THE EU PEAK FLOW METER

Size: px
Start display at page:

Download "THE EU PEAK FLOW METER"

Transcription

1 01 THE EU PEAK FLOW METER 04 THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS

2 THE EU PEAK FLOW METER 01 THE EU PEAK FLOW METER: HISTORICAL PERSPECTIVE, WHY CHANGE AND WHAT S AVAILABLE M.R. Miller Correspondence Dr Martin R Miller University Hospital Birmingham NHS Trust Selly Oak Hospital Birmingham B29 6JD UK martin.miller@uhb.nhs.uk The measurement of peak expiratory flow (PEF) was first introduced about 50 years ago [1, 2] as a simple and convenient measure to aid the diagnosis and management of patients with airflow obstruction. Its clinical use has waxed and waned in the intervening years and now its role in asthma management is better defined. The purpose of this article is not to argue the evidence concerning the use of PEF monitoring in the management of asthma [3], but to discuss how the measurement of PEF has changed recently and which types of meter may best suit certain applications. After covering the background to where PEF meters are today, the article will give an understanding of the characteristics of various types of meter for recording PEF, followed by a discussion concerning the clinical settings where PEF monitoring is used, so that the best match of device to purpose can be made. Background The first PEF meters [1, 2] represented a tremendous step forward in the ability to record aspects of lung function in a simple and repeatable way in order to help manage common lung disorders. The original calibration of these devices was undertaken to the best degree available at the time. Subsequent work found that when systems were available that could generate accurate and repeatable known flows, the readings from these meters were inaccurate in the midrange, with a significant overreading [4, 5]. While some believed this error was relatively unimportant for a device used to look at changes in PEF within individuals, in fact the measurement of within-subject PEF variability a hallmark of asthma is importantly distorted by this error such that clinical decisions based on PEF monitoring had in the past underestimated asthma severity [6]. New standards for accuracy were set [7] and have now been refined in a European Commission standard [8] that means all PEF meters sold in European Union countries now meet accuracy requirements and no longer need their results correcting for the known error in older instruments. Furthermore, the new standard covers aspects THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS 05

3 01 THE EU PEAK FLOW METER Figure 1. Some of the devices available. (Left to right and top to bottom: MicroLife, Personal Best, PiKo-1, Pocket Peak, Microplus, Pulmolife, Micro, Mechanical Peak Flow Meters, Asma-1). concerning frequency response (vide infra) that previously had been ignored. PEF meter characteristics For a PEF meter to function correctly, it must be accurate within the required flow range, it should offer a suitably low resistance to flow and it must be able to record the types of flow input signals that patients produce, i.e. the frequency response characteristics must be adequate for purpose. In addition, the meter must be reliable and appropriately robust for its purpose. There are now many types of meter available that are of small size and practical to use (fig. 1). These use a range of transduction systems to convert the flow of gas through the device into a signal that can be measured. As the performance characteristics of meters are dependent on the type of transduction system, this aspect will be discussed first, followed by the performance requirements. Transduction systems For a PEF device to work, the rapid flow of gas through the instrument has to be converted into a signal that can be measured. The methods most commonly used in devices available today are described below. Variable-orifice devices Figure 2 shows the design principle for these devices. At the start, a 6 THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS

4 THE EU PEAK FLOW METER 01 PRODUCT LISTINGS EU Standard EN13826 Designed specifically to meet EN13826 scale standard Movable asthma management colour zones that won t wash off In-built internal check valve, avoiding cross-infection concerns Available with either European or ATS scale Micro Medical Ltd PO Box 6, Rochester, Kent ME1 2AZ, United Kingdom Tel +44 (0) Fax +44 (0) MicroPeak Peak Flow Meter Micro DiaryCard Spirometer The new standard in Peak Flow measurement A new powerful and flexible replacement for paper diaries and mechanical PEF meters Designed specifically for European standard (EN13826) High visibility EU scale ATS scale version also available Wash-proof Colour-Zone asthma management system Built-in check valve Fully customisable for promotional purposes Easily cleaned Measures and stores PEF, FEV 1 and FVC, Flow/Volume curves and symptom responses All data electronically date and time stamped Easily configured for clinical trial protocols Simple to use for patients at home Secure audit path for regulatory clinical trials Phone: +44 (0) Phone:+44 (0) THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS 7

5 01 THE EU PEAK FLOW METER Flow Figure 2. Variable-orifice device such as the mini-wright meter. Flow Flow P1 > P2 Figure 3. The Venturi or fixed-orifice principle. Figure 4. The stator/rotor device. a) b) P1 > P2 Figure 5. a) Fleisch and b) Lilly pneumotachographs. P1 > P2 baffle occludes the orifice of the mouthpiece. When the subject blows through the meter, the baffle moves forward, uncoiling a spring. As the baffle progressively moves up the meter it exposes more and more of an exhaust orifice (shaded grey in the figure). The area of the orifice exposed is proportional to the peak flow. Some devices have a bending piece of metal (Pocket Peak), rather than a baffle attached to a coiled spring, that opens up the area of the orifice (mini-wright, Micro Peak, Vitalograph mechanical meter). The reading from these devices is dependent on gas density and so readings are affected by marked changes in atmospheric pressure such as at high altitude. The usual ambient weather variations in barometric pressure and temperature are not important in this respect. Fixed-orifice or Venturi devices Figure 3 shows the principles for devices where the pressure drop across a fixed orifice is measured. This drop is proportional to the square of the flow. The flow is turbulent and so the signal is dependent on gas density (i.e. altitude). These devices require pressure transducers to be very accurate in the low-pressure range, as the signal is very small at low flows, such as when recording at the end of the manoeuvre. The nonlinearity of these devices can be corrected by electronics holding an array of calibration factors for the range of flows. Stator/rotor devices Figure 4 shows the principle of devices where the flow of gas is turned into a vortex, usually by passing it through a fixed set of angled blades. This vortex then rotates a finely balanced vane that is very light with low inertia and set with precision bearings. The movement of the vane can be measured by its rorotation, which causes the path of a visible light or infra-red beam to be interrupted. These interruptions are counted and the frequency of the counts is related to flow. These devices have a stable calibration and the output is not altered by changes in gas composition which makes them suitable for recordings at altitude. The engineering of the bearings is critical to prevent the vane from either stopping too early or overrunning, either of which would affect the registration of low flows. Pneumotachographs Pneumotachographs were the first truly accurate flow meters to be used in respiratory practice. They are essentially of two designs: the Fleisch flow head (fig. 5a) and the Lilly mesh screen (fig. 5b). The pressure drop recorded is dependent on the viscosity of the gas passing through and not its density, so the signal is not altered by altitude but is sensitive to temperature and condensation building up in the flow head. The Lilly has less thermal mass but is intrinsically not quite as linear as the Fleisch [9]. The latter has considerable thermal mass and the flow head can change in temperature over a period of testing, which will influence the results. To avoid condensation and keep the flow-sensing element at a stable temperature, it is best practice to place these devices in the flow stream of a fan between blows [10]. Heating of the flow head can be undertaken to help prevent condensation and maintain thermal stability, but the heating controller must be extremely sophisticated to avoid temperature changes [11], so using flow heads unheated when recording single expiratory manoeuvres is more accurate. These devices need accurate and regular calibration and so are essentially only suitable for a laboratory, workplace or primary care setting when used by technical staff adequately trained in their maintenance. 8 THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS

6 THE EU PEAK FLOW METER 01 Table 1. A list of manufacturers, their devices with characteristics and estimated cost. Manufacturer / Supplier Website Device Transduction PEF only Price Logging Clement Clarke mini Wright variable orifice yes 15 no International airzone variable orifice yes 15 no Edinburgh Way mini Wright digital fixed orifice no 35 yes Harlow one flow tester fixed orifice no 535 yes Essex CM20 2TT, UK Micro Medical MicroPeak variable orifice yes 10 no PO BOX 6 PulmoLife stator/rotor no 165 no Rochester Micro Spirometers stator/rotor no 335 yes Kent. ME1 2AZ United Kingdom (part of VIASys Healthcare Inc.) nspire Health Pocket Peak variable orifice yes 25 no Louisville, Colorado, PiKo-1 fixed orifice no 25 yes USA KoKo fixed orifice no 25 yes (formerly Ferraris Respiratory) ndd Medizintechnik Easyone ultrasonic no 1875 Technoparkstrasse 1 CH-8005 Zürich, Switzerland Respironics peakflowfamily.respironics.com Assess variable orifice yes 15 no 1010 Murry Ridge Lane AsthmaCheck/Mentor variable orifice yes 15 no Murrysville, Personal best variable orifice yes 15 no PA , USA Vitalograph mechanical PEF meters variable orifice yes 10 no Maids Moreton, Asma-1 stator/rotor no 28 yes Buckingham Vitalograph Micro pneumotacho- no 500 yes MK18 1SW, UK graph Table 2.. Clinical settings for recording PEF and instrument requirements Setting Accident and emergency, outpatient dept, ward Home, workplace Lung function laboratory, primary care Requirements to consider Low cost, robust, multi-patient use, fixed calibration Low cost, robust, fixed calibration, data logging Absolute accuracy, multi-patient use Flow Figure 6. The ultrasonic flow meter. Ultrasonic Figure 6 shows the design for devices where an ultrasonic beam is passed diagonally across a flow of gas and the transit of the beam is affected by the gas flow according to the Doppler effect. By sending beams in both directions, various corrections can be made to render this an extremely accurate and low-resistance device. The calibration is stable and the output is affected by properties that determine the speed of sound, such as gas density. The crosssectional area of the beam needs to be adequately matched to the diameter of the tube through which the gas flow profile is passing. For example, measuring the Doppler effect on a very narrow beam passing across the flow in a much wider-bore tube would not give a signal representative of the whole of the flow profile. These devices are technically sophisticated and offer promise for many applications. Their calibration is stable, which has advantages for non-laboratory applications. Accuracy All PEF meters in the EU must meet the CE standard PrEN13826 [8], which means the accuracy of readings from handheld PEF meters will now agree with the standards met by larger laboratory-based equipment. In addition, the meters have to meet a drop test requirement to show they are suitably robust for patient selfadministered use. PEF meters in use in the USA do not have to meet THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS 9

7 01 THE EU PEAK FLOW METER PRODUCT LISTINGS PiKo Monitors Mini-Wright Digital nspire Health introduces PiKo Lung Health Monitors, the world s smallest electronic PEF, FEV1, and FEV6 meters. All PiKo monitors provide automatic test quality alerts, electronic data storage of 96 patient tests, single button review, and offer optional PiKoNET software for extended tracking and trending. PiKo monitors are downloadable to a personal computer. Revolutionary in their design, these miniaturized devices exceed all published accuracy standards, and remarkably, is priced at a similar cost to mechanical peak flow meters. For more information call +44 (0) or visit THE NAME SPEAKS VOLUMES It s Intelligent. It s Simple. It s Accurate. It s Compact. It s Easy to Clean. It s a Mini-Wright. Not only the name speaks volumes, so does the product. Accurate PEF and FEV1 measurements for treating both asthma and COPD patients at the touch of a button. Easily customised, intelligent patient monitoring in the palm of your hand. A water tight seal for simple sterilisation required under industry standards. And a name synonymous with accuracy, reliability and performance. Visit to find out more about the digital peak flow meter that speaks volumes or for the distributor nearest you. Phone: +44 (0) Phone: Telehealthcare SIBEL DATOSPIR PEAK-10 A modern medicine technique with definitive solutions Remote Patient Monitoring (RPM) Effective monitoring of chronic disease Effective therapy control Cost efficient Remote diagnostic possibilities Selective emergency management The DATOSPIR PEAK-10 peak flow meter provides an objective measurement of your Peak expiratory Flow (PEF) and is valuable information of your lung functions. It includes a colour zone asthma management. The correct use of this instrument will help you and your doctor to follow your condition closely and will enable him/her to provide the most beneficial treatment for you. It is important to keep a record of your peak flow. Readings should be taken twice a day; first thing in the morning and at bed time, or as directed by your physician. You should also measure and note your peak flow if you suddenly become wheezy or breathless at any time of the day or night. The peak flow meter DATOSPIR PEAK-10 has been designed and manufactured according to the Medical Device Directive 93/42/CEE. This device meets new EN scale standard. This device has been verified and calibrated at SIBEL, S.A., according to the verification and adjustment procedure. Phone: +49 (0) Phone: THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS

8 THE EU PEAK FLOW METER 01 this standard but are tested against a less demanding standard from the American Thoracic Society [7]. The readings from old devices based on the original Wright meter scale [1, 2] can be clinically misleading [6], depending on where across their range the reading is being made [4, 5]. The CE standard has addressed these accuracy issues and also addressed the frequency characteristics of these devices for the first time. It had been noted by paediatric clinicians that astute children could generate falsely high PEF readings by tongueing the delivery of the blow a bit like spitting into the meter which in under-damped meters would give a high flow. This new aspect of PEF meter testing found that some meters available in the USA were severely under-damped [12], that is they over-read with flow profiles that have a rapid rise up to and fall off from the peak. These meters do not meet the new EC standard, which is the only standard to test this aspect of performance. Calibration Most hand held PEF meters do not need day-to-day calibration. This is ideal for unsupervised patient use of the meter. The more sophisticated flow measuring devices common in lung function laboratories do need frequent calibration, since their transduction systems for converting flow into a measurement signal are usually not so stable and are influenced by ambient conditions. Although devices are marketed as not requiring calibration, it is possible with misuse or accident that the device no longer reads correctly. Most manufacturers state that if a device gives surprising readings then a calibration check procedure is required [13]. For handheld devices that are only giving a flow output, such as variable orifice meters, a calibration check is beyond the capability of all but the most sophisticated laboratories, since a pump system is needed that can deliver calibrated flows [14]. Human calibration can be used where, for instance, lung function laboratory staff who are free of lung disease regularly check their own PEF on a meter. This can soon detect an unexpected trend in the output of the device. If the device records volume as well as flow, then discharging a 3-L syringe several times through the device using a range of flows can give an indication if there is a significant error in the device. Resistance to flow PEF meters offer a complex impedance to flow, which will be referred to here as a resistance. When compared with variableorifice meters, some devices, such as pneumotachographs or ultrasonic devices, offer very little resistance. This can be easily appreciated when blowing through one such device and then through a variable-orifice meter. Standards for PEF meters state what is the acceptable limit for resistance [8, 13]. Turbine devices offer greater resistance at higher flows, while for variable orifice meters it is the reverse [15]. A highresistance meter has been found to reduce the achieved PEF by an average of 8% compared with PEF measured with a low-resistance device, but the PEF readings were also less variable [16], which may be an advantage in some settings. Reliability The variable-orifice meters such as the original mini-wright meter have working parts that have been found to work correctly for up to 15 years of continuous patient use without loss of accuracy [17]. However, if a PEF meter is mistreated then it can fail to work properly even when new. Thus instruction to the user about how to use and care for the meter is a key part of the clinician s role in suggesting the patient uses THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS 11

9 01 THE EU PEAK FLOW METER a PEF meter to help manage their asthma. This is equally true for healthcare professionals in accident and emergency depts or in primary care who use PEF meters. Cleaning of PEF devices can cause more harm than benefit and so the manufacturer s instructions must be followed. Disassembling a device can lead to damage to important key components, such as the spring in the mini-wright meter and other variable-orifice meters, which then render all subsequent readings invalid. Many devices are affected by excess heat and so leaving a device close to a heater may distort its component parts, rendering it useless. Environmental conditions Most PEF meters are used at home, at work or in a hospital setting. The ambient conditions in these settings do not usually present a problem. If accurate PEF recordings are needed at altitude, then turbine devices or pneumotachographs would be the best choice, as their readings are not affected by the changes in gas density at altitude [14]. Clinical settings Accident and emergency depts and outpatient depts In this setting, the instrument needs to be readily available and suitable for multi-patient use. Devices through which the patient cannot inhale can be used for different patients without the necessity for ensuring cleaning of the insides of the instrument itself. One-way mouthpieces can also be used to facilitate the use of one instrument for difference patients. Ideally this environment also requires rugged, robust and calibration-free instruments. In-patient wards Ward patients with asthma will be recording their PEF for several days while they recover and can individually use a single instrument for this time. A number of patients will take this instrument home on discharge to facilitate the monitoring of their condition. The environment is otherwise as demanding as the accident and emergency dept. Cleaning of such meters should be undertaken, especially for high-risk patients such as those with cystic fibrosis and tuberculosis. Suggestions have been made about how this can be done for simple mechanical variable orifice meters [18]. The manufacturer s instructions must be followed to ensure the PEF readings remain accurate after cleaning. Home, school, work These settings require the PEF meter to be used by the patient without supervision. Thus the instrument has to be reliable, robust and calibration free. Most devices for patient use fulfil these requirements. However, the results from unsupervised recordings can be less than accurately registered for a number of reasons [19, 20]. Electronic data logging meters may then be necessary to obtain faithful results. Another consideration is that if the readings are to be taken in open social environments, the instruments should not be awkward to use. For children the device may need to look cool to be acceptable and so increase the likelihood it will be used appropriately. Primary care Increasingly, it is hoped that satisfactory lung function measurements will be made in the primary care setting rather than in a hospital laboratory. There are tensions between the technical requirements for using instruments for diagnosis, which require that their calibration and results are accurate, and the simplicity of cheaper calibration-free instruments. In primary care other lung function indices, such as forced expiratory volume in one second (FEV1) may 12 THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS

10 THE EU PEAK FLOW METER 01 also be required from the instrument in order to facilitate making a diagnosis in chronic obstructive pulmonary disease or asthma. When measurements are made at a single interaction that then lead to a diagnosis being made, there is a necessity for the measurements to be undertaken with the highest technical standards being achieved [13]. Laboratory In a pulmonary function laboratory, PEF is often recorded incidentally at the same time as other spirometric measures. Evidence suggests that it is essential that the device must be recording flow as the primary measure in order to get an accurate representation of PEF. Differentiation of volume measurements with respect to time to obtain flow accentuates any noise on the signal and complex smoothing algorithms have been recommended [7, 21] to obtain an estimate of PEF from volume measuring equipment. This practice is not recommended since the frequency response characteristics to record PEF [12] are beyond that which volume measuring equipment can readily achieve. Table 1 includes a nonexhaustive list of manufacturers and their devices with some essential characteristics that can help determine the best choice for the patient or setting. Table 2 outlines the various clinical settings for recording PEF and the likely characteristics for choice of device. Conclusion PEF meters are useful in a wide variety of settings. From its origins in small, cheap, handheld PEF meters the recording of PEF has moved on to now include equally small and useable devices which are more sophisticated, with additional features such as recording FEV1 and data logging for subsequent download into a computer. It is remarkable and a credit to engineering that for extremely low cost there are very accurate and reliable devices available. With the information contained in this article, it should be possible to choose the best device for the required clinical application. REFERENCES 1. Higgins ITT. Respiratory symptoms, bronchitis and ventilatory capacity in a random sample of agricultural population. BMJ 1957; 2: Wright BM, McKerrow CB. Maximum forced expiratory flow rate as a measure of ventilatory capacity with a description of a new portable instrument measuring it. BMJ 1959; 2: The BTS/SIGN British Guideline on the management of asthma. Thorax 2003; 58: Supplement 1. Updated Nov 2005 from: /asthmaupdatenov05.pdf. Date last accessed: June 14, Miller MR, Dickinson SA, Hitchings DJ. The accuracy of portable peak flow meters. Thorax 1992; 47: Gardner RM, Crapo RO, Jackson BR, Jensen RL. Evaluation of accuracy and reproducibility of peak flow meters at 1400 m. Chest 1992; 101: Miles JF, Tunnicliffe W, Cayton RM, Ayres JG, Miller MR. Potential effects of correction of inaccuracies of the mini- Wright peak expiratory flow meter on the use of an asthma self-management plan. Thorax 1995; 51: American Thoracic Society. Standardization of spirometry update. Am J Respir Crit Care Med 1995; 152: European Standard pren Peak Flow Meters. London, British Standards Institute, Fry DL, Hyatt RE, McCall B, Mallos AJ. Evaluation of three types of respiratory flowmeters. J Appl Physiol 1957; 10: Miller MR, Sigsgaard T. Prevention of thermal and condensation errors in pneumotachographic recordings of the maximal forced expiratory manoeuvre. Eur Respir J 1994; 7: Miller MR, Pincock AC. Linearity and temperature control of the Fleisch pneumotachograph. J Appl Physiol 1986; 60: Miller MR, Atkins PR, Pedersen OF. Inadequate peak expiratory flow meter characteristics detected by a computerised explosive decompression device. Thorax 2003; 58: Miller MR, Hankinson J, Brusasco V, et al. Standardisation of spirometry. Eur Respir J 2005; 26: Miller MR, Jones B, Xu Y, Pedersen OF, Quanjer PH. Peak expiratory flow profiles delivered by pump systems: limitations due to wave action. Am J Respir Crit Care Med 2000; 161: Pedersen OF, Miller MR, Sigsgaard T, Tidley M, Harding RM. Portable peak flow meters: physical characteristics, influence of temperature, altitude, and humidity. Eur Respir J 1994; 7: Pedersen OF, Rasmussen TR, Omland Ø, Sigsgaard T, Quanjer PH, Miller MR. Eur Respir J 1996; 9: Miles JF, Bright P, Ayres JG, Cayton RM, Miller MR. The performance of Mini Wright peak flow meters after prolonged use. Respir Med 1995; 89: Bell J. Peak flow meter manufacturer s comment single patient use? Prim Care Respir J 2002; 11: Verschelden P, Cartier A, L Archevêque J, Trudeau C, Malo JL. Compliance with and accuracy of daily selfassessment of peak expiratory flows (PEF) in asthmatic subjects over a three month period. Eur Respir J 1996; 9: Malo JL, Trudeau C, Ghezzo H, L Archevêque J, Cartier A. Do subjects investigated for occupational asthma through serial peak expiratory flow measurements falsify their results? J Allergy Clin Immunol 1995; 96: American Thoracic Society. Standardization of spirometry: 1987 update. Am Rev Respir Dis 1987; 136: THE BUYERS GUIDE TO RESPIRATORY CARE PRODUCTS 13

11

Forced vital capacity: maximal volume of air exhaled with maximally forced effort from a maximal inspiration.

Forced vital capacity: maximal volume of air exhaled with maximally forced effort from a maximal inspiration. SOP Spirometry 1. General considerations Spirometry serves as a physiological test to quantify pulmonary disease severity and to assess clinical change in respiratory function over time. Standard spirometric

More information

Vitalograph Products and Services Guide

Vitalograph Products and Services Guide Vitalograph Products and Services Guide Vitalograph is a world leading provider of outstanding quality cardio-respiratory devices, clinical trial services and medical equipment servicing. With a pioneering

More information

Vmax Vyntus SPIRO. Advance your lung function testing experience

Vmax Vyntus SPIRO. Advance your lung function testing experience Vmax Vyntus SPIRO Advance your lung function testing experience The Vmax Vyntus SPIRO with SentrySuite software is designed to advance the lung function testing experience for both the patient and the

More information

SPIROMETRY FOR HEALTH CARE PROVIDERS Global Initiative for Chronic Obstructive Lung Disease (GOLD)

SPIROMETRY FOR HEALTH CARE PROVIDERS Global Initiative for Chronic Obstructive Lung Disease (GOLD) SPIROMETRY FOR HEALTH CARE PROVIDERS Global Initiative for Chronic Obstructive Lung Disease (GOLD) CONTENTS I. INTRODUCTION II. BACKGROUND INFORMATION A. What Is Spirometry? B. Why Perform Spirometry?

More information

PULMONARY FUNCTION TESTS A Workshop on Simple Spirometry & Flow Volume Loops

PULMONARY FUNCTION TESTS A Workshop on Simple Spirometry & Flow Volume Loops PULMONARY FUNCTION TESTS A Workshop on Simple Spirometry & Flow Volume Loops YOU SHOULD READ THE FOLLOWING MATERIAL BEFORE Tuesday March 30 Interpretation of PFTs Learning Objectives 1. Specify the indications

More information

microlife PF 100 Electronic Asthma Monitor Instruction Manual (1-15) Moniteur d'asthme électronique Mode d'emploi (16-31) Monitor electrónico para el

microlife PF 100 Electronic Asthma Monitor Instruction Manual (1-15) Moniteur d'asthme électronique Mode d'emploi (16-31) Monitor electrónico para el microlife PF 100 Electronic Asthma Monitor Instruction Manual (1-15) Moniteur d'asthme électronique Mode d'emploi (16-31) Monitor electrónico para el asma Manual de instrucciones (32-47) Elektronischer

More information

Airways Resistance and Airflow through the Tracheobronchial Tree

Airways Resistance and Airflow through the Tracheobronchial Tree Airways Resistance and Airflow through the Tracheobronchial Tree Lecturer: Sally Osborne, Ph.D. Department of Cellular & Physiological Sciences Email: sosborne@interchange.ubc.ca Useful links: www.sallyosborne.com

More information

MicroLab. The world s most versatile. portable Spirometer

MicroLab. The world s most versatile. portable Spirometer MicroLab The world s most versatile portable Spirometer MicroLab A new generation Spirometer designed for Child incentive animatio the professional Developed for the professional, the new generation MicroLab

More information

Accurate Air Flow Measurement in Electronics Cooling

Accurate Air Flow Measurement in Electronics Cooling Accurate Air Flow Measurement in Electronics Cooling Joachim Preiss, Raouf Ismail Cambridge AccuSense, Inc. E-mail: info@degreec.com Air is the most commonly used medium to remove heat from electronics

More information

Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers

Laminar and Turbulent flow. Flow Sensors. Reynolds Number. Thermal flow Sensor. Flow and Flow rate. R = Mass Flow controllers Flow and Flow rate. Laminar and Turbulent flow Laminar flow: smooth, orderly and regular Mechanical sensors have inertia, which can integrate out small variations due to turbulence Turbulent flow: chaotic

More information

A Telehomecare monitoring project in Cystic Fibrosis

A Telehomecare monitoring project in Cystic Fibrosis ICEHTM 2011 1st International Conference on E-HEALTH and TELEMEDICINE 10-12 October 2011 Nicosia - North Cyprus A Telehomecare monitoring project in Cystic Fibrosis Murgia F, Alghisi F, Majo F, Montemitro

More information

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc.

Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc. ASGMT / Averaging Pitot Tube Flow Measurement Michael Montgomery Marketing Product Manager Rosemount Inc. Russ Evans Manager of Engineering and Design Rosemount Inc. Averaging Pitot Tube Meters Introduction

More information

GCE AS/A level 1661/01A APPLIED SCIENCE UNIT 1. Pre-release Article for Examination in January 2010 JD*(A09-1661-01A)

GCE AS/A level 1661/01A APPLIED SCIENCE UNIT 1. Pre-release Article for Examination in January 2010 JD*(A09-1661-01A) GCE AS/A level 1661/01A APPLIED SCIENCE UNIT 1 Pre-release Article for Examination in January 2010 JD*(A09-1661-01A) 2 BLANK PAGE 3 Information for Teachers The attached article on asthma is based on some

More information

Grant Agreement No. 228296 SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme

Grant Agreement No. 228296 SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme Grant Agreement No. 228296 SFERA Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME Capacities Specific Programme Research Infrastructures Integrating Activity - Combination of

More information

Virginia Tech Departmental Policy 27 Sports Medicine Key Function:

Virginia Tech Departmental Policy 27 Sports Medicine Key Function: Virginia Tech Departmental Policy 27 Sports Medicine Key Function: Review: Yearly Director of Athletic Training Title: Management of Asthma in Athletes Section: Treatment S-A Safety POLICY STATEMENT: This

More information

Spirodoc. One Touch Laboratory for respiratory analysis two functional modes: doctor and patient. Spirometer with Touch Screen display

Spirodoc. One Touch Laboratory for respiratory analysis two functional modes: doctor and patient. Spirometer with Touch Screen display MEDICAL INTERNATIONAL RESEARCH Spirodoc One Touch Laboratory for respiratory analysis two functional modes: doctor and patient Spirometer with Touch Screen display Pulse Oximeter intelligent with on screen

More information

10EaZy SW White Paper Choosing the right hardware

10EaZy SW White Paper Choosing the right hardware 10EaZy SW White Paper Choosing the right hardware Selecting proper hardware for the 10EaZy SW This white paper describes important elements to consider before purchasing hardware intended for sound level

More information

GRADE 11F: Biology 3. UNIT 11FB.3 9 hours. Human gas exchange system and health. Resources. About this unit. Previous learning.

GRADE 11F: Biology 3. UNIT 11FB.3 9 hours. Human gas exchange system and health. Resources. About this unit. Previous learning. GRADE 11F: Biology 3 Human gas exchange system and health UNIT 11FB.3 9 hours About this unit This unit is the third of six units on biology for Grade 11 foundation. The unit is designed to guide your

More information

The photoionization detector (PID) utilizes ultraviolet

The photoionization detector (PID) utilizes ultraviolet Chapter 6 Photoionization Detectors The photoionization detector (PID) utilizes ultraviolet light to ionize gas molecules, and is commonly employed in the detection of volatile organic compounds (VOCs).

More information

Lothian Guideline for Domiciliary Oxygen Therapy Service for COPD

Lothian Guideline for Domiciliary Oxygen Therapy Service for COPD Lothian Guideline for Domiciliary Oxygen Therapy Service for COPD This document describes the standard for clinical assessment, prescription, optimal management and follow-up of patients receiving domiciliary

More information

Medicare Pulmonary Rehabilitation (PR) Benefit Frequently Asked Questions June 2010 (Latest Updates: December 18, 2013 and February 12, 2014)

Medicare Pulmonary Rehabilitation (PR) Benefit Frequently Asked Questions June 2010 (Latest Updates: December 18, 2013 and February 12, 2014) Medicare Pulmonary Rehabilitation (PR) Benefit Frequently Asked Questions June 2010 (Latest Updates: December 18, 2013 and February 12, 2014) Coverage Criteria Q. CMS has stated that only patients with

More information

Aeration Air & Digester Gas Flow Metering Using Thermal Mass Technology. HWEA 2011 Conference Craig S. Johnson

Aeration Air & Digester Gas Flow Metering Using Thermal Mass Technology. HWEA 2011 Conference Craig S. Johnson Aeration Air & Digester Gas Flow Metering Using Thermal Mass Technology HWEA 2011 Conference Craig S. Johnson Presentation Overview Introduction Aeration Air & Digester gas challenges Gas flow metering

More information

Data Bulletin. Mounting Variable Frequency Drives in Electrical Enclosures Thermal Concerns OVERVIEW WHY VARIABLE FREQUENCY DRIVES THERMAL MANAGEMENT?

Data Bulletin. Mounting Variable Frequency Drives in Electrical Enclosures Thermal Concerns OVERVIEW WHY VARIABLE FREQUENCY DRIVES THERMAL MANAGEMENT? Data Bulletin April 2001 Raleigh, NC, USA Mounting Variable Frequency Drives in Electrical Enclosures Thermal Concerns OVERVIEW Variable frequency drives are available from manufacturers as enclosed engineered

More information

Entrained Gas Diagnostic with Intelligent Differential Pressure Transmitter

Entrained Gas Diagnostic with Intelligent Differential Pressure Transmitter January Page Entrained Gas Diagnostic with Intelligent Differential Pressure Transmitter Dave Wehrs - Director, Pressure Engineering Andrew Klosinski - Application Engineer, Pressure Diagnostics Emerson

More information

A simple solution for your complex patients. The market-leading servo ventilation device System One BiPAP autosv

A simple solution for your complex patients. The market-leading servo ventilation device System One BiPAP autosv A simple solution for your complex patients The market-leading servo ventilation device System One BiPAP autosv Advanced simplifies treating complex sleep-disordered breathing patients Developed for your

More information

THE THERMAL FLOW METER, A GAS METER FOR ENERGY MEASUREMENT

THE THERMAL FLOW METER, A GAS METER FOR ENERGY MEASUREMENT THE THERMAL FLOW METER, A GAS METER FOR ENERGY MEASUREMENT Kazuto Otakane, Tokyo Gas Co., Ltd Katsuhito Sakai, Tokyo Gas Co., Ltd Minoru Seto, Tokyo Gas Co., Ltd 1. INTRODUCTION Tokyo Gas s new gas meter,

More information

Using meters to measure steam flow

Using meters to measure steam flow Page 1 of 6 April 1998 Plant Engineering SELECTING NEW TECHNOLOGY: PROBLEM SOLVING Using meters to measure steam flow Jesse L. Yoder, Ph.D., Senior Analyst, Automation Research Corp., Dedham, MA Key Concepts

More information

Heating Water by Direct Steam Injection

Heating Water by Direct Steam Injection Heating Water by Direct Steam Injection Producing hot water by direct steam injection provides a solution where large volumes of hot water at precise temperatures are required, and where energy and space

More information

Energy and Flow Measurement for Hydronic Systems

Energy and Flow Measurement for Hydronic Systems Energy and Flow Measurement for Hydronic Systems Presented By: George Szyszko Applications Consultant MCR for ONICON Incorporated Why Measure Thermal Energy? To control something, you must first measure

More information

Variable Capacity Compressors, a new dimension for refrigeration engineers to explore

Variable Capacity Compressors, a new dimension for refrigeration engineers to explore Variable Capacity Compressors, a new dimension for refrigeration engineers to explore By: Marcos G. Schwarz, VCC Group Leader Corporate Research & Development, EMBRACO SA Abstract: This paper presents

More information

1993-2013 Twenty years of innovative solutions in Spirometry, Oximetry, Telemedicine

1993-2013 Twenty years of innovative solutions in Spirometry, Oximetry, Telemedicine 1993-2013 Twenty years of innovative solutions in Spirometry, Oximetry, Telemedicine Established in 1993 MIR has become a landmark for innovation in portable diagnostic devices. High quality and ease of

More information

Flow Measurement Options for Pipeline and Open Channel Flow

Flow Measurement Options for Pipeline and Open Channel Flow Flow Measurement Options for Pipeline and Open Channel Flow October 2013 Presented by Molly Skorpik - 2013 Montana Association of Dam and Canal Systems Conference Irrigation Training and Research Center

More information

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS Selection and use of Ultrasonic Ceramic Transducers The purpose of this application note is to aid the user in the selection and application of the Ultrasonic

More information

Spirodoc. 3D Laboratory for respiratory analysis two functional modes: doctor and patient. Four devices in one. Spirometer with Touch Screen display

Spirodoc. 3D Laboratory for respiratory analysis two functional modes: doctor and patient. Four devices in one. Spirometer with Touch Screen display Spirodoc 3D Laboratory for respiratory analysis two functional modes: doctor and patient Four devices in one Spirometer with Touch Screen display Pulse Oximeter intelligent with on screen results 3D Accelerometer

More information

Why and How we Use Capacity Control

Why and How we Use Capacity Control Why and How we Use Capacity Control On refrigeration and air conditioning applications where the load may vary over a wide range, due to lighting, occupancy, product loading, ambient weather variations,

More information

An Overview of Asthma - Diagnosis and Treatment

An Overview of Asthma - Diagnosis and Treatment An Overview of Asthma - Diagnosis and Treatment Asthma is a common chronic disorder of the airways that is complex and characterized by variable and recurring symptoms, airflow obstruction, bronchial hyperresponsiveness,

More information

Chapter 10. Flow Rate. Flow Rate. Flow Measurements. The velocity of the flow is described at any

Chapter 10. Flow Rate. Flow Rate. Flow Measurements. The velocity of the flow is described at any Chapter 10 Flow Measurements Material from Theory and Design for Mechanical Measurements; Figliola, Third Edition Flow Rate Flow rate can be expressed in terms of volume flow rate (volume/time) or mass

More information

Pulmonary Rehabilitation in Chronic Obstructive Pulmonary Disease (COPD)

Pulmonary Rehabilitation in Chronic Obstructive Pulmonary Disease (COPD) Pulmonary Rehabilitation in Chronic Obstructive Pulmonary Disease (COPD) Development of disability in COPD The decline in airway function may initially go unnoticed as people adapt their lives to avoid

More information

Relative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration

Relative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration Relative Humidity Measurement and Calibration The widest range of humidity products from a pioneer of RH sensor development The best in Relative Humidity technology to optimize your process Whether you

More information

DTU-52 Series. High-Power Digital TV Transmitter. Features

DTU-52 Series. High-Power Digital TV Transmitter. Features DTU-52 Series High-Power Digital TV Transmitter Features Small footprint - 590 (W) x 2,000 (H) x 1,000 (D)mm for 8.7kW High efficiency - Excellent broadband matching circuit at amplifier modules New TRPA

More information

How to measure absolute pressure using piezoresistive sensing elements

How to measure absolute pressure using piezoresistive sensing elements In sensor technology several different methods are used to measure pressure. It is usually differentiated between the measurement of relative, differential, and absolute pressure. The following article

More information

Spirometry Testing in Occupational Health Programs. Best Practices for Healthcare Professionals

Spirometry Testing in Occupational Health Programs. Best Practices for Healthcare Professionals Spirometry Testing in Occupational Health Programs Best Practices for Healthcare Professionals OSHA 3637-03 2013 Occupational Safety and Health Act of 1970 To assure safe and healthful working conditions

More information

Flow System Calibration in Medical and Bio-medical Production Processes.

Flow System Calibration in Medical and Bio-medical Production Processes. Flow System Calibration in Medical and Bio-medical Production Processes. Speaker / Author: Mark Finch Fluke Calibration Norfolk, Norwich, NR6 6JB, UK e-mail: mark.finch@fluke.com Phone: + 44 (0)1908 678

More information

Advances in Thermal Dispersion Mass Flow Meter Accuracy

Advances in Thermal Dispersion Mass Flow Meter Accuracy Advances in Thermal Dispersion Mass Flow Meter Accuracy By Dan McQueen President Fluid Components International (FCI) Advances in Thermal Dispersion Mass Flow Meter Accuracy By Dan McQueen President Fluid

More information

Fiscal Measurement Natural Gas

Fiscal Measurement Natural Gas White Paper FSG-WP-0012 February 2014 Fiscal Measurement Natural Gas Fiscal Measurement How much? Fiscal Measurement must not be confused with Custody Transfer; in fact, fiscal measurement is a more general

More information

Intelligence down to the last detail.

Intelligence down to the last detail. Intelligence down to the last detail. We question what is already there. That is how we create something new. Those who have developed high-quality products are faced with two options: to declare themselves

More information

Cooking at the Speed of light!

Cooking at the Speed of light! Cooking at the Infrared Cooking & Colouring Infrabaker is a modular infrared continuous cooking system developed by Infrabaker International. The machine is designed to cook and/or put colour on a wide

More information

PTE Pediatric Asthma Metrics Reporting Updated January 2015

PTE Pediatric Asthma Metrics Reporting Updated January 2015 PTE Pediatric Asthma Metrics Reporting Updated January 20 Introduction: The Maine Health Management Coalition s (MHMC) Pathways to Excellence (PTE) Program is preparing for its next round of PTE Pediatric

More information

How to Choose the Right Air Compressor

How to Choose the Right Air Compressor How to Choose the Right Air Compressor Air compressors have been around for well over 100 years and have as many uses as there are tools that use air. One reason for their popularity is because air as

More information

COPD. Information brochure for chronic obstructive pulmonary disease.

COPD. Information brochure for chronic obstructive pulmonary disease. COPD Information brochure for chronic obstructive pulmonary disease. CONTENTS What does COPD mean?...04 What are the symptoms of COPD?...06 What causes COPD?...09 Treating COPD...10 Valve therapy in COPD...12

More information

COPD and Asthma Differential Diagnosis

COPD and Asthma Differential Diagnosis COPD and Asthma Differential Diagnosis Chronic Obstructive Pulmonary Disease (COPD) is the third leading cause of death in America. Learning Objectives Use tools to effectively diagnose chronic obstructive

More information

COMBIMASS. Technical Data COMBIMASS eco-bio +

COMBIMASS. Technical Data COMBIMASS eco-bio + COMBIMASS Technical Data THE SYSTEM COMBIMASS The field transmitters of the COMBIMASS eco series are suitable for gas flow measurement and cover a wide range of different applications. The instruments

More information

Spirometry Workshop for Primary Care Nurse Practitioners

Spirometry Workshop for Primary Care Nurse Practitioners Spirometry Workshop for Primary Care Nurse Practitioners Catherine Casey S. Jones PhD, RN, AE-C, ANP-C Certified Adult Nurse Practitioner Texas Pulmonary & Critical Care Consultants P.A. and Visiting Assistant

More information

Various Technics of Liquids and Solids Level Measurements. (Part 3)

Various Technics of Liquids and Solids Level Measurements. (Part 3) (Part 3) In part one of this series of articles, level measurement using a floating system was discusses and the instruments were recommended for each application. In the second part of these articles,

More information

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions

The Versatile Differential Pressure Transmitter. By David Gunn Honeywell Process Solutions The Versatile Differential Pressure Transmitter By David Gunn Honeywell Process Solutions The Versatile Differential Pressure Transmitter 2 Table of Contents Abstract... 3 Pressure Fundamentals... 3 Applications...

More information

Gas Custody Transfer Calibration

Gas Custody Transfer Calibration Gas Custody Transfer Calibration Using multi variable temperature / pressure calibrators for flowmeter calibration 2013 Introduction Gas custody transfer flow computers require special calibration to perform

More information

Respiratory Clinical Trial Services. the difference is clear. Your respiratory clinical trials partner

Respiratory Clinical Trial Services. the difference is clear. Your respiratory clinical trials partner Respiratory Clinical Trial Services the difference is clear Your respiratory clinical trials partner Our heritage We at Vitalograph are proud of our 50 year heritage as a world leading respiratory specialist.

More information

Service catalog Flow products

Service catalog Flow products Catalog CG/FLOW/001 EN Rev. B catalog Flow products Whatever your need our team will keep your measurement and plant performing Measurement made easy Introduction When it comes to ensuring your flowmeters

More information

Averaging Pitot Tubes; Fact and Fiction

Averaging Pitot Tubes; Fact and Fiction Averaging Pitot Tubes; Fact and Fiction Abstract An experimental investigation has been undertaken to elucidate effects of averaging stagnation pressures on estimated velocities for pressure averaging

More information

Flow Measurement with Respironics Flow Sensors

Flow Measurement with Respironics Flow Sensors W H I T E P A P E R Flow Measurement with Respironics Flow Sensors ABSTRACT Flow measurement issues with mechanically ventilated patients are reviewed with particular emphasis on the Respironics flow sensors.

More information

Everything you need for protection scheme testing

Everything you need for protection scheme testing Power System Simulator for Testing Protection Relays and Schemes Everything you need for protection scheme testing The is the only instrument with the high power, flexibility and software to perform full

More information

Liquid Flowmeters A Guide for Selecting a Flowmeter for Pressurized Systems

Liquid Flowmeters A Guide for Selecting a Flowmeter for Pressurized Systems Liquid Flowmeters A Guide for Selecting a Flowmeter for Pressurized Systems Selecting a meter requires a first-rate technical assessment of the pressurized system and a thorough evaluation of the capabilities

More information

Product brochure. GMS600 monitoring system for generator circuit-breaker State-of-the-art monitoring

Product brochure. GMS600 monitoring system for generator circuit-breaker State-of-the-art monitoring Product brochure GMS600 monitoring system for generator circuit-breaker State-of-the-art monitoring GMS600: One step ahead The ABB group is a leader in power and automation technologies. The company operates

More information

Fundamentals of Mass Flow Control

Fundamentals of Mass Flow Control Fundamentals of Mass Flow Control Critical Terminology and Operation Principles for Gas and Liquid MFCs A mass flow controller (MFC) is a closed-loop device that sets, measures, and controls the flow of

More information

PSAC/SAIT Well Testing Training Program Learning Outcomes/Objectives Level B

PSAC/SAIT Well Testing Training Program Learning Outcomes/Objectives Level B s/objectives Level B CHEM 6014: Properties and Characteristics of Natural Gases Discuss the pertinent properties and uses of natural gases as they occur in gas processing. 1. Describe the physical properties

More information

Severe asthma Definition, epidemiology and risk factors. Mina Gaga Athens Chest Hospital

Severe asthma Definition, epidemiology and risk factors. Mina Gaga Athens Chest Hospital Severe asthma Definition, epidemiology and risk factors Mina Gaga Athens Chest Hospital Difficult asthma Defined as asthma, poorly controlled in terms of chronic symptoms, with episodic exacerbations,

More information

EPM3. Phase Sequence and Motor Rotation Tester. Users Manual

EPM3. Phase Sequence and Motor Rotation Tester. Users Manual EPM3 Phase Sequence and Motor Rotation Tester Users Manual 1 L1 L2 L3 2 3 A B C CAT 600V 3-PHASE TESTER 5 TEST M1 M2 BATT M3 EPM3 MOTOR ROTATION TESTER EPM3 MOTOR ROTATION DETERMINED WHILE FACING MOTOR

More information

Medication and Devices for Chronic Obstructive Pulmonary Disease (COPD)

Medication and Devices for Chronic Obstructive Pulmonary Disease (COPD) Medication and Devices for Chronic Obstructive Pulmonary Disease (COPD) Patients with COPD take a wide variety of medicines to manage their symptoms these include: Inhaled Short Acting Bronchodilators

More information

MAXIMUM HEAT LOAD TEMPERATURE TESTING ( Differential Temperature Testing )

MAXIMUM HEAT LOAD TEMPERATURE TESTING ( Differential Temperature Testing ) MAXIMUM HEAT LOAD TEMPERATURE TESTING ( Differential Temperature Testing ) The Concept Maximum Heat Load Temperature Testing is a powerful air-conditioning diagnostic and evaluation technique. It is also

More information

A Guide to Performing Quality Assured Diagnostic Spirometry

A Guide to Performing Quality Assured Diagnostic Spirometry A Guide to Performing Quality Assured Diagnostic Spirometry PCC Contents 1. Introduction 4 2. Performing Spirometry 6 3. Interpreting and Reporting of Results 10 4. TOP TEN TIPS for reporting spirometry

More information

Centrifugal Fans and Pumps are sized to meet the maximum

Centrifugal Fans and Pumps are sized to meet the maximum Fans and Pumps are sized to meet the maximum flow rate required by the system. System conditions frequently require reducing the flow rate. Throttling and bypass devices dampers and valves are installed

More information

The Effect of Forced Air Cooling on Heat Sink Thermal Ratings

The Effect of Forced Air Cooling on Heat Sink Thermal Ratings zpero 1 The Effect of Forced Air Cooling on Heat Sink Thermal Ratings By Paul Bachman, Fellow Engineer & Ronnie Haiduk, Applications Engineer, Crydom, Inc. ABSTRACT A heat sink s ability to dissipate thermal

More information

SPIROMETRY IN PRACTICE

SPIROMETRY IN PRACTICE Spirometry_168x244_stg5 14/4/05 12:48 PM Page 2 SPIROMETRY IN PRACTICE A PRACTICAL GUIDE TO USING SPIROMETRY IN PRIMARY CARE Second Edition Spirometry_168x244_stg5 14/4/05 12:48 PM Page 3 Written by Dr

More information

Coding Guidelines for Certain Respiratory Care Services July 2014

Coding Guidelines for Certain Respiratory Care Services July 2014 Coding Guidelines for Certain Respiratory Care Services Overview From time to time the AARC receives inquiries about respiratory-related coding and coverage issues through its Help Line or Coding Listserv.

More information

Printing and Publishing Energy Savings Guide

Printing and Publishing Energy Savings Guide Printing and Publishing Energy Savings Guide Oregon printing and publishing plants face challenges of rising operating costs, environmental and other regulations, outdated equipment and customer demand

More information

Chapter 5. Components, Symbols, and Circuitry of Air-Conditioning Wiring Diagrams

Chapter 5. Components, Symbols, and Circuitry of Air-Conditioning Wiring Diagrams Chapter 5 Components, Symbols, and Circuitry of Air-Conditioning Wiring Diagrams Objectives Upon completion of this course, you will be able to: Explain what electrical loads are and their general purpose

More information

Application Note. So You Need to Measure Some Inductors?

Application Note. So You Need to Measure Some Inductors? So You Need to Measure Some nductors? Take a look at the 1910 nductance Analyzer. Although specifically designed for production testing of inductors and coils, in addition to measuring inductance (L),

More information

Interpretation of Pulmonary Function Tests

Interpretation of Pulmonary Function Tests Interpretation of Pulmonary Function Tests Dr. Sally Osborne Cellular & Physiological Sciences University of British Columbia Room 3602, D.H Copp building 604 822-3421 sally.osborne@ubc.ca www.sallyosborne.com

More information

RESULTS OF ICARUS 9 EXPERIMENTS RUN AT IMRA EUROPE

RESULTS OF ICARUS 9 EXPERIMENTS RUN AT IMRA EUROPE Roulette, T., J. Roulette, and S. Pons. Results of ICARUS 9 Experiments Run at IMRA Europe. in Sixth International Conference on Cold Fusion, Progress in New Hydrogen Energy. 1996. Lake Toya, Hokkaido,

More information

THE HUMIDITY/MOISTURE HANDBOOK

THE HUMIDITY/MOISTURE HANDBOOK THE HUMIDITY/MOISTURE HANDBOOK Table of Contents Introduction... 3 Relative Humidity... 3 Partial Pressure... 4 Saturation Pressure (Ps)... 5 Other Absolute Moisture Scales... 8 % Moisture by Volume (%M

More information

RESPIRATORY VENTILATION Page 1

RESPIRATORY VENTILATION Page 1 Page 1 VENTILATION PARAMETERS A. Lung Volumes 1. Basic volumes: elements a. Tidal Volume (V T, TV): volume of gas exchanged each breath; can change as ventilation pattern changes b. Inspiratory Reserve

More information

E190Q Lecture 5 Autonomous Robot Navigation

E190Q Lecture 5 Autonomous Robot Navigation E190Q Lecture 5 Autonomous Robot Navigation Instructor: Chris Clark Semester: Spring 2014 1 Figures courtesy of Siegwart & Nourbakhsh Control Structures Planning Based Control Prior Knowledge Operator

More information

Tests. Pulmonary Functions

Tests. Pulmonary Functions Pulmonary Functions Tests Static lung functions volumes Dynamic lung functions volume and velocity Dynamic Tests Velocity dependent on Airway resistance Resistance of lung tissue to change in shape Dynamic

More information

CMS National Coverage Policy

CMS National Coverage Policy LCD ID Number L32764 LCD Title Pulmonary Rehabilitation (PR) Programs Contractor s Determination Number L32764 AMA CPT/ADA CDT Copyright Statement CPT only copyright 2002-2011 American Medical Association.

More information

Oxygen Therapy. A guide for the patient. Please bring to any outpatient appointments. Patients name. Exceptional healthcare, personally delivered

Oxygen Therapy. A guide for the patient. Please bring to any outpatient appointments. Patients name. Exceptional healthcare, personally delivered A guide for the patient Please bring to any outpatient appointments Patients name Exceptional healthcare, personally delivered Introduction This leaflet has been written for patients starting, or who have

More information

STEAM HEATING SYSTEM TROUBLESHOOTING GUIDE

STEAM HEATING SYSTEM TROUBLESHOOTING GUIDE Page 1 of 9 PURPOSE Steam is the most commonly used heating medium for maintaining process temperatures. Compared to other heating media, steam contains a significant amount of heat energy, and this heat

More information

Chapter 3.5: Fans and Blowers

Chapter 3.5: Fans and Blowers Part I: Objective type questions and answers Chapter 3.5: Fans and Blowers 1. The parameter used by ASME to define fans, blowers and compressors is a) Fan ration b) Specific ratio c) Blade ratio d) Twist

More information

A descriptive definition of valve actuators

A descriptive definition of valve actuators A descriptive definition of valve actuators Abstract A valve actuator is any device that utilizes a source of power to operate a valve. This source of power can be a human being working a manual gearbox

More information

Environment-Laboratory Ambient Conditions

Environment-Laboratory Ambient Conditions ISSUE 3.0: MADE MINOR CLARIFICATION CHANGES. FOR CLIENT LABS Purpose To assure laboratory conditions are stable during the performance of testing and data collection. Why this requirement is important?

More information

MATLAB AS A PROTOTYPING TOOL FOR HYDRONIC NETWORKS BALANCING

MATLAB AS A PROTOTYPING TOOL FOR HYDRONIC NETWORKS BALANCING MATLAB AS A PROTOTYPING TOOL FOR HYDRONIC NETWORKS BALANCING J. Pekař, P. Trnka, V. Havlena* Abstract The objective of this note is to describe the prototyping stage of development of a system that is

More information

Ventilation Perfusion Relationships

Ventilation Perfusion Relationships Ventilation Perfusion Relationships VENTILATION PERFUSION RATIO Ideally, each alveolus in the lungs would receive the same amount of ventilation and pulmonary capillary blood flow (perfusion). In reality,

More information

Instructions Manual. Electromagnetic sensor Series FLOMID FX. instrumentation for fluids. R-MI-FlomidFX Rev.: 0 English version

Instructions Manual. Electromagnetic sensor Series FLOMID FX. instrumentation for fluids. R-MI-FlomidFX Rev.: 0 English version instrumentation for fluids Electromagnetic sensor Series FLOMID FX Instructions Manual! Conforms with the Pressure Equipment Directive 97/23/EC. 0830 This equipment is considered as being a pressure accessory

More information

ADVANCED CONTROL TECHNIQUE OF CENTRIFUGAL COMPRESSOR FOR COMPLEX GAS COMPRESSION PROCESSES

ADVANCED CONTROL TECHNIQUE OF CENTRIFUGAL COMPRESSOR FOR COMPLEX GAS COMPRESSION PROCESSES ADVANCED CONTROL TECHNIQUE OF CENTRIFUGAL COMPRESSOR FOR COMPLEX GAS COMPRESSION PROCESSES by Kazuhiro Takeda Research Manager, Research and Development Center and Kengo Hirano Instrument and Control Engineer,

More information

Telehealth. Putting the patient at the heart of the journey

Telehealth. Putting the patient at the heart of the journey Telehealth Putting the patient at the heart of the journey Why telehealth? 1 Telehealth is the remote monitoring of a patient s vital signs and symptoms in their own home proven to enhance the quality

More information

Welcome to this presentation on LED System Design, part of OSRAM Opto Semiconductors LED 101 series.

Welcome to this presentation on LED System Design, part of OSRAM Opto Semiconductors LED 101 series. Welcome to this presentation on LED System Design, part of OSRAM Opto Semiconductors LED 101 series. 1 To discuss the design challenges of LED systems we look at the individual system components. A basic

More information

DEVELOPMENT OF HIGH SPEED RESPONSE LAMINAR FLOW METER FOR AIR CONDITIONING

DEVELOPMENT OF HIGH SPEED RESPONSE LAMINAR FLOW METER FOR AIR CONDITIONING DEVELOPMENT OF HIGH SPEED RESPONSE LAMINAR FLOW METER FOR AIR CONDITIONING Toshiharu Kagawa 1, Yukako Saisu 2, Riki Nishimura 3 and Chongho Youn 4 ABSTRACT In this paper, we developed a new laminar flow

More information

Fläkt Woods Building Future

Fläkt Woods Building Future Energy, Expertise and Environment with Fläkt Woods Fläkt Woods Building Future Saving energy in building ventilation You must save 20% on energy before 2020. Following the Kyoto protocol the European Union

More information

Gas Flow Measurement Software A Report for

Gas Flow Measurement Software A Report for Gas Flow Measurement Software A Report for National Measurement System Directorate Department of Trade & Industry 151 Buckingham Palace Road London, SW1W 9SS Report No: 332/99 Date: September 1999 The

More information

Application Requirement

Application Requirement Hazardous Gas Monitors Sensor Selection Overview Safety Monitoring (LEL) Toxic Limit Detection (PEL) Leak Detection Personal Safety Application Requirement Exposure Assessment (TWA) Ambient Air Quality

More information

Custody Transfer Measurement. with the V-Cone Flowmeter

Custody Transfer Measurement. with the V-Cone Flowmeter Custody Transfer Measurement with the V-Cone Flowmeter Stephen A. Ifft McCrometer Inc. Hemet, California, USA Abstract This paper will discuss the approval of the McCrometer V-Cone flowmeter for custody

More information