Flow Cytometry 101. Jodi Moore, PhD Jan 2012 Goldenson 113

Size: px
Start display at page:

Download "Flow Cytometry 101. Jodi Moore, PhD Jan 2012 Goldenson 113"

Transcription

1 Flow Cytometry 101 Jodi Moore, PhD Jan 2012 Goldenson 113

2 Origins:? Dye Chemistry s Electronics Microscopy Automated Clinical Cytology ID Apps Computers

3 Cytometry vs. Flow Cytometry Microspectrophotometry Confocal Microscopy Scanning Cytometry Image Analysis LocalizaBon of anbgen is possible Poor enumerabon of cell subtypes within a populabon Limited number of simultaneous measurements? LocalizaBon of anbgen is not possible Detailed subset idenbficabon and enumerabon MulBple correlated parameters measured simultaneously Can analyze many cells rapidly

4 Why is Flow Cytometry a Powerful Research Tool?? Single Cell pro caspase caspase EGF Ca 2+ Ca 2+ Ca 2+ Ca 2+ Ca 2+ Ca 2+ Ca 2+ Ca 2+ Ca 2+ Stat Stat Stat JAK Correlated Measurements of: Lineage/Phenotype Cellular FuncBon Apoptosis/Necrosis Enzyme AcBvaBon Intracellular Proteins DNA Content/Ploidy Cell Cycle Protein PhosphorylaBon Protein- Protein InteracBons Gene Expression SorBng Highly Pure PopulaBon

5 Requirements for Flow Cytometry Any suspended parbcle or cell from uM is suitable for analysis. Sample must be in a single- cell suspension! Samples are labeled with fluorescent probes or fluorophore- conjugated anbbodies, and the free reagent is washed away. ExcitaBon sources (lasers) must match the excitabon maxima of the fluorochromes, and collecbon opbcs must correlate with the emission spectra of the fluorochromes.

6 What Does a Flow Cytometer Do? Rapid detecbon and quanbtabon of light sca7er and rela8ve fluorescence intensity of a parbcle or cell. Sca_er properbes (FSC and SSC) give us informabon about the relabve size and internal complexity of the parbcle, respecbvely. Fluorescence can be generated from a number of sources, intrinsic and extrinsic to cells.

7 How Does a Flow Cytometer Work? Fluidics: Introduce and focus the cells for interrogabon. Op8cs:? ExcitaBon- Laser and lenses, which shape and focus the laser beam. CollecBon- CollecBon lens (fibers) and a system of opbcal mirrors and filters that route specified wavelengths of collected light to opbcal detectors. Electronics: Convert the opbcal signals to proporbonal electronic signals and digibze them for computer analysis. Monitor and control the operabon of the cytometer.

8 Overview

9 Fluidics: Sheath Flow Principle 1. HYDRODYNAMIC FOCUSING: Refers to the restricbon of the cells to the center of the stream. It is facilitated by the reducbon in the cross- secbonal area of the vessel. Furthermore, the increased speed of the sheath stream versus that of the sample core stream at the injecbon point impinges on the core stream, thereby reducing its cross- secbonal width. FLOW CELL 2. LAMINAR FLOW: CondiBon whereby two adjacent, non- turbulent streams of fluid maintain their separabon with minimal mixing. Sheath Sheath Sample Central Core

10 OpBcs ExcitaBon OpBcs: Lasers which excite fluorochromes, fluorescent proteins, detecbon reagents in the assay Standard flavors: 488nm, 405nm,644nm and UV More exobc flavors: 532nm, 561nm, 592nm, 457nm Fiber opbc cables that carry beams to steering prisms, which then direct the beams to the flow cell and fluid stream; unless air launched CollecBon OpBcs: Fiber opbc cables that direct the emi_ed light to the appropriate emission block Filters/Mirrors that direct the signals in the emission block to the appropriate PMT

11 Laser Delays cell enters and leaves the first laser intercept blue t=0 cell enters and leaves the second laser intercept violet t= cell enters and leaves the third laser intercept Bme (μsec) yellow t= SETTING: TIME DELAY

12 OpBcal Filters Longpass Shortpass Bandpass LP 500 SP /50 Transmits all wavelengths greater than specified wavelength Transmits all wavelengths less than specified wavelength Transmits a specific band of wavelengths i.e. 500/50 = nm (500 +/- 25)

13 Dichroic Lenses Can be a long or short pass filter Filter is placed at a 45 o angle to the incident light Part of the light is reflected at 90 o to the incident light, and part of the light is transmi_ed Detector 1 Detector 2 Dichroic Lens

14 BD s Detector Arrays: Default ConfiguraBon Lowest Energy InnovaBve design for increased sensibvity is based on the principle that light reflecbon is more efficient than light transmission, resulbng in detecbon with minimum light loss.

15 Photons In How Is Fluorescence Detected? Photocathode Dynodes Anode e- Focusing Electrode Secondary Emission Amplified Signal Out: Current The role of the PMT is to convert an opbcal signal, comprised of photons, into a proporbonate amplified electrical signal, or current PhotomulBplier tubes (PMTs) ublize the photoelectric effect, whereby photons striking the photocathode sbmulate the producbon of photoelectrons, which are mulbplied as they pass through a series of electrodes (dynodes) in a process known as secondary emission. At the anode, the accumula8on of charge results in a sharp current pulse, which is transmi7ed to the electronics. This signal is then processed though log or linear amplifiers and converted to a voltage pulse. PMTs are ublized to detect SSC and all FL signals. FSC is detected by a less sensibve photo diode.

16 CreaBon of a Voltage Pulse Laser InterrogaBon Stream Voltage A voltage pulse is created when a cell or parbcle enters the laser beam and starts to sca_er light or fluoresce. The PMT converts light into an electrical pulse. Laser InterrogaBon Laser InterrogaBon Voltage Voltage Time Time Max. Height The highest part of the pulse occurs when the cell is in the center of the beam and the maximum amount of sca_er or fluorescence is achieved. The magnitude of the voltage pulse depends on the number of photons detected, the PMT voltage, and the amplifier gain. Time

17 Electronics Convert analog signals to proporbonal digital signals Compute area and height for each pulse Perform compensabon, calculate rabos and width Interface with the computer for data transfer Perform sorbng decisions including gabng, conflict resolubon, and drop charging

18 Pulse Measurements voltage area height The pulse processors measure pulses by three characterisbcs: height, area, and width. Pulse height is the maximum digibzed intensity measured Pulse area is an integrabon of the digibzed measures over Bme Pulse width calculates: Area/height x 64,000 threshold Bme window gate: width

19 Current produced by the PMT generates a voltage pulse; these electrical signals are digibzed using Analog to Digital Converters (ADCs). Data CollecBon & Display DigiBzed signals are processed by addibonal electronics that calculate peak, area, and width signals.. and generates histograms, dot plots, and FCS 3.0 files (Flow Cytometry Standard). Time FL1 Time Data Processor FL2 Time FL3

20 Signal GeneraBon in Flow Light Sca_er Signals Forward Sca_er Side Sca_er Fluorescence

21 What are Light Sca_er ProperBes? Side Scatter Detector α Cell Complexity Incident Light Source Forward Scatter Detector α Cell Surface Area* Forward Sca_er (FSC/small angle sca_er) diffracted light Detected along axis of incident light in the forward direcbon (Angles o ) Related to relabve cell surface area and refracbve index of the cell; not a quanbtabve biological equivalent to cell size* Oten used as a threshold to trigger signal processing Side Sca_er (SSC/wide angle sca_er) reflected and refracted light Detected at 90 to the laser beam (Angles o ) Related to cell granularity and internal complexity Occurs at any interface where there is a change in the refracbve index

22 PopulaBons in Flow Cytometry Each event that we measure is a single cell in flow cytometry, however, what we actually evaluate in flow cytometry is POPULATIONS or DISTRIBUTIONS of cells. We employ pa_ern recognibon regularly in discriminabng flow cytometry data.

23 Typical Sca_er Display Increasing Size Side Sca_er ApoptoBc Cells Dead Cells/ Noise Larger Cells Increasing Granularity (SSC) Live Cells *RelaBve Differences THRSHLD Forward Sca_er (FSC)

24 What InformaBon Can be Gained from Measurement of Sca_er ProperBes? Side Scatter Neutrophils Monocytes Lymphocytes Dead Viable Forward Scatter Cellular Subset DifferenBaBon/ RELATIVE Size/Internal substructure Cell Viability

25 When we return... Fluorescence

26 Fluorescence: Electronic Energy TransiBon HEAT, VIBRATION OUT (S n ) Excited Singlet State (S 1 ) Relaxed Singlet State E Excitation E Emission LIGHT IN HIGHER ENERGY LIGHT OUT LOWER ENERGY Ground State (S 0 )

27 Fluorescence Stoke s Shit = Δλ peak 525nm- 480nm=45nm λ= 488nm λ= 525nm Ex. Max = 480nm Em. Max = 525nm Incident Laser Light FITC Emi_ed Fluorescence CharacterisBc excitabon and emission spectra for each fluorochrome ExcitaBon and emission curves depict the probability of absorpbon or emission of a photon by the fluorochrome molecule at any given wavelength

28 Fluorescence Quantum Yield (Φ) Φ = Number of photons emitted Number of photons absorbed Defines the efficiency of the fluorescence process Maximum yield is 1.0 (100%) Measured compared to a standard; the quinine salt quinine sulfate in a sulfuric acid solubon is a common fluorescence standard, Φ = 0.58 GFP Φ = 0.79, AF488 Φ = 0.92, mcherry Φ = 0.22

29 Molar ExBncBon Coefficient (ε) Measures of the light absorbing capacity of a dye at a given wavelength per molar concentrabon Dyes with large molar exbncbon coefficients are efficient absorbers GFP ε = 55,000 M - 1 cm - 1, AF488 ε = 71,000 M - 1 cm - 1 BRIGHTNESS is proporbonate to the product of the exbncbon coefficient and quantum yield Brightness = ε*φ/1000

30 Typical Fluorescence Display Count NegaBve PopulaBon Dimmer PosiBve PopulaBon Brighter Channel Number (x- axis: represents the parameters signal value in channel numbers)

31 250 Emitted Fluorescence Intensity Binding Sites # Cells Number of Events FITC FITC FITC FITC FITC FITC FITC FITC FITC FITC 50 0 Negative GFP 90.4 Positive GFP <FITC-A>: GFP Mean/Median Fluorescence Intensity (x- axis: Mean or Median Fluorescence Intensity = MFI)

32 MulBple Correlated Parameters Single PosiBve PE PopulaBon Double PosiBve PopulaBon PE FL NegaBve PopulaBon FITC FL Single PosiBve FITC PopulaBon

33 Typical Fluorescence Display Log amplification is normally used for fluorescence displays because it expands weak signals and compresses strong signals, resulting in a distribution that is easy to display on a histogram. Linear scaling is preferable where there is not such a broad range of signals.

34 Data Display Histograms (Single Parameter) Overlays Contour Plot Density Plot Dot Plot (Dual Parameters)

35 Cell Cycle 1000 G 0 /G # Cells S G 2 M FL2-A Dean/Jett/Fox: %G1 = 55.8; %S = 32; %G2 = Cell Cycle DNA Dye Cycling Control Event Count: 8105 A stoichiometric DNA dye is quantitatively representative of the amount of DNA present. The Encyclopedia of Science Online

36 GaBng P1 SINGLETS P1 SINGLETS Used to idenbfy subsets of data or populabons. Only looking at data from your populabon of interest. Gate- defined populabons can be used to generate stabsbcs and limit the number of events displayed, collected (stopping gate) or saved (storage gate). GaBng is only used as a visualizabon tool during acquisibon or analysis; all events are saved unless using a storage gate to limit events saved.

37 ReporBng Values in Flow Cytometry Data # Cells 400 # Cells APC-A Ungated Specimen_001_DAPI only.fcs Event Count: APC-A Ungated Specimen_001_bTUB 10DF.fcs Event Count: % % = 91% APC+

38 ReporBng Values in Flow Cytometry Data The median is less sensibve to extreme scores than the mean and this makes it a be_er measure than the mean for highly skewed distribubons % of Max /68 = 13- fold increase in MFI (median fluorescence intensity) APC-A Specimen_001_DAPI only.fcs 68.2 Specimen_001_bTUB 10DF.fcs 862

39 ReporBng Values in Flow Cytometry Data 10 5 Hypodiploid 31.5 G1A 24.7 SA 21.7 G2MA DsRed-A 10 3 DsRed-A G1 3.5 S 2.75 G2M K 100K 150K 200K 250K Pacific Blue-A Singlets EL4_6h 50uM Viable VV & YOPRO 10df.fcs Event Count: FITC-A Ungated Specimen_001_GFP,3a,Venus,3a,Cherry,3a,Cerulean,3a,Sapphire.fcs Event Count: 30000

40 ReporBng Values in Flow Cytometry Data # Cells FL2-A Dean/Jett/Fox: %G1 = 55.8; %S = 32; %G2 = Cell Cycle Cycling Control Event Count: 8105

41 ReporBng Values in Flow Cytometry Data 250K 200K Ratio: FITC-A/PerCP-A 250K 200K 150K 100K 50K Time live 05 FURA X Calcium_VEGF1,2f,Iono.fcs Event Count: K FITC-A: FLUO4 PerCP-A: FURA RED 150K 100K 50K 0 250K 200K 150K 100K 50K Time Time % Cells Responding Ratio: FITC-A/PerCP-A 200K 150K 100K 50K Time live EC1 05 FURA X Calcium_VEGF1,2f,Iono.fcs Event Count: Ratio: FITC-A/PerCP-A Time Sample 05 FURA X Calcium_VEGF1,2f,Iono.fcs EC1 05 FURA X Calcium_VEGF1,2f,Iono.fcs

42 Common Fluorochrome Emission Spectra PacBlue AF405 ViViolet (V) FITC AF488 Sytox Blue (V) PE AF430 ViAqua (V) PE- TR PE- AF610 PI (V) 7- AAD (V) AF647 APC PE- Cy5 PacOrange APC- Cy5.5 AF700 Cy5 PE- Cy5.5 APC- Cy7 APC- AF750 PE- Cy7

43 Spectral Overlap/Spillover Occurs because fluorescent materials emit over a fairly broad range of wavelengths. Filters do not completely block spillover into adjacent detectors; therefore, a process to subtract the spectral overlap must be performed.

44 Filters: FITC: 530/30 PE: 575/26 PCPCy5.5: 695/40 FITC Spillover Filters: PECy7: 780/60 APC: 660/20 AF700: 730/45 Filters: APCCy7: 780/60 PB: 450/40 FITC Beads Posi8ve Beads = Green Unstained Beads = Blue Spillover = Red

45 FITC/PE Compensation-Log Display FITC only PE only x mean = 4.0 y mean = 40.0 Raw Data y mean = 2.0 x mean = 2.0 Compensated y mean 2.0 x mean 2.0

46 CompensaBon Refers to the process of mathemabcal subtracbon of dye spectral spillover into adjacent collecbon channels (detectors) Serves to align stained populabons in dye space without bias from spectral overlap Requires a NEGATIVE and SINGLE- COLOR controls for each fluorochrome/fp in the assay The posibve control must have enough posibves for an accurate count versus the negabve (at least 25% posibve) and must be as bright (MFI) as your brightest sample 150 Beads versus Cells Hardware (Pre- AcquisiBon Comp) Sotware (Post- AcquisiBon Comp) # Cells Negative Positive PE-A

47 FITC/PE Compensation Log versus Bi-exponential Transformation Also known as logicle displays Incorporates linear scaling for low and negative values together with log scaling for high values Bi-exponential transformation only changes the visualization of the data; doesn t effect gating or change the overlap of two populations Facilitates data display in an aesthetic, intuitive manner

48 Data Spread As A Result of CompensaBon (Digital Data) 575nm 695nm 780nm 719nm 660nm 660nm 660nm 660nm Bright fluorescent signals tend to spread in the dimensions against which they are compensated. Since dim (photon- deficient) fluorescent populabons have higher variance than bright (photon- rich) populabons, they will exhibit more data spread. Symmetry about the axes of the populabons indicates that they are properly compensated. CounBng errors are influenced by signal intensity and the degree of spillover, and are minimized by using bright dyes with as li_le spectral overlap as possible in cells with coordinately expressed markers.

49 NegaBve Values in Flow? How? Fluorescence from which an esbmated value of background light and electronic noise is subtracted, in conjuncbon with measurement error. Photon counbng error in photon- dim channels affects fluorescence spillover and contributes to negabve data spread following compensabon.

50 Facility InstrumentaBon

51 BD FACSAriaIIu 3 Lasers: 405, 488 & 594nm Simultaneous 9- Color Analysis/SorBng 1-4 way sort Temperature control of sample and collecbon tubes SorBng into plates with ACDU AutomaBc CompensaBon calculabon Digital Electronics High Speed/Variable Pressure SorBng

52 BD SORP LSRII w/hts 3 Lasers: 405, 488 & 594nm 2 Sca_er Parameters 12 FL Parameters Tube, 96- & 384- well plate input

53 BD FACSCalibur 2 Lasers: 488nm & 633nm 2 Sca_er Parameters 4 FL Parameters 12x75mm tube input.

54 Stratedigm S1000EX: 4 Lasers, 13 Colors!"#$%&'$()*+*#,%*)-&.).,(/0) Legend: T-Molding perimeter Top shelf with 3 supports Integrated bench light Integrated power strip, 8 outlets Integrated, adjustable monitor arm Deluxe, adjustable keyboard & mouse Tray 7. Integrated roll out shelf for fluidics tanks 8. Drawers Dimensions: Height to bench top: 30 Height to top shelf: 67 Length : 96 Depth: 36 The Future of Flow Cytometry

55 Vocabulary Configuration: The physical location of dichroic mirrors and filters in your detector array; if more than one, the others are given names that identify changes that have been made to the default configuration Acquisition: When you are viewing OR recording events in your sample Stopping Gate: Can be numeric (# of events) or time (800 seconds) that designates when sample recording stops Parameters: The name that designates each PMT detector PMT Voltages: The numbers that adjust the mean channel number or MFI of your population; these are usually adjusted with regard to a negative sample QC: Quality control; procedures performed on the instrument to ensure reproducible, peak performance Morphologic Gates: Gates that can be defined by FSC and SSC clustering patterns Pre-Acquisition Compensation, also Automatic Compensation: Hardware compensation performed pre-acquisition by the instrument Post-Acquisition Compensation: Compensation, done after data is acquired, by a software package CV: Coefficient of variation; a statistical measure of the dispersion of data points; used in QC applications MFI: Mean or Median Fluorescence Intensity, represented by a linear/log channel number Threshold: Establishes a level by which you have determined that an event represents a cell, rather than electronic noise; triggers the cytometer to display events at and above that level FCS File: Flow Cytometry Standard file format (Current v3.0)

56 Ready for a break??

Introduction to flow cytometry

Introduction to flow cytometry Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression

More information

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service

Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?

More information

MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:

MEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as: INTRODUCTION Flow Cytometry involves the use of a beam of laser light projected through a liquid stream that contains cells, or other particles, which when struck by the focused light give out signals

More information

Introduction to Flow Cytometry

Introduction to Flow Cytometry Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: daisy_kuo@bd.com BDBiosciences Application Examples

More information

These particles have something in common

These particles have something in common These particles have something in common Blood cells Chromosomes Algae Protozoa Certain parameters of these particles can be measured with a flow cytometer Which parameters can be measured? the relative

More information

Flow cytometry basics fluidics, optics, electronics...

Flow cytometry basics fluidics, optics, electronics... Title Flow cytometry basics fluidics, optics, electronics... RNDr. Jan Svoboda, Ph.D. Cytometry and Microscopy Core Facility IMB, CAS, v.v.i Vídeňská 1083 Fluorescence Fluorescence occurs when a valence

More information

Introduction to Flow Cytometry

Introduction to Flow Cytometry Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer

More information

BD FACSComp Software Tutorial

BD FACSComp Software Tutorial BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS

More information

Three Rules for Compensation Controls

Three Rules for Compensation Controls From FlowJo s Daily Dongle Three Rules for Compensation Controls First and foremost, there must be a single-stained control for every parameter in the experiment! In addition, there are three rules for

More information

Introduction to Flow Cytometry:

Introduction to Flow Cytometry: Introduction to Flow Cytometry: A Learning Guide Manual Part Number: 11-11032-01 April, 2000 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 1-800-448-2347 Introduction to Flow Cytometry: A Learning

More information

COMPENSATION MIT Flow Cytometry Core Facility

COMPENSATION MIT Flow Cytometry Core Facility COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,

More information

Flow Cytometry A Basic Overview

Flow Cytometry A Basic Overview Flow Cytometry A Basic Overview Overview Flow cytometry is a powerful technology for investigating many aspects of cell biology and for isolating cells of interest. Flow cytometry utilizes highly focused,

More information

CyAn : 11 Parameter Desktop Flow Cytometer

CyAn : 11 Parameter Desktop Flow Cytometer CyAn : 11 Parameter Desktop Flow Cytometer Cyan ADP 3 excitation lines 488nm, 635nm, and UV or violet 11 simultaneous parameters FSC, SSC, and 7-9 colors with simultaneous width, peak, area, and log on

More information

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each

More information

Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking

Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Mark KuKuruga Senior Technical Applications Specialist BD Biosciences 23-14462-00 Outline

More information

INSIDE THE BLACK BOX

INSIDE THE BLACK BOX FLOW CYTOMETRY ESSENTIALS INSIDE THE BLACK BOX Alice L. Givan Englert Cell Analysis Laboratory of the Norris Cotton Cancer Center Dartmouth Medical School HOW NOT TO BE A FLOW CYTOMETRIST Drawing by Ben

More information

Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and

Using BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and Using BD FACSDiva CST To Evaluate Cytometer Performance, Create Custom Assay Settings and Implement Cross-Instrument and Cross-Site Standardization of Assays PART 2 Alan M. Stall Director, Advanced Cytometry

More information

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE

CELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each

More information

Instrument Characterization and Performance Tracking for Digital Flow Cytometers

Instrument Characterization and Performance Tracking for Digital Flow Cytometers Instrument Characterization and Performance Tracking for Digital Flow Cytometers BD Biosciences Cytometer Set-up & Tracking (CS&T) System For Research Use Only. Not for use in diagnostic or therapeutic

More information

Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose

Application Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose Selecting Reagents for Multicolor Flow Cytometry Holden Maecker and Joe Trotter BD Biosciences, San Jose Contents 1 The basics: Know your instrument 2 Fluorochromes: Go for the bright 3 Colors and specificities:

More information

Flow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement

Flow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement What is Flow Cytometry? Flow Cytometry Basic Principle and Applications BD Biosciences Daisy Kuo (daisy_kuo@bd.com) Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements are made on cells,

More information

Introduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size

Introduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size Introduction to Flow Cytometry -- BD LSR II Daisy Kuo 郭 正 佼 Assistant Product Manager BD Biosciences Daisy_kuo@bd.com What is Flow Cytometry? Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements

More information

Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer

Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Presented by Clare Rogers, MS Senior Marketing Applications Specialist BD Biosciences 23-13660-00 Webinar Overview Multicolor

More information

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis

Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Compensation and gating

More information

Immunophenotyping peripheral blood cells

Immunophenotyping peripheral blood cells IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune

More information

BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY

BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY IMPORTANT NOTES BEFORE READING THIS TUTORIAL This is a very expensive piece of equipment so PLEASE treat it with respect! After you are done

More information

Cell Cycle Tutorial. Contents

Cell Cycle Tutorial. Contents Cell Cycle Tutorial Contents Experimental Requirements...2 DNA Dyes...2 Protocols...3 PI Parameter & Analysis Setup...4 PI Voltage Adjustments...6 7-AAD Parameter Setup...6 To-Pro3 Parameter Setup...6

More information

Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract

Technical Bulletin. Standardizing Application Setup Across Multiple Flow Cytometers Using BD FACSDiva Version 6 Software. Abstract arch 212 Standardizing Application Setup Across ultiple Flow Cytometers Using BD FACSDiva Version 6 Ellen einelt, ervi Reunanen, ark Edinger, aria Jaimes, Alan Stall, Dennis Sasaki, Joe Trotter Contents

More information

Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers

Technical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers BD Biosciences, San Jose, CA Contents 2 Introduction to compensation 5 Tips for setting up compensation 10 Tools for easy,

More information

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer

No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer Introduction Standard methods for isolating and detecting leukocytes

More information

Compensation Basics - Bagwell. Compensation Basics. C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc.

Compensation Basics - Bagwell. Compensation Basics. C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc. Compensation Basics C. Bruce Bagwell MD, Ph.D. Verity Software House, Inc. 2003 1 Intrinsic or Autofluorescence p2 ac 1,2 c 1 ac 1,1 p1 In order to describe how the general process of signal cross-over

More information

Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...

Immunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation... Immunophenotyping Flow Cytometry Tutorial Contents Experimental Requirements...1 Data Storage...2 Voltage Adjustments...3 Compensation...5 Experimental Requirements For immunophenotyping with FITC and

More information

UNIVERSITY OF PÉCS MEDICAL SCHOOL FLOW CYTOMETRY AND CELL SEPARATION BIOPHYSICS 2. 2015 4th March Dr. Beáta Bugyi Department of Biophysics Flow cytometry and cell separation FLOW = STREAM OF FLUID in a

More information

Flow Cytometry. flow cytometer DNA apoptosis ph

Flow Cytometry. flow cytometer DNA apoptosis ph Flow Cytometry flow cytometer DNA apoptosis ph flow cytometry flow cytometer (a) (b) cell counting instrument (c) 1960 ink-jet technology 17 19 1940 1950 fluorescence microscopy fluorescent dye DNA polyclonal

More information

Compensation Controls Data Visualization Panel Development Gating Quantum Dots

Compensation Controls Data Visualization Panel Development Gating Quantum Dots Compensation Controls Data Visualization Panel Development Gating Quantum Dots FITC Single Stain Control FITC PE Argon Laser FL1 FL2 450 500 550 600 FITC Compensation Control Unwanted signal detected in

More information

Fluorescence Compensation. Jennifer Wilshire, PhD jennifer@flowjo.com

Fluorescence Compensation. Jennifer Wilshire, PhD jennifer@flowjo.com Fluorescence Compensation Jennifer Wilshire, PhD jennifer@flowjo.com Outline 1. Compensation Basics 2. Practical Applications -selecting comp controls (cells, stains?) 3. Multicolor Issues -spreading due

More information

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy.

Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. At the end of this session, the participant will be able to: 1. Describe the components of a flow cytometer 2. Describe the gating

More information

Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH))

Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH)) Automated Quadratic Characterization of Flow Cytometer Instrument Sensitivity (flowqb Package: Introductory Processing Using Data NIH)) October 14, 2013 1 Licensing Under the Artistic License, you are

More information

Principles of Flowcytometry

Principles of Flowcytometry Objectives Introduction to Cell Markers: Principles of Flowcytometry Michelle Petrasich NZIMLS Scientific Meeting August 24, 2010, Paihia What are cell markers How do we detect them Production of Monoclonal

More information

APPLICATION INFORMATION

APPLICATION INFORMATION DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL

More information

Boundary-breaking acoustic focusing cytometry

Boundary-breaking acoustic focusing cytometry Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory

More information

Let the BD FACSAria II accompany your best performance ever

Let the BD FACSAria II accompany your best performance ever Let the BD FACSAria II accompany your best performance ever Meet the new BD FACSAria II high-speed, fixed-alignment cuvette cell sorter. The first generation BD FACSAria system brought the complex world

More information

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835

Instructions for Use. CyAn ADP. High-speed Analyzer. Summit 4.3. 0000050G June 2008. Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Instructions for Use CyAn ADP High-speed Analyzer Summit 4.3 0000050G June 2008 Beckman Coulter, Inc. 4300 N. Harbor Blvd. Fullerton, CA 92835 Overview Summit software is a Windows based application that

More information

FACS Laboratory. BD LSRs Operator Training. Training Guide. http://london-research-institute.org.uk/technologies/facs

FACS Laboratory. BD LSRs Operator Training. Training Guide. http://london-research-institute.org.uk/technologies/facs FACS Laboratory facs_lab@cancer.org.uk Training Guide http://london-research-institute.org.uk/technologies/facs BD LSRs Operator Training LRI FACS lab 1 Components 1.1 Basic parts overview.pg3-6 Overview

More information

Flow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE

Flow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE Flow Data Analysis Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA Flow Data Analysis From Cells to Data - data measurement and standard Data display and gating - Data Display options -

More information

Compensation in Flow Cytometry

Compensation in Flow Cytometry Compensation in UNIT.4 The term compensation, as it applies to flow cytometric analysis, refers to the process of correcting for fluorescence spillover, i.e., removing the signal of any given fluorochrome

More information

123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.

123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.

More information

Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility

Supplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility 10.1071/RD14043_AC CSIRO 2015 Supplementary Material: Reproduction, Fertility and Development, 2015, 27(8), 1187 1196. Supplementary Material Free-radical production after post-thaw incubation of ram spermatozoa

More information

FlowSight. Flow cytometry with vision

FlowSight. Flow cytometry with vision FlowSight Flow cytometry with vision Flow cytometry with vision Introducing FlowSight Capable: Sensitive and flexible for every need Intuitive: Easy-to-use, with imagery for every cell Affordable: Designed

More information

Zecotek S Light Projection Network Marketing

Zecotek S Light Projection Network Marketing White Paper Zecotek Visible Fiber Laser Platform Enabling the future of laser technology Zecotek Photonics Inc. (TSX- V: ZMS; Frankfurt: W1I) www.zecotek.com is a Canadian photonics technology company

More information

Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer

Technical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Contents 2 Thresholds 2 Setting the Threshold When analyzing small particles, defined as particles smaller than 3.0 µm, on the

More information

1. Instrument Layout pg 2. 2. Summit Overview pg 3. 3. Data Storage and Data Files pgs 6-7. 4. Saving Data pg 7. 5. Start Up pg 8

1. Instrument Layout pg 2. 2. Summit Overview pg 3. 3. Data Storage and Data Files pgs 6-7. 4. Saving Data pg 7. 5. Start Up pg 8 Table of Contents 1. Instrument Layout pg 2 2. Summit Overview pg 3 3. Data Storage and Data Files pgs 6-7 4. Saving Data pg 7 5. Start Up pg 8 6. Choosing and Naming Parameters pg 9 7. Creating Histograms

More information

Spherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1

Spherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1 78 Irma Lee Circle, Unit 0, Lake Forest, Illinois 00 SPHERO TM Technical Note STN- Rev B 008 DETERMINING PMT LINEARITY IN FLOW CYTOMETERS USING THE SPHERO TM PMT QUALITY CONTROL EXCEL TEMPLATE Introduction

More information

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Fundamentals of modern UV-visible spectroscopy. Presentation Materials Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms

More information

Compensation: An Instrumental Perspective

Compensation: An Instrumental Perspective Compensation: An Instrumental Perspective Why Digital? Visualization Issues Boston User s Group September 10, 2003 Joe Trotter Joe_Trotter@bd.com Review: Instrument Sensitivity Measuring Sensitivity -

More information

Analyzer Experiment setup guide for LSRII and Canto s

Analyzer Experiment setup guide for LSRII and Canto s Analyzer Experiment setup guide for LSRII and Canto s 1. Check the instrument configuration on our website to determine the most appropriate cytometer for your experiment, https://depts.washington.edu/flowlab/instrumentation.html,

More information

FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS. By: Douaa Moh. Sayed

FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS. By: Douaa Moh. Sayed FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS By: Douaa Moh. Sayed Definition Flow cytometry is a technique for counting, examining, and sorting microscopic particles suspended in a stream of fluid. It allows

More information

Molecular Spectroscopy

Molecular Spectroscopy Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended

More information

How to operate the BD FACSCanto flow cytometer

How to operate the BD FACSCanto flow cytometer How to operate the BD FACSCanto flow cytometer Preface Dear colleague, the BD FACSCanto flow cytometer is use to operate, however, it is a delicate instrument for measuring fluorescence of single cells.

More information

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement

Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University

More information

use. 3,5 Apoptosis Viability Titration Assays

use. 3,5 Apoptosis Viability Titration Assays Introduction Monitoring cell viability can provide critical insights into the effectiveness of biological protocols and the stability of biological systems including: the efficacy of cell harvesting procedures,

More information

Science is hard. Flow cytometry should be easy.

Science is hard. Flow cytometry should be easy. Science is hard. Flow cytometry should be easy. TABLE OF CONTENTS 1 INTRODUCTION TO BD ACCURI C6 SOFTWARE... 1 1.1 Starting BD Accuri C6 Software... 1 1.2 BD Accuri C6 Software Workspace... 2 1.3 Opening

More information

Standardization, Calibration and Quality Control

Standardization, Calibration and Quality Control Standardization, Calibration and Quality Control Ian Storie Flow cytometry has become an essential tool in the research and clinical diagnostic laboratory. The range of available flow-based diagnostic

More information

Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM.

Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM. Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM. Some limitations (disadvantages) of NSOM A trade-off

More information

DNA Detection. Chapter 13

DNA Detection. Chapter 13 DNA Detection Chapter 13 Detecting DNA molecules Once you have your DNA separated by size Now you need to be able to visualize the DNA on the gel somehow Original techniques: Radioactive label, silver

More information

FACSCanto RUO Special Order QUICK REFERENCE GUIDE

FACSCanto RUO Special Order QUICK REFERENCE GUIDE FACSCanto RUO Special Order QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW

More information

Chapter 12 Filters for FISH Imaging

Chapter 12 Filters for FISH Imaging Chapter 12 Filters for FISH Imaging Dan Osborn The application of in situ hybridization (ISH) has advanced from short lived, non-specific isotopic methods, to very specific, long lived, multiple color

More information

powerful. Intuitive. Customizable. The new star of benchtop flow cytometry

powerful. Intuitive. Customizable. The new star of benchtop flow cytometry powerful. Intuitive. Customizable. The new star of benchtop flow cytometry About ACEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.

More information

Flow Cytometry Protocols

Flow Cytometry Protocols METHODS IN MOLECULAR BIOLOGY TM TM Volume 263 Flow Cytometry Protocols SECOND EDITION Edited by Teresa S. Hawley Robert G. Hawley 1 Flow Cytometry An Introduction Alice L. Givan Summary A flow cytometer

More information

CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping

CyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping CyFlow SL Healthcare Immunology Portable FCM System for 3-colour Immunophenotyping 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results

More information

BD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin

BD LSR II and FACSDiVa Software. Dr. Jens Fleischer, Basel Dr. Norbert Leclere,, Berlin BD LSR II and FACSDiVa Software, Basel Dr. Norbert Leclere,, Berlin Overview Electronics Covers Fluidics Connectors Controlpanel Sample Port On/Off The Control Panel Ease of Use The basic procedures to

More information

Institut für Experimentelle Immunologie und Bildgebung

Institut für Experimentelle Immunologie und Bildgebung Page 1 von 6 Starting Up the System Start up the computer, start Windows and log in as IMCES_user start BD FACSDiva software with your Groupname Turn on the cytometer main power Wait a moment and Check

More information

Islet Viability Assessment by Single Cell Flow Cytometry

Islet Viability Assessment by Single Cell Flow Cytometry Islet Viability Assessment by Single Cell Flow Cytometry Page 1 of 8 Purpose: To comprehensively assess the viability of the islet cell preparation prior to transplantation. Tissue Samples: A sample containing

More information

Fluorescent dyes for use with the

Fluorescent dyes for use with the Detection of Multiple Reporter Dyes in Real-time, On-line PCR Analysis with the LightCycler System Gregor Sagner, Cornelia Goldstein, and Rob van Miltenburg Roche Molecular Biochemicals, Penzberg, Germany

More information

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs TracePro Opto-Mechanical Design Software s Fluorescence Property Utility TracePro s Fluorescence Property

More information

Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance:

Problem Set 6 UV-Vis Absorption Spectroscopy. 13-1. Express the following absorbances in terms of percent transmittance: Problem Set 6 UV-Vis Absorption Spectroscopy 13-1. Express the following absorbances in terms of percent transmittance: a 0.051 b 0.918 c 0.379 d 0.261 e 0.485 f 0.072 A = log P o /P = log1/t = - log T

More information

CyFlow SL. Microbiology. Detection and Analysis of Microorganisms and Small Particles

CyFlow SL. Microbiology. Detection and Analysis of Microorganisms and Small Particles CyFlow SL Microbiology Detection and Analysis of Microorganisms and Small Particles 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results

More information

GETTING STARTED. 6. Click the New Specimen icon.

GETTING STARTED. 6. Click the New Specimen icon. BD FACSDiva Software Basic Experiment Guide Please note, this guide is not intended to take the place of in-person training but is to be used as a reference as needed after completing training with Unified

More information

CyFlow Space Your flexible flow cytometer

CyFlow Space Your flexible flow cytometer Your flexible flow cytometer www.sysmex-partec.com its flexibility gives you the space you need for your work Analysing cells and particles, be it from blood, plasma, tissue, plants, cell cultures or other

More information

EdU Flow Cytometry Kit. User Manual

EdU Flow Cytometry Kit. User Manual User Manual Ordering information: (for detailed kit content see Table 2) EdU Flow Cytometry Kits for 50 assays: Product number EdU Used fluorescent dye BCK-FC488-50 10 mg 6-FAM Azide BCK-FC555-50 10 mg

More information

ArC Amine Reactive Compensation Bead Kit

ArC Amine Reactive Compensation Bead Kit ArC Amine Reactive Compensation Bead Kit Catalog no. A1346 Table 1. Contents and storage information. Material Amount Composition Storage Stability ArC reactive beads (Component A) ArC negative beads (Component

More information

Zeiss 780 Training Notes

Zeiss 780 Training Notes Zeiss 780 Training Notes 780 Start Up Sequence Do you need the argon laser, 458,488,514nm lines? No Turn on the Systems PC Switch Turn on Main Power Switch Yes Turn on the laser main power switch and turn

More information

A Basic Guide. Institute of Medical Biology, University of Southern Denmark. Ungated Gate 1 (= R1) Gate 4 (=R4) Gate3 (=R3) Gate 2 (= R2) R3 R4

A Basic Guide. Institute of Medical Biology, University of Southern Denmark. Ungated Gate 1 (= R1) Gate 4 (=R4) Gate3 (=R3) Gate 2 (= R2) R3 R4 Flow Cytometry A Basic Guide R1 Ungated Gate 1 (= R1) R2 R3 R4 Gate 2 (= R2) Gate3 (=R3) Gate 4 (=R4) Graham Leslie, PhD Institute of Medical Biology, University of Southern Denmark Graham Leslie, PhD

More information

UCHC Guide to CellQuest Pro. FACS Calibur instruments

UCHC Guide to CellQuest Pro. FACS Calibur instruments UCHC Guide to CellQuest Pro FACS Calibur instruments Startup: Turn on Cytometer using green switch on the side of the cytometer Cytometer Power Switch Within a minute of turning on the cytometer; turn

More information

Attune NxT Acoustic Focusing Cytometer

Attune NxT Acoustic Focusing Cytometer USER GUIDE Attune NxT Acoustic Focusing Cytometer Catalog Number A24858, A24859, A24860, A24861, A24862, A24863, A24864 Publication Number 100024235 Revision B.0 For Research Use Only. Not for use in diagnostic

More information

Becton-Dickinson FACSCalibur Flow Cytometer Instructions

Becton-Dickinson FACSCalibur Flow Cytometer Instructions Start-up 1. Turn on flow cytometer by pushing green button on right side of instrument, toward the back. Give it 30 minutes to warm up before collecting data. 2. Turn on monitor. 3. Turn on computer. If

More information

Experiment #5: Qualitative Absorption Spectroscopy

Experiment #5: Qualitative Absorption Spectroscopy Experiment #5: Qualitative Absorption Spectroscopy One of the most important areas in the field of analytical chemistry is that of spectroscopy. In general terms, spectroscopy deals with the interactions

More information

Introduction To Real Time Quantitative PCR (qpcr)

Introduction To Real Time Quantitative PCR (qpcr) Introduction To Real Time Quantitative PCR (qpcr) SABiosciences, A QIAGEN Company www.sabiosciences.com The Seminar Topics The advantages of qpcr versus conventional PCR Work flow & applications Factors

More information

FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data. Monisha Sundarrajan, PhD BD Biosciences

FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data. Monisha Sundarrajan, PhD BD Biosciences FCAP Array v3.0 Software: A New Tool to Analyze BD Cytometric Bead Array (CBA) Data Monisha Sundarrajan, PhD BD Biosciences 23 13795 00 Agenda Introduction Overview of BD CBA bead based immunoassays FCAP

More information

Standard Operating Procedure

Standard Operating Procedure 1.0 Purpose: 1.1 The characterisation of of main leukocyte subsets in peripheral blood cells from mice by flow cytometry. Reliable values of frequencies of leukocyte clusters are very much dependent on

More information

Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction

Application Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction Application Note 10 Measurement of Cell Recovery After Sorting with a Catcher-Tube-Based Cell Sorter Introduction In many experiments using sorted cells, it is important to be able to count the number

More information

LIVE/DEAD Fixable Dead Cell Stain Kits

LIVE/DEAD Fixable Dead Cell Stain Kits USER GUIDE LIVE/DEAD Fixable Dead Cell Stain Kits Pub. No. MAN0002416 (MP34955) Rev. A.0 Table 1. Contents and storage Material Amount Storage Stability Individual Kits: Blue, violet, aqua, yellow-, green,

More information

Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation

Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation Roche Applied Science Technical Note No. LC 19/2004 Purpose of this Note Color The LightCycler System is able to simultaneously detect and analyze more than one color in each capillary. Due to overlap

More information

Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility

Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility The mention of commercial products, their sources, or their use in connection

More information

Easy to use. Compact. Flexible. Sorting Chip

Easy to use. Compact. Flexible. Sorting Chip SH800Z Cell Sorter The SH800Z is a compact, affordable, and easy to use cell sorter developed using the state- of-the art technologies from Sony. An array of automated mechanisms makes it an ideal sorter

More information

Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T

Stepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom Cat. Reference: STP-25T Reagent provided:: 25 Stepcount tubes for 25 test INTENDED USE. Immunostep

More information

BD FACSCanto Flow Cytometer. Efficient Easy Powerful

BD FACSCanto Flow Cytometer. Efficient Easy Powerful BD FACSCanto Flow Cytometer Efficient Easy Powerful The BD FACSCanto System: Optimal Workflow and Time Management Automated compensation algorithms with optional off-line compensation, streamlined sample

More information

Copyright 1999 2010 by Mark Brandt, Ph.D. 12

Copyright 1999 2010 by Mark Brandt, Ph.D. 12 Introduction to Absorbance Spectroscopy A single beam spectrophotometer is comprised of a light source, a monochromator, a sample holder, and a detector. An ideal instrument has a light source that emits

More information

Applications of confocal fluorescence microscopy in biological sciences

Applications of confocal fluorescence microscopy in biological sciences Applications of confocal fluorescence microscopy in biological sciences B R Boruah Department of Physics IIT Guwahati Email: brboruah@iitg.ac.in Page 1 Contents Introduction Optical resolution Optical

More information

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required.

PROTOCOL. Immunostaining for Flow Cytometry. Background. Materials and equipment required. PROTOCOL Immunostaining for Flow Cytometry 1850 Millrace Drive, Suite 3A Eugene, Oregon 97403 Rev.0 Background The combination of single cell analysis using flow cytometry and the specificity of antibody-based

More information