OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD

Similar documents
REAL-TIME DUST AND AEROSOL MONITORING

CONTAMINATION CONTROL FOR ELECTRONIC AND SEMICONDUCTOR APPLICATIONS

HEATED INLET SAMPLE CONDITIONING EFFECTS ON REMOTE DUST MONITORING

FITPRO+ SOFTWARE V3 IMPORTING DATA FROM FITPRO V2 & FITPLUS3

APPLICATION PROGRAM INTERFACE (API) FOR AIRASSURE WEB PM2.5 SOFTWARE

NEW APPROACHES AND TECHNOLOGIES IN PERSONAL SAMPLING PUMPS

TSI BIOTRAK REAL-TIME VIABLE PARTICLE COUNTER FACILITY MONITORING SYSTEMS

Printing Fit Test Cards with an ID Card Printer

AEROSOL STATISTICS LOGNORMAL DISTRIBUTIONS AND dn/dlogd p

THE LEADER IN PERFORMANCE INDOOR AIR QUALITY MEASUREMENTS

PERCENT OUTDOOR AIR (%OA) CALCULATION AND ITS USE

Indoor air quality monitoring made easy

ab vacuum manifold for 96 well filter plates This product is for research use only and is not intended for diagnostic use.

PREMIUM CLEAN ROOM MONITOR MODELS 8630-CRM-S 8630-CRM-P

Particle testing in cleanroom high-pressure gas lines to ISO made easy with the MET ONE 3400 gas calibrations

Export. Exporting Client Addresses

How To Install Safari Antivirus On A Dv8000 Dv Recorder On A Pc Or Macbook Or Ipad (For A Pc) On A Microsoft Dv8 (For Macbook) On An Ipad Or Ipa (


Oilfield Services Measurement Solutions

PARTICLE GENERATOR MODEL 8026

Sub-100 nm calibration for size and number

Coarse PM Methods Evaluation Study

Small Animal Respiratory Exchange Ratio (RER)

Integrating CaliberRM with Software Configuration Management Tools

Density Measurement. Technology: Pressure. Technical Data Sheet INTRODUCTION. S min =1.0 S max =1.2 CONSTANT LEVEL APPLICATIONS

Technical Cleaning Ó

Taking Attendance in Salesforce with the Attendance Tracker

XCal-View user manual

How to operate the BD FACSCanto flow cytometer

LOAD BALANCING 2X APPLICATIONSERVER XG SECURE CLIENT GATEWAYS THROUGH MICROSOFT NETWORK LOAD BALANCING

Calibration Certificate

ILLUMINATE ASSESSMENT REPORTS REFERENCE GUIDE

FACILITY MONITORING SYSTEM DESIGN RECOMMENDATIONS

How to send s triggered by events

Chemical Measurement Solutions

SYSTEC Degasser WatchDog

Sample preparation for X-ray fluorescence analysis

2009 Braton Groupe sarl, All rights reserved.

Environmental Monitoring

Application Note #231 Revision B November Using a VPN Connection to Remotely Access Lutron Systems. Overview. Key Terms. Before You Begin...

DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL

Snap 9 Professional s Scanning Module

Image Lab Software for the GS-900 Densitometer

Trimble CFX-750 Display with the Yield Monitoring System

INPEX CONTRACT MANAGEMENT SYSTEM


Quick Start Guide Using OneDisk with the Tappin Service

Scanning Mobility Particle Sizer

Weight Loss Determined from Mass Spectrometry Trend Data in a Thermogravimetric/Mass Spectrometer System

FiberOptix TM. AP Sensor Zeroing and Calibration Guide INTRA-AORTIC BALLOON PRODUCT SUPPORT HOTLINE. U.S. and Canada: IABP (4227)

Activity P13: Buoyant Force (Force Sensor)

RITTER Multiple Sonic Nozzle Calibration System

Quality Assurance for the Analysis of Steel by Gas Component Analysis

Air Pollution Control Division. Technical Services Program

ADVANCES IN AUTOMATIC OPTICAL INSPECTION: GRAY SCALE CORRELATION vs. VECTORAL IMAGING

Report: Salesforce and Microsoft Dynamics CRM Outlook Integration Compared

QIAsymphony RGQ Protocol Sheet

DomainSherpa.com: The Domain Name Authority. Google AdWords Keyword Planner for Domain Name Investors

Glassware Calibration Guidelines Laura B. Secor and Dwight R. Stoll, 02/01/2012 Adapted from National Bureau of Standards Document

Standard Operating Procedure for Training for Staff Working on the PM 2.5 Chemical Speciation Program

Clearview Customer Web Access

A fully quantitative research method for the analysis of lead in whole blood using the Thermo Scientific icap Q ICP-MS

FACILITY MONITORING SYSTEM DESIGN RECOMMENDATIONS

Guide to in situ leak testing of HEPA filter configurations that cannot be conventionally scan tested

On the re-calibration of Ni-foil reference standards

Manual Analysis Software AFD 1201

Agilent High Sensitivity DNA Kit Guide


Activity P13: Buoyant Force (Force Sensor)

Air and Water Balancing - Sound and Vibration Testing - Cleanroom Services - Building Commissioning Building Envelope Testing Thermal Imaging

TECHNIQUES FOR NATURAL GAS SAMPLING A DISCUSSION OF FIELD METHODS FOR OBTAINING SPOT SAMPLES. Randall Messman SW Regional Sales Manager

Setting Up My Business Account - Wireless

M24LRxx/CR95HF application software installation guide

Experiment (13): Flow channel

Determining the Quantity of Iron in a Vitamin Tablet. Evaluation copy

Configuring Outlook 2016 for Windows

- 2 - IMPORTANT SAFETY REMINDERS

Thermo Scientific ClinQuan MD Software For In Vitro Diagnostic Use. Confidence in Results With Data Integrity

Accelerating Microsoft Exchange Servers with I/O Caching

Cell Sorting. S3e Cell Sorter Quick Guide

HOW THE CONTAMINATION IS MEASURED CONTAMINATION CLASSES ACCORDING TO ISO 4406:1999

Gas Custody Transfer Calibration

Office 365 Mail: Mac Mail OS X

1805 Series Relief Valves

Filtering Spam Using Outlook s Rule

Air Shower Biosafety Changing Station with Pass-Through Box Pass-Through Box Automatic atomizer for no-contact hand disinfection

Jump Start: Aspen HYSYS Dynamics V7.3

EQUIPMENT OVERVIEW... 4 SETTING UP CALL MANAGER...5

Experiment #4 Sugar in Soft Drinks and Fruit Juices. Laboratory Overview CHEM August 2010

LX650 Lamp/Filter Replacement Instruction Sheet

Agilent 4339B/4349B High Resistance Meters

InfplusWin OPERATING MANUAL

ValveLink Mobile Software

Lamp and Filter Replacement Instructions

PLEUR EVAC CHEST DRAINAGE SYSTEMS. Great Innovations to Enhance Chest Drainage

Millikan Oil Drop Experiment Matthew Norton, Jurasits Christopher, Heyduck William, Nick Chumbley. Norton 0

JustClust User Manual

Transcription:

OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD APPLICATION NOTE OPS-001 The Optical Particle Sizer (OPS) is calibrated for size with Polystyrene Latex (PSL) spheres (per ISO 12501-1/4) at TSI. This method of calibration is common for all optical particle sizing instruments. In order to calculate mass from this size information there are assumptions that must be made. Standard assumptions are that the particles are spheres and by supplying a density the mass of the particles can be determined. In reality, the particles may not be spheres and the light scatter of the particles can be significantly different than that of the calibration particles (white PSL spheres). Corrections for these differences may be incorporated into an effective density which is the material density multiplied by a correction factor for the aerosol being measured. The calibration consists of three main steps: 1. Create a mass reference 2. Calculate the effective density 3. Apply the effective density to the OPS Note that the OPS Pump Always On and Dead Time Correction options should be enabled. The former is to avoid contamination of the optical chamber by the test aerosols, and the latter is to minimize coincidence error. During the test, the total particle concentrations also need to be lower than the OPS upper concentration limit which is 3000 particles/cm 3. The following steps illustrate how to calibrate the OPS for mass measurement. Once calibrated, the OPS will measure mass with reasonable accuracy as long as the aerosols of interest are the same as the aerosol used for calibration, and the size distributions remain the same. If the aerosol types or size distributions are different, a new calibration may be required. Note: Mass is a post-processing calculation. Density information can be entered either before or after data has been taken and it will be reflected in the graphs, tables and statistics. Create a Mass Reference The reference mass can be created using the OPS internal filter or an external filter. The external filter method is preferred since the mass collected by the OPS internal 37 mm internal filter may be biased by transport losses inside the OPS. Aerosol Instrument Manager, TSI, and TSI logo are trademarks of TSI Incorporated.

External Filter (Preferred Method) The calibration steps are outlined below: 1. Setup an external filter measurement system. To avoid sampling bias, the flow rate and sample time of the external filter measurement system should be the same or very close to the OPS flow rate and sample time. 2. Weight the new filter with a microbalance. You may need to minimize the static charge effect by exposing the filter to a bipolar ion source before weighting it. 3. Clean the filter cartridge before placing the new filter in it. 4. Co-locate the two instruments as shown in figure 1. Aerosol Souce Optical Particle Sizer OPS To Flow Control and Vacuum Figure 1. Experimental setup for gravimetric calibration 5. To avoid test aerosol filling up the optical chamber, DO NOT hook up the test aerosol to the OPS before starting the OPS pump. 6. Start the OPS. If the Pump Always On is enabled, the pump should be running after warm up. 7. Introduce the aerosol of interest into the OPS and the external filter measurement system, and start the measurement. Since the amount of mass collected by the filter depending on the aerosol density, size distribution and concentration, you need to adjust the sample time appropriately to get reliable mass measurement. 8. Stop the measurement. Take the filter out and measure it with the microbalance. 9. Using the mass collected on the external filter, calculate the mass concentration by using the equation shown below: Mass Collected Mass concentration = Flow Rate Sample Time Internal Filter The OPS has a 37-mm filter downstream of the optical chamber that can be used for mass calibration. A new filter is required for every calibration. Note that the mass collected by this filter may be biased by the transport losses inside the OPS, so it is recommended to use external filter measurement system if possible. The calibration steps are outlined below: -2-

1. Weight the new filter with a microbalance. You may need to minimize the static charge effect by exposing the filter to a bipolar ion source before weighting it. 2. Clean the filter cartridge before placing the new filter in it. Put the assembly back to the OPS. Make sure the assembly is secured to avoid leaking. 3. To avoid test aerosol filling up the optical chamber, DO NOT hook up the test aerosol to the OPS before the pump starts. 4. Start the OPS. If the Pump Always On is enabled, the pump should be running after warm up. 5. Introduce the aerosol of interest into the OPS, and start the measurement. Since the amount of mass collected by the filter depending on the aerosol density, size distribution and concentration, you need to adjust the sample time appropriately to get reliable mass measurement. 6. Stop the measurement. Take the filter out and measure it with the microbalance. 7. Using the mass collected on the 37 mm filter, calculate the mass concentration by using the equation shown below: Mass Collected Mass concentration = Flow Rate Sample Time The flow rate and sample time are the OPS flow rate and the OPS sample time. Calculate the effective density 1. Set the OPS particle density to 1 g/cm3, and record the integrated mass concentration. The effective density can be calculated by: Effective density = mass concentration from the filter 3 1 g/cm OPS integrated mass concentration Apply the effective density to the OPS (on instrument) 1. Set the OPS particle density to the effective density calculated above. a. On the OPS select the Setup tab at the bottom of the screen. b. Select the Sampling icon. c. Select the User Calibration icon. d. Enter your effective density in the Density box as shown in Figure 2. -3-

Figure 2. User Calibration screen 2. The OPS will now use this effective density in the displayed mass results both in graphs and tables. Apply the effective density to the OPS (in software) 1. If data is being collected using Aerosol Instrument Manager software, the density information is entered using the OPS Properties Dialog Box. 2. When you start the Aerosol Instrument Manager software and select the OPS 3330 Data type, the OPS 3330 Properties window opens. Select the Data Setup tab and select the Set Up Channels button. 3. If you select the Particle Density check box, you should then enter your effective density in the Density box. 4. If density varies by particle size, it is possible to do a more advanced correction and supply different densities for each size channel. If you uncheck the box in Particle Density portion of the Channel Setup window, you can enter different densities for each of the Channels. Figure 3 shows an example of Channel Setup using different densities. -4-

Figure 3. Channel Setup Screen with different densities 5. The Aerosol Instrument Manager software will now reflect your effective density value(s) in the mass results both in graphs, tables and statistics. -5-

TSI Incorporated Visit our website www.tsi.com for more information. USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 4 91 11 87 64 Germany Tel: +49 241 523030 India Tel: +91 80 67877200 China Tel: +86 10 8251 6588 Singapore Tel: +65 6595 6388 OPS-001 Rev. B 2012 TSI Incorporated Printed in U.S.A.