Using HyperTerminal with Agilent General Purpose Instruments



Similar documents
To perform Ethernet setup and communication verification, first perform RS232 setup and communication verification:

Cable Pinouts. SRP I/O Module

How To Connect A Directsofl To A Powerpoint With An Acd With An Ctel With An Dm-Tel Modem On A Pc Or Ipad Or Ipa (Powerpoint) With A Powerline 2 (Powerline

Cable Specifications and Information

1.1 Connection Direct COM port connection. 1. Half duplex RS232 spy cable without handshaking

Accessing Diagnostics using a Dialup Modem

Connecting and Setting Up Your Laptop Computer

Modbus Communications for PanelView Terminals

Sending SMS Messages and s

2-Port RS232/422/485 Combo Serial to USB2.0 Adapter (w/ Metal Case and Screw Lock Mechanism) Installation Guide

IP SERIAL DEVICE SERVER

How to setup a serial Bluetooth adapter Master Guide

MODBUS TCP to RTU/ASCII Gateway. User s Manual

Programming and Using the Courier V.Everything Modem for Remote Operation of DDF6000

PM1122 INT DIGITAL INTERFACE REMOTE

Knowledge Base POS/C31A Troubleshooting

RJ45 Shielded (standard) port pinout. CS9000, Jetstream , Lanstream 2000, RTA8/RJX, RRC16, MTA8/RJX & SXDC8/RJX

SETTING UP A REMOTE MODEM CONNECTION Application Note 27

16-Port RS232 to USB2.0 High Speed Multi Serial Adapter (w/ Metal Case) Installation Guide

Introduction: Implementation of the MVI56-MCM module for modbus communications:

User s Manual TCP/IP TO RS-232/422/485 CONVERTER. 1.1 Introduction. 1.2 Main features. Dynamic DNS

Configuring Serial Terminal Emulation Programs

Virtual Integrated Design Getting started with RS232 Hex Com Tool v6.0

The Analyst RS422/RS232 Tester. With. VTR, Monitor, and Data Logging Option (LOG2) User Manual

L Firmware And FPGA Configuration Upgrade

RS-232 COMMUNICATIONS

Dial-Up / Leased-Line Modem. User Manual. AGM Electronics, Inc Dial-Up / Leased-Line Modem, Series ( ) Manual Rev A + - DLM CTS RTS DTR DSR

How to setup a serial Bluetooth adapter

Encore Controller to Router Connections

Instructions on How to Use HyperTerminal to Test Serial Ports ...

SUDT AccessPort TM Advanced Terminal / Monitor / Debugger Version 1.37 User Manual

In order to print in ESC/P mode, the serial connection should be activated.

One Port Serial Server Users Manual Model ESP901, ESP901E

MCB3101 (Class I) WiRobot Serial Bluetooth Wireless Module User Manual

Date Rev. Details Author

Connecting the Console Port to a PC

LAN / WAN Connection Of Instruments with Serial Interface By Using a Terminal Server

Monitor Program (Windows version) Instruction Manual

CONCEPT1 RS232 COMMUNICATION

Aquadyne TechTip TITLE: TROUBLESHOOTING PC COM PORT AND MODEM PRODUCTS AFFECTED SYMPTOMS POSSIBLE CAUSES

TRP-C31M MODBUS TCP to RTU/ASCII Gateway

Using IDENT M System T with Modbus/TCP

RS232 Board datasheet

Toshiba Serial Driver Help Kepware Technologies

USB TO SERIAL ADAPTER

Professional USB to Serial Adapter Hub with COM Retention

Application Note 2. Using the TCPDIAL & TCPPERM Commands to Connect Two TransPort router Serial Interfaces Over TCP/IP.

Connecting the DG-102S VoIP Gateway to your network

Data Bulletin. Communications Wiring for POWERLINK G3 Systems Class 1210 ABOUT THIS BULLETIN APPLICATION INTRODUCTION.

Stevens Cell-Net CDMA & GSM M2M 3G Data Modem. Product Manual

Serial to Bluetooth Adapter

Fanuc 0 M/T Model C Serial (RS232) Connection Guide

Application Note AN_242

RS232C < - > RS485 CONVERTER S MANUAL. Model: LD15U. Phone: / 97 / 98 (M)

LOW COST GSM MODEM. Description. Part Number

Single channel data transceiver module WIZ2-434

PCMCIA 1 Port RS EDITION OCTOBER 1999

How-To Set Custom Scale Format for FedEx Ship Manager

Technical Manual. For use with Caller ID signaling types: Belcore 202, British Telecom, & ETSI

s!nus-elektrotechnikai bt. Industrial IT & Automation

32 Transferring Projects and Data

Bluetooth UART/RS232 Module

Quick Note 32. Using Digi RealPort with a Digi TransPort Router. UK Support September 2012

Cabling Guide for Console and AUX Ports

MAGELLAN CONNECTING TO CASH REGISTER EXPRESS

TCP/IP Converter DDS EX-9132 Operation Manual for 8051 Series

ISHIDA BC Scale to Scale Communications

LS-101 LAN to Serial Device server. User s Manual

USB-to-Serial RS-232 Hub USB-to-Serial RS-422/485 Hub USER MANUAL UC2322 / UC2324 / UC4852 / UC4854

2-port RS-232 High Speed Low Profile Universal PCI Serial Board

KSH-3000-MI. Swipe Card Reader. ( Magnetic / IC Card two in One Reader ) Revision Reason Date V

Why you need to monitor serial communication?

KOYO DIRECT. HMI Setting: PLC Setting: Supported Series: KOYO DirectLogic series PLC DL05, DL06, DL105, DL205, DL305, and DL405 series.

SMS Alarm Messenger. Setup Software Guide. SMSPro_Setup. Revision [Version 2.2]

ModBus Server - KNX. Gateway for integration of KNX equipment into Modbus (RTU and TCP) control systems.

Liebert IntelliSlot Web Cards

T3 Mux M13 Multiplexer

CDI-S100 SERIAL INTERFACE CARD

PC Utility User s Manual

With a barcode reader connected to the printer, labels and barcodes can easily be printed by scanning barcodes.

Inmarsat B Application Note

POS Integration. Prepared by: Binh Nguyen

Using a Laptop Computer with a USB or Serial Port Adapter to Communicate With the Eagle System

Cypress CY7C64225 USB-to-UART Setup Guide Version 1.3

How To Set Up Aport 211 On A Pc Or Mac Or Ipa (For Microsoft) With Aport-211 On A Microsoft Mac Or Mac (For Mac) With An Ipa Or Ipo (For Pc Or Ipom

Auxiliary equipment for communication with the EuroProt protective devices

RN-WIFLY-EVAL-UM. WiFly Evaluation Kit Roving Networks. All rights reserved. RN-WIFLY-EVAL-UM Version 1.32r 10/9/2012 USER MANUAL

Keep it Simple Timing

Table of Contents. Creating a VC1000 Network... 3

TCP/IP MODULE CA-ETHR-A INSTALLATION MANUAL

UDS-10 External Network Interface Setup. User's Guide

HP Point of Sale (POS) Peripherals Configuration Guide ap5000 VFD Windows (non-opos)

Using Bluetooth-Enabled PosiTector 6000 with Statistical Process Control Software

MAX T1/E1. Quick Start Guide. VoIP Gateway. Version 1.0

SEC 4100 SYSTEM MONITOR Programmable Serial Interface Card Series 2

Cable Specifications and Information

IPG/7700 Hardware Manual SYSTECH. Document number Revision A

Docklight V2.1 User Manual 08/2015. Copyright 2015 Flachmann und Heggelbacher GbR

Quectel Cellular Engine

Making a DB to RJ45 adapter.

Transcription:

Using HyperTerminal with Agilent General Purpose Instruments Windows HyperTerminal can be used to program most General Purpose Instruments (not the 531xx series counters) using the RS-232 Serial Bus. Instrument and HyperTerminal Setup Procedure 1. In order to use the instrument with RS-232, the correct cable must be used. It is a common mistake that any 9-pin RS-232 cable can be used. This is not true. Most Agilent instruments use the Null modem cable that is shipped with the instrument. If unsure about the cable, check the connections, using an ohmmeter, to the connections shown on page 6 in this document or in the instrument's User's manual. 2. Setup the instrument to the factory default RS-232 settings. Refer to the instrument User's manual for the settings. The settings for some typical instruments are as follows: Agilent 34970A Data Acquisition/Switch Unit: Baud Rate: 57600 (selectable) Data bits: 8 (selectable) Parity: None (selectable) Stop bits: 1 (fixed) Flow control: XON/XOFF Agilent 34401A Digital Multimeter: Baud Rate: 9600 (selectable) Data bits: 8 (selectable) Parity: None (selectable) Stop bits: 2 (fixed) Flow control: DTR/DSR (fixed) Agilent 33120A Function Generator/Arbitrary Waveform Generator: Baud Rate: 9600 (selectable) Data bits: 8 (selectable; use this setting to download arbitrary waveforms) Parity: None (selectable; use this setting to download arbitrary waveforms) Stop bits: 2 (fixed) Flow control: DTR/DSR (fixed) Agilent 33250A Function Generator/Arbitrary Waveform Generator: Baud Rate: 57600 (selectable) Data bits: 8 (selectable; use this setting to download arbitrary waveforms) Parity: None (selectable; use this setting to download arbitrary waveforms) Stop bits: 1 (fixed) Flow control: DTR/DSR (use any flow control, except XON/XOFF, to download arbitrary waveforms) For other Agilent instruments not listed here, refer to the instrument's User's Manual. Page 1

3. Select HyperTerminal; use the instructions for your operating system to select it. 4. After HyperTerminal is enabled, select the communication with the instrument, as follows: A. Enter a connection name B. Click on "OK" 5. Select the COM port, as follows: A. Select the port (COM1, COM, etc.) B. Click on "OK" Page 2

6. Set the COM port parameters. The following assumes that COM1 port is used and HyperTerminal is set to the 34970A's default configuration. A. Select the parameters B. Click on "OK" 7. Setup the HyperTerminal, as follows: A. Click on "File" C. Click on "Settings" tab B. Click on "Properties" Page 3

D. Click on "ASCII Setup" E. Select "Send line ends with line feeds" and Select "Echo typed character locally" F. Click on "OK" G. Click on "OK" Page 4

8. Type in "*IDN?" to make sure the instrument returns the ID string, as follows: Type in "*IDN? And then press the "Enter" button (on the main keyboard, not the number pad) Typical returned ID string Storing Returned Data/Readings into a Text File Returned data from an instrument can be stored into a text file using HyperTerminal. For example, the "FETCh?" command used by the 34970A and other instruments, returns the readings from instrument memory. During the time the data is placed into the HyperTerminal window, it can automatically be stored into a text file. The following procedure shows how to do that. Setup the HyperTerminal to transfer the data to a file, as follows: A. Click on "Transfer" B. Click on "Capture Text" C. Select a Folder and File Name D. Click on "Start" Page 5

E. Type in the "FETCh?" command, or any other command that is appropriate for your instrument The returned readings are displayed and also stored into the test file. Typical RS-232 Cable for General Purpose Instruments The following shows the cable used for most of the Agilent General Purpose Instruments, like the 33120A, 33250A, 34401A, and 34970A. These instruments use a cable that can be ordered from Agilent's Parts Ordering department using part number: RS232-61601. For other instruments, check in the user's manual for the part number and connections. RS-232 Cable Instrument Port Computer Port DCD RX TX DTR GND DSR RTS CTS RI 1 2 3 4 5 6 7 8 9 1 2 3 4 5 6 7 8 9 DCD RX TX DTR GND DSR RTS CTS RI Male Female Female Male Page 6