1 PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biological Imaging Center& Ahmanson Translational Imaging Division David Geffen School of Medicine at UCLA Los Angeles 1 Learning Objectives Summarize the signal processing steps for coincidence detection Understand the components of a daily QA procedure Identify and trouble shoot possible sources of failure in daily QA List the recommended frequency of QA/QC tests Describe the process of scanner calibration Name the different components of the NEMA test used for PET acceptance testing Understand the meaning of the results of the NEMA tests. 2 Outline PET Basics PET Detectors and data flow Daily QC Procedures QC Evaluation Techniques Examples of Common Problems Quarterly and Annual QC Procedures 3 1
2 Quality Control in PET Verify the operational integrity of the system Ensure Stability of the Detectors Ensure Stability of the Acquisition Electronics Maintain consistent and high image quality Minimize chances for image artifacts Catch potential problems early Maintain quantitative accuracy Eliminate unnecessary repeat scans 4 Coincidence Detection Coincidence Detection Amp. Amp. PHA 511 kev? 511 kev? PHA Coinc. Counter Time Resolution: 2-3 ns 2
3 Coincidence Detection Data Flow in a PET System 8 Time-of-flight PET R R Det 2 Det 1 x R - x R + x s = v t R + x = vt 1 R x = vt 2 2 x = v( t2 t1) x = c t Time Resolution: ~300 ps 2 3
4 PET Detectors PMT B D << : < ~100% PMT B PMT D : ~0% PET Detectors To identify the detector elements, X- and Y- positions are calculated for each event: Element (4,4) Element (5,3) Pixelated Detector System Continuous-pixelated detector 4
5 PET Detector Calibration Balancing of PMTs Identification of Peaks/Detectors in Flood Histogram Acquire Energy Spectra / Set Energy Thresholds Time Alignment Challenge: Large number of detector blocks (~200) Large number of detector elements (>30,000) 13 Daily QC Detect small drifts in the system Detector Gain Drifts (Efficiency, Scatter Fraction, Randoms) Timing Quantification Challenges: Large number of detector blocks (~200) Large number of detector elements (>30,000) Time constraint (< 1 hr) 15 5
6 Daily QC Scan Transmission/Rotating Rod Sources Uniform cylinder phantom (20 cm Ø) Also used for calibration (Counts Bq/ml) Water Phantom Calibrated w. Dose Calibrator Solid 68 Ge Phantom Pre-calibrated 16 Sinogram θ All coincidence lines that are parallel at a given angle form a projection in the sinogram. ρ Sinogram θ All coincidence lines (or lines of response) for a given detector form a diagonal trace in the sinogram. ρ 6
7 Daily QC / Blank Scan Visually Inspect the sinograms for: Apparent streaks (hot or cold) Consistency Comparison to a reference scan PET Daily QA Scan O. Mawlawi MDACC 7
8 Normalization θ = z ρ Normalization After Detector Tuning Before PM Normalization After Detector Tuning w. norm Before PM w. norm 8
9 When things go wrong Normal Detector Failure Detector Controller Failure Sorting Memory Failure Detector Block Failure Detector Block Failure 9
10 Buchert et. al. JNM 40, 1999 PET Daily QA Scan O. Mawlawi MDACC 30 10
11 31 System Calibration Calculates Conversion Factor : Image A.U. Bq/ml Scan uniform 20 cm Ø 68 Ge or 18 F Cylinder Calibrated Activity, Time & Volume Emission & Transmission / CT Reconstruct: All corrections applied Standard reconstruction parameters Calibration Factor : Calibrated Act. Conc. / ROI of Image (A.U.) 32 Daily QC Scan / Siemens mct Acquire 68 Ge Cylinder Scan ( Mcts) Generation of Normalization Factors Evaluation of Daily QC Components Centered! Block Noise Block Efficiency Measured Randoms Scanner Efficiency Scatter Ratio Calibration Factor) & Image Plane Efficiency Timing Offset, Width and Time Alignment Fit 33 11
12 Daily QC Scan / Siemens mct 34 Scatter Fraction / Energy Threshold
13 CT Daily QA Scan Normal operations include the following 3 tasks (in order): Tube Warmup- A built-in prep scan that gradually increases heat loading in the X-ray tube in order to prevent thermal cracking and eliminate the potential for an arc to occur. It includes a series of exposures made at incrementing kvp Daily Air Cals- A built-in prep scan that performs a series of exposures at varying techniques in order to normalize the detector response using air as the attenuating media. These scans essentially adjust the detector gains to achieve a uniform response Daily QA Phantom scan - Provides data for 3 areas of concern in daily quality assurance O. Mawlawi MDACC Quality Control CT daily regimen Scan water layer Measure water HU CTAC Check for artifacts Ring artifacts Redo conditioning Redo Air Cal Austin Health Courtesy: Stefan Eberl, RPA Quarterly QC Procedures Quantification Check Gantry alignment (for PET/CT) Other cross calibrations (well counter, etc) 39 13
14 Quantification Scan Scan uniform 20 cm Ø 68 Ge or 18 F Cylinder Emission & Transmission / CT Reconstruct: All corrections applied Standard reconstruction parameters Visual inspection Compare image ROI activity to calibrated activity Always perform after any service or detector adjustments 40 Quantification Scan 18 F Cylinder µci/ml FBP OSEM Calculated Atten. Corr % % Meas. Atten. Corr. CT % % Meas. Atten. Corr. Rods % % 14
15 PET/CT Gantry Alignment CT PET 15
16 ERROR: undefined OFFENDING COMMAND: f ~ STACK:
PET/CT QC/QA. Quality Control in PET. Magnus Dahlbom, Ph.D. Verify the operational integrity of the system. PET Detectors
Quality Control in PET PET/CT QC/QA Magnus Dahlbom, Ph.D. Division of Nuclear Medicine Ahmanson Biochemical Imaging Clinic David Geffen School of Medicine at UCLA Los Angeles Verify the operational integrity
QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET
Centers for Quantitative Imaging Excellence (CQIE) LEARNING MODULE QUANTITATIVE IMAGING IN MULTICENTER CLINICAL TRIALS: PET American College of Radiology Clinical Research Center v.1 Centers for Quantitative
Rb 82 Cardiac PET Scanning Protocols and Dosimetry. Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals
Rb 82 Cardiac PET Scanning Protocols and Dosimetry Deborah Tout Nuclear Medicine Department Central Manchester University Hospitals Overview Rb 82 myocardial perfusion imaging protocols Acquisition Reconstruction
INTRODUCTION. A. Purpose
New York State Department of Health Bureau of Environmental Radiation Protection Guide for Radiation Safety/Quality Assurance Programs Computed Radiography INTRODUCTION A. Purpose This guide describes
Performance Comparison of Four Commercial GE Discovery PET/CT Scanners: A Monte Carlo Study Using GATE
Performance Comparison of Four Commercial GE Discovery PET/CT Scanners: A Monte Carlo Study Using Parham Geramifar 1,3,4, Mohammad Reza Ay 2,3,4, Mojtaba Shamsaie Zafarghandi 1, George Loudos 5, Arman
Improvements in quality and quantification of 3D PET images
Improvements in quality and quantification of 3D PET images The resolution of PET depends on some effects impossible to be readily corrected The implementation of 3D spatially-variant PSF (radially asymmetric)
Quantitative Imaging In Clinical Trials Using PET/CT: Update
Quantitative Imaging In Clinical Trials Using PET/CT: Update Paul Kinahan, Robert Doot Imaging Research Laboratory Department of Radiology University of Washington, Seattle, WA Supported by RSNA Quantitative
Prepublication Requirements
Issued Prepublication Requirements The Joint Commission has approved the following revisions for prepublication. While revised requirements are published in the semiannual updates to the print manuals
Physics testing of image detectors
Physics testing of image detectors Parameters to test Spatial resolution Contrast resolution Uniformity/geometric distortion Features and Weaknesses of Phantoms for CR/DR System Testing Dose response/signal
Performance testing for Precision 500D Classical R/F System
Performance testing for Precision 500D Classical R/F System John M. Boudry, Ph.D. Image Quality Systems Engineer GE Healthcare Technologies Outline System background Image Quality Signature Test (IQST)
Quality Control and Maintenance Programs
Quality Control and Maintenance Programs Cari Borrás, D.Sc., FACR, FAAPM Visiting Professor DOIN-DEN / UFPE Recife, Pernambuco, Brazil Co-Chair, IUPESM Health Technology Task Group 1 Medical Imaging Equipment
Truly digital PET imaging
Truly digital PET imaging Philips proprietary Digital Photon Counting technology Vereos PET/CT is the first commercially available scanner to offer truly digital PET, resulting in significantly improved
Truly digital PET imaging
Advanced Molecular Imaging Vereos PET/CT Truly digital PET imaging Philips proprietary Digital Photon Counting technology Vereos PET/CT is the first commercially available scanner to offer truly digital
CT scanning. By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU. (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E.
1 Overview CT scanning By Mikael Jensen & Jens E. Wilhjelm Risø National laboratory Ørsted DTU (Ver. 1.2 4/9/07) 2002-2007 by M. Jensen and J. E. Wilhjelm) As it can be imagined, planar X-ray imaging has
SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS
Note: This is a courtesy copy and is not the official version of this rule. The official, legally effective version of this rule is available through www.lexisnexic.com/bookstore (Phone: (800) 223-1940).
X- and Gamma Ray Imaging Systems based on CdTe-CMOS Detector Technology
X- and Gamma Ray Imaging Systems based on dte-mos Detector Technology Konstantinos Spartiotis, Ray Durrant, nssi Leppänen, Henrik Lohman, Olli Männistö, Tuomas Pantsar, Jouni Pyyhtiä, Tom Schulman, Oy
Performance Characteristics of PET Scanners
6 Performance Characteristics of PET Scanners A major goal of the PET studies is to obtain a good quality and detailed image of an object by the PET scanner, and so it depends on how well the scanner performs
How To Improve Your Ct Image Quality
Translating Protocols Between Scanner Manufacturer and Model Cynthia H. McCollough, PhD, FACR, FAAPM Professor of Radiologic Physics Director, CT Clinical Innovation Center Department of Radiology Mayo
CT RADIATION DOSE REPORT FROM DICOM. Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH
CT RADIATION DOSE REPORT FROM DICOM Frank Dong, PhD, DABR Diagnostic Physicist Imaging Institute Cleveland Clinic Foundation Cleveland, OH CT Patient comes out... Patient goes in... Big Black Box Radiology
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13
Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis Tushita Patel 4/2/13 Breast Cancer Statistics Second most common cancer after skin cancer Second leading cause of cancer
Spiral CT: Single and Multiple Detector Systems. AAPM Refresher Course Nashville, TN July 28,1999
Spiral CT: Single and Multiple Detector Systems AAPM Refresher Course Nashville, TN July 28,1999 Mike McNitt-Gray, PhD, DABR Assistant Professor UCLA Radiological Sciences [email protected] X-Ray
3D SCANNERTM. 3D Scanning Comes Full Circle. s u n. Your Most Valuable QA and Dosimetry Tools A / B / C. The 3D SCANNER Advantage
3D SCANNERTM 3D Scanning Comes Full Circle Relative 3D Dosimetry offering the easiest setup, most objectivity, and best consistency available The 3D SCANNER Advantage Advanced Design Ring and diameter
MDCT Technology. Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington
MDCT Technology Kalpana M. Kanal, Ph.D., DABR Assistant Professor Department of Radiology University of Washington Seattle, Washington ACMP Annual Meeting 2008 - Seattle, WA Educational Objectives Historical
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations
kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations Moyed Miften, PhD Dept of Radiation Oncology University of Colorado Denver Questions Disease Stage (local, regional,
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22
REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22 New Jersey Department of Environmental Protection Bureau of Radiological Health PO Box 415 Trenton NJ
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS
AN INVESTIGATION INTO THE USEFULNESS OF THE ISOCS MATHEMATICAL EFFICIENCY CALIBRATION FOR LARGE RECTANGULAR 3 x5 x16 NAI DETECTORS Frazier L. Bronson CHP Canberra Industries, Inc. 800 Research Parkway,
WHERE IN THE WORLD JILL LIPOTI?
WHERE IN THE WORLD IS JILL LIPOTI? HELLO FROM NEW JERSEY CRCPD - National Symposium on Fusion Imaging and Multimodalities February 18-20, 2004 Kansas City, Missouri New Jersey s Requirements As They Pertain
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008.
Amptek Application Note XRF-1: XRF Spectra and Spectra Analysis Software By R.Redus, Chief Scientist, Amptek Inc, 2008. X-Ray Fluorescence (XRF) is a very simple analytical technique: X-rays excite atoms
Laue lens for Nuclear Medicine
Laue lens for Nuclear Medicine PhD in Physics Gianfranco Paternò Ferrara, 6-11-013 Supervisor: prof. Vincenzo Guidi Sensors and Semiconductors Lab, Department of Physics and Earth Science, University of
Purchasing a cardiac CT scanner: What the radiologist needs to know
Purchasing a cardiac CT scanner: What the radiologist needs to know Maria Lewis ImPACT St George s Hospital, London [email protected] CT scanner development Slice wars 1998 Increased z-coverage
Quality Control of Full Field Digital Mammography Units
Quality Control of Full Field Digital Mammography Units Melissa C. Martin, M.S., FACMP, FACR, FAAPM [email protected] 310-612-8127 ACMP Annual Meeting Virginia Beach, VA May 2, 2009 History of
Cone Beam Reconstruction Jiang Hsieh, Ph.D.
Cone Beam Reconstruction Jiang Hsieh, Ph.D. Applied Science Laboratory, GE Healthcare Technologies 1 Image Generation Reconstruction of images from projections. textbook reconstruction advanced acquisition
Fundamentals of Cone-Beam CT Imaging
Fundamentals of Cone-Beam CT Imaging Marc Kachelrieß German Cancer Research Center (DKFZ) Heidelberg, Germany www.dkfz.de Learning Objectives To understand the principles of volumetric image formation
Breast MRI Quality Control
Donna M. Reeve, MS, DABR, DABMP Department of Imaging Physics Educational Objectives Discuss the importance of breast MRI quality control (QC). Provide an overview of the new ACR Breast MRI Accreditation
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions
CT Protocol Optimization over the Range of CT Scanner Types: Recommendations & Misconceptions Frank N. Ranallo, Ph.D. Associate Professor of Medical Physics & Radiology University of Wisconsin School of
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements
CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index Keith J. Strauss, MSc, FAAPM, FACR Clinical Imaging Physicist Cincinnati Children s Hospital University of Cincinnati College of
RapidArc QA Program in Prince of Wales Hospital. Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong
RapidArc QA Program in Prince of Wales Hospital Michael L. M. Cheung, Physicist Prince of Wales Hospital Hong Kong Hardware and Software 3 out of 5 Linacs capable of delivering RapidArc: 2 Varian Clinac
PET and PET/CT in Clinical Trials
PET and PET/CT in Clinical Trials Nathan C. Hall, M.D., Ph.D. The Ohio State University Medical Center CALGB Imaging Core Lab CALGB Imaging Committee Outline Introduction to PET Positron Emitter Physics
Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission
Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission MAVEN Science Community Workshop December 2, 2012 Particles and Fields Package Solar Energetic Particle Instrument (SEP) Davin Larson and the SEP
Next Generation Small Animal PET Technology
G4 PET/X-ray P R O D U C T N O T E Pre-clinical in vivo Imaging Key Benefits: High performance, ultra-sensitive PET technology designed for pre-clinical imaging 3D PET and X-ray whole body images Purpose
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES
Personal Identity Verification (PIV) IMAGE QUALITY SPECIFICATIONS FOR SINGLE FINGER CAPTURE DEVICES 1.0 SCOPE AND PURPOSE These specifications apply to fingerprint capture devices which scan and capture
Turnkey Solutions for Radiation Protection in PET Laboratories R A D I A T I O N P R O T E C T I O N
Turnkey Solutions for Radiation Protection in PET Laboratories R A D I A T I O N P R O T E C T I O N Berthold Technologies has more than 60 years experience in manufacturing radiation protection measuring
Ultrasound Phantoms: B-Mode, Doppler and Others. Heather M. Pierce, B.S. Norfolk, VA
Ultrasound Phantoms: B-Mode, Doppler and Others Heather M. Pierce, B.S. Norfolk, VA Overview Design of ultrasound phantoms Types of phantoms available Use of phantoms in a QA/QC program Advantages and
American College of Radiology CT Accreditation Program. Testing Instructions
American College of Radiology CT Accreditation Program Testing Instructions (Revised July 24, 2015) This guide provides all of the instructions necessary for clinical tests, phantom tests and general submission
State-of-the-Art Technology in Cardiac CT
1 2 Next Step Evolution or Revolution? State-of-the-Art Technology in Cardiac CT Stefan Ulzheimer, PhD Global Director of Collaborations CT Siemens Medical Solutions Major Innovations in CT Head The 80
R/F. Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening. 3. Utility of Chest Tomosynthesis. 1.
R/F Efforts to Reduce Exposure Dose in Chest Tomosynthesis Targeting Lung Cancer Screening Department of Radiology, National Cancer Center Hospital East Kaoru Shimizu Ms. Kaoru Shimizu 1. Introduction
Contents. X-ray and Computed Tomography. Characterization of X-rays. Production of X-rays
J. E. Wilhjelm Ørsted TU Technical University of enmark, Bldg. 348, K-2800 Kongens Lyngby, enmark. X-ray and Computed Tomography Contents History and characterization of X-rays Conventional (projection)
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy
An Overview of Digital Imaging Systems for Radiography and Fluoroscopy Michael Yester, Ph.D. University of Alabama at Birmingham Outline Introduction Imaging Considerations Receptor Properties General
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation
Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation LORAD Stereotactic Breast Biopsy System AAPM Spring Clinical Meeting Phoenix, AZ March 17, 2013 Melissa C. Martin, M.S., FACR,
Evaluation Tools for the Performance of a NESTOR Test Detector
Evaluation Tools for the Performance of a NESTOR Test Detector G. Bourlis, A. Leisos, S. E. Tzamarias and A. Tsirigotis Particle and Astroparticle Physics Group School of Science and Technology Hellenic
TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-
QUALITY Tissue Benefits Mimicking of s RMI recognized RMI as the ultrasound standard phantoms for quality are performance the ultrasound labs ofand hospitals, manufacturers. clinics Sophisticated and ultrasound
Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)
Q.A. Collectible Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7) Mammography Phantom Image Quality Evaluation (from the American College of Radiology 1999 Mammography Quality
Proton tracking for medical imaging and dosimetry
Proton tracking for medical imaging and dosimetry J.Taylor, P.Allport, G.Casse For the PRaVDA Consortium 1 Background and motivation - What is the PRaVDA experiment? - Why are we using Monte Carlo? GEANT4
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis
Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School Digital Breast Tomosynthesis
Experimental study of beam hardening artefacts in photon counting breast computed tomography
Experimental study of beam hardening artefacts in photon counting breast computed tomography M.G. Bisogni a, A. Del Guerra a,n. Lanconelli b, A. Lauria c, G. Mettivier c, M.C. Montesi c, D. Panetta a,
Chapter 8. Low energy ion scattering study of Fe 4 N on Cu(100)
Low energy ion scattering study of 4 on Cu(1) Chapter 8. Low energy ion scattering study of 4 on Cu(1) 8.1. Introduction For a better understanding of the reconstructed 4 surfaces one would like to know
Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques
Recognition I will take this opportunity to recognize and thank the following people s contributions to this presentation. Considerations for: Diagnostic Radiology, Nuclear Medicine and, Oncology M. S.
TOF-PET DETECTOR CONCEPT BASED ON ORGANIC SCINTILLATORS
TOF-PET DETECTOR CONCEPT BASED ON ORGANIC SCINTILLATORS P. Moskal, T. Bednarski, P. Białas, M. Ciszewska, E. Czerwiński, A. Heczko, M. Kajetanowicz, Ł. Kapłon, A. Kochanowski, G. Konopka-Cupiał, G. Korcyl,
Optimization of image quality through online motion correction
Advanced imaging technologies a revolution in diagnosis and treatment Optimization of image quality through online motion correction Oline Vinter Olesen December 10, 2010 Medical Solutions, Siemens Institute
Routine Ultrasound Equipment Tests for Quality Assurance. Contents
Routine Ultrasound Equipment Tests for Quality Assurance James A. Zagzebski, Ph.D. Nick Rupert, M.S. Ryan DeWall, M.S. Tomy Varghese, Ph.D. Dept. of Medical Physics University of Wisconsin, Madison, WI,
Cynthia H. McCollough b) and Michael R. Bruesewitz Department of Radiology, Mayo Clinic College of Medicine, Rochester, Minnesota 55905
The phantom portion of the American College of Radiology ACR Computed Tomography CT accreditation program: Practical tips, artifact examples, and pitfalls to avoid a Cynthia H. McCollough b) and Michael
CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems
CT for the production floor: Recent developments in Hard- and Software for industrial CT Systems GE s Industrial X-ray and CT Forum, Cologne July 7 th 2015 Oliver Brunke GE Sensing & Inspection Technologies
Overview. Creation of 3D printed phantoms for clinical radiation therapy 7/7/2015. Eric Ehler, PhD Assistant Professor University of Minnesota
Creation of 3D printed phantoms for clinical radiation therapy Eric Ehler, PhD Assistant Professor University of Minnesota ehler [email protected] Overview Background of 3D Printing Practical Information Current
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems
A software tool for Quality Assurance of Computed / Digital Radiography (CR/DR) systems Nikunj Desai a and Daniel J Valentino a,b a icr Company Inc, 2580 West 237th Street, Torrance, CA 90505, USA b Department
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy
M D Anderson Cancer Center Orlando TomoTherapy s Implementation of Image-guided Adaptive Radiation Therapy Katja Langen, PhD Research supported by TomoTherapy Inc. Today s Lecture Introduction to helical
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM
MODELING AND IMPLEMENTATION OF THE MECHANICAL SYSTEM AND CONTROL OF A CT WITH LOW ENERGY PROTON BEAM João Antônio Palma Setti, [email protected] Pontifícia Universidade Católica do Paraná / Rua Imaculada
The accurate calibration of all detectors is crucial for the subsequent data
Chapter 4 Calibration The accurate calibration of all detectors is crucial for the subsequent data analysis. The stability of the gain and offset for energy and time calibration of all detectors involved
Fundamentals of modern UV-visible spectroscopy. Presentation Materials
Fundamentals of modern UV-visible spectroscopy Presentation Materials The Electromagnetic Spectrum E = hν ν = c / λ 1 Electronic Transitions in Formaldehyde 2 Electronic Transitions and Spectra of Atoms
CMS Tracking Performance Results from early LHC Running
CMS Tracking Performance Results from early LHC Running CMS PAPER TRK-10-001 L. Spiegel, K. Stenson, M. Swartz 1 First Collisions Tracking Paper Provide description of tracker, tracking algorithm, and
Digital Breast Tomosynthesis QC Requirements
Digital Breast Tomosynthesis QC Requirements AAPM Spring Clinical Meeting March 8, 2015 Michael S Glaser, MS, DABR Alliance Medical Physics, LLC Learning Objectives 1. GE SenoClaire - Physicist & Technologist
Atomic and Nuclear Physics Laboratory (Physics 4780)
Gamma Ray Spectroscopy Week of September 27, 2010 Atomic and Nuclear Physics Laboratory (Physics 4780) The University of Toledo Instructor: Randy Ellingson Gamma Ray Production: Co 60 60 60 27Co28Ni *
2/28/2011. MIPPA overview and CMS requirements. CT accreditation. Today s agenda. About MIPPA. Computed Tomography
Today s agenda Computed Tomography Presented by: Dina Hernandez, BSRT, RT (R), CT, QM Krista Bush, RT, MBA Leonard Lucey, JD ACR Quality & Safety MIPPA overview and CMS requirements CT accreditation How
Data. microcat +SPECT
Data microcat +SPECT microcat at a Glance Designed to meet the throughput, resolution and image quality requirements of academic and pharmaceutical research, the Siemens microcat sets the standard for
How To Use An Ionsonic Microscope
Sample Analysis Design Element2 - Basic Software Concepts Scan Modes Magnetic Scan (BScan): the electric field is kept constant and the magnetic field is varied as a function of time the BScan is suitable
Protocol Management and Review Strategies The MD Anderson Experience
Protocol Management and Review Strategies The MD Anderson Experience Dianna Cody, Ph.D. Professor Dept. of Imaging Physics Univ. of Texas M.D. Anderson Cancer Center Objectives Setting up a policy for
The Importance of Following the PROTOCOL in Clinical Trials
The Importance of Following the PROTOCOL in Clinical Trials Presentation Objectives: Upon completion of this presentation, participants will be able to: Describe the following terms: Protocol, Protocol
(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier
(Amplifying) Photo Detectors: Avalanche Photodiodes Silicon Photomultiplier (no PiN and pinned Diodes) Peter Fischer P. Fischer, ziti, Uni Heidelberg, Seite 1 Overview Reminder: Classical Photomultiplier
Moving Forward What does this mean for the Medical Physicist and the Imaging Community?
Moving Forward What does this mean for the Medical Physicist and the Imaging Community? John M. Boone, Ph.D., FAAPM, FACR Professor and Vice Chairman of Radiology University of California Davis Medical
Positron Emission Tomography: Principles, Technology, and Recent Developments
Nuclear Physics A 752 (2005) 679c 687c Positron Emission Tomography: Principles, Technology, and Recent Developments Sibylle I. Ziegler Nuklearmedizinische Klinik, Klinikum rechts der Isar der Technischen
Computed Tomography Resolution Enhancement by Integrating High-Resolution 2D X-Ray Images into the CT reconstruction
Digital Industrial Radiology and Computed Tomography (DIR 2015) 22-25 June 2015, Belgium, Ghent - www.ndt.net/app.dir2015 More Info at Open Access Database www.ndt.net/?id=18046 Computed Tomography Resolution
Radiation Monitoring System DLMon
Radiation Monitoring System DLMon Fields of Application: Dose rate monitoring system for use in nuclear medicine Area monitoring ( e.g. manufacturing of radionuclides, laboratories, nuclear power plants,
Avizo Inspect New software for industrial inspection and materials R&D
Avizo Inspect New software for industrial inspection and materials R&D Reduce your design cycle, inspection times, and meet higher-level quality standards at a lower cost. Avizo Inspect software streamlines
Power Amplifier Gain Compression Measurements
Technical Brief Power Amplifier Gain Compression Measurements GPIB Private Bus Sweep Out Sweep In Pulse In AC Mod Out Blank/Marker Out Blanking In Overview The 1 db gain compression of an amplifier describes
Clinic. ED Trauma Trauma Stroke. OR Neuro/Spine. Critical Care. Neuro ENT. Diagnostic. Pediatric. Radiology. Plastics Thoracic. Neuro.
www.neurologica.com Critical Care Clinic Operating Room Radiology ED Trauma OR Neuro/Spine Pediatric Plastics Thoracic General Interventional Radiology ENT ED Trauma Trauma Stroke Critical Care Neuro Pediatric
Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz
Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar Network Analyzer combines a 90 db wide dynamic range with the accuracy and linearity
Tumor. An Brief Introduction to 4D CT Scanning. Outline. Three Types of Motion Artifacts. CT Artifacts in Free Breathing 3D Scan
An Brief Introduction to 4D CT Scanning Steve B. Jiang, Ph.D. Dept of Radiation Oncology Univ of California San Diego Outline Problems with free breathing 3D scanning What is 4D CT? How does it work? Acknowledgements
MEDICAL IMAGING 2nd Part Computed Tomography
MEDICAL IMAGING 2nd Part Computed Tomography Introduction 2 In the last 30 years X-ray Computed Tomography development produced a great change in the role of diagnostic imaging in medicine. In convetional
Laboratory Technologies, Inc. Multi-Wiper. Time Saving Multiple Detector Wipe Counter
Laboratory Technologies, Inc. Multi-Wiper Time Saving Multiple Detector Wipe Counter Features Wipe test sets In-vitro tests Patient library Isotope library AutoSpect efficiency correction Advanced 4096-channel
Ultrasonic Load Monitoring
Ultrasonic Load Monitoring Who We Are One Company, Total Support, Complete Solutions Over many years, Hydratight has provided world-class bolted joint solutions and continues to set international standards
Robot Perception Continued
Robot Perception Continued 1 Visual Perception Visual Odometry Reconstruction Recognition CS 685 11 Range Sensing strategies Active range sensors Ultrasound Laser range sensor Slides adopted from Siegwart
Description: Multiparameter System (4 or 8 channels)
Model MPA4, 8 ch acquisition system with 6 ns time tagging The MPA4 is a Multiparameter Data Acquisition System Description: Multiparameter System (4 or 8 channels) ADC Settings and Presets Dialog The
IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.
IAC Ch 41, p.1 641 41.7 (136C) X-ray machines used for stereotactically guided breast biopsy. 41.7(1) Definitions. In addition to the definitions provided in rules 641 38.2(136C), 641 40.2(136C), and 641
Introduction to Geiger Counters
Introduction to Geiger Counters A Geiger counter (Geiger-Muller tube) is a device used for the detection and measurement of all types of radiation: alpha, beta and gamma radiation. Basically it consists
Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
INTERNATIONAL SPECIALTY PRODUCTS 1361 Alps Road Wayne NJ 07470 Tel: 973-628-4000 Performance of the Vidar Red LED Dosimetry Pro Advantage : A scanner optimized for use with GAFCHROMIC EBT Dosimetry Film.
DICOM Correction Item
Correction Number DICOM Correction Item CP-626 Log Summary: Type of Modification Clarification Rationale for Correction Name of Standard PS 3.3 2004 + Sup 83 The description of pixel spacing related attributes