2 FLOW CYTOMETRY: PRINCIPLES AND APPLICATIONS By: Douaa Moh. Sayed
3 Definition Flow cytometry is a technique for counting, examining, and sorting microscopic particles suspended in a stream of fluid. It allows simultaneous multiparametric analysis of the physical and/or chemical characteristics of single cells flowing through an optical and/or electronic detection apparatus
4 BASIC PRINCIPLES Flow cytometry involves the analysis of the fluorescence and light scatter properties of single particles (e.g. cells, nuclei, chromosomes) during their passage within a narrow, precisely defined liquid stream.
5 Flow Cytometry - Principles Target: Evaluation of single cell Solution: hydrodynamic Focusing
6
7 When conditions are right, sample fluid flows in a central core that does not mix with sheath fluid. This termed Laminar flow.
8 Lasers Produce a single wavelength of high intensity Nearly all instruments fitted with an argon laser Gives blue light at 488 nm. Second or third laser may be fitted: - He-Ne- red light at 633 nm - He-Cd- UV at 325 nm
9 Light Scatter Forward angle light scatter: in a narrow angle from the direction of the laser beam (FALS or FS). Forward Scatter (FSC) < 10 detection Laser ~ Cell size
10 Light Scatter Side Scatter (SSC) 90 deflection ~ Cell structures Laser Right angle light scatter at right angles to the laser beam (RALS or SS).
11 Light Scatter FS tends to be more sensitive to the size and surface properties can be used to distinguish live from dead cells. SS tends to be more sensitive to inclusions within cells can be used to distinguish granulated cells from non-granulated
12 Scatter Diagram: Morphology
13 Instrument Parameters Side Scatter (SSC) 90 deflection ~ Cell structures Forward Scatter (FSC) < 10 detection Laser Y ~ Cell size Fluorescence Intensity Antigen Density
14 Fluorochromes Dye Excitation Emission Molecular Weight FITC 488 nm 520 nm 389 Da PE 488 nm 578 nm Da ECD 488 nm 613 nm Da PC5 488 nm 668 nm da PerCP 488 nm 688 nm Da APC 613 nm 665 nm Da
15 Common Fluorochromes for 488nm Excitation LASER 488 FITC 520 PE 575 ECD PI PC5 665 <390 λ λ λ λ λ λ λ >750 λ ultra- violet blue green yellow orange red infraviolet red FITC PE ECD PI PC5 = Fluorescein Isothiocyanate = Phycoerythrin (RD1) = Energy Coupled Dye = Propidium Iodide = Phycoerythrin Cyanin 5 (PC5)
16
17 Laminar Fluidic Sheath Core Sheath FITC FL PE FL Outer Sheath 488nm Sct
18 Each cell generates a quanta of fluorescence Photomultiplier Tubes (PMT s) PE FL FITC FL 488nm Sct Discriminating Filters Dichroic Lenses Confocal Lens Forward Light Scattering Detector
19 Multiparametric Analysis Simultaneous Detection of up to 20 Parameters Y Cell size Cell structure Dyes Time (follow enzyme reactions)
20 Flow Cytometry - Principles A typical flow cytometer consists of: Light source: argon laser 488 Fluidics Detectors/Computer system for analysis and storage of digitized data.
21 Amplification The electrical pulses originating from light detected by the PMTs are then processed by a series of linear and log amplifiers. Logarithmic amplification is most often used to measure fluorescence in cells. This type of amplification expands the scale for weak signals and compresses the scale for strong or specific fluorescence signals.
22 Data display Univariate histogram Dot plot (Bivariate display)
23 Data display Density plot Contour plot
24 Surface Plot The surface Plotmay be custom oriented smoothed rescaled Zoomed in and out by using the scroll button
25 Tomogram The Tomogram may be: rotated viewed in color precedence or density mode
26 Light Scattering, 2 Parameter Histogram Bigger 90 degree Light Scatter Y Axis Apoptotic Cells Dead Cells Bigger Cells More Granular X Axis Forward Light Scatter (FLS) Live Cells
27 1 Parameter Histogram Negative Positive Count Dimmer Brighter Channel Number Fluorescence picked up from the FITC PMT
28 Statistics Signal Analysis % Negative & % Positive Cells Antigen density ~ Mean Fluorescence Intensity ~ Channel number % negative cells mean fluorescence intensity % positive cells 0 Channel Number 1024
29 Single Positive PI Population 2 Parameter Histogram Double Positive Population PE FL Negative Population FITC FL Single Positive FITC Population
30 Compensation Spectral overlap of emission spectra s of different dyes
31 Compensation Simple rule for Adjustment of Color Compensation B1 B2 Compensation: FL1 versus FL2 Fluorescence 2 B3 B4 Mean-Channel Region B3 = B4 Compensation: FL2 versus FL1 Fluorescence 1 Mean-Channel Region B1 = B3
32 Compensation
33 Compensation
34 Compensation
35 Compensation
36 Quality Control - CDC-Panel Internal Checks Internal Consistency X % CD3 + cells if all tubes +/- 3 % Lymphocytes sums (Purity?) % CD3 + cells + % CD19 + cells + % CD16 + /CD56 + cells n lymphocyte gate = 100 % (+/- 5%) % CD3 + /CD4 + cells + % CD3 + /CD8 + cells = % CD3 + cells
37
38 Gating Set a region on a histogram or cytogram If cell in region then show another property cell selection
39 Gating-Option ungated Dual Parameter Correlation: FSC = Cell size SSC = Cell structure gated Analysis of Cell Sub-populations i.e. Lymphocytes
40 Flow Cytometry Data Smaller Region, Live cells mostly Larger Region includes all cells
41 Cell sorting Flow cytometry can be used to select and purify a specific subset of cells within a population cell sorting based on physical, biochemical and antigenic traits.
42 Flow Cytometry and sorting
43 Measurable parameters volume and morphological complexity of cells cell pigments such as chlorophyll or phycoerythrin DNA (cell cycle analysis, cell kinetics, proliferation etc.) RNA chromosome analysis and sorting (library construction, chromosome paint) protein expression and localization transgenic products in vivo, particularly the Green fluorescent protein or related fluorescent proteins cell surface antigens (Cluster of differentiation (CD) markers) intracellular antigens (various cytokines, secondary mediators etc.) nuclear antigens
44 Measurable parameters enzymatic activity ph, intracellular ionized calcium, magnesium, membrane potential membrane fluidity apoptosis (quantification, measurement of DNA degradation, mitochondrial membrane potential, permeability changes, caspase activity) cell viability monitoring electropermeabilization of cells oxidative burst characterising multidrug resistance (MDR) in cancer cells glutathione various combinations (DNA/surface antigens etc.)
45 Flow Cytometry Applications molecular biology, The specific antibodies bind to antigens on the target cells and help to give information on specific characteristics of the cells pathology, immunology, plant biology and marine biology: the auto-fluorescent properties of photosynthetic plankton can be exploited by flow cytometry in order to characterise abundance and community structure. protein engineering: flow cytometry is used in conjunction with yeast display and bacterial display to identify cell surface-displayed protein variants with desired properties.
46 Clinical applications in flow cytometry Immunophenotyping of leukemia and lymphoma DNA & cell cycle analysis Minimal residual disease PNH-diagnostics RNA content (reticulocytes) Immune-deficiencies
47 Clinical applications in flow cytometry Platelet Function Analysis: Platelet associated immunoglobulins platelet-associated IgG quantitation for the diagnosis of immune thrombocytopenias platelet cross-matching in transfusion reticulated platelet assay to detect stress platelets fibrinogen receptor occupancy studies for monitoring the clinical efficacy of plateletdirected anticoagulation in thrombosis detection of activated platelet surface markers, cytoplasmic calcium ion measurements, and platelet microparticles for the assessment of hypercoagulable states.
48 Clinical applications in flow cytometry Measurement of the Efficacy of Cancer Chemotherapy Multi-Drug-Resistance assays of proliferative survival using bromodeoxyuridine (BrdU) incorporation Ligand, antigen, or molecule-targeted biological therapy utilizing monoclonal antibodies
49 Clinical applications in flow cytometry Cell Function Analysis every event that occurs during the process of lymphocyte activation can be measured by flow cytometry [tyrosine phosphorylation (multiplex bead technology), calcium flux, oxidative metabolism, neoantigen expression (CD11b/CD18 and CD154), and cellular proliferation) Applications in Transfusion Medicine Fetal red cells in maternal blood Quality control of blood products
50 Clinical applications in flow cytometry Organ Transplantation and Hematopoietic Cell Therapy HLA-typing and cross- matching, enumeration of CD34+ hematopoietic stem cells pre-transplantation determinations of the efficacy of ex vivo T-cell graft depletion, post-transplantation evaluation of immune recovery, graft rejection, graft-versus host disease, and the graft-versus-leukemia effect.
51 Clinical applications in flow cytometry Applications in Microbiology detection of bacteria, fungi, parasites and viruses quantitative procedures to assess antimicrobial susceptibility and drug cytotoxicity sperm sorting in IVF Apoptosis PI incorporation TdT Assay Annexin V APO2.7
52
53 Advantages high speed analysis (> s -1 ) Measures single cells Measures large number of cells simultaneous analysis of multiple parameters (up to 20) Identifies small subpopulations quantification of fluorescence intensities sorting of predefined cell populations (up to s -1
54 Disadvantages very expensive and sophisticated instruments Need single particle Tissue architecture is lost Little information about intra-cellular distributions
55 THANKS
56 THANKS
57 THANK YOU FOR YOUR ATTENTION
Introduction to flow cytometry
Introduction to flow cytometry Flow cytometry is a popular laser-based technology. Discover more with our introduction to flow cytometry. Flow cytometry is now a widely used method for analyzing the expression
More informationKatharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service
Introduction into Flow Cytometry Katharina Lückerath (AG Dr. Martin Zörnig) adapted from Dr. Jörg Hildmann BD Biosciences,Customer Service How does a FACS look like? FACSCalibur FACScan What is Flow Cytometry?
More informationMEASURABLE PARAMETERS: Flow cytometers are capable of measuring a variety of cellular characteristics such as:
INTRODUCTION Flow Cytometry involves the use of a beam of laser light projected through a liquid stream that contains cells, or other particles, which when struck by the focused light give out signals
More informationIntroduction to Flow Cytometry
Outline Introduction to Flow Cytometry Basic Concept of Flow Cytometry Introduction to Instrument Subsystems Daisy Kuo Assistant Product Manager E-mail: daisy_kuo@bd.com BDBiosciences Application Examples
More informationIntroduction to Flow Cytometry
Introduction to Flow Cytometry presented by: Flow Cytometry y Core Facility Biomedical Instrumentation Center Uniformed Services University Topics Covered in this Lecture What is flow cytometry? Flow cytometer
More informationUses of Flow Cytometry
Uses of Flow Cytometry 1. Multicolour analysis... 2 2. Cell Cycle and Proliferation... 3 a. Analysis of Cellular DNA Content... 4 b. Cell Proliferation Assays... 5 3. Immunology... 6 4. Apoptosis... 7
More informationFlow Cytometry. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size. Flow = Fluid Cyto = Cell Metry = Measurement
What is Flow Cytometry? Flow Cytometry Basic Principle and Applications BD Biosciences Daisy Kuo (daisy_kuo@bd.com) Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements are made on cells,
More informationFlow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy.
Flow Cytometry for Everyone Else Susan McQuiston, J.D., MLS(ASCP), C.Cy. At the end of this session, the participant will be able to: 1. Describe the components of a flow cytometer 2. Describe the gating
More informationUNIVERSITY OF PÉCS MEDICAL SCHOOL FLOW CYTOMETRY AND CELL SEPARATION BIOPHYSICS 2. 2015 4th March Dr. Beáta Bugyi Department of Biophysics Flow cytometry and cell separation FLOW = STREAM OF FLUID in a
More informationFlow Cytometry. flow cytometer DNA apoptosis ph
Flow Cytometry flow cytometer DNA apoptosis ph flow cytometry flow cytometer (a) (b) cell counting instrument (c) 1960 ink-jet technology 17 19 1940 1950 fluorescence microscopy fluorescent dye DNA polyclonal
More informationPrinciples of Flowcytometry
Objectives Introduction to Cell Markers: Principles of Flowcytometry Michelle Petrasich NZIMLS Scientific Meeting August 24, 2010, Paihia What are cell markers How do we detect them Production of Monoclonal
More informationCyAn : 11 Parameter Desktop Flow Cytometer
CyAn : 11 Parameter Desktop Flow Cytometer Cyan ADP 3 excitation lines 488nm, 635nm, and UV or violet 11 simultaneous parameters FSC, SSC, and 7-9 colors with simultaneous width, peak, area, and log on
More informationINSIDE THE BLACK BOX
FLOW CYTOMETRY ESSENTIALS INSIDE THE BLACK BOX Alice L. Givan Englert Cell Analysis Laboratory of the Norris Cotton Cancer Center Dartmouth Medical School HOW NOT TO BE A FLOW CYTOMETRIST Drawing by Ben
More informationAPPLICATION INFORMATION
DRAFT: Rev. D A-2045A APPLICATION INFORMATION Flow Cytometry 3-COLOR COMPENSATION Raquel Cabana,* Mark Cheetham, Jay Enten, Yong Song, Michael Thomas,* and Brendan S. Yee Beckman Coulter, Inc., Miami FL
More informationFlow Cytometry A Basic Overview
Flow Cytometry A Basic Overview Overview Flow cytometry is a powerful technology for investigating many aspects of cell biology and for isolating cells of interest. Flow cytometry utilizes highly focused,
More informationCyFlow SL. Microbiology. Detection and Analysis of Microorganisms and Small Particles
CyFlow SL Microbiology Detection and Analysis of Microorganisms and Small Particles 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results
More informationIntroduction to Flow Cytometry:
Introduction to Flow Cytometry: A Learning Guide Manual Part Number: 11-11032-01 April, 2000 BD Biosciences 2350 Qume Drive San Jose, CA 95131-1807 1-800-448-2347 Introduction to Flow Cytometry: A Learning
More informationCELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
More informationThese particles have something in common
These particles have something in common Blood cells Chromosomes Algae Protozoa Certain parameters of these particles can be measured with a flow cytometer Which parameters can be measured? the relative
More informationImmunophenotyping peripheral blood cells
IMMUNOPHENOTYPING Attune Accoustic Focusing Cytometer Immunophenotyping peripheral blood cells A no-lyse, no-wash, no cell loss method for immunophenotyping nucleated peripheral blood cells using the Attune
More informationFlow Cytometry Protocols
METHODS IN MOLECULAR BIOLOGY TM TM Volume 263 Flow Cytometry Protocols SECOND EDITION Edited by Teresa S. Hawley Robert G. Hawley 1 Flow Cytometry An Introduction Alice L. Givan Summary A flow cytometer
More informationToo Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry
Too Many B Cells: Chronic Lymphocytic Leukemia and the Role of Flow Cytometry by Debby R. Walser-Kuntz Biology Department Carleton College, Northfield, MN Taylor goes in to see her doctor, Dr. Chavez,
More informationApplication Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement
Application Information Bulletin: Set-Up of the CytoFLEX Set-Up of the CytoFLEX* for Extracellular Vesicle Measurement Andreas Spittler, MD, Associate Professor for Pathophysiology, Medical University
More informationCOMPENSATION MIT Flow Cytometry Core Facility
COMPENSATION MIT Flow Cytometry Core Facility Why do we need compensation? 1) Because the long emission spectrum tail of dyes causes overlap like with the fluorophores FITC and PE. 2) For sensitivity reasons,
More informationIntroduction to Flow Cytometry -- BD LSR II. What is Flow Cytometry? What Can a Flow Cytometer Tell Us About a Cell? Particle Size
Introduction to Flow Cytometry -- BD LSR II Daisy Kuo 郭 正 佼 Assistant Product Manager BD Biosciences Daisy_kuo@bd.com What is Flow Cytometry? Flow = Fluid Cyto = Cell Metry = Measurement A variety of measurements
More informationCyFlow SL. Healthcare Immunology. Portable FCM System for 3-colour Immunophenotyping
CyFlow SL Healthcare Immunology Portable FCM System for 3-colour Immunophenotyping 01 COMPANY Flow Cytometry made in Germany New sophisticated applications and increasing requirements for reliable results
More informationFlowSight. Flow cytometry with vision
FlowSight Flow cytometry with vision Flow cytometry with vision Introducing FlowSight Capable: Sensitive and flexible for every need Intuitive: Easy-to-use, with imagery for every cell Affordable: Designed
More informationFlow cytometry basics fluidics, optics, electronics...
Title Flow cytometry basics fluidics, optics, electronics... RNDr. Jan Svoboda, Ph.D. Cytometry and Microscopy Core Facility IMB, CAS, v.v.i Vídeňská 1083 Fluorescence Fluorescence occurs when a valence
More informationCELL CYCLE BASICS. G0/1 = 1X S Phase G2/M = 2X DYE FLUORESCENCE
CELL CYCLE BASICS Analysis of a population of cells replication state can be achieved by fluorescence labeling of the nuclei of cells in suspension and then analyzing the fluorescence properties of each
More informationMulticolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer
Multicolor Flow Cytometry: Setup and Optimization on the BD Accuri C6 Flow Cytometer Presented by Clare Rogers, MS Senior Marketing Applications Specialist BD Biosciences 23-13660-00 Webinar Overview Multicolor
More informationSmall-Particle Analysis
John Sharpe, PhD, and Sonja Wulff Introduction Flow cytometry instrumentation is a powerful tool for the analysis and identification of complex samples of particles of significantly varying size. However,
More informationEdU Flow Cytometry Kit. User Manual
User Manual Ordering information: (for detailed kit content see Table 2) EdU Flow Cytometry Kits for 50 assays: Product number EdU Used fluorescent dye BCK-FC488-50 10 mg 6-FAM Azide BCK-FC555-50 10 mg
More informationuse. 3,5 Apoptosis Viability Titration Assays
Introduction Monitoring cell viability can provide critical insights into the effectiveness of biological protocols and the stability of biological systems including: the efficacy of cell harvesting procedures,
More informationInternational Beryllium Conference, Montreal, Canada March 10, 2005
Alternative Lymphocyte Proliferation Tests: BrdU and Flow Cytometry Based Tests International Beryllium Conference, Montreal, Canada March 10, 2005 Tim K. Takaro Department of Environmental and Occupational
More informationCell Cycle Tutorial. Contents
Cell Cycle Tutorial Contents Experimental Requirements...2 DNA Dyes...2 Protocols...3 PI Parameter & Analysis Setup...4 PI Voltage Adjustments...6 7-AAD Parameter Setup...6 To-Pro3 Parameter Setup...6
More informationBelow is a list of things you should be aware of before you schedule your sort.
Sorting at the Flow Cytometry Facility At the present time, the assistance of a trained cell sorter operator is needed for all sorting applications. As such, if you are planning a first time sort you will
More informationStandardization, Calibration and Quality Control
Standardization, Calibration and Quality Control Ian Storie Flow cytometry has become an essential tool in the research and clinical diagnostic laboratory. The range of available flow-based diagnostic
More informationBoundary-breaking acoustic focusing cytometry
Boundary-breaking acoustic focusing cytometry Introducing the Attune NxT Acoustic Focusing Cytometer a high-performance system that s flexible enough for any lab One of the main projects in my laboratory
More informationUsing the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking
Using the BD TM Cytometer Setup and Tracking (CS&T) System for Instrument Characterization and Performance Tracking Mark KuKuruga Senior Technical Applications Specialist BD Biosciences 23-14462-00 Outline
More informationStepcount. Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom. Cat. Reference: STP-25T
Product Description: Closed transparent tubes with a metal screen, including a white matrix at the bottom Cat. Reference: STP-25T Reagent provided:: 25 Stepcount tubes for 25 test INTENDED USE. Immunostep
More informationBD FACSComp Software Tutorial
BD FACSComp Software Tutorial This tutorial guides you through a BD FACSComp software lyse/no-wash assay setup run. If you are already familiar with previous versions of BD FACSComp software on Mac OS
More informationCyFlow Space Your flexible flow cytometer
Your flexible flow cytometer www.sysmex-partec.com its flexibility gives you the space you need for your work Analysing cells and particles, be it from blood, plasma, tissue, plants, cell cultures or other
More informationThree Rules for Compensation Controls
From FlowJo s Daily Dongle Three Rules for Compensation Controls First and foremost, there must be a single-stained control for every parameter in the experiment! In addition, there are three rules for
More informationCompensation Controls Data Visualization Panel Development Gating Quantum Dots
Compensation Controls Data Visualization Panel Development Gating Quantum Dots FITC Single Stain Control FITC PE Argon Laser FL1 FL2 450 500 550 600 FITC Compensation Control Unwanted signal detected in
More informationapoptosis detection Apoptosis assays for the Attune Acoustic Focusing Cytometer Apoptosis detection Attune Acoustic Focusing Cytometer
apoptosis detection ttune coustic Focusing Cytometer poptosis detection poptosis assays for the ttune coustic Focusing Cytometer poptosis is a carefully regulated process of cell death that occurs as a
More informationNo-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer
APPLICATION NOTE Attune NxT Flow Cytometer No-wash, no-lyse detection of leukocytes in human whole blood on the Attune NxT Flow Cytometer Introduction Standard methods for isolating and detecting leukocytes
More informationDELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES
DELPHI 27 V 2016 CYTOMETRY STRATEGIES IN THE DIAGNOSIS OF HEMATOLOGICAL DISEASES CLAUDIO ORTOLANI UNIVERSITY OF URBINO - ITALY SUN TZU (544 b.c. 496 b.c) SUN TZU (544 b.c. 496 b.c.) THE ART OF CYTOMETRY
More informationHoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-806)
Hoechst 33342 HSC Staining and Stem Cell Purification Protocol (see Goodell, M., et al. (1996) J Exp Med 183, 1797-86) The Hoechst purification was established for murine hematopoietic stem cells (HSC)
More informationSupplementary Material. Free-radical production after post-thaw incubation of ram spermatozoa is related to decreased in vivo fertility
10.1071/RD14043_AC CSIRO 2015 Supplementary Material: Reproduction, Fertility and Development, 2015, 27(8), 1187 1196. Supplementary Material Free-radical production after post-thaw incubation of ram spermatozoa
More informationImmunophenotyping Flow Cytometry Tutorial. Contents. Experimental Requirements...1. Data Storage...2. Voltage Adjustments...3. Compensation...
Immunophenotyping Flow Cytometry Tutorial Contents Experimental Requirements...1 Data Storage...2 Voltage Adjustments...3 Compensation...5 Experimental Requirements For immunophenotyping with FITC and
More informationCell Viability Assays: Microtitration (MTT) Viability Test Live/Dead Fluorescence Assay. Proliferation Assay: Anti-PCNA Staining
Cell Viability Assays: Microtitration (MTT) Viability Test Live/Dead Fluorescence Assay Proliferation Assay: Anti-PCNA Staining Spring 2008 1 Objectives To determine the viability of cells under different
More informationFlow Data Analysis. Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA FLOWJO CYTOMETRY DATA ANALYSIS SOFTWARE
Flow Data Analysis Qianjun Zhang Application Scientist, Tree Star Inc. Oregon, USA Flow Data Analysis From Cells to Data - data measurement and standard Data display and gating - Data Display options -
More information123count ebeads Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use.
Page 1 of 1 Catalog Number: 01-1234 Also known as: Absolute cell count beads GPR: General Purpose Reagents. For Laboratory Use. Normal human peripheral blood was stained with Anti- Human CD45 PE (cat.
More informationThe NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee
January, 1999 From: To: Subject: The NIAID Flow Cytometry Advisory Committee; the Guidelines Subcommittee NIAID DAIDS Flow Cytometry Laboratories Comparison study information for labs wishing to switch
More informationLab 2. Isolation of mononuclear cells from peripheral blood and separation into subpopulations
Lab 2 Isolation of mononuclear cells from peripheral blood and separation into subpopulations Supervisors: Sissela Broos sissela.broos@immun.lth.se tel: 222 96 78 Niclas Olsson niclas.olsson@immun.lth.se
More informationUsing BD FACSDiva CST To. Evaluate Cytometer Performance, Create Custom Assay Settings. and
Using BD FACSDiva CST To Evaluate Cytometer Performance, Create Custom Assay Settings and Implement Cross-Instrument and Cross-Site Standardization of Assays PART 2 Alan M. Stall Director, Advanced Cytometry
More informationSpherotech, Inc. 27845 Irma Lee Circle, Unit 101, Lake Forest, Illinois 60045 1
78 Irma Lee Circle, Unit 0, Lake Forest, Illinois 00 SPHERO TM Technical Note STN- Rev B 008 DETERMINING PMT LINEARITY IN FLOW CYTOMETERS USING THE SPHERO TM PMT QUALITY CONTROL EXCEL TEMPLATE Introduction
More informationMolecular Spectroscopy
Molecular Spectroscopy UV-Vis Spectroscopy Absorption Characteristics of Some Common Chromophores UV-Vis Spectroscopy Absorption Characteristics of Aromatic Compounds UV-Vis Spectroscopy Effect of extended
More informationIslet Viability Assessment by Single Cell Flow Cytometry
Islet Viability Assessment by Single Cell Flow Cytometry Page 1 of 8 Purpose: To comprehensively assess the viability of the islet cell preparation prior to transplantation. Tissue Samples: A sample containing
More informationTechnical Bulletin. Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer
Threshold and Analysis of Small Particles on the BD Accuri C6 Flow Cytometer Contents 2 Thresholds 2 Setting the Threshold When analyzing small particles, defined as particles smaller than 3.0 µm, on the
More informationTechnical Bulletin. An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers
An Introduction to Compensation for Multicolor Assays on Digital Flow Cytometers BD Biosciences, San Jose, CA Contents 2 Introduction to compensation 5 Tips for setting up compensation 10 Tools for easy,
More informationFlow Cytometry 101. Jodi Moore, PhD Jan 2012 Goldenson 113
Flow Cytometry 101 Jodi Moore, PhD Jan 2012 Goldenson 113 Origins:? Dye Chemistry 1940-1980 s Electronics Microscopy Automated Clinical Cytology ID Apps Computers Cytometry vs. Flow Cytometry Microspectrophotometry
More informationApplication Note. Selecting Reagents for Multicolor Flow Cytometry. Holden Maecker and Joe Trotter BD Biosciences, San Jose
Selecting Reagents for Multicolor Flow Cytometry Holden Maecker and Joe Trotter BD Biosciences, San Jose Contents 1 The basics: Know your instrument 2 Fluorochromes: Go for the bright 3 Colors and specificities:
More informationInstrument Characterization and Performance Tracking for Digital Flow Cytometers
Instrument Characterization and Performance Tracking for Digital Flow Cytometers BD Biosciences Cytometer Set-up & Tracking (CS&T) System For Research Use Only. Not for use in diagnostic or therapeutic
More informationCompensation in Flow Cytometry
Compensation in UNIT.4 The term compensation, as it applies to flow cytometric analysis, refers to the process of correcting for fluorescence spillover, i.e., removing the signal of any given fluorochrome
More informationCAP Accreditation Checklists 2015 Edition
CAP Accreditation Checklists 2015 Edition The College of American Pathologists (CAP) accreditation checklists contain the CAP accreditation program requirements, developed on more than 50 years of insight
More informationSelected Topics in Electrical Engineering: Flow Cytometry Data Analysis
Selected Topics in Electrical Engineering: Flow Cytometry Data Analysis Bilge Karaçalı, PhD Department of Electrical and Electronics Engineering Izmir Institute of Technology Outline Compensation and gating
More information1. Instrument Layout pg 2. 2. Summit Overview pg 3. 3. Data Storage and Data Files pgs 6-7. 4. Saving Data pg 7. 5. Start Up pg 8
Table of Contents 1. Instrument Layout pg 2 2. Summit Overview pg 3 3. Data Storage and Data Files pgs 6-7 4. Saving Data pg 7 5. Start Up pg 8 6. Choosing and Naming Parameters pg 9 7. Creating Histograms
More informationpowerful. Intuitive. Customizable. The new star of benchtop flow cytometry
powerful. Intuitive. Customizable. The new star of benchtop flow cytometry About ACEA Established in 2002, ACEA Biosciences, Inc. develops cutting-edge cell analysis platforms for life science research.
More informationResults: A plot showing propidium iodide and Annexin is used to determine apoptotic (Annexin only +) from necrotic (PI+ and Annexin +).
Results: A plot showing propidium iodide and Annexin is used to determine apoptotic (Annexin only +) from necrotic (PI+ and Annexin +). This group should be a low percentage. Surface markers can then be
More informationa Beckman Coulter Life Sciences: White Paper
a Beckman Coulter Life Sciences: White Paper Violet-Excited nim-da Allows Efficient and Reproducible Cell Cycle Analysis on the Gallios Flow Cytometer Authors: Valdez, Ben 1. Carr, Karen 2. Norman, John
More informationAnnexin V-EGFP Apoptosis Detection Kit
ab14153 Annexin V-EGFP Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of apoptosis in various samples This product is for research use only and is not intended
More informationA Basic Guide. Institute of Medical Biology, University of Southern Denmark. Ungated Gate 1 (= R1) Gate 4 (=R4) Gate3 (=R3) Gate 2 (= R2) R3 R4
Flow Cytometry A Basic Guide R1 Ungated Gate 1 (= R1) R2 R3 R4 Gate 2 (= R2) Gate3 (=R3) Gate 4 (=R4) Graham Leslie, PhD Institute of Medical Biology, University of Southern Denmark Graham Leslie, PhD
More informationAn Overview of Cells and Cell Research
An Overview of Cells and Cell Research 1 An Overview of Cells and Cell Research Chapter Outline Model Species and Cell types Cell components Tools of Cell Biology Model Species E. Coli: simplest organism
More informationChapter 10 Immunofluorescence
Chapter 10 Immunofluorescence J. Paul Robinson PhD, Jennifer Sturgis BS and George L. Kumar PhD Immunofluorescence (IF) is a common laboratory technique used in almost all aspects of biology. This technique
More informationBD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY
BD FACSDiva 4.1 - TUTORIAL TSRI FLOW CYTOMETRY CORE FACILITY IMPORTANT NOTES BEFORE READING THIS TUTORIAL This is a very expensive piece of equipment so PLEASE treat it with respect! After you are done
More informationMACSQuantify Software guide. Basics of flow cytometric analysis and software guide. Version 1 EN Original instructions
MACSQuantify Software guide Basics of flow cytometric analysis and software guide Version 1 EN Original instructions Miltenyi Biotec GmbH Friedrich-Ebert-Straße 68 51429 Bergisch Gladbach Germany Phone:
More informationThe use of FACS for the selection of cell lines with superior productivity characteristics
ESACT-UK, Cambridge 2007 The use of FACS for the selection of cell lines with superior productivity characteristics Jonathan Welsh Synopsis Why use FACS in bioprocess development Issues regarding use of
More informationApplication Note 10. Measurement of Cell Recovery. After Sorting with a Catcher-Tube-Based. Cell Sorter. Introduction
Application Note 10 Measurement of Cell Recovery After Sorting with a Catcher-Tube-Based Cell Sorter Introduction In many experiments using sorted cells, it is important to be able to count the number
More informationCell Cycle Phase Determination Kit
Cell Cycle Phase Determination Kit Item No. 10009349 Customer Service 800.364.9897 * Technical Support 888.526.5351 www.caymanchem.com TABLE OF CONTENTS GENERAL INFORMATION 3 Materials Supplied 3 Safety
More informationScience is hard. Flow cytometry should be easy.
Science is hard. Flow cytometry should be easy. TABLE OF CONTENTS 1 INTRODUCTION TO BD ACCURI C6 SOFTWARE... 1 1.1 Starting BD Accuri C6 Software... 1 1.2 BD Accuri C6 Software Workspace... 2 1.3 Opening
More informationMulticolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility
Multicolor Bead Flow Cytometry Standardization Heba Degheidy MD, PhD, QCYM DB/OSEL/CDRH/FDA Manager of MCM Flow Cytometry Facility The mention of commercial products, their sources, or their use in connection
More informationFACS Laboratory. BD LSRs Operator Training. Training Guide. http://london-research-institute.org.uk/technologies/facs
FACS Laboratory facs_lab@cancer.org.uk Training Guide http://london-research-institute.org.uk/technologies/facs BD LSRs Operator Training LRI FACS lab 1 Components 1.1 Basic parts overview.pg3-6 Overview
More informationPartec PAS. Your Flow Cytometer.
Partec PAS Your Flow Cytometer. The Challenge... New sophisticated applications and increasing requirements for reliable routine results in the shortest possible time continue to impose a challenge for
More informationBD FACSCanto Flow Cytometer. Efficient Easy Powerful
BD FACSCanto Flow Cytometer Efficient Easy Powerful The BD FACSCanto System: Optimal Workflow and Time Management Automated compensation algorithms with optional off-line compensation, streamlined sample
More informationSetting up and calibration of a flow cytometer for multicolor immunophenotyping
Journal of Biological Regulators and Homeostatic Agents Setting up and calibration of a flow cytometer for multicolor immunophenotyping J. KRAAN 1, J.W. GRATAMA 1, M. KEENEY 2, J.L. D HAUTCOURT 3 1 Department
More informationApplications of confocal fluorescence microscopy in biological sciences
Applications of confocal fluorescence microscopy in biological sciences B R Boruah Department of Physics IIT Guwahati Email: brboruah@iitg.ac.in Page 1 Contents Introduction Optical resolution Optical
More informationBD FACSCalibur. The flow cytometer for your routine cell analysis needs
BD FACSCalibur The flow cytometer for your routine cell analysis needs A system with a rich application basis and a modular approach that continues to meet evolving needs of cell analysis worldwide. The
More informationab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis
ab139418 Propidium Iodide Flow Cytometry Kit for Cell Cycle Analysis Instructions for Use To determine cell cycle status in tissue culture cell lines by measuring DNA content using a flow cytometer. This
More informationFITC Annexin V/Dead Cell Apoptosis Kit with FITC annexin V and PI, for Flow Cytometry
FITC Annexin V/Dead Cell Apoptosis Kit with FITC annexin V and PI, for Flow Cytometry Catalog no. V13242 Table 1. Contents and storage information. Material Amount Composition Storage* Stability FITC annexin
More informationFluorescence Compensation. Jennifer Wilshire, PhD jennifer@flowjo.com
Fluorescence Compensation Jennifer Wilshire, PhD jennifer@flowjo.com Outline 1. Compensation Basics 2. Practical Applications -selecting comp controls (cells, stains?) 3. Multicolor Issues -spreading due
More informationspecific B cells Humoral immunity lymphocytes antibodies B cells bone marrow Cell-mediated immunity: T cells antibodies proteins
Adaptive Immunity Chapter 17: Adaptive (specific) Immunity Bio 139 Dr. Amy Rogers Host defenses that are specific to a particular infectious agent Can be innate or genetic for humans as a group: most microbes
More informationChapter 6. Antigen-Antibody Properties 10/3/2012. Antigen-Antibody Interactions: Principles and Applications. Precipitin reactions
Chapter 6 Antigen-Antibody Interactions: Principles and Applications Antigen-Antibody Properties You must remember antibody affinity (single) VS avidity (multiple) High affinity: bound tightly and longer!
More informationBacterial Detection and Live/Dead Discrimination by Flow Cytometry
A P P L I C A T I O N N O T E Bacterial Detection and Live/Dead Discrimination by Flow Cytometry M I C R O B I A L C Y T O M E T R Y BD FACSCalibur, BD FACSort Flow Cytometers Introduction Accurate determination
More informationPROJECT GROUP FOR AUTOMATION IN MEDICINE AND BIOTECHNOLOGY IN MANNHEIM BIOPROCESS ENGINEERING
Fraunhofer Institute for Manufacturing Engineering and Automation IPA PROJECT GROUP FOR AUTOMATION IN MEDICINE AND BIOTECHNOLOGY IN MANNHEIM BIOPROCESS ENGINEERING VERGÜTUNG VON ERFINDUNGEN UND TECHNISCHEN
More informationArC Amine Reactive Compensation Bead Kit
ArC Amine Reactive Compensation Bead Kit Catalog no. A1346 Table 1. Contents and storage information. Material Amount Composition Storage Stability ArC reactive beads (Component A) ArC negative beads (Component
More information1.1.2. thebiotutor. AS Biology OCR. Unit F211: Cells, Exchange & Transport. Module 1.2 Cell Membranes. Notes & Questions.
thebiotutor AS Biology OCR Unit F211: Cells, Exchange & Transport Module 1.2 Cell Membranes Notes & Questions Andy Todd 1 Outline the roles of membranes within cells and at the surface of cells. The main
More informationOn the origin of giant multinuclear Reed-Sternberg cells and the role of CD4 T cells in Hodgkin lymphoma
On the origin of giant multinuclear Reed-Sternberg cells and the role of CD4 T cells in Hodgkin lymphoma Uber die Entstehung von multinuklearen Reed-Sternberg Riesenzellen und die Rolle von CD4 T-Zellen
More informationHow to operate the BD FACSCanto flow cytometer
How to operate the BD FACSCanto flow cytometer Preface Dear colleague, the BD FACSCanto flow cytometer is use to operate, however, it is a delicate instrument for measuring fluorescence of single cells.
More informationAnnexin V-FITC Apoptosis Detection Kit
ab14085 Annexin V-FITC Apoptosis Detection Kit Instructions for Use For the rapid, sensitive and accurate measurement of Apoptosis in living cells (adherent and suspension). This product is for research
More information