1 TSI and TSI logo are registered trademarks of TSI Incorporated. SmartTune is a trademark of TSI Incorporated. THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

2 IFA 300 Constant Temperature Anemometry (CTA) System SmartTune automatic bridge optimization technology High frequency response (with no tuning) Windows -2-

3 Expansion up to 16 Channels While the basic IFA 300 system can handle up to eight anemometer channels in one cabinet, a total of sixteen anemometers can be monitored and controlled by ThermalPro software with the addition of a second cabinet (and the appropriate digitizer). In case you need two systems, the two IFA 300 cabinets can be used as two separate systems with the purchase of a second A/D converter. Thermal Pro Data Acquisition and Analysis Software The IFA 300 is designed for complete computer control through the ThermalPro software. System operating functions, such as control of the overheat ratios and signal conditioner settings, are handled via an RS232 interface, while the high speed analog data from the anemometer output is managed with a high speed multichannel digitizer installed inside the computer. The ThermalPro software is supplied with the IFA 300 system to allow all control and data acquisition, data analysis and presentation of results. -3-

4 Complete IFA 300 CTA System The complete IFA 300 system is made up of the following components: IFA 300 CTA Multichannel digitizer ThermalPro software Hot Wire/Hot film probe and probe support The IFA 300 unit comes with a single CTA channel, when multiple channels are needed, additional CTA can be purchased (up to 8 channels of CTA). If more than 8 channels are required for your flow research, an IFA 300 auxiliary cabinet is also required. This allows a total of 16 channels to be configured. System Selection Guide Component System with 1 channel System with up to 8 channels System with up to 16 channels IFA 300 IFA 300 auxiliary cabinet (183102) SmartTune CTA module Multichannel digitizer (ADCPCI-4) Multichannel digitizer (ADCPCI-8) (qty of 2) ThermalPro software Hot wire/hot film probes Computer system with Win 7 32-bit OS Air velocity calibration systems Traverse system Air velocity calibration by TSI Water flow calibration by TSI Optional items -4-

5 IFA 300 Specifications Amplifier drift Amplifier input noise Frequency response Resistance measurement Operating resistance Maximum probe current Cable length Output Size Number of channels Input power Temperature measurement Signal conditioning Offset Gain Low pass filters High pass filters Output (analog) Accessories included 0.3 μv/ C 1.7 nv/-hz and 1.5 pa/-hz >250 khz (without tuning) 0.1% ±0.01 ohms 2 to 80 ohms 800 ma (1.6 A with high power setting) 5 m; RG-58 A/U (other lengths available on request) Bridge voltage to 11.5 volts 17.8 cm x 48 cm x 41 cm (7 in. x 19 in. x 16 in.); standard 19-inch rack mount Up to eight per cabinet; sixteen channels supported by standard software 100/120/220/240 VAC, Hz, no switching required Thermocouple circuit built-in for type-t thermocouple 0 to 10 V in 10 mv steps, 0.15% accuracy 1 to 1000 (1.0 MHz maximum bandwidth), 0.15% accuracy Linear phase, 12 Hz to 1 MHz; 13 settings; -60 db/decade 0.1, 1, 10 Hz; 60 db/decade 5 to +5 V One 5-meter probe cable per channel, one 2-meter output cable per channel (plus 2 extra per system), one Model 1340 thermocouple, one Model 1341 thermocouple cable, one BNC shorting cap (coaxial), one BNC T connector, power cable(s), one 4-meter (12 ft) RS232 cable, one shorting probe for singlesensor probe supports and one for dual-sensor probe supports -5-

6 Air Velocity Calibrators Three models of air velocity calibrators are offered. They are: Model 1127 Model 1128A Model 1129 Manual velocity calibrator without pressure transducer B and C Manual velocity calibrator with different pressure transducer (0 to 100 mm Hg; 0 to 100 mmhg and mmhg respectively) Automatic velocity calibrator with 0 to 100 mm Hg pressure transducer The velocity calibrator can be used with one-, two-, or three sensor probe and provides the probe mounts for yaw/pitch calibration. The calibrator is also integrated with the IFA 300 and the ThermalPro software so that the data acquisition and calibration can be perfumed with accuracy and convenience. When the Model 1129 automatic velocity calibrator is used, the calibration becomes very easy. Just define a velocity range in the ThermalPro software and attach a probe, the calibrator monitors the velocity through the output of a built-in pressure transducer, acquires sensor data, adjusts the flow rate to a new velocity, and repeats the process, until the entire velocity range is covered. The data is automatically integrated into the software calibration section. While the calibrator s most obvious benefit is automated control, it has several other advanced design features. For one, its exclusive flow nozzle design exhibits a flat velocity profile over a broad dynamic range. For another, it sports a probe mounting system that can handle any type probe and probe support with great positioning flexibility, essential for yaw and pitch calibration of probes. -6-

7 Computer Controlled Traverse Systems Three different computer-controlled traverses are offered, Models and 1193, each able to precisely locate and move a probe. As part of an anemometer system, they enable automated measurement or remote operation. Typically, they are controlled using ThermalPro software, where a customer defined matrix identifies the measurement points. They also operate independently using front panel controls and displays on the power supply. All traverses are built to ensure stable operation over the full translation range. They feature 600 mm travel with 10μm resolution, both absolute and relative homing capability, and a universal probe mount with tilt. The three models differ only in the number of axes of movement. Specifications of the Traverse Systems Resolution 10 μm Accuracy ±50 μm/ft Repeatability ±20 μm Traversing speed 10 mm/s Backlash compensation Yes Limit switch settings Adjustable Stepper motor 200 steps/revolution Remote control RS232 Front panel LED display Position of each axis Front panel control Pushbutton; jog/slew from front panel Power source 110 VAC ±10%, 60 Hz standard; 220 VAC ±10%, 50/60 Hz optional Probe mount Manual rotation and translation of probe Model 1191 Model 1192 Model 1193 Axes Horizontal Horizontal/vertical Three axis Traversing length by by 600 by 600 (mm) Weight (including 29 kg 54 kg 104 kg mount Dimensions (mm) excluding mount 1070 x 100 x x 1100 x x 1100 x

8 Calibration Services TSI provides calibration data, supplied on disk, for all probes it manufactures. When ordering, indicate the velocity range in air (0 to 5 m/s, 0 to 50 m/s, or 0 to 100 m/s) or water (0 to 1 m/s or 0 to 5 m/s). Start-up Anemometer Systems Model 1750 Constant Temperature Anemometer Module The Model 1750 Constant Temperature Anemometer (CTA) is designed for startup multi-channel applications. Users typically incorporate the 1750 in their existing data acquisition and control system, and therefore it does not include any software, output signal conditioner, or computer interface. The following components are needed to be used with the Model 1750 CTA module to make up the complete system: Model 1751 Power Supply Model 1304 Control Resistor Single sensor wire or film probe Model Probe adaptor, or equivalent Model CL-CAL1750AIR calibration (please specify your velocity range) Each 1750 interfaces with a single-sensor anemometer probe. The 1750 system does not include signal conditioning, hence any offset and gain applied to the output voltage needs to be within the range of your analog to digital converter (typically -5 to +5V or 0 to 10V). Optimization of the frequency response can be achieved by adjusting the Trim and Gain controls on the anemometer. Specifications Number of channels Amplifier input noise Frequency response Maximum probe current Maximum bridge voltage Output voltage range Analog output impedance Sensor operating resistance Cable length Size (including mounting flange) Input Voltage Operating temperature 0 to 50 C One 7 nv/ Hz DC to 100 khz(3.8 μm) diameter tungsten wire, 100 m/s velocity 0.75 VDC Supply 22 VDC Supply 8 to 22 VDC Supply 50 Ω From 4 Ω to 20 Ω 2 m 11.4 cm x 10.7 cm x 2.8 cm (4.5 in. x 4.2 in. x 1.1 in.) 12 to 25 VDC -8-

9 Model 1745 Constant Current Anemometer module The Model 1745 Constant Current Anemometer (CCA) module is a stand-alone anemometer designed to measure average as well as instantaneous temperatures in the 0 o C to 300 o C range. The module is to be used with wire or film sensor operating as cold probe to allow the rapid measurements of temperature. The output from the anemometer is analog voltage reading directly dependent upon the actual temperature of the flow. Zero adjust and gain features allow convenient scaling of the output during calibration and measurement. Specifications Input power +15 VDC ±10%, 250 m/a Output range -12 VDC to +12 VDC, BNC output or terminal strip Sensor current Up to 2 ma Gain Continuous from 50 to 1000 Frequency response Up to 1 khz Proudly represented in Australia and New Zealand by: TSI Incorporated Visit our website for more information. P/N Rev A. (10/25/2013) 2013 TSI Incorporated Printed in U.S.A.

Research RF Sputtering Packages

Research RF Sputtering Packages Research RF Sputtering Packages Shown with 3" Polaris Adjustable Position Source with Tilt and Shutter 300 or 600 Watt Low Cost Packages for Those With Limited Budgets Desiring Full Capability 13.56 MHz

More information

Isolated AC Sine Wave Input 3B42 / 3B43 / 3B44 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated AC Sine Wave Input 3B42 / 3B43 / 3B44 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM Isolated AC Sine Wave Input 3B42 / 3B43 / 3B44 FEATURES AC averaging technique used to rectify, amplify, and filter 50 Hz to 400 Hz sine-wave signals. Accepts inputs of between 20 mv to 550 V rms to give

More information

Physical Specifications (Custom design available.)

Physical Specifications (Custom design available.) Helmholtz Coils A useful laboratory technique for getting a fairly uniform magnetic field, is to use a pair of circular coils on a common axis with equal currents flowing in the same sense. For a given

More information

JNIOR. Overview. Get Connected. Get Results. JNIOR Model 310. JNIOR Model 312. JNIOR Model 314. JNIOR Model 410

JNIOR. Overview. Get Connected. Get Results. JNIOR Model 310. JNIOR Model 312. JNIOR Model 314. JNIOR Model 410 The INTEG is an Ethernet I/O (digital, analog) device that monitors and controls a small set of process signals. functions as both basic I/O for integration with another application or system AND as a

More information

Signal Conditioning Amplifier System

Signal Conditioning Amplifier System Micro-Measurements A2 FEATURES Strain gage, transducer, and thermocouple inputs Frequency response to 110 khz Analog output of ±10 VDC Operation with 12 to 15 VDC and 120/240 VAC power Scalable from 8

More information

Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz

Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz Technical Datasheet Scalar Network Analyzer Model 8003-10 MHz to 40 GHz The Giga-tronics Model 8003 Precision Scalar Network Analyzer combines a 90 db wide dynamic range with the accuracy and linearity

More information

MPC 4. Machinery Protection Card Type MPC 4 FEATURES. Continuous on-line Machinery Protection Card

MPC 4. Machinery Protection Card Type MPC 4 FEATURES. Continuous on-line Machinery Protection Card Machinery Protection Card Type FEATURES Continuous on-line Machinery Protection Card Real-time measurement and monitoring using state-of-the-art DSP techniques Fully VME-compatible slave interface Fully

More information

22.302 Experiment 5. Strain Gage Measurements

22.302 Experiment 5. Strain Gage Measurements 22.302 Experiment 5 Strain Gage Measurements Introduction The design of components for many engineering systems is based on the application of theoretical models. The accuracy of these models can be verified

More information

On-line Machine Condition Monitoring [The CMS System]

On-line Machine Condition Monitoring [The CMS System] On-line Machine Condition Monitoring [The CMS System] Strategic monitoring control your costs Avoid unplanned downtime Monitor your plant stoppers Early warning maintains profit Efficient and profitable

More information

The Universal DAQ Device. Connect and measure immediately!

The Universal DAQ Device. Connect and measure immediately! The Universal DAQ Device. Connect and measure immediately! XXL performance in mini format 8 universal connectors 1) on each QuantumX device suited for most common transducer technologies: 1) See technical

More information

Ordering Part Numbers: Universal Signal Converter: USC-CVB225-01 Universal Signal Converter with a 4-20mA input and +/- 10V output: USC-CVB225-B10V

Ordering Part Numbers: Universal Signal Converter: USC-CVB225-01 Universal Signal Converter with a 4-20mA input and +/- 10V output: USC-CVB225-B10V Description: The universal signal converter offers threeway isolation and translates one input control signal into one to three simultaneous outputs. Switches allow the user to select the desired input

More information

Current Probes. User Manual

Current Probes. User Manual Current Probes User Manual ETS-Lindgren L.P. reserves the right to make changes to any product described herein in order to improve function, design, or for any other reason. Nothing contained herein shall

More information

Measuring Temperature withthermistors a Tutorial David Potter

Measuring Temperature withthermistors a Tutorial David Potter NATIONAL INSTRUMENTS The Software is the Instrument Application Note 065 Measuring Temperature withthermistors a Tutorial David Potter Introduction Thermistors are thermally sensitive resistors used in

More information

Agilent 4339B/4349B High Resistance Meters

Agilent 4339B/4349B High Resistance Meters Agilent 4339B/4349B High Resistance Meters Technical Overview Within Budget Without Compromise Introducing the Agilent Technologies 4339B and 4349B High Resistance Meters Used for Making Ultra- High Resistance

More information

T-SERIES INDUSTRIAL INCLINOMETER ANALOG INTERFACE

T-SERIES INDUSTRIAL INCLINOMETER ANALOG INTERFACE T-SERIES INDUSTRIAL INCLINOMETER ANALOG INTERFACE T-Series industrial inclinometers are compact high performance sensors used to determine inclination in roll and pitch axes with excellent precision and

More information

1.44 kw Programmable DC Power Supplies XLN Series

1.44 kw Programmable DC Power Supplies XLN Series Data sheet 1.44 kw Programmable DC Power Supplies XLN Series New Family of High Density System Power Supplies The B&K Precision XLN series are compact, programmable, single-output DC power supplies, suitable

More information

WEATHER STATION FOR SOLAR FARM MONITORING

WEATHER STATION FOR SOLAR FARM MONITORING WEATHER STATION FOR SOLAR FARM MONITORING SOLAR FARM MONITORING SYSTEM: Measures global, horizontal, & background irradiance. Measures wind speed, wind direction, ambient temperature, and relative humidity.

More information

Contents. Document information

Contents. Document information User Manual Contents Document information... 2 Introduction... 3 Warnings... 3 Manufacturer... 3 Description... Installation... Configuration... Troubleshooting...11 Technical data...12 Device Scope: PCB

More information

An internal gyroscope minimizes the influence of dynamic linear acceleration on slope sensor readings.

An internal gyroscope minimizes the influence of dynamic linear acceleration on slope sensor readings. TECHNICAL DATASHEET #TDAX06070X Triaxial Inclinometer with Gyro ±180⁰ Pitch/Roll Angle Pitch Angle Rate Acceleration SAE J1939, Analog Output or RS-232 Options 2 M12 Connectors, IP67 with Electronic Assistant

More information

DPI 260 SERIES: Digital Pressure Indicators

DPI 260 SERIES: Digital Pressure Indicators DPI 260 SERIES: Digital Pressure Indicators INTRODUCTION The DPI 260, 261 and 262 range of digital pressure instruments measure and indicate pressure in any specified scale units and provide accuracies

More information

Tilt switches N3 / N4... MEMS technology, one or two axis. Use. Options. + long lifetime and highly reliability due to mechanism-free MEMS technology

Tilt switches N3 / N4... MEMS technology, one or two axis. Use. Options. + long lifetime and highly reliability due to mechanism-free MEMS technology Tilt switches MEMS technology, one or two axis N3 / N4... + long lifetime and highly reliability due to mechanism-free MEMS technology + combinable output signals + one or two axis measurement + tilt range

More information

High Voltage Power Supplies

High Voltage Power Supplies Source voltages up to 5kV and 10kV 1µA current measurement resolution Multi-channel programmable DC power supplies Low noise for precision sourcing and sensitive measurements; selectable filters reduce

More information

Agilent 4338B Milliohm Meter

Agilent 4338B Milliohm Meter Agilent 4338B Milliohm Meter 10 µω to 100 kω Technical Overview Introduction Ideal for precise measurements of extremely low resistances using an ac test signal, the Agilent Technologies 4338B suits bench-top

More information

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421

VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421 VOLTAGE/CURRENT CALIBRATOR ISO-TECH ILC-421 TABLE OF CONTENTS 1. FEATURES... 1 2. SPECIFICATIONS... 1 2-1 General Specifications...1 2-2 Electrical Specifications... 2 3. FRONT PANEL DESCRIPTION... 4 3-1

More information

ADC-20/ADC-24 Terminal Board. User Guide DO117-5

ADC-20/ADC-24 Terminal Board. User Guide DO117-5 ADC-20/ADC-24 Terminal Board User Guide DO117-5 Issues: 1) 8.11.05 Created by JB. 2) 13.12.05 p10: added 0V connection to thermocouple schematic. 3) 22.3.06 p11: removed C1. 4) 20.8.07 New logo. 5) 29.9.08

More information

LOXONE 12 Channel Amplifier

LOXONE 12 Channel Amplifier LOXONE 12 Channel Amplifier Item no.: 200110 Thank you for purchasing the Loxone Twelve Channel Amplifier. The versatility of the Amplifier makes it the perfect choice for almost every type of custom multi-room

More information

Differential Charge Amplifier

Differential Charge Amplifier Electronics Differential Type 58A... The differential charge amplifier is used for signal conversion of piezoelectric sensors with differential output. Executions Aluminum die-cast enclosure (IP64) Plastic

More information

MODEL 1211 CURRENT PREAMPLEFIER

MODEL 1211 CURRENT PREAMPLEFIER MODEL 1211 CURRENT PREAMPLEFIER Phone: (607)539-1108 Email: info@dlinstruments.com www.dlinstruments.com The Model 1211 Current Preamplifier was designed to provide all of the features required of a modern

More information

Process modules Digital input PMI for 24 V DC inputs for 120 V AC inputs

Process modules Digital input PMI for 24 V DC inputs for 120 V AC inputs E031026 000823 Process modules Digital input PMI for inputs for 120 V AC inputs PMI Input E4, E5, GND L- PMI 120 V AC Input E4, E5, Common C E6, E7, GND L- E6, E7, Common C LEDs for the inputs operation

More information

Electronic Control Devices The European Product Catalogue 2007

Electronic Control Devices The European Product Catalogue 2007 FX14 Field Controller (1/4) FX14 Field Controller The FX14 is an equipment field controller in the Facility Explorer range of products. The controller is designed specifically for commercial Heating, Ventilating,

More information

DS 600. A contact free flux gate based current measurement sensor 600A rms

DS 600. A contact free flux gate based current measurement sensor 600A rms DS 600 A contact free flux gate based current measurement sensor 600A rms DS 600 is member of the small housing sensor family. The family includes a 200A and a 600A version. 600A rms - 900A peak Maximum

More information

HIGH RELIABILITY POWER SUPPLY TESTING

HIGH RELIABILITY POWER SUPPLY TESTING HIGH RELIABILITY POWER SUPPLY TESTING Dale Cigoy Keithley Instruments, Inc. The reliability of a power supply must match or exceed the rest of the system in which it is installed. Generally, this requires

More information

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet

Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers. Data Sheet Agilent N8973A, N8974A, N8975A NFA Series Noise Figure Analyzers Data Sheet Specifications Specifications are only valid for the stated operating frequency, and apply over 0 C to +55 C unless otherwise

More information

Agilent 8753ET/8753ES Network Analyzers

Agilent 8753ET/8753ES Network Analyzers Agilent 8753ET/8753ES Network Analyzers 8753ET, 300 khz to 3 or 6 GHz 8753ES, 30 khz to 3 or 6 GHz Configuration Guide System configuration summary The following summary lists the main components required

More information

Application Note TMA Series

Application Note TMA Series 1W, SIP, Single & Dual Output DC/DC Converters Features SIP Package with Industry Standard Pinout Package Dimension: 19.5 x 10.2 x 6.1 mm (0.77 x 0.4 x 0.24 ) 5V&12V Models 19.5 x 10.2 x 7.1 mm (0.77 x

More information

9000 Series Accelerometers Specifications

9000 Series Accelerometers Specifications Technical Data 9000 Series Accelerometers Specifications Catalog Numbers EK-43781I, EK-43782I, EK-46255I, EK-45890I, EK-43784I, EK-47086I, EK-43792I, EK-47090I, EK-43799I, EK-43794I, EK-43795I, EK-43797I,

More information

Computer Controlled Generating Stations Control and Regulation Simulator, with SCADA SCE

Computer Controlled Generating Stations Control and Regulation Simulator, with SCADA SCE Technical Teaching Equipment Computer Controlled Generating Stations Control and Regulation Simulator, with SCADA SCE EDIBON SCADA System Teaching Technique used 4 5 2 Data Acquisition Board Cables and

More information

INTREPID Perimeter Intrusion Detection System

INTREPID Perimeter Intrusion Detection System TM INTREPID Perimeter Intrusion Detection System Next Generation Perimeter Protection icropoint Cable combines patented Southwest Microwave technology with microprocessor power and laptop computer convenience.

More information

TSI THERMAL ANEMOMETRY PROBES

TSI THERMAL ANEMOMETRY PROBES TSI THERMAL ANEMOMETRY PROBES FOR MAXIMUM FLEXIBILITY IN DESIGNING YOUR THERMAL ANEMOMETRY APPLICATION UNDERSTANDING, ACCELERATED TABLE OF CONTENTS Probe Selection Guide... Probe Specifications...4 Sensor

More information

Multi-Range Programmable DC Power Supplies 9115 Series

Multi-Range Programmable DC Power Supplies 9115 Series Data Sheet Multi-Range Programmable DC Power Supplies 1200 W / 3000 W Multi-Range DC Power Supplies Features & Benefits Any model can replace several supplies on your bench or in your rack. Unlike conventional

More information

Perimeter Security System

Perimeter Security System TM Perimeter Security System The Next Generation Perimeter Security System ICRONET combines patented Southwest Microwave technology with microprocessor power and laptop computer convenience. It is a sophisticated

More information

Lab E1: Introduction to Circuits

Lab E1: Introduction to Circuits E1.1 Lab E1: Introduction to Circuits The purpose of the this lab is to introduce you to some basic instrumentation used in electrical circuits. You will learn to use a DC power supply, a digital multimeter

More information

Multi-Range Programmable DC Power Supplies 9115 Series

Multi-Range Programmable DC Power Supplies 9115 Series Data Sheet 1200 W Multi-Range DC Power Supplies Features & Benefits Any 9115 series model can replace several supplies on your bench or in your rack. Unlike conventional supplies with fixed output ratings,

More information

PRODUCT PROFILE FEATURES

PRODUCT PROFILE FEATURES PRODUCT PROFILE Pressure Transmitters & Pneumatic Transducer Airflow Transmitter, Differential Air Pressure Transmitter and Differential & Gauge Liquid/Gas Pressure Transmitter FEATURES AIRFLOW TRANSMITTER

More information

GT Sensors Precision Gear Tooth and Encoder Sensors

GT Sensors Precision Gear Tooth and Encoder Sensors GT Sensors Precision Gear Tooth and Encoder Sensors NVE s GT Sensor products are based on a Low Hysteresis GMR sensor material and are designed for use in industrial speed applications where magnetic detection

More information

Precision, Unity-Gain Differential Amplifier AMP03

Precision, Unity-Gain Differential Amplifier AMP03 a FEATURES High CMRR: db Typ Low Nonlinearity:.% Max Low Distortion:.% Typ Wide Bandwidth: MHz Typ Fast Slew Rate: 9.5 V/ s Typ Fast Settling (.%): s Typ Low Cost APPLICATIONS Summing Amplifiers Instrumentation

More information

Description. Output Stage. 5k (10k) - + 5k (10k)

Description. Output Stage. 5k (10k) - + 5k (10k) THAT Corporation Low Noise, High Performance Audio Preamplifier IC FEATURES Low Noise: 1 nv/hz input noise (60dB gain) 34 nv/hz input noise (0dB gain) (1512) Low THD+N (full audio bandwidth): 0.0005% 40dB

More information

USB Smart Power Sensor

USB Smart Power Sensor 50Ω -30 dbm to +20 dbm, 9 khz to 4000 MHz The Big Deal Low cost USB HID device compatible with 32/64 Bit operating systems Includes Measurement Application GUI (Graphical User Interface) software with

More information

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S

OPERATION MANUAL. Pen type, separate electrode SOIL PH METER Model : PH-220S Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into the field of precision measurement. Although this METER is a complex and

More information

US-SPI New generation of High performances Ultrasonic device

US-SPI New generation of High performances Ultrasonic device US-SPI New generation of High performances Ultrasonic device Lecoeur Electronique - 19, Rue de Courtenay - 45220 CHUELLES - Tel. : +33 ( 0)2 38 94 28 30 - Fax : +33 (0)2 38 94 29 67 US-SPI Ultrasound device

More information

MODEL 2202IQ (1991-MSRP $549.00)

MODEL 2202IQ (1991-MSRP $549.00) F O R T H E L O V E O F M U S I C F O R T H E L O V E O F M U S I C MODEL 2202IQ (1991-MSRP $549.00) OWNER'S MANUAL AND INSTALLATION GUIDE INTRODUCTION Congratulations on your decision to purchase a LINEAR

More information

FREQUENCY RESPONSE ANALYZERS

FREQUENCY RESPONSE ANALYZERS FREQUENCY RESPONSE ANALYZERS Dynamic Response Analyzers Servo analyzers When you need to stabilize feedback loops to measure hardware characteristics to measure system response BAFCO, INC. 717 Mearns Road

More information

How To Use A Sound Card With A Subsonic Sound Card

How To Use A Sound Card With A Subsonic Sound Card !"## $#!%!"# &"#' ( "#' )*! #+ #,# "##!$ -+./0 1" 1! 2"# # -&1!"#" (2345-&1 #$6.7 -&89$## ' 6! #* #!"#" +" 1##6$ "#+# #-& :1# # $ #$#;1)+#1#+

More information

TAC I/NETTM MR-VAV-AX. Application Specific MicroRegulator TM

TAC I/NETTM MR-VAV-AX. Application Specific MicroRegulator TM TAC I/NETTM TM TAC s TM has an extensive range ofpre-engineered VAV control sequences, which combined with a fully integrated actuator and airflow transducer, makes it a low Total Installed Cost solution

More information

Digital Magnetometer Systems - DM Series

Digital Magnetometer Systems - DM Series The Digital Magnetometer DM Series DM 005, DM-010, DM 050 and DM-060 are digital 3 axes magnetometer systems for precise measurement of magnetic fields and also applicable for underwater operation. All

More information

Voltech DC1000. Precision DC Bias Current Source

Voltech DC1000. Precision DC Bias Current Source DC1000 Precision DC Bias Current Source DC1000 Precision DC Bias Current Source TM Integral Rack Mount Lugs High Contrast Green LED Display Easy Control Rotary Knob Output LED Indicator DC 1000 OUTPUT

More information

LT-82 Stationary IR Transmitter

LT-82 Stationary IR Transmitter LT-82 Stationary IR Transmitter Configuration LT-82-0 (North America) LT-82-02 (Asia, UK) LT-82-03 (Euro) The Listen LT-82 is the heart of a stationary IR listening system. It takes the desired audio signal

More information

Cisco Model DPC3010 DOCSIS 3.0 8x4 Cable Modem

Cisco Model DPC3010 DOCSIS 3.0 8x4 Cable Modem Cisco Model DPC3010 DOCSIS 3.0 8x4 Cable Modem The Cisco Model DPC3010 DOCSIS 3.0 8x4 Cable Modem (DPC3010) provides DOCSIS broadband service providers with a cost-effective solution for delivering high-speed

More information

DRTS 33. The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters

DRTS 33. The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters Testing all relay technologies: electromechanical, solid state, numerical and IEC61850 Manual

More information

Impedance 50 (75 connectors via adapters)

Impedance 50 (75 connectors via adapters) VECTOR NETWORK ANALYZER PLANAR TR1300/1 DATA SHEET Frequency range: 300 khz to 1.3 GHz Measured parameters: S11, S21 Dynamic range of transmission measurement magnitude: 130 db Measurement time per point:

More information

Zlinx Wireless I/O. Peer-to-Peer and Modbus I/O B&B ELECTRONICS PRODUCT INFORMATION

Zlinx Wireless I/O. Peer-to-Peer and Modbus I/O B&B ELECTRONICS PRODUCT INFORMATION Modular, Customizable Wire Replacement 128 / 256 Bit AES Encryption Software Selectable RF Transmit Power Software Selectable Over-the-air Data Rate Modbus ASCII /RTU Compatible Wide Operating Temperature

More information

Unidirectional Transmitter/ Receiver Units Introduction

Unidirectional Transmitter/ Receiver Units Introduction Unidirectional Transmitter/ Receiver Units Introduction Wireless Input/Output (I/O) connects directly to analog, discrete and pulse transducer signals. The signals are transmitted by radio and either re-created

More information

O C M III. Open Channel Flow Monitor

O C M III. Open Channel Flow Monitor O C M III Open Channel Flow Monitor Part 1. General 1.1 Scope A. This section describes the requirements for an ultrasonic flow transmitter with sensor and temperature compensation probe. B. Under this

More information

IRT Eurocard. Types DAX-3200

IRT Eurocard. Types DAX-3200 I R T Electronics Pty Ltd A.B.N. 35 000 832 575 26 Hotham Parade, ARTARMON N.S.W. 2064 AUSTRALIA National: Phone: (02) 9489 3744 Fax: (02) 9439 7439 International: +61 2 9439 3744 +61 2 9439 7439 Email:

More information

Dielectric Withstand and Ground Bond Testers For Production Line Safety Agency Compliance Testing

Dielectric Withstand and Ground Bond Testers For Production Line Safety Agency Compliance Testing Dielectric Withstand and Ground Bond Testers For Production Line Safety Agency Compliance Testing U.S. patent#: 5,548,501, 6,054,865. Other patents pending. Safety agency listed. ENHANCED GRAPHIC LCD The

More information

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE References for Nuclear Magnetic Resonance 1. Slichter, Principles of Magnetic Resonance, Harper and Row, 1963. chapter

More information

PORTER MASS FLOW MEASUREMENT AND CONTROL SYSTEMS

PORTER MASS FLOW MEASUREMENT AND CONTROL SYSTEMS PORTER MASS MEASUREMENT AND CONTROL SYSTEMS PORTER: THE STANDARD OF EXCELLENCE IN MASS Porter Mass Flow products reflect nearly three decades of experience in the design and manufacture of precision instruments

More information

14.5GHZ 2.2KW CW GENERATOR. GKP 22KP 14.5GHz WR62 3x400V

14.5GHZ 2.2KW CW GENERATOR. GKP 22KP 14.5GHz WR62 3x400V 14.5GHZ 2.2KW CW GENERATOR GKP 22KP 14.5GHz WR62 3x400V UTILIZATION OF GKP 22KP GENERATOR With its characteristics of power stability whatever the load, very fast response time at a pulse, low ripple,

More information

Product Information. Gateway For Connecting EnDat Encoders to PROFIBUS-DP

Product Information. Gateway For Connecting EnDat Encoders to PROFIBUS-DP Product Information Gateway For Connecting EnDat Encoders to PROFIBUS-DP April 2012 PROFIBUS Gateway For Connecting EnDat Encoders Encoders with EnDat interface for connection via gateway All absolute

More information

JUMPFLEX 857 Series. 0...20 ma 2...10 V. 4...20 ma. Transducers / Relay and Optocoupler Modules

JUMPFLEX 857 Series. 0...20 ma 2...10 V. 4...20 ma. Transducers / Relay and Optocoupler Modules 0888-0172/0002-3601 JUMPFLEX 857 Series 2.0 E 12/08 Printed in Germany Subject to design changes 2...10 V 0...20 ma 4...20 ma WAGO Kontakttechnik GmbH & Co. KG Postfach 2880 32385 Minden Hansastraße 27

More information

M2M I/O Modules. To view all of Advantech s M2M I/O Modules, please visit www.advantech.com/products.

M2M I/O Modules. To view all of Advantech s M2M I/O Modules, please visit www.advantech.com/products. M2M I/O Modules 14 M2M I/O Modules Overview 14-2 M2M I/O Modules Selection Guide 14-6 ADAM-2510Z Wireless Router Node 14-8 ADAM-2520Z Wireless Modbus RTU Gateway 14-9 ADAM-2031Z ADAM-2632Z ADAM-2017Z ADAM-2018Z

More information

Electronic Control Devices The European Product Catalogue 2007

Electronic Control Devices The European Product Catalogue 2007 FX07 Field Controller (1/4) FX07 Field Controller Display Model Dimensions in mm (inches) The FX07 is a terminal unit controller in the Facility Explorer range of products. The controller is designed specifically

More information

Agilent 8757D Scalar Network Analyzer 10 MHz to 110 GHz

Agilent 8757D Scalar Network Analyzer 10 MHz to 110 GHz Agilent 8757D Scalar Network Analyzer 10 MHz to 110 GHz Data Sheet Accurate measurement of transmission and reflection characteristics is a key requirement in your selection of a scalar network analyzer.

More information

Voltage Regulator SPAU 341 C. Product Guide

Voltage Regulator SPAU 341 C. Product Guide Issued: July 1998 Status: Updated Version: D/25.04.2006 Data subject to change without notice Features Comprehensive voltage regulation for power transformers with on-load tapchangers in distribution substations

More information

Thermal Flow Sensors. Anemometers. Calorimeters. Ensemble of both

Thermal Flow Sensors. Anemometers. Calorimeters. Ensemble of both Thermal Flow Sensors Anemometers Calorimeters Ensemble of both IST Presentation V4-2005 www.ist-ag.com 1 Constant Temperature Anemometer At the heart of a thermal anemometer are two sensors: an air velocity

More information

How To Sell A Talan

How To Sell A Talan The TALAN represents state-of-the-art capability to rapidly and reliably detect and locate illicit tampering and security vulnerabilities on both digital and analog telephone systems. Marketing Characteristics

More information

Product Specifications

Product Specifications Cisco Model EPR2320 EuroDOCSIS 2.0 Residential Gateway with Wireless Access Point The Cisco Model EPR2320 EuroDOCSIS 2.0 Residential Gateway (EPR2320) combines a cable modem, router, and wireless access

More information

Model #: SMART2200CRMXL

Model #: SMART2200CRMXL Tripp Lite 1111 West 35th Street Chicago, IL 60609 USA Telephone: +(773) 869 1234 E-mail: saleshelp@tripplite.com Model #: SMART2200CRMXL SmartPro Compact Rackmount UPS System - Precision Protection for

More information

Typical Aluminum GFM Mass Flow Meter

Typical Aluminum GFM Mass Flow Meter MASS FLOW METERS Design Features n Rigid metallic construction. n Maximum pressure of 1000 psig (70 bars). n Leak integrity 1 x 10-9 of helium. n NIST traceable certification. n Built-in tiltable LCD readout.

More information

NETWORK ENABLED EQUIPMENT MONITOR

NETWORK ENABLED EQUIPMENT MONITOR NETWORK ENABLED EQUIPMENT MONITOR Remotely Monitor Sensors over the Internet Connect Sensors to the Web to Remotely Monitor Equipment, Processes or Other Applications A Complete, Easy to Deploy, Stand-Alone

More information

Op Amp Circuit Collection

Op Amp Circuit Collection Op Amp Circuit Collection Note: National Semiconductor recommends replacing 2N2920 and 2N3728 matched pairs with LM394 in all application circuits. Section 1 Basic Circuits Inverting Amplifier Difference

More information

DSA800 Series Spectrum Analyzer

DSA800 Series Spectrum Analyzer DSA800 Series Spectrum Analyzer Configuration Guide This guide is used to help users to configure DSA800 series spectrum analyzer according to their requirements. You can get an overall understanding of

More information

DCS Series 1 kw, 1.2 kw, & 3 kw

DCS Series 1 kw, 1.2 kw, & 3 kw DCS Series 1 kw, 1.2 kw, & 3 kw AVAILABLE WITH LXI STANDARD COMPLIANT ETHERNET Applications Automotive Electronics Rackmount ATE System Battery Charging DCS Series DCS Series: 1 kw, 1.2 kw, & 3 kw DCS

More information

EFC 3600. Frequency converters

EFC 3600. Frequency converters 2 Bosch Rexroth AG Electric Drives and Controls Documentation Compact and complete: space saving side-by-side assembly, plug-in I/O terminals, with brake chopper and mains filter for ultra-simple installation

More information

Current Monitoring Kit

Current Monitoring Kit Current Monitoring Kit QUICK START GUIDE DO090-6 CONTENTS Issues: 1) 2.10.02 WP A4 format 2) 2.10.02 Added safety warning 3) 17.3.06 Word A5 format. S1: Removed relay modules. S2: Added MI010. S4.1: added

More information

Technical Data Sheet. 3DLevelScanner

Technical Data Sheet. 3DLevelScanner Changing the market from level to volume Technical Data Sheet APM`s 3DLevelScanner incorporates advanced technology for accurately measuring bulk solids and powder stored in silos and open bins of all

More information

www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters

www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters www.kem-kueppers.com info@kem-kueppers.com Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters Description The Tricor Mass Flow Meters measure simultaneously mass

More information

DC Electronic Loads 8500 series

DC Electronic Loads 8500 series Data Sheet DC Electronic Loads 8500 series 2400W 600 W - 1200 W 300 W Versatile & Economical DC Electronic Loads The 8500 series Programmable DC Electronic Loads can be used for testing and evaluating

More information

CONTENTS. PAGE Chapter 1 General Information 1. Chapter 2 Setup 4 DP450-VDC 7 DP450-VAC 7 DP450-DCC 7 DP450-F 8 DP450-HACC 9 DP450-HDCC 11 DP450-E 13

CONTENTS. PAGE Chapter 1 General Information 1. Chapter 2 Setup 4 DP450-VDC 7 DP450-VAC 7 DP450-DCC 7 DP450-F 8 DP450-HACC 9 DP450-HDCC 11 DP450-E 13 CONTENTS PAGE Chapter 1 General Information 1 Chapter 2 Setup 4 DP450-VDC 7 DP450-VAC 7 DP450-DCC 7 DP450-F 8 DP450-HACC 9 DP450-HDCC 11 DP450-E 13 Chapter 3 Alarm/Excitation Options 15 Wiring 17 Chapter

More information

Modular I/O System Analog and Digital Interface Modules

Modular I/O System Analog and Digital Interface Modules OPERATING INSTRUCTIONS Modular I/O System Analog and Digital Interface Modules Installation Operation Maintenance Document Information Document ID Title: Operating Instructions Modular I/O System Part

More information

NEW. EVEN MORE data acquisition and test stand automation

NEW. EVEN MORE data acquisition and test stand automation NEW EVEN MORE data acquisition and test stand automation the new class of data The plug&play complete package User benefits Expert Series is the latest generation of data acquisition Complete hardware

More information

K-Type Thermocouple Sensor User s Guide

K-Type Thermocouple Sensor User s Guide K-Type Thermocouple Sensor User s Guide 1 TABLE OF CONTENTS: 1 INTRODUCTION... 2 2 TYPICAL APPLICATION:... 2 3 INSTALLATION RULES:... 2 3.1 Connecting the sensor to M1/MD4 data logger:... 2 3.2 Connecting

More information

IRT Eurocard. Type DAX-3206. Audio Extractor for 270 Mb/s SDI

IRT Eurocard. Type DAX-3206. Audio Extractor for 270 Mb/s SDI I R T Electronics Pty Ltd A.B.N. 35 000 832 575 26 Hotham Parade, ARTARMON N.S.W. 2064 AUSTRALIA National: Phone: (02) 9439 3744 Fax: (02) 9439 7439 International: 61 2 9439 3744 61 2 9439 7439 Email:

More information

Duct Humidity Transmitter

Duct Humidity Transmitter SDC-H Duct Humidity Transmitter Features Replaceable sensor element Humidity measurement for air ducts Minimum and maximum value memory 0 0V, 0 0mA or 0V, 4 0mA measuring signals selectable with jumpers

More information

Application Note Noise Frequently Asked Questions

Application Note Noise Frequently Asked Questions : What is? is a random signal inherent in all physical components. It directly limits the detection and processing of all information. The common form of noise is white Gaussian due to the many random

More information

PIEZOMETER AIRFLOW MEASURING RING

PIEZOMETER AIRFLOW MEASURING RING PIEZOMETER AIRFLOW MEASURING RING M.K. Plastics is pleased to introduce The Piezometer Airflow Measuring Ring, now available as an accessory for our steel Axijet fans. The system consists of a Piezometer

More information

Product Information S N O. Portable VIP protection CCTV & Alarm System 2

Product Information S N O. Portable VIP protection CCTV & Alarm System 2 Product Information S N O Portable VIP protection CCTV & Alarm System 2 G O V E R N M E N T A L S E C U R I T Y S O L U T I VIP KIT Rapid Deployment VIP Protection Kit The VIP KIT has been designed to

More information

High Power Programmable DC Power Supplies PVS Series

High Power Programmable DC Power Supplies PVS Series Data Sheet High Power Programmable DC Power Supplies The PVS10005, PVS60085, and PVS60085MR programmable DC power supplies offer clean output power up to 5.1 kw, excellent regulation, and fast transient

More information

Cond-IS RF Conducted Immunity System

Cond-IS RF Conducted Immunity System Cond-IS RF Conducted Immunity System About RF Conducted Immunity In the global contest of EMC testing for residential and industrial EUTs (Equipments Under Test) RF Conducted Immunity compliance verification

More information

Product Overview 9100D

Product Overview 9100D Product Overview 9100D Portable Shaker Vibration Calibrator The Modal Shop 3149 East Kemper Road Cincinnati, OH 45241 Phone: +1.513.351.9919 Fax: 513.458.2171 info@modalshop.com Brief Overview We've taken