Real-time PCR: Understanding C t

Similar documents
2.500 Threshold e Threshold. Exponential phase. Cycle Number

Essentials of Real Time PCR. About Sequence Detection Chemistries

Introduction To Real Time Quantitative PCR (qpcr)

Highly specific and sensitive quantitation

Real-Time PCR Vs. Traditional PCR

Factors Influencing Multiplex Real-Time PCR

Data Analysis on the ABI PRISM 7700 Sequence Detection System: Setting Baselines and Thresholds. Overview. Data Analysis Tutorial

QPCR Applications using Stratagene s Mx Real-Time PCR Platform

PATHOGEN DETECTION SYSTEMS BY REAL TIME PCR. Results Interpretation Guide

Quando si parla di PCR quantitativa si intende:

Gene Expression Assays

amplification tech Optical Design of CFX96 Real-Time PCR Detection System Eliminates the Requirement of a Passive Reference Dye

Stratagene QPCR Mouse Reference Total RNA

Real time and Quantitative (RTAQ) PCR. so I have an outlier and I want to see if it really is changed

DNA Integrity Number (DIN) For the Assessment of Genomic DNA Samples in Real-Time Quantitative PCR (qpcr) Experiments

REAL TIME PCR SYBR GREEN

Path-ID Multiplex One-Step RT-PCR Kit

TaqMan Fast Advanced Master Mix. Protocol

QuantStudio 3D AnalysisSuite Software

Technical Note. Roche Applied Science. No. LC 18/2004. Assay Formats for Use in Real-Time PCR

TITRATION OF raav (VG) USING QUANTITATIVE REAL TIME PCR

MystiCq microrna cdna Synthesis Mix Catalog Number MIRRT Storage Temperature 20 C

Co Extra (GM and non GM supply chains: Their CO EXistence and TRAceability) Outcomes of Co Extra

Validating Microarray Data Using RT 2 Real-Time PCR Products

Absolute Quantification Getting Started Guide

Gene Expression Assay Performance Guaranteed With the TaqMan Assays QPCR Guarantee Program

Thermo Scientific DyNAmo cdna Synthesis Kit for qrt-pcr Technical Manual

Oligonucleotide Stability Study

REAL TIME PCR USING SYBR GREEN

Reagent Guide. Applied Biosystems StepOne and StepOnePlus Real-Time PCR Systems

Creating Standard Curves with Genomic DNA or Plasmid DNA Templates for Use in Quantitative PCR

Beginner s Guide to Real-Time PCR

Real-time PCR handbook

Using Excel (Microsoft Office 2007 Version) for Graphical Analysis of Data

Fluorescent dyes for use with the

QIAGEN Rotor Gene Q: Software V2.0

RealStar HBV PCR Kit /2012

CopyCaller Software. User Guide

quantitative real-time PCR, grain, simplex DNA extraction: PGS0426 RT-PCR: PGS0494 & PGS0476

Protocol. Introduction to TaqMan and SYBR Green Chemistries for Real-Time PCR

DyNAmo cdna Synthesis Kit for qrt-pcr

Real-time quantitative RT -PCR (Taqman)

Scott Reierstad. Field Applications Scientist

ONLINE SUPPLEMENTAL MATERIAL. Allele-Specific Expression of Angiotensinogen in Human Subcutaneous Adipose Tissue

Quantifiler Human DNA Quantification Kit Quantifiler Y Human Male DNA Quantification Kit

Methods and Application Guide. Introduction to Quantitative PCR

Application Guide... 2

Speed Matters - Fast ways from template to result

Accurate and sensitive mutation detection and quantitation using TaqMan Mutation Detection Assays for disease research

A guide to the analysis of KASP genotyping data using cluster plots

RT 2 Profiler PCR Array: Web-Based Data Analysis Tutorial

SYBR Green Realtime PCR Master Mix -Plus-

Real-time qpcr Assay Design Software

User Bulletin #2 ABI PRISM 7700 Sequence Detection System

VLLM0421c Medical Microbiology I, practical sessions. Protocol to topic J10

Introduction to Quantitative PCR

AP Physics 1 and 2 Lab Investigations

QuantStudio 3D AnalysisSuite Software: Relative Quantification

Methylation Analysis Using Methylation-Sensitive HRM and DNA Sequencing

Technical Note. Roche Applied Science. No. LC 19/2004. Color Compensation

Applications Guide. Real-Time PCR Applications Guide

Qubit dsdna HS Assay Kits

PrimePCR Assay Validation Report

Absolute Quantitation Using Standard Curve Getting Started Guide

Human Herpes Virus 4 (Epstein Barr)

Critical Factors for Successful Real-Time PCR

User Manual. Reference Gene Panel Mouse Probe protocol. Version 1.1 August 2014 For use in quantitative real-time PCR

Application Note. Biotechnology Explorer Crime Scene Investigator PCR Basics. Kit: A Real-Time PCR Extension

Mir-X mirna First-Strand Synthesis Kit User Manual

Absolute Quantifi cation of Gene Expression using SYBR Green in the Eco Real-Time PCR System

Molecular Spectroscopy

SmartFlare RNA Detection Probes: Principles, protocols and troubleshooting

TaqMan Genotyper Software v1.0.1 TaqMan Genotyping Data Analysis Software

REST 2009 Software User Guide

LightCycler 480 Real-Time PCR System

Nucleic Acid Amplification Life Science Dashboard Series 3

Power SYBR Green PCR Master Mix and Power SYBR Green RT-PCR Reagents Kit

Quantitative Real Time PCR Protocol. Stack Lab

CABIOS. Q-RT-PCR: data analysis software for measurement of gene expression by competitive RT-PCR

REAL-TIME PCR: Put the odds in your favor with SuperScript RT. FROM THEORY TO PRACTICE

TaqMan Mutation Detection Assay

SYBR Green PCR Master Mix and SYBR Green RT-PCR Reagents Kit

Quantification of Propionibacterium acnes. Standard kit 150 tests

Assay Migration Studies for In Vitro Diagnostic Devices Guidance for Industry and FDA Staff

Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL)

JBS FUNDAMENT Thermofluor Screen

Plexor Systems Instrument Setup and Data Analysis for the Applied Biosystems 7300 and 7500 Real-Time PCR Systems

Pitfalls in qpcr. Primer and probe design and synthesis. Mary Span Eurogentec S.A.

EZValidation Online Tool. The easy way to apply guidelines. The EZValidation Online Tool. The implementation of new tests

Gene Expression Macro Version 1.1

Quantitative PCR Systems

Mir-X mirna First-Strand Synthesis and SYBR qrt-pcr

All-in-One mirna qrt-pcr Reagent Kits For quantitative detection of mature mirna

StepOne and StepOnePlus Real-Time PCR Systems. Remarkably Simple Systems. Simply Remarkable Results.

Mathematics Online Instructional Materials Correlation to the 2009 Algebra I Standards of Learning and Curriculum Framework

Molecular Biology Techniques: A Classroom Laboratory Manual THIRD EDITION

Using TaqMan Endogenous Control Assays to select an endogenous control for experimental studies

Consistent Assay Performance Across Universal Arrays and Scanners

NATIONAL GENETICS REFERENCE LABORATORY (Manchester)

Guide to using the Bio Rad CFX96 Real Time PCR Machine

Transcription:

APPLICATION NOTE Real-Time PCR Real-time PCR: Understanding C t Real-time PCR, also called quantitative PCR or qpcr, can provide a simple and elegant method for determining the amount of a target sequence or gene that is present in a sample. Its very simplicity can sometimes lead to problems of overlooking some of the critical factors that make it work. This review will highlight these 3.500 factors that must be considered when 3.000 setting up and evaluating a real-time 2.500 PCR reaction. 2.000 Rn 4.500 4.000 3.500 3.000 2500 2000 1500 1000 0500 1000e + 001 1000e + 000 1000e - 001 aseline 3.500 3.000 2.500 2.000 1500 1000 0500 0000-0500 Exponential phase t 1000e + 001 1000e + 000 1000e - 001 1000e - 002 1000e - 003 1000e - 004 1 2 3 4 1500 1000 1000e - 002 9 30 31 32 33 34 35 36 37 38 39 40 Factors that can influence C 0500 t Exponential phase C t (threshold cycle) is the intersection 0000 between an amplification curve and t a -0500 threshold line (Figure 1). It is a relative measure of the concentration of target in the PCR reaction. Many factors impact the absolute value of C t besides the concentration of the target. We will discuss the most common template-independent factors that can influence C t and describe how to evaluate the performance of a real-time PCR reaction. Figure 1 shows several parameters of the real-time reaction amplification plot. The exponential phase in Figure 1 corresponds to the linear phase in Figure 1C. The threshold must be set in the linear phase of the amplification plot in Figure 1C. The C t value increases with a decreasing amount of template. However, artifacts from the reaction mix or instrument that change the fluorescence measurements associated with the C t calculation will result in templateindependent changes to the C t value. Therefore, the C t values from PCR reactions run under different conditions or with different reagents cannot be compared directly. Figure 1. Graphical representation of real-time PCR data. Rn is the fluorescence of the reporter dye divided by the fluorescence of a passive reference dye; i.e.,rn is the reporter signal normalized to the fluorescence signal of ROX. (A) In this view, Rn is plotted against PCR cycle number. () ΔRn is Rn minus the baseline; ΔRn is plotted against PCR cycle number. (C) An amplification plot shows the variation of log (ΔRn) with PCR cycle number. Amplitude 1000e - 003 1000e - 004 6.00 E + 3 5.00 E + 3 4.00 E + 3 3.00 E + 3 2.00 E + 3 1.00 E + 3 0.00 Master Mix A Master Mix VIC 0 5 10 15 20 25 30 in ROX Figure 2. Raw fluorescence data obtained from one assay using two master mixes with the same ROX level. The difference in signal is due to the master mix composition. The reaction was performed on an Applied iosystems 7900HT Fast Real-Time PCR System using a VIC MG probe. The x-axis shows the emission wavelength of the fluorophore and the y-axis shows the intensity of the emission.

The effect of master mix components The fluorescence emission of any molecule is dependent on environmental factors such as the ph of a solution and salt concentration. Figure 2 shows the raw fluorescence data of a TaqMan probe in the background of two different master mixes. Note that the fluorescence intensity is higher in Master Mix A even though the target, probe, and ROX dye concentrations are the same in both cases. The resulting ΔRn value will, therefore, vary as shown in Figure 3. Note that the baseline fluorescence signals, in a templateindependent factor, are different for the two master mixes (Figure 3A). Variations in the C t value do not reflect the overall performance of the reaction system (Figure 3). Master mixes with equivalent sensitivity may have different absolute C t values. ROX passive reference dye The Rn value is calculated as the ratio of the fluorescence of FAM dye divided by the fluorescence of ROX dye. Therefore, a lower amount of ROX dye would produce a higher Rn value assuming fluorescence signal from FAM dye is unchanged. This would lead to an increase in baseline Rn and subsequently a smaller as well as a different C t value. The new C t value obtained by lowering the level of ROX dye has no bearing on the true sensitivity of the reaction, but can have other unintended consequences. Low concentrations of ROX dye can result in increased standard deviation of the C t value, as shown in Figure 4. The greater the standard deviation, the lower the confidence in distinguishing between small differences in target concentration (see the precision section on the next page). Efficiency of a PCR reaction The efficiency of a PCR reaction can also affect C t. A dilution series amplified under low efficiency conditions could yield a standard curve with a different slope from one amplified under high efficiency conditions. In Figure 5, two samples (X and Y) amplified under low and high efficiency conditions show different C t values for the same target concentration. In this example, although the high-efficiency condition (the blue curve in Figure 5) gives a later C t at high concentrations, it results in better sensitivity at low target concentrations. The PCR efficiency is dependent on the assay, the master mix performance, and sample quality. Generally, an efficiency between 90 and 110% is considered acceptable. The observation that the C t value produced from one sample is higher than that of A 0.866 Rn vs. cycle 1.000 0.766 Master Mix A Rn 0.666 0.566 aseline A log () 0.100 Master Mix C t Master Mix A 0.010 0.466 aseline Master Mix C ta 0.366 1 2 3 3 3 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 0.001 1 2 3 3 3 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 Figure 3. Amplification of RNase P in equal amounts of human gdna using Master Mix A and Master Mix. (A) Rn is plotted against cycle number and the baselines for both reactions are shown. () log () is plotted against cycle number. The threshold (green line) is set at the same level for both master mixes. The C t value of Master Mix (C t ) is earlier than that of Master Mix A (C t A) for identical concentrations of target, reflecting the lower baseline of Master Mix. All amplifications were performed using the Applied iosystems 7500 Real-Time PCR System. A 14.5 14.0 0.30 Ct (VIC ) 13.5 13.0 12.5 Standard deviation 0.20 0.10 0.00 0.6X 0.75X 1X 0.6X 0.75X 1X ROX concentration ROX concentration Figure 4. Amplification of TGF-beta using master mixes containing 3 different concentrations of ROX dye. The variation of (A) C t value and () standard deviation with ROX dye concentration is shown. Decreasing the ROX dye concentration gives an earlier C t, but increases the standard deviation. All amplifications were performed using the Applied iosystems 7500 Real-Time PCR System.

40 lue Green 1 Log 35 Slope 3.3x + 39.8 4x + 42 Efficiency 100% 78% 168% 30 C t C t 100% 25 70% 20 15 X Y 5 Logs 108% 100% 92% 1 10 100 1000 10,000 100,000 Quantity Dilution Figure 5. Variation of C t with PCR efficiency. The blue standard curve has an efficiency of 100% (the slope is 3.3). The green standard curve has an efficiency of 78% (the slope is 4). Amplification of quantity Y gives an earlier C t under low efficiency conditions (green) compared to the high efficiency condition (blue). With a lower quantity (X) there is an inversion and the low efficiency condition (green) gives a later C t than the high efficiency condition (blue). Figure 6. Accurate calculation of PCR efficiency depends on the range of template amount used for the dilution series. For a 2-fold dilution with 5 points (orange), the potential artifact is higher than for the 10-fold dilution with 5 points (blue). the other, could be valuable in concluding that there is less template in the first sample, assuming all other factors such as instruments, reagents, and assays are equal. However, this is not true if different instruments, reagents, primers and probes, or reaction volumes are involved in producing the two C t s. Therefore, the absolute C t value comparison is only meaningful when comparing experiments using the same reaction conditions as defined above. How to evaluate the performance of a real-time PCR reaction In order to compare two reactions where a condition is changed (for example two different master mixes or two different instruments), the following parameters must be evaluated. Dynamic range To properly evaluate PCR efficiency, a minimum of 3 replicates and a minimum of 5 logs of template concentration are necessary. The reason for this suggested level of rigor is illustrated in Figure 6, which demonstrates the possible mathematical variation of slope or efficiency obtained when testing dilutions over 1 log vs. 5 logs. Thus, even if the assay is 100% efficient, a range from 70 to 170% can be obtained when testing a dilution series of a single log, due to the standard deviation in one dilution. For the same number of dilutions or replicates on a 5-log range, the potential artifact is only ±8 %. That means for 94% efficiency on a 5-log range, the assay would have a range of 88% to 100% efficiency. To accurately determine the efficiency of a PCR reaction, a 5-log dilution series must be performed. A slope of 3.3 ±10% reflects an efficiency of 100% ±10%. A PCR reaction with lower efficiency will have lower sensitivity. R 2 value Another critical parameter to evaluating PCR efficiency is R 2, which is a statistical term that indicates how good one value is at predicting another. When R 2 is 1, the value of Y (C t ) can be used to accurately predict the value of X (Figure 7A). If R 2 is 0, the value of X cannot be predicted from the value of Y (Figure 7). An R 2 value >0.99 provides good confidence in correlating two values. Precision The standard deviation (square root of the variance) is the most common measure of precision. If many data points are close to the mean, the standard deviation is small; if many data points are far from the mean, the standard deviation is large. In practice, a data set with a sufficient number of replicates forms an approximately normal distribution. This is frequently justified by the classic central limit theorem which states that sums of many independent, identically distributed random variables tend towards the normal distribution as a limit. As shown in Figure 8A, approximately 68% of the values are within 1 standard deviation of the mean, 95% are within two standard deviations, and 99.7% lie within 3 standard deviations. If a PCR is 100% efficient, the C t difference between two successive concentrations in a 2-fold dilution is 1 (Figure 8). To be able to quantify a 2-fold dilution in more than 99.7% of cases, the standard deviation has to be 0.167. The greater the standard deviation, the lower the ability to distinguish between 2-fold dilutions. To be able to discriminate between a 2-fold dilution in more than 95% of cases, the standard deviation has to be 0.250 (Figure 8C). Sensitivity Any system capable of effectively amplifying and detecting one copy of starting template has achieved the ultimate level of sensitivity, regardless of the absolute value of the C t. As described earlier, efficiency is a key factor in determining the sensitivity of a reaction (Figure 5). Another important consideration when detecting very low copy numbers is that normal distribution of template is not expected. Instead, a Poisson distribution is followed, which predicts that in a large number of replicates containing an average of one copy of starting template, approximately 37% should actually have no copies, only 37% should contain one copy, and 18% should contain two copies (see Figure 9). Thus, for a reliable low copy detection, a large number of replicates are necessary to provide statistical significance and to overcome the Poisson distribution limitation.

A 7 R 2 =1 7 R 2 =0 6 6 5 5 Sample 2 4 3 Sample 2 4 3 2 2 1 1 0 0 1 2 3 4 5 6 Sample 1 0 0 1 2 3 4 5 6 7 Sample 1 Figure 7. Examples of R 2 values calculated for 2 straight lines. (A) There is a direct relation between x and y values. () There is no relation between x and y values. A 99.6% X Y 95.4% 68.2% 1σ 1σ 1σ 1σ 1σ 1σ 1σ 1C t C X Y 1σ 1σ 1σ 1σ 3σ 2σ 1σ 1σ 2σ 3σ 1C t Figure 8. Normal distribution and standard deviation. (A) Normal distribution of data is shown. For a PCR efficiency of 100%, the difference in C t between the means of two successive samples in a 2-fold dilution series is 1 (sample X and sample Y). () To be able to quantify both samples in 99.7% of cases, the standard deviation has to be less than 1 C t divided by 6 standard deviations (1/6 = 0.167). (C) To be able to quantify both samples in 95% of cases, the standard deviation has to be less than 1 C t divided by 4 standard deviations (1/4 = 0.25). Frequency for 64 replicates 20 18 16 14 12 10 8 6 4 2 0 34 35 36 37 38 39 40 C t 3.3 pg 6.6 pg Table 1. Performance evaluation of real-time PCR. Factors Recommendations Criteria Efficiency Serial dilution with 5-log dilutions Slope: ~ 3.3 R 2 >0.99 Precision Minimum of 3 replicates Standard deviation <0.167 Sensitivity High replicate number of reactions for low copy number sample input due to Poisson distribution Statistical test analysis Figure 9. Poisson distribution for low copy number. The blue curve represents Poisson distribution for 3.3 pg of DNA (1 copy of DNA). The pink curve represents Poisson distribution for 6.6 pg of DNA (1 cell, 2 copies of DNA).

Conclusion Efficiency, R 2, precision, and sensitivity are used to determine performance of a PCR reaction when comparing different reaction conditions. For a rigorous evaluation, all factors listed in Table 1 must be evaluated together. In addition to these factors, proper experimental controls (such as no template control, no RT control) and template quality must be evaluated and validated. APPENDIX Amplification plot An amplification plot is the plot of fluorescence signal versus cycle number. Reactions are characterized by the point in time during cycling when amplification of a PCR product is first detected. The higher the starting copy number of the nucleic acid target, the sooner a significant increase in fluorescence is observed. aseline In the initial cycles of PCR there is little change in fluorescence signal. This defines the baseline for the amplification plot. In these cycles the fluorescence background of the reaction is observed. This is subtracted from the results when setting the baseline. For information on how to set up the baseline, download the document Applied iosystems 7300/7500 Real-Time PCR System (Part no. 4347825) at www.appliedbiosystems.com. Delta Rn (ΔRn) ΔRn is the normalization of Rn obtained by subtracting the baseline: (ΔRn = Rn baseline). Passive reference A dye that provides an internal fluorescence reference to which the reporter dye signal can be normalized during data analysis. Normalization is necessary to correct for fluorescence fluctuations caused by change in concentration, volume, or sample effects. PCR efficiency The equations below describe the exponential amplification of PCR: C n = C i x (1 + E) n C i = initial copy number C n = copy number at cycle n n = number of cycles E = efficiency of target amplification When efficiency is maximum (=1) the equation is: C n = C i x 2 n and the fold increase will be 2 at each cycle. The quantity of PCR product generated at each cycle decreases with decreasing efficiency, and the amplification plot is delayed. The recommended efficiency is from 90 to 110%. Reporter dye The reporter dye is the dye attached to the 5 end of the TaqMan probe. The dye provides a fluorescence signal that indicates specific amplification. For SYR Green I, the reporter dye binds double-stranded DNA, and the increase in fluorescence is also a measure of amplification. Specificity should be checked with a melt curve (Power SYR Green PCR Master Mix and RT-PCR Protocol, Part no. 4367218) or gel analysis of the PCR product. Rn Normalized reporter is the ratio of the fluorescence emission intensity of the reporter dye to the fluorescence emission intensity of the passive reference dye. A level of ΔRn used for the determination of the threshold cycle (C t ) in real-time assays. The level is set above the baseline, but sufficiently low to be within the exponential growth region of the amplification curve. The threshold is the line whose intersection with the amplification plot defines the C t. For information on how to set up the threshold, download the document Applied iosystems 7300/7500 Real-Time PCR System (Part no. 4347825) at www. appliedbiosystems.com. cycle (C t ) The fractional cycle number at which the fluorescence passes the threshold.

Life Technologies offers a breadth of products DNA RNA PROTEIN CELL CULTURE INSTRUMENTS For Research Use Only. Not intended for any animal or human therapeutic or diagnostic use. 2011 Life Technologies Corporation. All rights reserved. The trademarks mentioned herein are the property of Life Technologies Corporation or their respective owners. TaqMan is a registered trademark of Roche Molecular Systems, Inc. CO16926 0211 Headquarters 5791 Van Allen Way Carlsbad, CA 92008 USA Phone +1.760.603.7200 Toll Free in the USA 800.955.6288 www.lifetechnologies.com