Method validation and quality control procedures

Size: px
Start display at page:

Download "Method validation and quality control procedures"

Transcription

1 Method validation and quality control procedures Department of Food Chemistry and Analysis, ICT Prague Vladimír Kocourek Prague, 2012

2 WHAT IS VALIDATION? Validation is a process, within which the method is demonstrated to be suitable for its purpose. It documents methods performance! During validation process, methods Performance characteristics are estimated. Validation documents, that the methods performance characteristics are capable of producing results in line with the needs of the analytical problem. Is it possible to detect pesticide residues at regulation levels using the method? Is it possible to correctly quantify the amount of residues in apple/orange/? Validation procedure (protocol) is related to a particular analyte and matrix Applicability

3 Analytical method for decision making Defining the task Selection of method Sample handling Analytical procedure measuring Calibration a Calculation Data processing Interpretation, Presentation

4 VALIDATION VALIDATION PARAMETERS PRECISION TRUENESS ACCURACY RANGE & LINEARITY LIMIT OF DETECTION & LIMIT OF QUANTIFICATION SPECIFITY & SELECTIVITY RUGGEDNESS

5 TRUENESS AND PRECISION = ACCURACY ERRORS OF MEASUREMENT What is included in result value (X)? X = μ + ε + Σδ RELATED TO TRUENESS RELATED TO PRECISION X mean result of repeated measurement µ true value real content of analyte in sample ε systematic error always the same value and sign (+/-) Σδ sum of random errors variable value and sign (+/-) Error should not be confused with a mistake!!!

6 TRUENESS AND PRECISION = ACCURACY ERRORS OF MEASUREMENT What is included in result value (X)? X = μ + ε + Σ δ True value is an idealized concept and true value cannot be known exactly! Hence the REFERENCE VALUE represents a true value in routine practice Reference value usually provided with reference to: Certified reference material Reference measurement procedure Known amount of analyte added into the sample (spike)

7 TRUENESS TRUENESS is closeness of agreement between the mean of an N number of replicate measured values and a REFERENCE (TRUE) value. Trueness is inversely related to systematic error: The lower the systematic error, the higher the trueness Estimate of a systematic error: bias In analytical chemistry: RECOVERY Correction of the result can be carried out using the recovery => compensation for an estimated systematic effect

8 Traceability to reference material: trueness Assessment of trueness using CRM SAMPLE CRM (matrix) RESULT Compare with certified value t- test : t exp < t crit (P, n-1)

9 Certified Reference Material (CRM, SRM)

10 TRUENESS HOW TO ESTIMATE TRUENESS (RECOVERY) (Certified) Reference materials are available CRM Y ± y Sample 1 Sample 2 Sample 3 Sample 4 Sample 5 n 6 ANALYTICAL PROCESS Value 1 Value 2 Value 3 Value 4 Value 5 COMPARISON STATISTICAL TESTING MEAN VALUE X ± x (C)RMs at concentration levels close to expected analyte levels should be used

11 TRUENESS HOW TO ESTIMATE TRUENESS (RECOVERY) (Certified) Reference materials are not available THE MOST COMMON CASE A) BLANK SAMPLE IS AVAILABLE SAMPLE BLANK Sample 1 Sample 2 Sample 3 Sample 4 Sample 5 n 6 INCUBATION ANALYTICAL PROCESS Value 1 Value 2 Value 3 Value 4 Value 5 Addition of known amount of an analyte to each replicate (SPIKE) MEAN VALUE X ± x Blank samples should be checked, multiple spiking levels should be used After spike addition sample should be incubated (analyte incorporation into the matrix)

12 Certified Reference Material (CRM, SRM)

13 TRUENESS HOW TO ESTIMATE TRUENESS (RECOVERY) (Certified) Reference materials are not available A) BLANK SAMPLE IS AVAILABLE RECOVERY (%) MEAN VALUE =. 100 ADDED AMOUNT Recovery values can be both below or above 100% Recoveries between 80 and 120 are usually acceptable.

14 TRUENESS HOW TO ESTIMATE TRUENESS (RECOVERY) (Certified) Reference materials are not available B) BLANK SAMPLE IS NOT AVAILABLE Sample 1 Sample 2 ANALYTICAL PROCESS Value 1 Value 2 Sample 3 Value 3 SAMPLE Sample 4 Sample 5 Value 4 Value 5 n 6 MEAN VALUE Sample 1 Value 1 SAMPLE Sample 2 Sample 3 Sample 4 Sample 5 n 6 INCUBATION Value 2 Value 3 Value 4 Value 5 MEAN VALUE SPIKE

15 TRUENESS Codex Alimentarius: numerical values for the criteria

16 PRECISION PRECISION is related to RANDOM ERRORS Component of measurement error that - in replicate measurements - varies in an unpredictable manner Random error = error - systematic error Correction of random error can not be done Some sources of random errors: Methods (procedure, calibration,...) Laboratory (facility, environment) Equipment and materials / reagents / calibrants Personnel Time

17 PRECISION PRECISION represents random errors of a set of replicate measurements PRECISION is calculated as a (relative) standard deviation of replicate measurements σ x Less precision is reflected by a larger standard deviation Precision depends critically on the conditions! REPEATABILITY and REPRODUCIBILITY conditions are particular sets of extreme conditions....nothing to do with true or reference value!

18 PRECISION REPEATABILITY AND REPRODUCIBILITY Repeatability: a set of conditions that includes the same measurement, procedure, operators, same measuring system, operating conditions and location, and replicate measurements on the same or similar objects over a short period of time Reproducibility: a set of conditions that includes different locations, operators, measuring systems, or even methods on the same or similar objects. Intermediate precision (intra-laboratory reproducibility): the same laboratory, method, procedure but within an extended period of time - may include new calibrations, calibrants, operators, measuring systems, etc.

19 PRECISION INCREASING NUMBER OF CONSIDERED RANDOM ERROR SOURCES REPEATABILITY INTRA-LABORATORY REPEATABILITY REPRODUCIBILITY SAMPLE: SAME OPERATOR: SAME INSTRUMENT: SAME TIME PERIOD: SHORT CALIBRATION: SAME LAB: SAME SAMPLE: SAME OPERATOR: DIFFERENT INSTRUMENT: SAME / DIFF. TIME PERIOD: LONG CALIBRATION: DIFFERENT LAB: SAME SAMPLE: SAME OPERATOR: DIFFERENT INSTRUMENT: DIFFERENT TIME PERIOD: LONG CALIBRATION: DIFFERENT LAB: DIFFERENT Precision value is related to a certain analyte and concentration level

20 PRECISION VARIABILITY OF RESULTS: 2 DIFFERENT OPERATORS LU Standard FLD1 A, Ex=248, Em=374, TT (C:\DATAHP~1\FLD154-4\DATA02\MJKOLONY MePyr B[a]A Integration 1 Integration 2 FLD1 A, Ex=248, Em=374, TT (C:\DATAHP~1\FLD154-4\DATA02\MT021016\ D) LU MePyr BaA Area: Area: Area: LU FLD1 A, Ex=248, Em=374, TT (C:\DATAHP~1\FLD154-4\DATA02\MT021016\ D) MePyr Area: Area: m m m Difference between results 42 % (concentration level ppb)

21 PRECISION VARIABILITY OF RESULTS: 2 DIFFERENT MEASUREMENT SYSTEMS % Same sample, conditions, operator, Two different chromatographs of the same type

22 PRECISION INCREASING NUMBER OF CONSIDERED RANDOM ERROR SOURCES REPEATABILITY INTRA-LABORATORY REPEATABILITY REPRODUCIBILITY Repeated analyses of a sample containing analyte(s) at: level close to expected concentration in analyzed matrix level close to regulatory limit low level close to limit of quantification of the method Appropriate number of repeats: 8 15 (at least 5) Calculated as standard deviation or relative standard deviation (RSD)

23 PRECISION INCREASING NUMBER OF CONSIDERED RANDOM ERROR SOURCES REPEATABILITY INTRA-LABORATORY REPEATABILITY REPRODUCIBILITY STANDARD DEVIATION σ = 1 (x i x) 2 n x = 1 n MEAN x i RELATIVE STANDARD DEVIATION RSD (%) = σ 100 x CAN BE EASILY CALCULATED IN MS EXCEL

24 PRECISION INCREASING NUMBER OF CONSIDERED RANDOM ERROR SOURCES REPEATABILITY INTRA-LABORATORY REPEATABILITY REPRODUCIBILITY RESULT VALUES RSD (%) = σ 100 x

25 PRECISION INCREASING NUMBER OF CONSIDERED RANDOM ERROR SOURCES REPEATABILITY INTRA-LABORATORY REPEATABILITY REPRODUCIBILITY Can be estimated within an inter-laboratory study Two components: σ 2 r intra-laboratory variance (typical) σ 2 L between laboratories variance σ 2 R = σ 2 r + σ 2 L

26 PRECISION INCREASING NUMBER OF CONSIDERED RANDOM ERROR SOURCES REPEATABILITY INTRA-LABORATORY REPEATABILITY REPRODUCIBILITY Can be estimated within an inter-laboratory study however inter-laboratory study is time-demanding and costly it is problematic to find sufficient number of competent laboratories in multi residue analysis it is almost impossible to perform this kind of study for all analytes / matrices / concentration levels

27 PRECISION REPRODUCIBILITY - HORWITZ Reproducibility can be alternatively estimated from an empirical model developed based on numerous interlaboratory studies (more than 150) Horwitz (Thompson) empirical model of precision the RSD can be expressed as a function of the concentration William Horwitz

28 PRECISION REPRODUCIBILITY - HORWITZ Relative standard deviation variation coefficient: lower concentration of analyte increasing RSD nature of analyte, matrix, analytical method etc.: less important even can be ignored! RSD = 2 (1-0.5*logX) X is an analyte concentration expressed as a mass ratio

29 PRECISION REPRODUCIBILITY - HORWITZ 50 RELATIVE STANDARD DEVIATION (%) CV (%) Horwitzova kř ivka ANALYTE CONCENTRATION (ppb) ppb HORWITZ CURVE

30 PRECISION REPRODUCIBILITY - HORWITZ Official requirements on precision trace analysis: Concentration (ppb) CV r (%) Codex Alimentarius RSD R (%) Horwitz < 1 35 > > < 22

31 TRUENESS AND PRECISION = ACCURACY RELATIONSHIPS BETWEEN TYPE OF ERROR, RELATED CHARACTERISTICS AND THEIR QUANTITATIVE EXPRESSION

32 LINEARITY AND CALIBRATION LINEARITY can be tested by linear regression of the responses on the concentrations in an appropriate calibration set R 2 = 0,9977 Peak area DON concentration (µg kg -1 )

33 LINEARITY, CALIBRATION AND RANGE GENERAL RECOMENDATIONS FOR LINEAR CALIBRATION There should be six or more calibration points (standards) Even spacing over the concentration range of interest The calibration range should encompass 0 150% or % of the concentration likely to be encountered in samples Calibration standards should be run at least in duplicate in random order VALIDATION RANGE is the interval of analyte concentration within which the method can be regarded as validated Typically narrower than linear range In practice, most methods will be validated at only one or two levels of concentration. The validated range may be taken as a reasonable extrapolation from these points at concentration scale

34 LINEARITY, CALIBRATION AND RANGE VALIDATED RANGE VALIDATED CONCENTRATION LEVEL Signal VALIDATED CONCENTRATION LEVEL Linear range Concentration

35 LIMIT OF DETECTION / QUANTIFICATION

36 LIMIT OF DETECTION / QUANTIFICATION Limit of Detection (LOD): the smallest concentration of analyte in the test sample which can be reliably distinguished from zero. LOD is concentration of analyte which induce signal (S) that is 3 times higher than the background noise level (N). S/N=3 Limit of Quantification (LOQ): the smallest concentration of analyte in the test sample which can be reliably quantified. LOQ is concentration of analyte which induce signal (S) that is 10 times higher than the background noise level (N). S/N=10, LOQ usually corresponds to lowest calibration point. S/N can be usually calculated in processing software

37 LIMIT OF DETECTION / QUANTIFICATION S/N > 2 S/N > 6 signal noise

38 DETECTABILITY how to lower LOD? Increase of S/N ratio in GC/MS-NCI analysis of nitronaphtalenes (parsley sample) due to increasing electromultiplyier voltage in detector (1-nitroNAP = 0.1 ppb, 2-nitroNap = 0.08 ppb) A - EMV = 1300 V B - EMV = 1800 V 1-nNap 2-nNap A nNap 2-nNap B Time--> Time--> Vladimir.Kocourek@vscht.

39 LIMIT OF DETECTION / QUANTIFICATION HOW TO ESTIMATE LOD & LOQ SPIKING OF BLANK MATRIX WITH DECREASING AMOUNT OF ANALYTES SAMPLE PREPARATION & MEASUREMENTS CALCULATION OF S/N VALUES STANDARD SOLUTION WITH DECREASING AMOUNT OF ANALYTES MEASUREMENTS CALCULATION OF S/N VALUES LOD LOQ

40 Signal at LOD vs. background noise 1,64*s blank 1,64*s blank = 5 % = 5 % s blank y blank Response to blank Y LOD Analyte signal

41 Codex Alimentarius: numerical values for the criteria Applicability: The method has to be applicable for the specified provision, specified commodity and the specified level: maximum and/or minimum - ML. The minimum applicable range of the method depends on the specified level (ML) to be assessed, and can either be expressed in terms of the reproducibility standard deviation (s R ) or in terms of LOD and LOQ. CL 2008/7-MAS, March 2008 Minimum applicable range: ML 0.1 mg/kg [ML - 3 s R, ML + 3 s R ] ML < 0.1 mg/kg [ML - 2 s R, ML + 2 s R ] Limit of Detection (LOD): ML 0.1 mg/kg LOD ML 1/10 ML < 0.1 mg/kg LOD ML 1/5 Limit of Quantification (LOQ): ML 0.1 mg/kg LOQ ML 1/5 ML < 0.1 mg/kg LOQ ML 2/5

42 SELECTIVITY AND SPECIFITY Selectivity: the degree to which a method can quantify the analyte accurately in the presence of interferents. Selective method the results are influenced by the sample matrix (interferents, cross-reactivity, matrix effects, ) Specific methods the results are not influenced by the presence of sample matrix The presence and influence of any sample matrix interference on method results should be tested and described

43 SELECTIVITY AND SPECIFITY MATRIX EFFECTS: DECREASE (INCERASE) AND PEAK SHAPE DETERIORATION High matrix content Low matrix content

44 RUGGEDNESS Ruggedness of an analytical method is the resistance to change in results when minor deviations are made from the experimental conditions described in the SOP. The aspects of the method that are likely to affect results should be identified and described in SOP Examples of factors relevant to ruggedness: ph of a solution stability of the instrumental system extraction time concentration of (derivatization) reagents temperature/time of (derivatization) reaction time allowed for completation of whole analytical process

45 VALIDATION PROCEDURE Development and optimization of the method Spiked samples Training Reference material Repeated measurement Independent method Inter-laboratory study REPEATABILITY TRUENESS REPRODUCIBILITY LOD/LOQ Range Performance characteristics Fit-for-purpose Selectivity Ruggedness

46 USEFUL DOCUMENTS

47 USEFUL DOCUMENTS Criteria for validation of methods used in official control of contaminants and residues in food and feed:

48 All documents and methods issued to personnel in the laboratory shall be reviewed and approved for use by authorized personnel prior to issue. Documents are periodically reviewed and, where necessary, revised to ensure continuing suitability and compliance with applicable requirements; Invalid or obsolete documents are promptly removed from all points of issue or use, or otherwise assured against unintended use; Procedures shall be established to describe how changes in documents maintained in computerized systems are made and controlled. All records and data shall be stored and retained in such a way that they are readily retrievable in facilities that provide a suitable environment to prevent damage or deterioration and to prevent loss. The laboratory shall have procedures to protect and back-up records stored electronically and to prevent unauthorized access to or amendment of these records.

49 The laboratory shall retain records of primary data, calculations, calibration records, staff records and a copy of each test report issued. The records for each test shall contain information to facilitate identification of factors affecting the uncertainty and to enable the test to be repeated under conditions as close as possible to the original. The records shall include the identity of personnel responsible for the performance of each test and checking of results. When mistakes occur in records, each mistake shall be crossed out, not erased, made illegible or deleted, and the correct value entered alongside. All such alterations to records shall be signed by the person making the correction. In the case of records stored electronically, equivalent measures shall be taken to avoid loss or change of original data.

50 The laboratory shall use test methods, including methods for sampling, which meet the needs of the customer and which are appropriate for the tests it undertakes. Methods published in international, regional or national standards shall preferably be used. Appropriate methods have been also published by reputable technical organizations, or in relevant scientific texts or journals, or specified by the manufacturer of the equipment. Laboratory-developed methods or methods adopted by the laboratory may also be used if they are appropriate for the intended use and if they are validated. The laboratory shall validate nonstandard methods, laboratory developed methods, standard methods used outside their intended scope, and modifications of standard methods to confirm that the methods are fit for the intended use. Testing laboratories shall apply procedures for estimating uncertainty of measurement based on the method validation data.

51 USEFUL DOCUMENTS

52 Pesticides standards Pure standards should be of known purity and each must be uniquely identified and the date of receipt recorded. They should be stored at low temperature, preferably in a freezer, with light and moisture excluded, i.e. under conditions that minimise degradation. The identity of freshly acquired pure standards should be checked if the analytes are new to the laboratory. When preparing stock standards of pure standards of analytes and internal standards, the identity and mass of the pure standard and the identity and amount of the solvent must be recorded. The solvent(s) must be appropriate to the analyte (solubility, no reaction) and method of analysis. Moisture must be excluded during equilibration of the pure standard to room temperature before use and concentrations must be corrected for the purity of the pure standard. Not less than 10 mg of the pure standard should be weighed using a 5 decimal place balance. The ambient temperature should be that at which the glassware is calibrated, otherwise preparation of the standard should be based on solvent-mass measurement. Existing stock and working solutions may be tested against newly prepared solutions by comparing the detector responses obtained from appropriate dilutions of individual standards or mixtures of standards. The means from at least three replicate measurements for each of two solutions (old and new) should not normally differ by more than ±10 %. The mean from the new solution is taken to be 100%. Differences in apparent concentration between old and new standards must be investigated.

53 Calibration Responses used to quantify residues must be within the dynamic range of the detector. Extracts containing high-level analytes may be diluted to bring them within the calibrated range. Validation of analytical methods should include determination of recovery at the proposed reporting limit. Calibration by interpolation between two levels is acceptable providing the difference between the 2 levels is not greater than a factor of 4, and where the mean response factors, derived from replicate determinations at each level, indicate acceptable linearity of response with the higher being not more than 110 % of the lower response factor. Single-level calibration may provide more accurate results than multi-level calibration if the detector response is variable with time. When single-level calibration is employed, the sample response should be within ± 20 % of the calibration standard response if the MRL is exceeded. The potential for matrix effects to occur should be assessed at method validation. They are variable in occurrence and intensity but some techniques are particularly prone to them. If the techniques used are not inherently free from such effects, calibration should be matrix-matched routinely. Where a calibration standard is a mixture of isomers, etc., of the analyte, detector response generally may be assumed to be similar, on a molar basis, for each component.

54 Contamination & interference Samples must be separated from each other, and from other sources of potential contamination, during transit to, and storage at, the laboratory. Pest control in, or near, the laboratory must be restricted to pesticides that will not be sought as residues. Volumetric equipment, such as flasks, pipettes and syringes must be cleaned scrupulously, especially for re-use. As far as practicable, separate glassware, etc., should be allocated to standards and sample extracts, in order to avoid cross-contamination. Avoid using excessively scratched or etched glassware. Equipment, containers, solvents (including water), reagents, etc., should be checked as sources of possible interference. Rubber and plastic items (e.g. seals, protective gloves, wash bottles) and lubricants are frequent sources. Vial seals should be PTFE-lined. Extracts should be kept out of contact with seals by keeping vials upright. Vial seals may have to be replaced quickly after piercing, if re-analysis of the extracts is necessary. Analysis of reagent blanks should identify sources of interference in the equipment or materials used.

55 Validation parameters and criteria Parameter What/How Criterion Linearity Through calibration curve Residuals < ±20 % Matrix effect LOQ Specificity Accuracy Precision (RSD r ) Comparison of response from solvent standards and matrix-matched standards Definition: lowest level for which it has been demonstrated that criteria for accuracy and precision have been met Response in reagent blank and control samples Determine average recovery for both spike levels Determine repeatability RSDr, determine for both spike levels MRL < 30 % of LOQ % 20 %

56 Method validation Within-laboratory method validation should be performed to provide evidence that a method is fit for the purpose for which it is to be used. Method validation is a requirement of accreditation bodies, and must be supported and extended by method performance verification during routine analysis (analytical quality control). The method must be tested for sensitivity, mean recovery (as a measure of trueness or bias), precision, and limit of quantification (LOQ). This means that spiked recovery experiments to check the accuracy of the method should be undertaken. A minimum of 5 replicates is required at both the reporting limit, and at least another higher level, perhaps an action level, for example the MRL. The (method) LOQ is defined as the lowest validated spike level meeting the method performance acceptability criteria: mean recovery for each pesticide in the range %, with a RSD r 20%. Exceptionally, where recovery is low but consistent (demonstrating good precision) and the basis for this is well established (e.g. due to pesticide distribution in partition), a mean recovery below 70% may be acceptable.

57 Acceptability of analytical performance Acceptable limits for a single recovery result should normally be within the range of the mean recovery +/- 2x RSD). The laboratory must participate regularly in relevant proficiency tests. When a low number of compounds are analysed with respect to the pesticides present in the test sample, false positive(s) or negative(s) are reported or the accuracy achieved in any of the tests is questionable or unacceptable, the problem(s) should be investigated. Unacceptable performance, has to be rectified before proceeding with further determinations of the analyte involved. Negative results for represented analytes are supported only indirectly by the recovery and LCL data for representative analytes and must be interpreted with caution. Positive results (residues at or above the RL) usually require additional confirmation. Suspected MRL exceedances or unusual residues must be identified. The use of a highly specific detection system, such as mass spectrometry, is recommended. Selective detectors employed with GC or LC such as ECD, FPD, NPD, DAD and fluorescence, offer only limited specificity. Their use, even in combination with different polarity columns, does not provide unambiguous identification.

58 Identification - confirmation Mass spectrometry in conjunction with chromatographic separation is a very powerful combination for identification of an analyte. It simultaneously provides: i. retention time ii. ion mass/charge ratio; and iii. abundance data. The retention time (or relative retention time) of the analyte in the sample extract must match that of the calibration standard within a specified window after taking into consideration the resolving power of the chromatographic system. The relative retention time of the analyte, should correspond to that of the calibration solution with a tolerance of ± 0.5 % for GC and ± 2.5 % for LC. Reference spectra for the analyte should be generated using the instruments and techniques employed for analysis of the samples. If major differences are evident between a published spectrum and that generated within the laboratory, the latter must be shown to be valid. To avoid distortion of ion ratios, the response of the analyte ions must not overload the detector.

59 Identification - confirmation Different types and modes of mass spectrometric detectors provide different degrees of selectivity, which relates to the confidence in identification:

60 Identification - confirmation Different types and modes of mass spectrometric detectors provide different degrees of selectivity, which relates to the confidence in identification: For a higher degree of confidence in identification, further evidence may be required. This can be achieved through additional mass spectrometric information, for example evaluation of full scan spectra, additional accurate mass (fragment) ions, additional product ions (in MS/MS), or accurate mass product ions.

61 Identification points Commission decision 2002/657/EC:

62 Method validation pesticides: conditions

63 Thank you for your attention

Validation and Calibration. Definitions and Terminology

Validation and Calibration. Definitions and Terminology Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an

More information

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1) INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY

More information

Guidance for Industry

Guidance for Industry Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:

More information

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5 European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND

More information

GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005

GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005 CAC/GL 56-2005 Page 1 of 6 GUIDELINES ON THE USE OF MASS SPECTROMETRY (MS) FOR IDENTIFICATION, CONFIRMATION AND QUANTITATIVE DETERMINATION OF RESIDUES CAC/GL 56-2005 CONFIRMATORY TESTS When analyses are

More information

A commitment to quality and continuous improvement

A commitment to quality and continuous improvement Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous

More information

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1 Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification

More information

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS. GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au

More information

METHOD VALIDATION AND QUALITY CONTROL PROCEDURES FOR PESTICIDE RESIDUES ANALYSIS IN FOOD AND FEED

METHOD VALIDATION AND QUALITY CONTROL PROCEDURES FOR PESTICIDE RESIDUES ANALYSIS IN FOOD AND FEED METHD VALIDATIN AND QUALITY CNTRL PRCEDURES FR PESTICIDE RESIDUES ANALYSIS IN FD AND FEED Document No. SANC/10684/2009 Supersedes Document No. SANC/3131/2007 Implemented by 01/01/2010 Coordinator Dr. Tuija

More information

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high

More information

Ecology Quality Assurance Glossary

Ecology Quality Assurance Glossary Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability

More information

ASSURING THE QUALITY OF TEST RESULTS

ASSURING THE QUALITY OF TEST RESULTS Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office

More information

Signal, Noise, and Detection Limits in Mass Spectrometry

Signal, Noise, and Detection Limits in Mass Spectrometry Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville

More information

APPENDIX N. Data Validation Using Data Descriptors

APPENDIX N. Data Validation Using Data Descriptors APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection

More information

Pesticide Analysis by Mass Spectrometry

Pesticide Analysis by Mass Spectrometry Pesticide Analysis by Mass Spectrometry Purpose: The purpose of this assignment is to introduce concepts of mass spectrometry (MS) as they pertain to the qualitative and quantitative analysis of organochlorine

More information

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:

Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard

More information

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA

FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so

More information

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes Brussels, 18th May 2005 CED No 494 rev 2 Final TAXUD/3711/2004 POETRY:

More information

Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL)

Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) 13 October 2008 Date of application: 13 April 2009 INTRODUCTION The scope

More information

ISO 17025. How to Meet. Requirements for Method Verification. Prepared by:

ISO 17025. How to Meet. Requirements for Method Verification. Prepared by: How to Meet ISO 17025 Requirements for Method Verification Prepared by: AOAC INTERNATIONAL 481 N. Frederick Ave, Suite 500 Gaithersburg, MD 20877, USA www.aoac.org The Scientific Association Dedicated

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number

More information

Analytical Test Method Validation Report Template

Analytical Test Method Validation Report Template Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation

More information

State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist

State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist This checklist is intended to help laboratories improve the quality of their testing programs with regard to meat and poultry testing samples, in order to ensure those laboratories are at least equal to

More information

Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories

Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Wilton Park House, Wilton Place, Dublin 2, Ireland Tel +353 1 607 3003 Fax +353 1 607 3109 E-mail inab@inab.ie Web

More information

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in

More information

18.6.1 Terms concerned with internal quality control procedures

18.6.1 Terms concerned with internal quality control procedures 18.6.1 Terms concerned with internal quality control procedures Quality assurance in analytical laboratories Quality assurance is the essential organisational infrastructure that underlies all reliable

More information

Design of Experiments for Analytical Method Development and Validation

Design of Experiments for Analytical Method Development and Validation Design of Experiments for Analytical Method Development and Validation Thomas A. Little Ph.D. 2/12/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax. Standard Method for Analysis of Benzene and Toluene Content in Hydrocarbon Waxes by Headspace Gas Chromatography EWF METHOD 002/03 (Version 1 Reviewed 2015) 1 Scope 1.1 This test method covers the qualitative

More information

Quality Management System Policy Manual

Quality Management System Policy Manual Quality Management System Burns Engineering, Inc. 10201 Bren Road East Minnetonka, MN 55343 4. 4.1 General Requirements The Quality Manual consists of the quality system policy documents and the quality

More information

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination

More information

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application

More information

Applying Statistics Recommended by Regulatory Documents

Applying Statistics Recommended by Regulatory Documents Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services steven@statisticaloutsourcingservices.com 301-325 325-31293129 About the Speaker Mr. Steven

More information

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES

DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Introduction This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample

More information

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external

More information

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,

More information

LC-MS/MS, the new reference method for mycotoxin analysis

LC-MS/MS, the new reference method for mycotoxin analysis LC-MS/MS, the new reference method for mycotoxin analysis What is necessary for the implementation of the multi-residue mycotoxin method in an analytical laboratory? Azel Swemmer azel@fdalab.co.za AFMA

More information

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference Abstract The advantages of mass spectrometry (MS) in combination with gas

More information

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

General and statistical principles for certification of RM ISO Guide 35 and Guide 34 General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in

More information

Performance and advantages of qnmr measurements

Performance and advantages of qnmr measurements Return to Web Version This is the first of a series of articles related to the launch of new organic CRMs certified by quantitative NMR under double accreditation. In this issue, we focus on CRMs intended

More information

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system

More information

NIST HANDBOOK 150 CHECKLIST

NIST HANDBOOK 150 CHECKLIST NIST HANDBOOK 150 CHECKLIST Instructions to the Assessor: This checklist addresses the general accreditation criteria prescribed in NIST Handbook 150, NVLAP Procedures and General Requirements (2006 edition).

More information

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,

More information

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,

More information

Validating Methods using Waters Empower TM 2 Method. Validation. Manager

Validating Methods using Waters Empower TM 2 Method. Validation. Manager Validating Methods using Waters Empower TM 2 Method Validation Manager (MVM) Anders Janesten Nordic Informatics Sales Specialist 2008 Waters Corporation Decision-centric centric method development information

More information

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory

More information

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography INTRODUCTION The analysis of soft drinks for caffeine was able to be performed using UV-Vis. The complex sample

More information

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater

Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters

More information

Luminus Devices, Inc Luminus Testing Laboratory Quality Management Systems Manual

Luminus Devices, Inc Luminus Testing Laboratory Quality Management Systems Manual Luminus Devices, Inc Luminus Testing Laboratory Quality Management Systems Manual ISO/IEC 17025 NVLAP, NIST HANDBOOKS 150,150-1 This document belongs to Luminus Devices, Inc. It cannot be reproduced without

More information

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb

Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb Document: AND Lead 100 7 2013 Lead Testing and On Site Calibration for Water Testing Detection Range: 2 100ppb July, 2013 Edition 1 ANDalyze, Inc., 2012. All rights reserved. Printed in USA. Table of Contents

More information

Quality Assurance/Quality Control in Acid Deposition Monitoring

Quality Assurance/Quality Control in Acid Deposition Monitoring Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial

More information

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters

Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters 1. SCOPE and APPLICATION 1.1 Gravimetric analysis is used to determine total suspended

More information

AMD Analysis & Technology AG

AMD Analysis & Technology AG AMD Analysis & Technology AG Application Note 120419 Author: Karl-Heinz Maurer APCI-MS Trace Analysis of volatile organic compounds in ambient air A) Introduction Trace analysis of volatile organic compounds

More information

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in

More information

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)

More information

AUDIT PROTOCOL FOR THE VICTORIAN WATER QUALITY MONITORING NETWORK

AUDIT PROTOCOL FOR THE VICTORIAN WATER QUALITY MONITORING NETWORK AUDIT PROTOCOL FOR THE VICTORIAN WATER QUALITY MONITORING NETWORK ENVIRONMENT PROTECTION AUTHORITY June 1999 AUDIT PROTOCOL FOR THE VICTORIAN WATER QUALITY MONITORING NETWORK, June 1999 David Robinson

More information

4.2 Bias, Standards and Standardization

4.2 Bias, Standards and Standardization 4.2 Bias, Standards and Standardization bias and accuracy, estimation of bias origin of bias and the uncertainty in reference values quantifying by mass, chemical reactions, and physical methods standard

More information

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850

UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure

More information

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories

Evaluating Laboratory Data. Tom Frick Environmental Assessment Section Bureau of Laboratories Evaluating Laboratory Data Tom Frick Environmental Assessment Section Bureau of Laboratories Overview of Presentation Lab operations review QA requirements MDLs/ PQLs Why do we sample? Protect physical,

More information

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)

METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine

More information

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract

More information

Quality Assurance (QA) & Quality Control (QC)

Quality Assurance (QA) & Quality Control (QC) Quality Assurance (QA) & Quality Control (QC) In the laboratory Georges Ruta Water Quality Scientist City West Water 1 City West Water Water Supply & Sewerage Authority Melbourne Pop. >800,000 Area 580

More information

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz

More information

AOAC Guidelines for Single Laboratory Validation of Chemical Methods for Dietary Supplements and Botanicals

AOAC Guidelines for Single Laboratory Validation of Chemical Methods for Dietary Supplements and Botanicals AOAC Guidelines for Single Laboratory Validation of Chemical Methods for Dietary Supplements and Botanicals CONTENTS 1.0 Introduction 1.1 Definitions 1.1.1 Validation 1.1.2 Method of analysis 1.1.3 Performance

More information

1 Quality Assurance and Quality Control Project Plan

1 Quality Assurance and Quality Control Project Plan 1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing

More information

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) Method OIV-MA-AS313-22 Type II method Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) 1. Introduction

More information

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler

More information

Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team

Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team Control Charts and Trend Analysis for ISO 17025 Speakers: New York State Food Laboratory s Quality Assurance Team 1 ISO 17025 Requirements, 5.9.1: The laboratory shall have quality control procedures for

More information

Assessment of Accuracy and Precision

Assessment of Accuracy and Precision 2 chapter Assessment of Accuracy and Precision S.S. Nielsen, Food Analysis Laboratory Manual, Food Science Texts Series, DOI 10.1007/978-1-4419-1463-7_2, Springer Science+Business Media, LLC 2010 9 Chapter

More information

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS

Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Unique Software Tools to Enable Quick Screening and Identification of Residues and Contaminants in Food Samples using Accurate Mass LC-MS/MS Using PeakView Software with the XIC Manager to Get the Answers

More information

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management

More information

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of carbohydrates in foodstuff (LC/RID method) Exercise guarantor: Assoc. Prof. Karel Cejpek, Ph.D. CONTENT Required knowledge...

More information

ALACC Frequently Asked Questions (FAQs)

ALACC Frequently Asked Questions (FAQs) Updated October 17, 2012 ALACC Contents Web Introduction to FAQ... 2 FAQ #6... 3 ALACC Section: How to Meet ISO 17025 Requirements for Method Verification (Page 9)... 3 FAQ #13... 4 ALACC Section ALACC

More information

Assay Qualification Template for Host Cell Protein ELISA

Assay Qualification Template for Host Cell Protein ELISA Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to

More information

How to Verify Performance Specifications

How to Verify Performance Specifications How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated

More information

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays

Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Increasing the Multiplexing of High Resolution Targeted Peptide Quantification Assays Scheduled MRM HR Workflow on the TripleTOF Systems Jenny Albanese, Christie Hunter AB SCIEX, USA Targeted quantitative

More information

COMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS

COMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS COMPLIANCE BY DESIGN FOR PHARMACEUTICAL QUALITY CONTROL LABORATORIES INSIGHT FROM FDA WARNING LETTERS Primer CONTENTS INTRODUCTION...3 QUALITY AND COMPLIANCE IN QUALITY CONTROL LABORATORIES...5 Compliance

More information

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database C146-E175A Technical Report Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database Hitoshi Tsuchihashi 1 Abstract: A sample consisting

More information

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,

More information

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS. PURPOSE The purpose of this report is to verify the performance characteristics of the Sievers M-Series TOC Analyzers. Performance was quantified

More information

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview.

Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM. Technical Overview. Strategies for Developing Optimal Synchronous SIM-Scan Acquisition Methods AutoSIM/Scan Setup and Rapid SIM Technical Overview Introduction The 5975A and B series mass selective detectors (MSDs) provide

More information

Assay Development and Method Validation Essentials

Assay Development and Method Validation Essentials Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 drlittle@dr-tom.com

More information

Proficiency Testing In A Global Analytical Environment Benefits & Challenges. AOAC September 2007

Proficiency Testing In A Global Analytical Environment Benefits & Challenges. AOAC September 2007 Proficiency Testing In A Global Analytical Environment Benefits & Challenges AOAC September 2007 1 The Coca-Cola Company The Coca-Cola Company is the largest manufacturer, distributor and marketer of nonalcoholic

More information

Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity

Samples RMN070714. Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity For the benefit of VaporFi, Inc. Samples RMN070714 Analytical Report (0714-63) GC/FID Analysis (RM01) Nicotine Purity GC/MS Analysis (QG02) Anabasine, Anatabine, Myosmine, Nornicotine Enthalpy Analytical,

More information

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon

More information

1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.

1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting. Document #: FDPD-QMS.024.003 Page 1 of 12 Table of Contents 1. Purpose 2. Scope 3. Responsibility 4. References 5. Related Documents 6. Definitions 7. Safety 8. Equipment/Materials Needed 9. Process Description

More information

Quantifying Uncertainty in Analytical Measurement

Quantifying Uncertainty in Analytical Measurement EURACHEM / CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition QUAM:000.P1 EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition Editors S L R Ellison

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 204 3R HANDLING AND USE OF REFERENCE STANDARDS IN THE OMCL NETWORK Full document title and reference Document type Legislative

More information

To avoid potential inspection

To avoid potential inspection Good Analytical Method Validation Practice Deriving Acceptance for the AMV Protocol: Part II To avoid potential inspection observations for test method validations by the regulatory agencies, it has become

More information

Alignment and Preprocessing for Data Analysis

Alignment and Preprocessing for Data Analysis Alignment and Preprocessing for Data Analysis Preprocessing tools for chromatography Basics of alignment GC FID (D) data and issues PCA F Ratios GC MS (D) data and issues PCA F Ratios PARAFAC Piecewise

More information

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for

More information

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection 52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

Mass Spectrometry Signal Calibration for Protein Quantitation

Mass Spectrometry Signal Calibration for Protein Quantitation Cambridge Isotope Laboratories, Inc. www.isotope.com Proteomics Mass Spectrometry Signal Calibration for Protein Quantitation Michael J. MacCoss, PhD Associate Professor of Genome Sciences University of

More information

Sample Analysis Design Isotope Dilution

Sample Analysis Design Isotope Dilution Isotope Dilution Most accurate and precise calibration method available Requires analyte with two stable isotopes Monoisotopic elements cannot be determined via isotope dilution Spike natural sample with

More information

Medical Cannabis Laboratory Approval Program

Medical Cannabis Laboratory Approval Program Medical Cannabis Laboratory Approval Program Application Process and Required Documentation After the publication of the Medical Cannabis Laboratory Application, currently expected by February 16, 2015,

More information

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation

International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. ludwig_huber@labcompliance.com Overview GMP requirements for Quality Control laboratories

More information

The Theory of HPLC. Gradient HPLC

The Theory of HPLC. Gradient HPLC The Theory of HPLC Gradient HPLC i Wherever you see this symbol, it is important to access the on-line course as there is interactive material that cannot be fully shown in this reference manual. Aims

More information

Interlaboratory studies

Interlaboratory studies Interlaboratory studies Department of Food Chemistry and Analysis, ICT Prague Vladimír Kocourek Prague, 2012 Interlaboratory studies Various titles: Interlaboratory proficiency test or studies Interlaboratory

More information

Introduction. The following definitions may help you better understand the components of the data report.

Introduction. The following definitions may help you better understand the components of the data report. Introduction This document is designed to help our clients understand the quality control requirements and limitations of data reporting. There are three sections to this document. The first section will

More information

Vitamin C quantification using reversed-phase ion-pairing HPLC

Vitamin C quantification using reversed-phase ion-pairing HPLC Vitamin C quantification using reversed-phase ion-pairing HPLC Thomas Grindberg and Kristy Williams Department of Chemistry, Concordia College, 901 8 th St S, Moorhead, MN 56562 Abstract Vitamin C, an

More information