INSIDE THE BLACK BOX
|
|
|
- Mervyn Ross
- 9 years ago
- Views:
Transcription
1 FLOW CYTOMETRY ESSENTIALS INSIDE THE BLACK BOX Alice L. Givan Englert Cell Analysis Laboratory of the Norris Cotton Cancer Center Dartmouth Medical School
2 HOW NOT TO BE A FLOW CYTOMETRIST Drawing by Ben Givan
3 KAMENTSKY, MELAMED, AND DERMAN: NEW INSTRUMENT FOR ULTRARAPID CELL ANALYSIS Kamentsky, Melamed, and Derman, Science 150: , 1965
4 PAP SMEAR BY FLOW CYTOMETRY: 1965 UV ABSORBANCE 410nm BACK SCATTER Kamentsky, Melamed, and Derman, Science 150: , 1965
5 MEDLINE REFERENCES TO "FLOW CYTOMETRY" DNA OTHER 0% 20% 40% 60% 80% 100% WHEN?
6 STARTING MATERIAL: STAINED CELLS: WHAT A FLOW CYTOMETER CANNOT TELL YOU U937 CELL STAINED WITH PROPIDIUM IODIDE
7 EVEN MORE THAT A FLOW CYTOMETER CANNOT TELL YOU MONOCYTES: STAINED WITH FITC AND PE FOR FC RECEPTORS: PAUL WALLACE, DEPT OF MICROBIOLOGY KEN ORNDORFF, ENGLERT CELL ANALYSIS LABORATORY
8 A NOBLE GAS ION LASER coherent color (eg argon ion laser: UV/488nm/514nm) coherent direction
9 THE GAUSSIAN LASER BEAM PROFILE
10 BEAM SHAPES: MORE OR LESS ABILITY TO ILLUMINATE CELLS EQUALLY EVEN IF THEY STRAY FROM THE CENTER OF THE BEAM AND MORE OR LESS ABILITY TO AVOID COINCIDENCE OF TWO CELLS IN THE BEAM EVEN IF CELLS FOLLOW EACH OTHER CLOSELY IN THE FLOW STREAM
11 FLOW CYTOMETRY FLUIDICS
12 IF YOU ACHIEVE A FAST FLOW RATE BY MAKING THE SAMPLE CORE WIDE,YOU WILL HAVE WIDE CVs AND MULTIPLE CELLS MAY COINCIDE IN THE LASER BEAM
13 core diameter as a function of cell concentrationnnn and cell flow rate (stream velocity 10 m/s) cells/ml 15 5x10 5 cells/ml cells/ml 5 5x10 6 cells/ml cell flow rate (cells/second)
14 ORTHOGONAL ALIGNMENT
15 LIGHT FROM THE ANALYSIS POINT
16 4-PARAMETER OPTICAL BENCH
17 PARTITIONING LIGHT WITH MIRRORS AND FILTERS
18 FCS DATA FILE $ F
19 FCS DATA FILE (TRANSLATED) CELLS IN SEQUENCE
20 SINGLE COLOR HISTOGRAM WITH STATS % pos % pos 43 8
21 TWO-DIMENSIONAL PLOTS
22 TWO-DIMENSIONAL SCATTER PLOTS FOR PHYSICAL CHARACTERISTICS OF CELLS PLOT IT ANY WAY YOU LIKE
23 ELECTRONICS: AMPLIFICATION OF THE SIGNALS
24 GATING ON FORWARD AND SIDE SCATTER MONOS PMNS LYMPHS
25 MONOS PMNS FLUORESCENCE OF GATED CELLS LYMPHS
26 INSTRUMENT VOCABULARY (1) LASER: illuminates cells SHEATH FLUID: surrounds the sample stream, aligning cells in the laser beam SAMPLE CORE: narrow stream carrying cells within the wider sheath stream NOZZLE/FLOW CELL: delivers cells into the sheath stream and then accelerates the flow so that cells arrive one by one at the analysis point ANALYSIS POINT: position at which the laser beam intersects the sheath stream and illuminates cells
27 INSTRUMENT VOCABULARY (2) FLUORESCENCE: light emitted from a fluorochrome after it absorbs light from the laser (always of a longer wavelength than the absorbed light) SCATTER: light that is reflected, refracted, or otherwise bounced off a cell (always the same color as the illuminating beam) FORWARD SCATTER (FSC or FALS): light scattered at small angles away from its original direction SIDE SCATTER (SSC or 90 o LS): light scattered to 90 o from its original direction
28 INSTRUMENT VOCABULARY (3) PHOTODETECTOR: detects light scattered or emitted by cells and converts it to an electronic signal AMPLIFIER: increases or decreases signals from cells so as to put them on scale COMPUTER: takes the light signal that has been detected by the photodetector, amplified by the amplifier, and digitized by the analog-to-digital converter, and stores it as a number in a file for analysis.
29 ANALYSIS VOCABULARY (1) PARAMETER: 4, 5,6, or more types of information that are provided by a flow cytometer about each cell -- for example, a cytometer with 5 photodetectors provides 5 kinds of information about each cell DATA FILE: list of numbers that form the total flow description of a cell suspension after it passes the analysis point HISTOGRAM/1-D HISTOGRAM: plots the data for cells in a data file with respect to one parameter DOT PLOT/2-D HISTOGRAM: plots the data for cells in a data file with respect to two correlated parameters
30 ANALYSIS VOCABULARY (2) REGION: a description, using flow cytometric intensity values, of a cluster of cells GATE: a restriction, using Boolean combinations of regions, on the cells that will be included in subsequent analysis MARKER: a more-or-less arbitrary dividing line between intensities that will dichotomize negative from positive cells QUADRANTS: markers in 2 dimensions that divide signals into negative, single positive (2), and double positive intensities
31 FCS DATA FILE (AGAIN) CELLS IN SEQUENCE
32 NEUTROPHIL STAINED for DNA, CD16, and IFNγ WHAT DOES A CELL REALLY LOOK LIKE? a cell (by flow cytometry) a cell (by confocal microscopy) FSC = 224 SSC = 603 FL1 = 467 FL2 = 894 FL3 = 126
33 Webster s New World Dictionary (1958) says that a "minimizer" is "one who tries to make religious or philosophical problems appear easily explained." With humility, we need to admit that flow cytometry is, essentially, a minimizing technique. It simplifies the uniqueness and elegance and complexity of a cell down to a set of 5 or so numbers. The power of flow cytometry comes from rapid, objective, and quantitative computing, allowing calculations, correlations, and statistical conclusions from those few numbers derived from each of many cells. Our intelligence and training should make us aware of all those important things that flow cytometry does not tell us about a cell.
Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service
Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?
Introduction to Flow Cytometry
Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer
Introduction to flow cytometry
Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression
Flow Cytometry Protocols
METHODS IN MOLECULAR BIOLOGY TM TM Volume 263 Flow Cytometry Protocols SECOND EDITION Edited by Teresa S. Hawley Robert G. Hawley 1 Flow Cytometry An Introduction Alice L. Givan Summary A flow cytometer
Introduction to Flow Cytometry:
Introduction to Flow Cytometry: A Learning Guide Manual Part Number: 11-11032-01 April, 2000 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 1-800-448-2347 Introduction to Flow Cytometry: A Learning
Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy.
Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. At the end of this session, the participant will be able to: 1. Describe the components of a flow cytometer 2. Describe the gating
These particles have something in common
These particles have something in common Blood cells Chromosomes Algae Protozoa Certain parameters of these particles can be measured with a flow cytometer Which parameters can be measured? the relative
CyAn : 11 Parameter Desktop Flow Cytometer
CyAn : 11 Parameter Desktop Flow Cytometer Cyan ADP 3 excitation lines 488nm, 635nm, and UV or violet 11 simultaneous parameters FSC, SSC, and 7-9 colors with simultaneous width, peak, area, and log on
Flow cytometry basics fluidics, optics, electronics...
Title Flow cytometry basics fluidics, optics, electronics... RNDr. Jan Svoboda, Ph.D. Cytometry and Microscopy Core Facility IMB, CAS, v.v.i Vídeňská 1083 Fluorescence Fluorescence occurs when a valence
Flow Cytometry. flow cytometer DNA apoptosis ph
Flow Cytometry flow cytometer DNA apoptosis ph flow cytometry flow cytometer (a) (b) cell counting instrument (c) 1960 ink-jet technology 17 19 1940 1950 fluorescence microscopy fluorescent dye DNA polyclonal
Flow Cytometry A Basic Overview
Flow Cytometry A Basic Overview Overview Flow cytometry is a powerful technology for investigating many aspects of cell biology and for isolating cells of interest. Flow cytometry utilizes highly focused,
MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:
INTRODUCTION Flow Cytometry involves the use of a beam of laser light projected through a liquid stream that contains cells, or other particles, which when struck by the focused light give out signals
Introduction to Flow Cytometry
Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: [email protected] BDBiosciences Application Examples
Flow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement
What is Flow Cytometry? Flow Cytometry Basic Principle and Applications BD Biosciences Daisy Kuo ([email protected]) Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements are made on cells,
CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
UNIVERSITY OF PÉCS MEDICAL SCHOOL FLOW CYTOMETRY AND CELL SEPARATION BIOPHYSICS 2. 2015 4th March Dr. Beáta Bugyi Department of Biophysics Flow cytometry and cell separation FLOW = STREAM OF FLUID in a
CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
COMPENSATION MIT Flow Cytometry Core Facility
COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,
Below is a list of things you should be aware of before you schedule your sort.
Sorting at the Flow Cytometry Facility At the present time, the assistance of a trained cell sorter operator is needed for all sorting applications. As such, if you are planning a first time sort you will
FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS. By: Douaa Moh. Sayed
FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS By: Douaa Moh. Sayed Definition Flow cytometry is a technique for counting, examining, and sorting microscopic particles suspended in a stream of fluid. It allows
Boundary-breaking acoustic focusing cytometry
Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Presented by Clare Rogers, MS Senior Marketing Applications Specialist BD Biosciences 23-13660-00 Webinar Overview Multicolor
CyFlow SL. Microbiology. Detection and Analysis of Microorganisms and Small Particles
CyFlow SL Microbiology Detection and Analysis of Microorganisms and Small Particles 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Compensation and gating
Cell Cycle Tutorial. Contents
Cell Cycle Tutorial Contents Experimental Requirements...2 DNA Dyes...2 Protocols...3 PI Parameter & Analysis Setup...4 PI Voltage Adjustments...6 7-AAD Parameter Setup...6 To-Pro3 Parameter Setup...6
Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility
10.1071/RD14043_AC CSIRO 2015 Supplementary Material: Reproduction, Fertility and Development, 2015, 27(8), 1187 1196. Supplementary Material Free-radical production after post-thaw incubation of ram spermatozoa
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Mark KuKuruga Senior Technical Applications Specialist BD Biosciences 23-14462-00 Outline
BD FACSComp Software Tutorial
BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS
Principles of Flowcytometry
Objectives Introduction to Cell Markers: Principles of Flowcytometry Michelle Petrasich NZIMLS Scientific Meeting August 24, 2010, Paihia What are cell markers How do we detect them Production of Monoclonal
No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer
APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer Introduction Standard methods for isolating and detecting leukocytes
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY IMPORTANT NOTES BEFORE READING THIS TUTORIAL This is a very expensive piece of equipment so PLEASE treat it with respect! After you are done
Science is hard. Flow cytometry should be easy.
Science is hard. Flow cytometry should be easy. TABLE OF CONTENTS 1 INTRODUCTION TO BD ACCURI C6 SOFTWARE... 1 1.1 Starting BD Accuri C6 Software... 1 1.2 BD Accuri C6 Software Workspace... 2 1.3 Opening
Flow Cytometry 101. Jodi Moore, PhD Jan 2012 Goldenson 113
Flow Cytometry 101 Jodi Moore, PhD Jan 2012 Goldenson 113 Origins:? Dye Chemistry 1940-1980 s Electronics Microscopy Automated Clinical Cytology ID Apps Computers Cytometry vs. Flow Cytometry Microspectrophotometry
Spherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1
78 Irma Lee Circle, Unit 0, Lake Forest, Illinois 00 SPHERO TM Technical Note STN- Rev B 008 DETERMINING PMT LINEARITY IN FLOW CYTOMETERS USING THE SPHERO TM PMT QUALITY CONTROL EXCEL TEMPLATE Introduction
CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping
CyFlow SL Healthcare Immunology Portable FCM System for 3-colour Immunophenotyping 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results
FlowSight. Flow cytometry with vision
FlowSight Flow cytometry with vision Flow cytometry with vision Introducing FlowSight Capable: Sensitive and flexible for every need Intuitive: Easy-to-use, with imagery for every cell Affordable: Designed
Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...
Immunophenotyping Flow Cytometry Tutorial Contents Experimental Requirements...1 Data Storage...2 Voltage Adjustments...3 Compensation...5 Experimental Requirements For immunophenotyping with FITC and
123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.
Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University
WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS
FOR IN-VITRO DIAGNOSTIC USE INVITROGEN CAL-LYSE TM Lysing Solution WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS CAL-LYSE TM CATALOG No. GAS-010 250 tests 25 ml CATALOG No. GAS-010S-100
Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction
Application Note 10 Measurement of Cell Recovery After Sorting with a Catcher-Tube-Based Cell Sorter Introduction In many experiments using sorted cells, it is important to be able to count the number
Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T
Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom Cat. Reference: STP-25T Reagent provided:: 25 Stepcount tubes for 25 test INTENDED USE. Immunostep
Instrument Characterization and Performance Tracking for Digital Flow Cytometers
Instrument Characterization and Performance Tracking for Digital Flow Cytometers BD Biosciences Cytometer Set-up & Tracking (CS&T) System For Research Use Only. Not for use in diagnostic or therapeutic
APPLICATION INFORMATION
DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL
How to operate the BD FACSCanto flow cytometer
How to operate the BD FACSCanto flow cytometer Preface Dear colleague, the BD FACSCanto flow cytometer is use to operate, however, it is a delicate instrument for measuring fluorescence of single cells.
Too Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry
Too Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry by Debby R. Walser-Kuntz Biology Department Carleton College, Northfield, MN Taylor goes in to see her doctor, Dr. Chavez,
Small-Particle Analysis
John Sharpe, PhD, and Sonja Wulff Introduction Flow cytometry instrumentation is a powerful tool for the analysis and identification of complex samples of particles of significantly varying size. However,
UCHC Guide to FACS DiVa. LSR II instruments
UCHC Guide to FACS DiVa LSR II instruments Start Up: Turn on the instrument using the Green power button. Fluidics pump power, LSR II-B & C LSR II-B and LSRII-C-> Turn on the fluidics pump and ensure that
Cell Cycle Phase Determination Kit
Cell Cycle Phase Determination Kit Item No. 10009349 Customer Service 800.364.9897 * Technical Support 888.526.5351 www.caymanchem.com TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 3 Safety
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This
Introduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size
Introduction to Flow Cytometry -- BD LSR II Daisy Kuo 郭 正 佼 Assistant Product Manager BD Biosciences [email protected] What is Flow Cytometry? Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements
Three Rules for Compensation Controls
From FlowJo s Daily Dongle Three Rules for Compensation Controls First and foremost, there must be a single-stained control for every parameter in the experiment! In addition, there are three rules for
Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH))
Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH)) October 14, 2013 1 Licensing Under the Artistic License, you are
CyFlow Space Your flexible flow cytometer
Your flexible flow cytometer www.sysmex-partec.com its flexibility gives you the space you need for your work Analysing cells and particles, be it from blood, plasma, tissue, plants, cell cultures or other
Standardization, Calibration and Quality Control
Standardization, Calibration and Quality Control Ian Storie Flow cytometry has become an essential tool in the research and clinical diagnostic laboratory. The range of available flow-based diagnostic
Bacterial Detection and Live/Dead Discrimination by Flow Cytometry
A P P L I C A T I O N N O T E Bacterial Detection and Live/Dead Discrimination by Flow Cytometry M I C R O B I A L C Y T O M E T R Y BD FACSCalibur, BD FACSort Flow Cytometers Introduction Accurate determination
Compensation Basics - Bagwell. Compensation Basics. C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc.
Compensation Basics C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc. 2003 1 Intrinsic or Autofluorescence p2 ac 1,2 c 1 ac 1,1 p1 In order to describe how the general process of signal cross-over
Microscopy. MICROSCOPY Light Electron Tunnelling Atomic Force RESOLVE: => INCREASE CONTRAST BIODIVERSITY I BIOL1051 MAJOR FUNCTIONS OF MICROSCOPES
BIODIVERSITY I BIOL1051 Microscopy Professor Marc C. Lavoie [email protected] MAJOR FUNCTIONS OF MICROSCOPES MAGNIFY RESOLVE: => INCREASE CONTRAST Microscopy 1. Eyepieces 2. Diopter adjustment
Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer
Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Contents 2 Thresholds 2 Setting the Threshold When analyzing small particles, defined as particles smaller than 3.0 µm, on the
BD FACSVerse. Simply brilliant
BD FACSVerse Simply brilliant BD FACSVerse: Simply Brilliant The BD FACSVerse flow cytometer was engineered from the ground up to offer remarkable performance, flexibility, and ease of operation for research
Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-806)
Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-86) The Hoechst purification was established for murine hematopoietic stem cells (HSC)
Institut für Experimentelle Immunologie und Bildgebung
Page 1 von 6 Starting Up the System Start up the computer, start Windows and log in as IMCES_user start BD FACSDiva software with your Groupname Turn on the cytometer main power Wait a moment and Check
MACSQuantify Software guide. Basics of flow cytometric analysis and software guide. Version 1 EN Original instructions
MACSQuantify Software guide Basics of flow cytometric analysis and software guide Version 1 EN Original instructions Miltenyi Biotec GmbH Friedrich-Ebert-Straße 68 51429 Bergisch Gladbach Germany Phone:
GETTING STARTED. 6. Click the New Specimen icon.
BD FACSDiva Software Basic Experiment Guide Please note, this guide is not intended to take the place of in-person training but is to be used as a reference as needed after completing training with Unified
Fluorescence Compensation. Jennifer Wilshire, PhD [email protected]
Fluorescence Compensation Jennifer Wilshire, PhD [email protected] Outline 1. Compensation Basics 2. Practical Applications -selecting comp controls (cells, stains?) 3. Multicolor Issues -spreading due
Lab 2. Isolation of mononuclear cells from peripheral blood and separation into subpopulations
Lab 2 Isolation of mononuclear cells from peripheral blood and separation into subpopulations Supervisors: Sissela Broos [email protected] tel: 222 96 78 Niclas Olsson [email protected]
A Basic Guide. Institute of Medical Biology, University of Southern Denmark. Ungated Gate 1 (= R1) Gate 4 (=R4) Gate3 (=R3) Gate 2 (= R2) R3 R4
Flow Cytometry A Basic Guide R1 Ungated Gate 1 (= R1) R2 R3 R4 Gate 2 (= R2) Gate3 (=R3) Gate 4 (=R4) Graham Leslie, PhD Institute of Medical Biology, University of Southern Denmark Graham Leslie, PhD
Easy to use. Compact. Flexible. Sorting Chip
SH800Z Cell Sorter The SH800Z is a compact, affordable, and easy to use cell sorter developed using the state- of-the art technologies from Sony. An array of automated mechanisms makes it an ideal sorter
Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract
arch 212 Standardizing Application Setup Across ultiple Flow Cytometers Using BD FACSDiva Version 6 Ellen einelt, ervi Reunanen, ark Edinger, aria Jaimes, Alan Stall, Dennis Sasaki, Joe Trotter Contents
BD CellQuest Pro Software Analysis Tutorial
BD CellQuest Pro Analysis Tutorial This tutorial guides you through an analysis example using BD CellQuest Pro software. If you are already familiar with BD CellQuest Pro software on Mac OS 9, refer to
Immunophenotyping peripheral blood cells
IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune
Analyzer Experiment setup guide for LSRII and Canto s
Analyzer Experiment setup guide for LSRII and Canto s 1. Check the instrument configuration on our website to determine the most appropriate cytometer for your experiment, https://depts.washington.edu/flowlab/instrumentation.html,
powerful. Intuitive. Customizable. The new star of benchtop flow cytometry
powerful. Intuitive. Customizable. The new star of benchtop flow cytometry About ACEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.
Zecotek S Light Projection Network Marketing
White Paper Zecotek Visible Fiber Laser Platform Enabling the future of laser technology Zecotek Photonics Inc. (TSX- V: ZMS; Frankfurt: W1I) www.zecotek.com is a Canadian photonics technology company
FACSCalibur System User s Guide
11-10823-02 Rev. A FACSCalibur System User s Guide 02-61760-02 August, 1996 Becton Dickinson Immunocytometry Systems 2350 Qume Drive San Jose, CA 95131-1807 Ordering Information (800) 223-8226 Customer
Partec PAS. Your Flow Cytometer.
Partec PAS Your Flow Cytometer. The Challenge... New sophisticated applications and increasing requirements for reliable routine results in the shortest possible time continue to impose a challenge for
UCHC Guide to CellQuest Pro. FACS Calibur instruments
UCHC Guide to CellQuest Pro FACS Calibur instruments Startup: Turn on Cytometer using green switch on the side of the cytometer Cytometer Power Switch Within a minute of turning on the cytometer; turn
flow cytometry reinvented
FLOW CYTOMETRY flow cytometry reinvented The Attune Acoustic Focusing Cytometer Precision and at all speeds for rare events and precious samples The Attune Acoustic Focusing Cytometer Precise high collection
Flow Cytometry Analysis of the Activity of Disinfecting Agents Tested with Staphylococcus aureus and Escherichia coli
Polish Journal of Microbiology 2005, Vol. 54, No 1, 21 26 Flow Cytometry Analysis of the Activity of Disinfecting Agents Tested with Staphylococcus aureus and Escherichia coli A. WO NIAK-KOSEK and J. KAWIAK*
1. Instrument Layout pg 2. 2. Summit Overview pg 3. 3. Data Storage and Data Files pgs 6-7. 4. Saving Data pg 7. 5. Start Up pg 8
Table of Contents 1. Instrument Layout pg 2 2. Summit Overview pg 3 3. Data Storage and Data Files pgs 6-7 4. Saving Data pg 7 5. Start Up pg 8 6. Choosing and Naming Parameters pg 9 7. Creating Histograms
DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES
DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES CLAUDIO ORTOLANI UNIVERSITY OF URBINO - ITALY SUN TZU (544 b.c. 496 b.c) SUN TZU (544 b.c. 496 b.c.) THE ART OF CYTOMETRY
Islet Viability Assessment by Single Cell Flow Cytometry
Islet Viability Assessment by Single Cell Flow Cytometry Page 1 of 8 Purpose: To comprehensively assess the viability of the islet cell preparation prior to transplantation. Tissue Samples: A sample containing
AS COMPETITION PAPER 2008
AS COMPETITION PAPER 28 Name School Town & County Total Mark/5 Time Allowed: One hour Attempt as many questions as you can. Write your answers on this question paper. Marks allocated for each question
BD FACSCanto Flow Cytometer. Efficient Easy Powerful
BD FACSCanto Flow Cytometer Efficient Easy Powerful The BD FACSCanto System: Optimal Workflow and Time Management Automated compensation algorithms with optional off-line compensation, streamlined sample
Setting up and calibration of a flow cytometer for multicolor immunophenotyping
Journal of Biological Regulators and Homeostatic Agents Setting up and calibration of a flow cytometer for multicolor immunophenotyping J. KRAAN 1, J.W. GRATAMA 1, M. KEENEY 2, J.L. D HAUTCOURT 3 1 Department
Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility
Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility The mention of commercial products, their sources, or their use in connection
Deep profiling of multitube flow cytometry data Supplemental information
Deep profiling of multitube flow cytometry data Supplemental information Kieran O Neill et al December 19, 2014 1 Table S1: Markers in simulated multitube data. The data was split into three tubes, each
Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and
Using BD FACSDiva CST To Evaluate Cytometer Performance, Create Custom Assay Settings and Implement Cross-Instrument and Cross-Site Standardization of Assays PART 2 Alan M. Stall Director, Advanced Cytometry
FACSCanto RUO Special Order QUICK REFERENCE GUIDE
FACSCanto RUO Special Order QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW
Becton-Dickinson FACSCalibur Flow Cytometer Instructions
Start-up 1. Turn on flow cytometer by pushing green button on right side of instrument, toward the back. Give it 30 minutes to warm up before collecting data. 2. Turn on monitor. 3. Turn on computer. If
A. FSC and SSC gating of total BM cells. B. Gating strategy used to identify the Lin -
Supplementary Figure legends Figure S1. Multiparametric analysis of HSC Populations. A. FSC and SSC gating of total BM cells. B. Gating strategy used to identify the Lin - cell population. BM cells were
HIGH PERFORMANCE FLOW CYTOMETER FOR EVERYONE
POWERFUL. INTUITIVE. CUSTOMIZABLE. HIGH PERFORMANCE FLOW CYTOMETER FOR EVERYONE ABOUT ACEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
EDF Extended Depth of Field
EDF Extended Depth of Field An upgrade for the ImageStream system Think outside the dot. Break the classic depth of field barrier Improve precision Enhance discrimination Simplify analysis Increase resolution
Let the BD FACSAria II accompany your best performance ever
Let the BD FACSAria II accompany your best performance ever Meet the new BD FACSAria II high-speed, fixed-alignment cuvette cell sorter. The first generation BD FACSAria system brought the complex world
Compensation Controls Data Visualization Panel Development Gating Quantum Dots
Compensation Controls Data Visualization Panel Development Gating Quantum Dots FITC Single Stain Control FITC PE Argon Laser FL1 FL2 450 500 550 600 FITC Compensation Control Unwanted signal detected in