Model 1818 Oxygen Scavenger Monitor
|
|
|
- Ruby Pope
- 10 years ago
- Views:
Transcription
1 Model 1818 Oxygen Scavenger Monitor INSTRUCTION MANUAL IMPORTANT NOTICE: Model 1818 Oxygen Scavenger Monitor is shipped WITHOUT SOFTWARE AND WILL NOT OPERATE UNTIL SOFTWARE IS INSTALLED. For use with Hydrazine, use software found in Cat. No. HYDRAK; for use with ELIMIN-OX, use software found in Cat. No. ELIMAK. ELIMIN-OX is a registered trademark of NALCO Chemical Co. A Thermo Electron business
2
3 TABLE OF CONTENTS I. General Information I. 1 Introduction I. 1 Principles of Operation I. 1 Principles of Calibration I. 2 Sample Requirements I. 3 Description of Model 1818AO Monitor I. 4 II. Instrument Preparation II. 1 Unpacking Instrument II. 1 Mounting, Plumbing and Wiring of Series 1800 Monitor II. 1 Reagent Diffusion Bottle II. 2 Installation of Iodide Electrode II. 3 Installation of Reference Electrode II. 3 Installation of ATC Probe II. 3 Flow Regulation II. 3 III. Instrument Operation III. 1 Description of Basic Unit Controls III. 1 Start Up/ Normal Instrument Operation III. 2 Initial Instrument Set-Up III. 2 Use of Test Mode III. 2 Error Mode III. 3 Shutdown and Start-Up Procedure III. 4 Flow Off III. 4 V. Calibration V. 1 VI. Instrument Maintenance VI. 1 Maintenance Schedule VI. 1 Weekly Maintenance VI. 1 Bi-Weekly Maintenance VI. 1 Monthly Maintenance VI. 2 Bi-Monthly Maintenance VI. 2 Quarterly Air Pump Check VI. 2 Yearly Preventive Maintenance VI. 3 VII. Troubleshooting VII. 1 VIII. Repair and Service VIII. 1 IX. Notice of Compliance IX. 1 X. Warranty X. 1 XI. Ordering Information XI. 1 XII. Mounting Dimensions XII. 1 XIII. Wiring Diagram XIII. 1 IV. Installation and Operation of Modules IV. 1 Signal Conditioner Module IV. 1 Removing Installed Module IV. 1 Replacing Installed Module IV. 1 Setting Module Output IV. 1 Electrical Connections IV. 2 Description of Signal Conditioner IV. 2 Programming Measuring Range IV. 3 Installation of Optional Alarm Module IV. 3 Setting Alarm Output IV. 4 Electrical Connections IV. 4 Description of Alarm Module Controls IV. 5 Alarm During Calibration IV. 5 Alarm After Power Failure IV. 5 Setting Alarms 1 and 2 IV. 5
4 I. GENERAL INFORMATION Introduction The ORION Model 1818 Oxygen Scavenger Monitor is one of a series of microprocessor-based analyzers designed to measure continuously trace levels of certain oxygen scavengers in relatively pure water. The Model 1818 Monitor continuously measures these oxygen scavengers at points in the steam/water circuit where dissolved oxygen control is critical. Common measuring points are in feedwater, boiler water, or economizer inlet, depending on system design and where the oxygen scavenger is added. In high and low pressure boiler systems, oxygen scavengers are added to react with dissolved oxygen present, many leaving behind a small residual. This residual ensures that any small amount of dissolved oxygen that enters the system through leakage will be removed. Residual oxygen scavengers should be carefully monitored to prevent costly overdosing, and yet allow enough of the reductant in the system to allow for metal passivation. Also, excess levels of certain oxygen scavengers decompose into ammonia, increasing system alkalinity. Efficient control of oxygen scavengers depends on maintaining a predetermined safety zone. A unique system for ph adjustment eliminates the need for expensive reagents and frequent maintenance. Instrument calibration is quick, simple, highly reliable and accurate. A temperature sensor, mounted in the sample stream, is used for automatic temperature compensation.the installed signal conditioner offers the choice between linear and logarithmic concentration output. Since the 1800 Series Monitors have been designed with few moving parts, they require little maintenance and are extremely reliable. Principles Of Operation Numbers refer to Figure 2. As show in Figure 2, the sample enters the monitor and passes through an inlet valve, 1, bypass filter, 2, pressure regulator, 3, flow meter, 5, and restrictor tubing, 6. The sample then passes through the fluid connector block into a reagent diffusion bottle, 15, where sample ph is adjusted and iodine added by means of passive diffusion. The ph adjusted sample then flows back through the connector block into the reagent mixing loop, 11, where the oxygen scavenger in the sample mixes with iodine reagent to form iodide. The sample then passes the iodide-sensing, 9, and reference, 8, electrodes in the top portion of the flow cell, 12. The sample then flows past the temperature, 10, probe into an atmospheric drain, 19. The oxygen scavenger present in the sample reacts with the iodine to form iodide ion: OS red + X I 2 = 2X I - + OS ox where: OS red = oxygen scavenger reduced OS ox = oxygen scavenger oxidized The iodide electrode responds logarithmically to changes in the resulting iodide concentration. This response is described by the Nernst equation: E = Eo +2.3 (RT/nF) log (C/Ciso) where: E = measured electrode potential, mv E o = potential, when C equals C(iso), mv R = ideal gas constant T = temperature of sample, degrees K n = valence of ionic species (-1 for iodide ion) F = Faraday s constant C = effective iodide ion concentration (activity) Ciso = concentration (activity) of iodide ion where potential E is temperature independent (isopotential point) The above equation indicates that the measured potential varies with both temperature and the concentration of the ion of interest. In order to eliminate error caused by fluctuations in sample temperature, the Series 1800 Monitor s microprocessor constantly updates temperature corrections from data supplied by the ATC probe. From the Nernst equation, the theoretical response of an iodide ion-selective electrode to a ten-fold change in concentration of 25 C is mv. This is referred to as the electrode slope(s). Most electrodes, however, do not exhibit a theoretical slope. Therefore, the instrument is calibrated to determine its actual value. Two standards are used to provide information necessary for the microprocessor to compute the actual slope and E o for use during sample analysis. Acid is added to the sample to prevent a competing reaction from occurring: 3I 2 + H 2 O = 5I - + IO H + Maintaining acidic conditions prevents the release of iodide which would cause high apparent reductant readings. I. 1
5 The ph adjustment and iodine addition are accomplished by passive diffusion. The sample passes through tubing in a reagent bottle containing both acid and iodine. Both substances diffuse through the tubing wall, into the sample, and redissolve. This passive method eliminates sample contamination problems due to reagent contamination, and makes the fluid handling apparatus simple and reliable. Principles Of Calibration Model 1818 allows the user the ability to perform a two-point calibration. In addition, with some oxygen scavengers, an off-line or one-point calibration can be performed. Verify the appropriate calibration method(s) for the reductant being measured by referring to the separate APPENDIX found in the appropriate Application Kit. Off-Line Calibration Since the oxygen scavenger concentration in feedwater is normally controlled over a rather narrow range, usually a single-point calibration procedure is sufficient to provide accurate results. The off-line calibration procedure has the advantage of being quick and easy to perform. If sample reductant concentrations are expected to vary widely from day to day or if on-line verification of monitor readings is desired, follow the Two-Point Calibration procedure. The off-line calibration feature of Model 1818 allows the user to adjust the monitor to values determined by alternate methods used in their laboratory, such as a standard colorimetric method. It is essentially a onepoint calibration. To perform off-line calibration, a sample is taken from the bypass of the instrument; the sample value is stored in memory; the sample is analyzed by an alternate method of choice; the previously stored reading is adjusted to the lab method result; and the instrument is then returned to the analysis mode. The term off-line calibration refers only to the fact that a sample from 1818 bypass is taken off-line for laboratory analysis; in fact, no downtime is experienced during the procedure and the instrument remains on-line throughout. pressure reducing valve valve flowmeter flowmeter flow valve flow valve bypass bypass filter filter inlet filter inlet valve restrictor calibration diverter valve bypass valve drain Figure 1 Block Diagram of Sample Flow Figure 2 Flow During Normal Operation 1 reagent addition system temperature probe probe reference electrode electrode sensing electrode electrode calibration level level measuring level measuring level Legend 1. Inlet Valve 9. Reference Electrode 2. Bypass Filter 10. Temperature Probe 3. Pressure Regulator 11. Reagent Mixing Loop 4. Flow Valve 12. Flow Cell 5. Flowmeter 15. Reagent Diffusion 6. Flow Restrictor Tubing Bottle 7. Electrolyte Reservoir 19. Drain 8. Reference Electrode 21. Bypass Valve I. 2
6 Two-Point Calibration This procedure provides maximum calibration accuracy and requires use of dynamic calibrator, Cat. No. 15DC15/115 V or 15DC20/220 V. In addition to calculation of electrode E o as is done in off-line calibration procedure, this procedure determines electrode slope. To perform a dynamic two-point calibration, the 1818 is first connected to a reductant-free sample stream. An appropriate diluted standard is then prepared. A supplied syringe is filled with diluted standard and mounted on the calibrator. The calibrator is then mounted near the monitor and the syringe tubing connected to the standard injection port. By adjusting calibrator pump settings, two different flow rates produce two known standards diluted into the sample background. By pressing the appropriate keys on the monitor when prompted, the monitor s microprocessor completes the calibration. If a mistake is made or problems occur during calibration, an error light will come on and alert the operator. By pressing the error key, the operator can access information on the display to help determine the problem. Sample Requirements Sample Inlet Connection - 1/4 NPTF. If particulate matter is present in sample, prefiltration is necessary. Moderate amounts of particulates will be removed by the 60 micron stainless steel filter located after inlet valve. Flow Rate - must be 50 ml/min. minimum at monitor sample inlet. Pressure psig. Consult ORION for details on sample handling if pressure is outside range. Temperature - Temperature must be between C. Oxygen Scavenger Level - Levels are read directly in ppb or ppm. Sample Alkalinity - Sample alkalinity should not be more than 5 ppm CaCO 3 equivalent to ensure two-month reagent life. If sample alkalinity is above 5 up to 10 ppm maximum as CaCO 3, reagent life is reduced to six weeks. For higher sample alkalinity, contact the ORION Technical Service. I. 3
7 Description of Model 1818AO NOTE: Numbers in the description refer to Figure 3. Sample Inlet Valve (1) - Accepts the sample stream via 1/4 inch NPTF connector. The customer must supply the sample with a pressure between 8 and 100 psig, and a sample flow rate to the analyzer of 50 ml/min. minimum. Bypass Filter Assembly (2) - 60 micron stainless steel filter traps particulate matter in sample stream. Pressure Regulator (3) - Adjusts flow on incoming sample stream. Flow Valve (4) - Used to turn off flow to flow cell. Flow Meter (5) - Measures sample flow rate. 40 ml/min. nominal flow is required through the analyzer. Flow Restrictor Tubing (6) - Maintains steady sample flow rate in conjunction with pressure regulator. Pressurized Electrolyte Reservoir (7) - Provides a constant flow of electrolyte solution through reference electrode for maximum stability. Reference Electrode (8) - Provides a constant reference potential and completes the measurement circuit. Must be placed in right hand side of flow cell. Iodide Electrode (9) - Senses iodide ions in sample stream and produces an electrical potential dependent on sample oxygen scavenger. Must be placed in left hand side of flow cell. ATC Probe (10) - Measures sample temperature and inputs data to microprocessor for automatic temperature compensation (ATC). Reagent Mixing Loop (11) - Mixes iodine reagent with oxygen scavenger in sample to form iodide, measured by iodide-sensing electrode. Flow Cell (12) - Contains reference electrode, sensing electrode, ATC probe and sample measurement reservoir. Thumbscrew (13) - Supports the reagent diffusion bottle. Fluid Connector Block (14) - Connects reagent diffusion bottle to sample stream and flow cell assembly. Reagent Diffusion Bottle (15) - Acidifies and adds iodine to the sample by means of passive diffusion. LED Display (16) - Provides digital readouts of concentration, temperature, millivolts and error codes. Keypad (17) - Consists of five mode keys, four prompt indicator lights, two scroll keys and one key for entering data. Mode and error indicators are also incorporated on keypad. On/Off Switch (18) - Controls all power to the electronics. Drain (19) - Atmospheric drain prevents back pressure on reference electrode. Standard Injection Port (20) - Allows connection of dynamic calibrator tubing to fluid connector block during two-point calibration. Bypass Valve (21) - Used to throttle flow in bypass system. Check Valve (22) - Prevents backflow of electrolyte from pressurized reservoir. Air Pump (23) - Provides air to pressurize the electrolyte reservoir. Bottle Clamp Band (24) - Holds reagent diffusion bottle securely in place. I. 4
8 Pure Water Monitors 1800 Series Release latches to access 18 Figure 3 Major Monitor Components I. 5
9 I. 6
10 II. INSTRUMENT PREPARATION Unpacking Instrument Refer to Figure 4. Report any obvious damage of shipping container to carrier and hold for inspection. The carrier (not Orion) is responsible for any damage incurred during shipment. 1. Open outer box. Remove foam corner support pieces. 2. Open inner box and remove cardboard retaining shell. Remove accessory boxes and instruction manual. 3. Unbolt the monitor from mounting board. Save all hardware for use during installation. 4. Carefully place the monitor at a convenient location. Do not pull or lift instrument by its fluidic components. Figure 4 Unpacking Monitor Mounting, Plumbing And Wiring of Series 1800 Monitor 1. Select a site for instrument that allows it to be permanently bolted in an upright position with ample height for atmospheric drain operation and ready access to both electronic controls and flow cell. The analyzer site must permit connection to sample line, a drain, an AC power supply, and any connections for output devices. 2. Prepare mounting holes. Carefully lift the Series 1800 Monitor and bolt into place. Do not lift instrument by pulling on any fluidic components. Be sure to use the four shoulder washers that are provided, they will resist future corrosion. 3. Connect a waste line to a drain of sufficient capacity. 4. Connect sample line to the 1/4-inch tube connector. It is recommended that a shutoff valve be installed at the sampling point. 5. Remove decorative panel inside the cabinet by unscrewing the four Phillips screws. Then remove protective plug and feed AC power line through hole. See Figure See APPENDIX for wiring diagram. Disconnect: jumper wire from TB2 terminal 7 and connect to: TB2 terminal 6 for 100 V AC 5 for 115 V AC 4 for 220 V AC 3 for 240 V AC Warning: Failure to connect jumping wire to the proper tab will permanently damage the monitor. 7. Connect: COMMON to TB2 terminal 1 (most left) LINE to TB2 terminal 2 GROUND to ground stud adjacent to terminal 1 using captive type terminal at the end of ground wire. Warning: Failure to connect chassis to a suitable ground may result in hazardous conditions. Failure to use captive type terminal could result in potential shock hazard and may jeopardize CSA approval. II. 1
11 8. All electrical wire feeding through the chassis must pass through appropriate electrical fittings in order to maintain the design integrity of the splash-proof electronic chassis. Since different types of fittings are required at various installations, this feed through is to be supplied by the user. common line Warning: A grounded metal conduit is required for FCC compliance. 9. Replace decorative panel if no further wiring is required. Warning: Failure to replace decorative panel could result in potential shock hazard. ground 240V 220V 110V 100V jumper Reagent Diffusion Bottle Numbers refer to Figure 3, unless otherwise noted. NOTE: This procedure requires use of ORION Reagent (Cat. No ). CAUTION: Care must be taken to avoid contact of skin or clothes with this reagent which contains concentrated acetic acid and iodine. Avoid breathing vapor. Wear rubber gloves and goggles when performing this procedure. In case of skin contact, flush skin immediately with water to prevent burns. 1. Support the bottom of the reagent bottle, 15, release holding clamp, 24, with one hand. Turn thumbscrew, 13, counterclockwise to release diffusion bottle. Unscrew white bottle cap and gray tubing connector assembly. (Keep bottle supplied with instrument as a spare.) 2. Take a new bottle of reagent to a wellventilated area such as a laboratory fume hood. Unscrew and remove white cap. 3. Connect diffusion tubing (use Cat. No ONLY) to tubing connector assembly to conform to Figure 6. Place the tubing assembly and cap into the bottle. 4. Replace gray cap and screw white cap on tightly. 5. Insert the four nipples on the top of the reagent diffusion bottle into the fluid connector block, 14, as shown in Figure Support the bottom of the reagent bottle with one hand and tighten the thumbscrew clockwise. 7. Reclamp bottle to panel to secure. AC power Figure 5 Electrical Wiring of 1800 Series Monitors barbed fittings reagent diffusion bottle tubing fluid connector block white cap Figure 6 Reagent Diffusion Bottle Assembly options nipples with small O-rings gray cap with large O-ring tubing should be connected to right-hand barbed fittings only(as seen from the front) II. 2
12 Installation of Iodide Electrode 1. Unpack iodide electrode (Cat. No ) and carefully remove protective cap. 2. Polish electrode according to instruction card with strips (Cat. No ). 3. Insert iodide electrode into sensing hole on left in flow cell. 4. Connect electrode to cable marked Sens. Elect Installation of Reference Electrode 1. Unpack reference electrode, 8, (Cat. No ) and tubing from shipping box. Remove protective caps from bottom and sidearm of electrode. Save caps for future storage of electrode. Inspect internal of electrode (item 7 in Figure 7). Verify it is at least half filled with the internal fill solution. If not, replace the electrode. 2. Shake out as much filling solution as possible through sidearm, 30, into sink or drain. 3. While passing 1/8-inch tubing, 31, into electrode through sidearm, slide 1/4-inch tubing, 32, over sidearm. The outside tubing should extend 3/8-1/2-inch over sidearm. 4. Remove cap and fluid seal from electrolyte fill solution bottle, 33. Hold bottle in upright position. Check that rubber gasket, 34, is properly aligned, then connect cap end of tubing assembly to bottle. The 1/8-inch tubing should extend to the bottom of the bottle. 5. Dry off ceramic frit on base of electrode with tissue paper. Squeeze bottle for a few seconds. A small amount of filling solution should bead up on frit surface, indicating good electrolyte flow. If no moisture is visible, electrode is clogged and should be replaced. 6. Invert electrolyte bottle and snap into clip, 35. Locate discharge tube of air pump on fluidic panel. End of tubing has hollow push pin for pressurizing reference reservoir. Puncture base of bottle with pin and push pin down until its PVC base abuts bottle. Mark and date level of fill solution in reservoir. Then electrolyte solution will begin to flow into electrode. 7. Connect electrode to electronic chassis by keeper cable labeled Ref. Elect. Insert electrode into reference hole on right in flow cell. See Figure Figure 7 Reference Electrode Assembly Installation Of ATC Probe The automatic temperature compensation (ATC) probe is already connected to the electronics housing. Insert into hole of 3/8-inch tee fitting draining flowcell, located behind electrodes. Flow Regulation When the Model 1818AO Monitor is first commissioned, it is advisable to flush out fluidics system at least one hour prior to initial calibration and use. The electronics need not be turned on at this time. Refer to Figure Open sample inlet valve, Pull out red locking ring of pressure regulator, 3, then adjust black knob so that ball of flow meter reads 40ml/min.. Push in red locking ring to secure the setting. 3. Open bypass valve, 21. Check for appropriate sample flow. Readjust pressure regulator if required. 4. Wait at least one hour before calibrating analyzer. II. 3
13 II. 4
14 III. INSTRUMENT OPERATION Description of Basic Unit Controls On/Off Switch - Controls powers to all electronics and air pump. It is located on electronic chassis bottom and includes an integral circuit breaker. LCD Display - Displays four-and-one-half digit plus polarity sign used to read concentration, slope, E o, millivolts, temperature, error messages, and diagnostic information. Mode Indicating LED - (Located to left of LCD display.) Indicates range (ppb or ppm), error codes, and electrode stable in sample mode. Displays temperature ( C), millivolts, electrode slope, and E o on LCD in test mode. Prompt Indicating LED - Prompts the user during calibration. Four LED s located in keypad region are used. These are: add STD 1, add STD 2, begin calibration, two-point calibration complete. /\ and \/ Keys - increases or decreases displayed values that can be operator-changed such as alarm level, analog output range, and off-line calibration values. Enter/done Key - Enters value displayed on LCD into permanent memory for later use. Key also indicates to microcomputer that a required calibration step has been completed. Sample Key - Puts Series 1800 Monitor into sample mode and (re)activates any optional modules. This is also its default mode, e.g., instrument automatically enters sample mode when first turned on and after calibration. Cal Key - Starts Series 1800 Monitor into calibrate mode. Operator is then prompted through the steps necessary for calibration. Test Key - Places Series 1800 Monitor into test mode where LCD displays temperature, millivolts, Eo and slope. Each successive push of key, steps instrument through this sequence, and an LED on the left of display indicates value displayed. Error - If error LED is lit (soft error), then pressing key causes LCD to display an error code. Note, in the case of hard error, the LCD would cease to display normal output but would display error code only. Refer to Error Code Table. Program Key - Used to program expected calibration concentration increments. Also used in conjunction with Cal key for off-line calibration. a, b, c, and d Keys - For future expansion. test sample C ppm mv ppb slope E o error electrode stable Figure 8 Front Panel Keypad a b c d amplifier enter/ done done sample test program cal. cal. error error begin fill/ flow cal. off add add std. 1 std. two drain/ pt. cal. complete flow on add add std. 2 std. ORION 2A01 III. 1
15 Start Up/Normal Instrument Operation 1. Install the Model 1818AO Monitor according to instructions in INSTRUMENT PREPARATION. Power supply should be wired for proper voltage and instrument suitably grounded. 2. Turn on flow at sample inlet and flush for at least one hour. 3. Power up Model 1818AO by toggle switch on bottom of electronics case. See Figure LCD will display following information sequence: a ; b.1818 while LEDs sequentially light; NOTE: Mode indicating LEDs and test LED on modules are not tested during start-up sequence. c. Revision number of software program (e.g., -2.3-). 5. After completion of steps 1-4, the Monitor is now in normal operational mode sample. Initial Instrument Set-Up Before first sample measurements on Model 1818AO can be performed, calibration concentration increments must be programmed into memory. Program Key Table is found on inside instrument door and following. Refer to the appropriate APPENDIX for the particular oxygen scavenger being measured to find correct P2 and P3 default values. Table 1 Program Key Table Code Meaning P0 Off-line calibration value P1 (Not used on Model 1818) P2 Concentration increase for STD 1 addition P3 Concentration increase for STD 2 addition 1. Press program key three times until the LCD displays P2. Then the current value for the first concentration increment will be displayed. 2. Change value, if desired, by using keys marked /\ and \/. Press enter/done key to store appropriate value in memory. Note P2 cannot exceed the P3 value. 3. Press program key until the LCD displays P3 for two seconds. Then the current value for the second concentration increment will be displayed. 4. Change value, if desired, by using /\ and \/. Press enter/done key to store appropriate value in memory and unit will automatically return to sample mode. 5. Battery backup of memory will maintain programmed variables for at least one month if battery is fully charged. Use of Test Mode When test key is pressed, the LCD displays temperature, millivolts, slope and E o. Every time the test key is pressed, the monitor steps through this sequence. An LED to the left of display indicates value displayed. If test mode is entered from sample mode, then sample and test LED are lit. If test mode is entered from calibration mode, then calibrate and test LED are lit. Note that while LCD displays test information, the instrument still continues normal sampling or calibration sequence. Therefore, when sample and test modes are pressed, signal conditioning and optional alarm modules are active. When calibrate and test modes are pressed, calibration prompt LEDs will light as usual. To exit test mode, press sample or calibrate key as desired. LCD will now display instantaneous computed concentration. If sample key is pressed while in the calibration mode, the current calibration will abort with no data saved. III. 2
16 Error Mode The Model 1818AO Monitor diagnoses two types of errors which are termed hard and soft. In the event of hard error, the analyzer cannot compute any meaningful concentration values, error LED will be lit and LCD will display an error code. The signal conditioner module will indicate zero scale and alarms will be deactivated. Hard error must be corrected prior to resumption of normal operation (E40 and E50) by performing a complete calibration. If error LED is active but instrument continues to display concentration, this is termed soft error. Soft error can affect accuracy and/or precision of displayed concentration but not interfere with instrument s ability to compute sample concentration. To determine cause of soft error, press error key and LCD will display error code. Table 2 Error Codes Error Code Meaning E00 No errors E01 E02 E03 E04 E05 E10 E20 Default values are used. The monitor has not been calibrated since start-up or the reset button was pushed. Sample is outside the temperature range of C Combination of E01 and E02 Faulty ATC probe, or related circuitry Combination of E01 and E04 After calibration, the new slope is outside the expected range. Calibration due. Last calibration performed 30 days ago. A reminder to perform another calibration. E30 Calibration due and slope error. This indicates that the last calibration performed was at least 30 days ago and the electrode slope was out of specification at that time. E40 Calibration overdue. It has been at least 6 weeks since the last calibration. The instrument beyond this point could be out of specification. This is a hard error. The only way to exit this error is to perform a complete two-point or off-line calibration. E50 Calibration overdue and slope error. This indicates that it has been at least 6 weeks since the last calibration and the electrode slope was out of specification at that time. The instrument beyond this point could be out of specification. This is a hard error. The only way to exit this error is to perform a complete two-point calibration (or off-line calibration depending on reductant used.) The most common error codes are combinations of the above codes, but other errors may be displayed: E12 E10 and E02 E14 E10 and E04 E21 E20 and E01 E22 E20 and E02 III. 3
17 Shutdown And Start-Up Procedure The following steps should be taken if a loss of sample flow is expected for more than two days. These procedures will prevent possible build-up of acidic reagent vapors in the monitor. NOTE: At cycling power plants or on units where sample flow to the monitor is frequently interrupted, the installation of a backup sample is recommended. Connect a suitable high purity water source (such as condensate storage tank) into the sample line just upstream of the monitor s sample inlet. By putting shutoff valves in both sample and backup lines at this point, backup or actual sample flow can be quickly valved in or out as required. In this way, continuous flow to the monitor can be supplied and the need for the following startup/shutdown, equilibration and recalibration procedures is eliminated. (Note, however, that for sample flow loss of only a few hours, monitor recalibration should not be required.) Shutdown 1. Shut off sample flow prior to the monitor inlet. 2. Turn power OFF. WARNING: Wear rubber gloves and safety goggles to avoid possible injury from reagent residuals in the system. 3. Loosen clamp securing reagent bottle to fluidics panel. Support bottom of bottle with one hand, and turn thumbscrew on top of fluid connector block to release reagent bottle. Store in a well-ventilated area such as a laboratory fume hood. If bottle is to be stored more than two days, remove top cap and tubing assembly and store in a large beaker of tap water. Otherwise, tubing will become brittle and may break when analyzer is restarted. Start-Up 1. If reagent bottle has been stored longer than two days without removal of tubing, tubing must be replaced. If acid reagent is more than four to six weeks old depending on sample alkalinity, or age is unknown, it should be replaced. Use a fume hood, rubber gloves and goggles for these procedures. Unscrew bottle cap and carefully lift out gray cap with tubing connected. Remove old diffusion tubing from nipples. Replace with new tubing, replace reagent if required, and reinstall bottle on analyzer. 2. Polish iodide electrode per Monthly Maintenance instructions. Looking at monitor, reinstall carefully in the left section of the electrode holder. 3. Restore sample flow to analyzer. If necessary, adjust pressure and flow rate through analyzer to normal ranges. 4. Remove protective end cap from reference electrode. Looking at the monitor, reinstall electrode in the right section of the electrode holder, being careful not to disconnect reservoir tubing from electrode sidearm. 5. When meter reading stabilizes (about one hour), recalibrate according to instructions in appropriate oxygen scavenger APPENDIX. Flow Off If the analyzer is expected not to have flow for time periods of less than twenty-four (24) hours, leave the instrument on and neglect its output. If the time is expected to be greater than twenty-four hours, follow Shutdown procedures. 4. Carefully pull iodide and reference electrodes out of the top of the electrode holder and let them hang by their connectors. Locate protective end cap from reference electrode kit and place on base of reference electrode. This will prevent reference electrode from drying out. III. 4
18 IV. INSTALLATION AND OPERATION OF MODULES Signal Conditioner Module (Cat. No ) The Model 1818AO Monitor is shipped with signal conditioner module (Cat. No ) factory installed. Read section Setting Module Output and if you need to change the factory settings, see Removing Installed Module. When factory settings are correct for your instrument, then proceed to section Electrical Connections - Signal Conditioner Module Removing Installed Module 1. Ensure power to the Series 1800 Monitor is turned off at power source. 2. Unscrew module keypad from the black brackets using the two screws on the keypad. 3. Make certain the plungers are pulled back so that the grommets are in a closed position. 4. Slide out module far enough to access DIP switches. Replacing Installed Module 1. Ensure power to the Series 1800 Monitor is turned off at power source. 2. Open door of electronic assembly. Use No. 2 Phillips screwdriver to remove four screws on decorative panel. 3. Remove module keypad from shipping box. 4. Unscrew module keypad from the black brackets using the two screws on the keypad. 5. Make certain the plungers are pulled back so that the grommets are in a closed position. Figure 9 Dip Switches on Signal Conditioner Board Setting Module Output The signal conditioner has two voltage outputs (0-5 or 0-10 volts) and two current outputs (4-20 or 0-20 ma isolated). Switches are preset by manufacturer at 0-5 volts and 4-20 ma but can be reset by DIP switch (see Figure 9) on signal conditioner printed circuit board. When required, alter preset output values before sliding the module into the monitor as follows (also refer to Table 3). 1. Current output, factory set at 4-20 ma ma is obtained when switching S1-2 into OFF position and S1-3 into ON position. 2. Voltage output, factory set at 0-5V DC. 0-10V DC is obtained when switching S1-1 into ON position. 3. Slide PC Board onto instrument guide rails. Rear of PC Board should touch backplane and plug should fit securely into its socket. Check that module bracket is against both rails. 4. Push plungers in. 5. Make certain flex cable is attached to keypad and to PC Board. If not, carefully slide header into socket. Table 3 Signal Conditioner Output Modes S1-1 S1-2 S1-3 S V OFF X X X 0-10 V ON X X X 4-20 ma X ON OFF X 0-20 ma X OFF ON X X = No effect IV. 1
19 Electrical Connections 1. If module keypad is still attached, then unscrew module keypad from black brackets using the two screws on the keypad. 2. Feed wire through hole in chassis bottom labeled option. See Figure Run wire along wire trough to signal conditioner. 4. Connect wires to voltage (0-5 V or 0-10 V) or current (0-20 ma or 4-20 ma) output. See Figure Double check the flex cable connections. 6. Screw the keypad on the black metal brackets. 7. Double check the flex cable connections. 8. Measure to ensure appropriate number of panels has been removed. 9. Align decorative panel against four stand-offs and reattach with Phillips screws. Decorative panel should not cover any module. Warning: For safety reasons, the Series 1800 Monitor should not be operated without the decorative panel in place. Failure to replace the panel could expose the user to dangerous voltages. Table 4 Outputs Of Signal Conditioner Terminals Output Terminals Voltage ( + ) 1 ( ) 2 Current ( ) 7 ( + ) 8 Description Of Signal Conditioner Set Zero - Sets lower level concentration output. In this mode LED above set zero is active. To set output to either linear or logarithmic, hold key down to toggle between these two output forms. Set Full Scale - Sets upper level concentration output. In this mode LED above set full scale is active. Test - Generates zero or full scale output to set up an external recording device. Test key deactivates signal conditioner module and either set zero or full scale LED will be active. After pressing test key, there will be an output delay of up to three seconds. Press sample key to reactive module. Mode Indicating LED - (located on left of module keypad). Indicates linear of logarithmic signal output. When set zero key is held, the mode toggles between these output forms. log output linear output signal output test Figure 11 Signal Conditioner Keypad set zero set full scale ORION 2B to 5V DC or 0 to 10v DC Output 1K Ω Load Min. NC NC NC NC ma to 20 ma or 4 ma to 20 ma Current Output 1K Ω Load Min. Figure 10 Wiring Diagram of Signal Conditioner Module IV. 2
20 Programming Measuring Range 1. Press set zero key. LED above set zero will activate and LCD will display current programmed value. Present default value is 1 ppb for log range and 1 ppb for linear range. 2. Alter value by using /\ and \/ keys. Lowest possible value is 1 ppb. 3. Press enter/done key when the correct value is displayed. 4. Set upper range by pressing set full scale then follow above procedure. Preset default value is 100 ppb. Note that it is not possible to set zero value higher than full scale, and a range of at least two decades should be used. 5. To send 0% and 100% of full scale signal to an external recording device, press test key on signal conditioner module. The first time causes 0% signal to be sent, the second time causes full scale signal. Remember, there may be up to a three second delay when test key is pressed. 6. To alternate from log to linear mode: Press set zero key. The output mode will toggle between log and linear, indicated by the LED. The log and linear zero and full scale are independent and should be set separately. Press enter/done when in desired mode. Table 5 Default Values - Model 1818 P0 Off-line CAL Value 000 P1 Not Used 000 P2 STD 1 * P3 STD 2 * Set Zero (log) 1 ppb Set Zero (linear) 1 ppb Set Full Scale (log) 100 ppb Set Full Scale (linear) 100 ppb Set Alarm ppb Set Alarm ppm Slope * E o * Installation Optional Alarm Module (Cat. No ) The optional module is mounted in the instrument in the following manner: 1. Ensure power to the Model 1818 Monitor is turned off at power source. 2. Open door of electronic assembly. Use No. 2 Phillips screwdriver to remove four screws on decorative panel. 3. Remove optional module keypad assembly from shipping box. 4. Unscrew module keypad from the black brackets using the two screws on the keypad. 5. Make certain the plungers on the black metal frame are pulled back so that the grommets are in a closed position. * See appropriate oxygen scavenger APPENDIX for correct default values for P2, P3, Slope, and E o. IV. 3
21 Setting Alarm Output For both alarms 1 and 2, NO and NC conditions are available simultaneously. For alarms disabled condition, either NO (factory set) or NC is available. To change factory setting, perform the following steps: 1. Soldered the appropriate jumper, to change between the NO and NC mode for an alarms disabled remote indication (terminals 7 and 8). See Figure Place PC Board onto instrument guide rails. Rear of PC Board should touch backplane and plug should fit securely into its socket. Check that module bracket is against both rails. Press the two plastic plunger down to lock module in place. 3. Make certain flex cable is attached to keypad and to PC Board. If not, carefully slide header into socket. Table 6 Terminals 7 And 8 Alarms Disabled Mode NO NC Jumper E7 to E8 (factory set) E6 to E8 Electrical Connections 1. Feed wire through hole in chassis bottom labeled options (see Figure 5). 2. Run wire along wire trough to alarm module. 3. Connect wires output as required (see Figure 13). 4. Double check the flex cable connections. 5. Screw the keypad on the black metal brackets. 6. Double check the flex cable connections. 7. Measure to ensure appropriate number of panels have been removed. 8. Align decorative panel against four stand-offs and reattach with Phillips screws. Decorative panel should not cover any optional module. WARNING: For safety reasons, the Series 1800 Monitor should not operate without the decorative panel in place. Failure to replace the panel could expose the user to dangerous voltages. Table 6 Outputs Of Alarm Module Terminals Outputs Terminals Alarm 1 Comm 1 NC 2 NO 3 Alarm 2 Comm 4 NC 5 NO 6 Disable Comm 7 NO/NC 8 Figure 12 Jumper Connections on Alarm Module PC Board NC ALARM 1 Alarm contacts 10A 250VAC max NC NO NO NO ALARM 2 Alarm contacts 10A 250VAC max Figure 13 Wiring Diagram of Alarm Module NC E6 E8 E7 Disable output contacts Choose either NC or NO contacts 10A 250VAC m IV. 4
22 Description of Alarm Module Controls Set alarm 1 - Sets alarm 1 parameter. When set alarm 1 is pressed, LED on key is active. See Table 5 for default values or appropriate oxygen scavenger APPENDIX. Set alarm 2 - Sets alarm 2 parameter. When set alarm 2 is pressed, LED on key is active. See Table 5 for default values. Test - Activates and deactivates alarm 1 and alarm 2 and deactivates relays. Test LED will be lit along with corresponding mode key. o reactivate module, press sample key on front panel keypad. Mode indicating LED - (located to left of keypad.) Indicates if either alarm goes off. If alarms are not active, deactivated LED blinks; for example, during calibration. For both alarms, the LED indicating activation is on whenever concentration is above the set point. Since each alarm output has a normally open and normally closed position, any external alarm can be set to go off when concentration is either above or below the set point. NOTE: The corresponding LED for either alarms turns on when the concentration exceeds the set point. Simultaneously, the relay closes and the NO position closes. alarm one alarm two deactivated alarm module test Figure 14 Alarm Module Keypad alarm set alarm one alarm two ORION Alarm During Calibration During calibration the alarm module is automatically deactivated. After the completion of calibration the Series 1800 Monitor automatically enters the sample mode, however, the alarm module remains deactivated. To reactivate, press sample key. The alarm module is deactivated to avoid the high concentration alarm that would result from the standards used to calibrate. Alarm After Power Failure After a power failure, the alarm module remains deactivated until sample key on the front panel keypad (see Figure 8) is pressed. The alarm settings are not affected by power provided for remote indication. Setting Alarms 1 and 2 Alarm limits work on 5 decade ranges only. For example if an alarm is set to 1.0 ppb, the alarm will go off at 1.0 ppb and will continue to remain on until the 10 ppm value is reached. Anything above the 5 decade range will turn the alarm off. 1. Press set alarm 1 key. LED above set alarm 1 will activate and LCD will display present alarm 1 value. 2. Change displayed value by use of /\ and \/ keys. 3. Press enter/done key when correct value is displayed. Alarm 1 value is now set. 4. Press set alarm 2 key. LED above set alarm 2 will activate and LCD will display present alarm 2 value. 5. Change displayed value by use of /\ and \/ keys. 6. Press enter/done key when correct value is displayed. Alarm 2 value is now set. 7. If an external alarm circuit is to be tested, press test key. The first press of the test key deactivates both alarms, second press activates alarm 1, and third press activates alarm 2. Press sample key to exit test mode. IV. 5
23 IV. 6
24 V. CALIBRATION For complete calibration instructions for the specific oxygen scavenger you wish to measure, see the separate APPENDIX in specific Application Kit. V. 1
25 V. 2
26 VI. INSTRUMENT MAINTENANCE Maintenance Schedule The Model 1818AO Monitor is designed for simple maintenance. Follow instructions in this section to ensure proper operation of your instrument. Table 8 outlines our recommended maintenance schedule. See Ordering Information in separate APPENDIX for recommended supplies for the particular oxygen scavenger you wish to measure. IMPORTANT: To ensure proper maintenance and good analyzer performance, a service log book should be kept. Record date and type of maintenance done. Mark and date fluid levels of reagent and reference reservoir and date when replaced. Tag each electrode cable with installation date of electrode. This system has proven to be successful by ORION Service Department, and is especially useful to plants who rotate instrument operators. ORION s Service Department provides a periodic checkout, calibration, and operator training service on-site to certify analyzer performance to published specifications. This service can be tailored to fit individual customer needs. If interested, please contact ORION Service toll-free at Table 8 Recommended Maintenance Schedule Frequency Weekly Bi-Weekly MonthIy Bi-MonthIy Quarterly Yearly Operation Check flow rate Visual inspection. Check reference electrode filling solution and refill if necessary Polish sensing electrode Clean or replace filter element Calibrate Replace reagent, diffusion tubing, and reagent bottle O-rings Air Pump Check Replace reference electrode Replace iodide electrode Replace O-rings Replace restrictor tubing Weekly Maintenance 1. Check that sample flow rate is between 35 and 45 ml/min. Correct flow rate is dependent on oxygen scavenger being measured. Verify correct flow rate by referring to appropriate APPENDIX. To alter flow rate, pull out red locking ring of pressure regulator, then rotate black knob to increase (clockwise) or decrease (counterclockwise) sample flow. Push in red locking ring to set rate. 2. Inspect unit for leakage. Diffusion tubing leakage is indicated by a rise in level of reagent. 3. Check that there are no error indications and displayed concentration is reasonable. 4. Check that reference electrolyte solution is adequate. Bi-Weekly Maintenance Replacement of Reference Electrode Filling Solution 1. Refer to Figure 7. Use of Cat. No Reference Electrode requires use of fill solution. With the addition of a pressurized assembly for the reference electrode, it is estimated that usage of fill solution will be approximately 1.5 bottles per month. If reservoir is less than one quarter full, replace solution as follows: 2. Remove spent electrolyte bottle from clip. Unscrew cap and discard bottle. 3. Remove cap and seal from new electrolyte fill solution. Check that rubber gasket is properly aligned, then connect cap/tubing assembly to bottle. The 1/8 tubing should extend into the bottle. 4. Invert electrolyte bottle and snap into clip. 5. Use special push pin attached to pressurized reference assembly on monitor to pierce hole in bottom of bottle and to re-pressure reservoir by air pump. 6. Dry frit at a base of reference electrode. Check frit to see bead of fill solution form, indicating good flow. VI. 1
27 Monthly Maintenance Polish Iodide Electrode 1. Polish iodide electrode according to instruction card with strips, Cat. No Replace electrode and reconnect cable. Replacement/Cleaning of Sample Inlet Filter 1. Turn off sample flow. 2. Remove filter hex cap using 1-inch open-ended wrench. 3. Separate filter cap with bypass valve assembly from filter body. 4. Remove filter element and soft gasket (if necessary). 5. Replace gasket and press new sintered element into filter body. 6. Replace hex cap and bypass valve assembly and reposition bypass drain line into drain assembly. 7. Retighten hex cap to approximately 75 oz. in torque. 8. Turn on sample flow and check for leaks. 9. Readjust bypass flow. 10. To clean sintered filter element, clean with agitated dilute hydrochloric acid. Calibrate Unit See APPENDIX in appropriate Application Kit for instructions. Bi-Monthly Maintenance Replacement of Reagent, Tubing and O-rings 1. Replace acid reagent and diffusion tubing. Shut off analyzer prior to inlet. Remove reagent diffusion bottle by unscrewing the black thumbscrew at top of the fluid connector block. Support bottle at its base, unsnap clip around middle, and remove bottle from analyzer. CAUTION: Since reagent contains concentrated acetic acid and iodine which is very corrosive to skin and eyes, take to wellventilated area such as a laboratory fume hood. Wear rubber gloves and eye goggles. Unscrew cap and pour old reagent into a suitable waste container. 2. Remove cap, lift out gray insert, and remove old diffusion tubing from nipples. Rinse cap with deionized water. Replace five small O-rings on tubing connector barbs and one large O-ring on reagent bottle. Install new reagent and tubing. Re-install reagent bottle on analyzer, mark level and date. Restore sample flow to monitor. Wait at least one hour to allow analyzer to equilibrate prior to recalibration. Quarterly Air Pump Check Reference electrode fill solution level should drop at least 1/8 per month, indicating good flow through the electrode s ceramic frit. If flow rate is low, air pump should be tested for proper operation. To test air pump, remove push pin and tubing from reference reservoir. Place tip of pin in beaker of water. Vigorous bubbling should be observed. (Normally, air pumps deliver at least 500 cc/min. of air). Replace air pump if weak or no bubbling action is observed. To ensure tight seal on reservoir, replace with a fresh bottle of fill solution, ORION Cat. No VI. 2
28 Yearly Preventive Maintenance Replace Reference Electrode (Cat. No ) See Instrument Preparation for step-by-step instructions. Replace Sensing Electrode (Cat. No ) See Instrument Preparation for step-by-step instructions. Replacement of O-rings 1. Replace flow cell O-ring and microvalve O-ring yearly in addition to monthly reagent and tubing connector assembly barb O-ring changes. Remove top of flow cell to replace O-ring. Yearly, replace O-ring on microvalve (standard injection port). Both O-rings may be found in Yearly Consumables Kit. Replacement of Restrictor Tubing 1. Close sample inlet valve. 2. Loosen fitting connector cap on top of flow meter assembly and pull out restrictor adapter and restrictor tube assembly. 3. Unscrew restrictor from fluid manifold inlet by loosening 5/8 white fitting. 4. Remove gray PVC adapter from end of old restrictor tube and save for installation on new tubing. 5. Install adapter on new restrictor tube and connect to fluid manifold. 6. Push adapter into fitting connector on top of flow meter assembly and hand tighten. 7. Turn on sample inlet valve and check for leaks. Hand tighten where necessary. restrictor adapter fitting restrictor tube assembly flow meter clamp Figure 15 Restrictor Tubing Assembly O-ring to flow cell manifold VI. 3
29 VI. 4
30 VII. TROUBLESHOOTING The following section covers troubleshooting that can be performed without special tools or skills. Note that in most installations, the temperature should read between C, millivolt reading between to -200 should be observed. Slopes between - 40 to -75 mv, and E o between -50 to +150 should be obtained. Within the U.S., ORION Service Personnel may be consulted for troubleshooting advice at (800) Within Massachusetts, call (978) Outside the U.S., contact your local ORION Dealer. Troubleshooting Checklist Malfunction Possible Cause Remedy Low Slope Improper calibration technique Recalibrate Sample electrode defective Replace and recalibrate Inaccurate standards Recalibrate with fresh standards Electronics failure Consult Orion Dynamic calibrator failure Check dynamic calibrator flow rates. Out-of-Range Electronics failure Try resetting computer (see APPENDIX), Slope or consult ORION. Reference and sensing cables Swap cables. interchanged Improper calibration technique Recalibrate. Sample electrode defective Replace. Inaccurate standards Recalibrate with fresh standards. Dynamic calibrator failure Check dynamic calibrator flow rates. Slow responding sensing electrode Polish sensing electrode and recalibrate Noisy Reference electrolyte not flowing properly Dry frit of reference electrode.filling solution should form small bead on frit within seconds. If no flow, check air pump flow. If OK, reference electrode is clogged and needs replacement. Electrode failure Replace electrode. Loose ground wire Check ground wire at flow cell. Monitor OK/Recorder defective Change recorder. Signal conditioner Compare LCD output with module output. If LCD is quiet while recorder is noisy, contact ORION. Temperature probe Engage test key to check temperature stability. If noisy, replace probe. Electronics failure Try resetting computer (see APPENDIX), or consult ORION VII. 1
31 Malfunction Possible Cause Remedy Excessive Drift Loose ground Check ground wire connection at flow cell. Tighten if necessary. Reference electrode not Make sure electrolyte bottle is flowing properly connected to air pump. Reference electrode clogged Replace reference electrode. Iodide electrode defective Replace electrode. Temperature probe Engage test key to check temperature stability. If noisy, replace probe. Burst diffusion tubing Monitor liquid level in reagent bottle. If liquid level rises, replace reagent and diffusion tubing.tubing connector may need replacement. Electronics failure Try resetting computer (see APPENDIX), or consult ORION. Low Flow Sample pressure below 8 psi Check sample pressure. If less Rate than 8 psi, increase. Pressure regulator set too low Increase pressure by pulling on red locking ring and turning black knob clockwise. Bypass filter clogged Replace or clean filter. Restrictor tubing crimped or clogged Replace restrictor tubing. Does Not Inaccurate standards or Use new standard solution, Calibrate Properly defective calibrator (2-point cal.) check dynamic calibrator flow rate. Reagent spent Replace reagent. Flow meter inaccurate Check monitor flow rate at drain, adjust to 40mL/min if required. Electrode failure Replace one or both electrodes. Noisy during calibration Check syringe for smooth plunger action; replace if required. Temperature probe failure Replace. Electronics failure Try resetting computer (see APPENDIX), or consult ORION. High or Low Monitor out of calibration Perform off-line or two-point calibration. Readings If off-line calibration performed, verify accuracy of alternate method value. Inlet filter just replaced Flush one hour until reading stabilizes. Reagent just changed Recalibrate after one hour. VII. 2
32 Malfunction Possible Cause Remedy Default Power loss while back-up battery was not Allow battery to charge overnight. fully charged Reprogram and recalibrate. Battery failure Consult ORION. Default values displayed Ensure enter/done key pressed after calibration after two pt. cal. complete key is lit. Incorrect Keyboard Computer glitch Turn main power off and on. Display Recalibrate. Electronics failure Try resetting computer (see APPENDIX), or consult ORION. Reads 1 Electrode cable failure, or not Loosen four screws holding main keypad connected to computer board Check that cables are attached securely. Electronics failure Try resetting computer (see APPENDIX), or consult ORION. VII. 3
33 VIII. REPAIR AND SERVICE Field Service is dispatched in the U.S. from the home office of ORION. In the United States, call (800) Within Massachusetts, call (978) In Europe, the Middle East, and Africa consult your authorized ORION dealer, or write: ORION RESEARCH, INC. 500 Cummings Center Beverly, MA USA Tel: Dom. Fax: Int l. Fax: ORION EUROPE. York Street Cambridge, England CB1 2PX Tel: Int l. Fax: ORION FAR EAST Room 904, Federal building 369 Lockhart Road Wanchai, Hong Knog Tel: Int l. Fax: ORION INDIA 302, 3rd floor, 9/2 East Patel Nagar New Delhi , India Tel: Fax: ORION CUSTOMER SUPPORT Toll Free: WWW: Tel: Int l. VIII. 1
34 IX. NOTICE OF COMPLIANCE This meter may generate radio frequency energy and if not installed and used properly, that is, in strict accordance with the manufacturer s instructions, may cause interference to radio and television reception. It has been type-tested and found to comply with the limits for a Class B computing device in accordance with specifications in Subpart J of Part 15 of FCC Rules, which are designed to provide reasonable protection against such interference in a residential installation. However, there is no guarantee that interference will not occur in a particular installation. If the meter does cause interference to radio or television reception, which can be determined by turning the unit off and on, the user is encouraged to try to correct the interference by one or more of the following measures: Reorient the receiving antenna Relocate the meter with respect to the receiver Move the meter away from the receiver Plug the meter into a different outlet so that the meter and receiver are on different branch circuits. If necessary, the user should consult the dealer or an experienced radio/television technician for additional suggestions. The user may find the following booklet prepared by the Federal Communications Commission helpful: How to Identify and Resolve Radio-TV Interference Problems. This booklet is available from the U.S. Government Printing Office, Washington, DC Stock No This digital apparatus does not exceed the Class B limits for radio noise emissions from digital apparatus set out in the Radio Interference Regulations of the Canadian Department of Communications. Le présent appareil numérique n émet pas de bruits radioélectriques dépassant les limites applicables aux appareils numériques de la class A prescrites dans le Règlement sur le brouillage radioélectrique édicté par le ministère des Communications du Canada. IX. 1
35 X. WARRANTY The ORION Model 1818AO Monitor is warranted to be free from defects in material and workmanship for a period of twelve (12) months from the date of installation or eighteen (18) months from the date of shipment from ORION, whichever is earlier, when used under normal operating conditions and in accordance with the operating limitations and maintenance procedures in the instruction manual, and when not having been subjected to accident, alteration, misuse, or abuse. This warranty is also conditioned upon expendable and consumable items (diffusion tubing, electrodes, and all solutions) being stored at temperatures between 59 F and 104 F (15 C and 45 C) in a non-corrosive atmosphere. Industrial electrodes are warranted to be free from defects in material and workmanship for a period of three (3) months from the date of installation or eighteen (18) months from the date of shipment when used under normal operating conditions and in accordance with the operating limitations and maintenance procedures given in the instruction manual and when not having been subjected to accident, alteration, misuse, or abuse. In the event of failure within the warranty period, ORION, or its authorized dealer will, at ORION s option, repair or replace the product non-conforming to the above warranty, or will refund the purchase price of the unit. THE WARRANTY DESCRIBED ABOVE IS EXCLU- SIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER STATUTORY, EXPRESS OR IMPLIED INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE AND ALL WAR- RANTIES ARISING FROM THE COURSE OF DEAL- ING OR USAGE OF TRADE. THE BUYER S SOLE AND EXCLUSIVE REMEDY IS FOR REPAIR, OR REPLACEMENT OF THE NON-CONFORMING PRODUCT OR PART THEREOF, OR REFUND OF THE PURCHASE PRICE, BUT IN NO EVENT SHALL ORION (ITS CONTRACTORS AND SUPPLIERS OF ANY TIER) BE LIABLE TO THE BUYER OR ANY PERSON FOR ANY SPECIAL INDIRECT, INCIDEN- TAL OR CONSEQUENTIAL DAMAGES WHETHER THE CLAIMS ARE USED IN CONTRACT, IN TORT (INCLUDING NEGLIGENCE) OR OTHERWISE WITH RESPECT TO OR ARISING OUT OF THE PRODUCT FURNISHED HEREUNDER. Industrial products used at nuclear facilities are also subject to ORION nuclear terms and conditions. Contact ORION if you do not have a copy. Representations and warranties made by any person, including its authorized dealers, distributors, representatives, and employees of ORION, which are inconsistent or in addition to the terms of this warranty shall not be binding upon ORION unless in writing and signed by one of its officers. X. 1
36 XI. ORDERING INFORMATION Cat. No. 1818AO HYDRAK ELIMAK Description Microprocessor-based Oxygen Scavenger Monitor, complete with iodide electrode (100022), pressurized assembly with attached push pin for use with reference electrode (100056), one instruction manual, and signal output module. Does not include software for operation. Requires appropriate Application Kit for Oxygen Scavenger being measured and reagent (181811). Depending on oxygen scavenger and calibration method, dynamic calibrator (15DC15/20) may be required. Wired for 100/115/220/240 V AC, 50/60 Hz. Application Kit for measurement of HYDRAZINE. Includes software, application-specific mini door label, and APPENDIX to 1818 Instruction Manual. Application Kit for measurement of ELIMIN-OX. Includes software, application-specific mini door label, and APPENDIX to 1818 Instruction Manual Iodide Electrode for use with detachable cable Reference Electrode for use with detachable cable Pressure Regulator A01 Flow Meter Assembly (181124) Restrictor Tube Assembly Temperature Sensor (located in drain on 1818) Inlet Filters, stainless steel, 60 micron filters with gaskets, pack of Filter Gaskets for use with , pack of Air Pump Assembly Signal Conditioner Optional Alarm Module. ELIMIN-OX is a registered trademark of Nalco Chemical Co. XI. 1
37 Consumables Cat. No. Description * Reagent (three 1-liter bottles), includes three sets diffusion tubing and O-rings. Six month supply Diffusion tubing (three 6 ft. lengths) with fifteen small O-rings for tubing connector barbs, and three large O-rings for reagent bottle Reference electrode internal filling solution, five 2-oz. bottles. Average three month supply when used with pressurized reference assembly * Consumables kit for one year s operation, consists of two sets of three 1-liter bottles reagent with diffusion tubing and O-rings, (2 x ), iodide electrode (100022), reference electrode (100056), four sets of five 2-oz. bottles reference electrode internal fill solution (4 x ) for use with pressurized reference assembly, O-rings for flow cell and microvalve, one restriction tube assembly (181125), two 60 micron inlet filters (181170), one filter gasket, and check valve. (Does not contain standard [Cat. No ] needed for two point calibration.) * Contains hazardous material. For Two Point Calibration Part No. Description 15DC15/15DC20 Dynamic calibrator for use with multiple instruments. Includes syringe kits (Cat. Nos and ) and mounting bracket ( S01 for use with either 1500 or 1800 Series monitors.) Standardizing Solution, one pint Syringe Kit with 20 cc syringe for use with 1818 during two point calibration. Spare Parts Part No. Description A01 Electrode cable A01 Power board A01 Keypad with driver board A01 Power supply to PMA cable Power switch A01 Electrode holder, excluding electrodes A01 Tubing connector assembly with gray cap and barbs, and white reagent bottle cap A01 Connector block assembly with thumb screw and injection port Standard injection port. 1SMXXX Thumbscrew (on fluid block) with three washers A01 Pressurized reference assembly for use with Pressurizing needle A01 Tubing spool assembly. * NOTE: For sample alkalinities up to 5 ppm, Reagent life is extended to two months. The reagent life is six weeks with sample alkalinities from above 5 to 10 ppm as CaCO 3. * Contains hazardous material. XI. 2
38 XII. MOUNTING DIMENSIONS NOTE: Two Point Calibration If dynamic calibrator and optional mounting shelf are used, leave 12 clearance beneath analyzer ".375" DIA " 24.75" 22.75" 23.50" 26.00" FRONT VIEW 7.00" 10.06" 1.53" TB " SIDE VIEW 3.00" 10.00" 5.03".88" 10.50" cm 0.125" cm " 3/4" NPT-male 2.25" 1/4" NPTF 5.25" 10.06" 9.32" " 1.125" BOTTOM VIEW 3.00" 12.50" 15.75".88" DIA. 3 HOLES XII. 1
39
40 XIII. WIRING DIAGRAM TRANSFORMER 240V 220V 115V 100V COMM RED GRN-YEL ORN BLU } TO AIR PUMP POWER CABLE POWER SUPPLY BD WHT BLK POWER IN REF { HOT COMM GND GROUND TB J1 TB V 220V 115V 100V COMM SEE NOTE 1 NOTE 1 Jumper - At installation, remove one side of jumper and reconnect it to one of four terminals (3,4,5, or 6) to obtain desired voltage. CB1 3 1 LINE 4 2 XIII. 1
41
42
43 Rev. C
TABLE OF CONTENTS. V. Calibration 23 Three-Point Calibration Procedure 24 One-Point Calibration Procedure 24 Off-line Calibration Procedure 24
TABLE OF CONTENTS I. General Information 1 Introduction 1 Principles of Operation 1 Principles of Calibration 2 Sample Requirements 3 Description of Orion 1817LL Monitor 4 II. Instrument Preparation 7
Orion High-Level Chloride Monitor
Orion 1817HL Orion High-Level Chloride Monitor INSTRUCTION MANUAL rifos noglib rifos dome glid off a b c d Pure Water MONITORS 1800 series release latches to access Analyze Detect Measure Control AQUAfast,
Model 1811EL Low-Level Sodium Monitor
Model 1811EL Low-Level Sodium Monitor INSTRUCTION MANUAL test values C mv slope E 0 sample values ppm ppb error stable enter/ done sample test program cal. begin cal two pt. cal. complete error add std.
OPL BASIC. Dosing System for Professional Laundry machines. Contents
OPL BASIC Dosing System for Professional Laundry machines Contents 1 Getting Started. Page 2 2 Installation. Page 4 3 Set Up & Operation. Page 8 4 Maintenance & Accessories. Page 10 5 Troubleshooting Page
Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01
Milwaukee Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: SM600 Smart DO Meter Authorized Dealer: ISMIL600 11/01 PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery
RI-215A Operator s Manual. Part Number: 71-0045RK Revision 0 Released: 10/3/05
RI-215A Operator s Manual Part Number: 71-0045RK Revision 0 Released: 10/3/05 Warranty RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials and workmanship,
Milwaukee USER MANUAL
Milwaukee USER MANUAL PORTABLE DISSOLVED OXYGEN METER MODEL: MW600 Smart DO Meter PROBE PREPARATION: The meter is supplied with a 9V battery. Slide off the battery compartment cover on the back of the
Mobile Data Power Model: MDP-25
Mobile Data Power Model: MDP-25 Topic Section Features... 2 Operational Features Summary... 2 Back-up Battery Power Internal Charger Voltage Spike Protection RF Noise Filtering Warning of Imminent Loss
3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE
4 3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE A. Changing the Control Board B. Replacing a Heating Element C. Thermistor Replacement D. High Limit Switch Replacement E. Level Detector Replacement
User Guide. Heavy Duty Dissolved Oxygen Meter. Model 407510
User Guide Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
User Manual. Thermo Scientific Orion 2111LL
User Manual Thermo Scientific Orion 2111LL Low Level Sodium Analyzer UM-263639-004 Revision C October 2013 Contents Section 1 General Information... 6 Introduction... 6 Markets... 6 Applications... 7
Express5800/120Ed. Rack Mount Kit Installation Procedures PN: 455-01607-001
Express5800/120Ed Rack Mount Kit Installation Procedures PN: 455-01607-001 Proprietary Notice and Liability Disclaimer The information disclosed in this document, including all designs and related materials,
OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE
OEM Manual MODEL 2350 ELECTRONIC DUAL CYLINDER SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA &
BASYS INSTALLATION INSTRUCTIONS FOR ELECTRONIC CAPACITANCE SENSING LAVATORY FAUCETS EFX-3XX Pedestal, Sensor Activated Lavatory Faucets
BASYS INSTALLATION INSTRUCTIONS FOR ELECTRONIC CAPACITANCE SENSING LAVATORY FAUCETS EFX-3XX Pedestal, Sensor Activated Lavatory Faucets Any Application. Any Environment. Code No: 0816407 Rev. 4 (07/12)
User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending
User Guide ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens Patent Pending Getting Started Remove the cap from the bottom of the ExStik TM to expose the electrode bulb and reference
Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE
Waters Corporation Waters 2690/5 USER & TROUBLESHOOTING GUIDE Contents 2690/5 Theory Setup procedures. Troubleshooting the 2690/5 User maintenance of the 2690/5 Spare Parts 2 2690/5 Theory 2690/5 Solvent
Service Guide 12/27/03 TESTING, SERVICE & REPAIR GUIDE (For SH Space Heating Models & RA Water Heating Models)
TESTING, SERVICE & REPAIR GUIDE (For SH Space Heating Models & RA Water Heating Models) WARNING - HIGH VOLTAGE AC electrical circuits are connected to this heater. Do not attempt any service work on the
VOYAGER 570G. 744A Sprayer Control
VOYAGER 570G 744A Sprayer Control U S E R M A N U A L U S E R M A N U A L Table of Contents CHAPTER 1 - INTRODUCTION...1 SYSTEM CONFIGURATIONS...1 KIT CONTENTS...3 CONTROL HOUSING ASSEMBLY...5 CHAPTER
Advantium 2 Plus Alarm
ADI 9510-B Advantium 2 Plus Alarm INSTALLATION AND OPERATING INSTRUCTIONS Carefully Read These Instructions Before Operating Carefully Read These Controls Corporation of America 1501 Harpers Road Virginia
ExStik TM ph Waterproof Meters
User Guide ExStik TM ph Waterproof Meters Models PH100 & PH110 Patent Pending ExStik TM Description Front Panel Controls 1. Battery compartment cap 2. LCD Display 3. MODE / HOLD button 4. CAL / RECALL
5800 Temperature Sensor Cable Assembly
5800 Temperature Sensor Cable Assembly Removal and Replacement Instruction Sheet #60-4702-070 Revision D, January 14, 2013 Overview The 5800 has two refrigeration temperature sensors, one attached to the
MCR1900 Media Converter 19-Slot Chassis
MCR1900 Media Converter 19-Slot Chassis Installation Guide Part #5500304-11 Copyright Statement This document must not be reproduced in any way whatsoever, either printed or electronically, without the
B150 Series Commercial Brewer. Direct Plumb Kit
B150 Series Commercial Brewer Welcome Keurig Authorized Dealer! Direct Plumb Kit This manual provides: Installation of the Direct Plumb Kit (DPK) that allows you to connect your customer s B150 brewer
BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS
BUILT-IN DISHWASHER INSTALLATION INSTRUCTIONS PLEASE READ COMPLETE INSTRUCTIONS BEFORE YOU BEGIN LEAVE INSTALLATION INSTRUCTIONS AND USER'S GUIDE WITH OWNER ALL ELECTRIC WIRING AND PLUMBING MUST BE DONE
Portable Air Conditioner
Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE
User's Manual. Heavy Duty Dissolved Oxygen Meter Model 407510
User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved
Uninterruptible Power Supply. 500 kva - 750 kva
9 315 Uninterruptible Power Supply 500 kva - 750 kva T1 and T3 Installation Addendum 164201244-001 Rev. A ------------------------------------------------------------------------ ------------------------------------------------------------------------
http://waterheatertimer.org/troubleshoot-rheem-tankless-water-heater.html
http://waterheatertimer.org/troubleshoot-rheem-tankless-water-heater.html TECHNICAL SERVICE DEPARTMENT Removal, Cleaning, & Reinstallation of the Burner Assembly For models 74 & GT199 Required tools -
DISSOLVED OXYGEN PROBE
51970-88 DISSOLVED OXYGEN PROBE Introduction This instruction sheet provides information on the Hach Dissolved Oxygen Probe for use with the sension 6 Dissolved Oxygen Meter. The Dissolved Oxygen Probe
AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR. Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002
AIR QUALITY SURVEILLANCE BRANCH ACCEPTANCE TEST PROCEDURE (ATP) FOR Teledyne Advanced Pollution Instruments Model 400 E Ozone Analyzer AQSB ATP 002 First Edition MONITORING AND LABORATORY DIVISION August
TS93 EMR T/PT/TDE. Surface applied door closer
TS EMR T/PT/TDE Surface applied door closer Installation instructions: Pull side track mount door closer with smoke detector (EMR T) Push side track mount door closer with smoke detector (EMR PT) Double
IMPORTANT SAFETY RULES TO FOLLOW
WARNING FLOOR & CARPET CLEANER Any piece of equipment can be dangerous if not operated properly. YOU are responsible for the safe operation of this equipment. The operator must carefully read and follow
Installation Instructions 6028.801
DAZZLE Installation Instructions 08.80 Spread Lavatory Faucet with Speed Connect Drain* Congratulations on purchasing your American Standard faucet with Speed Connect drain, a feature found only on American
ETC TWO STAGE ELECTRONIC TEMPERATURE CONTROL
RANCO INSTALLATION INSTRUCTIONS ETC TWO STAGE ELECTRONIC TEMPERATURE CONTROL Relay Electrical Ratings PRODUCT DESCRIPTION The Ranco ETC is a microprocessor-based family of electronic temperature controls,
Technical Update TAA.TU.11093 Rev. 1
http://www.gambro.com/en/usa_tech/ 800-525-2623 303-222-6500 Technical Update TAA.TU.11093 Rev. 1 Effective: 05 APR 2013 CO# 13084 Product: Subject: From: Phoenix Dialysis System Required Electrical Safety
with MERCURY FREE 1 HP Relays ! WARNING Before using this product read and understand instructions.
B Installation & Maintenance Instructions MM-414 Series 150E and 157E Low Water Cut-Off/Pump Controllers For Steam Boilers and Other Level Control Applications A Typical Applications: Primary or secondary
i ChatterBox! Motorcycle Security
i Before you Start the Installation * Please read this manual to become familiar with the requirements necessary to complete the installation. * Use a high quality multi-meter to test all wires before
PART 1 - INTRODUCTION...
Table of Contents PART 1 - INTRODUCTION... 3 1.1 General... 3 1.2 Sensor Features... 3 1.3 Sensor Specifications (CDE-45P)... 4 Figure 1-1 CDE-45P Sensor Dimensions (standard, convertible style)... 4 PART
Installation Instructions 4508 4508S
SYMPHONY Spread Lavatory Faucet with Speed Connect Drain Congratulations on purchasing your American Standard faucet with Speed Connect drain, a feature found only on American Standard faucets. Speed Connect
PRODUCT: WASHER / WASHER-DRYER COMBO MODEL: AW 120 / AW 122 / AW 125 AWD 120 / AWD 121 / AWD 129
PRODUCT: WASHER / WASHER-DRYER COMBO MODEL: The information included in this Splendide Repair Manual may change without notice. Please see our web site www.splendide.com/service/docs.html for updates,
Fleck 4650. Service Manual INSTALLATION AND START-UP PROCEDURE TABLE OF CONTENTS JOB SPECIFICATION SHEET
Fleck 4650 Service Manual TABLE OF CONTENTS JOB SPECIFICATION SHEET...1 INSTALLATION AND START-UP PROCEDURE...1 CONTROL VALVE DRIVE ASSEMBLY...2 CONTROL DRIVE ASSEMBLY FOR CLOCK...3 BYPASS VALVE ASSEMBLY...4
ATS Overhead Table Shelf System INSTRUCTION MANUAL
ATS Overhead Table Shelf System INSTRUCTION MANUAL ATS Overhead Table Shelf System Instruction Manual Warranty Newport Corporation warrants this product to be free of defects in material and workmanship
Daker DK 1, 2, 3 kva. Manuel d installation Installation manual. Part. LE05334AC-07/13-01 GF
Daker DK 1, 2, 3 kva Manuel d installation Installation manual Part. LE05334AC-07/13-01 GF Daker DK 1, 2, 3 kva Index 1 Introduction 24 2 Conditions of use 24 3 LCD Panel 25 4 Installation 28 5 UPS communicator
User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600
User's Guide ExStik DO600 Dissolved Oxygen Meter II RECALL Dissolved Oxygen DO600 Introduction Congratulations on your purchase of the ExStik DO600 dissolved oxygen / temperature meter which simultaneously
GPS AutoSteer System Installation Manual
GPS AutoSteer System Installation Manual Supported Vehicles John Deere Sprayers 4720 4630 4730 4830 AutoTrac Ready PN: 602-0227-01-A LEGAL DISCLAIMER Note: Read and follow ALL instructions in this manual
OASIS-PLUS 120V READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE
OASIS-PLUS 120V INFORMATION & OPERATING INSTRUCTIONS READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE 56041852 WARNING:
Installation Instructions 4662.003
Installation Instructions.00 SINGLE CONTROL KITCHEN FAUCET Thank you for selecting American-Standard...the benchmark of fine quality for over 00 years. To ensure that your installation proceeds smoothly--please
Powers Controls TH 192 HC Heating/Cooling Room Thermostat
Powers Controls TH 192 HC Heating/Cooling Room Thermostat Technical Instructions Document No. 155-066P25 TH 192-2 50 60 70 80 70 TH0356R1 60 80 POWERS Description The TH 192 HC thermostats are proportional
MODEL 2202IQ (1991-MSRP $549.00)
F O R T H E L O V E O F M U S I C F O R T H E L O V E O F M U S I C MODEL 2202IQ (1991-MSRP $549.00) OWNER'S MANUAL AND INSTALLATION GUIDE INTRODUCTION Congratulations on your decision to purchase a LINEAR
SECTION G2: CABLE PROCESSOR MODULE MAINTENANCE
SECTION G2: CABLE PROCESSOR MODULE MAINTENANCE Cable Processor Module overview WARNING! When tipping the Cable Processor Module back, (after removing the toggle arm pin), use extreme caution not to drop
Dissolved Oxygen Sensors
Instruction Sheet PN 51A-HxDO/rev.F June 2008 Models Hx438 and Gx448 Dissolved Oxygen Sensors For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit
Trace Dissolved Oxygen Sensor
Instruction Sheet PN 51A-499ATRDO/rev.G September 2010 Model 499A TrDO Trace Dissolved Oxygen Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION
Air-Oxygen Blender. Service Manual. Model No. PM5200 Series PM5300 Series (shown)
Service Manual Model No. PM5200 Series PM5300 Series (shown) SAVE THESE INSTRUCTIONS 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com
Duct Humidity Transmitter
SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers
Measurement by Ion Selective Electrodes
Measurement by Ion Selective Electrodes Adrian Vazquez May 8, 2012 Why Use Ion Selective Electrodes? Responsive over a wide concentration range Not affected by color or turbidity of sample Rugged and durable
Manual for Fire Suppression & Methane Detection System
Manual for Fire Suppression & Methane Detection System Fogmaker North America Post address: 150 Gordon Dr Exton, PA 19341 Delivery address: 150 Gordon Dr Exton, PA 19341 Tel: 610-265-3610 Fax: 610-265-8327
Your safety and the safety of others are very important.
NATURAL GAS TO PROPANE CONVERSION KIT 090 INSTALLATION INSTRUCTIONS FOR ALTITUDES 0 -,00 FT. ONLY PROPANE CONVERSION KIT SAFETY... INSTALLATION REQUIREMENTS... Tools and Parts... LP Gas Requirements...
FUEL CELL CAR SCIENCE KIT ASSEMBLY GUIDE. Battery operation instructions:
FUEL CELL CAR SCIENCE KIT ASSEMBLY GUIDE Battery operation instructions: 1. The removing and inserting of batteries is to be conducted by the adults only. Unscrew the screw holding the battery pack s cover
Installation and Troubleshooting Instructions for Electric Tankless Residential Water Heaters.
Model Number: Serial Number: Information Manual Installation and Troubleshooting Instructions for Electric Tankless Residential Water Heaters. ATTENTION: IF YOU ARE NOT A LICENSED PLUMBER OR A LICENSED
site monitoring Kit Site Monitoring Kit User Manual we prove it.
site monitoring Kit 1 User Manual Site Monitoring Kit User Manual we prove it. 2 site monitoring Kit Content 1. Introduction 3 2. Content of the Site Monitoring Kit 4 3. Preparation 6 1. Positioning of
www.servicechamp.com
1-800-221-0216 Fax: 1-800-472-2281 www.servicechamp.com Service Champ Part 52081 Service Interval every 30,000 miles / Chevrolet 1992-2002 6.5 liter Description and Operation The fuel filter element separates
INSTALLATION AND OPERATING INSTRUCTIONS For Model GL1 Gate Locks
Securitron Magnalock Corp. www.securitron.com ASSA ABLOY, the global leader Tel 800.624.5625 [email protected] in door opening solutions INSTALLATION AND OPERATING INSTRUCTIONS For Model GL1 Gate
Service manual. Website: www.andico.com.au CAUTION - BEFORE SERVICING THE UNIT, READ THE SAFETY - PRECAUTIONS IN THIS MANUAL.
Website: www.andico.com.au Service manual CAUTION - BEFORE SERVICING THE UNIT, READ THE SAFETY - PRECAUTIONS IN THIS MANUAL. - ONLY FOR AUTHORISED SERVICE PERSONNEL. MODELS: MPK1-09CR-QB8 MPK1-12ER-QB6
430 Power/Electronics Replacement
Replacing the main board WARNING Before proceeding, turn off the main power switch and unplug the power cord. Caution Make sure you are properly grounded with an ESD strap before continuing. The main printed
prestige Condensing Water Boiler SERVICE TECHNICIAN S TROUBLE SHOOTING GUIDE 2008-36 TSG-PRESTIGE Date revised: 12/6/10
prestige Condensing Water Boiler SERVICE TECHNICIAN S TROUBLE SHOOTING GUIDE Date revised: 12/6/10 2008-36 TSG-PRESTIGE Table of Contents INTRODUCTION Page 1-2 SERVICING TIPS AND INSTRUCTIONS Page 3-4
Automatic Phone-Out Home Monitoring Systems
Automatic Phone-Out Home Monitoring Systems Power Outage and Freeze Alarm Model Number: THP202 Power Outage, Freeze and Flood Alarm Product Description Model Number: THP201 These monitoring systems are
ph Measurement and Control
ph Measurement and Control Tech Tip #14 1997 I. The Need for ph II. ph System Requirements III. ph Electrode Conditioning Procedures IV. How the electrode works V. Identifying the Electrode Components
DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR. Installation and Operation Manual
DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR Installation and Operation Manual 1 ENGLISH Safety Instructions 2 Features 2-3 Specifications 3 Installation 4-5 Wiring Diagrams 6-7 Warranty 8
Periodic Verification System Model 775PVS
Periodic Verification System Model 775PVS Instruction Manual Contents 1 Description... 1 775PVS System... 1 Model 775 Fieldmeter... 2 Model 775C Charger... 2 Model 775P Plate Assembly... 3 2 Operation...
Name of Equipment Silver King Model SKMCD1P/C1. This equipment chapter is to be inserted in the appropriate section of the Equipment Manual.
Name of Equipment Silver King Model SKMCD1P/C1 This equipment chapter is to be inserted in the appropriate section of the Equipment Manual. Manufactured exclusively for McDonald s By Silver King Refrigeration,
Line to Refrigerator Ice/Water Dispenser
Standard 18 Line to Refrigerator Ice/Water Dispenser Pump Module Bottled Water How The System Works The FLOJET Bottled Water Dispensing System was designed to pump water from a commercially available 5-gallon
123 Industrial Loop Road Paynesville, MN 56362 Phone: 1-800-864-1649 www.master-mfg.com MASTER MANUFACTURING MASTER GARDNER
123 Industrial Loop Road Paynesville, MN 56362 Phone: 1-800-864-1649 www.master-mfg.com MASTER MANUFACTURING MASTER GARDNER Part Number PCD E3 009B MM Rev 1 Nov. 2010 INTRODUCTION The purpose of this manual
EVANS ELECTRONIC TEMPERATURE CONTROL TROUBLESHOOTING GUIDE for systems equipped with electric coolant valve and external PC board.
EVANS ELECTRONIC TEMPERATURE CONTROL TROUBLESHOOTING GUIDE for systems equipped with electric coolant valve and external PC board. This Troubleshooting Guide covers the electric coolant valve and control
OWNER S MANUAL FORCE 10 MARINE COMPANY 23080 HAMILTON ROAD RICHMOND, BC CANADA V6V 1C9 TEL: (604) 522-0233 FAX: (604) 522-9608
Electric Water Heater OWNER S MANUAL FORCE 10 MARINE COMPANY 23080 HAMILTON ROAD RICHMOND, BC CANADA V6V 1C9 TEL: (604) 522-0233 FAX: (604) 522-9608 If your water Heater is Damaged or you have questions
UPGRADING AND SERVICING GUIDE
UPGRADING AND SERVICING GUIDE HPTouchSmart 610 PC Computer features may vary by model. Removing and Replacing a CD/DVD Drive...2 Removing the CD/DVD Drive... 2 Installing a New CD/DVD Drive... 5 Removing
Water Leak Detection System
Water Leak Detection System Installation and Operating Manual 505-334-5865 ph 505-334-5867 fax www.rodisystems.com email:[email protected] 936 Highway 516 Aztec, NM 87410-2828 Manual Revisions and Copyright
THERMOSTATIC MIXING VALVE TRIM K-T9493, K-T9494
THERMOSTATIC MIXING VALVE TRIM K-T9493, K-T9494 1. BEFORE YOU BEGIN IMPORTANT INSTRUCTIONS READ AND SAVE FOR THE CONSUMER WARNING: Risk of scalding or other severe injury. Before completing installation,
Packaged Terminal Air Conditioner Wall Sleeve Installation
Installation & Maintenance Data IM 1196 Group: PTAC Part Number: 910141799 Date: October 2013 Packaged Terminal Air Conditioner Installation x 42" PGAN with Top-Mounted Hydronic Heat Note: Installation
INSTALLATION GUIDE. Card Reader & Controller with KIM Swipe Reader for Solitaire 850 / 950 / 850L Learnlok PK2930
INSTALLATION GUIDE Card Reader & Controller with KIM Swipe Reader for Solitaire 850 / 950 / 850L Learnlok PK2930 Card Reader and Controller Model 3.5 with KIM Swipe Reader Table of Contents 1. Features..................................
Medical Air Systems: Scroll
GENERAL The Chemetron skid mount scroll medical air system is designed to provide medical breathing air for hospital and medical institutions. This system meets NFPA 99 requirements for Level 1 breathing
Bubble King. User Manual
Bubble King User Manual TABLE OF CONTENTS 1. Before You Begin... 3 What Is Included... 3 Unpacking Instructions... 3 Claims... 3 Text Conventions... 3 Icons... 3 Document Information... 3 Product at a
HP UPS R1500 Generation 3
HP UPS R1500 Generation 3 Installation Instructions Part Number 650952-001 NOTE: The rating label on the device provides the class (A or B) of the equipment. Class B devices have a Federal Communications
Refrigerant Charging Unit ICOGD. 020AH1000 Operating Manual. FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00
E Refrigerant Charging Unit ICOGD 020AH1000 Operating Manual FR.8.2.4-09 İ-COLD 12.03.2014 Rev. 00 Contents Technical Specifications... 20 Safety... 21 A/C System... 22 Components... 23 Control Panel...
RAY-MAX Integrated Solar Power Strip
RAY-MAX Integrated Solar Power Strip 600008, 600009, 600010, 600208, 600209, 600210 Owner s Manual NEXTRONEX, INC. Revision Date: 10/27/14 Contents 1. Safety Instructions... 3 2. General Equipment Warnings...
Installation Operation Maintenance Troubleshooting
Installation Operation Maintenance Troubleshooting Version 3, 2.1, June/09 Apr/09 1416 1402 Table of Contents Important information 2 Weights & Dimensions 3 Operator s Instructions 4 Flow Diagram 5 Purging
Oil and Coolant Circulating Heating System. Model - OCSM
Oil and Coolant Circulating Heating System Model - OCSM Installation & Operation Manual 216280-000 REV 2 Identifying Your System The HOTSTART heating system is designed to heat fluids for use in marine
Pro Logic Version 4.10. Diagnostics Manual
Pro Logic Version 4.10 Diagnostics Manual Turbo Cell & Control Electronics 2009 Hayward Industries Table of Contents Important safety instructions Pg. 1 No Cell Power 1 & No Cell Power 2 Pg. 2 No Cell
G670 Intermittent Pilot Ignition Controls
Installation Sheets Manual 121 Gas Combustion Combination Controls and Systems Section G Technical Bulletin Issue Date 0300 Intermittent Pilot Ignition Controls Figure 1: Intermittent Pilot Ignition Control
NELSON VOLTAGE MONITOR INSTALLATION & PROGRAMMING MANUAL
NELSON VOLTAGE MONITOR INSTALLATION & PROGRAMMING MANUAL CONTENTS GENERAL INFORMATION...3 INSTALLATION...3 FIELD WIRING...4 PROGRAMMING...4 Circuit Monitor Options...5 Power Frequency...5 Alarm Silence
Submersion/Insertion Toroidal Sensor
Instruction Sheet PN 51A-228/rev.P December 2010 Model 228 Submersion/Insertion Toroidal Sensor For additional information, please visit our website at www.emersonprocess.com/raihome/liquid/. CAUTION SENSOR/PROCESS
ENGINE FUEL FUEL FILTER... FUEL HEATER... INJECTOR... SUPPLY PUMP... COMMON RAIL... FUEL PRESSURE LIMITTER...
FUEL FILTER............................ FUEL HEATER.......................... INJECTOR.............................. SUPPLY PUMP.......................... COMMON RAIL.......................... FUEL PRESSURE
Reliance DuraStar ORDINARY LOCATION INSTALLATION, OPERATING & MAINTENANCE INSTRUCTIONS
STORAGE and HANDLING The LED Illuminator meets or exceeds all applicable specifications when shipped from the factory. All units should be inspected upon receipt to ensure that no damage has been incurred
JNIOR. Overview. Get Connected. Get Results. JNIOR Model 310. JNIOR Model 312. JNIOR Model 314. JNIOR Model 410
The INTEG is an Ethernet I/O (digital, analog) device that monitors and controls a small set of process signals. functions as both basic I/O for integration with another application or system AND as a
CEILING CASSETTE FAN COIL UNIT
CEILING CASSETTE FAN COIL UNIT Operation and Installation Manual MHCFC4W-04, 08, 12, 16 Four way ceiling cassette fan coil unit Please read this manual before using the fan coil unit. Installation and
KUSTER ELECTRONIC GAUGES
Kuster Company 2900 E. 29th Street Long Beach, CA USA 90806 Telephone 562-595-0661 Fax 562-426-7897 E-Mail [email protected] KUSTER ELECTRONIC GAUGES K8 EMR 25, K8 EMR 39, K4 EMR ¾, K4 EMR ¾ QUARTZ SERVICE
INSTALLATION INSTRUCTIONS GEO-BOOSTER TM
Table of Contents General Description 1 Parts list 1 Installation 2 New Ground Loop 2 Retrofit 4 Temporary 6 Troubleshooting 8 INSTALLATION INSTRUCTIONS GEO-BOOSTER TM General Description The Geo-Booster
TABLE OF CONTENTS. I. TROUBLESHOOTING... 2 - Section 1.01: Common Problems/Solutions... 2
BAL Accu-Slide System I. Table of Contents TABLE OF CONTENTS I. TROUBLESHOOTING... 2 - Section 1.01: Common Problems/Solutions... 2 II. GETTING STARTED... 5 - Section 2.01: Tools You Will Need... 5 - Section
Installation Operation Maintenance Troubleshooting Version 2.1, Apr/09 DAYTONA
Installation Operation Maintenance Troubleshooting Version 2.1, Apr/09 1607 DAYTONA Congratulations! All of us at Summit Dental Systems want you to know that your 1607 Daytona Cuspidor has been built with
Watts Pure Water Systems
Watts Pure Water Systems WM-120-PT Installation and Operation Manual Premier Water Systems Do not use with water that is microbiologically unsafe or of unknown quality without adequate disinfection before
Overview PARTS LIST. B. Lever mounting base C. Flush handle assembly D. Grey/Blue float stop E. Grey float (Full Flush) F. Flush valve washer
Overview READ ENTIRE INSTRUCTIONS BEFORE STARTING INSTALLATION PARTS LIST A. Flush valve B. Lever mounting base C. Flush handle assembly D. Grey/Blue float stop E. Grey float (Full Flush) F. Flush valve
