Back to Basics: Proper Care and Maintenance for Your Cell Culture Incubator
|
|
|
- Warren Lang
- 10 years ago
- Views:
Transcription
1 Back to Basics: Proper Care and Maintenance for Your Cell Culture Incubator Mary Kay Bates, M.S., Global Cell Culture Specialist; Jennifer D Onofrio, Ph.D., Application Specialist; Molly Love Parrucci, Global Product Manager, CO 2 Incubators; Douglas Wernerspach, Global Business Director, CO 2 Incubators and Constant Temperature Thermo Fisher Scientific Technical Note: TNCO2CAREFEED 0514 There are many different CO 2 incubators today, in different labs around the world. They all share some similarities. You trust your work your cells to your incubator. Having a well designed, properly functioning incubator means that you will never have to worry about it. In such an incubator, your cells will grow well, contamination problems are rare, and the incubator is easy to clean and maintain. Everyone using a cell culture incubator needs to understand a few important points in order to keep any incubator working its very best. The information presented here will help you to avoid common mistakes that can result in project delays. Please note that while these recommendations are generally applicable for any CO 2 incubator, we suggest that you consult your user manual to ensure you are also following the manufacturer s best practices. Positioning the CO 2 incubator Where you place the incubator can alter its performance. Ventilation, air flow, room temperature, and direct sunlight can all affect the temperature and humidity functions. When installing a CO 2 incubator, be sure to lift only by the sides of the bottom, never from front and back and do not ever lift using the door. Do not place any incubator directly on the floor. Instead, use a support stand. If the incubator sits directly on the floor, the air movement that is created each time the door is opened will sweep dirt, dust and contaminants directly into the incubator chamber. When stacking incubators, you should only stack similar brands. If you stack dissimilar units, you risk the stack tipping over or the top unit slipping off, since there is generally no mechanism to secure incubators that don t match. Next, ensure there is adequate clearance on all sides of the incubator. This allows for ventilation and access to power cords and connectors including gas hookup. Try to place the incubator away from traffic areas, but avoid damp, humid corners that may harbor fungal growth. Make sure the unit is sheltered from ventilation and other airstreams, because these air currents can direct contamination-carrying dirt and dust into the incubator. Do not position the incubator in direct sunlight, because fluctuating temperatures outside can affect the anti-condensation functions. Finally, ensure that the incubator is level both front to back and side to side by adjusting the leveling feet or stand.
2 2 Installation Each incubator model is different regarding connections and programming, but there are some general guidelines that apply to all incubators. Properly installing the incubator from the beginning will help provide the best conditions for your cells going forward. After the incubator is properly positioned, clean and disinfect the incubator interior, the shelves and shelf supports, and air ducts if applicable (see below for cleaning recommendations). After disinfection, install these and any other internal components, following the manufacturer s directions. Note: If an automated heat decontamination/sterilization cycle is available, you can skip the manual disinfection step and simply run the automated program assuming the program is proven to eliminate accepted biological indicator organisms. An automated high heat program also eliminates the need for autoclaving any internal parts. It s important to mention that while recent Thermo Scientific models with a high temperature cycle require no handling of parts to run the cycle, some earlier models and many from other manufacturers have sensors that must be removed first. When the incubator is clean, disinfected, and the shelf parts are in place, you can install any additional optional parts such as an oxygen sensor or HEPA filter. An in-chamber HEPA filtration system from Thermo Fisher Scientific will filter the entire chamber air volume every 60 seconds, establishing ISO-5 cleanroom air quality in the chamber within 5 minutes after every door opening 1. Note: A HEPA filter that is not located inside the chamber will not experience the same conditions as your cells and cannot provide the same level of assurance. Handle HEPA filters and gas sensors very carefully, as they are easily damaged. If you have a water jacketed incubator, fill the water jacket at this time. Also, fill the humidity pan with sterile distilled water, to ½ inch (1.25 cm) from the top. Keeping the water pan full is important for good, consistent humidity and best growth conditions. Now, using the electronics and following the manufacturer s directions, set the temperature, over-temperature, undertemperature, CO 2 and O 2 setpoints, and the appropriate alarms. If you have a Thermo Scientific incubator with the ican interface, you can also set the data logging parameters at this time. Depending on your settings, the ican will collect data to be stored onboard for up to a week, or exported to Microsoft Excel. This information can often be very useful, not only for GMP monitoring, but also in helping to pinpoint the start of any problems. CO 2 Monitoring There are two kinds of CO 2 gas sensors. The first is the T/C (thermal conductivity) sensor. Thermo Scientific incubators offer a T/C sensor and electronics that are very stable, accurate and economical. In fact, T/C sensors are our most popular method of CO 2 control. They are long-lasting, comparatively inexpensive and robust. An IR (infrared) sensor is preferred for some applications including GMP (good manufacturing practice) monitoring, or when the incubator door is opened repeatedly over a short period; for example, when performing time lapse expression analyses. Some manufacturers have a design that places the sensors outside the incubation chamber to avoid the need to remove them during sterilization; this is not a good option because exterior sensors will not experience the same conditions as your cells do. Thus, these incubators will not respond quickly to changes that affect cell growth. CO 2 Gas Hookups For connecting CO 2 tanks to Thermo Scientific incubators, you must use a two-stage CO 2 pressure regulator on the outlet valve of the gas cylinder. This is because the input pressure on our incubators must be maintained at 15 +/- 5 psi (pounds per square inch) (1 barr) for the proper functioning of the CO 2 control system. A single stage regulator simply will not maintain this pressure, with resulting inaccurate CO 2 levels. We recommend using industrial grade gas that is at least 99.5% pure, because impurities in the gas can negatively affect your cells. The CO 2 gas tank should not contain siphon tubes. Use a 2-stage regulator on the tank, and set the input pressure to the incubator at 15 psi (1 barr). Connect the tubing to the air filter and then to the labeled serrated fitting on the back of the incubator, and be sure to use a hose clamp and to check for gas leaks, which could be dangerous to you and your lab mates. When the incubator has been operating normally for 24 hr, check the calibration of the temperature using a NIST or other certified thermometer, and check the CO 2 using a fyrite or external IR tester.
3 Adding Extra Equipment Some researchers want to add electrical equipment such as shakers, rotators or stirrers to the CO 2 incubator. The difficulty with this is that too much heat from added equipment raises the internal temperature in the incubator, and that extra heat makes it challenging for the incubator electronics to compensate. Shakers create vibration as well as heat, so it s important to not set the rotation faster than about 120 rpm (rotations per minute), to minimize motion that can affect adherent cell attachment and growth. If you know you will be using electrical equipment inside, consider purchasing a water jacketed incubator configured with a cooling coil and chiller that will compensate for the extra heat generated. 3 Technical Note: TNCO2CAREFEED 0514 As an alternative to a shaker, the Thermo Scientific Heracell 240i offers a roller bottle assembly, which does not produce heat. Each substitutes for a shelf, so you can add up to 4 of these units inside, giving you added culturing flexibility. Reducing Contamination To reduce chances of contamination in your cultures, the cleanliness of your lab is important. Dust and dirt can fly around in the lab, carried by air currents created when people move about in the room or open and close doors. Normal indoor room air contains 100-1,000 microorganisms per cubic meter 2, all circulating at any given moment. Most of this comes from the trillions of normal flora that live in and on our skin. This means that every time you open the incubator door, contaminants can enter. So it s important to minimize contaminants and dirt in the lab by cleaning the lab often; at least one to two times per month. Clean and disinfect the biological safety cabinet (BSC), the water bath, the centrifuge, and microscope. Eliminate cardboard storage in or around refrigerators and freezers because the cardboard can get wet and then breed fungi. Do not store items on top of the incubator, because dust and dirt among these items can be swept inside the chamber with air currents created during a door opening. Remember to also clean the corners of the lab and on top of and under equipment where dust can collect. Incubator Disinfectants Many researchers ask: which disinfectants are okay to use in the CO 2 incubator? The truth is that there are many disinfectant options available, but not all are safe for your cells. Some strong disinfectants may give off fumes that enter the incubator and then affect cell growth. These fumes contain VOCs (volatile organic chemicals) that can induce expression of heat shock and other stress proteins. Common laboratory chemicals such as phenol, isoamyl alcohol and betamercaptoethanol are VOCs, but also laboratory cleaning products and disinfectants, and even floor cleaners and waxes produce harmful vapors. In short, if it smells strong or bad to you, it can also be bad for your cultured cells. Recognizing that VOCs are an increasing concern for cell culturists, Thermo Scientific incubators with HEPA filtration will accept specialized HEPA filters that also capture VOCs. Thermo Fisher Scientific s Technology Team investigated and tested a number of different types of disinfectants. We looked for a disinfectant that fulfilled the following criteria: broadly effective against a range of microorganisms; and harmless (non-corrosive) to incubator components. Of all those tested, the compound that best fulfilled our requirements is a quaternary ammonium disinfectant. This basic type is available in all regions from different manufacturers. Some examples include Lysol No Rinse (formerly Roccall-D), Conflikt (North America) and Fermicidal-D (Europe). Use a 2% solution of the same quaternary ammonium disinfectant that you used to disinfect the incubator interior, as a disinfection additive in the water pan 3. We must stress that you should never use bleachcontaining cleaners. Chlorine bleach and its derivatives with oxidizing activity corrode stainless steel and copper. In addition, these chemicals are very toxic to your cells!
4 4 Procedure for Cleaning the Incubator Regular cleaning of the incubator, while sometimes a chore, is a necessity to help protect your cells from contamination and to keep the incubator functioning properly. Carefully following these simple steps will keep your incubator clean, help reduce chances of contamination becoming established, and help keep your cells growing well. If you use an automated heat decontamination cycle that is proven to eliminate bacteria and fungi, you should not need to use a disinfectant, but you should still periodically clean the incubator to eliminate spilled liquids and dirt. 1. Move all the cultures to a different incubator. Or if the cleaning will not take very long, you can store them in a clean, disinfected plastic box that will keep them warm. Then, turn off the incubator, including turning off the gas supply. 2. Remove all the shelves, the shelf supports, and any brackets or air ducts. Empty the water reservoir and wipe it dry with a clean, lint-free cloth. 3. Clean all the internal surfaces, ducts, shelves, shelf supports, inner door, fan and door gaskets with mild soapy water. A mild dish detergent works well for this. Be sure to reach all the corners and crevices where dirt, dust and germs can hide. An incubator with coved corners makes this task easier. 4. Rinse these surfaces and parts using distilled water and wipe them dry again using a clean, lint free cloth. 5. Wipe the interior surfaces and parts with a diluted quaternary ammonium disinfectant. Follow this by wiping with 70% alcohol to remove any remaining traces of the disinfectant. Again, be sure to reach all the corners, and remember to treat the door gasket as well. Replace the internal parts. 6. Now you can turn the incubator heat back on and allow the incubator to dry completely. This should only take a moment or two, so do not leave the door open that would only reintroduce new dust and contaminants! 7. If you have an automated decontamination cycle, you can run it now. After that cycle is complete, fill the water reservoir with sterile distilled water, and turn on the gas supply. 8. Remember to clean the incubator exterior to eliminate dirt and microorganisms that could find their way inside. If the top of your incubator is very dusty, it s likely that dust is passing into the incubator when you open the door, so it s important to clean the top as well. Use a lint-free cloth dampened in mild soapy water. Then wipe clean using a clean cloth slightly dampened in clear water. Dry the outside with a clean, dry cloth. Pay special attention to the door handles where everyone touches. Do not use any liquids or spray cleaners to clean a touch sensitive surface or display. Instead, use a dry microfiber cloth to clean these. That s it, you re done! When the incubator has reached the set conditions, return your cells inside. We are often asked how do I clean my 100% pure copper incubator? You can use the same procedure to clean copper that you would with stainless steel, including mild soapy water followed by a 70% ethanol wipe. It is okay to use a quaternary ammonium disinfectant, but it is not necessary due to copper s inherent properties. Importance of Correct Water To provide the proper humidity required by your cultured cells, we recommend only sterile distilled water. Tap water with even small amounts of chlorine can corrode stainless steel or pure copper. Also, tap water can contain lots of bacteria and minerals. Since salts and minerals are precisely balanced in cell growth media, adding minerals via the water to the humidified atmosphere can cause poor cell growth. DI (de-ionized) or ultra-pure Type 1 water is very aggressive. It corrodes stainless steel because the water, containing very few ions, actively pulls ions from the stainless steel, pure copper, glass door, and other incubator components. RO (reverse osmosis) water can vary tremendously in terms of quality because the purification is a percent-removal process. Thus, if the starting water has 500 ppm (parts per million), the finished water might be 50 ppm, but if the starting water has 150 ppm, the finished water would only have 15 ppm. Our Technology Team has done extensive testing on water quality. Based on these tests and for long life of the incubator, we recommend sterile distilled water with a ph of 7-9 and a conductivity of 1-20 microsiemens (resistivity of 50 K-1 M ohm). Even sterile distilled water can have a ph that is too low, so be sure to test this. If only DI or ultra-pure water is available in the lab, we recommend using water with a resistance range of 1-10 Mega ohms/cm. Ensure that the ph is 7-9, and then sterilize the water before using it in the incubator. Note that sometimes your building source of Type 1, DI, or ultra-pure water actually came from a still to start with, so as long as the parameters we have laid out are met, this water would be acceptable. If you do not have access to distilled water, one option for using Type 1, DI or ultrapure water in the incubator is to add a little sodium bicarbonate to the water to raise the ph and to add some ions. But it must be a sterile solution of the salt, and you must ensure the final ph of the water is in the 7-9 range. Thermo Scientific Barnstead Classic Still
5 CO 2 Calibration You can calibrate your own CO 2, but there are a couple points you should know. If you have a T/C sensor, be sure to do any calibration first thing in the morning, because this provides at least 12 hours of uninterrupted equilibration to ensure stable temperature and humidity. Since the T/C measurement works in conjunction with temperature and humidity, it is important to keep the water pan full. If the water runs out, that will likely affect the T/C sensor calibration. If you have a Heracell incubator, the AutoStart cycle will automatically zero the sensors. In addition, the Heracell incubator will automatically compensate for pressure changes that can affect the concentration of the gases inside the chamber. In case you have a tri-gas incubator with variable oxygen control, you must calibrate the O 2 first and let it stabilize before calibrating the CO 2. This is because changing the oxygen will change the CO 2 calibration. Probably the most important thing you can do to maintain your incubator is to maintain a high level of water in the water pan. If the pan runs out of water, it is very bad for your cell cultures because when the water pan dries out, the humidity drops and that leads to evaporation of water from your culture medium. Thus, without proper humidity, the carefully balanced salts, minerals, amino acids, etc. in your growth medium become too concentrated, which can lead to toxicity and cell death 4. But beyond the danger to your cells, too low humidity can also damage the CO 2 sensor, as explained earlier. 5 Technical Note: TNCO2CAREFEED 0514 How often should you calibrate? You can calibrate the CO 2 once per month, using fyrite or an external IR tester. However, if you properly maintain the incubator and keep the water pan full, you should only need to calibrate a few times per year, for example, once per quarter. And remember, if you have any concerns, you can contact Thermo Fisher Scientific s factory trained service specialists. General Maintenance Ongoing maintenance of your incubator, other than an occasional CO 2 check and calibration, is very minor. If your incubator is equipped with HEPA filtration, you should replace the HEPA filter between 6 months to 1 year. The life of a filter will depend on the number of people using the incubator and the general cleanliness of it and the laboratory, as well as the manufacturer s design. You should also replace the gas inlet filters (where the CO 2 gas enters the incubator) between 6 months to 1 year, and simply follow the manufacturer s recommendations for that procedure, or schedule a qualified service specialist to do it for you. How often should you run the heat sterilization cycle? The answer depends on the cleanliness of your lab generally, how many people use the same incubator, how often the door is opened and how convenient it is for you to shut down an incubator overnight. Most users use this function from one time per month to once every six months.
6 References: 1. Thermo Scientific. Importance of Class 100 Air in a CO 2 Incubator. Thermo Scientific 2009, WPCO2Class United States Department of Labor, Occupational Safety & Health Administration. OSHA Technical Manual (OTM) Section III: Chapter 2. TED , Freshney, RI. Culture of Animal Cells. Hoboken, NJ: John Wiley & Sons, Inc., Esser, P and Weitzmann, L. Evaporation From Cell Culture Plates. Thermo Scientific 2011, TILSPNUNCBU Technical Note: xxxx 0414 thermoscientific.com/co Thermo Fisher Scientific Inc. All rights reserved. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Australia Austria Belgium China or France Germany national toll free Germany international India toll free India Italy Japan Netherlands New Zealand Nordic/Baltic/CIS countries Russia Spain/Portugal Switzerland UK/Ireland USA/Canada Other Asian countries Countries not listed TNCO2CAREFEED 0514
Thermo Scientific Rotor Care Guide. Protect your rotor investment
Thermo Scientific Rotor Care Guide Protect your rotor investment Thermo Scientific Rotor Care Guide Thermo Scientific rotors meet the high standards we set for all of our laboratory products. The precision-engineered
Procedures for the Effective Use of Biological Safety Cabinets
Procedures for the Effective Use of Biological Safety Cabinets APPROVAL DATE BY BIOHAZARDS SUBCOMMITTEE: August 10, 2012 SUPERSEDES: March 1, 2012 This corporate guideline/procedure is intended as a minimum
Thermo Scientific Barnstead TII Pure Water Systems. Pure water. Pure research. Total confidence.
Thermo Scientific Barnstead TII Pure Water Systems Pure water. Pure research. Total confidence. Thermo Scientific Barnstead TII Pure Water Systems Ideal for any daily application needs for pure water,
Multi Tube Rotator. Model No. 4632Q 4632-1CE 4632-1CECN 057-213-00 11/05/11
Multi Tube Rotator Model No. 4632Q 4632-1CE 4632-1CECN 057-213-00 11/05/11 Table of Contents Safety Information...3 Alert Signals...3 Description...4 Specifications...5 Power Requirements...5 Shaker Motion...5
Facility Requirements for the elan2 Office Liquid Nitrogen Generator
Facility Requirements for the elan2 Office Liquid Nitrogen Generator Overview In order to ensure the proper operation of the elan2 Liquid Nitrogen Generator system so that the maximum production of liquid
Welcome to chapter 10. The following chapter is called Equipment maintenance and monitoring. The author is Ronny Janssens.
Welcome to chapter 10. The following chapter is called Equipment maintenance and monitoring. The author is Ronny Janssens. 1 After finishing this chapter, the student should be familiar with - the requirements
Thermo Scientific Ultra-Low Temperature TSX Series Freezers
Thermo Scientific Ultra-Low Temperature TSX Series Freezers High sample security, low environmental impact White Paper Key Words Cold storage, sustainability, sample security, sample protection, low environmental
CLEANING AND SANITIZING
CLEANING AND SANITIZING Cleaning and sanitizing procedures must be part of the standard operating procedures that make up your food safety program. Improperly cleaned and sanitized surfaces allow harmful
Smart. Scalable. Simple.
Thermo Scientific Wireless Monitoring Solution Smart. Scalable. Simple. providing the protection you need with the simplicity you deserve Thermo Scientific Wireless Monitoring Solution Optimized Sample
Thermo Scientific Biological Safety Cabinets Fumigation Methodologies
Thermo Scientific Biological Safety Cabinets Fumigation Methodologies Due to its use in hazardous processes, the biological safety cabinet (BSC) is a highly regulated device. Generally speaking, regulatory
Decontamination and Waste Management www.biosecurity.sandia.gov
Decontamination and Waste Management www.biosecurity.sandia.gov SAND No. 2006-3684C Sandia is a multiprogram laboratory operated by Sandia Corporation, a Lockheed Martin Company, for the United States
Dehumidifier Users manual. For Models: DH45S DH65S
Dehumidifier Users manual For Models: DH45S DH65S 950-0062-revD Jan. 9 2007 FORWARD The appearance of the units that you purchase might be slightly different from the ones described in the Manual, but
3. Maintenance of Laboratory Equipment
- 52 - Mainten: of Lab Equipment / National Guidelines SLCM SLCM National Guidelines / Mainte: of Lab Equipment - 53-3. Maintenance of Laboratory Equipment Introduction Equipment maintenance is one important
Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance
Bench Autoclave Standard Operating Procedure For Installation, Use and Maintenance 1. Introduction This SOP is intended for use with the following model, in a laboratory context: Type: Nuve Bench Top Steam
OASIS-PLUS 120V READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE
OASIS-PLUS 120V INFORMATION & OPERATING INSTRUCTIONS READ ALL INSTRUCTIONS BEFORE OPERATING READ ALL INSTRUCTIONS BEFORE OPERATING OZONE IS A POWERFUL OXIDIZER AND MUST BE USED WITH CARE 56041852 WARNING:
Vertical Display and Storage B1350-2. SKOPE Gen2: Three Door Chiller
Vertical Display and Storage User Manual MAN1227 Rev. 3.0 March 2008 edition CONTACT ADDRESSES Designed and Manufactured by New Zealand SKOPE INDUSTRIES LIMITED PO Box 1091, Christchurch New Zealand Freephone:
Cal Water 1961 Petra Lane, Placentia, CA 92870 (800) CAL-WATER (800) 225-9283 FAX: (714) 792-0794 http://www.cal-water.com
Industrial Water Purification MAINTAINING A STERILE HIGH PURITY WATER SYSTEM Introduction Maintaining a microbe free deionized water system is like trying to maintain a vacuum, and nature abhors a vacuum.
PREVENTATIVE MAINTENANCE
PREVENTATIVE MAINTENANCE PREVENTIVE MAINTENANCE Sandwich Unit: Proper Operation Turn on at least 45 minutes before loading Load only pre-chilled food 34-37 F Move quickly from walk-in to sandwich unit
Lab Exercise 3: Media, incubation, and aseptic technique
Lab Exercise 3: Media, incubation, and aseptic technique Objectives 1. Compare the different types of media. 2. Describe the different formats of media, plate, tube etc. 3. Explain how to sterilize it,
SAFE HEALTHY CHILDREN A Health and Safety Manual for Childcare Providers
TABLE OF CONTENTS Chapter 7: Food Storage and Preparation Safe Food: Food Storage and Preparation 7.1 Requirements of the Day Nurseries Act... 7.2 Recommendations of the Medical Officer of Health... 7.2
ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL
1 of 8 The quality assurance plan as presented in the Quality Manual should assure that constant and consistent test conditions are met and verified and should be tailored to the laboratory s activities.
DESERT RESEARCH INSTITUTE MOLD ASSESSMENT AND REMEDIATEION PROCEDURE
1.0 Introduction DESERT RESEARCH INSTITUTE This Desert Research Institute (DRI) Mold Assessment and Remediation Procedure was established to assure that consistent and uniform information is provided to
Thermo Scientific Forma Steri-Cycle CO2 Incubators Total contamination control and elimination
Thermo Scientific Forma Steri-Cycle CO2 Incubators Total contamination control and elimination Part of Thermo Fisher Scientific Thermo Scientific Forma Steri-Cycle CO 2 Incubators Contamination control
Protocol for Disinfection of Cell Culture and Tissue Culture in Media:
Protocol for Disinfection of Cell Culture and Tissue Culture in Media: Location: Hickory Hall 001 Director: Dr. Guido Verbeck DECONTAMINATION OF CELL CULTURE WASTE Cell culture has become a common laboratory
North Carolina State University Emergency Facilities Closure Checklist- Part I
North Carolina State University Emergency Facilities Closure Checklist- Part I Unplanned or spontaneous events often disrupt daily operations on campus. In the event that an incident may interrupt your
Thermo Scientific Heratherm Microbiological Incubators. Safe, easy and efficient
Microbiological Incubators Safe, easy and efficient Microbiological Incubators your samples We are obsessed with sample integrity. Each of our new laboratory microbiological incubators is designed with
SITE REQUIREMENTS AND INSTALLATION INSTRUCTIONS
ADVANTAGE PLUS SITE REQUIREMENTS SITE REQUIREMENTS AND INSTALLATION INSTRUCTIONS 50096-927 Rev A For additional information contact your Medivators Pg 1 of 18 SITE REQUIREMENTS The purpose of this section
RBS detergents for laboratories
RBS detergents for laboratories High performance solutions For reliable and reproducible results Competence & experience RBS products are mainly intended for water-based cleaning processes of glassware
Incubating Eggs. Percent Hatchability is the percentage of fertile eggs which actually hatch out as live young.
Incubating Eggs Phillip J. Clauer, Poultry Extension Specialist, Animal and Poultry Sciences Introduction Many domestic bird owners incubate eggs to help sustain their flock over time. This fact sheet
Cleaning. By the end of this chapter, you will be able to: Introduction. Definitions. Chapter 9
Chapter 9 By the end of this chapter, you will be able to: l Define the terms cleaning, decontamination, disinfection l List the factors that affect disinfection l Describe the aims of a cleaning schedule
Portable Air Conditioner
Portable Air Conditioner Owner's Manual Model:3 in 1 12,000 Btu/h Series 3 Please read this owner s manual carefully before operation and retain it for future reference. CONTENTS 1. SUMMARY...1 2. PORTABLE
COMMON LABORATORY APPARATUS
COMMON LABORATORY APPARATUS Beakers are useful as a reaction container or to hold liquid or solid samples. They are also used to catch liquids from titrations and filtrates from filtering operations. Bunsen
Environmental Monitoring of Clean Rooms
Environmental Monitoring of Raul Duarte President DDK Scientific, Corp. Copyright DDK Scientific, Corp. 2008, 2009, DDK Scientific, Corp. Proprietary A manufacturing facility for pharmaceutical products
Thermo Scientific Forma Steri-Cult CO 2 Incubators Ultimate contamination control for ideal culturing conditions
Thermo Scientific Forma Steri-Cult CO 2 Incubators Ultimate contamination control for ideal culturing conditions 2 The innovative Thermo Scientific Forma Steri-Cult brings leading edge technology into
USER S MANUAL FH052EAV1 FH070EAV1. System Air Conditioner (Cooling and Heating) ENGLISH ESPAÑOL FRANÇAIS ITALIANO PORTUGUÊS DEUTSCH E HNIKA
USER S MANUAL FH052EAV1 FH070EAV1 E HNIKA PORTUGUÊS ENGLISH ESPAÑOL ITALIANO DEUTSCH FRANÇAIS System Air Conditioner (Cooling and Heating) E S F I P D G DB98-29263A(1) Safety Precautions Register your
Bathroom Installation Guide
Bathroom Installation Guide Step-by-step installation of your cabinets We do a lot of groundwork to make it as simple as possible for you to assemble and install your new bathroom furniture yourself. We
Thermo Scientific Rotor Care Guide. Protect your rotor investment
Thermo Scientific Rotor Care Guide Protect your rotor investment pg 2 pg 4 pg 6 pg 8 Routine Evaluation and Care of Your Rotor Rotor Maintenance Rotor Inspection Services Rotor Life Management Thermo Scientific
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration
Thermo Scientific Nunc Products for InVitro Fertilization. security, safety and reproducibility. for your valuable. samples
Thermo Scientific Nunc Products for InVitro Fertilization security, safety and reproducibility for your valuable samples The criteria for an IVF product is that it is reliable and does not offer any threat
Campus Services Hints and tips for residents
Campus Services Hints and tips for residents We have compiled this webpage to help University residents with simple tips and hints about cleaning kitchen areas - which are residents responsibility. We
Thermo Scientific VARIOKLAV Modular Laboratory Autoclaves Versatile, space-saving performance for steam sterilization applications
Thermo Scientific VARIOKLAV Modular Laboratory Autoclaves Versatile, space-saving performance for steam sterilization applications Part of Thermo Fisher Scientific Thermo Scientific Varioklav Laboratory
Service Guide. Evaporator Coil Cleaning and Condensate Drain Maintenance Guidelines. ORDER No: UN-SVG001A-EN DATE: October 9, 2007
Service Guide Evaporator Coil Cleaning and Condensate Drain Maintenance Guidelines ORDER No: UN-SVG001A-EN DATE: October 9, 2007 Introduction This document defines American Standard and Trane s recommendations
ph Measurements of Common Substances
Chem 100 Section Experiment 10 Name Partner s Name Introduction ph Measurements of Common Substances The concentration of an acid or base is frequently expressed as ph. Historically, ph stands for the
IX. Decontamination and Spills
IX. Decontamination and Spills IX.1. Definitions Sterilization: the act or process, physical or chemical, which destroys or eliminates all forms of life, especially microorganisms. Decontamination: reduction
CelCulture. CelCulture CO2 Incubator Model CCL-170_-_. CO2 Incubator with Integrated Cooling System Solution for Highly Specialized Applications
1 CO2 Incubator Model CCL-170_-_ CO2 Incubator with Integrated Cooling System Solution for Highly Specialized Applications Welcome to Esco Esco s Vision is to provide enabling technologies for scientific
PACUC Guidelines. Rodent Survival Surgery
PACUC Guidelines Rodent Survival Surgery Guidelines for rodent survival surgery are based on the National Research Council Guide for the Care and Use of Laboratory Animals 8 th edition and the Animal Welfare
NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS
NewAir AC-10100E / AC-10100H Portable Air Conditioner Owner s Manual PLEASE READ AND SAVE THESE INSTRUCTIONS ELECTRICAL SAFETY This appliance is for indoor use only. Always turn off the unit and unplug
GTS. New! Plus: PR0DUCT CATALOG. Vapor-logic 4 controller with: and LonTalk. New models GTS-500 and GTS-700. Gas-to-Steam Humidification System
GTS Gas-to-Steam Humidification System PR0DUCT CATALOG New! Vapor-logic 4 controller with: and LonTalk Plus: New models GTS-500 and GTS-700 Gas has lowest energy operating costs The industry s first and
CARE HOMES AND NURSING HOMES
Care Homes and Nursing Homes: This document provides guidelines and recommendations for cleaning care homes and nursing homes in the event of a disease outbreak in the vicinity. Cleaning and disinfection
BIOLOGICAL SAFETY CABINET PROGRAM
BIOLOGICAL SAFETY CABINET PROGRAM TABLE OF CONTENTS: 1. INTRODUCTION 2. PURPOSE AND SCOPE 3. SELECTION AND INSTALLATION 4. USE OF BIOLOGICAL SAFETY CABINETS 5. CERTIFICATION OF BIOLOGICAL SAFETY CABINETS
Laboratory & Scientific Line. Bulk Steam Sterilizers. Since
Laboratory & Scientific Line Bulk Steam Sterilizers Since 1925 Bulk Steam Sterilizers for Laboratory & Scientific Applications The Tuttnauer line of bulk autoclaves are designed to handle the most critical
RULE 1102. DRY CLEANERS USING SOLVENT OTHER THAN PERCHLOROETHYLENE
(Adopted January 6, 1978)(Amended August 3, 1979)(Amended July 11, 1980) (Amended August 3, 1990)(Amended December 7, 1990) (Amended November 17, 2000) RULE 1102. DRY CLEANERS USING SOLVENT OTHER THAN
National Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook
National Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook Trainee:... Training Provider:... Year:... Trainee Contact no:... Competency Based Skill Training Logbook Introduction
FACT SHEET : Using Autoclaves Safely
CSULA Environmental Health and Safety Biosafety Office FACT SHEET : Using Autoclaves Safely Most science research laboratories on campus require the use of autoclaves. The primary purpose of the autoclave
Federal Wage System Job Grading Standards for Air Conditioning Equipment Operating, 5415. Table of Contents
Federal Wage System Job Grading Standards for Air Conditioning Equipment Operating, 5415 Table of Contents WORK COVERED... 2 WORK NOT COVERED...2 TITLES... 2 GRADE LEVELS... 2 HELPER AND INTERMEDIATE JOBS...
Thermo Scientific Guide for Vaccine Storage. Four steps for. safe and secure vaccine storage
4 Thermo Scientific Guide for Vaccine Storage Four steps for safe and secure vaccine storage Selecting the right cold storage equipment is a critical decision for vaccine storage Refrigeration problems
STERILIZATION AND DISINFECTION
STERILIZATION AND DISINFECTION Importance of hand washing shown by Semmelweis STERILIZATION A physical or chemical process that destroys or eliminates all forms of microbial life, including spores. A satisfactory
Biological Safety Program
Risk Management & Safety Main Office, Wyoming Hall Phone: (307) 766-3277 Fax: (307)766-6116 Regulated Materials Management Center Phone: (307)766-3696 Fax: (307)766-3699 Web: www.uwyo.edu/ehs Email: [email protected]
GENANO 4500 MEDICAL AIR PURIFICATION DEVICE
GENANO AIR PURIFICATION TECHNOLOGY GENANO 4500 MEDICAL AIR PURIFICATION DEVICE OPERATING INSTRUCTION 1 GENANO LTD Kimmeltie 3 FI- 02110 Espoo, Finland Tel + 358 9 7743 870 Fax + 358 9 7533 130 www.genano.fi
SPILLS & SPILL KITS. Spills -General Guidelines:
Spills -General Guidelines: If handled properly, a spill may be nothing more than a nuisance. If handled improperly, a spill can seriously disrupt your activities and the work of your colleagues. At worst,
EQUIPMENT CALIBRATION/MAINTENANCE
EQUIPMENT CALIBRATION/MAINTENANCE The following chart provides a guide for calibration of standard equipment used in the Fish Health Laboratory. All Intervals suggested are subject to individual manufacturer
CHILDCARE SETTINGS. Childcare Settings
Childcare Settings This document provides guidelines and recommendations for cleaning childcare settings such as nurseries, crèches and schools in the event of a disease outbreak in the vicinity. Cleaning
Nu G Medical Waste System Technology (Pyrolysis / Thermal Decomposition)
Product Description: Nu G Medical Waste System Technology (Pyrolysis / Thermal Decomposition) The NU G System uses pyrolysis thermal decomposition to treat infectious wastes typically generated in hospitals.
Meaco 30L and Meaco 40L dehumidifier instruction manual
Meaco 30L and Meaco 40L dehumidifier instruction manual Please read this instruction manual before using the dehumidifier and keep safe for future reference SAFETY INSTRUCTIONS PLEASE READ ALL INSTRUCTIONS
Incubator with mechanical convection
BF series 400 Incubators and cooling incubators Incubator with mechanical convection The BINDER incubator of the BF series with mechanical convection is suitable for all gentle incubation applications,
Chemical Fume Hoods. Environmental Health and Safety 974-5084
Chemical Fume Hoods Environmental Health and Safety 974-5084 Regulatory References Prudent Practices in the Laboratory 2011 (National Research Council of The National Academies): Laboratory chemical hoods
Lab Safety and Standard Operating Procedures. Faculty of Dentistry And School of Biomedical Engineering
Lab Safety and Standard Operating Procedures Faculty of Dentistry And School of Biomedical Engineering Introduction It is the requirement that students working in research laboratories at Dalhousie have
Please read this owner s Manual carefully before operating the unit. - Cooling - Heating - Dehumidifying - Fan
Please read this owner s Manual carefully before operating the unit. - Cooling - Heating - Dehumidifying - Fan TABLE OF CONTENTS INTRODUCTION 2 IMPORTANT SAFEGUARDS...2 PACKAGE CONTAINS..2 NAMES OF PARTS.3
USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102
USER INSTRUCTIONS FOR 10 LITRE PORTABLE DEHUMIDIFIER MODEL NO. DHMD102 THANK YOU FOR CHOOSING YOUR NEW DEHUMIDIFIER. BEFORE USING THE UNIT READ THESE INSTRUCTIONS FULLY AND RETAIN THEM FOR FUTURE REFERENCE
Dew Point Tester. Instruction Manual. CVS Regular Chiller Model A-2. CVS Regular Chiller Model A-2
Instruction Manual Dew Point Tester The CVS Dew Point Tester is a chilled mirror apparatus, which operates by attaining the conditions necessary by the definition of dew point. When a gas is cooled out
GLYCOL CHILLER SYSTEM TROUBLESHOOTING 101
GLYCOL CHILLER SYSTEM TROUBLESHOOTING 101 By Jim VanderGiessen Jr, Pro Refrigeration Inc. There is a good chance your glycol chiller system is often completely forgotten about, until that dreaded day you
Sample Pharmacy CLEANING AND SANITIZING PROCEDURE P-304.1
P-304.1 1.0 Definition and Purpose: To standardize and define the process by which the controlled Cleanroom environments (ISO Class 5, 7, and 8) and the general pharmacy prep area is cleaned, disinfected,
BOSTON COLLEGE ENVIRONMENTAL HEALTH AND SAFETY LABORATORY GUIDE FOR RENOVATIONS, REMODELS, MOVES AND TERMINATIONS
BOSTON COLLEGE ENVIRONMENTAL HEALTH AND SAFETY LABORATORY GUIDE FOR RENOVATIONS, REMODELS, MOVES AND TERMINATIONS I. Policy for Termination of Laboratory Use of Hazardous Materials The Principal Investigator
Treatment options for hydrogen sulfide. Testing for hydrogen sulfide
Sometimes hot water will have a sour smell, similar to that of an old damp rag. This smell often develops when the thermostat has been lowered to save energy or reduce the potential for scalding. Odor-causing
Instruction Manual. Image of SP-3015 & SP-3815. Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function
Important Safeguards READ ALL INSTRUCTIONS BEFORE USE. Instruction Manual Automatic Dispensing Hot Water Pot with Reboil Function Image of SP-3015 & SP-3815 SP-3015: 3.0L SP-3815: 3.8L SP-3017: 3.0L (Stainless
THE UNIVERSITY OF NEWCASTLE- SCHOOL of BIOMEDICAL SCIENCES
Page: 1 of 7 1. Purpose: 1.1. To describe the procedures to be used when dealing with chemical or microbiological spills. 2. Equipment: 2.1. Spill Kit 2.2. Miscellaneous items as listed 3. Materials: 3.1.
Wastewater Reuse. Typical treated wastewater is:
Wastewater Reuse Most metal finishing industries have in-house wastewater treatment to economically dispose of the acids, alkali, oils, and dissolved metals in the rinse water and occasional tank solution
Troubleshooting Your Refrigerator
Troubleshooting Your Refrigerator With conscientious care, a refrigerator will perform for years with little or no repairs. Refrigerators are categorized according to the way frost is removed from them.
2 (. 3 4 5 . -,.. -,. 3. : 1 2 3 > 4, :!! 1,, 17 Lt MAX 2 3 4,, > 5.,,, 6 :, :1)! 2)! 7,, 1, :,. 2,, Bottom of steam generator 7 8 9 10 11 12 automatically and this phenomenon will continue for
User s Guide. Oreck Air Purifier with HEPA Filtration. Important! Read this manual carefully, and keep for future reference.
User s Guide Oreck Air Purifier with HEPA Filtration Important! Read this manual carefully, and keep for future reference. Enjoy Congratulations on your purchase of the Oreck Air Purifier with HEPA Filtration.
Instruction Manual. Window Wiper AFW900. (Read carefully before use)
Instruction Manual Window Wiper AFW900 (Read carefully before use) When using electrical appliances, basic precautions should always be followed, including the fact READ ALL INSTRUCTIONS IN THIS MANUAL
SWIMMING POOL HEAT PUMP
SWIMMING POOL HEAT PUMP Installation & User Manual Model HP40B HP50B HP65B Hayward Pool Products Canada, Inc. T: 1-888-238-7665 www.haywardpool.ca CONTENT I. Application 4 II. Features 4 III. Technical
CO 2 -Incubator. Professional Cell Culture CO 2 incubator with UV sterilisation. Contamination-Controlled. Downtime-None.
Professional Cell Culture CO 2 incubator with UV sterilisation. Contamination-Controlled. Downtime-None. The new Automatic CO 2 Incubator provides a stable cell culture environment where contamination
FORMATION DAMAGE AND WELL TREATMENT SYSTEM
FORMATION DAMAGE AND WELL TREATMENT SYSTEM FDWTS-10K Includes ability to test Liquid Permeability & Gas Permeability Fines migration & critical velocity Static Filtration Drilling mud invasion evaluation
SPACE TYPE: LABORATORY (WET) FOR CHEMISTRY AND BIOLOGY
SPACE TYPE: LABORATORY (WET) FOR CHEMISTRY AND BIOLOGY Construction Criteria The unit costs for (Wet) space types are based on the construction quality and design features in the following table. This
Assembly and Installation Procedures
Assembly and Installation Procedures for Pall Pharmaceutical Grade Capsule Assemblies 1. Introduction The following procedures must be followed for the installation of Pall pharmaceutical grade capsule
Steam Sterilization Cycles for Lab Applications
Steam Sterilization Cycles for Lab Applications Presented by Gary Butler STERIS Life Sciences August 2009 Early Steam Sterilizers Koch Upright Sterilizer First Pressurized Sterilizer First built by Pasteur-Chamberlain
Stenner Pump Chlorinator Installation & Start Up Guide
Clean Water Made Easy www.cleanwaterstore.com Stenner Pump Chlorinator Installation & Start Up Guide Thank you for purchasing a Clean Water System! With proper installation and a little routine maintenance
UB1 AIR CONDITIONING UNIT INSTALLATION INSTRUCTIONS
UB1 AIR CONDITIONING UNIT INSTALLATION INSTRUCTIONS INSTALLATION INSTRUCTIONS: Carefully read these instructions before installing your new air-conditioner. AUSTRALIAN AUTOMOTIVE AIR AL00500054E 1 Table
Safety information...3 Important Information...3 Alert Signals...3 Introduction...4 General Usage...4 Principles of Operation...4 Installation...
9/3/10 Table of Contents Safety information...3 Important Information...3 Alert Signals...3 Introduction...4 General Usage...4 Principles of Operation...4 Installation...4 General Specifications...5 Environmental
Mechanical Systems Competency 1.20
Competency 1.20 Mechanical systems personnel shall demonstrate a working level knowledge of the safety and health fundamentals of mechanical systems and/or components. 1. Supporting Knowledge and Skills
Rusty Walker, Corporate Trainer Hill PHOENIX
Refrigeration 101 Rusty Walker, Corporate Trainer Hill PHOENIX Compressor Basic Refrigeration Cycle Evaporator Condenser / Receiver Expansion Device Vapor Compression Cycle Cooling by the removal of heat
Pall in the Brewery (more value per hectolitre) DE Free Clarification FB1795
Pall in the Brewery (more value per hectolitre) DE Free Clarification FB1795 The Pall World Pall Corporation is a world leader in the design, production and supply of filters, membranes and systems for
Coating and Extraction of Honeycomb Denuders
QSM Approval: Coating and Extraction of Honeycomb Denuders 1 INTRODUCTION 1.1 The following procedures are used for the coating of honeycomb denuders with citric acid and sodium carbonate solutions, and
1-888-448-7250 GENERAL CARE PLUMBING & VALVE INFORMATION TUB & SHOWER DOOR INFORMATION FOR QUESTIONS CALL OUR SERVICE DEPARTMENT
GENERAL CARE PLUMBING & VALVE INFORMATION TUB & SHOWER DOOR INFORMATION FOR QUESTIONS CALL OUR SERVICE DEPARTMENT 1-888-448-7250 or email [email protected] caring for your new bath fitter unit Thank
NITROUS TRANSFER PUMP INSTRUCTIONS
NITROUS TRANSFER PUMP INSTRUCTIONS SAFETY TIPS Never directly inhale nitrous oxide. When inhaled in large quantities, nitrous oxide can cause respiratory ailments or in extreme cases, death by suffocation.
Thermo Scientific Heratherm Microbiological Incubators. Safe,
Thermo Scientific Heratherm Microbiological Incubators Safe, easy and efficient Thermo Scientific Heratherm Microbiological Incubators your samples We are obsessed with sample integrity. Each of our new
Instruction manual. VWR unstirred water baths
Instruction manual VWR unstirred water baths Introduction VWR laboratory thermostatic baths provide the means to maintain samples at a precise temperature from just above room temperature up to 99 0 C.