sonde and updates and removes software drivers. 5
8 Visit us at
9 Section 2 Startup 1. Connect the Data Cable from the computer and to the sonde. 2. Start Hydras 3 LT. The software will automatically scan for sondes. All detected sondes are displayed in the Connected sondes list in the Main window displayed below. If a sonde is not found, press RE-SCAN FOR SONDES. If a connection can not be established, refer to Appendix A Troubleshooting on page 29. 7
10 2.1 Downloading Data After a log file is created in the Log Files tab, the files can be downloaded by checking the appropriate Log File box and clicking DOWNLOAD SELECTED FILES. Multiple files can be downloaded at once. The downloaded log files are stored in the location indicated at the bottom of the screen (typically C:\Program Files\HYDRAS3LT\LogFiles ). The location can be changed by clicking EDIT and specifying a new directory. 2.2 Operating a Sonde Note: To enter a higher level security mode, click on the Level button from the System Tab and enter the password. 1. From the Main Hydras 3 LT menu, highlight the appropriate sonde and press OPERATE SONDE. The sonde window will appear. The window defaults to Security Level The sonde window contains 7 function tabs; System, Online Monitoring, Log Files, Parameter Setup, Calibration, Settings, and Software. 8
11 2.2.1 System Tab The System Tab sets up general information and configures the system components. # Function Description 1 Instrument ID Enter a description of the sonde. This will be used as an output on the data file. 2 Sonde Information Displays static information about the sonde. 3 Clock Sets the sonde date and time to match your PC clock or manually set the date and time (this is useful when monitoring in different time zones). 4 Circulator Select whether to start or stop the circulator. 5 Audio Turns on/off the beeping sound the sonde makes when taking measurements. 6 Security Level Click on LEVEL 0, LEVEL 1, LEVEL 2, or LEVEL 3 and enter the password to view that security mode. While in Level 3, click on PASSWORDS to set the passwords for Level 1, Level 2, and Level 3. The default Level 3 password from the factory is Hydrolab. 9
12 2.2.2 Online Monitoring Tab The Online Monitoring Tab has three modes for capturing data; Manual, Time Series, and Vertical Profile (sections , , and ). Criteria checks can be set before capturing the data (section ) and then the results can be graphed (section ). # Function Description 1 Monitoring Mode The monitoring mode must be set before starting Online Monitoring. Select Time Series for logging values at regular time intervals. Select Vertical Profile for logging values at user-defined depth levels. Or select Manually for User-defined logging. 2 Monitoring Interval Select the time interval or depth level between samples as well as the stability check for Vertical Profile and Manual settings (see section ). 3 Parameters Check the parameters to be monitored. 4 Arrow buttons Select a parameter and use the UP and DOWN arrow buttons to move the parameter in the list. 10
13 5 Data Buttons 1 (Monitoring must begin before the data button functions can be used) New Graph: Opens a graphic window with the selected parameters (max. 6) plotted as a time series. See Real-time Graph Functions on page 15 for more information on changing graph properties. New Depth Graph: Opens a graphic window with the selected parameters (max. 6) displayed at the corresponding depth. New Table: Displays the selected monitored data in a tabular view (real-time). Export Excel: Microsoft Excel must be installed on the computer for this function to work. The monitored data is transferred to an Excel application running in the background. If Excel is not running, it is automatically started and a new worksheet is created. After the data is exported, the Excel file is not updated with new real-time data. Hydras 3LT continues to log such data for graphing and storage purposes. Export Textfile: Writes the monitored data to a text file. After the data is exported, the file is not updated with new real-time data. Transfer to Database: Not available with Hydras 3 LT. 1 The data buttons become available when at least one parameter in the list is selected Time Series A time-based logging can be set by selecting Time Series as the Monitoring Mode. The user must select the time interval and parameters. Click START to begin monitoring data (see section for a screen shot). 11
14 Vertical Profile A vertical profile can be set by selecting Vertical Profile as the Monitoring Mode. The user must select the depth increment for the vertical profile, the units of measure for the profile, and whether the profile is ascending (up arrow) or descending (down arrow). The user should also indicate whether or not a stability check is to be used and how many samples per measurement point are preferred. 12
15 Manual Profile A manual profile can be set by selecting Manually as the Monitoring Mode. The user should also indicate whether or not a stability check is to be used and how many samples per measurement point are preferred. The data is captured by either pressing the space bar or by clicking on the capture button 13
16 Stability Check When monitoring in Depth Mode or Manually, the user can set up stability criteria which must be met prior to Hydras 3LT recording a measurement. The user needs to select a given parameter such as SpCond, the maximum delta (± this number from an average value), the units of measurement, and the number of samples that must meet this stability criteria. To select a parameter of interest, click on the parameter name in the Stability Criteria window. The check box in front of the parameter name is used to indicate whether that parameter will be used as part of the stability criteria check. When more than one stability criteria is checked, all conditions must be met for data to be collected. Click on CONFIGURATION in the Online Monitoring screen to enter the stability criteria screen. 14
17 Real-time Graph Functions Up to six selected parameters can be displayed with different colors and an individual axis range for every sensor. In the graphic options (F2), the user can select if every parameter has its own vertical axis displayed or that only 2 vertical axis are used (one left, one right). To change the scale on the y-axis, click on the minimum and maximum values on the graph. A dialog box will appear, enter the minimum and maximum y-axis values. Press Ctrl-A to autoscale the y-axis at any time while the data is plotted on the graph. Right-click on the graph to display a context menu. See the table below for context menu options. Function Span Grid Overlay Statistics Table Ruler Description Select the time span that is displayed until the current value (1 min, 5 min, 10 min, 30 min, 1 h, 3 h, 6 h, 12 h, or 24 h). Turn the horizontal and vertical grid on or off. Select overlay statistics to show simple statistics on the graph such as max, min, and average. Select table to view the graph in a tabular form. Turn the two rulers on or off. 15
18 2.2.3 Log Files Tab Creating a Log File Users frequently like to use the same Log File settings for multiple sondes or for redeploying a sonde after retrieving a stored log file. The Log Files tab allows the user to share and retrieve template Log File settings using a right-click while the cursor is in the Log Files screen. Note: A log file must be created and then enabled before data can be collected. 1. Click the CREATE button. 2. Enter the name for the new log file. The empty log file is now created. Note: To delete a log file, select the log file in the Log File drop-down menu and click the DELETE button. 3. Enter the start and end time of the log, the logging interval, the sensor warm-up time before logging, and how long before logging the circulator will be turned on, and if audio signals will be used while logging. 4. Select the parameters in the Parameter in sonde list and click the ADD button to place them into the Parameters in log file list. Change the order of the parameters using the UP and DOWN arrow buttons. 5. Click SAVE SETTINGS to send the configuration to the sonde. 6. Click ENABLE to start collecting data. Click DISABLE to stop collecting data during logging. A fully completed logging run will automatically disable at the end of the run. 7. Click DOWNLOAD to download and display the log file. Select Printable or Spreadsheet format. 16
19 Storing a Template File Note: In order to create a stored template file, a sonde must be connected to Hydras 3LT. 1. Open an existing log file by using the Log File pull-down menu or create a new log file by following steps 1 4 in section While the cursor is in the Log Files screen, right-click to open the template selection screen. 3. Select STORE TEMPLATE. 4. Choose a name for the template. It will be stored automatically in the template folder. Reusing a template name will overwrite the old template. 5. Select SAVE. 6. The log file template is now saved and can be used to speed up the creation of future log files. 17
20 Recalling Template Files Note: In order to use a template file, the sonde used to create the template and the sonde used when retrieving the template should be configured in a compatible manner. That is, all parameters used in the template file should be available in the sonde to which the template file is being applied. 1. Click the CREATE button to create a new log file or open an existing log file by using the Log File pull-down menu. 2. Enter a name for the new log file, if necessary. An empty log file is created. 3. While the cursor is in the Log Files screen, right-click to open the template selection screen. 4. Select Load Template. 5. The template will load onto the Log Files screen. It will be identical to the stored template except the Start Logging start time will be the current PC time. 6. Verify that all items are prepared as needed, including the Start Logging and Stop Logging times. 7. Click SAVE SETTINGS to send the configuration to the sonde. 8. Click ENABLE to start collecting data. 18
21 2.2.4 Parameter Setup Tab Click on the appropriate parameter and enter the setup information for each parameter. Press SAVE SETTINGS when all parameter setup information is complete. Refer to the instrument manual for more information for each parameter. 19
22 2.2.5 Calibration Tab Click on the appropriate parameter and enter the calibration information for each parameter. Press CALIBRATE when all the calibration information is complete. Refer to the instrument manual for more calibration information. 20
23 2.2.6 Settings Tab # Function Description 1 Baud Rate 2 SDI 3 MODBUS Select the instrument baud rate (1200, 2400, 4800, 9600, or 19200). Note: If the automatic baud rate scan is enabled (section DataSonde Tab on page 25), Hydras3LT will only scan for 9600 and baud rates. Enable SDI when multiple sondes or other SDI-12 enabled sensors are connected to a single SDI-12 controller. Set the delay to allow the sensors to warm-up and stabilize for accurate measurements. The transmitter factory default SDI-12 address is 0 for all parameters. If continuous mode is enabled, the unit will never enter sleep mode and measurements will be available immediately upon receiving an SDI-12 data request. The Modbus address default is 1. When using multiple sondes, assign a unique address (1 to 247) for each sonde. The Modbus works with even-parity, 8 data bit, and 1 stop bit. 4 Log Files When Auto Log is selected, it works as a back-up logging file. Auto Log captures a reading of all available parameters, battery voltages, and turns on the audio and circulator with a two-minute warm-up (if installed) once every hour. No log files must be selected in the Log Files tab for Autolog to run. 5 Statistics Press the STATISTICS button to view a summary of the sonde. 6 TTY Mode Use this option only when backwards compatibility is required. TTY mode provides data and limited menu access for external devices that interface with earlier generations of sondes. If TTY mode is entered, a user may exit TTY mode by opening a terminal screen and then pressing the space bar followed by a Q or q to quit. 7 Options See Hydras 3 Options (F2) on page
24 Define SDI Parameter Order The user can define the order in which parameters are reported for an SDI 12 data request. In the HydrolabsondeParameterOrderForm screen, the user can add parameters to be reported using SDI 12 data requests. The user can also order the parameters by using the UP and DOWN arrows to arrange the parameters in the desired order. Note: The parameter order set in this screen will also be utilized by TTY mode for reporting data values. 22
25 2.2.7 Software Tab Important Note: Level 3 Security Required. Software version 3.35 or higher is required to use the features on the Software screen. If the sonde has an older version of software, upgrade to the newest software version using terminal mode. Refer to the instrument manual for more information on updating the sonde software using terminal mode. # Function Description 1 Software Version Drivers/Remove To update the firmware, click on the UPDATE button in the Main group box. A prompt to select a file name will appear and then the selected file is uploaded to the sonde. After the upload, the sonde is reset to default values and the sonde window is created again. To remove a driver from a slot, select the driver and click the REMOVE button. After the driver is removed the sonde is reset and the sonde window is created again. 2 To update a driver, select the slot of the driver and click the UPDATE button in the Drivers group box. Select a file name when prompted. The selected file is then uploaded to the selected slot in Drivers/Update the sonde. If Reset Driver After Update is checked, all driver parameters are set to default values. 3 Reset Driver Resets all driver calibration parameters to default values. 4 Reset Sonde Resets all system parameters to default values. 23
26 2.2.8 Hydras 3 Options (F2) Graphics Tab To set the sonde Options select File>Options or press F2 from the sonde startup screen. The options menu contains two tabs, Graphics and DataSonde (sections and ). The Graphics tab adjusts the settings and appearance of the graph (ruler size, labeling, type of grid, and multiple graph functions). Up to six selected parameters can be displayed, with different colors and an individual axis range for every sensor. In the graphic options (F2), the user can select if every parameter has its own vertical axis displayed or that only 2 vertical axis are used (one left, one right). 24
27 DataSonde Tab The user is able to define which serial ports to scan for sondes using the Options screen (F2). Hydras 3LT can be set to scan either all ports or only those ports indicated in the ports list. A default PC baud rate can also be set. An extended timeout can be used when the user is connecting to older Series 4a sondes, typically those built before If a user is having difficulty connecting to a sonde, it is recommended that they set the Communication screen to Auto Scan (all serial ports), check the automatic baud rate scan, and use the extended timeout. The DataSonde tab is used to select the PC baud rate (1200, 2400, 4800, 9600, or 19200), measurement units (Celsius, Fahrenheit, Kelvin, depth, and battery voltage), log file formats (spreadsheet or printable), and format options such as date order, digits for year, date separator, time separator, and radix (decimal point or comma). 25
28 Visit us at
29 Section 3 Contact Information U.S.A. Customers By Telephone: 6:30 a.m. to 5:00 p.m. MST Monday through Friday (800) or (970) By Fax: (970) Ordering Information Required By Mail: Hach Environmental P.O. Box 389 Loveland, Colorado U.S.A. Ordering information by [email protected] Account number (if available) Billing address Your name and phone number Shipping address Purchase order number Catalog number Brief description or model number Quantity International Customers To locate the representative nearest you, send an to: [email protected] or contact: Hach Environmental World Headquarters; Loveland, Colorado, U.S.A. Telephone: (970) ; Fax: (970) Technical and Customer Service (U.S.A. only) Call or [email protected] 27
30 Visit us at
31 Appendix A Troubleshooting If Hydras 3LT does not automatically detect the Sonde when launched, the following screen will appear after a period of time: Note: For some users with more than one active serial port, it will take some time for Hydras 3LT to try to auto-detect sondes on all ports. It may take up to 30 seconds per active serial port to do the auto-scan. If the sondes is not automatically detected, press the RE-SCAN FOR SONDES button. If communication is still not established after several attempts, try the following: Verify the Hardware. 1. Check power cables and connections: a. Verify that your PC and multiprobe are properly connected to the wall outlet or external battery if used. 2. Verify that the input voltage to the multiprobe is between 7V and 14V. 3. If your multiprobe is equipped with an internal battery pack, check the batteries' polarity and voltages. 29
32 Troubleshooting Verify the Hydras3LT Communication Settings. 1. Select File>Options from the Hydras 3LT Connection Screen. 2. If the Sondes/PC baud rate and COM port are known, disable the COM Port and baud rate auto scans and set Hydras3LT to use the known values. This requires Hydras 3LT to concentrate scan activities on only the serial ports of interest. 30
33 Troubleshooting 3. In some cases, extending the connection timeout from 10 to 15 seconds, will improve the connection procedure. This option allows time for additional retries. It is normally disabled to decrease the time taken to scan for a sonde. 31
34 Visit us at
Ovation Operator Workstation for Microsoft Windows Operating System Data Sheet
Ovation Operator Workstation for Microsoft Windows Operating System Features Delivers full multi-tasking operation Accesses up to 200,000 dynamic points Secure standard operating desktop environment Intuitive
User Manual. Thermo Scientific Orion
User Manual Thermo Scientific Orion Orion Star Com Software Program 68X637901 Revision A April 2013 Contents Chapter 1... 4 Introduction... 4 Star Com Functions... 5 Chapter 2... 6 Software Installation
Cart Interface Installation & Control Board Manual for M38 XP and RX Carts and Legacy M39 Carts
Rubbermaid Medical Solutions Cart Interface Installation & Control Board Manual for M38 XP and RX Carts and Legacy M39 Carts REVISED FEBRUARY 25, 2010 Filename: Cart Interface Installation & Control Board
File Management Utility. T u t o r i a l
File Management Utility T u t o r i a l Contents System Requirements... 2 Preparing Files for Transfer to GlobalMark... 2 Application Launch... 2 Printer Setup... 2 Communication Status... 4 Communication
SubCue Analyzer Software - Introduction
SubCue Analyzer Software - Introduction The SubCue Analyzer is a tool to initialize and download the temperature data from implantable SubCue Dataloggers. To begin, open the SubCue Analyzer program from
Model 288B Charge Plate Graphing Software Operators Guide
Monroe Electronics, Inc. Model 288B Charge Plate Graphing Software Operators Guide P/N 0340175 288BGraph (80207) Software V2.01 100 Housel Ave PO Box 535 Lyndonville NY 14098 1-800-821-6001 585-765-2254
User Guide Win7Zilla
User Guide Win7Zilla Table of contents Section 1: Installation... 3 1.1 System Requirements... 3 1.2 Software Installation... 3 1.3 Uninstalling Win7Zilla software... 3 Section 2: Navigation... 4 2.1 Main
User Manual WatchPower
User Manual WatchPower Management Software for Inverter Table of Contents 1. WatchPower Overview... 1 1.1. Introduction... 1 1.2. Features... 1 2. WatchPower Install and Uninstall... 1 2.1. System Requirement...
SA-9600 Surface Area Software Manual
SA-9600 Surface Area Software Manual Version 4.0 Introduction The operation and data Presentation of the SA-9600 Surface Area analyzer is performed using a Microsoft Windows based software package. The
Real Time Monitor. A Real-Time Windows Operator Interface. DDE Compliant. (for remote data display)
Real Time Monitor A Real-Time Windows Operator Interface DDE Compliant (for remote data display) TABLE OF CONTENTS 1. INTRODUCTION...1 1.1 INSTALLATION...2 1.2 FIRST START UP - DDE CONFIGURE...2 1.3 AUTO-STARTUP...2
Data Logging Software and Cable Bundle Model DL-SC3 User Manual. Overview
Data Logging Software and Cable Bundle Model DL-SC3 User Manual Overview Eco Sensors Digital Products are capable of transmitting data to a computer via an RS-232 serial connection. Eco Sensors DL-SC3
TimeValue Software Due Date Tracking and Task Management Software
User s Guide TM TimeValue Software Due Date Tracking and Task Management Software File In Time Software User s Guide Copyright TimeValue Software, Inc. (a California Corporation) 1992-2010. All rights
Global Water Instrumentation, Inc.
Global Water Instrumentation, Inc. 151 Graham Road P.O. Box 9010 College Station, TX 77842-9010 Telephone : 800-876-1172 International : (979) 690-5560, Fax : (979) 690-0440 e-mail : [email protected]
ViewPower. User s Manual. Management Software for Uninterruptible Power Supply Systems
ViewPower User s Manual Management Software for Uninterruptible Power Supply Systems Table of Contents 1. ViewPower Overview... 2 1.1. Introduction...2 1.2. Structure...2 1.3. Applications...2 1.4. Features...2
Navigator Software. Contents 1. NAVIGATOR SOFTWARE 2. INSTALLATION 3. USING NAVIGATOR SOFTWARE 3.1 STARTING THE PROGRAM 3.
Navigator Software Contents 1. NAVIGATOR SOFTWARE 2. INSTALLATION 3. USING NAVIGATOR SOFTWARE 3.1 STARTING THE PROGRAM 3.2 SYSTEM SET UP 3.3 LOAD DATA FILE 3.3.1 LOADING PARTIAL FILES 3.4 DATA TABLE WINDOW
Quick Start Using DASYLab with your Measurement Computing USB device
Quick Start Using DASYLab with your Measurement Computing USB device Thank you for purchasing a USB data acquisition device from Measurement Computing Corporation (MCC). This Quick Start document contains
Guide to Installing BBL Crystal MIND on Windows 7
Guide to Installing BBL Crystal MIND on Windows 7 Introduction The BBL Crystal MIND software can not be directly installed on the Microsoft Windows 7 platform, however it can be installed and run via XP
10 Thomas, Irvine, CA 92618 USA Tel: (949) 465-0900 Fax: (949) 465-0905 Toll Free: (800) 23 FUTEK
Table of Contents Software Overview... - 4 - Product Key... - 5 - Model Selection... - 6 - Device Selection... - 7 - Menu... - 8 - File... - 8 - Edit... - 8 - View... - 8 - Format... - 8 - Help... - 9
HOBOmobile User s Guide Android
HOBOmobile User s Guide Android Onset Computer Corporation 470 MacArthur Blvd. Bourne, MA 02532 www.onsetcomp.com Mailing Address: P.O. Box 3450 Pocasset, MA 02559 3450 Phone: 1 800 LOGGERS (1 800 564
Data Analysis Software
TRAKPRO Data Analysis Software User s Manual November 1996 P/N 1980200 Rev. D TSI Incorporated TRAKPRO Data Analysis Software User s Manual November 1996 P/N 1980200 Rev. D SHIP TO: TSI Incorporated 500
WAMLocal. Wireless Asset Monitoring - Local Food Safety Software. Software Installation and User Guide BA/WAM-L-F
Wireless Asset Monitoring - Local Food Safety Software BA/WAM-L-F Software Installation and User Guide System Overview The BAPI Wireless Asset Monitoring Local (WAM Local) Software receives temperature
User s Manual. Management Software for Inverter
WatchPower User s Manual Management Software for Inverter Table of Contents 1. WatchPower Overview... 2 1.1. Introduction... 2 1.2. Features... 2 2. WatchPower Install and Uninstall... 2 2.1. System Requirement...
SMS Alarm Messenger. Setup Software Guide. SMSPro_Setup. Revision 090210 [Version 2.2]
SMS Alarm Messenger SMSPro_Setup Revision 090210 [Version 2.2] ~ 1 ~ Contents 1. How to setup SMS Alarm Messenger?... 3 2. Install the SMSPro_Setup software... 5 3. Connection Type... 6 4. Connection Port
JEROME COMMUNICATIONS SOFTWARE (JCS) INSTALLATION and OPERATION MANUAL V1.02 for Windows XP
AZI P/N 700-0042-B Last update May 2009 USER MANUAL JEROME COMMUNICATIONS SOFTWARE (JCS) INSTALLATION and OPERATION MANUAL V1.02 for Windows XP May 2009 ARIZONA INSTRUMENT LLC 3375 N Delaware Street Chandler,
Software Manual. For. MadgeTech 4 Data Logger Software
Software Manual For MadgeTech 4 Data Logger Software Introduction 7 About MadgeTech, Inc... 8 Software Overview 9 Getting Started... 10 System Requirements...10 Software Installation: Installing the MadgeTech
OPERATION MANUAL. MV-410RGB Layout Editor. Version 2.1- higher
OPERATION MANUAL MV-410RGB Layout Editor Version 2.1- higher Table of Contents 1. Setup... 1 1-1. Overview... 1 1-2. System Requirements... 1 1-3. Operation Flow... 1 1-4. Installing MV-410RGB Layout
Installation and Operation Manual Portable Device Manager, Windows version
Installation and Operation Manual version version About this document This document is intended as a guide for installation, maintenance and troubleshooting of Portable Device Manager (PDM) and is relevant
User s Manual. Management Software for Uninterruptible Power Supply Systems
PowerShield NetGuard User s Manual Management Software for Uninterruptible Power Supply Systems Table of Contents 1. NetGuard Overview...2 1.1. Introduction...2 1.2. Structure...2 1.3. Applications...3
User s Manual. Management Software for ATS
ATS Monitor User s Manual Management Software for ATS Table of Contents 1. ATS Monitor Overview... 2 2. ATS Monitor Install and Uninstall... 2 2.1. System Requirement... 2 2.2. Software Install... 2 2.3.
Using Avaya Flare Experience for Windows
Using Avaya Flare Experience for Windows Release 9.0 Issue 02.01 September 2013 Contents Chapter 1: About Flare Experience... 5 About Flare Experience... 5 Main window... 6 Button descriptions... 10 Chapter
BMV-602 Data Link Manual
BMV-602 Data Link Manual 1 Contents of the kit 1 x BMV-602 data link cable. 1 x Installation CD 2 System Requirements 2.1 Hardware requirements A BMV-602 with software version 2.03 or higher. A spare RS-232
User's Guide DylosLogger Software Version 1.6
User's Guide DylosLogger Software Version 1.6 The DylosLogger software allows users of Dylos Air Quality Monitors equipped with PC interface to easily record, download, and graph data. The COM port is
Chapter 7 Event Log. 7.1 Event Log Management
Chapter 7 Event Log... 2 7.1 Event Log Management... 2 7.1.1 Excel Editing... 4 7.2 Create a New Event Log... 5 7.2.1 Event (Alarm) Log General Settings... 5 7.2.2 Event (Alarm) Log Message Settings...
Online Monitoring User Guide
High Resolution Temperature Sensing Strip Online Monitoring User Guide 888.637.3282 www.nerdata.com Page 1 of 26 Contents The Aurora Online Monitoring System... 3 1. Creating an Account on the Aurora Online
User s Guide for version 5 Page
Network Management Interface Catalog VPi 32.15E For use with these ATS products: 7000 & 4000 Series (Group 5 Controller) Series 300 (Group 1 Controller), and 940, 962, 436, 434, 447, 448 (Group 6A/7A Control
Veriteq. vlog 4.4 > > > > > User s Guide. w w w. v e r i t e q. c o m. w w w. v a i s a l a. c o m
Veriteq > > > > > vlog 4.4 User s Guide w w w. v e r i t e q. c o m w w w. v a i s a l a. c o m Copyright Copyright 2010 Veriteq Instruments, Inc. All rights reserved. No copying, distribution, publication,
NVMS-1200. User Manual
NVMS-1200 User Manual Contents 1 Software Introduction... 1 1.1 Summary... 1 1.2 Install and Uninstall... 1 1.2.1 Install the Software... 1 2 Login Software... 3 2.1 Login... 3 2.2 Control Panel Instruction...
Gigabyte Management Console User s Guide (For ASPEED AST 2400 Chipset)
Gigabyte Management Console User s Guide (For ASPEED AST 2400 Chipset) Version: 1.4 Table of Contents Using Your Gigabyte Management Console... 3 Gigabyte Management Console Key Features and Functions...
Table of Contents. 1. Overview... 3. 1.1 Materials Required. 3 1.2 System Requirements. 3 1.3 User Mode 3. 2. Installation Instructions..
Table of Contents 1. Overview..... 3 1.1 Materials Required. 3 1.2 System Requirements. 3 1.3 User Mode 3 2. Installation Instructions.. 4 2.1 Installing the On Call Diabetes Management Software. 4 2.2
DOC026.53.80098. PharmSpec 3. Software Manual. 10/2013, Edition 3
DOC026.53.80098 PharmSpec 3 Software Manual 10/2013, Edition 3 Table of Contents Section 1 PharmSpec 3 description...3 1.1 What is new in PharmSpec 3?...3 1.2 Software components...3 Section 2 Installation...5
DAS202Tools v1.0.0 for DAS202 Operating Manual
DAS202Tools v1.0.0 for DAS202 Operating Manual DAT102Tools 1.0.0 Manual Table of context 2 Table of Contents 1 General Information... 3 2 PC... Configuration Requirements 3 3 Software Installation... 3
Legal Notes. Regarding Trademarks. 2012 KYOCERA Document Solutions Inc.
Legal Notes Unauthorized reproduction of all or part of this guide is prohibited. The information in this guide is subject to change without notice. We cannot be held liable for any problems arising from
Capture Pro Software FTP Server System Output
Capture Pro Software FTP Server System Output Overview The Capture Pro Software FTP server will transfer batches and index data (that have been scanned and output to the local PC) to an FTP location accessible
STIM202 Evaluation Kit
Table of contents: 1 FEATURES... 2 2 GENERAL DESCRIPTIONS AND SYSTEM CONTENTS... 2 3 SYSTEM REQUIREMENTS... 2 4 GETTING STARTED... 3 4.1 INSTALLATION OF NI-SERIAL CABLE ASSEMBLY DRIVER... 3 4.2 INSTALLATION
INF-USB2 and SI-USB Quick Start Guide
INF-USB2 and SI-USB Quick Start Guide Please follow these instructions carefully. DO NOT connect the INF-USB2 or SI-USB module to your computer before running the setup program. After running Setup and
SW43W. Users Manual. FlukeView Power Quality Analyzer Software Version 3.20 onwards
SW43W FlukeView Power Quality Analyzer Software Version 3.20 onwards Users Manual 4822 872 30778 December 2005, Rev.1, 04/06 2005 Fluke Corporation, All rights reserved. All product names are trademarks
HP Point of Sale (POS) Peripherals Configuration Guide 2D Imaging / Linear / Presentation Scanner
HP Point of Sale (POS) Peripherals Configuration Guide 2D Imaging / Linear / Presentation Scanner Document Version 3.42 December 2011 1 Copyright 2007-2011 Hewlett-Packard Development Company, L.P. The
How To Use A Water Quality Monitor (Wqmsw)
Water Quality Monitor WQM Host Software User s Guide The user s guide is an evolving document. If you find sections that are unclear, or missing information, please let us know. Please check our website
Analyser Data Manager Operating Manual
Analyser Data Manager Operating Manual Geotechnical Instruments (UK) Ltd Sovereign House Queensway Leamington Spa Warwickshire CV31 3JR England Tel: +44 (0)1926 338111 Fax: +44 (0)1926 338110 Email: [email protected]
Productivity Software Features
O P E R A T I O N S A N D P R O C E D U R E S F O R T H E P R O D U C T I V I T Y S O F T W A R E Productivity Software Features Remote CS-230 calibration and set-up on a personal computer. CS-230 calibration
Accounting Manager. User Guide A31003-P1030-U114-2-7619
Accounting Manager User Guide A31003-P1030-U114-2-7619 Our Quality and Environmental Management Systems are implemented according to the requirements of the ISO9001 and ISO14001 standards and are certified
LOREX CLIENT Remote Software 4.0
LOREX CLIENT Remote Software 4.0 Instruction Manual English Version 2.0 MODEL: L20WD800 Series www.lorexcctv.com Copyright 2008 LOREX Technology Inc. Table of Contents Table of Contents Software Installation...
ViewPower User Manual
ViewPower User Manual Management Software for Uninterruptible Power Supply Systems Table of Contents 1. ViewPower Overview... 1 1.1. Introduction... 1 1.2. Structure... 1 1.3. Applications... 1 1.4. Features...
Monitoring Network DMN
Monitoring Network DMN User Manual Table of contents Table of contents... 2 1. Product features and capabilities... 3 2. System requirements... 5 3. Getting started with the software... 5 3-1 Installation...
Remote Access. Table Of Contents. Rev: 3.30, 8/03
Remote Access Rev: 3.30, 8/03 Table Of Contents Welcome to Remote Access... 1 Using Remote Access... 1 Using Remote Access... 3 Selecting an OPLC model... 3 Communication-PC Settings... 3 Remote Access
DUKANE Intelligent Assembly Solutions
PC Configuration Requirements: Configuration Requirements for ipc Operation The hardware and operating system of the PC must comply with a list of minimum requirements for proper operation with the ipc
Data Retrieval. Chapter III. In This Chapter
Chapter III. Data Retrieval In This Chapter This chapter gives you instructions for setting up and using TradingExpert Pro s automatic data retrieval system which is used for downloading end-of-day data
Allworx OfficeSafe Operations Guide Release 6.0
Allworx OfficeSafe Operations Guide Release 6.0 No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form or by any means, electronic, mechanical, photocopy,
NVMS - 1000 USER MANUAL. Version 2.1.0
NVMS - 1000 USER MANUAL Version 2.1.0 Contents 1 Software Introduction... 1 1.1 Summary... 1 1.2 Operation Environment... 1 1.3 Install and Uninstall... 2 1.3.1 Install the Software... 2 1.3.2 Uninstall
SPC-1. Single Point Controller. Operation Manual
SPC-1 Single Point Controller Operation Manual NOTICE: PPM Technology Limited shall not be liable for errors that may appear herein or for incidental or consequential damages in connection with or arising
Using AORUS Notebook for the First Time
V2.0 Congratulations on your purchase of the AORUS Notebook! This Manual will help you to get started with setting up your notebook. For more detailed information, please visit our website at http://www.aorus.com.
SNMP Web card. User s Manual. Management Software for Uninterruptible Power Supply Systems
SNMP Web card User s Manual Management Software for Uninterruptible Power Supply Systems Table of Contents 1. Overview... 3 1.1 Introduction... 3 1.2 Features... 3 1.3 Overlook... 3 1.4 Installation and
Temperature Controller Support Software
Temperature Controller Support Software CX-Thermo Ver.4.0 Online Help Contents Chapter 1 Introduction 1.1 What is CX-Thermo? 1.2 Compatible Device Models and Computers 1.3 Hardware Setups 1.4 Basic Use
Installing S500 Power Monitor Software and LabVIEW Run-time Engine
EigenLight S500 Power Monitor Software Manual Software Installation... 1 Installing S500 Power Monitor Software and LabVIEW Run-time Engine... 1 Install Drivers for Windows XP... 4 Install VISA run-time...
1 of 10 1/31/2014 4:08 PM
1 of 10 1/31/2014 4:08 PM copyright 2014 How to backup Microsoft SQL Server with Nordic Backup Pro Before creating a SQL backup set within Nordic Backup Pro it is first necessary to verify that the settings
SOFTWARE GUIDE PM1405 DATA PROCESSING SOFTWARE
SOFTWARE GUIDE PM1405 DATA PROCESSING SOFTWARE Equipment Software Manufacturer Survey meter РМ1405 "РМ1405 Software" Polimaster Copyright Polimaster 2015. All rights reserved. According Copyright protection
How To Connect A300 To A300 Time Clock On A Pc Or Mac Computer
1 TABLE OF CONTENTS INTRODUCTION CHAPTER 1 ESTABLISHING A CONNECTION TO TIME CLOCK 1.1 INSTALLATION OF SOFTWARE 1.2 CONNECT TIME CLOCK TO PC 2 1.2.A..VIA USB CABLE CONNECTION 2-5 1.2.B... VIA TCP/IP NETWORK
GX-2009 Data Logger Management Program Operator s Manual
GX-2009 Data Logger Management Program Operator s Manual Part Number: 71-0163RK Revision: C Released: 3/19/12 www.rkiinstruments.com Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by
Modbus RTU (RS-485) TCP/IP (Ethernet)
Eaton Innovative Technology PEM PC Software Installation and Operation Instruction Manual: IM01005006E Rev. 2 March 2007 Modbus RTU (RS-485) TCP/IP (Ethernet) Effective Date 3/2007 For more information
Device configurator DRC200
Operating manual 42/49-27 EN Engineer IT Device configurator DRC200 R&C Process Data Management Software Impressum Device configurator DRC200 Operating manual Document No. 42/49-27 EN Date of issue: 11.02
<User s Guide> Plus Viewer. monitoring. Web
Plus Viewer 1 < Plus Viewer (web ) > 1-1 Access Method The user can access the DVR system through the web. 1 Enter the IP for the DVR system in the address field of the web browser. 2 The
LP 2844 Thermal Printer
Quick Installation Guide FedEx Ship Manager Software FedEx Ship Manager at fedex.com LP 2844 Thermal Printer Step 1: Confirm Printer Kit Contents Your printer kit contains: 1. Thermal Printer 4. Serial
Power Log PC Application Software
Power Log PC Application Software Users Manual May 2007 Rev. 1, 5/08 2007-2008 Fluke Corporation. All rights reserved. Specifications are subject to change without notice. All product names are trademarks
Multi-Site 4200. Remote Viewing Software. User Guide
Multi-Site 4200 Remote Viewing Software User Guide Document 800-12064V1 Rev A 09/2012 User Guide Revisions Issue Date Revisions A 07/2012 New document. V1 Rev A 09/2012 Updated the description of the
3.5 EXTERNAL NETWORK HDD. User s Manual
3.5 EXTERNAL NETWORK HDD User s Manual Table of Content Before You Use Key Features H/W Installation Illustration of Product LED Definition NETWORK HDD Assembly Setup the Network HDD Home Disk Utility
Universal Simple Control, USC-1
Universal Simple Control, USC-1 Data and Event Logging with the USB Flash Drive DATA-PAK The USC-1 universal simple voltage regulator control uses a flash drive to store data. Then a propriety Data and
User s Manual C R Y S T A L. engineering corporation
DaTa LoGGerXPTM User s Manual C R Y S T A L engineering corporation DataLoggerXP User s Manual Page Contents Introduction........................................................ 1 How it works........................................................
Page 1 S5.5 Monitoring Software Manual Rev 1.2 for Software Ver 1.0
Trantec S5.5 Monitoring Software Manual Initial Setup Minimum PC Requirements 1.6 GHz CPU, 256 MB Memory, 1 available USB port, Windows XP Recommended PC >2.0 GHz CPU, 512 MB (XP) 1 GB (Vista) memory,
Chapter 1 Downloading Images to Your Computer (Windows)
Software User Guide For the safe use of your camera, be sure to read the Safety Precautions thoroughly before use. Types of software installed on your computer varies depending on the method of installation
NortechCommander Software Operating Manual MAN-00004 R6
NortechCommander Software Operating Manual MAN-00004 R6 If the equipment described herein bears the symbol, the said equipment complies with the applicable European Union Directive and Standards mentioned
Rapid Assessment Key User Manual
Rapid Assessment Key User Manual Table of Contents Getting Started with the Rapid Assessment Key... 1 Welcome to the Print Audit Rapid Assessment Key...1 System Requirements...1 Network Requirements...1
eshuttle Dictate Manual Software Version 1.0
eshuttle Dictate Manual Software Version 1.0 eshuttle software is designed for securely emailing dictation and transcription files between a dictator and transcriptionist. eshuttle Dictate by Bytescribe,
TFA. KlimaLogg Pro. User Manual. Revision: 0.1
TFA KlimaLogg Pro User Manual Revision: 0.1 Table of Contents Overview... 2 Getting Started... 2 Setting up the KlimaLogg Pro device... 2 Installing the USB-Stick... 2 Installation... 2 System Requirements...
VisionMate Flat Bed Scanner 2D Tube Barcode Reader
VisionMate Flat Bed Scanner 2D Tube Barcode Reader User s Manual Page 1 Catalog #3111 MAN-21256 Rev G Contact Information North America: Tel: 800.345.0206 email: [email protected] Europe: Tel:
Release Notes. R&S ZVH4/ZVH8 Cable and Antenna Analyzer
Release Notes Revision: 01 R&S ZVH4/ZVH8 Cable and Antenna Analyzer Firmware Release V1.50 These Release Notes describe the following models and options of the R&S Cable and Antenna Analyzer: R&S Cable
Manual. EN Appendix. Blue Solar PWM Pro setup and monitoring software
Manual EN Appendix Blue Solar PWM Pro setup and monitoring software EN Appendix Contents 1. Region and language settings 2 2. Download the software 2 3. UnRAR the software 2 4. Install the software 3
Gigabyte Content Management System Console User s Guide. Version: 0.1
Gigabyte Content Management System Console User s Guide Version: 0.1 Table of Contents Using Your Gigabyte Content Management System Console... 2 Gigabyte Content Management System Key Features and Functions...
Novell ZENworks Asset Management 7.5
Novell ZENworks Asset Management 7.5 w w w. n o v e l l. c o m October 2006 USING THE WEB CONSOLE Table Of Contents Getting Started with ZENworks Asset Management Web Console... 1 How to Get Started...
Embedded Web Server. User Guide. General usage guide lines to the ISONAS Embedded Web Server available on all ISONAS PowerNet reader-controllers
Embedded Web Server User Guide General usage guide lines to the ISONAS Embedded Web Server available on all ISONAS PowerNet reader-controllers 06/12/09 Contents Revision history... 4 Scope... 4 Introduction...
User s Manual for Fingerprint Door Control Software
User s Manual for Fingerprint Door Control Software Foreword The naissance of F7 indicated that fingerprint reader enters into professional door control domain. That s why we developed this software to
Smart Card Authentication. Administrator's Guide
Smart Card Authentication Administrator's Guide October 2012 www.lexmark.com Contents 2 Contents Overview...4 Configuring the applications...5 Configuring printer settings for use with the applications...5
Centurion C4 Transfer Guide using C4 File Transfer Utility
Centurion C4 Transfer Guide using C4 File Transfer Utility 1010536 02 23 11 Section 50 1.0 Background: 1.1 A valid configuration is required for the Centurion system. Configuration files are downloaded
Custom Reporting System User Guide
Citibank Custom Reporting System User Guide April 2012 Version 8.1.1 Transaction Services Citibank Custom Reporting System User Guide Table of Contents Table of Contents User Guide Overview...2 Subscribe
X Series Application Note 43:
X Series Application Note 43: Using the Remote Viewing & Web Pages of the X - Series & GR Series Recorders The Remote Viewing function of the X-Series and GR Series Recorders provide the user with the
Diamante WiFi Wireless Communication User Guide. CradlePoint CTR35
Diamante WiFi Wireless Communication User Guide CradlePoint CTR35 Release: July 2011; March 2011 Patent Pending. Copyright 2011, Stenograph, L.L.C. All Rights Reserved. Printed in U.S.A. Stenograph, L.L.C.
JVC Monitor Calibration 2 User's Guide
JVC Monitor Calibration 2 User's Guide Table of Contents 1. What is JVC Monitor Calibration 2? 2. System Requirement 3. Installation (JVC Monitor Calibration 2) 4. Uninstall 5. Installation (EyeOne Display
Using GIGABYTE Notebook for the First Time
Congratulations on your purchase of the GIGABYTE Notebook. This manual will help you to get started with setting up your notebook. The final product configuration depends on the model at the point of your
Event recording (or logging) with a Fluke 287/289 Digital Multimeter
Event recording (or logging) with a Fluke 287/289 Digital Multimeter Our thanks to Fluke Corp for allowing us to reprint the following article. One of the major features of the Fluke 280 Series digital
OnDemand Version 1.7. Installation Manual and User s Guide. AST Technology. Sophienstrasse 4 32051 Herford Germany
OnDemand Version 1.7 Installation Manual and User s Guide AST Technology Sophienstrasse 4 32051 Herford Germany Contents Overview of OnDemand Software... 1 System and Software Requirements... 1 CVe Monitor
Strategic Asset Tracking System User Guide
Strategic Asset Tracking System User Guide Contents 1 Overview 2 Web Application 2.1 Logging In 2.2 Navigation 2.3 Assets 2.3.1 Favorites 2.3.3 Purchasing 2.3.4 User Fields 2.3.5 History 2.3.6 Import Data