Four Color Compensation

Size: px
Start display at page:

Download "Four Color Compensation"

Transcription

1 Cytometry (Communications in Clinical Cytometry) 38: (1999) Four Color Compensation Carleton C. Stewart* and Sigrid J. Stewart Laboratory of Flow Cytometry, Roswell Park Cancer Institute, Buffalo, New York Four-color immunophenotyping can now be routinely performed using either a single laser or dual laser flow cytometer. When a single laser instrument is used, the fluorochromes evaluated are usually FITC, PE, PE-TR and PE-CY5 (or PerCP). For two-laser excitation APC is generally used in place of PE-TR. Since each tandem dye construct contains PE, three of the four detectors are affected and compensation can be problematic. In this report we show that each tandem conjugated antibody, whether different batches from the same supplier or conjugates from different suppliers all require unique compensation. This inconsistency results in erroneous data, negates the use of single labeled particles as a method for providing adequate compensation and requires dual and triple labeled cells of known pattern to verify compensation. It is also shown that improper compensation can reduce or eliminate completely the detection of fluorescence emission from PECY5 conjugated antibodies. These problems are caused by a variation in energy transfer between PE and either TR or CY5 because the chemistry involved in preparation and conjugation to antibodies is not sufficiently controlled to produce reagents with uniform compensation requirements. The variation in tandem dye compensation can be addressed by either using the same tandem conjugated antibody, by using the same second step tandem reagent to an appropriate first step antibody or by using software compensation. The latter provides an easy solution because a unique compensation matrix can be produced for each antibody tandem conjugate. Cytometry (Comm. Clin. Cytometry) 38: , Wiley-Liss, Inc. Key terms: flow cytometry; instrument and software compensation; tandem dyes; antibodies; single laser; dual laser Multiparameter flow cytometry is a powerful tool for the objective measurement of antigen co-expression on populations of hematopoietic cells (1 6). Over the last decade, we have learned that, by combining two antibodies together, patterns emerge in the form of cell clusters that represent the expression of antigens on the cell s surface. When three antibodies are added together, even more clusters can be resolved representing additional subsets of cells (7). Further increasing the dimensionality of the data by adding four antibodies not only results in increasing the resolving power for important subsets, but can also markedly decrease the cost of specimen processing by reducing the number of tubes analyzed and reducing the number of redundant antibodies used. Each time another reagent is added, the complexity of the analysis process is increased. For example, the potential for antibody co-blocking (8,9), or fluorochrome quenching (10), or energy transfer (11) by epitopes in close proximity must be thoroughly explored before accepting any data as valid. It is now possible to routinely perform four-color immunophenotyping using either a single (Beckman Coulter XL) or dual (BDIS FACSCalibur) laser instrument. When a single laser instrument is used, the most frequently employed fluorochromes, all excited at 488 nm by an argon laser, are FITC, PE, PE-Texas-Red (PE-TR) and PE-CY5. PerCP conjugated antibodies can be used instead of PE-CY5 conjugated ones. Currently, a limited number of antibodies directly conjugated with PE-TR in the form of ECD are available from Beckman Coulter. Otherwise, biotinylated or unconjugated antibodies can be used in combination with the second reagents conjugated with PE-TR, products available from Life Technologies. When a dual laser instrument is used, three fluorochromes (usually FITC, PE and PE-CY5 or PerCP) are excited at 488 nm by an argon laser and one fluorochrome (usually APC) is excited at 635 nm by a diode or HeNe laser. A large selection of antibodies is available directly conjugated with APC from many reagent suppliers. Grant sponsor: New York State Department of Health; Grant sponsor: National Cancer Institute (NCI); Grant numbers: #2 P30-CA16056 and #5RO1CA *Correspondence to: Carleton C. Stewart, Ph.D., Laboratory of Flow Cytometry, Roswell Park Cancer Institute, Elm and Carlton Streets, Buffalo, NY Stewart@sc3101.med.buffalo.edu Received 8 January 1999; Accepted 6 April Wiley-Liss, Inc.

2 162 STEWART AND STEWART Compensation represents a special problem for fourcolor flow cytometry because PE is present in two or three of the four fluorochromes used. While it is customary to provide compensation standards using cells separately labeled with each fluorochrome, this approach is not adequate for four-color compensation adjustments. When each fluorescently labeled cell is interrogated, compensation signals will be present at the appropriate electronic summation junctions for the multiple compensations required. Because only single labeled cells are used, only one color at a time is present at these junctions, and the actual effect of one fluorochrome on another will not be seen. Even if a mixture of the four populations in a single tube is analyzed, each color is interrogated one at a time because each single labeled cell appears in the laser beam at a different time, and the effect seen will still be as if they were run separately. Only when the same cell is labeled with two or more fluorochromes can the true compensation requirement be evaluated and the effect of compensating one fluorochrome on another fluorochrome appreciated. This means that control cells must be labeled with antibodies whose binding to the cells within the suspension is known. In this report, we show that the fluorochrome emission from PE-TR in the single laser configuration or PE-CY5 or PerCP fluorochrome emission in the dual laser configuration is exquisitely sensitive to improper instrument compensation. Because both of the tandem constructs (but not PerCP) can exhibit significant variations in their amount of PE leakage, proper compensation is unique for each batch of fluorochrome and antibody to which it is conjugated. This problem can produce significant logistic problems in acquiring correctly compensated data when samples contain different antibodies conjugated with the tandem dyes. A method for verifying the correct instrument compensation is provided. In addition, we provide a strategy, using software compensation that can be used to easily solve the compensation problems associated with the variation in tandem fluorochromes conjugated to different antibodies. MATERIALS AND METHODS For these studies peripheral blood from healthy donors was used to provide daily assessment for immunophenotyping and instrument performance (12,13). The use of excess human products for research purposes is performed using approved protocols by the Institutional Review Board at RPCI. Blood was collected in heparinized vacutainers (BD) and the cellularity was determined using a Coulter M430 hematology counter. One part of blood was washed with 14 parts of PBS containing 10 units of heparin per ml. Throughout this report a wash is defined as centrifuging the cells for 3 min. at 1500 G, aspirating the supernatant and resuspending the pellet in residual buffer. The cells were washed a second time with 14 volumes of PBS without heparin. This procedure is used to reduce platelet aggregation by removing clotting factors and eliminating immunoglobulins or soluble cell surface antigens to which selected antibodies may be directed. The cells were resuspended in PBS for the experiments, at a concentration of cells per ml. The antibodies used for these studies were conjugated with fluorescein isothiocyanate (FITC), phycoerythrin (PE), PE-Texas Red (ECD), PE-cyanin 5 (PE-CY5), peridinin chlorophyll protein (PerCP) or allophycocyanin (APC). FITC-CD3, FITC-CD45, PE-CD3, PE-CD4, PerCP-CD45 and APC-CD45 were obtained from Becton Dickinson Immunocytometry Systems (BDIS; San Jose, CA). PE-CD19, PE- CD64, PE-CY5-CD2, PE-CY5-CD3, PE-CY5-CD8, PE-CY5- CD10, PE-CY5-CD22, PE-CY5-CD33 and PE-CY5-CD64 were obtained from CALTAG Laboratories (Burlingame, CA). PE-CY5-CD19 was obtained from Immunotech, Beckman- Coulter (Hialeah, FL). All antibodies were titered before use as previously described (14). Cells were stained as previously described in detail (15,16). Briefly, antibodies in the appropriate combinations in a final volume of 30µl were added to mm polypropylene tubes (Falcon #2054) followed by 50µl of cells. The final volume of the sample (antibodies plus cells) was rigorously maintained at 80µl. Cells were incubated 15 min. in an ice bath and then 3.5 ml freshly prepared lysing reagent (4.13g Ammonium Chloride (NH 4 CL) -- Sigma A-5666, 0.5g Potassium Bicarbonate (KHCO 3 ) -- Sigma P-4913, g Tetra Sodium EDTA - Sigma ED4SS) pre-warmed to 25 C was added and incubated 5 min. After centrifuging at 1500 g for three minutes, cells were washed with 3.5 ml PBS and fixed in 300µl of 2% formaldehyde (1:5 dilution in PBS of Poly- Sciences #04018 Ultrapure EM Grade 10% formaldehyde) for a minimum of 30 min., but no longer than 72 hours, prior to data acquisition. A FACScan (BDIS) was modified to acquire four-color data using a single laser (17). Briefly, a fourth PMT with associated electronics was added. Instrument compensation was made possible for all combinations of detectors, including those not available on a commercial FACScan. Instrument compensation could also be turned off so the user has a choice between hardware and software compensation. Data acquisition for the fourth color was made possible by using the DDE pulse processing capability of Lysys II Software and assigning the area parameter to FL4. A FACSCalibur dual laser flow cytometer (BDIS) was used for all the two laser excitation studies. Both instruments underwent a quality assessment each morning using DNA check microspheres (BC part # ). The acceptable variation in the peak channel was less than 5% and the coefficient of variation (CV) was less than 3% for all parameters except side scatter where the CV was less than 5%. Interlaser time for the FACSCalibur was calibrated at least three times per day according to the manufacturer s procedure, and the sheath fluid level was never allowed to decrease past one-third full. We have noticed instability in interlaser timing, which could be completely resolved, if the sheath tank is never allowed to decrease beyond one third full. When the instrument is functioning properly the timing verification passes on the first try, using the software supplied. If repeated tries are necessary, there is a timing problem that must be found

3 FIG. 1. Lymphogated cells for compensation. A lymphogate was applied to data from processed human blood to provide cells for compensation standards. and fixed. Usual causes for timing instability are low sheath tank volume, a partially blocked flow cell, high debris count, the red laser requires replacement or air bubbles in the sample stream. Preliminary instrument compensation was performed using cells stained with either FITC-CD45, PE-CD4, ECD-CD4 or APC-CD45 and with each reagent conjugated with PE-CY5 or PerCP. For each sample, 10,000 ungated events were acquired using a FSC threshold set at the extreme left side of the lymphocyte cluster (see Fig. 1). For the forward scatter gain settings we use, this threshold was 200. Acquired data was transferred to a PC based computer using Apple Talk FTP, where it is permanently archived on CD-ROM discs. The data was analyzed using WinList version 4.0 (Verity Software House, Topsham, ME). This program provides software compensation using matrix algebra and has been previously described (18). A lymphogate was used for analysis of all reagents except CD33 and CD64 where monocytes were also included in the gate and CD10 where granulocytes were also included. RESULTS For these studies, a lymphogate was used, as shown in Figure 1. Lymphocytes are ideal because they represent a relatively homogeneous population of small cells that exhibit low autofluorescence. Using single laser excitation and cells stained with only one antibody prior to mixing them together, the instrument compensation is shown in Figure 2 for the six possible bivariate displays of FITC versus PE, ECD versus PE, PE-CY5 versus PE, FITC versus ECD, PE-CY5 versus ECD and PE-CY5 versus FITC. Cells stained with each reagent are wholly contained within their appropriate single color region. One would expect that this instrument is properly compensated. Using the combination PE-CD8, ECD-CD4 and PE-CY5-CD2, to stain lymphocytes, all CD4 and CD8 positive cells should be CD2 and CD4 and CD8 positive cells should be mutually exclusive populations. As shown in Figure 3, the instrument is not properly compensated (top row); because some CD8 or CD4 cells are CD2 - and some CD8 cells are CD4 dim. Both of these problems can FOUR COLOR COMPENSATION 163 be corrected, as shown in Figure 3 (bottom row) after making an adjustment to the compensation, shown in the Figure 3 caption. It is not possible to see this change when a single labeled compensation standard is used because the acceptable limits are too broad (11.5% vs 9%). Beckman/ Coulter provides reagent combinations called Cyto- Comp reagent kit (part # ) for use with their XL flow cytometer for verifying compensation. They also produce PE-TR (ECD) tandem conjugated antibodies, with consistent compensation requirements, but the number of reagents is limited. It may still be necessary to stain cells with the current batch, of each reagent combination, to verify that the Cyto-comp kit behaves the same way as the reagents being used to stain cells. This verification would only be necessary when new batches of reagents are initiated. Instead of using instrument compensation, software compensation, such as that provided by Beckman Coulter with their XL flow cytometer or WinList by Verity Software House, can be used. When using WinList software compensation, single color compensation standards are prepared as described above and uncompensated files collected. This program uses matrix algebra previously described in detail (18) for determining compensation. Nevertheless, a multicolor compensation standard like that shown in Figure 3 must still be evaluated to obtain the correct compensation in software. When using software compensation, the samples are stained with the desired antibody combinations and the data is acquired uncompensated, as shown in Figure 4, top row and the compensation matrix applied to this data to produce the compensated data shown in the second row. For comparison, the bottom row shows the same specimen re-run after instrument compensation. These results clearly show that there is no difference in the patterns produced between instrument and software compensation. The percentages of positive cells for software versus instrument compensation, compared in Table 1, are in good agreement. One feature of software compensation is that the matrix can be named so that it can be incorporated into a data analysis macro. In this way a unique compensation matrix which can be automatically associated with its unique fluorochrome antibody conjugation batch. This capability provides for custom compensation described in more detail later when using the four-color, two laser applications. The most serious problem associated with either three or four-color immunophenotyping is the variation in compensation required for the tandem complexes. This variation is illustrated in Figure 5, and it is due to the inability to produce a consistent tandem of either PE-TR (not shown) or PE-CY5 that has exactly the same amount of PE leakage. When a second laser is used to excite APC, the CY5 in the PE-CY5 tandem is also excited. Since the CY5 compensation will differ from the APC compensation, this additional problem must also be appropriately dealt with. PerCP reagents are also excited to a lesser extent by the second laser. As shown in Figure 5 and Table 2, the PE-CY5-tandem variations cover a range from 0 to 8.9% for FL2-%FL3 and

4 164 STEWART AND STEWART FIG. 2. Instrument compensation. Blood was aliquoted into four tubes which contained either FITC-CD45 (FL1), PE-CD4 (FL2), PE-CY5-CD8 (FL3) or ECD-CD4 (FL4). After processing, data was acquired on a FACScan modified to collect four colors of fluorescence and the compensation adjusted so each population was within the appropriate quadrant. The six-parameter combinations of bivariate data are displayed. Compensation settings (in percent) are FL1 - %FL2 0.5, FL2 - %FL3 12, FL2 - %FL4 26, FL3 - %FL4 15, FL4 - %FL1 0.5 and FL4 - %FL3 50. The PMT high voltages were FL1 675, FL2 600, FL3 715 and FL from 16% to 42% for FL4-%FL3 on our FACSCalibur. These percentages will vary depending on the high voltage adjustments for each PMT and the batch of PE-CY5 used. When two different batches of CD8 were compared, batch 801 exhibited 1.1% compensation for FL2-%FL3 and 35.8% for FL4-%FL3. The mean values for this batch are shown in Table 2. Batch 901 (not shown in Table 2) exhibited a similar FL2-%FL3 compensation, but a much higher FL4- %FL3 compensation of 61.3%. A second batch of PE-CY5- CD19 exhibited a similar variation to the batch shown in Table 2. For this batch FL2-%FL3 was 1% and FL4-%FL3 was 39.1%. Antibodies conjugated with PerCP, Table 2, required no compensation for FL3-%FL2 and 3% compensation for FL4-%FL3. No variation in compensation requirements was found among different antibodies conjugated with PerCP. The problems illustrated in Figure 6, associated with using cells (or microspheres) stained with a single color for obtaining the correct compensation that were described for a four-color single laser instrument are also found with a two-laser instrument. As shown in the top row, four PE-CY5 reagents were compensated using single antibody stained cells. The amount of compensation for each reagent is reported in the caption. For each reagent the x axis fluorescence intensity (also shown in the view) is much higher than that found for the same reagent in row 2 when the PE-CY5 conjugated antibody was combined with APC-CD45. Indeed, very few CD8 T-cells and no CD64 monocytes were found. When this same combination is prepared with FITC-CD45, as shown in the third row, the PE-CY5 conjugated antibody fluorescence intensity is identical to that found when it is evaluated as a single reagent (top row). As shown in the fourth row, this situation does not occur if each antibody is conjugated with PE, rather than a PE-CY5 and then combined with APC-CD45. This problem illustrated in row 2 is unique to the combination of PE-CY5 conjugated antibodies combined with APC-CD45. It is caused by overcompensation of FL3-% FL4 when using a single color compensation sample. So why is a single color compensation sample inadequate? When a single color is used for compensation, there are no other signals at the summation junctions within the compensation circuits except those from one fluorochrome. Even if cells stained separately with different fluorochromes are mixed together, they appear one at a time for interrogation. When cells are dually positive, fluorescence signals from each fluorochrome will appear at the summation junctions simultaneously. Since the first fluorochrome will have a component of the second s color and vice versa, each of them will produce a different compensation requirement when evaluated on doubly stained cells compared to singly stained cells. The degree of difference in the compensation settings between the single versus double labeled cells will depend on the intensity of each fluorochrome and the percent of their respective

5 FOUR COLOR COMPENSATION 165 FIG. 3. PE-TX-Red & PE-CY5 compensation. Blood was processed using the antibody combination PE-CD8, ECD-CD4 and PE-CY5-CD2 (TC). The top row shows the pattern obtained after single color compensation. The appropriate adjustments were made to the compensation to obtain the correct pattern shown in the bottom row. All compensation and high voltage settings are listed in the caption for Figure 2 except FL3 - %FL2 and FL3 - %FL4. These are the most sensitive compensation parameters. The compensation for FL3 - %FL2 for TC-CD2 vs PE-CD8 in the column 1, top view using single labeled cells was 11.5% and for column 1, bottom view using double-labeled cells was 9.0%. The compensation for FL3 - %FL4 for TC-CD2 vs ECD-CD4 in column 2, top view, using single labeled cells was 58% and for column 2, bottom view, using double-labeled cells was 40%. spectra that overlap. (A detailed description of why there is a difference may be found in the appendix.) This means that a dual stained cell preparation must be used to verify compensation, as shown in Figure 7. To obtain the correct compensation, a very bright APC antibody, such as APC-CD45 that will stain all cells, is combined with a PE-CY5 or PerCP antibody, such as CD8, that stains only a subset. As illustrated in Figure 7-A, the correct compensation is obtained by adjusting for the maximum separation of the negative and the brightest stained clusters (shown by the line). This is the same position for single labeled CD8 positive cells, shown in Figure 6. CD8 is a good antibody to use because both bright and dim populations of CD8 cells are present and each should be well resolved. In Figure 7B, the sample has been overcompensated by only 4%, and the CD8 appears to be heterogeneously expressed, the cells are not as bright and the dim cells are poorly resolved as they merge with the bright cells. In Figure 7C, the sample is not compensated and the bright (b), dim (d) and negative (n) populations are still resolved. When the dual stained standard is used, the optimal separation between positive and negative cells can be obtained. Once adjusted, the setting is good for any PE-CY5/APC antibody combination. Unfortunately, even the two-color combination is inadequate and a three-color compensation standard is needed because PE-CY5 emission occurs at the summation junctions of FL2, FL3 and FL4. By combining mutually exclusive PE (CD4) and PE-CY5 (CD8) bright antibodies with a bright APC antibody coexpressed (CD45) by both, the optimal separations for all populations can be achieved through compensation. The reasons why this is necessary are illustrated in Figure 8. When a single color sample such as PE-CD4 is used for compensation (top row of Fig. 8) PE (FL2) vs PE-CY5 (FL3) (A) and PE (FL2) vs APC (FL4) (B) and APC (FL4) vs PE-CY5 (FL3) (C) look like they are correctly compensated. When cells stained simultaneously with three bright antibodies such as PE-CD4, PE-CY5-CD8 and APC-CD45, are evaluated, as shown in the second row, PE-CD4 (FL2) (D) is clearly under compensated and there are a high number of apparently CD8 dim cells (F) shown in the ovals. This

6 166 STEWART AND STEWART FIG. 4. Comparison of compensation methods. Blood was processed with the four-color antibody combination FITC-CD3, PE-CD8, ECD-CD4 and PE-CY5-CD2 (TC). Data was acquired using no compensation (top row), software compensation (middle row) or instrument compensation (bottom row). The six-parameter combinations of bivariate data are displayed. See Figure 2 and Figure 3 captions for high voltage and compensation settings. Table 1 Comparison of Cluster Event Count in Software Compensated and Instrument Compensated Data Cluster a % of gated cells compensated CD3 CD8 CD2 CD4 Software Instrument a Populations 1% not included. appears to be caused by under compensation of either FL3-%FL2 or FL3-%FL4. The tendency might be to increase compensation for FL3-%FL2 to correct the under-compensation seen in Figure 8D. When this is done, and the single stained cells are reevaluated, third row, PE-CD4 stained cells (G) are correctly compensated, but they are now under-compensated for FL4-%FL2 (H, I). Clearly the selection of FL3 - %FL2 was the wrong choice. If, instead, FL3-%FL4 is adjusted and the single stained cells reevaluated (J, K, L) as well as the multicolor stained samples, M, N, O correct compensation has been achieved and verified. When performing this compensation make sure that the PE-CY5 positive cells are in the same location as they were in Figure 7 (vertical line) i.e. the maximum separation has been achieved. Therefore, it is necessary to evaluate a three-color stained sample to verify correct compensation. This antibody combination can be evaluated every time four-color samples are evaluated to verify that the instrument is properly compensated. We have defined two different problems. The first problem is encountered when each TC reagent requires a different amount of compensation and the second is encountered when we try to accurately adjust compensation using single color standards. What is a simple solution to these problems? Because each PE-CY5 conjugated antibody and any PerCP conjugated antibody requires its own unique compensation and it is desired to use several different PE-CY5 antibodies and PerCP antibodies in separate combinations with other antibodies that will be evaluated on the same specimen, we recommend using software compensation. This provides for a compensation matrix unique for each PE-CY5 antibody that has been combined with three other reagents. When a unique compensation matrix is generated in software for each

7 FOUR COLOR COMPENSATION 167 FIG. 5. Variation in compensation for PE-CY5 reagents. Separate samples are processed with PE-CY5 conjugated CD2, CD19 or two different batches of CD8. The data were acquired using no compensation or after adjusting the compensation to the value shown. The high voltage was FL3 690V, FL4 730V. Table 2 Variation Among and Stability of Compensation Requirements for Third Color Reagents FL2-%FL3 FL4-%FL3 a Antibody Supplier Mean % STD Mean % STD PE-CY5-CD2 Caltag PE-CY5-CD8 Caltag PE-CY5-CD10 Caltag PE-CY5-CD19 Immunotech PE-CY5-CD22 Caltag PE-CY5-CD33 Caltag PerCP-CD45 BDIS PE-CY5-CD62L Caltag PE-CY5-CD64 Caltag a The amount of compensation is dependent on the high voltage and may vary markedly from those shown here and still be correct. For this experiment, the high voltages were FL2 640V, FL3 690V and FL4 730V. The mean and % STD for each reagent measured monthly for seven months is shown. combination, it can be applied to uncompensated acquired data at the time the list mode files are analyzed. Is it necessary to create a new compensation matrix every time (e.g. daily) that a PE-CY5 conjugated antibody is evaluated? We determined the stability of 8 separate PE-CY5 antibodies and a PerCP antibody over a period of seven months by staining cells monthly and recorded the compensation required. The results, summarized in Table 2, demonstrate that compensation for each antibody is stable and varies 5% for FL4 - %FL3 when properly stored. Variation was greater than 5% for FL2 - %FL3, but this is because so little compensation is needed, that a change in only 0.1% would make a large statistical difference. Light exposure, however, can break down the PE-CY5 tandem, causing increased PE leakage and this would invalidate the matrix. Such an occurrence, however, would show up during periodic evaluation, as controls would fall outside acceptable thresholds established for each PE-CY5 antibody. This procedure offers a means for quality assessment of stored reagents on a routine basis. To perform the quality assessment, for both maximum separation of negative and bright clusters and the unique compensation for (FL3-%FL4), we suggest staining two groups of cells. The first group is stained with each PE-CY5 antibody alone. The second group is one tube of cells that have been stained with the antibody combination PE-CD4, PE-CY5-CD8 and APC-CD45. All three markers are very bright. A different combination might be chosen, but the important criteria is that all are bright markers, FL2 and FL3 cells are all FL4, but FL2 cells do not co-express FL3. Each PE-CY5 antibody stained cells is mixed with the cells stained with the combination. The pattern generated is shown in Figure 9. In the top row, CD4, CD8 and CD45 are properly compensated and maximum separation of CD45 CD4 and CD45 CD8 clusters from those cells that are only CD45 is achieved. TC-CD2 cells have not yet been properly compensated. Both FL2-%FL3 and FL4-

8 168 STEWART AND STEWART FIG. 6. Over compensation of PE-CY5 and APC-CD45. Blood was stained with each antibody combination shown on the axis. Each of them was compensated using the single stained cells shown in the top row. FL4 - % FL3 for PE-CY5-CD19 was 32.8%, PE-CY5-CD8 was 50%, PE-CY5-CD3 was 31.5% or PE-CY5-CD64 was 28.6% (for this antibody, monocytes were included in the gate) as a single antibody (top row) or combined with APC-CD45 (second row), or FITC-CD45, third row. In the fourth row, PE-CD19, PE-CD8, PE-CD3 or PE-CD64 was combined with APC-CD45. Note that PE-CD8 is brighter than PE-CY5-CD8, PE-CD3 is brighter than PE-CY5-CD3 and PE-CD64 is very much brighter than PE-CY5-CD64. The high voltage settings were FL1 690V, FL2 640V, FL3 690V and FL4 730V. FIG. 7. Using a double labeled compensation standard. Blood was stained with PE-CY5-CD8 and APC-CD45. When properly compensated (A, FL3 - %FL4 20%), the maximum spatial separation of each cluster representing CD8 negative (n), dim (d) and bright (b) cells, occurs. Over compensation (B, FL3 - %FL4 24%) significantly reduces the spatial separation and produces erroneous data (note position of cells relative to the vertical line in A vs B). In C, no compensation has been applied, but the three populations can still be seen. Proper compensation can only be achieved using cells stained with both antibodies so the maximal separation of n and b can be determined. The high voltage settings were FL3 690V and FL4 730V. %FL3 are adjusted to properly compensate the CD2 cells and the pattern shown in the bottom row is found. The instrument has been properly compensated for PE-CD4, APC-CD45 and TC-CD2. In this paradigm, instrument compensation is set for all antibodies except the TC reagents. For these the unique compensation matrix for the TC reagent is determined and subsequently used for software compensation.

9 FOUR COLOR COMPENSATION 169 FIG. 8. Using a Triple Labeled Compensation Standard. Blood was stained as a single label with PE-CD4, PE-CY5-CD8 or APC-CD45. Unstained cells were used to set the high voltage which was FL2 640V, FL3 690V and FL4 730V. The compensation was adjusted to FL2-%FL3 1%, FL3-%FL2 12.4%, FL3-%FL4 18% and FL4-%FL3 41%. A, B and C show that PE-CD4 cells appear to be properly compensated. Cells were also stained with the combination containing PE-CD4, PE-CY5-CD8 and APC-CD45. D, E and F show the bivariate histograms of these cells. FL3-%FL2 was increased to 16.3% using the mutually exclusive combination of PE-CD4 vs PE-CY5-CD8 (D). G, H and I show the reacquisition of PE-CD4 at this compensation setting. FL3-%FL2 was set to the original setting (12.4%) and FL3-%FL4 adjusted to 20% (from 18%) and J, K and L show reacquisition of PE-CD4 at this compensation setting and M, N and O show reacquisition of the triple labeled compensation standard.

10 170 STEWART AND STEWART FIG. 9. Verification of compensation. Blood is processed in two tubes. The first tube contains the antibody combination PE-CD4, PE-CY5-CD8 and APC-CD45 and the second tube contains only PE-CY5-CD2. After fixation, the contents of both tubes were combined and data acquired. Since the PE-CY5-CD8 requires little compensation for FL2-%FL3, it is used in the three-color antibody combination for insuring maximum separation of CD45 versus FL3. PE-CY5 reagents stained separately and mixed will appear to be uncompensated for FL4-%FL3 (A, B, C). When properly compensated, either in software or hardware, the pattern shown in D, E, F will appear thereby verifying correct compensation has been achieved. In this example FL2 - %FL3 10 and FL4 - %FL3 20 at high voltages for FL2 640V, FL3 690V and FL4 730V. Each PE-CY5 is verified monthly and whenever a new batch of reagent is introduced. DISCUSSION These results show that every PE-CY5 tandem requires its own unique compensation, but the stability for any single batch is quite good. PerCP, a natural tandem reagent does not exhibit this variation, as all PerCP conjugated antibodies require the same low compensation, nor does the 635 nm laser efficiently excite it. Unfortunately it is not as bright as the PE-CY5 reagents. The variation in compensation of PE-CY5 reagents is caused by both the leakage of PE photons due to a variation in energy transfer to CY5 among different tandem batches as well as conjugation to each antibody and to the excitation of CY5 in the tandem by the diode laser. Some suppliers of TC reagents have controlled their product s compensation for PE-%TC compensation to about 1%; others supply TC reagents that may require 5 10% PE-%TC compensation. For any single TC reagent, its compensation requirement is stable providing the reagent is not exposed to light during storage. The amount of TC found in the APC channel is much higher and this is because the CY5 is excited by the 635 nm laser. While the exact physical chemical relationships may not have been elucidated, the basis for this effect is the inability to produce conjugations of PE-CY5 to antibodies that are identical each time. The amount of PE-%TC and APC-%TC compensation for a given construct will depend on the tandem chemistry, the conjugation chemistry of the tandem to the antibody and the conjugated antibody s binding to its epitope on the cell. All three categories are variable with each preparation and that causes an effect on the efficiency of energy transfer from PE to CY5, which affects compensation. The variation among reagents creates a practical logistic problem in evaluating specimens when multiple tandem conjugated antibodies are combined with other antibodies in a multiple tube screen. The problem can be solved in three ways: 1. Always use the same PE-CY5 reagent. Since only one tandem is used, there is no variation, and the same compensation can be used for all samples regardless of the other three antibodies combined with it. For example, CD45-PE-CY5 (or PerCP) might be chosen as the common reagent. The problem with this solution is that the

11 investigator no longer has the flexibility to use other tandem antibodies in the four-color combinations. 2. Use a biotinylated antibody and a PE-CY5 avidin. An unconjugated antibody and a second antibody conjugated with the tandem could also be used. Since the tandem avidin or second antibody is always the same, there is no variation in compensation. Thus, the investigator now has the choice of any antibody in the four-color combination, but a second step is required. 3. Use software compensation. Using WinList, version 4 (or 5), a compensation matrix unique to each PE-CY5 or PerCP conjugated antibody can be produced. The appropriate compensation matrix is applied to each unique antibody among the combinations in the screen. This problem does not occur with PerCP reagents, as all of them will require the same FL4-%FL3 compensation and they require no FL2-%FL3 compensation. Nevertheless, both FL4-%FL3 and FL2-%FL3 compensation will be different for each PE-CY5 reagent and between the PE-CY5 and PerCP reagents. Over or under compensation is a problem that must be recognized by the laboratorian. A small change in the percent compensation can make a well-resolved cluster become heterogeneous in shape as shown in Figure 7. Cells with low epitope density that renders them dimly fluorescent may become completely negative, Figure 6, row 2. To insure that each parameter is properly compensated, cells should be evaluated every time data is collected using antibody combinations that will allow for verifying that the maximum difference between positive and negative cell clusters has been achieved. An example of how to accomplish this was shown in Figures 7, 8 and 9. For software compensation, the files are used to validate the compensation matrix. The compensation problems identified here will be even more acute in the future as laboratorians make their own combinations using analyte specific reagents (ASRs). It is likely that the manufacturers will introduce fewer FDA approved combinations (in vitro diagnostic device products), because it will be much less expensive for them to provide ASRs. All clinical laboratories performing flow cytometric evaluations using ASR combinations prepared by them must implement the regulations, which went into effect on November 23, Since all compensation combinations are available in software, it is important that the correct ones are used. It is also important to realize that a population of cells along one of the six pairs of axes may be a projection of an uncompensated population and not one that is actually positive for the antibody. All six bivariate plots of four-color data must be viewed before determining which pair requires compensation, because the bivariate that is not shown may be the one that actually shows the proper parameters to compensate. Only the relevant bivariates have been shown in this report to save space. The importance of choosing the proper parameters is best illustrated when PE, PE-CY5 and APC antibodies produce bright cellular staining such as the combination of PE-CD4, PE-CY5-CD8 and APC-CD45. As was shown in FOUR COLOR COMPENSATION 171 Figure 8, after compensation using cells stained with only a single antibody (top row), it looked as though the instrument was correctly compensated. But when cells stained with all three combined antibodies were evaluated, a high frequency of CD4 dim, CD8 cells were found. This situation will only occur when the FL4 cells are also FL2. If extra compensation is added to FL3 - %FL2 to correct this problem, a summation occurs that actually produces under compensation of CD4 vs CD45 (Figure 8H). Correct compensation can be achieved by increasing FL3-%FL4 (Figure 8K). This illustrates the interaction of the compensation circuits with one another and the need to understand how they work, so that the correct parameters are compensated. A discussion of how compensation works may be found in the Appendix. We recommend software compensation because a matrix can be prepared for each antibody combination and then applied to the sample during analysis of the data. Uncompensated data can be retrospectively compensated in software, but over compensated data obtained using hardware compensation is lost forever. Using WinList,a combination specific macro can be written that will automatically bring up the correct matrix as the sample is analyzed. Providing the PE-CY5 reagents are properly stored and not exposed to bright light, the compensation matrix for each batch of any PE-CY5 reagent is stable for at least six months. Further, anytime a new batch of reagent is initiated, a new compensation matrix must be generated because it is highly unlikely compensation requirements will be the same. There are two software compensation approaches to consider. The first is to acquire all data uncompensated and do all the compensation in software. The advantage of this approach is that all data can be compensated retrospectively. A corollary to this approach would be to use instrument compensation for FL1 and FL2 because they are not affected by the problems encountered with the tandems. Alternatively, instrument compensation for all parameters as described in Figure 9, can be performed, except for FL2-%FL3 and FL4 - %FL3, which is performed in software for each PE-CY5 reagent and for PerCP reagents. While there is no advantage to this approach, it is perhaps more traditional because all the other compensation conditions are performed on the instrument. There are currently two software packages available that are capable of compensation, WinList and FloJo (19). We have no experience with FloJo. An interaction between the PMT voltage (gain) and the percentage of that signal appearing at the summation junction from a second PMT produce compensation. The percentage compensation is usually adjusted by changing the attenuation of the compensation network. There is a perception that a high percentage of compensation is too much and this is certainly true if it is 99% wherein the network is out of range. In practice, any desired percentage compensation can be set and the events brought into compliance by adjusting the appropriate PMT voltage. This procedure will insure that the compensation network (or software compensation matrix) is in

12 172 STEWART AND STEWART range. Thus, the FL4-%FL3 compensation requirement for CD8 (batch 901) could be reduced to 30% by increasing the FL3 PMT voltage or decreasing the FL4 voltage. (See Table 2) Adjusting the PMT voltage, however, will produce a change in the position of unstained cells. Four-color immunophenotyping significantly improves our ability to resolve unique subsets of cells within a complex population. To insure that the data has been properly acquired, the laboratorian must have an understanding of the entire process, including the instrumentation, so potential artifacts are recognized. While some may feel increasing the number of antibodies is too complex to pursue, this application continues to expand our understanding of subsets of hematopoietic cells and in reducing the cost of processing. Using appropriate combinations of antibodies it is possible to both identify cells and determine what they are doing. Never before has the acquisition of so much relevant data been possible using wellestablished procedures. The addition of even more antibodies in the combination (19) will further expand this knowledge base as we strive to understand the biology of hematopoiesis in health and disease. ACKNOWLEDGMENTS The authors wish to thank Mary Beth Dell and David Sheedy for their expert assistance in specimen processing and data analysis. LITERATURE CITED 1. Jennings CD, Foon KA. Recent advances in flow cytometry: application to the diagnosis of hematologic malignancy. Blood 1997; 90: Roth G, Schmitz G. Consensus protocol for the flow cytometric immunophenotyping of hematopoietic malignancies. Working group on flow cytometry and image analysis. Leukemia 1996; 10: Braylan RC, Borowitz MJ, Davis BH, Stelzer GT, Stewart CC. U.S. -Canadian consensus recommendations on the immunophenotypic analysis of hematologic neoplasia by flow cytometry: data reporting. Cytometry 1997; 30: Davis BH, et al. U.S. -Canadian consensus recommendations on the immunophenotypic analysis of hematologic neoplasia by flow cytometry: medical indications. Cytometry 1997; 30: Borowitz MJ, et al. U.S. -Canadian consensus recommendations on the immunophenotypic analysis of hematologic neoplasia by flow cytometry: data analysis and interpretation. Cytometry 1997; 30: Loken M, Terstappen L, Civin C, Falker M. Flow cytometric characterization of erythroid, lymphoid and monomyeloid lineages in normal human bone marrow. In: O. Laerum, Bjerhnes R, editors. Flow cytometry in hematology, New York; Academic Press, Stewart CC, Stewart SJ. Multiparameter analysis of leukocytes by flow cytometry. In: Darzynkiewicz Z, Robinson J, Crissman H, editors. Methods in cell biology, New York; Academic Press, Inc., 1994; 41: Ierino FL, Hulett MD, McKenzie IF, Hogarth PM. Mapping epitopes of human Fc gamma RII (CDw32) with monoclonal antibodies and recombinant receptors. J Immunol 1993; 150: Stein R, Belisle E, Hansen HJ, Goldenberg DM. Epitope specificity of the anti-(b-cell lymphoma) monoclonal antibody, LL2. Cancer Immunol Immunother 1993; 37: Deka C, Lehnert BE, Lehnert NM, Jones GM, Sklar LA, Steinkamp JA. Analysis of fluorescence lifetime and quenching of FITC-conjugated antibodies on cells by phase-sensitive flow cytometry. Cytometry 1996; Waggoner AS. Fluorescent probes for cytometry, In: Melamed MR, Lindmo T, Mendelsohn ML, editors. Flow cytometry and sorting, New York; Wiley-Liss, 1990; p Owens M, Loken M. Flow cytometry principles for clinical laboratory practice. Quality assurance for quantitative immunophenotyping. New York; Wiley-Liss, Stewart CC, Stewart SJ. Phenotypic analysis. In: Robinson JP, Darzynkiewicz Z, Dean P, Dressler L, Rabinovitch P, Stewart C, Tanke H, Wheeless L, editors. Current protocols in cytometry, New York, John Wiley & Sons, Inc., 1997; Stewart CC, Stewart SJ. Titering antibodies. In: Robinson JP, Darzynkiewicz Z, Dean P, Dressler L, Rabinovitch P, Stewart C, Tanke H, Wheeless L, editors, Current protocols in cytometry, New York, John Wiley & Sons, Inc., 1997; Stewart CC, Stewart SJ. Cell Preparation for the identification of leukocytes. In: Darzynkiewicz Z, Robinson J, Crissman H, editors. Methods in cell biology, New York; Academic Press, Inc., 1994; 41: Stewart CC, Stewart SJ. The use of directly and indirectly labeled monoclonal antibodies in flow cytometry. In: Davis WC editor, Methods in molecular biology. Humana Press, Inc., 1995; 45: Timm EA, Podniesinski E, Duckett L, Cardott J, Stewart CC. Amplification and detection of a y-chromosome DNA sequence by fluorescence in situ polymerase chain reaction and flow cytometry using cells in suspension. Cytometry 1995; 22: Bagwell CB, Adams EG. Fluorescence spectral overlap compensation for any number of flow cytometric parameters. In: Landay A, Ault KA, Bauer RD, Rabinovitch, editors. Clinical flow cytometry, Annals of the New York Academy of Sciences, New York, 1993: 677:94: Roederer M, De Rosa S, Gerstein R, Anderson M, Bigos M, Stovel R, Nozaki T, Parks D, Herzenberg L, Herzenberg L. 8 color, 10-parameter flow cytometry to elucidate complex leukocyte heterogeneity. Cytometry 1997; 29: APPENDIX The purpose of this appendix is to provide a description of how compensation works. This may be helpful in understanding of the results presented in this report. There are two compensation combinations for two-color, six for three-color and 12 for four-color. As the number of parameters increases, the complexity and cost of electronic compensation increases, therefore, software compensation becomes a more attractive alternative. Since higher order compensation is an array of the basic twocolor compensation, this discussion will be limited to two-color compensation. The emission spectra for two fluorochromes are shown in Figure A-1. The spectrum is an intensity (probability) distribution of emitted photons (intensity) of differing energies (wavelength or color). The parallel lines represent the colors that are determined by band pass filters for detection by the photomultiplier tubes (PMT and preamplifier). A simplified schematic representation of a compensation circuit is shown in Figure 2. The actual circuit is more complex for several electronic design reasons. In Figure A-2, FLA is a PMT and pre-amplifier with a band pass filter for selection of photons whose energy is the color green. Similarly, FLB is a PMT and pre-amplifier with a band pass filter for selecting photons whose energy is yellow orange. We will call it red for discussion. The pre-amplifiers amplify the PMT output for processing by the compensation circuits. Each of the fluorochromes emit some light that will pass through both optical filters. The amount of this light depends on the emission spectrum of the dye and the characteristics for the optical filters used for each PMT. Compensation corrects for the light that enters the secondary detector. Since this light is directly proportional to the light observed in the primary detector, it can be removed by subtracting a proportion of the primary signal from the secondary detector. This can be performed for multiple detectors simultaneously.

13 FOUR COLOR COMPENSATION 173 FIG. A-1. Emission Spectra of Two Fluorochromes. Emission of photons from fluorochromes excited by light occurs over a broad range of energies. This emission is described by an intensity distribution of photons at different wavelengths. The spectra of two fluorochromes may overlap one another and the photons from either that have the same wavelength (or color) are indistinguishable. Optical filtration, shown by the parallel lines representing the filters bandwidth, can be used to restrict the color that is transmitted through an optical detection system. Nevertheless, overlap of part of the unwanted fluorochromes occurs. When cells are labeled with a single color (Fig. A-2a), photons from the fluorescent cells are focused through optical filters onto both PMTs. FLA will produce a greater output signal for green photons than FLB used to detect the red photons because the band pass filters in front of each PMT selectively filters the appropriate color. The signal height at the output of each PMT/pre-amp will be proportional to the photon intensity of each selected color and this is measured in volts. There are two summation circuits, %A and %B in Figure A-2. The summation device is called a differential operational amplifier, OPAMP, which can process two signals by inverting one of them (at the negative input) and adding them together. For compensation, the signal from FLA goes to the positive input of OPAMP A. The output signal from OPAMP A is routed through an adjustable circuit %A, into the negative input of OPAMP B. This negative signal from OPAMP A is adjusted so that it is exactly equal to the height of the signal at the output of FLB, thereby canceling the unwanted signal in FLB. Thus, we have taken a percentage of signal A and subtracted it from signal B, often written, B - %A. Note: Compensation is a linear process. The signal may then be amplified through a logarithmic amplifier. In a similar manner, when the tube of cells is labeled only with the second color, whose primary detector is FLB (Fig. A-2b), FLB will produce a greater output signal than FLA. As before, the bandpass filters selectively determine the amount of colored light detected by each PMT. Now the FLB signal is routed to the positive input of OPAMP B. The output of FLB is routed through the adjustable negative circuit, %B, where the height of signal B is adjusted to the same height as the unwanted signal A. Thus, we have taken a percentage of signal B and subtracted it from signal A or A - %B, thereby canceling the unwanted signal FLA. An important, often overlooked, corollary to this description is the high voltage adjustment. This amount of subtraction depends upon the relative sensitivity of the two detectors to the two fluorochromes. This changes when optical filters are changed or when the high voltage is changed. If the high voltage on one of the PMTs is increased the signal heights in that detector will be proportionally higher. This will cause the signal height at the negative input of the OPAMP to be higher and unless it is readjusted, over-compensation will occur. Similarly, if the high voltage of the PMT is lowered, its signal height will be lower, causing the adjusted signal height at the negative input of the OPAMP to be lower and undercompensation will result. Thus, the compensation circuit could be adjusted to a fixed value and the cells actually compensated by adjusting the high voltage. The high voltage and compensation circuits work together; adjusting one will always affect the other. What is different when the same cells express both colors? There is a difference and it occurs for any color combination and becomes increasingly more problematic as the number of measured colors increases. The awareness of the problem has not been high because it is less noticeable when only two or three fluorochromes are used with fairly well separated emission spectra. To see quantitatively what is occurring, we can use voltage to represent signal height by letting green and red fluorescence each equal 5000 mv as illustrated in Figure A-3. Since each fluorochrome has a constant proportion of photons in each detector s energy range, 5% of the red fluorochrome might overlap the green fluorochrome and

14 174 STEWART AND STEWART FIG. A-2. Compensation for single labeled cells. A conceptual diagram of compensation for cells labeled with fluorescent dyes. FLA and FLB represent the combined PMT and pre-amp. The triangular devices are OPAMPs whose function is to invert any signal at the negative input and then add it to any signal at the positive input. In Figure A-2a, when only green cells are acquired, the signals from FLA will represent the green component of emission and from FLB the red component from the green fluorochrome. In Figure A-2b, when only red cells are acquired, the signals from FLA will represent the green component from the red fluorochrome and the red component is detected by FLB. FIG. A-3. Compensation for double-labeled cells. The component of each fluorochrome is represented by individual signal heights that are added together. The latter being what the detector actually measures. The higher first signal is from the fluorochrome primarily detected and the lower second signal from the overlapping component of the other fluorochrome. The signal processing is identical to that shown in Figure A-2.

15 FOUR COLOR COMPENSATION 175 Table A-1 OPAMP Input Voltages (ma) for Single-Labeled Cells OPAMP A (mvolts) OPAMP B (mvolts) Input Green cells 5000 A %B B %A 1000 Output Red cells 250 A %B B %A 50 Output Table A-2 OPAMP Input Voltages (ma) for Double-Labeled Cells OPAMP A (mvolts) OPAMP B (mvolts) Input Green 5000 A %B B %A 1000 Red Total 5250 A %B B %A 1200 Output Table A-1 OPAMP Input Voltages (ma) for Single-Labeled Cells Input Output Green labeled cells OPAMP A 5000 A 5% B OPAMP B 1000 B 20% A Red labeled cells OPAMP A 250 A 5% B OPAMP B 5000 B 20% A the compensation for it so that A -%B 5%. The green fluorochrome might overlap the red fluorochrome by 20% sothatb-%a 20%. (The actual percentage of compensation will depend on the high voltage.) In general, emission spectra are skewed to the red so that blue-green dyes appear in orange-red detectors rather than visa versa. The voltages (in millivolts, mv) at each OPAMP are shown in Table A-1 for single color compensation. Referring to Figure A-2 for green cells, FLB sees 20% of A, so that the signal at B is 1000 mv. Since B 20%A 1000 mv, OPAMP B has no output because the exact voltage has been subtracted from FLB and the output signal for OPAMP A is 5000 mv. Similarly for red cells, FLA detects only 5% of B so that the signal at A is 5% of 5000 mv or 250 mv. Since A 5%B 250 mv, the output of OPAMP A is zero and the compensated output signal from OPAMP B is 5000 mv. What happens if cells are double labeled with 5000 mv of green fluorescence and 5000 mv of red fluorescence at the same compensation settings shown above? The voltages at each OPAMP for doubly labeled cells are shown in Table A-2. The FLA detector has a signal voltage equal to green 5% of red or 5000 mv 250 mv or 5250 mv and the FLB detector has a signal voltage equal to red 20% of green or 5000 mv 1000 mv 6000 mv. The compensation (A - 5%B) produces a negative input voltage for OPAMP A of 300 mv ( ) and the negative input for (B-20%A) OPAMP B is 1050 mv ( ). This Table A-2 OPAMP Input Voltages (ma) for Double-Labeled Cells Input Output OPAMP A 5250 A 5% B OPAMP B 6000 B 20% A produces an output from A of 4950 mv ( ). The output from OPAMP B is also 4950 mv ( ). Thus, each fluorochrome has an output voltage that is somewhat less than when they were tested as single labels. This will not be noticeable when both antibodies produce very bright fluorescence, but will produce an increasingly noticeable effect as under-compensation when fluorochromes have increasingly disparate intensities. This can be easily demonstrated by substituting different voltage values for the FLA and FLB voltages and computing the output voltage for each OPAMP as shown in Table 2. As we increase the number of fluorochromes with overlapping spectra and their brightness, the voltage differences for double, triple and quadruple labeled cells become more complex and exacerbated, producing increasing divergence between single and multiple labeled cells. This becomes increasingly noticeable and problematic. It is hoped that this brief description will be useful in understanding the effects of multiple fluorescent signals on compensation. The degree to which these effects become noticeable will depend on the fluorochromes relative intensities on the cells being measured. By using appropriate compensation controls, the precise optimal settings for reliable measurement of fluorescence can be achieved. Questions are welcome and may be addressed to: stewart@sc3101.med.buffalo.edu. ACKNOWLEDGMENTS The authors wish to thank Drs. Michael Loken and Gerald Marti for their review and contributions to the appendix.

Introduction to flow cytometry

Introduction to flow cytometry Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression

More information

APPLICATION INFORMATION

APPLICATION INFORMATION DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL

More information

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?

More information

BD FACSComp Software Tutorial

BD FACSComp Software Tutorial BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS

More information

The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee

The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee January, 1999 From: To: Subject: The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee NIAID DAIDS Flow Cytometry Laboratories Comparison study information for labs wishing to switch

More information

Introduction to Flow Cytometry

Introduction to Flow Cytometry Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer

More information

Immunophenotyping peripheral blood cells

Immunophenotyping peripheral blood cells IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune

More information

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Compensation and gating

More information

Compensation in Flow Cytometry

Compensation in Flow Cytometry Compensation in UNIT.4 The term compensation, as it applies to flow cytometric analysis, refers to the process of correcting for fluorescence spillover, i.e., removing the signal of any given fluorochrome

More information

Three Rules for Compensation Controls

Three Rules for Compensation Controls From FlowJo s Daily Dongle Three Rules for Compensation Controls First and foremost, there must be a single-stained control for every parameter in the experiment! In addition, there are three rules for

More information

Titering Antibodies INTRODUCTION Materials Procedure for titering antibodies to extracellular antigens A. Directly conjugated antibodies

Titering Antibodies INTRODUCTION Materials Procedure for titering antibodies to extracellular antigens A. Directly conjugated antibodies C.C. Stewart, S.J. Stewart, Titering Antibodies. In: Current Protocols in Cytometry, (J.P. Robinson, Z. Darzynkiewicz, P. Dean. L. Dressler, P.Rabinovitch, C. Stewart, H. Tanke, L. Wheeless, eds.) J.Wiley

More information

Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T

Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom Cat. Reference: STP-25T Reagent provided:: 25 Stepcount tubes for 25 test INTENDED USE. Immunostep

More information

Introduction to Flow Cytometry

Introduction to Flow Cytometry Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: daisy_kuo@bd.com BDBiosciences Application Examples

More information

Fluorescence Compensation. Jennifer Wilshire, PhD jennifer@flowjo.com

Fluorescence Compensation. Jennifer Wilshire, PhD jennifer@flowjo.com Fluorescence Compensation Jennifer Wilshire, PhD jennifer@flowjo.com Outline 1. Compensation Basics 2. Practical Applications -selecting comp controls (cells, stains?) 3. Multicolor Issues -spreading due

More information

Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose

Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose Selecting Reagents for Multicolor Flow Cytometry Holden Maecker and Joe Trotter BD Biosciences, San Jose Contents 1 The basics: Know your instrument 2 Fluorochromes: Go for the bright 3 Colors and specificities:

More information

COMPENSATION MIT Flow Cytometry Core Facility

COMPENSATION MIT Flow Cytometry Core Facility COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,

More information

Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer

Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Presented by Clare Rogers, MS Senior Marketing Applications Specialist BD Biosciences 23-13660-00 Webinar Overview Multicolor

More information

Standardization, Calibration and Quality Control

Standardization, Calibration and Quality Control Standardization, Calibration and Quality Control Ian Storie Flow cytometry has become an essential tool in the research and clinical diagnostic laboratory. The range of available flow-based diagnostic

More information

Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction

Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction Application Note 10 Measurement of Cell Recovery After Sorting with a Catcher-Tube-Based Cell Sorter Introduction In many experiments using sorted cells, it is important to be able to count the number

More information

Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers

Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers BD Biosciences, San Jose, CA Contents 2 Introduction to compensation 5 Tips for setting up compensation 10 Tools for easy,

More information

MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:

MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as: INTRODUCTION Flow Cytometry involves the use of a beam of laser light projected through a liquid stream that contains cells, or other particles, which when struck by the focused light give out signals

More information

Minimal residual disease detection in Acute Myeloid Leukaemia on a Becton Dickinson flow cytometer

Minimal residual disease detection in Acute Myeloid Leukaemia on a Becton Dickinson flow cytometer Minimal residual disease detection in Acute Myeloid Leukaemia on a Becton Dickinson flow cytometer Purpose This procedure gives instruction on minimal residual disease (MRD) detection in patients with

More information

WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS

WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS FOR IN-VITRO DIAGNOSTIC USE INVITROGEN CAL-LYSE TM Lysing Solution WHOLE BLOOD LYSING SOLUTION FOR FLOW CYTOMETRIC APPLICATIONS CAL-LYSE TM CATALOG No. GAS-010 250 tests 25 ml CATALOG No. GAS-010S-100

More information

These particles have something in common

These particles have something in common These particles have something in common Blood cells Chromosomes Algae Protozoa Certain parameters of these particles can be measured with a flow cytometer Which parameters can be measured? the relative

More information

Introduction to Flow Cytometry:

Introduction to Flow Cytometry: Introduction to Flow Cytometry: A Learning Guide Manual Part Number: 11-11032-01 April, 2000 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 1-800-448-2347 Introduction to Flow Cytometry: A Learning

More information

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy.

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. At the end of this session, the participant will be able to: 1. Describe the components of a flow cytometer 2. Describe the gating

More information

CyAn : 11 Parameter Desktop Flow Cytometer

CyAn : 11 Parameter Desktop Flow Cytometer CyAn : 11 Parameter Desktop Flow Cytometer Cyan ADP 3 excitation lines 488nm, 635nm, and UV or violet 11 simultaneous parameters FSC, SSC, and 7-9 colors with simultaneous width, peak, area, and log on

More information

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer Introduction Standard methods for isolating and detecting leukocytes

More information

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each

More information

RPCI 004 v.002 Staining Procedure For all Directly Conjugated Reagents (Whole Blood Method)

RPCI 004 v.002 Staining Procedure For all Directly Conjugated Reagents (Whole Blood Method) Immune Tolerance Network RPCI 004 v.002 Staining Procedure For all Directly Conjugated Reagents (Whole Blood Method) Author: Paul Wallace, Director, RPCI Laboratory of Flow Cytometry Approved by: Paul

More information

123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.

123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.

More information

Boundary-breaking acoustic focusing cytometry

Boundary-breaking acoustic focusing cytometry Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory

More information

Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking

Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Mark KuKuruga Senior Technical Applications Specialist BD Biosciences 23-14462-00 Outline

More information

Standard Operating Procedure

Standard Operating Procedure 1.0 Purpose: 1.1 The characterisation of of main leukocyte subsets in peripheral blood cells from mice by flow cytometry. Reliable values of frequencies of leukocyte clusters are very much dependent on

More information

Flow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement

Flow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement What is Flow Cytometry? Flow Cytometry Basic Principle and Applications BD Biosciences Daisy Kuo (daisy_kuo@bd.com) Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements are made on cells,

More information

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University

More information

Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer

Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Contents 2 Thresholds 2 Setting the Threshold When analyzing small particles, defined as particles smaller than 3.0 µm, on the

More information

Compensation: An Instrumental Perspective

Compensation: An Instrumental Perspective Compensation: An Instrumental Perspective Why Digital? Visualization Issues Boston User s Group September 10, 2003 Joe Trotter Joe_Trotter@bd.com Review: Instrument Sensitivity Measuring Sensitivity -

More information

Principles of Flowcytometry

Principles of Flowcytometry Objectives Introduction to Cell Markers: Principles of Flowcytometry Michelle Petrasich NZIMLS Scientific Meeting August 24, 2010, Paihia What are cell markers How do we detect them Production of Monoclonal

More information

Spherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1

Spherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1 78 Irma Lee Circle, Unit 0, Lake Forest, Illinois 00 SPHERO TM Technical Note STN- Rev B 008 DETERMINING PMT LINEARITY IN FLOW CYTOMETERS USING THE SPHERO TM PMT QUALITY CONTROL EXCEL TEMPLATE Introduction

More information

EdU Flow Cytometry Kit. User Manual

EdU Flow Cytometry Kit. User Manual User Manual Ordering information: (for detailed kit content see Table 2) EdU Flow Cytometry Kits for 50 assays: Product number EdU Used fluorescent dye BCK-FC488-50 10 mg 6-FAM Azide BCK-FC555-50 10 mg

More information

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each

More information

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required.

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required. PROTOCOL Immunostaining for Flow Cytometry 1850 Millrace Drive, Suite 3A Eugene, Oregon 97403 Rev.0 Background The combination of single cell analysis using flow cytometry and the specificity of antibody-based

More information

Compensation Basics - Bagwell. Compensation Basics. C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc.

Compensation Basics - Bagwell. Compensation Basics. C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc. Compensation Basics C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc. 2003 1 Intrinsic or Autofluorescence p2 ac 1,2 c 1 ac 1,1 p1 In order to describe how the general process of signal cross-over

More information

Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and

Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and Using BD FACSDiva CST To Evaluate Cytometer Performance, Create Custom Assay Settings and Implement Cross-Instrument and Cross-Site Standardization of Assays PART 2 Alan M. Stall Director, Advanced Cytometry

More information

Compensation Controls Data Visualization Panel Development Gating Quantum Dots

Compensation Controls Data Visualization Panel Development Gating Quantum Dots Compensation Controls Data Visualization Panel Development Gating Quantum Dots FITC Single Stain Control FITC PE Argon Laser FL1 FL2 450 500 550 600 FITC Compensation Control Unwanted signal detected in

More information

ArC Amine Reactive Compensation Bead Kit

ArC Amine Reactive Compensation Bead Kit ArC Amine Reactive Compensation Bead Kit Catalog no. A1346 Table 1. Contents and storage information. Material Amount Composition Storage Stability ArC reactive beads (Component A) ArC negative beads (Component

More information

Islet Viability Assessment by Single Cell Flow Cytometry

Islet Viability Assessment by Single Cell Flow Cytometry Islet Viability Assessment by Single Cell Flow Cytometry Page 1 of 8 Purpose: To comprehensively assess the viability of the islet cell preparation prior to transplantation. Tissue Samples: A sample containing

More information

Introduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size

Introduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size Introduction to Flow Cytometry -- BD LSR II Daisy Kuo 郭 正 佼 Assistant Product Manager BD Biosciences Daisy_kuo@bd.com What is Flow Cytometry? Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements

More information

Instrument Characterization and Performance Tracking for Digital Flow Cytometers

Instrument Characterization and Performance Tracking for Digital Flow Cytometers Instrument Characterization and Performance Tracking for Digital Flow Cytometers BD Biosciences Cytometer Set-up & Tracking (CS&T) System For Research Use Only. Not for use in diagnostic or therapeutic

More information

Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-806)

Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-806) Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-86) The Hoechst purification was established for murine hematopoietic stem cells (HSC)

More information

Setting up and calibration of a flow cytometer for multicolor immunophenotyping

Setting up and calibration of a flow cytometer for multicolor immunophenotyping Journal of Biological Regulators and Homeostatic Agents Setting up and calibration of a flow cytometer for multicolor immunophenotyping J. KRAAN 1, J.W. GRATAMA 1, M. KEENEY 2, J.L. D HAUTCOURT 3 1 Department

More information

Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit

Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit 2 Analyzing Samples for CD34 Enumeration Using the BD Stem Cell Enumeration Kit Presented by Ellen Meinelt, MS MLS(ASCP) CM, Technical Applications Specialist and Calin Yuan, Product Course Developer,

More information

BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY

BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY IMPORTANT NOTES BEFORE READING THIS TUTORIAL This is a very expensive piece of equipment so PLEASE treat it with respect! After you are done

More information

Absolute counting using flow cytometry. Bright Count Microspheres

Absolute counting using flow cytometry. Bright Count Microspheres Absolute counting using flow cytometry Single-platform system for absolute counts I PRODUCT MANUAL [REF] IQP-391 [REF] IQP-392 s 50 tests s 100 tests [IVD] In Vitro Diagnostic medical device IQ Products

More information

Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility

Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility The mention of commercial products, their sources, or their use in connection

More information

Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility

Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility 10.1071/RD14043_AC CSIRO 2015 Supplementary Material: Reproduction, Fertility and Development, 2015, 27(8), 1187 1196. Supplementary Material Free-radical production after post-thaw incubation of ram spermatozoa

More information

Flow Cytometry A Basic Overview

Flow Cytometry A Basic Overview Flow Cytometry A Basic Overview Overview Flow cytometry is a powerful technology for investigating many aspects of cell biology and for isolating cells of interest. Flow cytometry utilizes highly focused,

More information

Cell Cycle Tutorial. Contents

Cell Cycle Tutorial. Contents Cell Cycle Tutorial Contents Experimental Requirements...2 DNA Dyes...2 Protocols...3 PI Parameter & Analysis Setup...4 PI Voltage Adjustments...6 7-AAD Parameter Setup...6 To-Pro3 Parameter Setup...6

More information

How To Read Flow Cytometry Data

How To Read Flow Cytometry Data 26 Nature Publishing Group http://www.nature.com/natureimmunology Interpreting flow cytometry data: a guide for the perplexed Leonore A Herzenberg, James Tung, Wayne A Moore, Leonard A Herzenberg & David

More information

CRITICAL ASPECTS OF STAINING FOR FLOW CYTOMETRY

CRITICAL ASPECTS OF STAINING FOR FLOW CYTOMETRY 页 码,1/6 CRITICAL ASPECTS OF STAINING FOR FLOW CYTOMETRY From Givan, A.L. (2000), chapter in In Living Color: Protocols in Flow Cytometry and Cell Sorting (R. Diamond and S. DeMaggio, eds). Springer, Berlin,

More information

CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping

CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping CyFlow SL Healthcare Immunology Portable FCM System for 3-colour Immunophenotyping 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results

More information

Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...

Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation... Immunophenotyping Flow Cytometry Tutorial Contents Experimental Requirements...1 Data Storage...2 Voltage Adjustments...3 Compensation...5 Experimental Requirements For immunophenotyping with FITC and

More information

CyTOF2. Mass cytometry system. Unveil new cell types and function with high-parameter protein detection

CyTOF2. Mass cytometry system. Unveil new cell types and function with high-parameter protein detection CyTOF2 Mass cytometry system Unveil new cell types and function with high-parameter protein detection DISCOVER MORE. IMAGINE MORE. MASS CYTOMETRY. THE FUTURE OF CYTOMETRY TODAY. Mass cytometry resolves

More information

CFSE Cell Division Assay Kit

CFSE Cell Division Assay Kit CFSE Cell Division Assay Kit Item No. 10009853 Customer Service 800.364.9897 * Technical Support 888.526.5351 www.caymanchem.com TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 4 Precautions

More information

INSIDE THE BLACK BOX

INSIDE THE BLACK BOX FLOW CYTOMETRY ESSENTIALS INSIDE THE BLACK BOX Alice L. Givan Englert Cell Analysis Laboratory of the Norris Cotton Cancer Center Dartmouth Medical School HOW NOT TO BE A FLOW CYTOMETRIST Drawing by Ben

More information

Lab 2. Isolation of mononuclear cells from peripheral blood and separation into subpopulations

Lab 2. Isolation of mononuclear cells from peripheral blood and separation into subpopulations Lab 2 Isolation of mononuclear cells from peripheral blood and separation into subpopulations Supervisors: Sissela Broos sissela.broos@immun.lth.se tel: 222 96 78 Niclas Olsson niclas.olsson@immun.lth.se

More information

FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data. Monisha Sundarrajan, PhD BD Biosciences

FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data. Monisha Sundarrajan, PhD BD Biosciences FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data Monisha Sundarrajan, PhD BD Biosciences 23 13795 00 Agenda Introduction Overview of BD CBA bead based immunoassays FCAP

More information

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that

More information

A practical approach to multicolor flow cytometry for immunophenotyping

A practical approach to multicolor flow cytometry for immunophenotyping Journal of Immunological Methods 243 (2000) 77 97 www.elsevier.nl/ locate/ jim A practical approach to multicolor flow cytometry for immunophenotyping Nicole Baumgarth, Mario Roederer a, b, * a Department

More information

Novel Use of the BD FACS TM SPA to Automate Custom Monoclonal Antibody Panel Preparations for Immunophenotyping

Novel Use of the BD FACS TM SPA to Automate Custom Monoclonal Antibody Panel Preparations for Immunophenotyping Cytometry Part B (Clinical Cytometry) 66B:40 45 (2005) TECHNICAL INNOVATION Novel Use of the BD FACS TM SPA to Automate Custom Monoclonal Antibody Panel Preparations for Immunophenotyping Abigail S. Kelliher,

More information

Science is hard. Flow cytometry should be easy.

Science is hard. Flow cytometry should be easy. Science is hard. Flow cytometry should be easy. TABLE OF CONTENTS 1 INTRODUCTION TO BD ACCURI C6 SOFTWARE... 1 1.1 Starting BD Accuri C6 Software... 1 1.2 BD Accuri C6 Software Workspace... 2 1.3 Opening

More information

DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES

DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES CLAUDIO ORTOLANI UNIVERSITY OF URBINO - ITALY SUN TZU (544 b.c. 496 b.c) SUN TZU (544 b.c. 496 b.c.) THE ART OF CYTOMETRY

More information

Flow cytometry basics fluidics, optics, electronics...

Flow cytometry basics fluidics, optics, electronics... Title Flow cytometry basics fluidics, optics, electronics... RNDr. Jan Svoboda, Ph.D. Cytometry and Microscopy Core Facility IMB, CAS, v.v.i Vídeňská 1083 Fluorescence Fluorescence occurs when a valence

More information

FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS. By: Douaa Moh. Sayed

FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS. By: Douaa Moh. Sayed FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS By: Douaa Moh. Sayed Definition Flow cytometry is a technique for counting, examining, and sorting microscopic particles suspended in a stream of fluid. It allows

More information

Flow Cytometry. flow cytometer DNA apoptosis ph

Flow Cytometry. flow cytometer DNA apoptosis ph Flow Cytometry flow cytometer DNA apoptosis ph flow cytometry flow cytometer (a) (b) cell counting instrument (c) 1960 ink-jet technology 17 19 1940 1950 fluorescence microscopy fluorescent dye DNA polyclonal

More information

UNIVERSITY OF PÉCS MEDICAL SCHOOL FLOW CYTOMETRY AND CELL SEPARATION BIOPHYSICS 2. 2015 4th March Dr. Beáta Bugyi Department of Biophysics Flow cytometry and cell separation FLOW = STREAM OF FLUID in a

More information

Utility of flow cytometric κ and λ light chain analysis of peripheral blood

Utility of flow cytometric κ and λ light chain analysis of peripheral blood JBUON 2015; 20(5): 1322-1326 ISSN: 1107-0625, online ISSN: 2241-6293 www.jbuon.com E-mail: editorial_office@jbuon.com ORIGINAL ARTICLE Utility of flow cytometric κ and λ light chain analysis of peripheral

More information

Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation

Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation Roche Applied Science Technical Note No. LC 19/2004 Purpose of this Note Color The LightCycler System is able to simultaneously detect and analyze more than one color in each capillary. Due to overlap

More information

ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis

ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This

More information

Below is a list of things you should be aware of before you schedule your sort.

Below is a list of things you should be aware of before you schedule your sort. Sorting at the Flow Cytometry Facility At the present time, the assistance of a trained cell sorter operator is needed for all sorting applications. As such, if you are planning a first time sort you will

More information

Uses of Flow Cytometry

Uses of Flow Cytometry Uses of Flow Cytometry 1. Multicolour analysis... 2 2. Cell Cycle and Proliferation... 3 a. Analysis of Cellular DNA Content... 4 b. Cell Proliferation Assays... 5 3. Immunology... 6 4. Apoptosis... 7

More information

Flow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE

Flow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE Flow Data Analysis Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA Flow Data Analysis From Cells to Data - data measurement and standard Data display and gating - Data Display options -

More information

Chapter 6. Antigen-Antibody Properties 10/3/2012. Antigen-Antibody Interactions: Principles and Applications. Precipitin reactions

Chapter 6. Antigen-Antibody Properties 10/3/2012. Antigen-Antibody Interactions: Principles and Applications. Precipitin reactions Chapter 6 Antigen-Antibody Interactions: Principles and Applications Antigen-Antibody Properties You must remember antibody affinity (single) VS avidity (multiple) High affinity: bound tightly and longer!

More information

Zecotek S Light Projection Network Marketing

Zecotek S Light Projection Network Marketing White Paper Zecotek Visible Fiber Laser Platform Enabling the future of laser technology Zecotek Photonics Inc. (TSX- V: ZMS; Frankfurt: W1I) www.zecotek.com is a Canadian photonics technology company

More information

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit

Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit Rat Creatine Kinase MB isoenzyme,ck-mb ELISA Kit Catalog No: E0479r 96 Tests Operating instructions www.eiaab.com FOR RESEARCH USE ONLY; NOT FOR THERAPEUTIC OR DIAGNOSTIC APPLICATIONS! PLEASE READ THROUGH

More information

BD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin

BD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin BD LSR II and FACSDiVa Software, Basel Dr. Norbert Leclere,, Berlin Overview Electronics Covers Fluidics Connectors Controlpanel Sample Port On/Off The Control Panel Ease of Use The basic procedures to

More information

March 19, 2014. Dear Dr. Duvall, Dr. Hambrick, and Ms. Smith,

March 19, 2014. Dear Dr. Duvall, Dr. Hambrick, and Ms. Smith, Dr. Daniel Duvall, Medical Officer Center for Medicare, Hospital and Ambulatory Policy Group Centers for Medicare and Medicaid Services 7500 Security Boulevard Baltimore, Maryland 21244 Dr. Edith Hambrick,

More information

Whole Blood Flow Cytometry

Whole Blood Flow Cytometry Whole Blood Flow Cytometry y Nailin Li Department t of Medicine, i Clinical i l Pharmacology Unit Karolinska Institute/University Hospital, 171 76 Stockholm Department of Pathology & Pathophysiology Zhejiang

More information

Fluorescent dyes for use with the

Fluorescent dyes for use with the Detection of Multiple Reporter Dyes in Real-time, On-line PCR Analysis with the LightCycler System Gregor Sagner, Cornelia Goldstein, and Rob van Miltenburg Roche Molecular Biochemicals, Penzberg, Germany

More information

Becton-Dickinson FACSCalibur Flow Cytometer Instructions

Becton-Dickinson FACSCalibur Flow Cytometer Instructions Start-up 1. Turn on flow cytometer by pushing green button on right side of instrument, toward the back. Give it 30 minutes to warm up before collecting data. 2. Turn on monitor. 3. Turn on computer. If

More information

International Beryllium Conference, Montreal, Canada March 10, 2005

International Beryllium Conference, Montreal, Canada March 10, 2005 Alternative Lymphocyte Proliferation Tests: BrdU and Flow Cytometry Based Tests International Beryllium Conference, Montreal, Canada March 10, 2005 Tim K. Takaro Department of Environmental and Occupational

More information

Too Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry

Too Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry Too Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry by Debby R. Walser-Kuntz Biology Department Carleton College, Northfield, MN Taylor goes in to see her doctor, Dr. Chavez,

More information

a Beckman Coulter Life Sciences: White Paper

a Beckman Coulter Life Sciences: White Paper a Beckman Coulter Life Sciences: White Paper Violet-Excited nim-da Allows Efficient and Reproducible Cell Cycle Analysis on the Gallios Flow Cytometer Authors: Valdez, Ben 1. Carr, Karen 2. Norman, John

More information

Annexin V-EGFP Apoptosis Detection Kit

Annexin V-EGFP Apoptosis Detection Kit ab14153 Annexin V-EGFP Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of apoptosis in various samples This product is for research use only and is not intended

More information

Flow Cytometry Protocols

Flow Cytometry Protocols METHODS IN MOLECULAR BIOLOGY TM TM Volume 263 Flow Cytometry Protocols SECOND EDITION Edited by Teresa S. Hawley Robert G. Hawley 1 Flow Cytometry An Introduction Alice L. Givan Summary A flow cytometer

More information

Flow Cytometric Diagnosis of Low Grade B-cell Leukemia/Lymphoma

Flow Cytometric Diagnosis of Low Grade B-cell Leukemia/Lymphoma Flow Cytometric Diagnosis of Low Grade B-cell Leukemia/Lymphoma Maryalice Stetler-Stevenson, M.D., Ph.D. Flow Cytometry Unit, Laboratory of Pathology, DCS, NCI,NIH DEPARTMENT OF HEALTH & HUMAN SERVICES

More information

2. SUMMARY AND EXPLANATION

2. SUMMARY AND EXPLANATION M3630IV_C 2003-09-01 1 / 29 English Stem-Kit Reagents 1-29 1. INTENDED USE 2 2. SUMMARY AND EXPLANATION 2 3. PRINCIPLE OF TEST 3 4. CLINICAL RELEVANCE 3 5. REAGENTS AND CONTENTS 3 6. STATEMENT OF WARNING

More information

TITLE: TRU-COUNT ENUMERATION OF T-CELLS IN WHOLE

TITLE: TRU-COUNT ENUMERATION OF T-CELLS IN WHOLE TITLE: TRU-COUNT ENUMERATION OF T-CELLS IN WHOLE PRINCIPLES OF THE PROCEDURE: When whole blood is added to the reagent, the fluorochrome labeled antibodies in the reagent bind specifically to leukocyte

More information

Reagent kit for the flow cytometric quantitation of platelet associated immunoglobulin. Antibodies and reagents for 25 tests

Reagent kit for the flow cytometric quantitation of platelet associated immunoglobulin. Antibodies and reagents for 25 tests page 1 of 8 Reagent kit for the flow cytometric quantitation of platelet associated immunoglobulin Antibodies and reagents for 25 tests Key to symbols used Glycotope Biotechnology GmbH Czernyring 22 69115

More information