Understanding Exposure and Deviation Indices
|
|
- Dwayne Riley
- 7 years ago
- Views:
Transcription
1 Imaging Educational Course - Radiography III Understanding Exposure and Deviation Indices David L. Leong 1,2 Marie-Louise Butler 2 1 Analogic Corporation, Peabody, MA, USA 2 UCD School of Medicine and Medical Sciences, University College Dublin, Dublin, Ireland AAPM 2011 WE-A Learning Objectives 1. Understand AAPM recommendations, 2. Understand the IEC standard, 3. Understand factors that affect Exposure and Deviation Index values. 4. Understand DICOM support for Exposure and Deviation Index values. 2 1
2 State of Confusion Manufacturer Specific Indices Rex S lgm EI DEI EXI EVI f# A range of units and relationships to exposure. Calibration conditions ranged from 70 to 80 kvp with different filtration selections 3 AAPM Task Group 116 TG116 was started in July 25, 2004 Report published June individuals participated in the Task Group Full report available on AAPM Website Executive summary published in Medical Physics S. J. Shepard et al. "An exposure indicator for digital radiography: AAPM Task Group 116 (Executive Summary)," Med. Phys. 36, (2009)
3 IEC Working Group 43/62b Preparatory work in the German (DIN) standardization organization Proposal for standard by the German National Committee June 2006 WG43 was started in September 2006 Experts from Germany, Belgium, Canada, China, Japan, Netherlands, United States 16 members, 6 people overlapped with TG116 First meeting October 2006 in Hamburg, Germany IEC published August 2008 Stability date: Keys Parts Exposure Index Target Exposure Index Deviation Index Calibration Condition 6 3
4 Exposure Index Name IEC Exposure Index (EI) AAPM TG116 Indicated Equivalent Air Kerma (K ind ) Units Unitless μgy Precision Not specified 3 significant Different definitions, but substantially the same. K ind typically less than 1 as it is μgy EI is designed to be greater than 1 7 Exposure Index Formulas IEC: CAL 1 1 K f V g V f V V f K CAL CAL CAL -1 1 EI c0 g V 100μGy f V TG116: K STD K STD Q K 1000 log log10 K Q 1000 IND 0 0 K K 10, where K Gy 8 4
5 Target Exposure Index Target values: K TGT (TG116) and EI T (IEC) EI T expected value of EI when the detector is properly exposed Manufacturer / Site Specified Needs to be adjustable by user Can be established by professional societies Body part and view specific targets (recommended by TG116) 9 Deviation Index IEC AAPM TG116 Name Deviation Index (EI) Deviation Index (EI) Units Unitless Unitless Precision Not specified 1 decimal place TG116 and IEC are the same: DI EI 10 log EI T More important clinically than EI 10 5
6 Action Levels (TG116) DI Table 2. Exposure Indicator DI Control Limits for Clinical Images Range Action > +3.0 Excessive patient radiation exposure Repeat only if relevant anatomy is clipped or burned out Require immediate management followup. +1 to +3.0 Overexposure: Repeat only if relevant anatomy is clipped or burned out 0.5 to +0.5 Target range Less than 1.0 Underexposed: Consult radiologist for repeat Less than 3.0 Repeat 11 Calibration IEC kvp: Based on RQA5 from IEC HVL: 6.8 ± 0.3 mm Al Added filter of either 21 mm aluminum or 0.5 mm copper and 2 mm aluminum Adjustment of the kvp from 66 to 74 is permitted to achieve the target HVL AAPM kvp: 70 ± 4 HVL: 6.8 ± 0.25 mm Al Added filter of either 21 mm aluminum or 0.5 mm copper and 0 to 4 mm aluminum*, Adjustment of the kvp and filter permitted to achieve the target HVL *Type 1100 Document the added filter and X-ray tube voltage used for calibration 12 6
7 Calibration - Recommendation Start as close to 70 kvp as possible Measure non-invasively without added filtration If your meter support Practical Peak Voltage (PPV), use it Use at least 1 meter Source to Detector distance Adjust added filtration to achieve 6.8 mm HVL No less than 0.5 mm Cu or 18 mm Al No more than 0.5 mm Cu + 4 mm Al or 23 mm Al Raise or lower kv as a last resort Record the final filter thickness and generator kvp On repeat calibrations, start with the previous filter thickness 13 Summary Exposure Indicator Name IEC Exposure Index (EI) AAPM TG116 Indicated Equivalent Air Kerma (K ind ) Exposure Indicator Units Unitless μgy Exposure Indicator Precision Not specified 3 significant EI EI T K KTGT IND Deviation Index formula DI 10 log DI 10*log Deviation Index Units Unitless Unitless Deviation Index Precision Not specified 1 decimal place Beam Conditions 6.8 ± 0.3 mm Al 6.8 ± 0.25 mm Al 14 7
8 Factors that affect EI Calculation of EI based on anatomical region of interest. (ROI) - Relevant Image Region Method to determine ROI and area of ROI affect EI value Either define a specific procedure or method 15 ROI and Exposure Variation 16 8
9 30 Exposures No Changes Bone and Soft Tissue Mostly Bone Bone Only Mean Std Chest Exposures All Image Anatomy Mean Std
10 ROI Effects on DI Bone and Soft Tissue Mostly Bone Bone Only 4 mas mas DICOM Changes CP1024 written to support new Exposure Indices Became Final Text June 21, 2010 Use is not mandatory 20 10
11 DICOM Changes Legacy Exposure Indices Attribute Name Tag Type Attribute Description Relative X-Ray Exposure (0018,1405) 3 Relative x-ray exposure on the plate. Meaning of values is implementation specific. May be used to adjust the dynamic range of the plate digitizer (scanner). Note: This value is manufacturerspecific. DICOM specifies standard Attributes in Table 10.x-1 Exposure Index Macro which are recommended. Sensitivity (0018,6000) 3 Read out sensitivity. Note: This value is manufacturerspecific. DICOM specifies standard Attributes in Table 10.x-1 Exposure Index Macro which are recommended. 21 DICOM Changes New Exposure Indices Attribute Name Tag Type Attribute Description Exposure Index (0018,1411) 3 Measure of the detector response to radiation in the relevant image region of an image acquired with a digital x-ray imaging system as defined in IEC Notes: 1. A string rather than binary Value Representation is used for this Attribute, in order to allow the sender to control the precision of the value as suggested in the report of AAPM Task Group This index value is scaled as defined by IEC Target Exposure Index (0018,1412) 3 The target value used to calculate the Deviation Index (0018,1413) as defined in IEC Deviation Index (0018,1413) 3 A scaled representation of the difference of the Exposure Index compared to the Target Exposure Index as defined in IEC and the report of AAPM TG
12 Recommendations Insist that all new CR and DR equipment support IEC Both EI and DI Push your PACS vendor to support and display DI on all workstations (CP1024) Attempt to upgrade older CR and DR systems to add support for IEC Develop a set of Target Exposure Values Setup a training program on the use of EI, DI and action limits 23 Learning Objectives 1. Understand AAPM recommendations, 2. Understand the IEC standard, 3. Understand factors that affect Exposure and Deviation Index values. 4. Understand DICOM support for Exposure and Deviation Index values
13 Final Comment It is essential that t all aspects of good clinical technique be adhered to and an appropriate DI value should not be interpreted as proof of good work. TG116 Report Section X 25 Thank You UCD School of Medicine & Medical Science Scoil an Leighis agus Eolaíocht An Leighis UCD 13
The Exposure Index and Its Standardization
The Exposure Index and Its Standardization Ulrich Neitzel Philips Medical Systems Clinical Science Hamburg, Germany ulrich.neitzel@philips.com Content The Exposure Index (EI) What is it? What is it good
More informationStudy the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter
Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt.
More informationPhysics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
More informationUnderstanding Digital Modalities: System Integration and Use
Understanding Digital Modalities: System Integration and Use Donald J. Peck, PhD Henry Ford Health System Detroit, MI donaldp@rad.hfh.edu Dicom Digital Imaging and Communications in Medicine Part 3: Information
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MAMMOGRAPHY) SECTION
More informationIrradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm
Proceedings of the Ninth EGS4 Users' Meeting in Japan, KEK Proceedings 200-22, p.5-8 MONTE CARLO SIMULATION ANALYSIS OF BACKSCATTER FACTOR FOR LOW-ENERGY X-RAY K. Shimizu, K. Koshida and T. Miyati Department
More informationProduction of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9
Module 9 This module presents information on what X-rays are and how they are produced. Introduction Module 9, Page 2 X-rays are a type of electromagnetic radiation. Other types of electromagnetic radiation
More informationX-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission
Principles of Imaging Science I (RAD119) X-ray Production & Emission X-ray Production X-rays are produced inside the x-ray tube when high energy projectile electrons from the filament interact with the
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOGRAPHY)
More informationINTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
More informationName: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography
Name: Date: Team: Lab Experiment # 3 Focused Grid Positioning Errors Computed Radiography and Direct Digital Radiography Purpose This experiment is designed to demonstrate the effect of off-level error,
More informationDiagnostic Exposure Tracking in the Medical Record
Diagnostic Exposure Tracking in the Medical Record J.A. Seibert, Ph.D. Department of Radiology Sacramento, California USA Vancouver. British Columbia Relevant disclosures None Learning objectives Understand
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY DIAGNOSTIC (VETERINARY) SECTION 1:
More informationRole of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
More informationRadiography: 2D and 3D Metrics of Performance Towards Quality Index
AAPM COMP 011 Radiography: D and 3D Metrics of Performance Towards Quality Index Ehsan Samei, Sam Richard Duke University Learning objectives Outlook 1. To understand methods for D and 3D resolution measurements..
More informationCT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of
More informationQuality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
More informationCHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC
CHAPTER 5 QC Test For Radiographic Equipment Prepared by:- Kamarul Amin bin Abdullah @ Abu Bakar School of Medical Imaging KLMUC Lesson Outcomes Describe the objectives of each QC test done.(importance)
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (MOBILE RADIOSCOPY)
More informationDigital radiography has become the norm in U.S.
Expanded content at www.asrt.org/publications. Image Gently: Using Exposure Indicators To Improve Pediatric Digital Radiography Quentin T Moore, MPH, R.T.(R)(T)(QM) Steven Don, MD Marilyn J Goske, MD,
More informationGE Healthcare. DoseWatch. Gathering Radiation Dose Data
GE Healthcare DoseWatch Gathering Radiation Dose Data Gathering Radiation Dose Data After your organization crosses the bridge of Why? for patient radiation dose management either due to regulation, risk
More informationManual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment
Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment Atomic Energy Regulatory Board Mumbai 400 094 1 DEFINITIONS 1. Quality Assurance (QA): Planned and systematic actions
More informationA software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department
More informationProduction of X-rays and Interactions of X-rays with Matter
Production of X-rays and Interactions of X-rays with Matter Goaz and Pharoah. Pages 11-20. Neill Serman Electrons traveling from the filament ( cathode) to the target (anode) convert a small percentage
More informationDigital Breast Tomosynthesis QC Requirements
Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist
More informationDICOM Correction Item
Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes
More informationRadiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids
Principles of Imaging Science II (RAD 120) Radiographic Grids 1 Image-Forming X-Rays Four X-ray paths a. X-rays interact with patient and scatter away from the receptor b. X-rays interact and are absorbed
More informationCT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
More information5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION
5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION 5.2.1 Task The bremsstrahlung produced by the X-ray tube has a continuous spectrum, limited by the set and spreads
More informationSECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS
Department of Health and Human Services Population Health Radiation Protection Act 2005 Section 17 CERTIFICATE OF COMPLIANCE: STANDARD FOR RADIATION APPARATUS - X-RAY MEDICAL DIAGNOSTIC (BIOPSY) SECTION
More informationEPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography
Environment Protection Authority EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography Doc. No. RP91 Prepared by the Radiation Protection Branch For further information
More informationDigital Image Processing with DragonflyPACS
Digital Image Processing with DragonflyPACS The Smarter PACS The smartest image processing system in veterinary medicine. Outstanding image quality that s compliant with DICOM 3.0 standards. Compatible
More informationImage Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered
Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered vs. Wall-Mount X-Ray X Systems Edgar Bailey*, MSEHE, CHP Consultant Joel Gray*, PhD, FAAPM DIQUAD, LLC
More informationPage: 1 of 6 Page: 1 of 6
Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique
More informationPerformance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
More informationDesign and Implementation of an Institution-Wide Patient-Specific Radiation Dose
Design and Implementation of an Institution-Wide Patient-Specific Radiation Dose Monitoring Program for Computed Tomography, Digital Radiography, and Nuclear Medicine by Olav Christianson Medical Physics
More informationUse of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators
Technical Note: 9 Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators 1- Introduction The VALIDATOR, model TN-ID-60, is a compact, and stand-alone dosimetry
More informationMeasuring Size and Distance in Digital Mammography
On the matter of size and distance measurements in digital mammography David A. Clunie 2007/03/18 Radiologists reading mammograms have a need to measure size and distance in order to estimate interval
More informationUpdate on ACR Digital Mammography QC Manual
Update on ACR Digital Mammography QC Manual Priscilla F. Butler, M.S. Medical Physicist and Senior Director, ACR, Reston, VA (with thanks to Eric Berns, Ph.D.) Overview Phantom Specifications QC Manual
More informationSWABIK Project. Software Tools for DICOM Media Exchange in Clinical Research: the. SWABIK Project
Software Tools for DICOM Media Exchange in Clinical Research: the SWABIK Project Software Tools for DICOM Media Exchange in Clinical Research: the SWABIK Project O. ElGazzar 1, M. Onken 1, M. Eichelberg
More informationQuality control of CT systems by automated monitoring of key performance indicators: a two-year study
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 16, NUMBER 4, 2015 Quality control of CT systems by automated monitoring of key performance indicators: a two-year study Patrik Nowik, a Robert Bujila,
More informationGestión global de la dosis en TC. Sistema de registro y gestión
Gestión global de la dosis en TC. Sistema de registro y gestión IV Jornada de Protección Radiológica Hospitalaria SARH FEA Radiofísica Hospital Virgen de las Nieves. Granada jalmansa.lopez@gmail.com 22
More informationIGRT. IGRT can increase the accuracy by locating the target volume before and during the treatment.
DERYA ÇÖNE RADIOTHERAPY THERAPIST ACIBADEM KOZYATAGI HOSPITAL RADIATION ONCOLOGY DEPARTMENT IGRT IGRT (image-guided radiation therapy) is a technique that reduces geometric uncertainties by considering
More informationTracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives
Tracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives Gilbert W. Beebe Symposium The National Academies Katharine Grant, PhD Staff Scientist 8 December 2011 For internal
More informationUsing Automatic Exposure Control in Digital Radiography
Using Automatic Exposure Control in Digital Radiography A. Kyle Jones, Ph.D. U.T. M. D. Anderson Cancer Center Department of Imaging Physics 2008 AAPM Meeting, Houston, TX Automatic Exposure Control The
More informationQuality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects
Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects Tamás Porubszky *, Csaba Váradi, László Ballay, Olivér Turák, Géza Gáspárdy, István Turai Frédéric
More informationDose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida
Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida Average Glandular Dose Required measurement performed by medical physicist as part of Mammography
More informationQualitätsmanagement-Handbuch
U r Kalibrier- und Messmöglichkeiten Qualitätsmanagement-Handbuch Relative erweiterte Messunsicherheit, die sich aus der Standardmessunsicherheit durch Multiplikation mit dem Erweiterungsfaktor k = 2 ergibt.
More informationA PACS-Aware DICOM Image Object
Designing and Implementing A PACS-Aware DICOM Image Object For Digital X-ray, Mammography and Intraoral Applications David A. Clunie Quintiles Intelligent Imaging Clear Vision for the Healthcare Industry
More information7üVNORD. TÜV NORD Röntgentechnik
.. TÜV NORD Röntgentechnik Niedersachsen Nordrhein-Westfalen Hamburg Schleswig-Holstein Baden-Württemberg Mecklenburg-Vorpommern Bremen Sachsen-Anhalt Berlin Brandenburg Sachsen' Thüringen' Bayern' Hessen
More informationApplication of Digital Radiography for the Detection and Classification of Pneumoconiosis
Application of Digital Radiography for the Detection and Classification of Pneumoconiosis DEPARTMENT OF HEALTH AND HUMAN SERVICES Centers for Disease Control and Prevention National Institute for Occupational
More informationBest Practices in Digital Radiography
WHITE PAPER Best Practices in Digital Radiography Tracy L. Herrmann, M.Ed., R.T.(R); Terri L. Fauber, Ed.D., R.T.(R)(M); Julie Gill, Ph.D., R.T.(R)(QM); Colleen Hoffman, R.T.(R)(M)(CT); Denise K. Orth,
More informationThe effects of radiation on the body can be divided into Stochastic (random) effects and deterministic or Non-stochastic effects.
RADIATION SAFETY: HOW TO EDUCATE AND PROTECT YOURSELF AND YOUR STAFF John Farrelly, DVM, MS, ACVIM (Oncology), ACVR (Radiation Oncology) Cornell University Veterinary Specialists The Veterinary Cancer
More informationComparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations
Comparing dose and image quality of a conventional 400 speed class film/screen system with a CR system by assessing pediatric chest examinations Alzen,, G., Beiderwellen,, K., Volz,, M., Berthold,, L.D.,
More informationAn Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
More informationAssessing Radiation Dose: How to Do It Right
Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine
More informationACR AAPM SIIM PRACTICE PARAMETER FOR DIGITAL RADIOGRAPHY
The American College of Radiology, with more than 30,000 members, is the principal organization of radiologists, radiation oncologists, and clinical medical physicists in the United States. The College
More informationRADIATION THERAPY. Dosimetry Pioneers since 1922 NEW
RADIATION THERAPY Dosimetry Pioneers since 1922 NEW New Product Releases 2015 One phantom multiple options 4D Patient and Machine QA Modular Phantom for Stay flexible. Go modular. } Revolutionary new modular
More informationIAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)
IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe) Results of the Coordinated Research Project on Optimization
More informationIndirect X-ray photon counting image sensor with 27T pixels and 15 electrons RMS accurate threshold
Indirect X-ray photon counting image sensor with 27T pixels and 15 electrons RMS accurate threshold B. Dierickx 1,2, B. Dupont 1,3, A. Defernez 1, N. Ahmed 1 1 Caeleste, Antwerp, Belgium 2 Vrije Universiteit
More informationPatient Exposure Doses During Diagnostic Radiography
Patient Exposure Doses During Diagnostic Radiography JMAJ 44(11): 473 479, 2001 Shoichi SUZUKI Associated Professor, Faculty of Radiological Technology, School of Health Sciences, Fujita Health University
More informationDigital radiography exposure indices: A review
REVIEW ARTICLE Digital radiography exposure indices: A review Ursula Mothiram, BSc Diagnostic Radiography (Hons), MHlth Sc.(Ed), 1 Patrick C. Brennan, PhD, 1 Sarah J. Lewis, PhD, MEd, BAppSc, DR (Hons),
More informationImage Acquisition and Diagnostics
en examion aqs software examion aqs software Image Acquisition and Diagnostics Professional Image Acquisition and Diagnostic Workstation examion aqs software is the universal image acquisition software
More informationAgency for Health Care Administration
Page 1 of 20 FED - H0000 - INITIAL COMMENTS Title INITIAL COMMENTS CFR Type Memo Tag FED - H0005 - COMPLIANCE WITH FED/STATE/LOCAL LAWS Title COMPLIANCE WITH FED/STATE/LOCAL LAWS CFR 486.100 Type Condition
More informationDigital Dental X-ray. Midmark Animal Health Products
Digital Dental X-ray Midmark Animal Health Products Disease you can't see... 60% of periodontal disease lies below the gum line, and many issues aren't detectable via probing or visual inspection.* Don't
More informationMediCal QAWeb. Secured online service for Quality Assurance of medical displays
MediCal QAWeb Secured online service for Quality Assurance of medical displays The first secured online service for high-grade Quality Assurance MediCal QAWeb is the industry s first online service for
More informationThe convenient and compact CR system for veterinary practices
Divario CR-Fvet with dicom PACS DX-R Software CR Systems for the Future The convenient and compact CR system for veterinary practices Divario CR-F20/ F40/ F60/ F80 vet Divario CR-F20/ F40/ F60/ F80 vet
More informationElements of required physical infrastructures: space, schielding, and patient flow..
Elements of required physical infrastructures: space, schielding, and patient flow.. IAEA Following the IAEA guidelines, adapded by Anna Benini for workshop on Health Technology IUPESM Task Group, Porto
More informationBarracuda & QABrowser
Barracuda & QABrowser Reference Manual - English - Version 4.3A RTI article number: 9620501-00 Welcome to Barracuda and the QABrowser The Barracuda is an X-ray Analyser/Multimeter for everybody working
More informationDENTAL Cone beam 3D X-RAY SYSTEM with
VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic
More information3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses.
1 3701-72-03 Standards for accreditation of educational programs and approval of continuing education courses. (A) Any person may apply to the director for approval to conduct an educational program for
More informationComputed Radiography: Acceptance Testing and Quality Control
Computed Radiography: Acceptance Testing and Quality Control J. Anthony Seibert, Ph.D. University of California Davis Department of Radiology Sacramento, California Introduction CR is the primary means
More informationSecurity Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards
Security Imaging Systems (X-Ray): Design, Measurements Regulations, and Standards Presented: August 3, 2011 Daniel Kassiday United States Food and Drug Administration Center for Devices and Radiological
More informationMulti-slice Helical CT Scanning of the Chest
Multi-slice Helical CT Scanning of the Chest Comparison of different low-dose acquisitions Lung cancer is the main cause of deaths due to cancer in human males and the incidence is constantly increasing.
More informationQUALITY CONTROL IN DIAGNOSTIC RADIOLOGY
AAPM REPORT NO. 74 QUALITY CONTROL IN DIAGNOSTIC RADIOLOGY Report of Task Group #12 Diagnostic X-ray Imaging Committee Members S. Jeff Shepard, Chairman Pei-Jan Paul Lin, Co-Chairman John M. Boone Dianna
More informationI.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS
I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS A. Slavtchev 1, I. Manolov 2 1 National Centre of Radiobiology and Radiation Protection 2 Medrom Ltd. Abstract In recent years the number
More informationThinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL
Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,
More informationFailure Modes and Effects Analysis (FMEA)
Failure Modes and Effects Analysis (FMEA) Sasa Mutic Washington University School of Medicine St. Louis Missouri Failure Modes and Effects Analysis Objectives: To motivate the use of FMEA and to provide
More informationRecommendations for a technical quality control program for diagnostic X-ray equipment
Australasian Physical & Engineering Sciences in Medicine Volume 28 Number 2, 2005 ACPSEM POSITION PAPER Recommendations for a technical quality control program for diagnostic X-ray equipment D. A. Causer
More informationRequired RS Training Info
C-arm Radiation Safety at Tufts Required RS Training Info What are annual rad. dose limits? Who is our regulator? What should you do in an emergency? Are there health effects of radiation? C-arm beam awareness
More informationIntroduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.
Slide 1 Digital Mammography Detector QC Slide 2 Introduction What is FFDM QC and why is it important? Eric A. Berns, Ph.D. What to know before you start eric.berns@gmail.com Overview and compare QC tests
More informationEDXRF of Used Automotive Catalytic Converters
EDXRF of Used Automotive Catalytic Converters Energy Dispersive X-Ray Fluorescence (EDXRF) is a very powerful technique for measuring the concentration of elements in a sample. It is fast, nondestructive,
More informationPRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT
PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,
More informationX-Mind. Instinct for perfection
X-Mind Instinct for perfection X-Mind tubes are located at the back of the head which gives the patient better protection because the distance between the focal spot and the skin is 50% greater than in
More informationDigital radiography conquers the veterinary world
Digital radiography conquers the veterinary world Author: Dirk De Langhe Increasingly, veterinarians are using medical imaging to diagnose their patients. There is a corresponding tendency towards replacing
More informationThe Convenient and Compact CR System for Veterinary Practices
The Convenient and Compact CR System for Veterinary Practices Fire CR+ 20 / 40 / 60 / 80 vet with a throughput of 20 to 70 cassettes per hour and the option of a space - saving wall mount The small and
More informationX-ray (Radiography) - Abdomen
Scan for mobile link. X-ray (Radiography) - Abdomen Abdominal x-ray uses a very small dose of ionizing radiation to produce pictures of the inside of the abdominal cavity. It is used to evaluate the stomach,
More informationImplementation of Cone-beam CT imaging for Radiotherapy treatment localisation.
Implementation of Cone-beam CT imaging for Radiotherapy treatment localisation. Andrew Bridges Clinical Scientist Diagnostic Radiology & Radiation Protection Physics Overview What is CBCT? Use of CBCT
More informationChecklist for the Qualification of Digital Detector Array Systems
Paper No. 010-12 Checklist for the Qualification of Digital Detector Array Systems 1 February 2012 This document was created by the Federal Working Group on Industrial Digital Radiography. Reproduction
More informationEPEX. The Hologic EPEX General. Direct-to-Digital System for General Radiography. Versatility for a complete range of general radiographic exams
EPEX Direct-to-Digital System for General Radiography Featuring... Versatility for a complete range of general radiographic exams The Hologic EPEX General Radiography System is a direct-to-digital, fully
More informationEquipment. Lin muyan. Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764
Digital Medical Radiography Diagnostic (DR) system X-ray Equipment Presented by: CocticaUwititonze Email: costicauwitonze@yahoo.com Cell:+250 (0)7888 44764 Lin muyan 1. Goals 2. Introduction 3. X-ray machine
More informationSUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES
SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES Legislation There are two sets of regulations in the UK governing the use of ionizing radiation. They both form part of The Health and Safety at Work Act
More informationRADIATION SAFETY STANDARD
Radiation Safety Act 1999 RADIATION SAFETY STANDARD NM004:2010 Standard for radiation apparatus used to carry out industrial radiography Preface Under section 17 of the Radiation Safety Act 1999, a possession
More informationCAR Standards for Teleradiology
CAR Standards for Teleradiology The standards of the Canadian Association of Radiologists (CAR) are not rules, but are guidelines that attempt to define principles of practice that should generally produce
More informationRadiation Dose Measurements in Routine X ray Examinations
Tenth Radiation Physics & Protection Conference, 27-3 November 21, Nasr City - Cairo, Egypt Radiation Dose Measurements in Routine X ray Examinations H. Osman *1, A. Sulieman 1, I.I. Suliman 2 and A.K.
More informationAN INNOVATIVE & CUTTING-EDGE DIGITAL RADIOGRAPHY SYSTEM
SAMSUNG SMART DIGITAL RADIOGRAPHY 2012 Samsung Electronics All rights reserved. Samsung Electronics reserves the right to modify the design, packaging, specifications and features shown herein, without
More informationTA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1
TA3 Dosimetry and Instrumentation Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1 1 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil. ABSTRACT
More informationLearning Objectives: Quality Assurance (QA) is. Quality Assurance Procedures for Digital Radiography. Alternate definition of Quality Assurance (QA)
Quality Assurance Procedures for Digital Radiography Charles E. Willis, Ph.D., DABR Associate Professor Department of Imaging Physics The University of Texas M.D. Anderson Cancer Center Houston, Texas
More informationportable x-ray survey report
department of HealtH and Human services centers for medicare & medicaid services form approved omb no. 0938-0027 portable x-ray survey report provider number H1 date surveyed (1) initial (2) survey W H2
More information