12. Pressure sensors

Size: px
Start display at page:

Download "12. Pressure sensors"

Transcription

1 12. Pressure s Questions Task a) What is a piezo-resistive effect? b) What is the difference between absolute, gauge and differential pressure? c) What is the relation between output voltage (e.g. U X =5V) and applied pressure p[psi] if you know: excitation voltage (e.g. 10V), signal conditioning amplification A[-]=100 and sensitivity (e.g. 10mV/V.psi)? d) What is the advantage of four-arm active bridge? (4 active elements connected in bridge) 1) Measure the offset and sensitivity of a pressure gauge Honeywell 24PCCFA6G. The is built-in in the tool for blood pressure measurement. 2) Measure the blood pressure and beating frequency of a selected volunteer. Use an oscilloscope connected to the blood pressure measurement tool; record the signal from the oscilloscope (USB flash-disk). 3) Set-up the demonstration kit of ADC (Texas Instruments ADS1281) in Multiscope mode (samples and shows the input signal, calculates its mean value). 4) Connect an absolute pressure to the ADC. Measure the output voltage (proportional to atmospheric pressure) for different altitudes (0, 20, 40, 60, 80 cm above the table). 5) Connect a differential pressure to the ADC. The measures air-flow through a transparent tube in given setup. Measure the transfer characteristics: pressure difference versus air-flow (the relation between air-flow and ventilator rotation frequency is considered to be linear). Optional task (for curious students) 6) Check-out the three modes of operation of the ADC demonstration kit (Multi_Histogram, Multi_FFT, Multi_Scope), which are supported by the ADCPro software. Which values are used to characterize an AD converter? Measurement instructions 1) Measure the offset voltage for zero applied pressure (the valve at the pump balloon is fully opened). Close the valve; bend the tube leading to the arm-cuff and press it with fingers (the cuff should not been pressurized without the arm). Pressurize the system slowly to 200mmHg (see the mechanical gauge reading) and note the output voltage. Use the offset and sensitivity readings to calculate linear approximation of the voltage pressure transfer characteristics. 2) Connect the output of the tool to the oscilloscope input. Use two channels of the scope connected in parallel. First channel set to DC coupling; it will indicate the DC component of the pressure (sensitivity approximately 1V/div). The second channel set to an AC input coupling will indicate the so called Korotkoff sounds - pulses, see the picture below (sensitivity approximately 10mV/div). Set the scope to Roll-mode (time base of 2 seconds/div). Place the arm-cuff on volunteer s arm and pressurize it to 170mmHg. Release the pressure slowly and evenly by soft adjustment of the valve. Stop the measurement (STOP button on the scope) after you get a desired signal reading (Fig. 4). 1 (6)

2 Ch1 AC Ch2 DC OSC Fig. 1 Tool for blood pressure measurement 3) Power up the laptop computer (login: Student/Student), run the ADCPro application (Desktop). Switch on the power supply for AD converter demonstration kit (BK125, ±15V, +5V). Activate the option ADS1281DEVKIT in EVM menu. Select MultiScope mode in TEST menu. Press Acquire (Continuous) button to start the measurement. 4) Connect the BNC output connector of the absolute pressure signal conditioning unit to the ADC. Measure the output voltage in Multiscope mode for desired heights (altitudes). The requires power supply (BK125 ±15V). p 1 Absolute pressure POWER SUPPLY ±15VDC Constant current bridge excitation and output signal amplification Voltmeter ADS1281 Development Kit Fig. 2 Tool for altitude measurement 5) Connect the output of the differential pressure tool to the input of the AD converter. Switch on the power supply for the ventilators (~14V, 3A), connect A and B BNC connectors on the tool to the oscilloscope the two signals indicates the speed of rotation of the ventilators. Measure the dependency of differential pressure output voltage versus the air-flow speed (sum of A and B frequencies). Do the measurement for five different speeds of rotations. What happens if you block the airflow (with an open hand) for non-zero ventilator speed? POWER SUPPLY 14V DC 3A ROTATION SENSORS OUTPUTS SPEED OF Ch1 Ch2 ROTATION OSC f 1 +f 2 =? Air-flow direction Differential pressure p 1 p 2 orifice POWER SUPPLY ±15VDC Constant current bridge excitation and output signal amplification Voltmeter ADS1281 Development Kit Fig. 3 Tool for air-flow measurement 2 (6)

3 Notes for the measurement 1) The electrical circuit in the tool provides non-zero output voltage for zero applied pressure. The offset has to be calibrated. Ideally the transfer characteristics should be measured in several points, practically for our purpose two points calibration is sufficient. At first we measure U OFFSET (while p=0) and U 200 (while p=200mmhg). The pressure can be then calculated from the voltage readout U X : ( U X UOFFSET ) = ( ) 200 V p U 200 U mmhg ;, mmhg OFFSET V The gauge measures the pressure in the input port with respect to the atmosphere (the other side of the membrane is freely open). 2) Blood pressure measurement 1. increase the pressure in the arm-cup to 170 mmhg. Observe the signal on the oscilloscope. There are only small pulses visible for high pressure. The Korotkov pulses will appear if the pressure is decreased slowly and steadily (cca 2-3 mmhg/s). The pulses disappear if the pressure is further decreased. The automatic electronic tonometer measures the maximum amplitude of the pulses. There is a couple of definitions that can be used to determine the systolic and diastolic pressures. Probably the most common definition is presented on the picture. The systolic pressure Ps is estimated as an amplitude rise to 50% of the maximum value; while the diastolic pressure is measured for pressure decrease to 85% of the maximal amplitude. You can freeze the oscilloscopic screen by pressing RUN/STOP button and measure the desired values with cursors. Fig. 4 Korotkoff sounds definition of systolic and diastolic pressures 3) The ADS1281 IC made by Texas Instruments is currently one of the best (regulary commercially available) delta-sigma AD converters. It has 31-bits output word, effective resolution (ENOB Effective Number Of Bits) approximately 22-23bits (dependend on selected sampling frequency f S max 4kSa/s). The resolution of 23 bits corresponds roughly to 6.5 DMM. The ADCPro program in MultiFFT mode shows frequency spectrum of the acquired signal. The MultiHistogram mode shows the measured data in the form of column graph. The distribution of the code bins is used to characterize the ADC. The acquired signal is presented in timedomain in MultiScope mode. The most accurate measurements can be obtained for the lowest sampling frequency (250Sa/s). For record length of 256 samples we get an update of the mean value approximately each second. There was a 16 th order FIR filter (moving average) added to the program in order to suppress the noise. 3 (6)

4 Fig. 5 Texas Instruments ADS1281 internal structure ADCPro program, use menu EVM to run ADS1281EVM-PDK module (see picture below), then activate module MultiScope from Task menu. 4 (6) Fig. 8 ADCPro program (based on NI LabView) 4) Measurement of the absolute pressure of the atmosphere is often used to determine the altitude. The absolute contains a chamber with a reference pressure (often vacuum of various quality); on the other side of the membrane then acts the measured pressure. There is a drop of the atmospheric pressure of approximately 11.5 Pa for each ascended meter. Let s have a with pressure range of 15 psi (pound square inch) = Pa, with a sensitivity 90 mv/15 psi (for voltage excitation 12VDC), the instrumentation amplifier used for output amplification has an amplification of 25,7x. The is powered by constant current source (1.57 ma) in the tool (it removes the influence of wires resistance, it is good for temperature compensation); the mentioned current produces 7.35VDC on the. The output voltage should therefore decrease by 158µV for each ascended meter.

5 5) Differential pressure measures the difference between two pressures applied to two separate inputs. The output is ideally independent of the absolute pressure of the media (but the maximum absolute pressure is specified) A typical application for this is a flow measurement. The advantages of an orifice based flowmeter: very robust and simple, no moving parts, sufficient measurement accuracy. Drawbacks: there is a pressure drop across the orifice (more powerful ventilator must be used). The ventilators are powered by 12V constant voltage power supply and the speed is regulated by PWM in the laboratory tool. The ventilator pair (with counter rotating blades) provides and indication of the rotating speed (2 pulses per revolution). For simplicity we expect linear dependence between the speed (sum of speeds) and the generated air-flow. Suggested placement of the pressure sensing ports for flow measurement with an orifice. Air-flow and pressure profile in the tube. The pictures taken from: ISO Measurement of fluid flow by means of pressure differential devices inserted in circular-cross section conduits running full 5 (6)

6 Attachement Pressure s, the difference between absolute, gauge and differential pressure. p[kpa] 100 p A =169kPa p A =101kPa p A =64kPa p G =60kPa p G =20kPa p D =-20kPa atmospheric pressure p D =27kPa (absolute zero, 0 ideal vacuum) absolute p G =-68kPa gauge(relative) p D =40kPa differential p 1 vacuum (~0kPa) p 1 atmosphere (~101kPa) p 1 p 2 The electrical circuit used for constant current excitation and output signal conditioning. (voltage reference: REF V, op-amp OPA277, instrumentation amplifier AD8221). Advantages of four active elements bridge setup (e.g. with four piezo-resistors) linear output for both voltage and current excitation, four times more sensitive with respect to one sensing element, automatic compensation of temperature drifts for current excitation of the bridge (sensitivity compensation). Bridge topology with respect to non-bridge topology easy amplification (typically 10x-1000x) of the output signal. Amplification of a small signal rather than amplification of a small signal superimposed on some big offset signal. Piezo-resistive effect change of resistance of a semiconductor material due to applied mechanical stress tension. In contrast to piezo-electric effect no charge (voltage) generation takes place. 6 (6)

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES TSI and TSI logo are registered trademarks of TSI Incorporated. SmartTune is a trademark of TSI Incorporated. THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES IFA 300 Constant Temperature Anemometry

More information

Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz

Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar Network Analyzer combines a 90 db wide dynamic range with the accuracy and linearity

More information

Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement

Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement Chapter 2 Calibration and Use of a Strain-Gage-Instrumented Beam: Density Determination and Weight-Flow-Rate Measurement 2.1 Introduction and Objectives This laboratory exercise involves the static calibration

More information

Reading: HH Sections 4.11 4.13, 4.19 4.20 (pgs. 189-212, 222 224)

Reading: HH Sections 4.11 4.13, 4.19 4.20 (pgs. 189-212, 222 224) 6 OP AMPS II 6 Op Amps II In the previous lab, you explored several applications of op amps. In this exercise, you will look at some of their limitations. You will also examine the op amp integrator and

More information

Gas Custody Transfer Calibration

Gas Custody Transfer Calibration Gas Custody Transfer Calibration Using multi variable temperature / pressure calibrators for flowmeter calibration 2013 Introduction Gas custody transfer flow computers require special calibration to perform

More information

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE References for Nuclear Magnetic Resonance 1. Slichter, Principles of Magnetic Resonance, Harper and Row, 1963. chapter

More information

How to measure absolute pressure using piezoresistive sensing elements

How to measure absolute pressure using piezoresistive sensing elements In sensor technology several different methods are used to measure pressure. It is usually differentiated between the measurement of relative, differential, and absolute pressure. The following article

More information

Conversion Between Analog and Digital Signals

Conversion Between Analog and Digital Signals ELET 3156 DL - Laboratory #6 Conversion Between Analog and Digital Signals There is no pre-lab work required for this experiment. However, be sure to read through the assignment completely prior to starting

More information

Subminiature Load Cell Model 8417

Subminiature Load Cell Model 8417 w Technical Product Information Subminiature Load Cell 1. Introduction... 2 2. Preparing for use... 2 2.1 Unpacking... 2 2.2 Using the instrument for the first time... 2 2.3 Grounding and potential connection...

More information

Lab 1: The Digital Oscilloscope

Lab 1: The Digital Oscilloscope PHYSICS 220 Physical Electronics Lab 1: The Digital Oscilloscope Object: To become familiar with the oscilloscope, a ubiquitous instrument for observing and measuring electronic signals. Apparatus: Tektronix

More information

Power Amplifier Gain Compression Measurements

Power Amplifier Gain Compression Measurements Technical Brief Power Amplifier Gain Compression Measurements GPIB Private Bus Sweep Out Sweep In Pulse In AC Mod Out Blank/Marker Out Blanking In Overview The 1 db gain compression of an amplifier describes

More information

Electronics. Discrete assembly of an operational amplifier as a transistor circuit. LD Physics Leaflets P4.2.1.1

Electronics. Discrete assembly of an operational amplifier as a transistor circuit. LD Physics Leaflets P4.2.1.1 Electronics Operational Amplifier Internal design of an operational amplifier LD Physics Leaflets Discrete assembly of an operational amplifier as a transistor circuit P4.2.1.1 Objects of the experiment

More information

Pressure Transducer to ADC Application

Pressure Transducer to ADC Application Application Report SLOA05 October 2000 Pressure Transducer to ADC Application John Bishop ABSTRACT Advanced Analog Products/OpAmp Applications A range of bridgetype transducers can measure numerous process

More information

Inductors in AC Circuits

Inductors in AC Circuits Inductors in AC Circuits Name Section Resistors, inductors, and capacitors all have the effect of modifying the size of the current in an AC circuit and the time at which the current reaches its maximum

More information

Using Thermocouple Sensors Connecting Grounded and Floating Thermocouples

Using Thermocouple Sensors Connecting Grounded and Floating Thermocouples Connecting Grounded and Floating Thermocouples For best performance, Thermocouple sensors should be floating. This will ensure that no noise currents can flow in the sensor leads and that no common-mode

More information

HIGH RELIABILITY POWER SUPPLY TESTING

HIGH RELIABILITY POWER SUPPLY TESTING HIGH RELIABILITY POWER SUPPLY TESTING Dale Cigoy Keithley Instruments, Inc. The reliability of a power supply must match or exceed the rest of the system in which it is installed. Generally, this requires

More information

USB 2.0/3.0 Droop/Drop Test Fixture

USB 2.0/3.0 Droop/Drop Test Fixture Operations Manual USB-IF USB 2.0/3.0 Droop/Drop Test Fixture USB 2.0/3.0 Droop/Drop Test Fixture June 15, 2012 Revision 1.1 PACKAGE CONTENTS The USB-IF 2.0/3.0 Droop/Drop Test Fixture should come with

More information

Physical Specifications (Custom design available.)

Physical Specifications (Custom design available.) Helmholtz Coils A useful laboratory technique for getting a fairly uniform magnetic field, is to use a pair of circular coils on a common axis with equal currents flowing in the same sense. For a given

More information

1.44 kw Programmable DC Power Supplies XLN Series

1.44 kw Programmable DC Power Supplies XLN Series Data sheet 1.44 kw Programmable DC Power Supplies XLN Series New Family of High Density System Power Supplies The B&K Precision XLN series are compact, programmable, single-output DC power supplies, suitable

More information

Automotive Sensor Simulator. Automotive sensor simulator. Operating manual. AutoSim

Automotive Sensor Simulator. Automotive sensor simulator. Operating manual. AutoSim Automotive sensor simulator Operating manual AutoSim Contents Introduction.. page 3 Technical specifications.... page 4 Typical application of AutoSim simulator..... page 4 Device appearance... page 5

More information

PRODUCT PROFILE FEATURES

PRODUCT PROFILE FEATURES PRODUCT PROFILE Pressure Transmitters & Pneumatic Transducer Airflow Transmitter, Differential Air Pressure Transmitter and Differential & Gauge Liquid/Gas Pressure Transmitter FEATURES AIRFLOW TRANSMITTER

More information

4 Posiflow Proportional Solenoid Valves

4 Posiflow Proportional Solenoid Valves 4 Posiflow Solenoid Brass or Stainless Steel Bodies 1/4" to 1/2" NPT 2/2 Features Flow rates adjustable between 0% and 100% of rating Flow rate can also be regulated by a range of electrical inputs (sensors,

More information

Automation System TROVIS 6400 TROVIS 6493 Compact Controller

Automation System TROVIS 6400 TROVIS 6493 Compact Controller Automation System TROVIS 6400 TROVIS 6493 Compact Controller For panel mounting (front frame 48 x 96 mm/1.89 x 3.78 inch) Application Digital controller to automate industrial and process plants for general

More information

Transmitter Interface Program

Transmitter Interface Program Transmitter Interface Program Operational Manual Version 3.0.4 1 Overview The transmitter interface software allows you to adjust configuration settings of your Max solid state transmitters. The following

More information

9. Force Measurement with a Strain gage Bridge

9. Force Measurement with a Strain gage Bridge 9. Force Measurement with a Strain gage Bridge Task 1. Measure the transfer characteristics of strain-gage based force sensor in a range 0 10 kg 1. 2. Determine the mass of an unknown weight. 3. Check

More information

Impedance 50 (75 connectors via adapters)

Impedance 50 (75 connectors via adapters) VECTOR NETWORK ANALYZER PLANAR TR1300/1 DATA SHEET Frequency range: 300 khz to 1.3 GHz Measured parameters: S11, S21 Dynamic range of transmission measurement magnitude: 130 db Measurement time per point:

More information

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection w Technical Product Information Precision Miniature Load Cell with Overload Protection 1. Introduction The load cells in the model 8431 and 8432 series are primarily designed for the measurement of force

More information

Electrical Resonance

Electrical Resonance Electrical Resonance (R-L-C series circuit) APPARATUS 1. R-L-C Circuit board 2. Signal generator 3. Oscilloscope Tektronix TDS1002 with two sets of leads (see Introduction to the Oscilloscope ) INTRODUCTION

More information

EVAL-UFDC-1/UFDC-1M-16

EVAL-UFDC-1/UFDC-1M-16 Evaluation Board for Universal Frequency-to- Digital Converters UFDC-1 and UFDC-1M-16 EVAL-UFDC-1/UFDC-1M-16 FEATURES Full-Featured Evaluation Board for the Universal Frequency-to-Digital Converters UFDC-1

More information

LABORATORY 2 THE DIFFERENTIAL AMPLIFIER

LABORATORY 2 THE DIFFERENTIAL AMPLIFIER LABORATORY 2 THE DIFFERENTIAL AMPLIFIER OBJECTIVES 1. To understand how to amplify weak (small) signals in the presence of noise. 1. To understand how a differential amplifier rejects noise and common

More information

U1602A Handheld Oscilloscopes, 20 MHz

U1602A Handheld Oscilloscopes, 20 MHz Products & Services Technical Support Buy Industries About Agilent United States Home >... > Oscilloscopes > U1600A Series handheld oscilloscopes (2 models) > U1602A Handheld Oscilloscopes, 20 MHz Product

More information

Digital to Analog and Analog to Digital Conversion

Digital to Analog and Analog to Digital Conversion Real world (lab) is Computer (binary) is digital Digital to Analog and Analog to Digital Conversion V t V t D/A or DAC and A/D or ADC D/A Conversion Computer DAC A/D Conversion Computer DAC Digital to

More information

T0118 T2118 T3118. Instruction Manual

T0118 T2118 T3118. Instruction Manual Programmable indoor transmitter of temperature T0118 Programmable indoor transmitter of atmospheric pressure T2118 Programmable indoor transmitter of temperature, relative humidity and other derived humidity

More information

High Power Programmable DC Power Supplies PVS Series

High Power Programmable DC Power Supplies PVS Series Data Sheet High Power Programmable DC Power Supplies The PVS10005, PVS60085, and PVS60085MR programmable DC power supplies offer clean output power up to 5.1 kw, excellent regulation, and fast transient

More information

Application Note for SDP600 and SDP1000 Series Measuring Differential Pressure and Air Volume with Sensirion s CMOSens technology

Application Note for SDP600 and SDP1000 Series Measuring Differential Pressure and Air Volume with Sensirion s CMOSens technology Application Note for SDP600 and SDP1000 Series Measuring Differential Pressure and Air Volume with Sensirion s CMOSens technology Summary The increasing customer requirements for comfort and safety and

More information

Lock - in Amplifier and Applications

Lock - in Amplifier and Applications Lock - in Amplifier and Applications What is a Lock in Amplifier? In a nut shell, what a lock-in amplifier does is measure the amplitude V o of a sinusoidal voltage, V in (t) = V o cos(ω o t) where ω o

More information

Lab E1: Introduction to Circuits

Lab E1: Introduction to Circuits E1.1 Lab E1: Introduction to Circuits The purpose of the this lab is to introduce you to some basic instrumentation used in electrical circuits. You will learn to use a DC power supply, a digital multimeter

More information

WEA-Base. User manual for load cell transmitters. UK WEA-Base User manual for load cell transmitters Version 3.2 UK

WEA-Base. User manual for load cell transmitters. UK WEA-Base User manual for load cell transmitters Version 3.2 UK WEA-Base User manual for load cell transmitters 1 Contents 1. Technical data... 3 2. Assembly... 4 2.1 Power supply... 4 2.2 Load cells... 4 2.3 RS-485... 4 2.4 Relays... 5 2.5 Digital input... 5 2.6 Analogue

More information

4. Optical position sensor, photo-current measurement

4. Optical position sensor, photo-current measurement 4. Optical position sensor, photo-current measurement Task 1. Measure the output signal of photo-electric sensors Measure the relative dependency of output signal versus light intensity. The dependence

More information

The Sonometer The Resonant String and Timbre Change after plucking

The Sonometer The Resonant String and Timbre Change after plucking The Sonometer The Resonant String and Timbre Change after plucking EQUIPMENT Pasco sonometers (pick up 5 from teaching lab) and 5 kits to go with them BK Precision function generators and Tenma oscilloscopes

More information

The Ultimate British Valve Tester

The Ultimate British Valve Tester TEST EQUIPMENT The Ultimate British Valve Tester AVO VCM 163 by Kurt Schmid, DH3PJ To evaluate the quality of thermionic valves, ambitious valve testers (USA: tube tester) are able to diagnose a couple

More information

2/8.10.01. These vacuum switches are used in measuring ranges between -1 and 0 bar.

2/8.10.01. These vacuum switches are used in measuring ranges between -1 and 0 bar. CATALOGUE > Release 8.5 > Electronic vacuum switches Series SWM Electronic miniature vacuum switches Series SWM These vacuum switches are used in measuring ranges between -1 and 0 bar. Small dimension

More information

Wahl C50 Calibrator FEATURES

Wahl C50 Calibrator FEATURES NEW! On Site Multifunction Portable Far Superior Accuracy as Compared to Fluke! Simultaneous Measurement and Generation Quick Connect Terminals Anti-Shock Boot Quick Connect Terminals Navigator Power Supply,

More information

CruzPro VAF110. AC Volts, Amps, Frequency, kw Monitor

CruzPro VAF110. AC Volts, Amps, Frequency, kw Monitor CruzPro VAF110 AC Volts, Amps, Frequency, kw Monitor Table of Contents Introduction............................ 3 Specifications........................... 4 Installation..............................5

More information

THE STRAIN GAGE PRESSURE TRANSDUCER

THE STRAIN GAGE PRESSURE TRANSDUCER THE STRAIN GAGE PRESSURE TRANSDUCER Pressure transducers use a variety of sensing devices to provide an electrical output proportional to applied pressure. The sensing device employed in the transducers

More information

Practical Application of Industrial Fiber Optic Sensing Systems

Practical Application of Industrial Fiber Optic Sensing Systems Practical Application of Industrial Fiber Optic Sensing Systems John W. Berthold and David B. Needham Davidson Instruments, Inc. P.O. Box 130100, The Woodlands, TX 77393 ABSTRACT In this presentation,

More information

Lab 8: DC generators: shunt, series, and compounded.

Lab 8: DC generators: shunt, series, and compounded. Lab 8: DC generators: shunt, series, and compounded. Objective: to study the properties of DC generators under no-load and full-load conditions; to learn how to connect these generators; to obtain their

More information

Designing the Solution for Electrochemistry Potentiostat/Galvanostat І Battery Cycler І Fuel Cell Test Station +82-2-578-6516 І sales@wonatech.

Designing the Solution for Electrochemistry Potentiostat/Galvanostat І Battery Cycler І Fuel Cell Test Station +82-2-578-6516 І sales@wonatech. ZIVE Electrochemical Workstations ZIVE SP2 Designing the Solution for Electrochemistry Potentiostat/Galvanostat І Battery Cycler І Fuel Cell Test Station +82-2-578-6516 І sales@wonatech.com www.wonatech.com

More information

Engineering Sciences 151. Electromagnetic Communication Laboratory Assignment 3 Fall Term 1998-99

Engineering Sciences 151. Electromagnetic Communication Laboratory Assignment 3 Fall Term 1998-99 Engineering Sciences 151 Electromagnetic Communication Laboratory Assignment 3 Fall Term 1998-99 WAVE PROPAGATION II: HIGH FREQUENCY SLOTTED LINE AND REFLECTOMETER MEASUREMENTS OBJECTIVES: To build greater

More information

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON.

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON. WHY DIFFERENTIAL? Voltage, The Difference Whether aware of it or not, a person using an oscilloscope to make any voltage measurement is actually making a differential voltage measurement. By definition,

More information

RLC Series Resonance

RLC Series Resonance RLC Series Resonance 11EM Object: The purpose of this laboratory activity is to study resonance in a resistor-inductor-capacitor (RLC) circuit by examining the current through the circuit as a function

More information

Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist.

Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist. Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist. Abstract. I have been asked to put together a detailed article on a SWR meter. In this article I will deal

More information

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS 1 EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS The oscilloscope is the most versatile and most important tool in this lab and is probably the best tool an electrical engineer uses. This outline guides

More information

FREQUENCY RESPONSE ANALYZERS

FREQUENCY RESPONSE ANALYZERS FREQUENCY RESPONSE ANALYZERS Dynamic Response Analyzers Servo analyzers When you need to stabilize feedback loops to measure hardware characteristics to measure system response BAFCO, INC. 717 Mearns Road

More information

Tadahiro Yasuda. Introduction. Overview of Criterion D200. Feature Article

Tadahiro Yasuda. Introduction. Overview of Criterion D200. Feature Article F e a t u r e A r t i c l e Feature Article Development of a High Accuracy, Fast Response Mass Flow Module Utilizing Pressure Measurement with a Laminar Flow Element (Resistive Element) Criterion D200

More information

AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz

AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz AC Measurements Using the Oscilloscope and Multimeter by Mr. David Fritz 1 Sine wave with a DC offset f = frequency in Hz A = DC offset voltage (average voltage) B = Sine amplitude Vpp = 2B Vmax = A +

More information

Computer Controlled Generating Stations Control and Regulation Simulator, with SCADA SCE

Computer Controlled Generating Stations Control and Regulation Simulator, with SCADA SCE Technical Teaching Equipment Computer Controlled Generating Stations Control and Regulation Simulator, with SCADA SCE EDIBON SCADA System Teaching Technique used 4 5 2 Data Acquisition Board Cables and

More information

FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER

FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER 2014 Amplifier - 1 FREQUENCY RESPONSE OF AN AUDIO AMPLIFIER The objectives of this experiment are: To understand the concept of HI-FI audio equipment To generate a frequency response curve for an audio

More information

INSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS

INSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS INSTRUMENTATION AND CONTROL TUTORIAL 3 SIGNAL PROCESSORS AND RECEIVERS This tutorial provides an overview of signal processing and conditioning for use in instrumentation and automatic control systems.

More information

Operational Amplifier - IC 741

Operational Amplifier - IC 741 Operational Amplifier - IC 741 Tabish December 2005 Aim: To study the working of an 741 operational amplifier by conducting the following experiments: (a) Input bias current measurement (b) Input offset

More information

MATRIX TECHNICAL NOTES

MATRIX TECHNICAL NOTES 200 WOOD AVENUE, MIDDLESEX, NJ 08846 PHONE (732) 469-9510 FAX (732) 469-0418 MATRIX TECHNICAL NOTES MTN-107 TEST SETUP FOR THE MEASUREMENT OF X-MOD, CTB, AND CSO USING A MEAN SQUARE CIRCUIT AS A DETECTOR

More information

psi 2.901 0.001 fth20 6.690 0.001 mbar * 200.0 0.01/0.1 cmh20 203.9 0.1 inhg 5.906 0.001 bar 0.200 0.001

psi 2.901 0.001 fth20 6.690 0.001 mbar * 200.0 0.01/0.1 cmh20 203.9 0.1 inhg 5.906 0.001 bar 0.200 0.001 Manometer DD890 Introduction The device measures Gauge/Differential Pressure from -200 mbar to 200mbar (+-80.29 in H2O), it features 11 selectable units of measure: inh20, psi, bar, mbar, kpa, inhg, mmhg,ozin

More information

DATA LOGGING SYSTEM FOR PRESSURE MONITORING

DATA LOGGING SYSTEM FOR PRESSURE MONITORING DATA LOGGING SYSTEM FOR PRESSURE MONITORING Georgi Todorov Nikolov, Boyanka Marinova Nikolova, Marin Berov Marinov Department of Electronics, Technical University of Sofia, Studenstki Grad, TU-Sofia, block

More information

INSTALLATION INSTRUCTIONS BOOST CONTROLLER. Pro Control Input (Optional) Tach. Signal. Speed. Gray. Signal. Green Blue. Orange.

INSTALLATION INSTRUCTIONS BOOST CONTROLLER. Pro Control Input (Optional) Tach. Signal. Speed. Gray. Signal. Green Blue. Orange. 2650-1706-00 INSTALLATION INSTRUCTIONS BOOST CONTROLLER WARNING! The installation of the Auto Meter Boost Controller is recommended only for experienced technicians. This product may damage your engine

More information

E&I MAINTENANCE ENTRY TEST ENABLING OBJECTIVES. DESCRIBE hazards and precautions taken to avoid injury in the workplace.

E&I MAINTENANCE ENTRY TEST ENABLING OBJECTIVES. DESCRIBE hazards and precautions taken to avoid injury in the workplace. SAFETY Industrial DESCRIBE hazards and precautions taken to avoid injury in the workplace. Example #1: All of the following are common PPE used to perform maintenance activities EXCEPT: a. Safety Glasses

More information

SVPRIBOR ALFA DSL. Users Manual

SVPRIBOR ALFA DSL. Users Manual SVPRIBOR ALFA DSL Users Manual TABLE OF CONTENTS THE GENERAL INORMATION... 3 EXPLOITATION CONDITIONS... 3 CHARACTERISTICS... 3 PACKING LIST... 4 SOKCETS PANNEL... 5 CONTROL... 6 TURNING ON THE DEVICE...

More information

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b

DIODE CIRCUITS LABORATORY. Fig. 8.1a Fig 8.1b DIODE CIRCUITS LABORATORY A solid state diode consists of a junction of either dissimilar semiconductors (pn junction diode) or a metal and a semiconductor (Schottky barrier diode). Regardless of the type,

More information

DPI 260 SERIES: Digital Pressure Indicators

DPI 260 SERIES: Digital Pressure Indicators DPI 260 SERIES: Digital Pressure Indicators INTRODUCTION The DPI 260, 261 and 262 range of digital pressure instruments measure and indicate pressure in any specified scale units and provide accuracies

More information

QUICK START GUIDE FOR DEMONSTRATION CIRCUIT 956 24-BIT DIFFERENTIAL ADC WITH I2C LTC2485 DESCRIPTION

QUICK START GUIDE FOR DEMONSTRATION CIRCUIT 956 24-BIT DIFFERENTIAL ADC WITH I2C LTC2485 DESCRIPTION LTC2485 DESCRIPTION Demonstration circuit 956 features the LTC2485, a 24-Bit high performance Σ analog-to-digital converter (ADC). The LTC2485 features 2ppm linearity, 0.5µV offset, and 600nV RMS noise.

More information

Measuring Temperature withthermistors a Tutorial David Potter

Measuring Temperature withthermistors a Tutorial David Potter NATIONAL INSTRUMENTS The Software is the Instrument Application Note 065 Measuring Temperature withthermistors a Tutorial David Potter Introduction Thermistors are thermally sensitive resistors used in

More information

BOARD-LEVEL TEST PROCEDURE GYROSCOPE SUSPENSION SYSTEM (GSS) HV AMPLIFIER AND BRIDGE (HVA) ASSEMBLY. GP-B Procedure P0829 Rev A

BOARD-LEVEL TEST PROCEDURE GYROSCOPE SUSPENSION SYSTEM (GSS) HV AMPLIFIER AND BRIDGE (HVA) ASSEMBLY. GP-B Procedure P0829 Rev A P0829 Rev. A September 26, 2001 W. W. Hansen Experimental Physics Laboratory STANFORD UNIVERSITY STANFORD, CALIFORNIA 94305-4085 Gravity Probe B Relativity Mission BOARD-LEVEL TEST PROCEDURE GYROSCOPE

More information

RIGOL Data Sheet. DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D. Product Overview. Applications. Easy to Use Design

RIGOL Data Sheet. DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D. Product Overview. Applications. Easy to Use Design RIGOL Data Sheet DS1000E, DS1000D Series Digital Oscilloscopes DS1102E, DS1052E, DS1102D, DS1052D Product Overview DS1000E, DS1000D series are kinds of economical digital oscilloscope with high-performance.

More information

User s Guide DDS-3X25 USB ARBITRARY FUNCTION GENERATOR

User s Guide DDS-3X25 USB ARBITRARY FUNCTION GENERATOR User s Guide DDS-3X25 USB ARBITRARY FUNCTION GENERATOR Content General safety summary...1 Introduction...2 Chapter 1 Getting started...3 System Requirements...4 Installing Hardware...5 Installing Software...8

More information

22.302 Experiment 5. Strain Gage Measurements

22.302 Experiment 5. Strain Gage Measurements 22.302 Experiment 5 Strain Gage Measurements Introduction The design of components for many engineering systems is based on the application of theoretical models. The accuracy of these models can be verified

More information

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS Blood Pressure Project 9 Blood pressure is a measure of the fluid pressure within the circulatory system. This pressure is required to ensure the delivery of oxygen and nutrients to, and the removal of

More information

Dash 18X / Dash 18 Data Acquisition Recorder

Dash 18X / Dash 18 Data Acquisition Recorder 75 Dash 18X / Dash 18 Data Acquisition Recorder QUICK START GUIDE Supports Recorder System Software Version 3.1 1. INTRODUCTION 2. GETTING STARTED 3. HARDWARE OVERVIEW 4. MENUS & BUTTONS 5. USING THE DASH

More information

Evaluating AC Current Sensor Options for Power Delivery Systems

Evaluating AC Current Sensor Options for Power Delivery Systems Evaluating AC Current Sensor Options for Power Delivery Systems State-of-the-art isolated ac current sensors based on CMOS technology can increase efficiency, performance and reliability compared to legacy

More information

PHYSICS 111 LABORATORY Experiment #3 Current, Voltage and Resistance in Series and Parallel Circuits

PHYSICS 111 LABORATORY Experiment #3 Current, Voltage and Resistance in Series and Parallel Circuits PHYSCS 111 LABORATORY Experiment #3 Current, Voltage and Resistance in Series and Parallel Circuits This experiment is designed to investigate the relationship between current and potential in simple series

More information

Micro-Step Driving for Stepper Motors: A Case Study

Micro-Step Driving for Stepper Motors: A Case Study Micro-Step Driving for Stepper Motors: A Case Study N. Sedaghati-Mokhtari Graduate Student, School of ECE, University of Tehran, Tehran, Iran n.sedaghati @ece.ut.ac.ir Abstract: In this paper, a case study

More information

Multi-Range Programmable DC Power Supplies 9115 Series

Multi-Range Programmable DC Power Supplies 9115 Series Data Sheet 1200 W Multi-Range DC Power Supplies Features & Benefits Any 9115 series model can replace several supplies on your bench or in your rack. Unlike conventional supplies with fixed output ratings,

More information

TECHNICAL DATASHEET #TD1404AX PWM CONTROLLED SOLENOID DRIVER

TECHNICAL DATASHEET #TD1404AX PWM CONTROLLED SOLENOID DRIVER TECHNICAL DATASHEET #TD1404AX PWM CONTROLLED SOLENOID DRIVER (PWM Input, 1.2A or 2A Output, Metal Box or PCB) PCB Board - P/N: PWMC-PCB-2A, PWMC-PCB-1.2A Packaged Driver (metal box with 1.5 m (5 ft.) cable)

More information

Pulse Width Modulation Applications

Pulse Width Modulation Applications Pulse Width Modulation Applications Lecture 21 EE 383 Microcomputers Learning Objectives What is DTMF? How to use PWM to generate DTMF? How to use PWM to control a servo motor? How to use PWM to control

More information

NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706

NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706 (revised 4/21/03) NUCLEAR MAGNETIC RESONANCE Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706 Abstract This experiment studies the Nuclear Magnetic Resonance of protons

More information

I2C PRESSURE MONITORING THROUGH USB PROTOCOL.

I2C PRESSURE MONITORING THROUGH USB PROTOCOL. I2C PRESSURE MONITORING THROUGH USB PROTOCOL. Product Details: To eradicate human error while taking readings such as upper precision or lower precision Embedded with JAVA Application: Technology Used:

More information

X8 Option 2 - Technology

X8 Option 2 - Technology moog AC DC AC 4 to 450 A 4 to 210 A Compact MSD Servo Drive TTL in out Specification X8 Option 2 - Technology TTL Encoder Simulation / TTL Master Encoder moog Specification Option 2 - Technology TTL encoder

More information

AVR127: Understanding ADC Parameters. Introduction. Features. Atmel 8-bit and 32-bit Microcontrollers APPLICATION NOTE

AVR127: Understanding ADC Parameters. Introduction. Features. Atmel 8-bit and 32-bit Microcontrollers APPLICATION NOTE Atmel 8-bit and 32-bit Microcontrollers AVR127: Understanding ADC Parameters APPLICATION NOTE Introduction This application note explains the basic concepts of analog-to-digital converter (ADC) and the

More information

Servo Motors (SensorDAQ only) Evaluation copy. Vernier Digital Control Unit (DCU) LabQuest or LabPro power supply

Servo Motors (SensorDAQ only) Evaluation copy. Vernier Digital Control Unit (DCU) LabQuest or LabPro power supply Servo Motors (SensorDAQ only) Project 7 Servos are small, relatively inexpensive motors known for their ability to provide a large torque or turning force. They draw current proportional to the mechanical

More information

Current Probes. User Manual

Current Probes. User Manual Current Probes User Manual ETS-Lindgren L.P. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained herein shall

More information

Rackmount Intelligent Pressure Scanners

Rackmount Intelligent Pressure Scanners Rackmount Intelligent Pressure Scanners NetScanner System 9816/98RK-1 FEATURES ±0.05% Accuracy 100 Hz EU Throughput Autonegotiating 10/100/1G Base T Ethernet - TCP and UDP Protocol Pressure Ranges up to

More information

Three Axis TB6560 CNC Driver Users Manual

Three Axis TB6560 CNC Driver Users Manual Three Axis TB6560 CNC Driver Users Manual Revision 2.0 Oct. 16. 2009 1 Content 1. GENERAL INFORMATION... 3 1.1. Scope... 3 1.2. General Description... 3 2. Descriptions of 3-AXIS CNC Board... 3 2.1. Photo

More information

MAS.836 HOW TO BIAS AN OP-AMP

MAS.836 HOW TO BIAS AN OP-AMP MAS.836 HOW TO BIAS AN OP-AMP Op-Amp Circuits: Bias, in an electronic circuit, describes the steady state operating characteristics with no signal being applied. In an op-amp circuit, the operating characteristic

More information

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and

More information

Op Amp Circuit Collection

Op Amp Circuit Collection Op Amp Circuit Collection Note: National Semiconductor recommends replacing 2N2920 and 2N3728 matched pairs with LM394 in all application circuits. Section 1 Basic Circuits Inverting Amplifier Difference

More information

6/14/02 Chapter 14: Use of Electrical Test Equipment 1/20

6/14/02 Chapter 14: Use of Electrical Test Equipment 1/20 USE OF ELECTRICAL TEST EQUIPMENT Test equipment is necessary for determining proper set-up, adjustment, operation, and maintenance of electrical systems and control panels. The following is a general procedure

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK code Location code Customers Sites locations: 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK Calibration Centre Bolkiah Garrison BB3510 Negara Brunei Darussalam Contact: Mr Lim Tiong Thai Tel: +673-2-386475

More information

Performing Amplifier Measurements with the Vector Network Analyzer ZVB

Performing Amplifier Measurements with the Vector Network Analyzer ZVB Product: Vector Network Analyzer R&S ZVB Performing Amplifier Measurements with the Vector Network Analyzer ZVB Application Note This document describes typical measurements that are required to be made

More information

DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS

DDX 7000 & 8003. Digital Partial Discharge Detectors FEATURES APPLICATIONS DDX 7000 & 8003 Digital Partial Discharge Detectors The HAEFELY HIPOTRONICS DDX Digital Partial Discharge Detector offers the high accuracy and flexibility of digital technology, plus the real-time display

More information

Electronic Control Devices The European Product Catalogue 2007

Electronic Control Devices The European Product Catalogue 2007 FX14 Field Controller (1/4) FX14 Field Controller The FX14 is an equipment field controller in the Facility Explorer range of products. The controller is designed specifically for commercial Heating, Ventilating,

More information