1 Materials Science and Engineering C 30 (2010) Contents lists available at ScienceDirect Materials Science and Engineering C journal homepage: Evaluating mechanical properties and degradation of YTZP dental implants Pablo Sevilla a,, Clara Sandino a, Milena Arciniegas a, Jordi Martínez-Gomis b, Maria Peraire b, Francisco Javier Gil a a Biomaterials and Biomechanics Division, Department of Materials Science and Metallurgy, Technical University of Catalonia, Spain b Department of Prosthodontics, Faculty of Odontology, University of Barcelona, Spain article info abstract Article history: Received 28 April 2009 Received in revised form 10 July 2009 Accepted 4 August 2009 Available online 13 August 2009 Keywords: Y-TZP Dental implants Fatigue and fracture Degradation Nanoindentation Lately new biomedical grade yttria stabilized zirconia (YTZP) dental implants have appeared in the implantology market. This material has better aesthetical properties than conventional titanium used for implants but long term behaviour of these new implants is not yet well known. The aim of this paper is to quantify the mechanical response of YTZP dental implants previously degraded under different time conditions and compare the toughness and fatigue strength with titanium implants. Mechanical response has been studied by means of mechanical testing following the ISO for Standards for dental implants and by finite element analysis. Accelerated hydrothermal degradation has been achieved by means of water vapour and studied by X-ray diffraction and nanoindentation tests. The results show that the degradation suffered by YTZP dental implants will not have a significant effect on the mechanical behaviour. Otherwise the fracture toughness of YTZP ceramics is still insufficient in certain implantation conditions. Published by Elsevier B.V. 1. Introduction The zirconia ceramics are used in different applications due to their remarkable mechanical and physical properties, such as their very low thermal expansion coefficient and stability under very high temperature [1]. Important advances have been achieved in these materials, such as the increment of fracture toughness produced by the stabilization of the tetragonal phase with different oxides (i.e. yttria, ceria or magnesia) [2]. This material, called Yttria-stabilized tetragonal zirconia polycrystal (Y-TZP), exhibits a very high flexural strength (900 to 1200 MPa), a notable fracture toughness (K IC 7 to 10 MPa m 1/2 ) and a Young's modulus of 210 GPa [3] which has permitted to expand its applications to biomedical field, with special interest in load-transfer applications such as dental implants. The assumed advantages of the Y-TZP in comparison with the widely employed c.p. titanium or Ti6Al4V for dental implants are a better aesthetical outcome due to its shiny white when the implant shows through a thin mucosa or when it becomes visible following a mucosa retraction, and a great biocompatibility due to the absence of metal ions in the surrounding tissue [4]. The experience with the utilization of biomedical grade zirconia femoral heads for orthopaedic hip implants showed that a low temperature hydrothermal degradation, also called ageing, takes place when the material is exposed to physiological conditions. The current knowledge on ageing process and on its effect on the long term performance shows that there is a strong variability of zirconia to in vivo degradation as a consequence of different process related to microstructure [5]. Theuse Corresponding author. Tel.: ; fax: address: pablo.sevilla@upc.edu (P. Sevilla). of zirconia for dental implants is in its early phase and the issue of ageing needs further research. Therefore, the purpose of this study was to quantify the mechanical response of YTZP implants previously degraded under different time conditions and to compare the toughness and fatigue strength of YTZP implants with those of titanium implants. 2. Materials and methods 21 commercial biomedical grade yttria stabilized zirconia dental implants (Zlock3 411 YTZP A-BIO HIP, Z-System, Konstanz, Germany) were acquired and 43 c.p. grade 3 implants (SK2 Ti Klockner Implant Systems, Barcelona, Spain) were provided by the manufacturer (Fig. 1). The Zlock3 411 implants have a threaded length of 11 mm with a diameter of 4 mm and a diameter of the crest module of 6 mm. The SK2 implants have a threaded length of 10 mm with an average diameter of 3.8 mm and a diameter of the crest module of 4.2 mm. These two types of implants were chosen because they are the most used of both companies and have similar applications in mouth. The chemical composition of the two kinds of implants is shown in Tables 1 and 2, according to manufacturer's data. The zirconia implants were divided in two groups. The first group of 3 implants was submitted to nanoindentation test after a degradation process. The second group of 18 implants was submitted to mechanical tests Low-temperature autoclave ageing Each one of the 3 YTZP Zlock3 implants of the first group were cut with diamond disk to obtain 21 slices 1 mm thick and were /$ see front matter. Published by Elsevier B.V. doi: /j.msec

2 P. Sevilla et al. / Materials Science and Engineering C 30 (2010) Table 2 Chemical composition of the c.p. grade 3 titanium. O C N H Fe Ti Ti c.p. grade % 0.08% 0.05% 0.01% 0.30% Balance Percentages in weight. Fig. 1. Dental implants used in the study. Left: Zlock3 411 Z-Systems (Germany). Right: SK2 Klockner Implant System (Spain). mechanically polished with diamond paste from a solution with a particle size of 30 up to 1 µm and finished with colloidal silica to give an average surface roughness of <5 nm. Finally, all samples were washed with acetone and dried at room temperature. 18 slices were submitted to a degradation process conducted by means of an autoclave at 134 C under 2 Bar of pressure. The degradation times were 1, 2, 3, 4, 5 and 10 h and were applied to 3 slices each one [6]. 3 slices were left not degraded X-ray diffraction The martensitic transformation from tetragonal to monoclinic phase induced by the degradation was analysed by X-ray diffraction using a Bruker diffractometer with Cu Kα of The voltage, intensity and step-size applied were 40 kv, 20 ma and 0.02, respectively. The counting time per each step was 16 s. In order to quantify the transformed phase fraction the following equations were employed [7]: I m 111 ð Þ + I mð111þ X m = ð1þ I mð111þ + I mð111þ + I tð101þ where I t and I m represent the integrated intensity (area under peaks) of the tetragonal (101) and monoclinic (111) and ( 111) peaks. The volume of monoclinic phase is then given by [6]: V m = 1:311 X m 1+0:311 X m ð2þ oscillations. All tests were conducted at 22 C and the indenter was held in contact with the surface until the thermal drift was lower than 0.05 nm s 1. It is possible to know the thickness of the degraded layer by means of a model of thin layer properties that has in account the contribution of the substrate to the mechanical response of the surface layer submitted to nanoindentation and proposes the next equation [9]: 1 E c 2a = 1+ ð 2t = πaþ t πa 2 E f! + 1 2aE s where: E c is the Apparent Young modulus obtained in the test. E f E s a t is the Young modulus of the degraded layer is the substrate Young modulus. (210 GPa) is the contact radius corresponding to cylindrical indenter with the same contact area and it's defined as: A c =πa 2 is the thickness of the degraded layer 2.4. Mechanical tests The mechanical tests were carried out following the ISO Standard [10]. Five Zlock3 411 and five SK2 implants were embedded in an acrylic resin Tecnovit 4071, Sulzer (Switzerland) 3 mm over the nominal depth, simulating 3 mm of bone resorption, and placed in an angle of 30 with respect to the vertical axis of the implant (Fig. 2). Since the Zlock3 implants are implant and abutment in one piece, a special abutment were connected to SK2 implants to equal the test conditions. The testing machine was a Bionix 858, MTS (Minnesota, USA) controlled by MTS Testworks 4 software. The test was done firstly statically to determine the 30 flexure of the implants. Later ð3þ 2.3. Nanoindentation test Samples of each degradation time, as well as non degraded samples, were used for instrumented nanoindentation tests on an MTS XP System Nano Indenter. On each sample nine monotonic tests were performed, in which the penetration depth was linearly increased up to the maximum depth of 1600 nm. The tests were conducted with a Berkovich tip with a measured radius of 750 nm. The elastic modulus and contact pressure values were evaluated with Continuous Stiffness Measurement (CSM) module [8]. The CSM module overlaps high frequency oscillations to the P h curves, evaluating the elastic response from the unloading portion of such Table 1 Chemical composition of the commercial YTZP A-BIO HIP acquired for the studies. Oxides Percentage (in wt.%) ZrO 2 (+HfO 2 ) >95.5 Y 2 O Al 2 O Fig. 2. Scheme of the mechanical tests following ISO Standard.

3 16 P. Sevilla et al. / Materials Science and Engineering C 30 (2010) on, fatigue tests were carried out setting a maximum load of 40, 50, 60 and 80% of the failure load detected on the static tests. The minimum load was always 10% of the maximum load, the waveshape was sinusoidal and the frequency was established in 15 Hz. In each case 3 Zlock3 411 and 3 SK2 Implants were tested. A total of 18 SK2 implants and 18 Zlock3 411 implants were tested following the ISO Standard. Testing condition specified by ISO Standard are the most unfavourable conditions that a dental implant is supposed to stand. It was found interesting to know the mechanical response of both types of implants under not so demanding conditions. With this purpose, the 25 SK2 remaining implants were tested changing the slope of the long implant axis with respect to the vertical machine axis from 30 to 15 and 0. Also, at these 3 slopes implants were tested simulating bone resorption (3 mm over the nominal depth) and without simulating bone resorption. For each case 5 SK2 implants were tested. Zlock3 411 implants were not tested at conditions different from the specified by ISO Standard. Instead of it, a finite element model of Zlock3 411 implant was developed as it is explained in the next paragraph Finite element analysis A finite element model of the Zlock3 411 implant was created. Initially a 3d model of the implant was created by means of Solid Works 2007 software (USA), next the 3d model was meshed using Patran, MSC Software (USA) and finally the contour conditions and the calculus were carried out by means of Marc software, MSC Software (USA). The conditions studied by finite element study were: 0, 15 and 30 of slope with respect to the axis of the force and 3 mm over the nominal depth, simulating 3 mm of bone resorption as ISO standard recommends and the implant immersed until the nominal depth, simulating no bone resorption at all (Fig. 3). To validate the model the results of 30 slope with 3 mm of bone resorption were compared with the results of mechanical testing following the ISO Standard Degradation studies To determine how the hydrothermal degradation affects to the fracture of the Zlock3 implants submitted to flexure forces the stress intensity factor was calculated. A mathematical model was established for a theoretical bar submitted to a 3 point bending test to the stress calculated by finite element on the Zlock 3 implants with a Fig. 4. X-ray diffraction spectra of the different times hydrothermally aged samples. The monoclinic peak at 2θ=28.2 is marked with the letter m. The peaks belonging the tetragonal phase are marked with the letter t. surface crack of the length of the degraded layer thickness obtained in the nanoindentation test by the following equation [11]: p K I = σ ffiffiffiffiffiffiffiffiffi π a Fða= bþ ð4þ Fa= ð bþ = p 1 ffiffiffi π 1:99 a = bð1 a = b h i Þ 2:15 3:93a = b +2:7ða=bÞ 2 ð1+2a = bþð1 a=bþ 3 = 2 ð5þ where K I is the stress intensity factor; a, the crack length and b is the bar thickness, which was established in 4 mm. 3. Results 3.1. X-ray diffraction X-ray diffraction profiles of the degraded slices of zirconia implants are shown in Fig. 4. In this figure a fully tetragonal pattern is observed in the non degraded sample. In contrast, from the degraded samples a peak at 28.2 is detected, which is small under short time of degradation, but is increased as a function of the degradation time. The presence of this peak and the intensity changes are associated with the martensitic transformation from a tetragonal to monoclinic structure. The transformed phase fraction and volume values calculated for each sample are shown in Table 3. A notable increment of the transformed phase is produced as a function of the degradation time. Table 3 Monoclinic fraction of the hydrothermally aged samples calculated by Garvie and Nicholson method. Degradation time Monoclinic fraction (%) Monoclinic fraction in volume (%) Fig D FEA models of implant and acrylic resin in two different loading cases: Left: 3 mm elevate apex (simulating 3 mm bone resorption like ISO testing). Right: Immersed apex (without bone resorption). 0h h h h h

4 P. Sevilla et al. / Materials Science and Engineering C 30 (2010) Fig. 5. Hardness as a function of tip displacement in the nanoindentation test for different times of hydrothermally aging samples. Fig. 7. Degraded layer Young modulus as a function of tip displacement calculated with Bec's equation for different layer thicknesses. It fits to say that the penetration of the X-ray beam is around 1 µm. So this technique is useful in the range of the beam penetration to detect the quantity of phase transformation Nanoindentation tests Figs. 5 and 6 show constant hardness and elastic modulus values as functions of the penetration depth, respectively, obtained by nanoindentation on the non degraded sample, with an average value of 16.9±0.9 GPa and 245.7±11.6 GPa. Conversely, a significant reduction of these properties is presented by the all degraded samples, measured in first 1600 nm, with an increment of the values, up to reach similar values to those found from the non degraded sample, as the penetration into surface is raised (Figs. 5 and 6). The Young modulus of the degraded layer was determined using the results of the most degraded sample. Showing the graphic of the Film Young modulus (E f ) vs indentation deepness for different layer thicknesses (t) and choosing the value of E f and t that produces a constant value across the indentation deepness. In this case a value of 165 GPa was chosen as Fig. 7 shows. Finally it is possible to determine the degraded layer thickness comparing the curves of E c vs tip displacement calculated by the Bec equation for different values of degraded layer thickness t with the curves of E c vs tip displacement obtained on the nanoindentation test. Fig. 8 shows an example for the sample degraded during 10 h obtaining a degraded layer thickness of 0.7 µm. The values of the degraded layer thickness for each time of autoclave aging are shown in Table Mechanical testing Fig. 9 shows the mechanical response of the Zlock3 411 implant and the SK2 implant to a monotonic 30 flexure test. It can be seen that the slope (stiffness) of the curve on the YTZP implants is much higher than the slope on the titanium implants (108.8±8.78 N/mm for Zlock3 implant and 40.6±2.26 N/mm for the SK2 implants). Also the figure shows that the behaviour of the Zlock3 implants is completely fragile and it don't tolerate any plastic strain. Otherwise the SK2 implants can be deformed permanently during an overload without breaking. Finally it fits to say the failure force for both implants is very similar although the elastic limit of the SK2 implant is slightly higher than the maximum of Zlock3 411 implants Fig. 6. Young Modulus as a function of tip displacement in the nanoindentation test for different times of hydrothermally aging samples. Fig. 8. Apparent Young modulus as a function of tip displacement calculated with Bec's equation for different layer thicknesses and compared with nanoindentation results.

5 18 P. Sevilla et al. / Materials Science and Engineering C 30 (2010) Table 4 Thicknesses of the degraded layer for different hydrothermal aging time. Aging time Degraded layer thickness (µm) 0h 0 1 h h h h h h 0.70 ( of Zlock3: 736.3±65.3 N, Elastic limit of SK2: 812.2±25.0 N). For the fatigue testing the results are shown in Fig. 10. As it can be seen in the graphic the fatigue limit for both implants is very similar (about 300 N) and higher than the Standard recommendation (225 N). Otherwise the Zlock3 implant presents a ceramic behaviour where no very high overloads can be tolerated by the material while SK2 implants can tolerate many cycles of higher forces than the fatigue limit of the implant Comparison of mechanical response Table 5 shows the toughness of the implants in different conditions as it was specified in the methodology. The values obtained for SK2 implants are from mechanical testing while the values for Zlock3 411 implants are obtained by means of finite element analysis. It can be seen that in the worst possible conditions (angle of 30 and 3 mm of elevated crest module) the behaviour of Zlock3 411 implants is worse than the behaviour of SK2 implants. Nevertheless in not so demanding conditions (angles of 15 and 0 ) the Zlock3 implants have better response than SK2 implants and also when there is no bone resorption at all the flexure of the Zlock3 implants even doubles the maximum of the SK2 implants Degradation studies The calculus of the stress intensity factor of the Zlock3 implants is shown in Table 6. It can be seen that the stress intensity factor produced by the degraded layer is very low even on the highest times of hydrothermal ageing (10 h where the K I is 1.49 MPa m 1/2 ). To arrive at a stress intensity factor equal to the fracture toughness of the material (6 MPa m 1/2 ) it is necessary to make the hydrothermal Fig. 10. Fatigue curve of both implants studied done with ISO Standard recommendations. ageing treatment on the material during about 170 h which implies a degraded layer thickness of 12 µm. 4. Discussion A fully tetragonal pattern on non degraded samples demonstrates that the polishing method was correct and it didn't produce stresses on the samples surface [12]. The martensitic transformation from a tetragonal to a monoclinic structure observed with X-ray diffraction during the degradation process is attributed to the hydrothermal degradation that takes place when the material is in contact with a liquid medium, as it was reported [6,13,14]. The notable increment of the phase transformed, as a function of the degradation times is in agreement with the results reported by other authors [6,14]. The significant reduction of hardness and elastic modulus presented by the all degraded samples, measured in first 1600 nm, with an increment of the values, up to reach similar values to those found from the non degraded sample, as the penetration into surface is raised is a phenomenon well reported in literature [15] and it is attributed to a degradation process that is started on the surface due to the autoclave conditions, and continue through the material in a very homogeneous mode. Although the variation in hardness could be attributed to the appearance of the monoclinic phase, the marked variation in elastic modulus suggest that the mechanism responsible of this variation is the generation of microcracks. When a partial unload is carried out, the cracks open and the contact stiffness decreases on decreasing the effective area of load transfer. These microcracks are generated due to the volumetric expansion of the monoclinic phase and they are more evident when the materials are submitted to indentation [16]. Nevertheless, the problem of hydrothermal degradation that has affected many YTZP Hip prostheses in the past doesn't seem to be so critical on the application of dental implantology. It can be explained because the working mode of this material on a hip prosthesis and on Table 5 Mechanical of the studied implants in different mechanical conditions. Angle ( ) Elevated apex 3 mm (ISO Standard) Inmersed apex (ideal case) SK2 Zlock3 411 SK2 Zlock3 411 Elastic limit Elastic limit Fig. 9. Force as a function of displacement for both implants studied in the static 30 flexure test , , , ,106.2

6 P. Sevilla et al. / Materials Science and Engineering C 30 (2010) Table 6 Stress intensity factors calculated for different times of hydrothermal aging. Aging time Crack length (µm) K I (MPa m 1/2 ) 0h h h h h h h a dental implant is completely different. In a hip prosthesis the ball and the cup are wearing each walking cycle. When the water in contact with the YTZP components transforms the surface from tetragonal to monoclinic phase it becomes rougher and the friction force is increased performing grain pull-out leaving a new tetragonal surface to transform and provoking abrasive wear [17]. A dental implant has a structural function. It stays static without friction between parts and without grain pull-out. Comparing the design of Zlock3 411 and SK2 implants from Fig. 1 some differences can be observed. The most important difference relating to the mechanical properties of the implants is that the Zlock3 implant has a higher diameter (6 mm) on the crest module of the implant. This is a design solution to minimize the stress bore by the threaded zone which, because of the low fracture toughness of the YTZP, is the weakest point of the implant. It is very important for Zlock3 implants that the crest module will be in contact with the bone to be able to shield the stress on the lower zones. It can be seen in Fig. 11 from finite element analysis results which shows that an immersed implant has a maximum stress 4 times lower than an implant with a 3 mm exposed implant. This fact has a paramount importance in the case of a bone resorption occurs. The results of monotonic 30 flexure tests demonstrate that the stiffness of the YTZP implants is much higher than the titanium ones. These results suggest that the YTZP implants can provoke stress shielding [18,19] and, consequently, bone resorption more than titanium implants because the low deformation of the implant due to the applied forces will generate low transmission of forces to the cortical bone. Fig. 11. Stress plots of Zlock3 411 implant in the case of 30 flexion test. Left: 3 mm elevated apex. Right: Immersed apex. The arrows show the direction of the force. Horizontal dark blue lines indicate the position of the implant inside the resin. (For interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.) Fatigue tests have shown that although the Zlock3 411 implants have a good fatigue limit they are not as capable to resist one-off overloads as the titanium implants. It is important to take it in account when treating patients with bruxism. For the fatigue testing it is important to remark that the graphic doesn't show any static fatigue effect [20 23] given that the tests were too short in time to show this effect. Comparing new YTZP implants with conventional titanium implants the mechanical test have shown that for Zlock3 implants at 30 angle of inclination the mechanical response will be poorer than the response of SK2 implants. Clinically, in anterior situation or in posterior situation with lateral occlusal contacts, the force axis over the implant can be greater than 30 [24]. Nevertheless with lower angle of inclination the Zlock3 implants have higher than SK2 implants. It fits to say that the manufacturer recommends never implant the Zlock3 implants with a slope higher than 20 with respect to the force axis. 5. Conclusions In ideal implantation conditions: minimal inclination of the implant with respect to the force axis, crest module of the implant in intimate contact with bone and absence of overloads, the Zlock implant shows better than SK2 implant. Notwithstanding the low fracture toughness of YTZP implants make them not capable to stand one-off overloads. It is essential that the Zlock implant will be the more parallel to the force axis in order to avoid the stress concentration and the fracture risk. If bone resorption occurs, even minimal, the fracture risk increases noticeably. The hydrothermal degradation won't play a relevant role on the mechanical response of the YTZP implants. Acknowledgments We thank Klockner Implant System for funding this study and providing the test implants. References [1] P. Mohan, B. Yuan, T. Patterson, V.H. Desai, Y.H. Sohn, J. Am. Ceram. Soc. 90 (2007) [2] L. Jyung-Dong, D. Jenq-Gong, Mat. Chem. Phys. 78 (2002) 253. [3] C. Piconi, P. Maccauro, Biomaterials. 20 (1999) 1. [4] H.J. Wenz, J. Bartsch, S. Wolfart, M. Kern, Int. J. Prosthodont. 21 (2008) 27. [5] J. Chevalier, Biomaterials. 27 (2006) 535. [6] S. Deville, L. Gremillard, J. Chevalier, G. Fantozzi, J. Biomed. Mater. Res. 72B (2005) 239. [7] R.C. Garvie, P.S. Nicholson, J. Am. Ceram. Soc. 55 (1972). [8] G.M. Pharr, Mater. Sci. Eng., A 253 (1998) 151. [9] S. Bec, A. Tonck, J-M. Georges, E. Georges, J-L. Loubet, Philos. Mag. A. 74 (1996) [10] ISO 14801: Dentistry Implants Dynamic fatigue test for endosseous dental implants, [11] H. Tada, P.C. Paris, G.R. Irwin, The stress analysis of cracks handbook, Third edition, ASME Press, St Louis, 1985, p [12] S. Deville, J. Chevalier, L. Gremillard, Biomaterials. 27 (2006) [13] R.A. Kimel, J.H. Adair, J. Am. Ceram. Soc. 85 (2002) [14] L. Gremillard, J. Chevalier, T. Epicier, S. Deville, G. Fantozzi, J. Eur. Ceram. Soc. 24 (2004) [15] A. Feder, M. Anglada, J. Eur. Ceram. Soc. 25 (2005) [16] E. Jiménez-Piqué, L. Ceseracciu, Y. Gaillard, M. Barch, G. De Portu, M. Anglada, Phil. Mag. 86 (2006) [17] M. Fernández-Fairén, P. Sala, F.J. Gil, Biomed. Mater. Eng. 16 (2006) 415. [18] A.H. Glassman, J.D. Bobyn, M. Tanzer, Clin. Orthop. Relat. Res. 453 (2006) 64. [19] A. Gefen, Med. Biol. Eng. Comput. 40 (2002) 311. [20] D. Casellas, L. Llanes, M. Anglada, Fatigue of Yttria-stabilised zirconia ceramics: Influence of microstructure, Proceedings of the seventh international fatigue congress, Beijing, 1999, p [21] P. Zhu, L. Zhongqin, C. Guanlong, I. Kiyohiko. Int. J. Fatigue. 26 (2004) [22] S.-Y. Liu, I-W. Chen, J. Am. Ceram. Soc. 74 (1991) [23] S.-Y. Liu, I-W. Chen, J. Am. Ceram. Soc. 74 (1991) [24] S. Kulmer, A. Ruzicka, A. Niederwanger, I. Moschen, J. Oral. Rehabil. 26 (1999) 650.

Dentistry Fatigue test for endosseous dental implants. Report No.: 1 ASTRA System Ti Design

Dentistry Fatigue test for endosseous dental implants. Report No.: 1 ASTRA System Ti Design Dentistry Fatigue test for endosseous dental implants Report No.: 1 ASTRA System Ti Design Client: NT - Trading Time of execution: Mai 2011 Subject of work: - - S - 24285 Abutment Ti DesignTM - RESULTS

More information

ALL-CERAMIC DENTAL IMPLANT SOLUTIONS

ALL-CERAMIC DENTAL IMPLANT SOLUTIONS ALL-CERAMIC DENTAL IMPLANT SOLUTIONS Scientific Evidence Bone-to-implant contact of 78% at 3 months. One piece implant = no prosthetic connections References 1. One-year follow-up of first consecutive

More information

Bending, Forming and Flexing Printed Circuits

Bending, Forming and Flexing Printed Circuits Bending, Forming and Flexing Printed Circuits John Coonrod Rogers Corporation Introduction: In the printed circuit board industry there are generally two main types of circuit boards; there are rigid printed

More information

Structural Integrity Analysis

Structural Integrity Analysis Structural Integrity Analysis 1. STRESS CONCENTRATION Igor Kokcharov 1.1 STRESSES AND CONCENTRATORS 1.1.1 Stress An applied external force F causes inner forces in the carrying structure. Inner forces

More information

Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope

Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope Numerical Analysis of Independent Wire Strand Core (IWSC) Wire Rope Rakesh Sidharthan 1 Gnanavel B K 2 Assistant professor Mechanical, Department Professor, Mechanical Department, Gojan engineering college,

More information

ZIRCONIA INFORMATION FOR DENTISTS. FAQ S from the dental practice ENGLISH

ZIRCONIA INFORMATION FOR DENTISTS. FAQ S from the dental practice ENGLISH ZIRCONIA INFORMATION FOR DENTISTS FAQ S from the dental practice ENGLISH Zr Facts and Answers 2 Facts and Answers What is Zirconia? Zirconium (ZrSiO ) belongs to the mineral group of silicates and 4 was

More information

Tensile fracture analysis of blunt notched PMMA specimens by means of the Strain Energy Density

Tensile fracture analysis of blunt notched PMMA specimens by means of the Strain Energy Density Engineering Solid Mechanics 3 (2015) 35-42 Contents lists available at GrowingScience Engineering Solid Mechanics homepage: www.growingscience.com/esm Tensile fracture analysis of blunt notched PMMA specimens

More information

Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y.

Ando A., Nakamura Y., Kanbara R., Kumano H., Miyata T., Masuda T., Ohno Y. and Tanaka Y. 11. The Effect of Abutment Tooth Connection with Extracoronal Attachment using the Three Dimensional Finite Element Method - Part 2. The Construction of Finite Element Model from CT Data - Ando A., Nakamura

More information

September 13-16, 2009

September 13-16, 2009 September 13-16, 2009 Waikoloa, HI September 13-16, 2009, Waokoloa, Waikoloa, HI Biomimetic i Tooth Implants by Electron Beam Melting Mari Koike, DDS, PhD 1 ; Gilbert Chahine 2 ; Radovan Kovacevic, PhD

More information

Stress and deformation of offshore piles under structural and wave loading

Stress and deformation of offshore piles under structural and wave loading Stress and deformation of offshore piles under structural and wave loading J. A. Eicher, H. Guan, and D. S. Jeng # School of Engineering, Griffith University, Gold Coast Campus, PMB 50 Gold Coast Mail

More information

LABORATORY EXPERIMENTS TESTING OF MATERIALS

LABORATORY EXPERIMENTS TESTING OF MATERIALS LABORATORY EXPERIMENTS TESTING OF MATERIALS 1. TENSION TEST: INTRODUCTION & THEORY The tension test is the most commonly used method to evaluate the mechanical properties of metals. Its main objective

More information

Nanotribology of Hard Thin Film Coatings: A Case Study Using the G200 Nanoindenter

Nanotribology of Hard Thin Film Coatings: A Case Study Using the G200 Nanoindenter Nanotribology of Hard Thin Film Coatings: A Case Study Using the G200 Nanoindenter Application Note Introduction Thin film coatings are extensively used in various applications, such as microelectronics,

More information

Modelling of Contact Problems of Rough Surfaces

Modelling of Contact Problems of Rough Surfaces Modelling of Contact Problems of Rough Surfaces N. Schwarzer, Saxonian Institute of Surface Mechanics, Am Lauchberg 2, 04838 Eilenburg, Germany, Tel. ++49 (0) 3423 656639 or 58, Fax. ++49 (0) 3423 656666,

More information

The Study on Mechanical Performances of Thin SiO 2 Film by. Micro-Bridge Methods

The Study on Mechanical Performances of Thin SiO 2 Film by. Micro-Bridge Methods The Study on Mechanical Performances of Thin SiO Film by Micro-Bridge Methods WANG Yu 1, ZHANG Hai-ia 1*, ZHANG Tai-Hua 1 Institute of Microelectronics, Peking University State Key Laboratory of Nonlinear

More information

The atomic packing factor is defined as the ratio of sphere volume to the total unit cell volume, or APF = V S V C. = 2(sphere volume) = 2 = V C = 4R

The atomic packing factor is defined as the ratio of sphere volume to the total unit cell volume, or APF = V S V C. = 2(sphere volume) = 2 = V C = 4R 3.5 Show that the atomic packing factor for BCC is 0.68. The atomic packing factor is defined as the ratio of sphere volume to the total unit cell volume, or APF = V S V C Since there are two spheres associated

More information

Chapter Outline. Mechanical Properties of Metals How do metals respond to external loads?

Chapter Outline. Mechanical Properties of Metals How do metals respond to external loads? Mechanical Properties of Metals How do metals respond to external loads? Stress and Strain Tension Compression Shear Torsion Elastic deformation Plastic Deformation Yield Strength Tensile Strength Ductility

More information

Material data sheet. EOS CobaltChrome MP1. Description

Material data sheet. EOS CobaltChrome MP1. Description EOS CobaltChrome MP1 EOS CobaltChrome MP1 is a cobalt-chrome-molybdenum-based superalloy powder which has been optimized especially for processing on EOSINT M systems. This document provides information

More information

A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS

A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS A THREE-DIMENSIONAL FINITE ELEMENT ANALYSIS OF MANDIBULAR OVERDENTURE SUPPORTED BY CYLINDRICAL AND CONICAL IMPLANTS Evandro Afonso Sartori, evandroasartori@yahoo.com.br Josué Ricardo Broilo Luiz Oscar

More information

Metal Injection Molding (MIM) of components made of Titanium and its alloys

Metal Injection Molding (MIM) of components made of Titanium and its alloys Metal Injection Molding (MIM) of components made of Titanium and its alloys 1 Presentation content Introduction to Metal Injection Molding (MIM) Technology - explaination Products - examples Company brief

More information

different levels, also called repeated, alternating, or fluctuating stresses.

different levels, also called repeated, alternating, or fluctuating stresses. Fatigue and Dynamic Loading 1 Fti Fatigue fil failure: 2 Static ti conditions : loads are applied gradually, to give sufficient i time for the strain to fully develop. Variable conditions : stresses vary

More information

Surface Treatment of Titanium

Surface Treatment of Titanium Surface Treatment of Titanium Christiane Jung KKS Ultraschall AG, Steinen SZ The Tribology of Precision Swiss Tribology Meeting SBB Centre Löwenberg in Murten, Switzerland 16. November 2007 1 Outline:

More information

Numerical modelling of shear connection between concrete slab and sheeting deck

Numerical modelling of shear connection between concrete slab and sheeting deck 7th fib International PhD Symposium in Civil Engineering 2008 September 10-13, Universität Stuttgart, Germany Numerical modelling of shear connection between concrete slab and sheeting deck Noémi Seres

More information

Analysis of Slotted Counter Sunk (35NCD16 Steel) Aerospace Fasteners

Analysis of Slotted Counter Sunk (35NCD16 Steel) Aerospace Fasteners Analysis of Slotted Counter Sunk (35NCD16 Steel) Aerospace Fasteners A R Abelin Roy Deptt. of ME, Govt. Engineering College, Thrissur, India Christopher Solomon S MMD VSSC, ISRO Thiruvananthapuram, India

More information

Implants of high-performance ceramic. vk-lange 12.03.2007 14:44 Uhr Seite 1. One-piece precision implant. Made in Germany

Implants of high-performance ceramic. vk-lange 12.03.2007 14:44 Uhr Seite 1. One-piece precision implant. Made in Germany OMNIS Implants of high-performance ceramic vk-lange 12.03.2007 14:44 Uhr Seite 1 One-piece precision implant C M Y CM MY CY CMY K Dieter Lange Verkaufsleiter Labor Made in Germany creamed nge 12.03.2007

More information

Information on the product

Information on the product Information on the product Information on the product -Implantat.doc Page 1 of 15 This document composes of the following items 1. Name of the product 2. Raw materials 3. Structure and dimensions 3.1.

More information

ANALYTICAL AND EXPERIMENTAL EVALUATION OF SPRING BACK EFFECTS IN A TYPICAL COLD ROLLED SHEET

ANALYTICAL AND EXPERIMENTAL EVALUATION OF SPRING BACK EFFECTS IN A TYPICAL COLD ROLLED SHEET International Journal of Mechanical Engineering and Technology (IJMET) Volume 7, Issue 1, Jan-Feb 2016, pp. 119-130, Article ID: IJMET_07_01_013 Available online at http://www.iaeme.com/ijmet/issues.asp?jtype=ijmet&vtype=7&itype=1

More information

VeMet, Utrecht, NL «Solution in Wear Protection» 26.10.2011 Dipl.-Ing. Wolfgang Leichnitz. Quit

VeMet, Utrecht, NL «Solution in Wear Protection» 26.10.2011 Dipl.-Ing. Wolfgang Leichnitz. Quit VeMet, Utrecht, NL «Solution in Wear Protection» 26.10.2011 Dipl.-Ing. Wolfgang Leichnitz Quit Theory and Practice of Wear Definition In materials science, wear is the erosion of material from a solid

More information

COMPARATION BETWEEN PIII SUPERFICIAL TREATMENT AND CERAMIC COATING IN CREEP TEST OF Ti-6Al-4V ALLOY

COMPARATION BETWEEN PIII SUPERFICIAL TREATMENT AND CERAMIC COATING IN CREEP TEST OF Ti-6Al-4V ALLOY Materials Science Forum Vols. 660-661 (2010) pp 225-228 Online available since 2010/Oct/25 at www.scientific.net (2010) Trans Tech Publications, Switzerland doi:10.4028/www.scientific.net/msf.660-661.225

More information

TIE-31: Mechanical and thermal properties of optical glass

TIE-31: Mechanical and thermal properties of optical glass PAGE 1/10 1 Density The density of optical glass varies from 239 for N-BK10 to 603 for SF66 In most cases glasses with higher densities also have higher refractive indices (eg SF type glasses) The density

More information

COATED CARBIDE. TiN. Al 2 O 3

COATED CARBIDE. TiN. Al 2 O 3 COATED CARBIDE GENERAL INFORMATION CVD = Chemical Vapour Deposition coated grades GC2015, GC2025, GC2135, GC235, GC3005, GC3015, GC3020, GC3025, GC3115, GC4015, GC4025, GC4035, S05F, and CD1810. PVD =

More information

Computer Aided Engineering (CAE) Techniques Applied To Hip Implant

Computer Aided Engineering (CAE) Techniques Applied To Hip Implant International Journal Of Computational Engineering Research (ijceronline.com) Vol. 3 Issue. 3 Computer Aided Engineering (CAE) Techniques Applied To Hip Implant 1, M. S. Abo_Elkhair, 2, M. E. Abo-Elnor,

More information

BIOACTIVE COATINGS ON 316L STAINLESS STEEL IMPLANTS

BIOACTIVE COATINGS ON 316L STAINLESS STEEL IMPLANTS Trends Biomater. Artif. Organs. Vol. 17(2) pp 43-47 (2004) http//www.sbaoi.org BIOACTIVE COATINGS ON 316L STAINLESS STEEL IMPLANTS N. Ramesh Babu*,+, Sushant Manwatkar*, K. Prasada Rao* and T. S. Sampath

More information

Description of mechanical properties

Description of mechanical properties ArcelorMittal Europe Flat Products Description of mechanical properties Introduction Mechanical properties are governed by the basic concepts of elasticity, plasticity and toughness. Elasticity is the

More information

THE COMPOSITE DISC - A NEW JOINT FOR HIGH POWER DRIVESHAFTS

THE COMPOSITE DISC - A NEW JOINT FOR HIGH POWER DRIVESHAFTS THE COMPOSITE DISC - A NEW JOINT FOR HIGH POWER DRIVESHAFTS Dr Andrew Pollard Principal Engineer GKN Technology UK INTRODUCTION There is a wide choice of flexible couplings for power transmission applications,

More information

Outline of a quality system and standard for the certification of conformity

Outline of a quality system and standard for the certification of conformity Outline of a quality system and standard for the certification of conformity Although still in early steps, novel Additive Manufacturing (AM) processes are growing in expectations as enabling technology

More information

A R C H I V E S O F M E T A L L U R G Y A N D M A T E R I A L S Volume 57 2012 Issue 3 DOI: 10.2478/v10172-012-0095-3

A R C H I V E S O F M E T A L L U R G Y A N D M A T E R I A L S Volume 57 2012 Issue 3 DOI: 10.2478/v10172-012-0095-3 A R C H I V E S O F M E T A L L U R G Y A N D M A T E R I A L S Volume 57 2012 Issue 3 DOI: 10.2478/v10172-012-0095-3 K. TOPOLSKI, H. GARBACZ, P. WIECIŃSKI, W. PACHLA, K.J. KURZYDŁOWSKI MECHANICAL PROPERTIES

More information

Technical Data. 7. Bearing Fits. 7.1 Interference. 7.2 Calculation of interference F B LLLLLLLLL( A-54

Technical Data. 7. Bearing Fits. 7.1 Interference. 7.2 Calculation of interference F B LLLLLLLLL( A-54 Technical Data 7. Bearing Fits 7.1 Interference For rolling s the rings are fixed on the or in the housing so that slip or movement does not occur between the mated surface during operation or under. This

More information

UDDEHOLM VANADIS 30 SUPERCLEAN

UDDEHOLM VANADIS 30 SUPERCLEAN UDDEHOLM VANADIS 30 SUPERCLEAN UDDEHOLMS AB No part of this publication may be reproduced or transmitted for commercial purposes without permission of the copyright holder. This information is based on

More information

Short and Large Crack, Mixed-Mode Fatigue-Crack Growth Thresholds in Ti-6Al-4V 1

Short and Large Crack, Mixed-Mode Fatigue-Crack Growth Thresholds in Ti-6Al-4V 1 Short and Large Crack, Mixed-Mode Fatigue-Crack Growth Thresholds in Ti-6Al-4V 1 Mr. R. K. Nalla, Dr. J. P. Campbell and Professor R. O. Ritchie 2 Department of Materials Science and Engineering University

More information

Experimental assessment of concrete damage due to exposure to high temperature and efficacy of the repair system

Experimental assessment of concrete damage due to exposure to high temperature and efficacy of the repair system MATEC Web of Conferences 6, 06002 (2013) DOI: 10.1051/matecconf/20130606002 C Owned by the authors, published by EDP Sciences, 2013 Experimental assessment of concrete damage due to exposure to high temperature

More information

INVESTIGATION OF MECHANICAL PROPERTIES OF POLYESTER REINFORCED WITH HEMP FIBRE (LONG FIBRES AND MAT) FOR PATELLAR IMPLANT APPLICATION

INVESTIGATION OF MECHANICAL PROPERTIES OF POLYESTER REINFORCED WITH HEMP FIBRE (LONG FIBRES AND MAT) FOR PATELLAR IMPLANT APPLICATION International Journal of Mechanical Engineering and Technology (IJMET) Volume 7, Issue 1, Jan-Feb 2016, pp. 154-163, Article ID: IJMET_07_01_016 Available online at http://www.iaeme.com/ijmet/issues.asp?jtype=ijmet&vtype=7&itype=1

More information

Microscopy and Nanoindentation. Combining Orientation Imaging. to investigate localized. deformation behaviour. Felix Reinauer

Microscopy and Nanoindentation. Combining Orientation Imaging. to investigate localized. deformation behaviour. Felix Reinauer Combining Orientation Imaging Microscopy and Nanoindentation to investigate localized deformation behaviour Felix Reinauer René de Kloe Matt Nowell Introduction Anisotropy in crystalline materials Presentation

More information

FEM analysis of the forming process of automotive suspension springs

FEM analysis of the forming process of automotive suspension springs FEM analysis of the forming process of automotive suspension springs Berti G. and Monti M. University of Padua, DTG, Stradella San Nicola 3, I-36100 Vicenza (Italy) guido.berti@unipd.it, manuel.monti@unipd.it.

More information

The Mechanical Properties of Glass

The Mechanical Properties of Glass The Mechanical Properties of Glass Theoretical strength, practical strength, fatigue, flaws, toughness, chemical processes Glass Engineering 150:312 Professor Richard Lehman Department of Ceramics and

More information

EFFECT OF HARDNESS VARIATION ON SURFACE INTEGRITY OF CARBURIZED P20 STEEL

EFFECT OF HARDNESS VARIATION ON SURFACE INTEGRITY OF CARBURIZED P20 STEEL Copyright 2013 by ABCM EFFECT OF HARDNESS VARIATION ON SURFACE INTEGRITY OF CARBURIZED P20 STEEL Franciele Litvin franciele-litvin@hotmail.com Larissa França Madeira Manfrinato lari_manfrinato@hotmail.com

More information

Figure 1: Typical S-N Curves

Figure 1: Typical S-N Curves Stress-Life Diagram (S-N Diagram) The basis of the Stress-Life method is the Wohler S-N diagram, shown schematically for two materials in Figure 1. The S-N diagram plots nominal stress amplitude S versus

More information

North American Stainless

North American Stainless North American Stainless Long Products Stainless Steel Grade Sheet 2205 UNS S2205 EN 1.4462 2304 UNS S2304 EN 1.4362 INTRODUCTION Types 2205 and 2304 are duplex stainless steel grades with a microstructure,

More information

SEISMIC RETROFITTING TECHNIQUE USING CARBON FIBERS FOR REINFORCED CONCRETE BUILDINGS

SEISMIC RETROFITTING TECHNIQUE USING CARBON FIBERS FOR REINFORCED CONCRETE BUILDINGS Fracture Mechanics of Concrete Structures Proceedings FRAMCOS-3 AEDIFICA TIO Publishers, D-79104 Freiburg, Germany SEISMIC RETROFITTING TECHNIQUE USING CARBON FIBERS FOR REINFORCED CONCRETE BUILDINGS H.

More information

9. TIME DEPENDENT BEHAVIOUR: CYCLIC FATIGUE

9. TIME DEPENDENT BEHAVIOUR: CYCLIC FATIGUE 9. TIME DEPENDENT BEHAVIOUR: CYCLIC FATIGUE A machine part or structure will, if improperly designed and subjected to a repeated reversal or removal of an applied load, fail at a stress much lower than

More information

Contact: Steve Hurson VP, R & D Nobel Biocare 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 (714) 282-5072 steve.hurson@nobelbiocare.

Contact: Steve Hurson VP, R & D Nobel Biocare 22715 Savi Ranch Pkwy Yorba Linda, CA 92887 (714) 282-5072 steve.hurson@nobelbiocare. Use of CP Titanium and Titanium alloys for Dental Implants Steve Hurson, Nobel Biocare U.S.A. Yorba Linda CA Abstract Professor Per Ingvar Branemark of Sweden published his results on 15 years of research

More information

FRETTING FATIGUE OF STEELS WITH IFFERENT STRENGTH

FRETTING FATIGUE OF STEELS WITH IFFERENT STRENGTH FRETTING FATIGUE OF STEELS WITH IFFERENT STRENGTH Václav LINHART, Martin ČIPERA, Dagmar MIKULOVÁ SVÚM, a.s., Podnikatelská 565, 190 11 Praha 9- Běchovice,Czech Republic Abstract The investigation of fretting

More information

Welcome to the World of High Performance Ceramics

Welcome to the World of High Performance Ceramics Welcome to the World of High Performance Ceramics Agenda Company Introduction BIOLOX BIOLOX Inteligent Ceramic Material Future Material Hot Topics New Applications for Ceramics - Medical Engineering Division

More information

Quality of PVC sewage pipes in the Netherlands

Quality of PVC sewage pipes in the Netherlands TNO report MT-RAP-2008-01066/mso2 Quality of PVC sewage pipes in the Netherlands De Rondom 1 P.O. Box 6235 5600 HE Eindhoven The Netherlands www.tno.nl T +31 40 265 00 00 F +31 40 265 03 01 info-ient@tno.nl

More information

Wear-contact analysis of hip implant coated with functionally graded material

Wear-contact analysis of hip implant coated with functionally graded material 2014; 1(7): 439-443 IJMRD 2014; 1(7): 439-443 www.allsubjectjournal.com Received: 13-12-2014 Accepted: 30-12-2014 e-issn: 2349-4182 p-issn: 2349-5979 Impact Factor: 3.762 Vinod Pallapu M. Tech Student

More information

Modal and Vibration Analysis of Functionally Graded Dental Implant

Modal and Vibration Analysis of Functionally Graded Dental Implant British Journal of Medicine & Medical Research 12(7): 1-14, 2016, Article no.bjmmr.22139 ISSN: 2231-0614, NLM ID: 101570965 SCIENCEDOMAIN international www.sciencedomain.org Modal and Vibration Analysis

More information

ANALYSIS OF THE FATIGUE STRENGTH UNDER TWO LOAD LEVELS OF A STAINLESS STEEL BASED ON ENERGY DISSIPATION

ANALYSIS OF THE FATIGUE STRENGTH UNDER TWO LOAD LEVELS OF A STAINLESS STEEL BASED ON ENERGY DISSIPATION EPJ Web of Conferences 6, 6 38009 (00) DOI:0.05/epjconf/000638009 Owned by the authors, published by EDP Sciences, 00 ANALYSIS OF THE FATIGUE STRENGTH UNDER TWO LOAD LEVELS OF A STAINLESS STEEL BASED ON

More information

Titanium versus Zirconium Implants

Titanium versus Zirconium Implants Titanium versus Zirconium Implants A dental materials comparison Dr. Gerald Regni, DMD!!" Christian St. Phila. PA 1213" T: (""") %&'-!"#" E: Dr.Gerald_Regni@yahoo.com Titanium versus Zirconium Implants

More information

Material data sheet. EOS Aluminium AlSi10Mg. Description

Material data sheet. EOS Aluminium AlSi10Mg. Description EOS Aluminium AlSi10Mg EOS Aluminium AlSi10Mg is an aluminium alloy in fine powder form which has been specially optimised for processing on EOSINT M systems This document provides information and data

More information

METHODS FOR ACHIEVEMENT UNIFORM STRESSES DISTRIBUTION UNDER THE FOUNDATION

METHODS FOR ACHIEVEMENT UNIFORM STRESSES DISTRIBUTION UNDER THE FOUNDATION International Journal of Civil Engineering and Technology (IJCIET) Volume 7, Issue 2, March-April 2016, pp. 45-66, Article ID: IJCIET_07_02_004 Available online at http://www.iaeme.com/ijciet/issues.asp?jtype=ijciet&vtype=7&itype=2

More information

Material data sheet. EOS Aluminium AlSi10Mg_200C. Description

Material data sheet. EOS Aluminium AlSi10Mg_200C. Description EOS Aluminium AlSi10Mg_200C All information in this data sheet refers to the alloy EOS Aluminium AlSi10Mg_200C. This alloy is formed when the powder EOS Aluminium AlSi10Mg is processes at a building platform

More information

MODELING THE EFFECT OF TEMPERATURE AND FREQUENCY ON BITUMEN-COATED HELICAL CABLE ELEMENTS

MODELING THE EFFECT OF TEMPERATURE AND FREQUENCY ON BITUMEN-COATED HELICAL CABLE ELEMENTS MODELING THE EFFECT OF TEMPERATURE AND FREQUENCY ON BITUMEN-COATED HELICAL CABLE ELEMENTS Bjørn Konradsen 1 Technological Analyses Centre, Hybrid Underwater Cables, Nexans Norway Steinar V. Ouren Material

More information

The CVD diamond booklet

The CVD diamond booklet available at: www.diamond-materials.com/download Content 1. General properties of diamond... 2 2. Optical Properties... 4 Optical transparency...4 Absorption coefficient at 10.6 µm...5 Refractive index:

More information

DETERMINATION OF TIME-TEMPERATURE SHIFT FACTOR FOR LONG-TERM LIFE PREDICTION OF POLYMER COMPOSITES

DETERMINATION OF TIME-TEMPERATURE SHIFT FACTOR FOR LONG-TERM LIFE PREDICTION OF POLYMER COMPOSITES DETERMINATION OF TIME-TEMPERATURE SHIFT FACTOR FOR LONG-TERM LIFE PREDICTION OF POLYMER COMPOSITES K. Fukushima*, H. Cai**, M. Nakada*** and Y. Miyano*** * Graduate School, Kanazawa Institute of Technology

More information

Jacket crown. Advantage : Crown and Bridge

Jacket crown. Advantage : Crown and Bridge Crown and Bridge Lecture 1 Dr.Nibras AL-Kuraine Jacket crown It is a type of crown that is formed by a tooth colored material. It is mainly used as a single unit in the anterior quadrant of the mouth.

More information

CH 6: Fatigue Failure Resulting from Variable Loading

CH 6: Fatigue Failure Resulting from Variable Loading CH 6: Fatigue Failure Resulting from Variable Loading Some machine elements are subjected to static loads and for such elements static failure theories are used to predict failure (yielding or fracture).

More information

Simulation of Residual Stresses in an Induction Hardened Roll

Simulation of Residual Stresses in an Induction Hardened Roll 2.6.4 Simulation of Residual Stresses in an Induction Hardened Roll Ludwig Hellenthal, Clemens Groth Walzen Irle GmbH, Netphen-Deuz, Germany CADFEM GmbH, Burgdorf/Hannover, Germany Summary A heat treatment

More information

A NE ALW L G T E H N ATE R G A LI TIO T N E R O S F DEN I T S A L N C OTE R G A O M L I D CS by Mark Gervais

A NE ALW L G T E H N ATE R G A LI TIO T N E R O S F DEN I T S A L N C OTE R G A O M L I D CS by Mark Gervais A NEW ALL THAT GENERATION GLITTERS OF DENTAL IS NOT CERAMICS GOLD by Mark Gervais The next 28 minutes Material classifications Bonded veneers & partial coverage restorations Conventional crowns & bridges

More information

Objectives. Experimentally determine the yield strength, tensile strength, and modules of elasticity and ductility of given materials.

Objectives. Experimentally determine the yield strength, tensile strength, and modules of elasticity and ductility of given materials. Lab 3 Tension Test Objectives Concepts Background Experimental Procedure Report Requirements Discussion Objectives Experimentally determine the yield strength, tensile strength, and modules of elasticity

More information

15. MODULUS OF ELASTICITY

15. MODULUS OF ELASTICITY Chapter 5 Modulus of Elasticity 5. MODULUS OF ELASTICITY The modulus of elasticity (= Young s modulus) E is a material property, that describes its stiffness and is therefore one of the most important

More information

NANO INDENTERS FROM MICRO STAR TECHNOLOGIES

NANO INDENTERS FROM MICRO STAR TECHNOLOGIES NANO INDENTERS FROM MICRO STAR TECHNOLOGIES Micro Star makes a variety of nano indenters following defined standards or custom requested geometries and dimensions. Micro Star calibration laboratory complies

More information

Measurement of Soil Parameters by Using Penetrometer Needle Apparatus

Measurement of Soil Parameters by Using Penetrometer Needle Apparatus Vol.3, Issue.1, Jan-Feb. 2013 pp-284-290 ISSN: 2249-6645 Measurement of Soil Parameters by Using Penetrometer Needle Apparatus Mahmoud M. Abu zeid, 1 Amr M. Radwan, 2 Emad A. Osman, 3 Ahmed M.Abu-bakr,

More information

Solid shape molding is not desired in injection molding due to following reasons.

Solid shape molding is not desired in injection molding due to following reasons. PLASTICS PART DESIGN and MOULDABILITY Injection molding is popular manufacturing method because of its high-speed production capability. Performance of plastics part is limited by its properties which

More information

Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied

Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied Mechanical Properties of Metals Mechanical Properties refers to the behavior of material when external forces are applied Stress and strain fracture or engineering point of view: allows to predict the

More information

TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION

TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION Richard A. Sweet, M.D. Louisville Orthopaedic Clinic Louisville, KY TOTAL HIP REPLACEMENT FOR A LIFETIME: THE CEMENTLESS METAL ON METAL RECONSTRUCTION INTRODUCTION Total hip replacement surgery (THR) has

More information

The mechanical behavior of the bone microstructure around dental implants

The mechanical behavior of the bone microstructure around dental implants BUDAPEST UNIVERSITY OF TECHNOLOGY AND ECONOMICS DEPARTMENT OF STRUCTURAL MECHANICS The mechanical behavior of the bone microstructure around dental implants Summary and Theses of PhD dissertation ÉVA LAKATOS

More information

NEW TECHNIQUE FOR RESIDUAL STRESS MEASUREMENT NDT

NEW TECHNIQUE FOR RESIDUAL STRESS MEASUREMENT NDT NEW TECHNIQUE FOR RESIDUAL STRESS MEASUREMENT NDT E. Curto. p.i. Ennio Curto Via E. di Velo,84 36100 Vicenza Tel. 0444-511819 E-mail enniocurto@fastwebnet.it Key words: NDE Residual stress. New technique

More information

Coating Thickness and Composition Analysis by Micro-EDXRF

Coating Thickness and Composition Analysis by Micro-EDXRF Application Note: XRF Coating Thickness and Composition Analysis by Micro-EDXRF www.edax.com Coating Thickness and Composition Analysis by Micro-EDXRF Introduction: The use of coatings in the modern manufacturing

More information

PROPERTIES OF MATERIALS

PROPERTIES OF MATERIALS 1 PROPERTIES OF MATERIALS 1.1 PROPERTIES OF MATERIALS Different materials possess different properties in varying degree and therefore behave in different ways under given conditions. These properties

More information

EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR PRODUCTION ON THE PIERCED FLAT SHEET METAL USING LASER FORMING PROCESS

EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR PRODUCTION ON THE PIERCED FLAT SHEET METAL USING LASER FORMING PROCESS JOURNAL OF CURRENT RESEARCH IN SCIENCE (ISSN 2322-5009) CODEN (USA): JCRSDJ 2014, Vol. 2, No. 2, pp:277-284 Available at www.jcrs010.com ORIGINAL ARTICLE EXPERIMENTAL AND NUMERICAL ANALYSIS OF THE COLLAR

More information

Why an Exactech Hip is Right for You

Why an Exactech Hip is Right for You Why an Exactech Hip is Right for You Why do I need a total hip replacement? Which surgical approach is best for me? How long will it last? Which implant is right for me? Founded in 1985 by an orthopaedic

More information

Testing and appraisal of Lucobit polymer effect as an additive on asphalt mixture performance

Testing and appraisal of Lucobit polymer effect as an additive on asphalt mixture performance Abstract Testing and appraisal of polymer effect as an additive on asphalt mixture performance Hamid Sabbagh mollahosseini*,golazin Yadollahi**, Ershad Amoosoltani*** *, ***Executive of Engineering and

More information

nanovea.com MECHANICAL TESTERS Indentation Scratch Wear

nanovea.com MECHANICAL TESTERS Indentation Scratch Wear MECHANICAL TESTERS Indentation Scratch Wear nanovea.com MECHANICAL TESTER INTRO Nanovea Mechanical Testers provide unmatched multi-function Nano and Micro/Macro modules on a single platform. Both the Nano

More information

Implant Systems MAX LOGIC MONO. Today s name in tomorrow s implant technology

Implant Systems MAX LOGIC MONO. Today s name in tomorrow s implant technology Implant Systems MAX LOGIC MONO Today s name in tomorrow s implant technology Implant Systems Etgar Medical Implant Systems brings to the market an entire spectrum of dental implant products based on its

More information

Hardened Concrete. Lecture No. 14

Hardened Concrete. Lecture No. 14 Hardened Concrete Lecture No. 14 Strength of Concrete Strength of concrete is commonly considered its most valuable property, although in many practical cases, other characteristics, such as durability

More information

TENSILE BEHAVIOUR OF OPEN CELL CERAMIC FOAMS

TENSILE BEHAVIOUR OF OPEN CELL CERAMIC FOAMS Original papers TENSILE BEHAVIOUR OF OPEN CELL CERAMIC FOAMS LUKÁŠ ŘEHOŘEK*, **, IVO DLOUHÝ*,**, ZDENĚK CHLUP* *Institute of Physics of Materials ASCR, Žižkova 22, 616 62 Brno, Czech Republic **Institute

More information

STUDY OF THE BEHAVIOUR OF BITUMINOUS MIXTURES RESISTANT TO FUEL

STUDY OF THE BEHAVIOUR OF BITUMINOUS MIXTURES RESISTANT TO FUEL STUDY OF THE BEHAVIOUR OF BITUMINOUS MIXTURES RESISTANT TO FUEL INTRODUCTION Evaluation of the performance of asphalt mixtures applied on wearing courses of road and airport pavements when subjected to

More information

International Journal of Engineering Research-Online A Peer Reviewed International Journal Articles available online http://www.ijoer.

International Journal of Engineering Research-Online A Peer Reviewed International Journal Articles available online http://www.ijoer. RESEARCH ARTICLE ISSN: 2321-7758 DESIGN AND DEVELOPMENT OF A DYNAMOMETER FOR MEASURING THRUST AND TORQUE IN DRILLING APPLICATION SREEJITH C 1,MANU RAJ K R 2 1 PG Scholar, M.Tech Machine Design, Nehru College

More information

on-metal Hips: Device Mechanics and Failure Modes

on-metal Hips: Device Mechanics and Failure Modes 1 Metal-on on-metal Hips: Device Mechanics and Failure Modes Steven M. Kurtz, Ph.D., and Richard Underwood, Ph.D. Exponent, Inc., and Drexel University NIH R01 AR47904 2 Contracts: DePuy Orthopaedics,

More information

USE OF SIMULATION TO IMPROVE CARDIOVASCULAR STENT DEVELOPMENT

USE OF SIMULATION TO IMPROVE CARDIOVASCULAR STENT DEVELOPMENT USE OF SIMULATION TO IMPROVE CARDIOVASCULAR STENT DEVELOPMENT V. Astier, S. Bidal, V. Garitey, F. Mouret Abstract : A stent is a wire metal mesh tube placed into an artery or blood vessel to hold the structure

More information

Phase Characterization of TiO 2 Powder by XRD and TEM

Phase Characterization of TiO 2 Powder by XRD and TEM Kasetsart J. (Nat. Sci.) 42 : 357-361 (28) Phase Characterization of TiO 2 Powder by XRD and TEM Kheamrutai Thamaphat 1 *, Pichet Limsuwan 1 and Boonlaer Ngotawornchai 2 ABSTRACT In this study, the commercial

More information

Phenomenological aspects of a modified fragmentation of the ground material

Phenomenological aspects of a modified fragmentation of the ground material Phenomenological aspects of a modified fragmentation of the ground material Lucjan Dabrowski, Mieczyslaw Marciniak Warsaw University of Technology, Warsaw, Poland Summary: The main point of this paper

More information

Dental Updates. Excerpted Article e-mail: re777@comcast.net. Why Implant Screws Loosen Part 1. Richard Erickson, MS, DDS

Dental Updates. Excerpted Article e-mail: re777@comcast.net. Why Implant Screws Loosen Part 1. Richard Erickson, MS, DDS ¼ ½ ¾ µ mw/cm 2 Volume 17; 2007 Dental Updates "CUTTING EDGE INFORMATION FOR THE DENTAL PROFESSIONAL " 200 SEMINARS AND 30 JOURNALS REVIEWED YEARLY FOR THE LATEST, CUTTING EDGE INFORMATION Excerpted Article

More information

Effect of Sleeve Shrink-fit on Bearing Preload of a Machine Tool Spindle: Analysis using Finite Element Method

Effect of Sleeve Shrink-fit on Bearing Preload of a Machine Tool Spindle: Analysis using Finite Element Method Effect of Sleeve Shrink-fit on Bearing Preload of a Machine Tool Spindle: Analysis using Finite Element Method Aslam Pasha Taj 1, Chandramouli SR 2* ACE Designers Limited, Peenya Industrial Area, Bangalore-560058

More information

2. Deposition process

2. Deposition process Properties of optical thin films produced by reactive low voltage ion plating (RLVIP) Antje Hallbauer Thin Film Technology Institute of Ion Physics & Applied Physics University of Innsbruck Investigations

More information

* This work is an official contribution of the National Institute of Standards and Technology and

* This work is an official contribution of the National Institute of Standards and Technology and Variability in the Geometric Accuracy of Additively Manufactured Test Parts A.L. Cooke and J.A. Soons National Institute of Standards and Technology * Gaithersburg, MD, USA Abstract This paper describes

More information

Simulation in design of high performance machine tools

Simulation in design of high performance machine tools P. Wagner, Gebr. HELLER Maschinenfabrik GmbH 1. Introduktion Machine tools have been constructed and used for industrial applications for more than 100 years. Today, almost 100 large-sized companies and

More information

EFFECT OF SEVERE PLASTIC DEFORMATION ON STRUCTURE AND PROPERTIES OF AUSTENITIC AISI 316 GRADE STEEL

EFFECT OF SEVERE PLASTIC DEFORMATION ON STRUCTURE AND PROPERTIES OF AUSTENITIC AISI 316 GRADE STEEL EFFECT OF SEVERE PLASTIC DEFORMATION ON STRUCTURE AND PROPERTIES OF AUSTENITIC AISI 316 GRADE STEEL Ladislav KANDER a, Miroslav GREGER b a MATERIÁLOVÝ A METALURGICKÝ VÝZKUM, s.r.o., Ostrava, Czech Republic,

More information

Machine Design II Prof. K.Gopinath & Prof. M.M.Mayuram. Module 2 - GEARS. Lecture 17 DESIGN OF GEARBOX

Machine Design II Prof. K.Gopinath & Prof. M.M.Mayuram. Module 2 - GEARS. Lecture 17 DESIGN OF GEARBOX Module 2 - GEARS Lecture 17 DESIGN OF GEARBOX Contents 17.1 Commercial gearboxes 17.2 Gearbox design. 17.1 COMMERCIAL GEARBOXES Various commercial gearbox designs are depicted in Fig. 17.1 to 17.10. These

More information

CRITERIA FOR PRELOADED BOLTS

CRITERIA FOR PRELOADED BOLTS National Aeronautics and Space Administration Lyndon B. Johnson Space Center Houston, Texas 77058 REVISION A JULY 6, 1998 REPLACES BASELINE SPACE SHUTTLE CRITERIA FOR PRELOADED BOLTS CONTENTS 1.0 INTRODUCTION..............................................

More information

Dental Implants. You are correct, it is a labeling error. Here is your dental prosthesis!!!

Dental Implants. You are correct, it is a labeling error. Here is your dental prosthesis!!! Dental Implants You are correct, it is a labeling error. Here is your dental prosthesis!!! Manufacturing process of bridges, crowns, bone implants, fixtures, abutments. Manufacturing process of bridges,

More information