CT Scan Protocol DYONICS PLAN. Hip Impingement Planning System. CT Scan Protocol Rev. A

Size: px
Start display at page:

Download "CT Scan Protocol DYONICS PLAN. Hip Impingement Planning System. CT Scan Protocol Rev. A"

Transcription

1 CT Scan Protocol DYONICS PLAN Hip Impingement Planning System CT Scan Protocol Rev. A

2 Protocol Requirements Protocol Requirements The CT exam must adhere to the protocol requirements specified in this document. If it does not fully comply with these requirements, DYONICS PLAN may not be able to load the CT exam or it may compromise the validity of the analysis. If the CT exam is incomplete, DYONICS PLAN may fail to compute some anatomic measurements. Acquisitions Perform only two acquisitions: pelvis and knees. Use the same frame of reference for both acquisitions - i.e., from the same scout view. Required Scout/Topograms For the acquisitions, obtain each of the following scout/ topograms: AP/Coronal Lateral/Sagittal Other Reformatted and 3D Reconstruction Volumes The responsible radiologist should decide whether additional scans from the routine diagnostic protocol should be added to the exam to provide any additional information. Patient Preparation Position the patient so that they are supine, with no unnatural tilt of the pelvis, feet first, legs parallel in neutral position, toes pointing up. Instruct the patient on the importance of holding completely still during the scan. Patient stillness is critical for accurate 3D imaging and modeling. Required Reconstructions Create three or four reconstructions, including the pelvis, the knees, and the right and/or left unilateral hips. The following series must be included in the exam: Reconstruction 1.: Bilateral axial pelvis reconstruction Reconstruction 2.: Bilateral axial knees reconstruction Reconstruction 3.:Unilateral axial right hip reconstruction and/or Reconstruction 4.: Unilateral axial left hip reconstruction You must also include the AP/Coronal and Lateral/Sagittal scout views in the exam. Note: For the unilateral hip reconstruction(s), do not perform a separate acquisition/study for each hip side of interest. Instead, generate the unilateral hip reconstruction(s) from the full pelvis acquisition/study Rev. A DYONICS PLAN CT Scan Protocol

3 Image Acquisitions Image Acquisitions Bilateral Pelvis Start AIIS End Figure 1. ASIS Lesser Trochanter Bilateral pelvis Axial Scanning Range: Start: Midpoint between the anterior inferior iliac spine (AIIS) and anterior superior iliac spine (ASIS) End: 30 mm below lesser trochanter FOV Diameter: 500 mm. Use same X and Y centers for both pelvis and knee scans based on a single coronal scout Collimation Thickness: Minimum slice/row thickness available Pitch: As close to 1.0 as possible Tube Voltage: 120 kv Tube Current Time Product: Small patients: 100 mas Medium patients: 150 mas Large patients: 200 mas Or other standard guidelines, if required by your department or institution.* Bilateral Knees Axial Scanning Range: Start: 40 mm proximal to knee joint line. Figure 2. Start End Bilateral knees Joint Line End: Knee joint line (Scan must cover epicondyles and posterior condyles in their entirety) FOV Diameter: 500 mm. Use same X and Y centers for both pelvis and knee scans based on a single coronal scout Collimation Thickness: Minimum slice/row thickness available Pitch: As close to 1.0 as possible Tube Voltage: 120 kv Tube Current Time Product: Small patients: 100 mas Medium patients: 150 mas Large patients: 200 mas Or other standard guidelines, if required by your department or institution.* *CT scan compatibility with DYONICS PLAN has only been validated for the specified parameters. DYONICS PLAN CT Scan Protocol Rev. A 3

4 s s Reconstruction 1: Bilateral Pelvis Slice Thickness: mm Reconstruction Interval: mm Or Bone Tissue: Siemens: B70, GE: Bone, Toshiba: FC30, Philips: C, or equivalent* Figure 3. Bilateral pelvis reconstruction field of view Reconstruction 2: Bilateral Knees Slice Thickness: mm Reconstruction Interval: mm Or Bone Tissue: Siemens: B70, GE: Bone, Toshiba: FC30, Philips: C, or equivalent* Figure 4. Bilateral knees reconstruction field of view *CT scan compatibility with DYONICS PLAN has only been validated for the specified parameters Rev. A DYONICS PLAN CT Scan Protocol

5 s Reconstruction 3: Unilateral Right Hip Figure 5. Unilateral right hip reconstruction field of view Slice Thickness: Miniminum available (must be less than 1.25 mm) Reconstruction FOV: Less than 250 mm Reconstruction Interval: mm Note: Unilateral reconstructions do not require separate acquisitions. They are reconstructed from the bilateral pelvis acquisition. Note: Either the left or the right reconstruction may not be required. Reconstruction 4: Unilateral Left Hip Figure 6. Unilateral left hip reconstruction field of view Slice Thickness: Miniminum available (must be less than 1.25 mm) Reconstruction FOV: Less than 250 mm Reconstruction Interval: mm Note: Unilateral reconstructions do not require separate acquisitions. They are reconstructed from the bilateral pelvis acquisition. Note: Either the left or the right reconstruction may not be required. *CT scan compatibility with DYONICS PLAN has only been validated for the specified parameters. DYONICS PLAN CT Scan Protocol Rev. A 5

6 DICOM Fields Required DICOM Fields The following table lists the fields that must be present in the DICOM header. If a value is required, it is specified. Otherwise, it is not. Image Type (0008, 0008) Study Date (0008, 0020) Modality (0008, 0060) Patient s Name (0010, 0010) Patient s Birth Date (0010, 0030) Patient s Sex (0010, 0040) Slice Thickness (0018, 0050) Convolution Kernel (0018, 1210) Series Number (0020, 0011) Required Field Acquisition Number (0020, 0012) Image Position (Patient) (0020, 0032) Must contain label AXIAL = CT Must be unique for each series Image Orientation Patient (0020, 0037) = 1.0\0.0\0.0\0.0\1.0\0.0 Photometric Interpretation (0028, 0004) Rows (0028, 0010) = 512 Columns (0028, 0011) = 512 Pixel Spacing (0028, 0030) Bits Allocated (0028, 0100) Bits Stored (0028, 0101) High Bit (0028, 0102) Pixel Representation (0028, 0103) Rescale Intercept (0028, 1052) Rescale Slope (0028, 1053) = MONOCHROME2 Required Value Optional DICOM Fields The following table lists the fields that may be optionally included in the DICOM header. There are no required values. Required Field Accession Number (0008, 0050) Patient Id (0010, 0020) Required Value Endoscopy Smith & Nephew, Inc. 150 Minuteman Road Andover, MA USA Fax U.S. Customer Service 2013 Smith & Nephew, Inc. All rights reserved. 09/ Rev. A

PLAN Begins Here The revolutionary way to visualize FAI and develop a patient-specific preoperative plan to correct it.

PLAN Begins Here The revolutionary way to visualize FAI and develop a patient-specific preoperative plan to correct it. USER MANUAL AND FREQUENTLY ASKED QUESTIONS dyonicsplan.com PLAN Begins Here The revolutionary way to visualize FAI and develop a patient-specific preoperative plan to correct it. DYONICS PLAN Hip Impingement

More information

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH

CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology

More information

The disclaimer on page 1 is an integral part of this document. Copyright March 1, 2016 by AAPM. All rights reserved.

The disclaimer on page 1 is an integral part of this document. Copyright March 1, 2016 by AAPM. All rights reserved. DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION

More information

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions

CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of

More information

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved.

The disclaimer on page 1 is an integral part of this document. Copyright February 23, 2016 by AAPM. All rights reserved. DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION

More information

The disclaimer on page 1 is an integral part of this document. 1. Copyright December 14, 2015 by AAPM. All rights reserved.

The disclaimer on page 1 is an integral part of this document. 1. Copyright December 14, 2015 by AAPM. All rights reserved. DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION

More information

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL

Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL Thinking ahead. Focused on life. REALIZED: GROUNDBREAKING RESOLUTION OF 80 µm VOXEL X-ray ZOOM RECONSTRUCTION Flat Panel Detector (FPD) Automatic Positioning Function For ø 40 x H 40 mm, ø 60 x H 60 mm,

More information

Copyright March 1, 2016 by AAPM. All rights reserved.

Copyright March 1, 2016 by AAPM. All rights reserved. DISCLAIMER: TO THE EXTENT ALLOWED BY LOCAL LAW, THIS INFORMATION IS PROVIDED TO YOU BY THE AMERICAN ASSOCIATION OF PHYSICISTS IN MEDICINE, A NON-PROFIT ORGANIZATION ORGANIZED TO PROMOTE THE APPLICATION

More information

How To Improve Your Ct Image Quality

How To Improve Your Ct Image Quality Translating Protocols Between Scanner Manufacturer and Model Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo

More information

LONI De-Identification Policy

LONI De-Identification Policy The following defines how different file formats are de-identified with LONI tools. Each metadata attribute of each file can have the following operations performed: Operation keep Remove Replace Description

More information

Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality!

Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality! Temporomandibular Joint Imaging Using CBCT: Technology Now Captures Reality! Dale A. Miles BA, DDS, MS, FRCD (C) Diplomate, American Board of Oral and Maxillofacial Radiology Diplomate, American Board

More information

SITE IMAGING MANUAL ACRIN 6698

SITE IMAGING MANUAL ACRIN 6698 SITE IMAGING MANUAL ACRIN 6698 Diffusion Weighted MR Imaging Biomarkers for Assessment of Breast Cancer Response to Neoadjuvant Treatment: A sub-study of the I-SPY 2 TRIAL Version: 1.0 Date: May 28, 2012

More information

American College of Radiology CT Accreditation Program. Testing Instructions

American College of Radiology CT Accreditation Program. Testing Instructions American College of Radiology CT Accreditation Program Testing Instructions (Revised July 24, 2015) This guide provides all of the instructions necessary for clinical tests, phantom tests and general submission

More information

DENTAL Cone beam 3D X-RAY SYSTEM with

DENTAL Cone beam 3D X-RAY SYSTEM with VERSATILE INTUITIVE efficient DENTAL Cone beam 3D X-RAY SYSTEM with dedicated panoramic imaging With thirty years of experience in designing and manufacturing state-of-the-art dental panoramic and tomographic

More information

Data. microcat +SPECT

Data. microcat +SPECT Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for

More information

Chest CT protocols. Mannudeep K. Kalra, MD, DNB. Dianna D. Cody, PhD. Massachusetts General Hospital Harvard Medical School

Chest CT protocols. Mannudeep K. Kalra, MD, DNB. Dianna D. Cody, PhD. Massachusetts General Hospital Harvard Medical School Chest CT protocols Mannudeep K. Kalra, MD, DNB Dianna D. Cody, PhD Massachusetts General Hospital Harvard Medical School M.D. Anderson Cancer Center Specific principles Routine chest CT Lung nodule follow

More information

SWABIK Project. Software Tools for DICOM Media Exchange in Clinical Research: the. SWABIK Project

SWABIK Project. Software Tools for DICOM Media Exchange in Clinical Research: the. SWABIK Project Software Tools for DICOM Media Exchange in Clinical Research: the SWABIK Project Software Tools for DICOM Media Exchange in Clinical Research: the SWABIK Project O. ElGazzar 1, M. Onken 1, M. Eichelberg

More information

Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro.

Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro. www.neurologica.com Critical Care Clinic Operating Room Radiology ED Trauma OR Neuro/Spine Pediatric Plastics Thoracic General Interventional Radiology ENT ED Trauma Trauma Stroke Critical Care Neuro Pediatric

More information

Purchasing a cardiac CT scanner: What the radiologist needs to know

Purchasing a cardiac CT scanner: What the radiologist needs to know Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London maria.lewis@stgeorges.nhs.uk CT scanner development Slice wars 1998 Increased z-coverage

More information

Moving Forward What does this mean for the Medical Physicist and the Imaging Community?

Moving Forward What does this mean for the Medical Physicist and the Imaging Community? Moving Forward What does this mean for the Medical Physicist and the Imaging Community? John M. Boone, Ph.D., FAAPM, FACR Professor and Vice Chairman of Radiology University of California Davis Medical

More information

Stretching the Low Back THERAPIST ASSISTED AND CLIENT SELF-CARE STRETCHES FOR THE LUMBOSACRAL SPINE

Stretching the Low Back THERAPIST ASSISTED AND CLIENT SELF-CARE STRETCHES FOR THE LUMBOSACRAL SPINE EXPERT CONTENT by Joseph E. Muscolino photos by Yanik Chauvin body mechanics THE ESSENCE OF MOST MANUAL THERAPIES, and certainly clinical orthopedic massage therapy, is to loosen taut soft tissues, thereby

More information

MRI DATA PROCESSING. Compiled by: Nicolas F. Lori and Carlos Ferreira. Introduction

MRI DATA PROCESSING. Compiled by: Nicolas F. Lori and Carlos Ferreira. Introduction MRI DATA PROCESSING Compiled by: Nicolas F. Lori and Carlos Ferreira Introduction Magnetic Resonance Imaging (MRI) is a clinical exam that is safe to the patient. Nevertheless, it s very important to attend

More information

MOST Operations Manual page 1 KNEE AND FULL LIMB RADIOGRAPHY

MOST Operations Manual page 1 KNEE AND FULL LIMB RADIOGRAPHY MOST Operations Manual page 1 TABLE OF CONTENTS KNEE AND FULL LIMB RADIOGRAPHY 1. Introduction... 3 2. Background and rationale... 4 3. Equipment and supplies... 4 4. Inclusion/exclusion criteria and safety...

More information

DICOM Viewer 3.0 Instruction for Use

DICOM Viewer 3.0 Instruction for Use Philips Medical Systems Nederland BV ICAP-PF -1/34- DICOM Viewer 3.0 Instruction for Use KONINKLIJKE PHILIPS ELECTRONICS N.V. 2013 All rights are reserved. Reproduction or transmission in whole or in part,

More information

Gestión global de la dosis en TC. Sistema de registro y gestión

Gestión global de la dosis en TC. Sistema de registro y gestión Gestión global de la dosis en TC. Sistema de registro y gestión IV Jornada de Protección Radiológica Hospitalaria SARH FEA Radiofísica Hospital Virgen de las Nieves. Granada jalmansa.lopez@gmail.com 22

More information

Lower Body Exercise One: Glute Bridge

Lower Body Exercise One: Glute Bridge Lower Body Exercise One: Glute Bridge Lying on your back hands by your side, head on the floor. Position your feet shoulder width apart close to your glutes, feet facing forwards. Place a theraband/mini

More information

DICOM Conformance Statement. CT Tomoscan M/EG/EG Compact Release 4.0. Document Number 4522 220 59691 30 March 2000

DICOM Conformance Statement. CT Tomoscan M/EG/EG Compact Release 4.0. Document Number 4522 220 59691 30 March 2000 Philips Medical Systems DICOM Conformance Statement CT Tomoscan M/EG/EG Compact Release 4.0 Document Number 4522 220 59691 30 March 2000 Copyright Philips Medical Systems Nederland B.V. 2000 Philips Medical

More information

Assessing Radiation Dose: How to Do It Right

Assessing Radiation Dose: How to Do It Right Assessing Radiation Dose: How to Do It Right Michael McNitt-Gray, PhD, DABR, FAAPM Professor, Department of Radiology Director, UCLA Biomedical Physics Graduate Program David Geffen School of Medicine

More information

A d v a n c e d 1 2 8

A d v a n c e d 1 2 8 Advanced 128 SCENARIA Advanced 128 Hitachi Innovation at Work for You. Continuing its commitment to the SCENARIA scalable CT platform, Hitachi introduces cutting-edge 128-slice technology with the SCENARIA

More information

Achilles Tendon Repair, Operative Technique

Achilles Tendon Repair, Operative Technique *smith&nephew ANKLE TECHNIQUE GUIDE Achilles Tendon Repair, Operative Technique Prepared in Consultation with: C. Niek van Dijk, MD, PhD KNEE HIP SHOULDER EXTREMITIES Achilles Tendon Repair, Operative

More information

The Pilates Studio of Los Angeles / PilatesCertificationOnline.com

The Pilates Studio of Los Angeles / PilatesCertificationOnline.com Anatomy Review Part I Anatomical Terminology and Review Questions (through pg. 80) Define the following: 1. Sagittal Plane 2. Frontal or Coronal Plane 3. Horizontal Plane 4. Superior 5. Inferior 6. Anterior

More information

VERSYS HERITAGE CDH HIP PROSTHESIS. Surgical Technique for CDH Hip Arthroplasty

VERSYS HERITAGE CDH HIP PROSTHESIS. Surgical Technique for CDH Hip Arthroplasty VERSYS HERITAGE CDH HIP PROSTHESIS Surgical Technique for CDH Hip Arthroplasty SURGICAL TECHNIQUE FOR VERSYS HERITAGE CDH HIP PROSTHESIS CONTENTS ANATOMICAL CONSIDERATIONS....... 2 PREOPERATIVE PLANNING............

More information

Table of Contents. Scan acquisition and user interface basics. Dose modulation and reduction tools. Multi-Slice Detector Geometry

Table of Contents. Scan acquisition and user interface basics. Dose modulation and reduction tools. Multi-Slice Detector Geometry Table of Contents Scan acquisition and user interface basics Dose modulation and reduction tools Multi-Slice Detector Geometry Image Reconstruction and Display Contrast Media Tools Multi-planar formats

More information

ACL Reconstruction with Bone-Tendon-Bone Transplants Using the ENDOBUTTON CL BTB Fixation System

ACL Reconstruction with Bone-Tendon-Bone Transplants Using the ENDOBUTTON CL BTB Fixation System Knee Series Technique Guide ACL Reconstruction with Bone-Tendon-Bone Transplants Using the ENDOBUTTON CL BTB Fixation System Stuart Fromm, M.D. As described by: Stuart Fromm, M.D. Black Hills Orthopedic

More information

Radiographic Procedure Manual for Examinations of the Knee, Hand, Pelvis and Lower Limbs. Osteoarthritis Initiative: A Knee Health Study

Radiographic Procedure Manual for Examinations of the Knee, Hand, Pelvis and Lower Limbs. Osteoarthritis Initiative: A Knee Health Study Radiographic Procedure Manual for Radiographic Procedure Manual for Examinations of the Knee, Hand, Pelvis and Lower Limbs Osteoarthritis Initiative: A Knee Health Study SYN RC W H E R E S C I E N C E

More information

BODY BODY PEDICLE PEDICLE TRANSVERSE TRANSVERSE PROCESS PROCESS

BODY BODY PEDICLE PEDICLE TRANSVERSE TRANSVERSE PROCESS PROCESS Learning Objective Radiology Anatomy of the Spine and Upper Extremity Identify anatomic structures of the spine and upper extremities on standard radiographic and cross-sectional images Timothy J. Mosher,

More information

Requests. Who requests a DXA scan? DXA. DXA Technology. Adequate Clinical Information

Requests. Who requests a DXA scan? DXA. DXA Technology. Adequate Clinical Information Requests Define a protocol for validation and prioritisation of densitometry requests. Registered medical practitioners In writing Include adequate clinical information Unsuitable requests should be discussed

More information

IntelliSpace PACS 4.4. Image Enabled EMR Workflow and Enterprise Overview. Learning Objectives

IntelliSpace PACS 4.4. Image Enabled EMR Workflow and Enterprise Overview. Learning Objectives IntelliSpace PACS 4.4 Image Enabled EMR Workflow and Enterprise Overview Learning Objectives Use the Cerner EMR to view images and reports via IntelliSpace PACS: Image Launch Workflow (SUBI Image Viewer)

More information

HDI 4000 Ultrasound System

HDI 4000 Ultrasound System HDI 4000 Ultrasound System DICOM Conformance Statement Rev 1.0 Philips Medical Systems HDI 4000 DICOM Conformance Rev. 1.0 12/17/2001 Philips Medical Systems Table of Contents 0. Introduction...3 0.1 DICOM

More information

Multi-slice Helical CT Scanning of the Chest

Multi-slice Helical CT Scanning of the Chest Multi-slice Helical CT Scanning of the Chest Comparison of different low-dose acquisitions Lung cancer is the main cause of deaths due to cancer in human males and the incidence is constantly increasing.

More information

Introduction. I. Objectives. II. Introduction. A. To become familiar with the terms of direction and location.

Introduction. I. Objectives. II. Introduction. A. To become familiar with the terms of direction and location. E X E R C I S E Introduction I. Objectives A. To become familiar with the terms of direction and location. B. To become familiar with different types of planes and sections. C. To learn the names and locations

More information

The Exposure Index and Its Standardization

The Exposure Index and Its Standardization The Exposure Index and Its Standardization Ulrich Neitzel Philips Medical Systems Clinical Science Hamburg, Germany ulrich.neitzel@philips.com Content The Exposure Index (EI) What is it? What is it good

More information

Massage and Movement

Massage and Movement Massage and Movement Incorporating Movement into Massage Part One: Theory and Technique in Prone With Lee Stang, LMT NCBTMB #450217-06 1850 West Street Southington, CT 06489 860.747.6388 www.bridgestohealthseminars.com

More information

Anterior Superior Iliac Spine. Anterior Inferior Iliac Spine. head neck greater trochanter intertrochanteric line lesser trochanter

Anterior Superior Iliac Spine. Anterior Inferior Iliac Spine. head neck greater trochanter intertrochanteric line lesser trochanter Ilium Bones The Skeleton Ischium Pubis Sacro-iliac Joint Iliac Crest Anterior Superior Superior Pubic Ramus Anterior Inferior Acetabulum Obturator Foramen Ischio-pubic ramus Ischial tuberosity Pubic Crest

More information

Musculoskeletal MRI Technical Considerations

Musculoskeletal MRI Technical Considerations Musculoskeletal MRI Technical Considerations Garry E. Gold, M.D. Professor of Radiology, Bioengineering and Orthopaedic Surgery Stanford University Outline Joint Structure Image Contrast Protocols: 3.0T

More information

Laerdal' Human Anatomy Manual The Skeleton

Laerdal' Human Anatomy Manual The Skeleton Human Anatomy Manual The Skeleton Laerdal Texas P.O. Box 38.226 EM. 116 Gatesville,Texas U.S.A.76528 U.S.A.1-800-433-5539 IntemationaI1-254-865-7221 24 Hour Fax 254-865-8011 ~ Laerdal' TABLE OF CONTENTS

More information

The Whys, Hows and Whats of the Noise Power Spectrum. Helge Pettersen, Haukeland University Hospital, NO

The Whys, Hows and Whats of the Noise Power Spectrum. Helge Pettersen, Haukeland University Hospital, NO The Whys, Hows and Whats of the Noise Power Spectrum Helge Pettersen, Haukeland University Hospital, NO Introduction to the Noise Power Spectrum Before diving into NPS curves, we need Fourier transforms

More information

TransFx External Fixation System Large and Intermediate Surgical Technique

TransFx External Fixation System Large and Intermediate Surgical Technique TransFx External Fixation System Large and Intermediate Surgical Technique TransFx External Fixation System Large and Intermediate Surgical Technique 1 Surgical Technique For TransFx External Fixation

More information

Veraviewepocs 3D R100 & F40

Veraviewepocs 3D R100 & F40 Veraviewepocs 3D R100 & F40 Innovative 3D Reuleaux Full Arch FOV Thinking ahead. Focused on life. Veraviewepocs 3D R100 A New Frontier in X-ray Diagnostics Veraviewepocs 3D R100 has changed the shape of

More information

Anatomy and Pathomechanics of the Sacrum and Pelvis. Charles R. Thompson Head Athletic Trainer Princeton University

Anatomy and Pathomechanics of the Sacrum and Pelvis. Charles R. Thompson Head Athletic Trainer Princeton University Anatomy and Pathomechanics of the Sacrum and Pelvis Charles R. Thompson Head Athletic Trainer Princeton University Simplify Everything There are actually only three bones: Two innominates, one sacrum.

More information

THE BENJAMIN INSTITUTE PRESENTS. Excerpt from Listen To Your Pain. Assessment & Treatment of. Low Back Pain. Ben E. Benjamin, Ph.D.

THE BENJAMIN INSTITUTE PRESENTS. Excerpt from Listen To Your Pain. Assessment & Treatment of. Low Back Pain. Ben E. Benjamin, Ph.D. THE BENJAMIN INSTITUTE PRESENTS Excerpt from Listen To Your Pain Assessment & Treatment of Low Back Pain A B E N J A M I N I N S T I T U T E E B O O K Ben E. Benjamin, Ph.D. 2 THERAPIST/CLIENT MANUAL The

More information

Veraviewepocs 3D R100 & F40

Veraviewepocs 3D R100 & F40 Veraviewepocs 3D R100 & F40 Innovative 3D Reuleaux Full Arch FOV Thinking ahead. Focused on life. Veraviewepocs 3D R100 A New Frontier in X-ray Diagnostics Veraviewepocs 3D R100 has changed the shape of

More information

Construction of human knee bone joint model using FDM based 3D Printer from CT scan images.

Construction of human knee bone joint model using FDM based 3D Printer from CT scan images. Biomedical Research 2015; 26 (4): S1-5 ISSN 0970-938X www.biomedres.info Construction of human knee bone joint model using FDM based 3D Printer from CT scan images. Marlon Jones Louis a *, R. Malayalamurthi

More information

ABSTRACT INTRODUCTION

ABSTRACT INTRODUCTION UTILIZING COMPUTED TOMOGRAPHY SCANS FOR ANALYSIS OF MOTORCYCLE HELMETS IN REAL-WORLD CRASHES Kathryn L. Loftis 1, Daniel P. Moreno 1, Joshua Tan 2, Hampton C. Gabler 1, Joel D. Stitzel 1 1 Virginia Tech-

More information

Posterior Referencing. Surgical Technique

Posterior Referencing. Surgical Technique Posterior Referencing Surgical Technique Posterior Referencing Surgical Technique INTRO Introduction Instrumentation Successful total knee arthroplasty depends in part on re-establishment of normal lower

More information

Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Gantry Rotation Time. mas (Reg-Lg) 40-80

Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Gantry Rotation Time. mas (Reg-Lg) 40-80 Revisions Effective January 2012 Chest 1: Pulmonary Nodule Follow-up: Low-Dose Helical CT (Unenhanced) (Non-metastatic) Technologist Instructions Patient must cough several times prior to scan to clear

More information

(in: Skeletal Radiology - Vol. 30, n. 3 - March 2001)

(in: Skeletal Radiology - Vol. 30, n. 3 - March 2001) ACUTE AVULSION FRACTURES OF THE PELVIS IN ADOLESCENT COMPETITIVE ATHLETES: PREVALENCE, LOCATION AND SPORTS DISTRIBUTION OF 203 CASES COLLECTED Folco Rossi - Stefano Dragoni (in: Skeletal Radiology - Vol.

More information

Skeletal Surveys in Suspected Non- Accidental Injury (NAI)

Skeletal Surveys in Suspected Non- Accidental Injury (NAI) This is an official Northern Trust policy and should not be edited in any way Skeletal Surveys in Suspected Non- Accidental Injury (NAI) Reference Number: NHSCT/11/410 Target audience: This protocol is

More information

Musculoskeletal Ultrasound Technical Guidelines. IV. Hip

Musculoskeletal Ultrasound Technical Guidelines. IV. Hip European Society of MusculoSkeletal Radiology Musculoskeletal Ultrasound Technical Guidelines IV. Ian Beggs, UK Stefano Bianchi, Switzerland Angel Bueno, Spain Michel Cohen, France Michel Court-Payen,

More information

Cone Beam Implant Planning Manual

Cone Beam Implant Planning Manual Cone Beam Implant Planning Manual (Everything you want to know about CBCT planning) Page 2 Overview & Guide Selection Criteria Page 3 Scan Appliance basics When are they required? Page 4 Scan Appliance

More information

MET: Posterior (backward) Rotation of the Innominate Bone.

MET: Posterior (backward) Rotation of the Innominate Bone. MET: Posterior (backward) Rotation of the Innominate Bone. Purpose: To reduce an anterior rotation of the innominate bone at the SI joint. To increase posterior (backward) rotation of the SI joint. Precautions:

More information

Sports Radiology : Making a diagnosis or solving a problem

Sports Radiology : Making a diagnosis or solving a problem Sports Radiology : Making a diagnosis or solving a problem Professor Dr John George University of Malaya Research Imaging Centre University of Malaya Medical Centre Kuala Lumpur, Malaysia. Abstract: Aim:

More information

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington

MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical

More information

DICOM Correction Item

DICOM Correction Item Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes

More information

Arthroscopic Shoulder Instability Repair Using the SUTUREFIX ULTRA Suture Anchor and SUTUREFIX ULTRA Instrumentation System

Arthroscopic Shoulder Instability Repair Using the SUTUREFIX ULTRA Suture Anchor and SUTUREFIX ULTRA Instrumentation System *smith&nephew SHOULDER TECHNIQUE GUIDE Arthroscopic Shoulder Instability Repair Using the SUTUREFIX ULTRA Suture Anchor and SUTUREFIX ULTRA Instrumentation System KNEE HIP SHOULDER EXTREMITIES Arthroscopic

More information

Voxar 3D TM. A suite of advanced visualization and analysis software tools

Voxar 3D TM. A suite of advanced visualization and analysis software tools Voxar 3D TM A suite of advanced visualization and analysis software tools The power to deliver advanced visualization throughout the enterprise To effectively manage the rapid growth of large volumetric

More information

Etude POPART'MUS MRI Component

Etude POPART'MUS MRI Component TECHNICAL SURVEY Dear Investigators, This document is a Technical Survey which provides the teams of THERALYS and of the Pierre Wertheimer Neurological Hospital of Lyon with an overview of your site s

More information

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999

Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999 Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences mmcnittgray@mednet.ucla.edu X-Ray

More information

Presentation Objective: Benefits to New Employee. Goal for Participants. Benefits to the Administration

Presentation Objective: Benefits to New Employee. Goal for Participants. Benefits to the Administration Imaging Department Quick Step How to Create and Implement a Rapid Integration Orientation New Employee Orientation builds a bridge between the old job and the new job for better quality in health care

More information

OfficePACS Power. Digital Imaging for the Orthopaedic Workflow

OfficePACS Power. Digital Imaging for the Orthopaedic Workflow OfficePACS Power Digital Imaging for the Orthopaedic Workflow Find Patients Faster Find Studies Faster PowerSearch makes it easy to quickly search a wide range of demographic data from a single text box.

More information

Imaging of Thoracic Endovascular Stent-Grafts

Imaging of Thoracic Endovascular Stent-Grafts Imaging of Thoracic Endovascular Stent-Grafts Tariq Hameed, M.D. Department of Radiology and Imaging Sciences, Indiana University School of Medicine, Indianapolis, Indiana Disclosures: No relevant financial

More information

The power of Multislice CT Philips MX 6000 Dual

The power of Multislice CT Philips MX 6000 Dual The power of Multislice CT Philips MX 6000 Dual Philips MX systems bring new levels of quality, simplicity and reliability to CT imaging. Designed by the same people that brought the world Philips CT,

More information

Diagnostic Imaging Exams

Diagnostic Imaging Exams Guide for Chiropractors Diagnostic Imaging Exams CREATED FOR OUR CHIROPRACTIC PARTNERS This document has been prepared by the specialized, board-certified radiologists who interpret patient exams for Center

More information

Configuration and Use of the MR Knee Dot Engine

Configuration and Use of the MR Knee Dot Engine Configuration and Use of the MR Knee Dot Engine Kai Reiter (RT); Sebastian Auer (RT) Radiology Herne, Herne, Germany 1 42 MAGNETOM Flash 3/2011 www.siemens.com/magnetom-world We are a radiological joint

More information

The Dedicated MRI Factbook. Esaote MRI: A world of difference

The Dedicated MRI Factbook. Esaote MRI: A world of difference The Dedicated MRI Factbook Esaote MRI: A world of difference Esaote Dedicated MRI Measures of Excellence Creativity in Healthcare is Esaote s motto which represents the philosophy that has been adopted

More information

IAEA HUMAN HEALTH SERIES

IAEA HUMAN HEALTH SERIES IAEA HUMAN HEALTH SERIES No. 19 Quality Assurance Programme for Computed Tomography: Diagnostic and Therapy Applications IAEA HUMAN HEALTH SERIES PUBLICATIONS The mandate of the IAEA human health programme

More information

Integrating Postprocessed Functional MR Images with Picture Archiving and Communication Systems

Integrating Postprocessed Functional MR Images with Picture Archiving and Communication Systems Integrating Postprocessed Functional MR Images with Picture Archiving and Communication Systems Joseph A. Maldjian, John Listerud, and Satjeet Khalsa AJNR Am J Neuroradiol 23:1393 1397, September 2002

More information

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications)

Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Toshiba Excelart Vantage 1.5T MRI Tech Specs (Technical Specifications) Excelart Vantage Magnet Configuration: Ultra-short-bore Strength (or W x H): 1.5 T Homogeneity, ppm V-RMS: Dimensions of maximum

More information

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography

2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography Today s agenda Computed Tomography Presented by: Dina Hernandez, BSRT, RT (R), CT, QM Krista Bush, RT, MBA Leonard Lucey, JD ACR Quality & Safety MIPPA overview and CMS requirements CT accreditation How

More information

Page: 1 of 6 Page: 1 of 6

Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 Page: 1 of 6 CR Basics and FAQ Overview Computed Radiography is a term used to describe a system that electronically records a radiographic image. Computed Radiographic systems use unique

More information

A PACS-Aware DICOM Image Object

A PACS-Aware DICOM Image Object Designing and Implementing A PACS-Aware DICOM Image Object For Digital X-ray, Mammography and Intraoral Applications David A. Clunie Quintiles Intelligent Imaging Clear Vision for the Healthcare Industry

More information

GE Healthcare. DoseWatch. Gathering Radiation Dose Data

GE Healthcare. DoseWatch. Gathering Radiation Dose Data GE Healthcare DoseWatch Gathering Radiation Dose Data Gathering Radiation Dose Data After your organization crosses the bridge of Why? for patient radiation dose management either due to regulation, risk

More information

Addressing Pelvic Rotation

Addressing Pelvic Rotation 1 of 5 4/20/2008 10:31 AM http://www.strengthcoach.com Addressing Pelvic Rotation Aaron Brooks of Perfect Postures When trying to address your athlete's or client's limitations due to pain or joint restriction,

More information

Digital Image Processing with DragonflyPACS

Digital Image Processing with DragonflyPACS Digital Image Processing with DragonflyPACS The Smarter PACS The smartest image processing system in veterinary medicine. Outstanding image quality that s compliant with DICOM 3.0 standards. Compatible

More information

Printlink5-ID_IV. DICOM 3.0 Conformance Statement PRINT MANAGEMENT SYSTEM CODE NO. 1118. Manufacturer:

Printlink5-ID_IV. DICOM 3.0 Conformance Statement PRINT MANAGEMENT SYSTEM CODE NO. 1118. Manufacturer: PRINT MANAGEMENT SYSTEM Printlink5-ID_IV DICOM 3.0 Conformance Statement CODE NO. 1118 Manufacturer: 2970 Ishikawa-machi, Hachioji-shi Tokyo 192-8505 Japan DICOM Conformance Statement Important Notes Konica

More information

FUNCTIONAL STRENGTHENING

FUNCTIONAL STRENGTHENING FUNCTIONAL STRENGTHENING *This group of exercises are designed to increase leg and core endurance and stability as related to bipedal functional activities. *Perform this program for 4-6 weeks. It is ideal

More information

DSM Spine+Sport - Mobility

DSM Spine+Sport - Mobility To set yourself up for success, practice keeping a neutral spine throughout all of these movements. This will ensure the tissue mobilization is being applied to the correct area, and make the techniques

More information

MEDICAL IMAGING. 90/30 Remote Controlled RF System

MEDICAL IMAGING. 90/30 Remote Controlled RF System MEDICAL IMAGING 90/30 Remote Controlled RF System The Easy solution for your Remote Controlled room Apollo EZ represents the Easy approach for the radiographic/fluoroscopic room and it comes from the experience

More information

Tracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives

Tracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives Tracking Radiation Exposure From Medical Diagnostic Procedures: Siemens Perspectives Gilbert W. Beebe Symposium The National Academies Katharine Grant, PhD Staff Scientist 8 December 2011 For internal

More information

Siemens Computed Tomography

Siemens Computed Tomography Siemens Computed Tomography Innovation Leader. The Cosmopolitan Information Topic Waltraud Winter Siemens Copyright Siemens AG 2011. All rights reserved. Page 1 October 2012 Waltraud Winter An Innovation

More information

Development of pediatric CT protocols to obtain minimal doses in PET/CT studies a physicists approach.

Development of pediatric CT protocols to obtain minimal doses in PET/CT studies a physicists approach. Development of pediatric CT protocols to obtain minimal doses in PET/CT studies a physicists approach. Wibeke Nordhøy 1, Mehrnaz Sheikhaeri 2, Alise Larsen 1, Otto Glomset 1 Til slutt et utvalg fra foredraget

More information

Target Volumes for Anal Carcinoma For RTOG 0529

Target Volumes for Anal Carcinoma For RTOG 0529 Target Volumes for Anal Carcinoma For RTOG 0529 Robert Myerson, M.D. Ph.D.*, Lisa Kachnic, M.D.**, Jacqueline Esthappan, Ph.D.*,Parag Parikh M.D.*William Straube M.S.*, John Willins, Ph.D.** *Washington

More information

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT

PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT PRACTICAL TIPS IN ENSURING RADIATION SAFETY IN THE USE OF MEDICAL DIAGNOSTIC X-RAY EQUIPMENT Although the medical uses of X-rays to examine a patient without surgery became an amazing medical breakthrough,

More information

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals

Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Rb 82 Cardiac PET Scanning Protocols and Dosimetry Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Overview Rb 82 myocardial perfusion imaging protocols Acquisition Reconstruction

More information

Basic Radiographic Procedures

Basic Radiographic Procedures Basic Radiographic Procedures The GPT Method G. Patrick Thomas, Jr., DC, DACBR Copyright 2002 G. Patrick Thomas, Jr., DC, DACBR PO Box 1000 Blue Springs, MO 64013-1000 gptdacbr@juno.com Table of Contents

More information

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of

More information

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13 Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer

More information

DIGITAL X-RAY AUNTMINNIE.COM X-RAY PATIENT POSITIONING MANUAL HTTP://XRAY.AUNTMINNIE.COM. Proper positioning for the pelvis and proximal femur

DIGITAL X-RAY AUNTMINNIE.COM X-RAY PATIENT POSITIONING MANUAL HTTP://XRAY.AUNTMINNIE.COM. Proper positioning for the pelvis and proximal femur A R T I C L E S Proper positioning for the pelvis and proximal femur The lowdown on lumbar spine positioning Radiographic positioning techniques for the cervical spine Boning up on humerus, clavicle, and

More information

How To Fix A Radial Head Plate

How To Fix A Radial Head Plate Mayo Clinic CoNGRUENT RADIAL HEAD PLATE Since 1988 Acumed has been designing solutions to the demanding situations facing orthopedic surgeons, hospitals and their patients. Our strategy has been to know

More information

R/F. Applications and Present Issues of Tomosynthesis in Spine Surgery. 87th Annual Meeting of the Japanese Orthopaedic Association Evening Seminar 4

R/F. Applications and Present Issues of Tomosynthesis in Spine Surgery. 87th Annual Meeting of the Japanese Orthopaedic Association Evening Seminar 4 R/F Applications and Present Issues of Tomosynthesis in Spine Surgery 87th Annual Meeting of the Japanese Orthopaedic Association Evening Seminar 4 Professor and Chairman Department of Orthopaedic Surgery

More information