Fluorescence Compensation. Jennifer Wilshire, PhD

Size: px
Start display at page:

Download "Fluorescence Compensation. Jennifer Wilshire, PhD jennifer@flowjo.com"

Transcription

1 Fluorescence Compensation Jennifer Wilshire, PhD

2 Outline 1. Compensation Basics 2. Practical Applications -selecting comp controls (cells, stains?) 3. Multicolor Issues -spreading due to measurement error -Bi-Exponential scaling -FMO controls

3 Acknowledgements Mario Roederer - Vaccine Research Center, NIH Alan Saluk - Scripps Research Institute Carleton Stewart - Roswell Park Cancer Institute Sue DeMaggio - FloCyte Training Course Purdue Cytometry Resource

4 Importance of ACCURATE Compensation APC CD45 FL4 --> Uncompensated Compensated Over Compensated n d b n d b FL4 --> FL4 --> n b Where is the CD8 dull population?! FL3-Height --> FL3-Height --> PE-CY5-CD FL3-Height --> RPCI LFC n = negatives d = dim positives b = bright positives

5 FITC Single Stain Control FITC PE Argon Laser FL1 FL

6 Uncompensated FITC single stain control PE - no stain FITC CD3

7 Uncompensated FITC Single stain Control Unwanted signal detected in FL2 roughly 15% PE - no stain FITC CD3 Total signal detected in FL1

8 FITC Single Stain Control FITC PE Total signal detected in FL1 Unwanted signal detected in FL2 = roughly 15% FL1 FL2 True PE = Total FL2 15% FL1

9 FITC Single Stain Control Uncompensated Compensated PE - no stain FL2-15%FL1 PE - no stain FITC CD3 FITC CD3

10 Part 2: Compensation in Practice Compensation Controls - Single Stains -which cells? -which stains?

11 Compensation Controls Single Stain Controls

12 Single Stain Controls - Which cells? Does not matter as long as: The autofluorescence is the same in the negative and positive populations you are lining up. eg, Pre-gate on lymphocytes if you are using CD8 FITC as a single stain control The compensation values will be valid for ALL cell types, regardless of which type of cell is used to calculate the values. The compensation is specific for the fluorochrome, not the cell type

13 Single Stain Controls - which reagents? Use the same reagent (Ab-fluorochrome conjugate) as used in the experimental sample OR A different antibody may be substituted, as long as it is conjugated to the same fluorochrome. However

14 Single Stain Controls - which reagents? Caveats for substituting reagents: With tandem dyes (Cy5PE/Cy7PE etc.) it is necessary to use the exact same reagent spillover varies from reagent to reagent GFP, CFSE, and FITC are NOT the same fluorochrome even though they are all green! Controls should be as bright as possible As bright or brighter than the experimental stains

15 Using Beads to Compensate Antibody-capture beads Use reagent in use Lots positive Small CV, bright Some reagents won t work (IgL, EMA/PI) mix with regular comps

16 Part 3: Multicolor Issues Spreading due to measurement error -why do these populations look funny? Bi-Exponential scaling -new display options FMO controls -determine where to set your gates

17 Spreading due to Measurement Error Why do these populations look funny?

18 Multicolor Compensation Uncompensated Compensated 10 5 Lymphocytes 10 5 Lymphocytes Cy7PE-A: CD <Cy7PE-A>: CD PE-A: CD <PE-A>: CD8

19 The Actual Spread 10 5 Lymphocytes <Cy7PE-A>: CD <PE-A>: CD8

20 Bi-Exponential Scaling New Display Options

21 Bi-Exponential Scaling Compensated with New Display 10 5 Lymphocytes <Cy7PE-A>: CD <PE-A>: CD8

22 Gating controls FMO (Fluorescence Minus One)

23 Gating Controls Unstained cells or complete isotype control stains are improper controls for determining positive vs. negative expression in multi-color experiments. The best control is to stain cells with all reagents except the one of interest. FMO Control Fluorescence Minus One

24 FMO Controls FMO controls are necessary to identify cells which do or do not express a given antigen. FMO controls should be used whenever accurate discrimination is essential or when antigen expression is relatively low

25 Identifying CD4 cells with 3 colors -PBMC were stained as shown in a 3-color experiment. -Compensation was properly set for all spillovers FITC PE Cy5PE Unstained Control FMO Control Fully Stained CD3 CD8 CD3 CD4 CD PE 10 3 Unstained Bounds FMO Bounds FITC

26 3 Color Experimental Setup Example staining setup of a 3 color experiment: Tube # Description FL1 FL2 FL3 1 Experimental Sample CD3 FITC CD4 PE CD8 Cy5PE 2 Compensation Controls CD3 FITC (Single stains one for each - CD4 PE - fluorochrome used in the 4 experiment) - - CD8 Cy5PE 5 Gating Controls 6 (FMO leave out one CD3 FITC - CD8 Cy5PE fluorochrome at a time) 7 CD3 FITC CD4 PE - 8 Experimental Controls CD3 FITC CD4 PE CD8 Cy5PE 9 (fully stain healthy or CD3 FITC CD4 PE CD8 Cy5PE 10 CD3 FITC CD4 PE CD8 Cy5PE untreated samples to compare to experimental sample) * no stain added or add isotype matched control stain.

27 New Quad-Gates Log Fluor. #2 Single Positive Double Positive Single Positive Doub Positi Single Positive Log Fluorescence #1 P

28 Take Away Lessons Proper CONTROLS are essential DON T compensate by eye Use Median to adjust the populations if you must do it manually New DISPLAYS help view data intuitively

Compensation Controls Data Visualization Panel Development Gating Quantum Dots

Compensation Controls Data Visualization Panel Development Gating Quantum Dots Compensation Controls Data Visualization Panel Development Gating Quantum Dots FITC Single Stain Control FITC PE Argon Laser FL1 FL2 450 500 550 600 FITC Compensation Control Unwanted signal detected in

More information

Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer

Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Presented by Clare Rogers, MS Senior Marketing Applications Specialist BD Biosciences 23-13660-00 Webinar Overview Multicolor

More information

Flow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE

Flow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE Flow Data Analysis Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA Flow Data Analysis From Cells to Data - data measurement and standard Data display and gating - Data Display options -

More information

Compensation in Flow Cytometry

Compensation in Flow Cytometry Compensation in UNIT.4 The term compensation, as it applies to flow cytometric analysis, refers to the process of correcting for fluorescence spillover, i.e., removing the signal of any given fluorochrome

More information

Three Rules for Compensation Controls

Three Rules for Compensation Controls From FlowJo s Daily Dongle Three Rules for Compensation Controls First and foremost, there must be a single-stained control for every parameter in the experiment! In addition, there are three rules for

More information

Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers

Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers BD Biosciences, San Jose, CA Contents 2 Introduction to compensation 5 Tips for setting up compensation 10 Tools for easy,

More information

COMPENSATION MIT Flow Cytometry Core Facility

COMPENSATION MIT Flow Cytometry Core Facility COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,

More information

BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY

BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY IMPORTANT NOTES BEFORE READING THIS TUTORIAL This is a very expensive piece of equipment so PLEASE treat it with respect! After you are done

More information

Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose

Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose Selecting Reagents for Multicolor Flow Cytometry Holden Maecker and Joe Trotter BD Biosciences, San Jose Contents 1 The basics: Know your instrument 2 Fluorochromes: Go for the bright 3 Colors and specificities:

More information

APPLICATION INFORMATION

APPLICATION INFORMATION DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL

More information

Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and

Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and Using BD FACSDiva CST To Evaluate Cytometer Performance, Create Custom Assay Settings and Implement Cross-Instrument and Cross-Site Standardization of Assays PART 2 Alan M. Stall Director, Advanced Cytometry

More information

Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...

Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation... Immunophenotyping Flow Cytometry Tutorial Contents Experimental Requirements...1 Data Storage...2 Voltage Adjustments...3 Compensation...5 Experimental Requirements For immunophenotyping with FITC and

More information

Compensation: An Instrumental Perspective

Compensation: An Instrumental Perspective Compensation: An Instrumental Perspective Why Digital? Visualization Issues Boston User s Group September 10, 2003 Joe Trotter Joe_Trotter@bd.com Review: Instrument Sensitivity Measuring Sensitivity -

More information

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Compensation and gating

More information

Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking

Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Mark KuKuruga Senior Technical Applications Specialist BD Biosciences 23-14462-00 Outline

More information

Introduction to flow cytometry

Introduction to flow cytometry Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression

More information

BD FACSComp Software Tutorial

BD FACSComp Software Tutorial BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS

More information

How To Read Flow Cytometry Data

How To Read Flow Cytometry Data 26 Nature Publishing Group http://www.nature.com/natureimmunology Interpreting flow cytometry data: a guide for the perplexed Leonore A Herzenberg, James Tung, Wayne A Moore, Leonard A Herzenberg & David

More information

Immunophenotyping peripheral blood cells

Immunophenotyping peripheral blood cells IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune

More information

123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.

123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.

More information

Four Color Compensation

Four Color Compensation Cytometry (Communications in Clinical Cytometry) 38:161 175 (1999) Four Color Compensation Carleton C. Stewart* and Sigrid J. Stewart Laboratory of Flow Cytometry, Roswell Park Cancer Institute, Buffalo,

More information

Instrument Characterization and Performance Tracking for Digital Flow Cytometers

Instrument Characterization and Performance Tracking for Digital Flow Cytometers Instrument Characterization and Performance Tracking for Digital Flow Cytometers BD Biosciences Cytometer Set-up & Tracking (CS&T) System For Research Use Only. Not for use in diagnostic or therapeutic

More information

Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility

Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility The mention of commercial products, their sources, or their use in connection

More information

Compensation Basics - Bagwell. Compensation Basics. C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc.

Compensation Basics - Bagwell. Compensation Basics. C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc. Compensation Basics C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc. 2003 1 Intrinsic or Autofluorescence p2 ac 1,2 c 1 ac 1,1 p1 In order to describe how the general process of signal cross-over

More information

Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract

Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract arch 212 Standardizing Application Setup Across ultiple Flow Cytometers Using BD FACSDiva Version 6 Ellen einelt, ervi Reunanen, ark Edinger, aria Jaimes, Alan Stall, Dennis Sasaki, Joe Trotter Contents

More information

The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee

The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee January, 1999 From: To: Subject: The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee NIAID DAIDS Flow Cytometry Laboratories Comparison study information for labs wishing to switch

More information

Boundary-breaking acoustic focusing cytometry

Boundary-breaking acoustic focusing cytometry Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory

More information

CRITICAL ASPECTS OF STAINING FOR FLOW CYTOMETRY

CRITICAL ASPECTS OF STAINING FOR FLOW CYTOMETRY 页 码,1/6 CRITICAL ASPECTS OF STAINING FOR FLOW CYTOMETRY From Givan, A.L. (2000), chapter in In Living Color: Protocols in Flow Cytometry and Cell Sorting (R. Diamond and S. DeMaggio, eds). Springer, Berlin,

More information

FACSCanto RUO Special Order QUICK REFERENCE GUIDE

FACSCanto RUO Special Order QUICK REFERENCE GUIDE FACSCanto RUO Special Order QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW

More information

Standardization, Calibration and Quality Control

Standardization, Calibration and Quality Control Standardization, Calibration and Quality Control Ian Storie Flow cytometry has become an essential tool in the research and clinical diagnostic laboratory. The range of available flow-based diagnostic

More information

A practical approach to multicolor flow cytometry for immunophenotyping

A practical approach to multicolor flow cytometry for immunophenotyping Journal of Immunological Methods 243 (2000) 77 97 www.elsevier.nl/ locate/ jim A practical approach to multicolor flow cytometry for immunophenotyping Nicole Baumgarth, Mario Roederer a, b, * a Department

More information

Below is a list of things you should be aware of before you schedule your sort.

Below is a list of things you should be aware of before you schedule your sort. Sorting at the Flow Cytometry Facility At the present time, the assistance of a trained cell sorter operator is needed for all sorting applications. As such, if you are planning a first time sort you will

More information

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required.

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required. PROTOCOL Immunostaining for Flow Cytometry 1850 Millrace Drive, Suite 3A Eugene, Oregon 97403 Rev.0 Background The combination of single cell analysis using flow cytometry and the specificity of antibody-based

More information

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University

More information

CyAn : 11 Parameter Desktop Flow Cytometer

CyAn : 11 Parameter Desktop Flow Cytometer CyAn : 11 Parameter Desktop Flow Cytometer Cyan ADP 3 excitation lines 488nm, 635nm, and UV or violet 11 simultaneous parameters FSC, SSC, and 7-9 colors with simultaneous width, peak, area, and log on

More information

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that

More information

FlowJo Basic Tutorial

FlowJo Basic Tutorial FlowJo Basic Tutorial FlowJo is software for the analysis of flow cytometry data. This tutorial will quickly introduce you to FlowJo using an example experiment, where a FITC anti-cd8 reagent is titrated

More information

Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T

Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom Cat. Reference: STP-25T Reagent provided:: 25 Stepcount tubes for 25 test INTENDED USE. Immunostep

More information

Introduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size

Introduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size Introduction to Flow Cytometry -- BD LSR II Daisy Kuo 郭 正 佼 Assistant Product Manager BD Biosciences Daisy_kuo@bd.com What is Flow Cytometry? Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements

More information

Introduction into Flow Cytometry

Introduction into Flow Cytometry Introduction into Flow Cytometry Stephanie Gurka, Andreas Hutloff, Timo Lischke, Kroczek-Lab, Robert Koch Institute, Berlin Flow Cytometry - FCM FACS = Fluorescence Activated Cell Sorting (also used for

More information

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy.

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. At the end of this session, the participant will be able to: 1. Describe the components of a flow cytometer 2. Describe the gating

More information

BD StemFlow Human Pluripotent Stem Cell Sorting and Analysis Kit

BD StemFlow Human Pluripotent Stem Cell Sorting and Analysis Kit BD StemFlow Human Pluripotent Stem Cell Sorting and Analysis Kit Instruction Manual Catalog No. 560461 ii Human Pluripotent Stem Cell Sorting and Analysis Kit 2009, Becton, Dickinson and Company. All rights

More information

These particles have something in common

These particles have something in common These particles have something in common Blood cells Chromosomes Algae Protozoa Certain parameters of these particles can be measured with a flow cytometer Which parameters can be measured? the relative

More information

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?

More information

DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES

DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES CLAUDIO ORTOLANI UNIVERSITY OF URBINO - ITALY SUN TZU (544 b.c. 496 b.c) SUN TZU (544 b.c. 496 b.c.) THE ART OF CYTOMETRY

More information

Chapter 6. Antigen-Antibody Properties 10/3/2012. Antigen-Antibody Interactions: Principles and Applications. Precipitin reactions

Chapter 6. Antigen-Antibody Properties 10/3/2012. Antigen-Antibody Interactions: Principles and Applications. Precipitin reactions Chapter 6 Antigen-Antibody Interactions: Principles and Applications Antigen-Antibody Properties You must remember antibody affinity (single) VS avidity (multiple) High affinity: bound tightly and longer!

More information

1. Instrument Layout pg 2. 2. Summit Overview pg 3. 3. Data Storage and Data Files pgs 6-7. 4. Saving Data pg 7. 5. Start Up pg 8

1. Instrument Layout pg 2. 2. Summit Overview pg 3. 3. Data Storage and Data Files pgs 6-7. 4. Saving Data pg 7. 5. Start Up pg 8 Table of Contents 1. Instrument Layout pg 2 2. Summit Overview pg 3 3. Data Storage and Data Files pgs 6-7 4. Saving Data pg 7 5. Start Up pg 8 6. Choosing and Naming Parameters pg 9 7. Creating Histograms

More information

Notes on titering antibodies

Notes on titering antibodies Home Events This Week Bulletin Boards Lab Personnel Lab Policies Protocols MolBio Tetramers Proteins Flow Cytometry Immunoassays Instruments Computers Lab Publications Scientiific Literature Science on

More information

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer Introduction Standard methods for isolating and detecting leukocytes

More information

Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer

Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Contents 2 Thresholds 2 Setting the Threshold When analyzing small particles, defined as particles smaller than 3.0 µm, on the

More information

Setting up and calibration of a flow cytometer for multicolor immunophenotyping

Setting up and calibration of a flow cytometer for multicolor immunophenotyping Journal of Biological Regulators and Homeostatic Agents Setting up and calibration of a flow cytometer for multicolor immunophenotyping J. KRAAN 1, J.W. GRATAMA 1, M. KEENEY 2, J.L. D HAUTCOURT 3 1 Department

More information

Guidelines on the use of multicolour flow cytometry in the diagnosis of haematological neoplasms

Guidelines on the use of multicolour flow cytometry in the diagnosis of haematological neoplasms guideline Guidelines on the use of multicolour flow cytometry in the diagnosis of haematological neoplasms Ulrika Johansson, 1 David Bloxham, 2 Stephen Couzens, 3 Jennifer Jesson, 4 Ricardo Morilla, 5

More information

Playing Good Comp, Bad Comp. Vera S. Donnenberg, Petar Popovic, Erin McClelland, Anita Popovic, Melissa Merola, Cassandra Singer, Albert D.

Playing Good Comp, Bad Comp. Vera S. Donnenberg, Petar Popovic, Erin McClelland, Anita Popovic, Melissa Merola, Cassandra Singer, Albert D. Playing Good Comp, Bad Comp Vera S. Donnenberg, Petar Popovic, Erin McClelland, Anita Popovic, Melissa Merola, Cassandra Singer, Albert D. Donnenberg University of Pittsburgh Cancer Institute BUG 9/10/03

More information

Flow Cytometry 101. Jodi Moore, PhD Jan 2012 Goldenson 113

Flow Cytometry 101. Jodi Moore, PhD Jan 2012 Goldenson 113 Flow Cytometry 101 Jodi Moore, PhD Jan 2012 Goldenson 113 Origins:? Dye Chemistry 1940-1980 s Electronics Microscopy Automated Clinical Cytology ID Apps Computers Cytometry vs. Flow Cytometry Microspectrophotometry

More information

Potency Assays for an Autologous Active Immunotherapy (Sipuleucel-T) Pocheng Liu, Ph.D. Senior Scientist of Product Development Dendreon Corporation

Potency Assays for an Autologous Active Immunotherapy (Sipuleucel-T) Pocheng Liu, Ph.D. Senior Scientist of Product Development Dendreon Corporation Potency Assays for an Autologous Active Immunotherapy (Sipuleucel-T) Pocheng Liu, Ph.D. Senior Scientist of Product Development Dendreon Corporation Sipuleucel-T Manufacturing Process Day 1 Leukapheresis

More information

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each

More information

FlowSight. Flow cytometry with vision

FlowSight. Flow cytometry with vision FlowSight Flow cytometry with vision Flow cytometry with vision Introducing FlowSight Capable: Sensitive and flexible for every need Intuitive: Easy-to-use, with imagery for every cell Affordable: Designed

More information

INSIDE THE BLACK BOX

INSIDE THE BLACK BOX FLOW CYTOMETRY ESSENTIALS INSIDE THE BLACK BOX Alice L. Givan Englert Cell Analysis Laboratory of the Norris Cotton Cancer Center Dartmouth Medical School HOW NOT TO BE A FLOW CYTOMETRIST Drawing by Ben

More information

Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction

Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction Application Note 10 Measurement of Cell Recovery After Sorting with a Catcher-Tube-Based Cell Sorter Introduction In many experiments using sorted cells, it is important to be able to count the number

More information

Islet Viability Assessment by Single Cell Flow Cytometry

Islet Viability Assessment by Single Cell Flow Cytometry Islet Viability Assessment by Single Cell Flow Cytometry Page 1 of 8 Purpose: To comprehensively assess the viability of the islet cell preparation prior to transplantation. Tissue Samples: A sample containing

More information

A Basic Guide. Institute of Medical Biology, University of Southern Denmark. Ungated Gate 1 (= R1) Gate 4 (=R4) Gate3 (=R3) Gate 2 (= R2) R3 R4

A Basic Guide. Institute of Medical Biology, University of Southern Denmark. Ungated Gate 1 (= R1) Gate 4 (=R4) Gate3 (=R3) Gate 2 (= R2) R3 R4 Flow Cytometry A Basic Guide R1 Ungated Gate 1 (= R1) R2 R3 R4 Gate 2 (= R2) Gate3 (=R3) Gate 4 (=R4) Graham Leslie, PhD Institute of Medical Biology, University of Southern Denmark Graham Leslie, PhD

More information

UCHC Guide to CellQuest Pro. FACS Calibur instruments

UCHC Guide to CellQuest Pro. FACS Calibur instruments UCHC Guide to CellQuest Pro FACS Calibur instruments Startup: Turn on Cytometer using green switch on the side of the cytometer Cytometer Power Switch Within a minute of turning on the cytometer; turn

More information

BD FACSamples and Recommendations

BD FACSamples and Recommendations Eesti FACSi Kasutajate Konverents Advances in Flow Cytometry 6-7.11.2012 Tallinn Meriton Grand Conference & Spa Hotel Eesti FACSi Kasutajate Konverents Advances in Flow Cytometry 6-7.11.2012 Tallinn Edited

More information

FACS Laboratory. BD LSRs Operator Training. Training Guide. http://london-research-institute.org.uk/technologies/facs

FACS Laboratory. BD LSRs Operator Training. Training Guide. http://london-research-institute.org.uk/technologies/facs FACS Laboratory facs_lab@cancer.org.uk Training Guide http://london-research-institute.org.uk/technologies/facs BD LSRs Operator Training LRI FACS lab 1 Components 1.1 Basic parts overview.pg3-6 Overview

More information

GETTING STARTED. 6. Click the New Specimen icon.

GETTING STARTED. 6. Click the New Specimen icon. BD FACSDiva Software Basic Experiment Guide Please note, this guide is not intended to take the place of in-person training but is to be used as a reference as needed after completing training with Unified

More information

BD FC Beads 8-Color Kit for BD OneFlow Assays

BD FC Beads 8-Color Kit for BD OneFlow Assays BD FC Beads 8-Color Kit for BD OneFlow Assays 7/2014 23-15785-00 IVD 5 tests per kit Catalog No. 658621 BD, BD Logo and all other trademarks are property of Becton, Dickinson and Company. 2014 BD Becton,

More information

Science is hard. Flow cytometry should be easy.

Science is hard. Flow cytometry should be easy. Science is hard. Flow cytometry should be easy. TABLE OF CONTENTS 1 INTRODUCTION TO BD ACCURI C6 SOFTWARE... 1 1.1 Starting BD Accuri C6 Software... 1 1.2 BD Accuri C6 Software Workspace... 2 1.3 Opening

More information

Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility

Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility 10.1071/RD14043_AC CSIRO 2015 Supplementary Material: Reproduction, Fertility and Development, 2015, 27(8), 1187 1196. Supplementary Material Free-radical production after post-thaw incubation of ram spermatozoa

More information

UCHC Guide to FACS DiVa. LSR II instruments

UCHC Guide to FACS DiVa. LSR II instruments UCHC Guide to FACS DiVa LSR II instruments Start Up: Turn on the instrument using the Green power button. Fluidics pump power, LSR II-B & C LSR II-B and LSRII-C-> Turn on the fluidics pump and ensure that

More information

QDot Nanocrystals Insights

QDot Nanocrystals Insights QDot Nanocrystals Insights Invitrogen Corporation Eugene, OR. HeLa cell labeled with three Qdot conjugates for nucleus, Golgi, and tubulin What are quantum dots? Highly fluorescent, nanometer-size, crystals

More information

Principles of Flowcytometry

Principles of Flowcytometry Objectives Introduction to Cell Markers: Principles of Flowcytometry Michelle Petrasich NZIMLS Scientific Meeting August 24, 2010, Paihia What are cell markers How do we detect them Production of Monoclonal

More information

Introduction to Flow Cytometry

Introduction to Flow Cytometry Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer

More information

Analyzer Experiment setup guide for LSRII and Canto s

Analyzer Experiment setup guide for LSRII and Canto s Analyzer Experiment setup guide for LSRII and Canto s 1. Check the instrument configuration on our website to determine the most appropriate cytometer for your experiment, https://depts.washington.edu/flowlab/instrumentation.html,

More information

MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:

MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as: INTRODUCTION Flow Cytometry involves the use of a beam of laser light projected through a liquid stream that contains cells, or other particles, which when struck by the focused light give out signals

More information

FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data. Monisha Sundarrajan, PhD BD Biosciences

FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data. Monisha Sundarrajan, PhD BD Biosciences FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data Monisha Sundarrajan, PhD BD Biosciences 23 13795 00 Agenda Introduction Overview of BD CBA bead based immunoassays FCAP

More information

Titering Antibodies INTRODUCTION Materials Procedure for titering antibodies to extracellular antigens A. Directly conjugated antibodies

Titering Antibodies INTRODUCTION Materials Procedure for titering antibodies to extracellular antigens A. Directly conjugated antibodies C.C. Stewart, S.J. Stewart, Titering Antibodies. In: Current Protocols in Cytometry, (J.P. Robinson, Z. Darzynkiewicz, P. Dean. L. Dressler, P.Rabinovitch, C. Stewart, H. Tanke, L. Wheeless, eds.) J.Wiley

More information

Lab 2. Isolation of mononuclear cells from peripheral blood and separation into subpopulations

Lab 2. Isolation of mononuclear cells from peripheral blood and separation into subpopulations Lab 2 Isolation of mononuclear cells from peripheral blood and separation into subpopulations Supervisors: Sissela Broos sissela.broos@immun.lth.se tel: 222 96 78 Niclas Olsson niclas.olsson@immun.lth.se

More information

BD FACSVerse. Simply brilliant

BD FACSVerse. Simply brilliant BD FACSVerse Simply brilliant BD FACSVerse: Simply Brilliant The BD FACSVerse flow cytometer was engineered from the ground up to offer remarkable performance, flexibility, and ease of operation for research

More information

IQ Corporation c/o Harold G. Haines, Ph.D. President SEP 1 f 2002 South Florida BioAssociates, Inc. 11511 SW 127h Street Miami, Florida 33 176

IQ Corporation c/o Harold G. Haines, Ph.D. President SEP 1 f 2002 South Florida BioAssociates, Inc. 11511 SW 127h Street Miami, Florida 33 176 **,lo 4 DEPARTMENT OF HEALTH & HUMAN SERVICES IQ Corporation c/o Harold G. Haines, Ph.D. President SEP 1 f 2002 South Florida BioAssociates, Inc. 11511 SW 127h Street Miami, Florida 33 176. Food and Drug

More information

EuroFlow standardization of flow cytometer instrument settings and immunophenotyping protocols

EuroFlow standardization of flow cytometer instrument settings and immunophenotyping protocols Leukemia (2012) 26, 1986 2010 All rights reserved 0887-6924/12 www.nature.com/leu SPECIAL REPORT EuroFlow standardization of flow cytometer instrument settings and immunophenotyping protocols T Kalina

More information

CFSE Cell Division Assay Kit

CFSE Cell Division Assay Kit CFSE Cell Division Assay Kit Item No. 10009853 Customer Service 800.364.9897 * Technical Support 888.526.5351 www.caymanchem.com TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 4 Precautions

More information

International Beryllium Conference, Montreal, Canada March 10, 2005

International Beryllium Conference, Montreal, Canada March 10, 2005 Alternative Lymphocyte Proliferation Tests: BrdU and Flow Cytometry Based Tests International Beryllium Conference, Montreal, Canada March 10, 2005 Tim K. Takaro Department of Environmental and Occupational

More information

ArC Amine Reactive Compensation Bead Kit

ArC Amine Reactive Compensation Bead Kit ArC Amine Reactive Compensation Bead Kit Catalog no. A1346 Table 1. Contents and storage information. Material Amount Composition Storage Stability ArC reactive beads (Component A) ArC negative beads (Component

More information

Institut für Experimentelle Immunologie und Bildgebung

Institut für Experimentelle Immunologie und Bildgebung Page 1 von 6 Starting Up the System Start up the computer, start Windows and log in as IMCES_user start BD FACSDiva software with your Groupname Turn on the cytometer main power Wait a moment and Check

More information

CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping

CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping CyFlow SL Healthcare Immunology Portable FCM System for 3-colour Immunophenotyping 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results

More information

Introduction to Flow Cytometry

Introduction to Flow Cytometry Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: daisy_kuo@bd.com BDBiosciences Application Examples

More information

BD FACSCanto Flow Cytometer. Efficient Easy Powerful

BD FACSCanto Flow Cytometer. Efficient Easy Powerful BD FACSCanto Flow Cytometer Efficient Easy Powerful The BD FACSCanto System: Optimal Workflow and Time Management Automated compensation algorithms with optional off-line compensation, streamlined sample

More information

--Verity Software House, ModFit for Proliferation http://www.vsh.com/ offers online tutorials

--Verity Software House, ModFit for Proliferation http://www.vsh.com/ offers online tutorials Customers may want to use another analysis software package with data collected on Attune. Reason for this include: they already have the SW and are confortable with using it, their institution may provide

More information

Standard Operating Procedure

Standard Operating Procedure 1.0 Purpose: 1.1 The characterisation of of main leukocyte subsets in peripheral blood cells from mice by flow cytometry. Reliable values of frequencies of leukocyte clusters are very much dependent on

More information

Getting Started with BD FACSDiva Software

Getting Started with BD FACSDiva Software Getting Started with BD FACSDiva Software For In Vitro Diagnostic Use bdbiosciences.com Part No. 642213 Rev. A June 2007 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 USA Tel (877) 232-8995 Fax

More information

BD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin

BD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin BD LSR II and FACSDiVa Software, Basel Dr. Norbert Leclere,, Berlin Overview Electronics Covers Fluidics Connectors Controlpanel Sample Port On/Off The Control Panel Ease of Use The basic procedures to

More information

BrdU Flow Kits. Instruction Manual. FITC BrdU Flow Kit. APC BrdU Flow Kit. BD Pharmingen. Cat. No. 559619 (50 tests) Cat. No. 557891 (4 50 tests)

BrdU Flow Kits. Instruction Manual. FITC BrdU Flow Kit. APC BrdU Flow Kit. BD Pharmingen. Cat. No. 559619 (50 tests) Cat. No. 557891 (4 50 tests) BD Pharmingen BrdU Flow Kits Instruction Manual FITC BrdU Flow Kit Cat. No. 559619 (50 tests) Cat. No. 557891 (4 50 tests) APC BrdU Flow Kit Cat. No. 552598 (50 tests) Cat. No. 557892 (4 50 tests) BD flow

More information

PRODUCT INFORMATION SHEET Monoclonal antibodies detecting human antigens

PRODUCT INFORMATION SHEET Monoclonal antibodies detecting human antigens www.iqproducts.nl PRODUCT INFORMATION SHEET Monoclonal antibodies detecting human antigens IFN- γ PURE [RUO] [REF] IQP-160P s 50 tests FITC [RUO] [REF] IQP-160F s 50 tests R-PE [RUO] [REF] IQP-160R s 50

More information

Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH))

Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH)) Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH)) October 14, 2013 1 Licensing Under the Artistic License, you are

More information

Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-806)

Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-806) Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-86) The Hoechst purification was established for murine hematopoietic stem cells (HSC)

More information

BD FACSCalibur. The flow cytometer for your routine cell analysis needs

BD FACSCalibur. The flow cytometer for your routine cell analysis needs BD FACSCalibur The flow cytometer for your routine cell analysis needs A system with a rich application basis and a modular approach that continues to meet evolving needs of cell analysis worldwide. The

More information

Introduction to Flow Cytometry:

Introduction to Flow Cytometry: Introduction to Flow Cytometry: A Learning Guide Manual Part Number: 11-11032-01 April, 2000 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 1-800-448-2347 Introduction to Flow Cytometry: A Learning

More information

Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit

Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit 2 Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit Presented by Ellen Meinelt, MS MLS(ASCP) CM, Technical Applications Specialist and Calin Yuan, Product Course Developer,

More information

How to operate the BD FACSCanto flow cytometer

How to operate the BD FACSCanto flow cytometer How to operate the BD FACSCanto flow cytometer Preface Dear colleague, the BD FACSCanto flow cytometer is use to operate, however, it is a delicate instrument for measuring fluorescence of single cells.

More information

powerful. Intuitive. Customizable. The new star of benchtop flow cytometry

powerful. Intuitive. Customizable. The new star of benchtop flow cytometry powerful. Intuitive. Customizable. The new star of benchtop flow cytometry About ACEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.

More information