Performance evaluation cobas b 123 POC system

Size: px
Start display at page:

Download "Performance evaluation cobas b 123 POC system"

Transcription

1 Summary These externally validated performance evaluations confirm that the results obtained with the cobas b 123 POC system for 17 commonly assessed critical care parameters are as true and accurate as those obtained from well-recognized and widely used analyzers, such as the cobas b 221 analyzer widely regarded as the gold standard BGE and the cobas 6 analyzer series. The is an attractive and proven solution for BGE at the point of care, designed for performance at low risk to the hospital. Performance evaluation Introduction This proof-of-performance white paper demonstrates that the cobas b 123 POC (point-of-care) system is as true and accurate in its measurement of 17 critical care parameters (see below) as well-established and widely used analyzers, such as the cobas b 221 analyzer, and the cobas 6 analyzer series. In order to demonstrate this, the document details the externally validated performance evaluations conducted with this POC system in order that potential users can have complete confidence that the results obtained for the critical care parameters assessed using this system are equivalent to those obtained with other well-recognised, widely-used analyzers. 17 parameters evaluated with the cobas b 123 POC system: ph, blood gas (BG: po2, pco2), electrolytes (ISE: Na, K, Cl, Ca), metabolites (Glu and Lac), COOX (hematocrit (Hct), total hemoglobin (thb), oxygen saturation (SO2), hemoglobin derivates (O2Hb, HHb, COHb, MetHb)) and bilirubin (total). Key conclusions The acceptance criteria for trueness and accuracy and the acceptance criteria repeatability and intermediate precision with all sites combined were fulfilled. These externally validated performance evaluations confirm that the results obtained with the for 17 commonly assessed critical care parameters are as true and accurate as those obtained from well-recognized and widely used analyzers, such as the cobas b 221 analyzer widely regarded as the gold standard BGE and the cobas 6 analyzer series. The is an attractive and proven solution for BGE at the point of care, designed for performance at low risk to the hospital. * For investigational use only in the U.S. CAUTION Investigational device limited by Federal (or United States) Law to investigational use The performance characteristics of this product have not been established. Not available in the U.S. pending FDA clearance. COBAS, COBAS B, COBAS C and LIFE NEEDS ANSWERS are trademarks of Roche. 211 Roche Roche Diagnostics Ltd. CH-6343 Rotkreuz Switzerland

2 Overview of methods for external proof of performance evaluations performed for the cobas b 123 POC system The objective of the external performance evaluation studies was to investigate the performance of cobas b 123 POC system against widely-used and validated analyzers. Performance evaluation study: a. Method comparisons of the with the for ph, pco2, po2, Na, K, Cl, Ca, Hct, O2Hb, HHb, COHb, MetHb, thb and SO2, and compared with cobas 6 analyzer series for glucose and lactate. These were performed externally at six sites in Europe [see below] (2 residual samples per lot; 1 per site and lot). b. Determination of the intermediate precision of the QC results (daily quality control) over 21 days performed externally at six sites in Europe for three different lots at the same test sites [Note; this excluded bilirubin]. The six European evaluation sites were: Ludwigshafen, Hamburg, Graz, Zwolle, Göteborg and Cambridge. External performance evaluation study, bilirubin: a. Method comparisons of the for bilirubin compared with the cobas c 51 module (part of the cobas 6 analyzer series) performed externally at three US sites (see below). b. Determination of the intermediate precision of the QC results (daily quality control) over 21 days performed externally in three US sites.* The three US evaluation sites were: Galveston, TX, Louisville, KY and San Diego, CA. Additional information on the studies Heparinised arterial or venous whole blood samples were used throughout and fresh arterial, venous or capillary lithium heparin blood samples for the bilirubin evaluation. For determination of results within the cobas b 123 POC system, 15μL samples were required unless otherwise stated. Quality controls were performed on each measuring day of the separate study phases. Samples were measured with both of the used sensor cartridge lots in three different QC levels over 21 days in duplicate determinations (Intermediate precision; daily quality control). In situations where over 21 measuring days were needed, they were tested twice daily once only. A familiarisation phase was performed prior to the studies with one lot only to ensure that the study operators were suitably familiar with the handling and procedures of the new cobas b 123 POC analyser. Results for each of the 17 parameters assessed externally for the in external proof of performance evaluations The following pages summarise the main results obtained in the performance evaluations when each of the 17 parameters [see page 1] analyzed with the were compared against the reference analyzer [as stated within the results of each parameter]. Parameter: ph Reference analyzer: n = 691, slope =.99, intercept =.8, r =.99, median absolute bias =.9, accuracy 1 % Table 1. Acceptance criteria for ph method comparison with blood samples outside the typical range ph 1 Potentiometry ph 1 Potentiometry Figure 1. Trueness overall and accuracy overall: ph Key point: The was shown to be equivalent to the cobas b 221 analyzer for assessment of ph for trueness overall and accuracy overall in this externally validated performance evaluation in 691 samples. Y = X = +.78 md(95) = ph 1 Potentiometry N = 691 r =.9865 t =.8943

3 Blood gases Parameter: po2 (mmhg) Reference analyzer: n = 677, slope =.97, r = 1., median absolute bias = -3.1 mmhg, accuracy 98 % Parameter: pco2 (mmhg) Reference analyzer: n = 687, slope =.95, r =.99, median absolute bias = -.2 mmhg, accuracy 1 % Table 2. Acceptance criteria for po2 (mmhg) method comparison with blood samples outside the typical range Table 3. Acceptance criteria for pco2 (mmhg) method comparison with blood samples outside the typical range Y =.9864 X = md(95) = N = 677 r =.9963 t =.9587 Y =.953 X = md(95) = 1.36 N = 687 r =.9933 t =.9162 po2 3 Amperometry pco2 4 Potentiometry po2 3 Amperometry po2 3 Amperometry pco2 4 Potentiometry pco2 4 Potentiometry Figure 2. Trueness overall and accuracy overall: po2 Figure 3. Trueness overall and accuracy overall: pco2 POC system and the for po2 was good in the 677 samples assessed in this externally validated performance evaluation, and the accuracy for po2 was 98 %. POC system and the for pco2 was good, and the accuracy for pco2 was 1 % in 687 samples assessed in this externally validated performance evaluation.

4 Electrolytes Parameter: sodium (mmol/l) Reference analyzer: n = 691, slope = 1.3, r =.94, mean absolute bias =.2 %, accuracy 1 % Parameter: potassium (mmol/l) Reference analyzer: n = 682, slope =.96, r =.99, mean absolute bias = -.3 mmol/l, accuracy 99.9 % Table 4. Acceptance criteria for sodium (mmol/l) method comparison with blood samples outside the typical range Table 5. Acceptance criteria for potassium (mmol/l) method comparison with blood samples outside the typical range Y = 1.34 X = md(95) = N = 691 r =.9396 t =.76 Y =.956 X = md(95) = -.85 N = 682 r =.9947 t = Correction factor used! Na 6 Potentiometry (Y X) / X [%] 8. % % 4. % %. % -% -4.% -% -8.% K 7 Potentiometry Na 6 Potentiometry Na 6 Potentiometry K 7 Potentiometry K 7 Potentiometry Figure 4. Trueness overall and accuracy overall: sodium Figure 5. Trueness overall and accuracy overall: potassium Key point: Overall, the method comparison of the vs. showed comparable results for Na, with 1 % of the values within the 95 percentile. Key point: Overall, the method comparison of the vs. showed a close correlation for K, with 1 % of the values within the 95 percentile.

5 Parameter: chloride (mmol/l) Reference analyzer: n = 657, slope = 1.15, r =.98, mean absolute bias = 1.17 %, accuracy 1 % Parameter: calcium (mmol/l) Reference analyzer: n = 689, slope =.95, r =.94, mean absolute bias = -.1 mmol/l, accuracy 1 % Table 6. Acceptance criteria for chloride (mmol/l) method comparison with blood samples outside the typical range Table 7. Acceptance criteria for calcium (mmol/l) method comparison with blood samples outside the typical range Y = X = md(95) = 1.32 N = 657 r =.9832 t =.8821 Y =.948 X = +.48 md(95) =.53 N = 689 r =.9394 t = Correction factor used! 2.5 Correction factor used! Cl 9 Potentiometry (Y X) / X [%] 15.% 1.% 5.%.% -5. % -1.% Ca 8 Potentiometry % Cl 9 Potentiometry Cl 9 Potentiometry Ca 8 Potentiometry Ca 8 Potentiometry Figure 6. Trueness overall and accuracy overall: chloride Figure 7. Trueness overall and accuracy overall: calcium Key point: Overall, the method comparison of the vs. showed a close correlation for Cl, with 1 % of the values within the 95 percentile. Key point: Overall, the method comparison of the vs. showed comparable results for Ca, with 1 % of the values within the 95 percentile.

6 Metabolites Parameter: glucose (mmol/l) Reference platform: cobas 6 analyzer series n = 686, slope =.98, r =.98, median relative bias = -2.4 %, accuracy = 99.7 % Parameter: lactate (mmol/l) Reference platform: cobas 6 analyzer series n = 59, slope = 1., r =.99, median relative bias =.6 %, accuracy = 1 % Table 8. Acceptance criteria for glucose (mmol/l) method comparison with blood samples outside the typical range Table 9. Acceptance criteria for lactate (mmol/l) method comparison with blood samples outside the typical range Y =.982 X = -.41 md(95) =.581 N = 686 r =.9794 t =.8698 Y = 1.33 X = -.2 md(95) =.6 N = 59 r =.9816 t =.8666 GLUC 18 Photometry (Y X) / X [%] 3.% 2.% 1.%.% -1.% -2.% -3.% LAC 3 Amperometry GLUC 18 Photometry cobas 6 analyzer series GLUC 18 Photometry cobas 6 analyzer series LAC 3 Photometry cobas 6 analyzer series LAC 3 Photometry cobas 6 analyzer series Figure 8. Trueness overall and accuracy overall: glucose Figure 9. Trueness overall and accuracy overall: lactate Key point: Overall, there was good correlation between the and the cobas 6 analyzer series for glucose in the 686 samples evaluated, with the correlation coefficient for all lots being.98 for glucose. The bias of -2.4 % for glucose on the was within the acceptable range. Key point: Overall, there was good correlation between the cobas 6 analyzer series and the for lactate in the 59 samples evaluated, with the correlation coefficient for all lots being 1. for lactate and no samples falling outside the 95 % accuracy limit.

7 Hematocrit Parameter: hematocrit (%) Reference analyzer: n = 691, slope.95, r =.99, mean absolute bias = -.1 %, accuracy = 99.7 % Total hemoglobin Parameter: total hemoglobin (thb) (g/dl) Reference analyzer: n = 682, slope.95, r =.98, mean absolute bias =.3 g/dl, accuracy = 99.3 % Table 1. Acceptance criteria for hematocrit (%) method comparison with blood samples outside the typical range Table 11. Acceptance criteria for total hemoglobin (thb) (g/dl) method comparison with blood samples outside the typical range Hct 5 Conductivity Hct 5 Conductivity Figure 1. Trueness overall and accuracy overall: hematocrit Y = X = md(95) = Hct 5 Conductivity N = 691 r =.9891 t =.9227 thb Coox 1 Spectro-photometrical thb Coox 1 Spectro-photometrical Figure 11. Trueness overall and accuracy overall: total hemoglobin (thb) Y = X = md(95) =.444 N = 682 r =.9812 t = thb Coox 1 Spectro-photometrical POC system and the for hematocrit was good, and the accuracy for hematocrit was almost 1 %. POC system and the for thb was good, and the accuracy for thb was 99.3 %, with only few outliers.

8 Hemoglobin derivates Parameter: COHb (%) Reference analyzer: n = 682, slope 1., r =.97, mean absolute bias =. %, accuracy = 1 % Parameter: HHb (%) Reference analyzer: n = 668, slope 1.1, r =.99, mean absolute bias = -.6 %, accuracy = 98.8 % Table 12. Acceptance criteria for COHb (%) method comparison with blood samples outside the typical range Table 13. Acceptance criteria for HHb (%) method comparison with blood samples outside the typical range COHb 13 Spectro-photometrical Y = X = +. md(95) =.212 N = 682 r =.9678 t = HHb 16 Spectro-photometrical Y = X = md(95) = 1.41 N = 668 r =.9975 t = COHb 13 Spectro-photometrical COHb 13 Spectro-photometrical HHb 16 Spectro-photometrical HHb 16 Spectro-photometrical Figure 12. Trueness overall and accuracy overall: COHb Figure 13. Trueness overall and accuracy overall: HHb POC system and the for COHb was excellent. POC system and the for HHb was excellent, with a slight negative bias due to some outliers.

9 Parameter: MetHb (%) Reference analyzer: n = 682, slope 1., r =.67, mean absolute bias =.1 %, accuracy = 1 % Parameter: O2Hb (%) Reference analyzer: n = 666, slope 1., r =.99, mean absolute bias =.5 %, accuracy = 99.4 % Table 14. Acceptance criteria for MetHb (%) method comparison with blood samples outside the typical range Table 15. Acceptance criteria for O2Hb (%) method comparison with blood samples outside the typical range Y = 1. X = +.1 md(95) =.212 N = 682 r =.6665 t =.5692 Y = 1.7 X = md(95) = N = 666 r =.9973 t =.918 MetHb 14 Spectro-photometrical MetHb 14 Spectro-photometrical MetHb 14 Spectro-photometrical O2Hb 12 Spectro-photometrical O2Hb 12 Spectro-photometrical O2Hb 12 Spectro-photometrical Figure 14. Trueness overall and accuracy overall: MetHb Figure 15. Trueness overall and accuracy overall: O2Hb POC system and the for MetHb was good. The Pearson s coefficient is of limited value here. POC system and the for O2Hb was excellent, with good accuracy and very few outliers.

10 Oxygen saturation Parameter: SO2 (%) Reference analyzer: n = 666, slope 1.1, r =.99, mean absolute bias =.6 %, accuracy = 99.2 % Total bilirubin (neonatal) Parameter: total bilirubin (neonatal blood) (mg/dl) Reference analyzer: cobas c 51 module (part of cobas 6 analyzer series) n = 139, slope 1.26, r =.92, accuracy = 95.7 %, mean overall pooled bias = -.5, median overall pooled bias = Table 16. Acceptance criteria for SO2 (%) method comparison with blood samples outside the typical range Table 17. Acceptance criteria for total bilirubin (neonatal blood) (mg/dl) method comparison with blood samples outside the typical range Y = 1.15 X = md(95) = N = 666 r =.9973 t =.975 Y =.9816 X = md(95) = N = 148 r =.9855 t = SO2 11 Specto-photometrical Bili Photometry SO2 11 Specto-photometrical SO2 11 Specto-photometrical Doumas et al. Doumas et al. Figure 16. Trueness overall and accuracy overall: SO2 Figure 17. Trueness overall and accuracy overall: pooled data from three US sites, bilirubin POC system and the for SO2 was excellent, with good accuracy and very few outliers. Key point: Generally, there was a good agreement between bilirubin concentrations measured between and cobas c 51 module (part of cobas 6 analyzer series). The slope of the regression lines were close to 1, and no significant systematic bias between the two test methods was observed

GEM Premier 3000, now with iqm The New Standard for Critical Care Testing

GEM Premier 3000, now with iqm The New Standard for Critical Care Testing GEM Premier 3000, now with iqm The New Standard for Critical Care Testing ph, Blood Gas Electrolytes Metabolites Hematocrit CO-Oximetry Coagulation GEM Premier 3000, now with iqm Simple. Smart. Standar

More information

Portex Arterial Blood Sampling Kits

Portex Arterial Blood Sampling Kits Portex Arterial Blood Sampling Kits Safe Application. Simple Activation. Consistent Results. Portex Arterial Blood Sampling Kits > Compatible with Multi-parameter Analyzers > Balanced Heparin Provides

More information

C RITICAL C ARE 4. Blood Gas Analysis, Oxygen Transport, Electrolytes Balance, Glucose and Bilirubin Determination

C RITICAL C ARE 4. Blood Gas Analysis, Oxygen Transport, Electrolytes Balance, Glucose and Bilirubin Determination C RITICAL C ARE 4 R. Galli D. Colombo M. Ferrario Blood Gas Analysis, Oxygen Transport, Electrolytes Balance, Glucose and Bilirubin Determination A Short Survey on Measured, Calculated and Special Parameters

More information

Accu-Chek Inform II system. Professional glucose testing for the wireless age

Accu-Chek Inform II system. Professional glucose testing for the wireless age Accu-Chek Inform II system Professional glucose testing for the wireless age The Accu-Chek Inform II system Real time control The Accu-Chek Inform II System overview Quality Advanced chemistry with multiple

More information

OPTI CCA-TS Analyzer. Operator s Manual

OPTI CCA-TS Analyzer. Operator s Manual OPTI CCA-TS Analyzer Operator s Manual OPERATOR S MANUAL REVISION LOG (Please record any changes made to this manual) Revision Release Date Approved by Description A Sept 2004 Released per CO 040316 B

More information

510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2

510(K) SUMMARY FEB - 12008. Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2 Attachment #2 510 (k) Summary 510(K) SUMMARY FEB - 12008 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

More information

cobas 6500 urine analyzer series Fully automated urine work area on a modular platform

cobas 6500 urine analyzer series Fully automated urine work area on a modular platform cobas 6500 urine analyzer series Fully automated urine work area on a modular platform cobas 6500 urine analyzer series Fully automated urine work area on a modular platform cobas 6500 urine analyzer series

More information

Biochemistry Validation Form

Biochemistry Validation Form Biochemistry Validation Form Method: Piccolo Xpress Chemistry 13 panel Manufacturer: Abaxis Cat no: 400-0029 CE marked: Yes: x No: Location of bench book: X:\Bio\Patricia\Piccollo Xpress evaluation (August

More information

Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT

Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Alexander Belenky, PhD Director -Product Development Samsung - Nexus-Dx Edward Brennan, PhD

More information

Procedure Manual for the i-stat System

Procedure Manual for the i-stat System Procedure Manual for the i-stat System This Procedure Manual is intended to be a template for the Procedure Manual required by CLIA and laboratory accreditation bodies. This Procedure Manual should be

More information

HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB

HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB Hematocrit is determined conductometrically. The measured conductivity, after correction for electrolyte concentration, is inversely related to the hematocrit.

More information

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz

More information

How To Test For Creatinine

How To Test For Creatinine Creatinine is measured amperometrically. Creatinine is hydrolyzed to creatine in a reaction catalyzed by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to sarcosine in a reaction catalyzed

More information

Centro de Salud México España Centro de Salud San Francisco Culhuacán Mexico City, Mexico

Centro de Salud México España Centro de Salud San Francisco Culhuacán Mexico City, Mexico Centro de Salud México España Centro de Salud San Francisco Culhuacán Mexico City, Mexico Mexico City is one of the largest cities in the world and home to over 21 million people. The population is culturally

More information

Georgia Northwestern Technical College Practical Nursing Program CLINICAL DAILY ASSESSMENT WORKSHEET FOR MODULES I-IV STUDENT: CLINICAL INSTRUCTOR:

Georgia Northwestern Technical College Practical Nursing Program CLINICAL DAILY ASSESSMENT WORKSHEET FOR MODULES I-IV STUDENT: CLINICAL INSTRUCTOR: Georgia Northwestern Technical College Practical Nursing Program CLINICAL DAILY ASSESSMENT WORKSHEET FOR MODULES I-IV STUDENT: CLINICAL INSTRUCTOR: CLINICAL UNIT: ASSIGNMENT DATES: PATIENT (last initial):

More information

Implementation Guide

Implementation Guide HAWAII HEALTH INFORMATION CORPORATION HAWAII S SOURCE FOR HEALTHCARE Implementation Guide ASCII HL7 Specifications for Laboratory Observation Reporting Technical Specifications and Transmittal Instructions

More information

cobas c 111 analyzer Small box. Big performance

cobas c 111 analyzer Small box. Big performance cobas c 111 analyzer Small box. Big performance His journey back to health begins in your laboratory cobas solutions making a difference when it matters most Every patient prognosis depends on several

More information

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6 CentraCare Health (CCH) adopts the following policy/procedure for: St. Cloud Hospital Original: 5/02 Minor Revision: Full Review: 1/15 Replaces:

More information

Quality Requirements of a Point of Care Testing Service

Quality Requirements of a Point of Care Testing Service Quality Requirements of a Point of Care Testing Service Adil I. Khan MSc, PhD Assistant Professor of Pathology Director of Point of Care Testing & Clinical Chemistry Laboratory, Temple University Episcopal

More information

Blood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff

Blood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff Blood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff GUIDANCE guidance. Submit written comments to the Division of

More information

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

SKUP Scandinavian evaluation of laboratory equipment for primary health care

SKUP Scandinavian evaluation of laboratory equipment for primary health care SKUP Scandinavian evaluation of laboratory equipment for primary health care ACCU-CHEK Aviva Meter and test strips designed for glucose self-measurement Manufactured by Roche Diagnostics GmbH Report from

More information

Cedex Bio HT Bioprocess Analyzer. Be Confident. Be Assured.

Cedex Bio HT Bioprocess Analyzer. Be Confident. Be Assured. Cedex Bio HT Bioprocess Analyzer Be Confident. Be Assured. Next Generation Technology for Bioprocess Analytics Analyze Your Samples with Confidence. Be Assured of the Data Integrity. Metabolite Screening

More information

Reference Range: 0.5-1.6 mmol/l (arterial) 0.7-2.1 mmol/l (venous) CPT Code: 83605

Reference Range: 0.5-1.6 mmol/l (arterial) 0.7-2.1 mmol/l (venous) CPT Code: 83605 LACTIC ACID Fasting, arterial specimen preferred. Please note whether arterial or venous. 0.5 ml heparinized plasma. Green top or PST must be drawn, placed on ice, and spun within 15 minutes. Immediately

More information

And Now, Presenting...

And Now, Presenting... Presentation and Handling of Clinical Laboratory Data From Test Tube to Table Randall K. Carlson, Wilmington, DE and Nate Freimark, Lakewood, NJ Omnicare Clinical Research. Inc. INTRODUCTION In human clinical

More information

CTSC Clinical Laboratory. Clinical Laboratory Services CTSC. Clinical & Translational Science Center

CTSC Clinical Laboratory. Clinical Laboratory Services CTSC. Clinical & Translational Science Center Clinical Laboratory The Clinical Laboratory develops and carries out research-related sample analyses to UNM HSC investigators, researchers throughout the United States and world, as well as corporate

More information

Oxygen Dissociation Curve

Oxygen Dissociation Curve 122 Visit http://www.anaesthesiamcq.com for details Chapter 4 Dissociation Curve Can you draw the oxygen dissociation curve of normal adult haemoglobin? How many points on the curve can you indicate with

More information

Clinical Laboratory Parameters for Crl:CD(SD) Rats. March, 2006. Information Prepared by Mary L.A. Giknis, Ph.D. Charles B. Clifford, D.V.M., Ph.D.

Clinical Laboratory Parameters for Crl:CD(SD) Rats. March, 2006. Information Prepared by Mary L.A. Giknis, Ph.D. Charles B. Clifford, D.V.M., Ph.D. Clinical Laboratory Parameters for Crl:CD(SD) Rats March, 2006 Information Prepared by Mary L.A. Giknis, Ph.D. Charles B. Clifford, D.V.M., Ph.D. CHARLES RIVER LABORATORIES Clinical Laboratory Parameters

More information

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods

Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness

More information

UNIVERSITY OF KENTUCKY HOSPITAL LABORATORIES 3/26/14 Lexington, KY Page 1 of 6

UNIVERSITY OF KENTUCKY HOSPITAL LABORATORIES 3/26/14 Lexington, KY Page 1 of 6 Lexington, KY Page 1 of 6 Affected Sites: X Enterprise Chandler Good Samaritan CRITICAL VALUES: Critical Values are defined as an analytic result that suggests a clinical condition that may be lifethreatening

More information

I know my value. CoaguChek XS Plus system Smart INR monitoring at your practice. 1 of 6

I know my value. CoaguChek XS Plus system Smart INR monitoring at your practice. 1 of 6 I know my value CoaguChek XS Plus system Smart INR monitoring at your practice 1 of 6 CoaguChek XS Plus system A better choice for your patients and your practice There has been an extraordinary increase

More information

Equations for Primary FRCA

Equations for Primary FRCA Equations for Primary FA Pharmacology : concentration t: time Bioavailability Bioavailab ility AU AU OAL IV AU: area under concentration time curve Exponential Function d d or K. dt dt e.78 or 0. 37 e

More information

Validation of measurement procedures

Validation of measurement procedures Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation

More information

Suffolk County Community College School of Nursing NUR 133 ADULT NURSING I

Suffolk County Community College School of Nursing NUR 133 ADULT NURSING I Suffolk County Community College School of Nursing NUR 133 ADULT NURSING I Page # 1 Instructions for students: Case study # 1 For this lab, you are planning to provide care to the following client: CB

More information

Unleash Your Potential

Unleash Your Potential Unleash Your Potential Efficiency. Power. Flexibility. Small In Size Big In Performance. The respons 910 vet A Leap Forward For Your Laboratory. System solution consisting of the respons 910 vet analyzer,

More information

Arterial Blood Gas Case Questions and Answers

Arterial Blood Gas Case Questions and Answers Arterial Blood Gas Case Questions and Answers In the space that follows you will find a series of cases that include arterial blood gases. Each case is then followed by an explanation of the acid-base

More information

User Guide. Quality Check Messages and Codes (continued) (01)00054749001286. Message Cause Action

User Guide. Quality Check Messages and Codes (continued) (01)00054749001286. Message Cause Action Quality Check Messages and Codes (continued) Message Cause Action i-stat Analyzer Error, Use Electronic Simulator Analyzer Error, See Manual No display Cartridge Locked does not disappear after test cycle

More information

Blood Glucose Monitoring: The Facts about Accuracy

Blood Glucose Monitoring: The Facts about Accuracy Blood Glucose Monitoring: The Facts about Accuracy Summary The purpose of this document is to provide the facts about the accuracy of self-monitoring of blood glucose (SMBG) systems. Accuracy can be defined

More information

Name Class Date. Section: Calculating Quantities in Reactions. Complete each statement below by writing the correct term or phrase.

Name Class Date. Section: Calculating Quantities in Reactions. Complete each statement below by writing the correct term or phrase. Skills Worksheet Concept Review Section: Calculating Quantities in Reactions Complete each statement below by writing the correct term or phrase. 1. All stoichiometric calculations involving equations

More information

Contents: POCT Program Massachusetts General Hospital - Pathology Service 55 Fruit Street, Boston, MA 02114

Contents: POCT Program Massachusetts General Hospital - Pathology Service 55 Fruit Street, Boston, MA 02114 Title: i-stat Operations and Testing Procedure Cross References: i-stat Operator Training Checklist i-stat INR Fingerstick Training and Assessment Record i-stat Competency Assessment 6-month and Annual

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO

Application Sheet for Rivaroxaban (Xarelto ) Standard Range with. BIOPHEN Heparin LRT (#221011/221013) RUO Instrument Adaptation The attached instrument adaptation has been prepared and validated by the reagent manufacturer, Hyphen-Biomed. Instrument Siemens BCS-XP Product BIOPHEN Heparin LRT Analyte Rivaroxaban

More information

cobas 8100 automated workflow series 3-D intelligence in lab automation

cobas 8100 automated workflow series 3-D intelligence in lab automation cobas 8100 automated workflow series 3-D intelligence in lab automation Personalized Lab Automation Maximizing Testing Efficiency and Medical Value At Roche, laboratory automated solutions deliver the

More information

ACID- BASE and ELECTROLYTE BALANCE. MGHS School of EMT-Paramedic Program 2011

ACID- BASE and ELECTROLYTE BALANCE. MGHS School of EMT-Paramedic Program 2011 ACID- BASE and ELECTROLYTE BALANCE MGHS School of EMT-Paramedic Program 2011 ACID- BASE BALANCE Ions balance themselves like a see-saw. Solutions turn into acids when concentration of hydrogen ions rises

More information

Guide to Common Laboratory Tests for Eating Disorder Patients

Guide to Common Laboratory Tests for Eating Disorder Patients GuidetoCommonLaboratoryTestsforEatingDisorderPatients Test Measure Description Reference Range** CompleteBloodCount Seebelow Levelsofmultiple See (CBC) blood individual components counts WhiteBlood Cell(WBC)

More information

CHM1 Review for Exam 12

CHM1 Review for Exam 12 Topics Solutions 1. Arrhenius Acids and bases a. An acid increases the H + concentration in b. A base increases the OH - concentration in 2. Strong acids and bases completely dissociate 3. Weak acids and

More information

Unit 2: Quantities in Chemistry

Unit 2: Quantities in Chemistry Mass, Moles, & Molar Mass Relative quantities of isotopes in a natural occurring element (%) E.g. Carbon has 2 isotopes C-12 and C-13. Of Carbon s two isotopes, there is 98.9% C-12 and 11.1% C-13. Find

More information

Course outline. Code: MLS211 Title: Medical Biochemistry

Course outline. Code: MLS211 Title: Medical Biochemistry Course outline Code: MLS211 Title: Medical Biochemistry Faculty of: Science, Health, Education and Engineering Teaching Session: Semester 2 Year: 2015 Course Coordinator: Dr Mark Holmes Tel: 5430 2844

More information

CORRELATION ANALYSIS

CORRELATION ANALYSIS CORRELATION ANALYSIS Learning Objectives Understand how correlation can be used to demonstrate a relationship between two factors. Know how to perform a correlation analysis and calculate the coefficient

More information

Blood gas analyzers: retooled designs, more choices

Blood gas analyzers: retooled designs, more choices August 2012 CAP TODAY / 75 Blood gas analyzers: retooled designs, more choices Brendan Dabkowski N ew designs work wonders, says Mary Catherine Coyle, of Roche Diagnostics. And judging by the number of

More information

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System *

Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System * Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System * Summary Objectives: To demonstrate that the TRUE METRIX Self-Monitoring Blood Glucose System*, from

More information

Simple linear regression

Simple linear regression Simple linear regression Introduction Simple linear regression is a statistical method for obtaining a formula to predict values of one variable from another where there is a causal relationship between

More information

Quality Consultant for Oregon Medical Laboratories

Quality Consultant for Oregon Medical Laboratories F E A T U R E Karen Dalton-Beninato Critical Value tifications Are Never Welcome News Quality Consultant for Oregon Medical Laboratories Vivian Benfield, MT(ASCP)SBB, describes a memorable case: Our client

More information

Determinants of Blood Oxygen Content Instructor s Guide

Determinants of Blood Oxygen Content Instructor s Guide Determinants of Blood Oxygen Content Instructor s Guide Time to Complete This activity will take approximately 75 minutes, but can be shortened depending on how much time the instructor takes to review

More information

TRANSPORT OF BLOOD GASES From The Lungs To The Tissues & Back

TRANSPORT OF BLOOD GASES From The Lungs To The Tissues & Back TRANSPORT OF BLOOD GASES From The Lungs To The Tissues & Back Dr. Sally Osborne Department of Cellular & Physiological Sciences University of British Columbia Room 3602, D.H Copp Building 604 822-3421

More information

Diabetic Ketoacidosis: When Sugar Isn t Sweet!!!

Diabetic Ketoacidosis: When Sugar Isn t Sweet!!! Diabetic Ketoacidosis: When Sugar Isn t Sweet!!! W Ricks Hanna Jr MD Assistant Professor of Pediatrics University of Tennessee Health Science Center LeBonheur Children s Hospital Introduction Diabetes

More information

Lewis Dot Structures of Atoms and Ions

Lewis Dot Structures of Atoms and Ions Why? The chemical properties of an element are based on the number of electrons in the outer shell of its atoms. We use Lewis dot structures to map these valence electrons in order to identify stable electron

More information

An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and

An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and CoaguChek systems Cardiovascular disease: the #1 killer in

More information

tips Insulin Pump Users 1 Early detection of insulin deprivation in continuous subcutaneous 2 Population Study of Pediatric Ketoacidosis in Sweden:

tips Insulin Pump Users 1 Early detection of insulin deprivation in continuous subcutaneous 2 Population Study of Pediatric Ketoacidosis in Sweden: tips Top International Publications Selection Insulin Pump Users Early detection of insulin deprivation in continuous subcutaneous insulin infusion-treated Patients with TD Population Study of Pediatric

More information

CORD BLOOD COLLECTION / ANALYSIS- AT BIRTH

CORD BLOOD COLLECTION / ANALYSIS- AT BIRTH WOMEN AND NEWBORN HEALTH SERVICE King Edward Memorial Hospital CLINICAL GUIDELINES OBSTETRICS AND MIDWIFERY King Edward Memorial Hospital WOMEN AND NEWBORN HEALTH SERVICE INTRAPARTUM CARE SPECIMEN COLLECTION

More information

Basic Statistics and Data Analysis for Health Researchers from Foreign Countries

Basic Statistics and Data Analysis for Health Researchers from Foreign Countries Basic Statistics and Data Analysis for Health Researchers from Foreign Countries Volkert Siersma siersma@sund.ku.dk The Research Unit for General Practice in Copenhagen Dias 1 Content Quantifying association

More information

Molar Mass Worksheet Answer Key

Molar Mass Worksheet Answer Key Molar Mass Worksheet Answer Key Calculate the molar masses of the following chemicals: 1) Cl 2 71 g/mol 2) KOH 56.1 g/mol 3) BeCl 2 80 g/mol 4) FeCl 3 162.3 g/mol 5) BF 3 67.8 g/mol 6) CCl 2 F 2 121 g/mol

More information

Name: Teacher: Pd. Date:

Name: Teacher: Pd. Date: Name: Teacher: Pd. Date: STAAR Tutorial : Energy and Matter: Elements, Compounds, and Chemical Equations: 6.5C Differentiate between elements and compounds on the most basic level. 8.5F Recognize whether

More information

for patient monitoring

for patient monitoring Thermo Scientific Indiko Reagents for Immunosuppressant Drug Testing Complete solutions for patient monitoring The cost-effective Thermo Scientific Indiko benchtop analyzer and bar-coded reagent is designed

More information

IB Chemistry 1 Mole. One atom of C-12 has a mass of 12 amu. One mole of C-12 has a mass of 12 g. Grams we can use more easily.

IB Chemistry 1 Mole. One atom of C-12 has a mass of 12 amu. One mole of C-12 has a mass of 12 g. Grams we can use more easily. The Mole Atomic mass units and atoms are not convenient units to work with. The concept of the mole was invented. This was the number of atoms of carbon-12 that were needed to make 12 g of carbon. 1 mole

More information

Sepsis Reassess patient Monitor and maintain respiratory/ hemodynamic status

Sepsis Reassess patient Monitor and maintain respiratory/ hemodynamic status Patient exhibits two or more of the following SIRS criteria: Temperature greater than 38 o C (100.4 o F) or less SIRS than criteria 36 o C (96.8 o F) Heart Rate greater than 90 beats/minute Respiratory

More information

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory

More information

FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)

FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs) FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs) Barriers to Overcoming Interferences and Limitations Alan T. Cariski, MD, JD, and Mike Flis, AdvaMed BGM

More information

Oxygenation and Oxygen Therapy Michael Billow, D.O.

Oxygenation and Oxygen Therapy Michael Billow, D.O. Oxygenation and Oxygen Therapy Michael Billow, D.O. The delivery of oxygen to all body tissues is the essence of critical care. Patients in respiratory distress/failure come easily to mind as the ones

More information

PROCEDURE NO. POC.514.01 LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version.

PROCEDURE NO. POC.514.01 LBH. Printed copies are for reference only. Please refer to the electronic copy for the latest version. Department Of Pathology POC.514.01- Blood Glucose Monitoring Accu-Chek Inform II Procedure-LBH Version# 1 Department PROCEDURE NO. PAGE NO. Point-of-Care Testing POC.514.01 LBH 1 OF 7 Printed copies are

More information

Intelligent laboratory management Delivered by cobas IT middleware

Intelligent laboratory management Delivered by cobas IT middleware For technical specifications and additional information please see the enclosed datasheets. Intelligent laboratory management Delivered by cobas IT middleware To arrange a meeting or demonstration to learn

More information

Pointe Scientific, Inc. Instrument Application

Pointe Scientific, Inc. Instrument Application Test: ACID PHOS. (NP) Catalog # : A7503 TEST NAME: ACID PHOS. (NP) TEST CODE: [AP (NP)] UNITS: [U/L] PRECISION (DECIMAL): [0] Kinetic MATH MODEL: Fixed Factor FACTOR: 1318.00 [

More information

Accutrend Plus. User s Manual

Accutrend Plus. User s Manual Accutrend Plus User s Manual Date of issue: March 2007 2007, Roche Diagnostics All rights reserved. ACCU-CHEK, ACCUTREND, COBAS, SAFE-T-PRO and SOFTCLIX are trademarks of Roche. This instrument complies

More information

Module 3: Correlation and Covariance

Module 3: Correlation and Covariance Using Statistical Data to Make Decisions Module 3: Correlation and Covariance Tom Ilvento Dr. Mugdim Pašiƒ University of Delaware Sarajevo Graduate School of Business O ften our interest in data analysis

More information

Mole Notes.notebook. October 29, 2014

Mole Notes.notebook. October 29, 2014 1 2 How do chemists count atoms/formula units/molecules? How do we go from the atomic scale to the scale of everyday measurements (macroscopic scale)? The gateway is the mole! But before we get to the

More information

PROFICIENCY TESTING. Clinical Laboratory Improvement Amendments (CLIA) DOs and DON Ts. Brochure # 8

PROFICIENCY TESTING. Clinical Laboratory Improvement Amendments (CLIA) DOs and DON Ts. Brochure # 8 Clinical Laboratory Improvement Amendments (CLIA) Brochure # 8 PROFICIENCY TESTING DOs and DON Ts NOTE: Congress passed the Clinical Laboratory Improvement Amendments (CLIA) in 1988 establishing quality

More information

Chapter 2: The Chemical Context of Life

Chapter 2: The Chemical Context of Life Chapter 2: The Chemical Context of Life Name Period This chapter covers the basics that you may have learned in your chemistry class. Whether your teacher goes over this chapter, or assigns it for you

More information

Replacement post-filter (ml/hr) Blood flow (ml/min) Dialysate (ml/hr) Weight (kg)

Replacement post-filter (ml/hr) Blood flow (ml/min) Dialysate (ml/hr) Weight (kg) 404FM.1 CONTINUOUS RENAL REPLACEMENT THERAPY (CRRT) USING CITRATE Target Audience: Hospital only ICU (Based on Gambro and Kalmar Hospital protocols) CRRT using regional citrate anticoagulation This is

More information

EN ISO 22870 accreditation for blood gases

EN ISO 22870 accreditation for blood gases EN ISO 22870 accreditation for blood gases Michel Vaubourdolle SFBC Accreditation WG chairman EFLM ISO/A WG chairman Head of Department of Laboratory Medicine* and Pathology University Hospitals Paris-East,

More information

Accuracy Test Omnitest 3 Blood Glucose Monitoring System. B. Braun Diabetes Care

Accuracy Test Omnitest 3 Blood Glucose Monitoring System. B. Braun Diabetes Care Accuracy Test Omnitest Blood Glucose Monitoring System B. Braun Diabetes Care Accuracy Test Omnitest blood glucose monitoring system according to ISO 97 The accuracy study of Omnitest blood glucose monitoring

More information

Acid-Base Balance and the Anion Gap

Acid-Base Balance and the Anion Gap Acid-Base Balance and the Anion Gap 1. The body strives for electrical neutrality. a. Cations = Anions b. One of the cations is very special, H +, and its concentration is monitored and regulated very

More information

Title: Fingerstick Glucose by Accu-Chek Inform

Title: Fingerstick Glucose by Accu-Chek Inform Title: Fingerstick Glucose by Accu-Chek Inform Target Audience: This module is available to aid in assessing competency for all clinical staff who perform fingerstick glucose testing. Contents Instructions...

More information

Success criteria You should be able to write the correct formula for any ionic compound

Success criteria You should be able to write the correct formula for any ionic compound Chemical Formulas and Names of Ionic Compounds WHY? Going back to pre-historic times, humans have experimented with chemical processes that helped them to make better tools, pottery and weapons. In the

More information

A comparison of blood glucose meters in Australia

A comparison of blood glucose meters in Australia Diabetes Research and Clinical Practice 71 (2006) 113 118 www.elsevier.com/locate/diabres A comparison of blood glucose meters in Australia Matthew Cohen *, Erin Boyle, Carol Delaney, Jonathan Shaw International

More information

Clinical Reasoning Case Study: I. Data Collection Chief complaint/history of Present Illness:

Clinical Reasoning Case Study: I. Data Collection Chief complaint/history of Present Illness: Clinical Reasoning Case Study: I. Data Collection Chief complaint/history of Present Illness: What data is relevant that must be recognized as clinically significant to the nurse? Rationale: Personal/Social

More information

Hemoglobin detection kit Catalog # ADI-907-034

Hemoglobin detection kit Catalog # ADI-907-034 New Dual Range! Hemoglobin detection kit Catalog # ADI-907-034 2x96 wells For use with blood samples Store at 4 C. Check our website for additional protocols, technical notes and FAQs. For proper performance,

More information

Written by: Paul Pappagianopoulos Date: 6/01/2005

Written by: Paul Pappagianopoulos Date: 6/01/2005 Title: GEM 3500 Procedure Cross References: GEM 3500 Competency Documentation-6 month and Annual GEM 3500 Training Quiz GEM 3500 Corrective Action Log MGH POCT Inter-Laboratory Correlation Procedure MGH

More information

Description of the Mole Concept:

Description of the Mole Concept: Description of the Mole Concept: Suppose you were sent into the store to buy 36 eggs. When you picked them up you would get 3 boxes, each containing 12 eggs. You just used a mathematical device, called

More information

PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatine Kinase REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL

More information

Measurement Systems Correlation MSC for Suppliers

Measurement Systems Correlation MSC for Suppliers Measurement Systems Correlation MSC for Suppliers Copyright 2003-2007 Raytheon Company. All rights reserved. R6σ is a Raytheon trademark registered in the United States and Europe. Raytheon Six Sigma is

More information

A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved

A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved Overview and Objectives This module provides an introduction to

More information

Chemistry Diagnostic Questions

Chemistry Diagnostic Questions Chemistry Diagnostic Questions Answer these 40 multiple choice questions and then check your answers, located at the end of this document. If you correctly answered less than 25 questions, you need to

More information

Intraosseous Vascular Access and Lidocaine

Intraosseous Vascular Access and Lidocaine Intraosseous Vascular Access and Lidocaine Intraosseous (IO) needles provide access to the medullary cavity of a bone. It is a technique primarily used in emergency situations to administer fluid and medication

More information

SPSS Guide: Regression Analysis

SPSS Guide: Regression Analysis SPSS Guide: Regression Analysis I put this together to give you a step-by-step guide for replicating what we did in the computer lab. It should help you run the tests we covered. The best way to get familiar

More information

PROTONS AND ELECTRONS

PROTONS AND ELECTRONS reflect Imagine that you have a bowl of oranges, bananas, pineapples, berries, pears, and watermelon. How do you identify each piece of fruit? Most likely, you are familiar with the characteristics of

More information

Urinalysis from Roche What else

Urinalysis from Roche What else Urinalysis from Roche What else Ascorbic acid - Vitamin C Healthy, powerful and interfering Be attentive Consider that laboratories testing urine samples regularly found high amounts of ascorbic acid (

More information

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) WHITE PAPER SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR) LifeScan developed the SureStep Pro Linearity Solutions to assist institutional customers in meeting regulatory

More information

Chemical Calculations: Formula Masses, Moles, and Chemical Equations

Chemical Calculations: Formula Masses, Moles, and Chemical Equations Chemical Calculations: Formula Masses, Moles, and Chemical Equations Atomic Mass & Formula Mass Recall from Chapter Three that the average mass of an atom of a given element can be found on the periodic

More information

3% Sodium Chloride Injection, USP 5% Sodium Chloride Injection, USP

3% Sodium Chloride Injection, USP 5% Sodium Chloride Injection, USP PRESCRIBING INFORMATION 3% Sodium Chloride Injection, USP 5% Sodium Chloride Injection, USP IV Fluid and Electrolyte Replenisher Baxter Corporation Mississauga, Ontario L5N 0C2 Canada Date of Revision:

More information

Chapter 7: Simple linear regression Learning Objectives

Chapter 7: Simple linear regression Learning Objectives Chapter 7: Simple linear regression Learning Objectives Reading: Section 7.1 of OpenIntro Statistics Video: Correlation vs. causation, YouTube (2:19) Video: Intro to Linear Regression, YouTube (5:18) -

More information