Hiral N. Davea, Ashlesha G. Makwanab,*, and Bhanubhai N. Suhagiac



Similar documents
Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

High performance thin layer chromatographic method for estimation of deflazacort in tablet

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage Form

HIGH PERFORMANCE THIN LAYER CHROMATOGRAPHIC METHOD FOR DETERMINATION OF TADALAFIL IN TABLET DOSAGE FORM

PVV Satyanaryana et al., IJSID, 2012, 2 (1), International Journal of Science Innovations and Discoveries

Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry

Guidance for Industry

ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS

A NEW METHOD DEVELOPMENT AND VALIDATION FOR ANALYSIS OF RIVAROXABAN IN FORMULATION BY RP HPLC

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

UV-VIS SPECTROPHOTOMETRIC METHOD FOR ESTIMATION OF GABAPENTIN AND METHYLCOBALAMIN IN BULK AND TABLET

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

SUCRALOSE. White to off-white, practically odourless crystalline powder

International Journal of Pharma and Bio Sciences V1(2)2010

VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

Liquid Chromatographic Method for the estimation of Donepezil Hydrochloride in a Pharmaceutical Formulation

ORIENTAL JOURNAL OF CHEMISTRY An International Open Free Access, Peer Reviewed Research Journal.

By T.Sudha, T.Raghupathi

Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS

Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction

Journal of Chemical and Pharmaceutical Research

Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS

Photolytic-Thermal Degradation Study And Method Development Of Rivaroxaban By RP-HPLC

Journal of Chemical and Pharmaceutical Research

Research Article. Stress study and estimation of a potent anticoagulant drug rivaroxaban by a validated HPLC method: Technology transfer to UPLC

Determination of Letrozole in Tablet Formulations by Reversed Phase High Performance Liquid Chromatography

Determination of Food Dye Concentrations in an Unknown Aqueous Sample Using HPLC

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Scholars Research Library. A validated RP-HPLC method for estimation of Rivastigmine in pharmaceutical formulations

International Journal of Research and Reviews in Pharmacy and Applied science.

Research Article. Development and validation of a HPLC analytical assay method for dapoxetine tablets: A medicine for premature ejaculation

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

Overview. Purpose. Methods. Results

ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

DEVELOPMENT AND VALIDATION OF A STABILITY INDICATING HPLC METHOD FOR ANALYSIS OF PEFLOXACIN IN BULK DRUG AND TABLET DOSAGE FORM

GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.

Stability-indicating High Performance Thin Layer Chromatographic Determination of Clozapine in Tablet Dosage Form

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

Rao, et al., Int J Res Pharm Sci 2015, 5(2) ; Available online at Research Article

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

How To Test For Contamination In Large Volume Water

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

VALIDATION OF HPLC METHOD FOR DETERMINATION OF DOCOSAHEXAENOIC ACID (DHA) IN MULIIVITAMIN CAPSULE DIAH WIDOWATI*, ROS SUMARNY, ESTI MUMPUNI

A practical and novel standard addition strategy to screen. pharmacodynamic components in traditional Chinese medicine using

Development and Validation of RP-HPLC Method for the determination of Methylphenidate Hydrochloride in API

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Quality by Design Approach for the Separation of Naproxcinod and its Related Substances by Fused Core Particle Technology Column

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids

Transfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System Saving time and costs

Simultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC

ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE

Vitamin C quantification using reversed-phase ion-pairing HPLC

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Analysis of Various Vitamins in Multivitamin Tablets

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes

Determination of Caffeine in Beverages HPLC-1

UV Spectrophotometric estimation of Paracetamol and Lornoxicam in Bulk drug and Tablet dosage form using Multiwavelength

Online SPE-LC-APCI-MS/MS for the Determination of Steroidal Hormones in Drinking Water

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

DRAFT MONOGRAPH FOR THE INTERNATIONAL PHARMACOPOEIA PARACETAMOL ORAL SUSPENSION (September 2010)

A commitment to quality and continuous improvement

A Complete Solution for Method Linearity in HPLC and UHPLC

Analytical Method Development and Validation of Caffeine in Tablet Dosage Form By Using UV- Spectroscopy

Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

All the prepared formulations were subjected for following. evaluation parameters and obtained results were showed in Tables 6.3 &

Journal of Chemical and Pharmaceutical Research, 2012, 4(7): Research Article

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract

Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control

Simultaneous Qualitative and Quantitative Data Acquisition for Research of Diabetes Drugs

HIGH PERFORMANCE THIN LAYER CHROMATOGRAPHIC METHOD FOR DETERMINATION OF VALGACYCLOVIR IN TABLET DOSAGE FORM

Anion chromatography using on-line recycled eluents

Development and validation of a dissolution test for Candesartan cilexetil in tablet forms using reverse phase High performance liquid chromatography

Venkata S Rao Somisetty 1 *, D Dhachinamoorthi 1, S A Rahaman 2, CH M M Prasada Rao 1

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Pelagia Research Library

8. SUMMARY AND CONCLUSION

Applying Statistics Recommended by Regulatory Documents

Fundamentals of modern UV-visible spectroscopy. Presentation Materials

Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana*

Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup

A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS

Transcription:

International Journal of Applied Science and Engineering 2013. 11, 2: 125-135 Validated Reversed Phase High Performance Liquid Chromatographic Method for Determination of Three Novel Steroids in Bulk and Pressurized Metered - Dose Commercial Preparations Using a Common Mobile Phase Hiral N. Davea, Ashlesha G. Makwanab,*, and Bhanubhai N. Suhagiac a Department of Pharmaceutical chemistry, L. J. Institute of Pharmacy, Sanand Cross road, Ahmedabad, Gujarat, India b Department of Pharmaceutical chemistry, K.B. Institute of Pharmaceutical Education and Research, GH/6, Sector-23, Gandhinagar, Gujarat, India c Department of Pharmaceutical chemistry, L. M. College of Pharmacy, Opp. Gujarat University, Ahmedabad, Gujarat, India Abstract: A simple and sensitive reversed phase high performance liquid chromatography (RP HPLC) method has been proposed for determination of three widely used novel steroids in bulk as well as in their pressurized metered dose commercial preparations. The steroids are Budesonide, Fluticasone propionate and Ciclesonide. These steroids are widely used in treatment of different types of respiratory disorders mainly in the form of pressurized metered dose commercial formulations. The steroids were chromatographed using methanol: acetonitrile (60:40 v/v) as the mobile phase at a flow rate of 1.0 ml/min. The retention time (RT) of Budesonide, Fluticasone propionate and Ciclesonide was found to be 3.4161 ± 0.1305 min., 3.0512 ± 0.3694 min. and 4.841 ± 0.2421 min., respectively. The linearity of Budesonide, Fluticasone propionate and Ciclesonide were in the range of 1-10 µg/ml, 1-18 µg/ml and 2 20 µg/ml, respectively. The method was validated as per ICH guidelines and successfully applied for quantification of Budesonide, Fluticasone propionate and Ciclesonide in their pure and metered dose commercial formulations, contributing to improve the quality control and to assure the therapeutic efficacy. Keywords: Budesonide; ciclesonide; fluticasone propionate; novel steroids; pressurized metered - dose preparations; reversed phase high performance liquid chromatography (RP HPLC); analytical method validation. 1. Introduction Budesonide, chemically known as (11β, 16α) - 16, 17-[butytlidene bis (oxy)-11, 21dihydroxypregna-1, 4-diene-3, 20- dione]. It is used orally for the treatment of Crohn s disease. While it s inhaled formulation are used for the treatment of asthma and various types of respiratory disorders. Ciclesonide, chemically known as (11β, 16α) - 16, * Corresponding author; e-mail: dramakwana@yahoo.com 2013 Chaoyang University of Technology, ISSN 1727-2394 Received 1 August 2012 Revised 3 September 2012 Accepted 9 October 2012 125

Hiral N. Dave, Ashlesha G. Makwana, and Bhanubhai N. Suhagia 17-[(R)-cyclohexylmethylene] bi (oxy)-11-hydroxy-21-(2-methyl-1-oxopropoxy) pregna-1, 4-diene-3, 20-dione, is a new corticosteroid, for the prophylactic treatment of asthmatic attacks in adults. Fluticasone propionate, S-(fluoromethyl)-6α, 9-difluoro-11β, 17-dihydroxy-16α-methyl-3-oxoandrosta-1, 4-diene-17β-carbothioate, 17-propionate, is a synthetic steroid of the glucocorticoid family of drugs that is used to treat asthma and allergic rhinitis. The structures of Budesonide, Ciclesonide and Fluticasone propionate are represented in Figure 1, Figure 2 and Figure 3, respectively. Figure 1. Structure of budesonide Figure 2. Structure of ciclesonide Figure 3. Structure of fluticasone propionate 126

Validated Reversed Phase High Performance Liquid Chromatographic Method for the Determination of Three Novel Steroids in Bulk and Pressurized Metered - Dose Commercial Preparations Using a Common Mobile Phase As per literature, the methods available for determination of Budesonide include Uv spectroscopy [1], Stability indicating HPLC [2], LC/MS method [3, 4] and TLC with UV detection [5]. One UV spectroscopy [6] and HPLC [7] methods were also reported for simultaneous determination of Budesonide and Formoterol in their combine dosage forms. One LC/MS/MS method was reported as well for determination of Budesonide and Fluticasone in human sputam sample [8]. Simultaneous HPLC analysis was also reported for determination of Budesonide along with triamcinolone acetonide in microdialysate and rat plasma [9]. Similarly for determination of Fluticasone propionate, the reported methods include LC-MS/MS [10], Stability indicating LC [11] and visible spectroscopy [12]. A RP - HPLC method is reported for determination of Fluticasone propionate along with Salmeterol xinafoate [13]. For determination of Ciclesonide and Fluticasone propionate individually, one HPTLC method was also reported [14]. For determination of Ciclesonide, the available methods include LC-MS/MS [15] and stability indicating HPLC method [16]. One liquid chromatography method is reported for simultaneous determination of Ciclesonide and Formoterol fumarate in dry powder inhalers [17]. RP - HPLC is found to be the most effective and widely used method where sensitivity as well as accuracy is concerned. According to literature, any RP - HPLC method could not be traced for the analysis of Ciclesonide and Fluticasone propionate separately in their bulk as well as their pressurized metered dose preparations. Considering the wide importance of the novel steroids in the treatment of respiratory disorders, we developed simple, sensitive, accurate as well as precise RP - HPLC method for quantification of pure steroids and application of the developed method for determination of their metered dose commercial formulations were carried out. After development of the method, it was successfully validated as per validation parameters of ICH guidelines [18]. 2. Materials and method 2.1. Chemicals and reagents All chemicals used were of analytical grade and HPLC grade water was used throughout. Pure Budesonide (BUD), Fluticasone propionate (FP) and Ciclesonide (CIC) were of pharmacopeial grade. HPLC grade methanol and acetonitrile were used during the study (Ranchem, Rfcl Ltd., New Delhi, India). Water from Milli Q (Millipore Bedford, MA) was employed for total method development. Various pressurized metered dose preparations of Budesonide, Ciclesonide and Fluticasone propionate were obtained commercially. 2.2. Apperatus The instrumentation comprises of a Shimadzu s HPLC (LC-2010-HT, Shimadzu, Singapore) equipped with UV-Visible detector, ODS Phenomenex C18 column (250 mm X 4.6 mm i.d., 5 mm), Hamilton 20 μl glass syringe, a Citizen analytical balance (Sartorius Mechatronics India Pvt. Ltd., Bangalore, India.), an Equitron - ultra sonic sonicator (Labnet scientific services, Chennai, India) were used for method development. 2.3. Optimized chromatographic conditions Chromatographic estimations were performed under the following conditions: ODS Phenomenex C18 column (250 mm X 4.6 mm i.d., 5 mm) was used as a stationary phase at 127

Hiral N. Dave, Ashlesha G. Makwana, and Bhanubhai N. Suhagia ambient temperature. The mobile phase comprised methanol: acetonitrile (60:40, v/v), was pumped at a flow rate of 1 ml min-1. The mobile phase was filtered through Nylon 0.45 mm, 47 mm membrane filter and was degassed before use. The source of radiation was D2 lamp emitting a continuous ultra violet radiation between 180 nm to 400 nm. The elution was monitored at 243nm for Budesonide, 236 nm for Fluticasone propionate and 242 nm for Ciclesonide. The injection volume was 20 µl for each sample. 2.4. Preparation of the standard solutions All reagents were tested for stability in solution and during the actual analysis; the behavior of the analytes remained unchanged up to 48 h from their preparation at ambient temperature using the selected solvent as well as up to 96 h from the preparation in refrigerator using the same. The steroids were found to be stable during each kind of experimental measurements. Each measurement was done at ambient temperature. Standard solutions of the steroids were prepared individually having final concentration as 1 mg/ ml for each steroid. One ml of aliquots from these solutions were taken and diluted up to 10 ml using methanol (100 µg/ml). From these stock solutions, various dilutions were made for Budesonide, Ciclesonide and Fluticasone propionate separately for calibration curves. Evaluation was done by measuring peak areas with linear regression for each plate. 2.5. Procedure for pressurized metered dose commercial preparation analysis The preparation of the sample for metered dose formulation analysis was done exactly as was described in Indian Pharmacopoeia - 2007 [19]. The samples from the metered dose formulation were collected in methanol and the contents were shaken well. Solutions were filtered through Whatman filter paper No. 41 and same HPLC conditions were applied as mentioned above. All the brands of the pressurized metered dose preparations containing Budesonide, Ciclesonide and Fluticasone propionate were tested according to the procedure described above. The formulations were already available in the market and hence were procured from the registered medical store. 2.6. Validation parameters 2.6.1. Linearity and range A calibration curve was plotted over a concentration range of 1-10 µg/ml for Budesonide, 1 18 µg/ml for Fluticasone propionate and 2-20 µg/ml for Ciclesonide. Stock solutions of Budesonide, Ciclesonide and Fluticasone propionate, having concentration 100 µg/ml were prepared separately and further diluted for calibration curves of each drug separately. Accurately measured standard stock solution of Budesonide, standard stock solution of Fluticasone propionate and standard stock solution of Ciclesonide were used for further preparations of standard solutions. 20 µl of each solution was injected under operating chromatographic conditions described above. 2.6.2. Precision Intraday precision and inter day precision for the developed methods were measured in terms of % RSD. The experiments were repeated six times a day for intraday precision and on six 128

Validated Reversed Phase High Performance Liquid Chromatographic Method for the Determination of Three Novel Steroids in Bulk and Pressurized Metered - Dose Commercial Preparations Using a Common Mobile Phase different days for interday precision. The concentration values for both intraday precision and interday precision were calculated six times separately and percent relative standard deviation were calculated. Finally the mean of % RSD (% RSD = [S/X] 100, where S is standard deviation and X is mean of the sample analyzed) was calculated. The precision of the instrument was checked by repeated scanning of the same spot of both drugs six times without changing the condition of the instrument. 2.6.3. Recovery studies (Accuracy) Accuracy of proposed method and interference from excipients was determined by recovery experiments. Recovery experiments were carried out by the standard addition method (spiking method). This study was performed by addition of known amounts of Budesonide, Ciclesonide and Fluticasone propionate (80%, 100% and 120% to that of labeled claim of commercial formulations) to a known concentration of the pressurized metered dose preanalyzed formulations and % of standard drugs recovered, were calculated. 2.6.4. Limit of detection (LOD) Limit of detection was calculated using following equation as per ICH guidelines. LOD = 3.3 S/m where, S is the standard deviation of the peak areas of the drug and m is the slope of the corresponding calibration curve. It is expressed as signal to noise ratio of 3:1. 2.6.5. Limit of quantitation (LOQ) Limit of quantification (LOQ) was calculated using following equation as per ICH guidelines. LOQ = 10 S/m where, S is the standard deviation of the peak areas of the drug and m is the slope of the corresponding calibration curve. It is expressed as signal to noise ratio of 10:1. 2.6.6. System suitability Number of theoretical plates was determined by employing the formula, n =16(t/w)2 where t=retention time and w = width of the peak. Tailing factor was derived from the formula t = w/2t where w = half of the width, t = retention time. The retention time was also observed as a system suitability factor. 3. Results and discussion As per the literature, HPTLC method was reported for determination of two steroids, Ciclesonide and Fluticasone propionate individually [14]. As we know that HPTLC method development varies with various factors like size of the Plates, saturation of the Chamber, scanning position of the plates inside the scanner etc. RP-HPLC method provides more flexibility as compare to HPTLC in case of estimation of drugs and method development as well as validation. One of the most important purposes of the method development is application of the developed method for commercial formulation analysis. When we compare the results of formulation analysis by RP-HPLC and HPTLC for the same steroids, RP-HPLC proven to be more accurate and precise as compare to the previously developed HPTLC method. Excellent results of method validation as well as commercial formulation analysis were obtained in the 129

Hiral N. Dave, Ashlesha G. Makwana, and Bhanubhai N. Suhagia developed RP-HPLC method for all three Novel steroids. 3.1. RP - HPLC method optimization Budesonide, Fluticasone propionate and Ciclesonide were soluble in methanol; therefore methanol was selected as common solvent for the steroids. Attempts were made to optimize a common mobile phase for the novel steroids which would give fast retention as well as detection at lower concentration. Various mobile phases were tried and retention time as well as peak area and peak symmetry was observed. Best results were obtained using Methanol: ACN in the ratio of 60: 40 v/v as mobile phase using C-18 column. The retention time (RT) of Budesonide, Fluticasone propionate and Ciclesonide was found to be 3.4161 ± 0.1305, 3.0512 ± 0.3694 and 4.841 ± 0.2421, respectively. The selected mobile phase gives best result in terms of faster retention, least tailing of the peaks as well as accurate results for each kind of chromatographic measurements throughout the development of the entire method. The chromatogram of Budesonide, Ciclesonide and Fluticasone propionate were represented in Figure 4, Figure 5 and Figure 6 respectively. Figure 4. HPLC chromatogram of budesonide Figure 5. HPLC chromatogram of ciclesonide 130

Validated Reversed Phase High Performance Liquid Chromatographic Method for the Determination of Three Novel Steroids in Bulk and Pressurized Metered - Dose Commercial Preparations Using a Common Mobile Phase Figure 6. HPLC chromatogram of fluticasone propionate 3.2. Method validation 3.2.1. Linearity and range Linearity was found between the concentration range of 1-10 µg/ml, 1-18 µg/ml for Fluticasone propionate and 2-20 µg/ml for Ciclesonide. The average linear regression equation was represented as Y = 105.75x + 9.065 for Budesonide, y = 43.30x + 69.51 for Ciclesonide and same for Fluticasone propionate was found to be y = 66.24x + 14.304, where X is the concentration of drug Peak areas and concentration were subjected to least square linear regression analysis to calculate the calibration equation and correlation coefficient. The linearity was found to be 0.999 for Budesonide, 0.999 for Fluticasone propionate and 0.999 for Ciclesonide in terms of correlation of coefficient (R2). Relative standard deviations for determination of linearity of method were found to be 0.0174 for Budesonide, 0.0207 for Ciclesonide and 0.0276 for Fluticasone propionate determination (Table 1). 3.2.2. Precision The intra-day and inter-day precision for Budesonide was found to be 0.1392 and 0.6935 respectively and the same for Ciclesonide was found to be 0.3459 and 0.7309 respectively. For Fluticasone propionate, the intraday and inter day precision was found to be 0.5058 and 1.5267 respectively in terms of %RSD. These values indicate that the method is precise. The %RSD for precision of the instrument by measuring the peak area was found to be 0.2047 for Budesonide, 0.2125 for Ciclesonide and 0.2091 for Fluticasone propionate. The %RSD for measuring the peak area (less than 2%), ensured proper functioning of HPLC system. The results depicted in Table 1. 3.2.3. Recovery studies (Accuracy) The recovery study was carried out by spiking method. The formulations were preanalyzed and same types of sampling were made from the formulation for the recovery study. In the sample of formulations itself, 80%, 100% and 120% of the standard drugs were added for the respected steroids. Here at no addition, the estimated content of Budesonide, Ciclesonide and Fluticasone propionate were 47.26 µg/ml, 47.06 µg/ml and 47.22 µg/ml respectively. These are 131

Hiral N. Dave, Ashlesha G. Makwana, and Bhanubhai N. Suhagia the results of formulation analysis without spiking. After spiking 80%, 100% and 120% of the standards, the estimated contents were written in the table itself (column 2). Finally % of the standard that is recovered, were calculated, which is shown in the column 3 of Table 2. When the recovery study was carried out in preanalyzed sample of metered dose commercial formulations percentage recoveries were found to be 99.84 ± 0.8076, 99.46 ± 0.8787 and 99.71 ± 0.5262 for Budesonide, Ciclesonide and Fluticasone propionate respectively (Table 2). 3.2.4. Limit of detection (LOD) and limit of quantitation (LOQ) The Values of LOD and LOQ for Budesonide were found to be 0.1220 µg/ml and 0.3699 µg/ml. The value of LOD was found to be 0.5304 µg/ml for Ciclesonide and 0.2041 µg/ml for Fluticasone propionate. Similarly the values of LOQ were found to be 1.6075 µg/ml and 0.6186 µg/ml for Ciclesonide and Fluticasone propionate respectively (Table 1). Table 1. Validation parameters of developed RP - HPLC methods for budesonide (BUD), ciclesonide (CIC) and fluticasone propionate (FP) Validation parameters BUD CIC FP Linearity range (µg/ml) 1-10 2-20 1-18 Scanning wavelength 243 nm 242 nm 236 nm Regression equation Y = 105.75x + 9.065 y = 43.30x + 69.51 y = 66.24x + 14.304 Correlation coefficient (R2) 0.999 ± 0.0174 0.999 ± 0.0207 0.999 ± 0.0276 Accuracy (Recovery study) 99.84 ± 0.8076 99.46 ± 0.8787 99.71 ± 0.5262 LOD (µg/ml) 0.1220 0.5304 0.2041 LOQ (µg/ml) 0.3699 1.6075 0.6186 Precision - intraday (%RSD) 0.1392 0.3459 0.5058 Interday (%RSD) 0.6935 0.7309 1.5267 Repeatability (%RSD) 0.2047 0.2125 0.2091 RSD- Relative Standard Deviation Table 2. Results of recovery studies of budesonide (BUD), ciclesonide (CIC) and fluticasone propionate (FP) by the developed RP - HPLC method Std drug added Estimated content (mcg/ml) % Recoverya (%) BUD CIC FP BUD CIC FP 0 47.26 47.06 47.22 80 87.26 87.06 87.22 99.80 ± 0.8564 98.64 ± 0.5201 99.61 ± 0.9145 100 97.26 97.06 97.22 99.83 ± 0.9231 99.08 ± 0.7963 101.42 ± 0.3868 120 107.26 107.06 107.22 99.91 ± 0.6433 100.66 ± 1.3197 98.54 ± 0.2773 Mean recovery ± standard deviation 99.84 ± 0.8076 99.46 ± 0.8787 99.71 ± 0.5262 3.2.5. System suitability The number of theoretical plates, tailing factors and retention times were represented in Table 3. 132

Validated Reversed Phase High Performance Liquid Chromatographic Method for the Determination of Three Novel Steroids in Bulk and Pressurized Metered - Dose Commercial Preparations Using a Common Mobile Phase Table 3. Results of system suitability of budesonide, ciclesonide and fluticasone propionate by the developed RP HPLC method. System suitability Budesonidea Ciclesonidea Fluticasone propionatea parameters Retention time 3.4161 ± 0.1305(%RSD) 4.841 ± 0.2421(%RSD) 3.051 ± 0.3694(%RSD) Tailing factor 0.6313 ± 1.1724 (%RSD) 1.046 ± 0.3103 (%RSD) 1.043 ± 0.2288 (%RSD) Theoretical plates 9848.5 ± 0.7446 a - mean value of six determinations 11428 ± 0.7837 9531 ± 0.0913 3.3. Application of the developed method for analysis of pressurized meter dose commercial formulations Various brands of pressurized metered dose formulations containing Budesonide, Fluticasone propionate and Ciclesonide were tried and the results were shown in the table 4. Table 4. Application of the developed RP - HPLC methods for pressurized metered dose commercial formulations of budesonide (BUD), ciclesonide (CIC) and fluticasone propionate (FP) Formulation % Amount founda ± S.D. BUD CIC FP BUD CIC FP Brand A Brand C Brand E 95.84 ± 0.848 94.54 ± 1.061 94.12 ± 0.653 Brand B a Brand D Brand F 95.56 ± 0.690 93.70 ± 0.854 94.79 ± 1.004 - mean value of three determinations 4. Conclusion The newly developed HPLC method for estimation of Budesonide, Fluticasone propionate and Ciclesonide in bulk and pressurized metered dose formulations are found to be simple, rapid, sensitive, accurate, precise and reliable which indicates its adequacy for the routine pharmaceutical analysis. The statistical analysis proves that the developed method is suitable for determination of Budesonide, Ciclesonide and Fluticasone propionate as bulk drugs individually and in the commercial pressurized metered dose preparations without any interference from the excipients and with the use of common mobile phase. Acknowledgement The authors express their sincere thanks to Dr. K. Pundarikakshudu, Director, L. J. Institute of Pharmacy, Ahmedabad, Gujarat, India and Lok Jagruti Kendra, Ahmedabad, Gujarat, India for the giving the proper facilities for this research work. References [ 1] Sanap, D. D., Sisodia, A. M., Patil, S. H., and Janjale, M. V. 2011. Novel And Validated Spectrophotometric Determination Of Budesonide From Bulk And Tablets Using Mixed Hydrotropic Solubilization Technique. International journal of Pharmaceutical science and 133

Hiral N. Dave, Ashlesha G. Makwana, and Bhanubhai N. Suhagia [ 2] [ 3] [ 4] [ 5] [ 6] [ 7] [ 8] [ 9] [10] [11] [12] [13] [14] [15] [16] 134 research, 2, 9: 2419-2423. Hou, S., Hindle, M., and Byron, P. R. 2001. A stability-indicating HPLC assay method for Budesonide. Journal of Pharmaceutical and Biomedical Analysis, 24: 371-380. Deventer, K., Mikulikov, P., Hoecke, H. V., Eenoo, P. V., and Delbeke, F. T. 2006. Detection of budesonide in human urine after inhalation by liquid chromatography-mass spectrometry. Journal of Pharmaceutical and Biomedical Analysis, 42: 474-479. Lindberg, C., Blomqvist, A., and Paulson, J. 2005. Determination of (22R,S) budesonide in human plasma by automated liquid chromatography/thermospray mass spectrometry. Biological Mass Spectrometry, 21, 11: 525-533. Krzek, J., Hubicka, U., Dabrowska, M., and Leciejewicz E. 2002. Determination of budesonide R(+) and S( ) isomers in pharmaceuticals by thin-layer chromatography with UV densitometric detection. Chromatographia, 56, 11: 759-762. Prasad, A. V. S. S. 2006. Simultaneous spectrophotometric estimation of budesonide and formoterol fumarate in their combined dosage forms. Indian journal of chemical technology, 13: 81-83. Assi, K. H., Tarsin, W., and Chrystyn, H. 2006. High performance liquid chromatography assay method for simultaneous quantitation of formoterol and budesonide in Symbicort Turbuhaler. Journal of Pharmaceutical and Biomedical Analysis, 41: 325-328. Buscher, B. A., Jagfeldt, H., Sandman, H., Brust-van S. R., Schaik, F., and Brüll, L. P. 2012. The determination of budesonide and fluticasone in human sputum samples collected from COPD patients using LC-MS/MS. Journal of Chromatography B Analytical Technology and Biomedical Life Science, 880, 1: 6-11. Vieira, M. L. T., Singh, R. P., and Derendorf, H. 2010. Simultaneous HPLC analysis of triamcinolone acetonide and budesonide in microdialysate and rat plasma: Application to a pharmacokinetic study. Journal of Chromatography B, 878: 2967-2973. Laugher, L., Noctor, T. G., Barrow, A., Oxford, J. M., and Phillips, T. 1999. An improved method for the determination of fluticasone propionate in human plasma. Journal of Pharmaceutical and Biomedical Analysis, 21, 4: 749-58. Maximiliano, S., Daniele, R. N., Lucélia, M., Diogo, P. L., and Sergio, L. D. 2008. Validation of a Stability Indicating Reversed Phase LC Method for the Determination of Fluticasone Propionate in Pharmaceutical Formulations. Journal of Liquid Chromatography and Related Technologies, 31: 2113-2127. Ayad, M. M., Henawee, M. M., Abdellatef, H. E., and Sayed, H. 2005. Spectrophotometric determination of levonorgestrel, norethisterone acetate, mometasone furoate and fluticasone propionate using 4 amino-antipyrine, 43, 3: 24-28. Murnane, D., Martin, G. P., and Marriott, C. 2006. Validation of a reverse-phase high performance liquid chromatographic method for concurrent assay of a weak base salmeterol xinafoate and a pharmacologically active steroid fluticasone propionate. Journal of Pharmaceutical and Biomedical Analalysis, 40, 5: 1149-1154. Dave, H. N., Makwana, A. G., Bhatt, C., and Killambi, P. 2011. Validated HPTLC method for the determination of the two novel steroids in bulk and pressurized metered-dose preparations. International Journal of Applied Science and Engineering, 9, 3: 177-185. Maschera, H. J., Zechb, K., and Maschera, D. G. 2008. Sensitive simultaneous determination of ciclesonide, ciclesonide-m1-metabolite and fluticasone propionate in human serum by HPLC MS/MS with APPI. Journal of Chromatography B, 869, 1-2: 84-92. Elkady, E. F. and Fouad, M. A. 2011. Forced degradation study to develop and validate

Validated Reversed Phase High Performance Liquid Chromatographic Method for the Determination of Three Novel Steroids in Bulk and Pressurized Metered - Dose Commercial Preparations Using a Common Mobile Phase stability-indicating RP-LC method for the determination of ciclesonide in bulk drug and metered dose inhalers. Talanta, 87: 222-229. [17] Patil, A. T., Patil, S. D., and Shaikh, K. A. 2011. Sensitive LC method for simultaneous determination of ciclesonide and formoterol fumarate in dry powder inhaler. Journal of Liquid Chromatography and Related Technologies, 34: 1568-1577. [18] International Conference on Harmonization (ICH) of Technical Requirements for the Registration of Pharmaceutical for Human Use, Validation of Analytical Procedures: Text and Methodology Q2 (R1). 2005: 1-13. [19] Indian Pharmacopoeia, (Published by Indian Pharmacopoeia Commission- Ghaziabad), 2007, 1: 639-652. 135