Rao, et al., Int J Res Pharm Sci 2015, 5(2) ; Available online at Research Article
|
|
|
- Amos Thomas
- 10 years ago
- Views:
Transcription
1 International Journal of Research in Pharmacy and Science Available online at Research Article Stability-indicating UPLC method for determining related substances and degradants in Rivaroxaban Rao PSP 1,2, Cholleti VK 1, Reddy VR 1 1 Dr. Reddy's Laboratories Ltd. API - PDO, Bachupally, Hyderabad Andhra Pradesh, India. 2 Department of Chemistry, JNT University, Kukatpally, Hyderabad, Andhra Pradesh, India. Address for Correspondence P. Surya Prakash Rao [email protected] Received: Review completed: Accepted: Access this article online QR Code Website: ABSTRACT A simple, sensitive and reproducible method was developed on ultra performance liquid chromatography (UPLC) coupled with a photodiode array detector for the quantitative determination of Rivaroxaban (RVB) in drug substance and pharmaceutical dosage forms. The method is applicable for the quantification of related substances and assays of drug substances. Chromatographic separation was achieved on Acquity UPLC BEH HSS T mm, 2.1-mm, and 1.8-μm columns, and the gradient elution within a short runtime, i.e., within 12.0 min. The eluted compounds were monitored at 248 nm, the flow rate was 0.45 ml/min, and the column oven temperature was maintained at 25 C. The resolution between RVB and ten impurities was greater than 2.0. The high correlation coefficient (> ) value indicates the clear correlations between the investigated compound concentrations and their peak areas within the test ranges. The repeatability and intermediate precision expressed by the RSD were less than 6.5. The accuracy and validity of the method were further ascertained by performing recovery studies via a spike method. Rivaroxaban was subjected to the stress conditions of oxidative, acid, base, hydrolytic, thermal and photolytic degradation. Rivaroxaban was found to degrade significantly in acid and base stress conditions and stable in thermal, photolytic degradation, oxidative and hydrolytic conditions. The degradation products were well resolved from main peak and its impurities, proving the stability-indicating power of the method. The developed method was validated as per international conference on harmonization (ICH) guidelines. Key words: UPLC, Rivaroxaban, Forced degradation, Validation, Stability-indicating INTRODUCTION High performance liquid chromatography (HPLC) is well recognized as regular separation technique in the pharmaceutical industry. This is preferred mostly due to the many features offered by HPLC like robustness, ease of operation, well-understood separation principles, sensitivity and tunable selectivity. However, the main constraint of HPLC since many years has been the lack of high efficiency and the analysis time. The escalating need for impurity analysis in pharmaceuticals provides the basis for evaluating rapid, simple and reliable methods of analysis. UPLC provides the speed and resolving power to develop efficient separation methods. This new category of analytical separation science retains the practicality and principles of HPLC while increasing the overall interrelated attributes of speed, sensitivity and resolution. Using a 1.7 μm reversed-phase packing material and operating at high pressure (up to psi), makes Ultraperformance liquid chromatography (UPLC) a unique and highly sophisticated technique to achieve separations with very high-resolution within a short period of time and with little organic solvent consumption 1-3 which has attracted much attention of pharmaceutical and biochemical analysts. In the present work, this technology has been applied to the method development and validation study of related substance and assay determination of RVB bulk drug and dosage forms. The chemical name of Rivaroxaban is 5-Chloro-N-({(5S)-2-oxo-3-[4-(3-oxo-4- morpholinyl)phenyl]-1,3-oxazolidin-5-yl}methyl)-2 thio phenecarboxamide (Fig.1A). Rivaroxaban used for potent anticoagulant and antithrombotic effects 4,5.Very few methods appeared in the literature for the determination of RVB in pharmaceutical dosage forms, determination of RVB in human plasma by using high-performance liquid chromatography -tandem mass spectrometry (HPLC MS/MS) 6, determination of RVB by using 17
2 spectrophotometric methods in pharmaceutical formulations and bulk dosage forms 7, few RP-HPLC method for determination of Rivaroxaban in pharmaceutical dosage forms 8-10 and a chiral HPLC method 11. Furthermore, there is no stability indicating HPLC/UPLC method is available in literature for estimation of RVB along with related substances and degradents. Currently, the determination of impurities is one of the most difficult tasks for pharmaceutical analysis during method development, especially if increasing numbers of impurities are required to be determined. In RVB, separation of process related impurities and degradents are very critical and there is no HPLC/UPLC method reported in the literature that can adequately separate these process related impurities along with degradants. It is, therefore, felt necessary to develop a new stability indicating method for the related substance determination and quantitative estimation of RVB which can separate all the impurities in short run time without compromising the resolution and sensitivity. Hence a reproducible stability-indicating RP UPLC method was developed for the quantitative determination of RVB and its ten impurities namely Imp-1, 2, 3, 4, 5, 6, 7, 8, 9 and 10 (Fig.1B-K). This method was successfully validated according to the International Conference Harmonization (ICH) guidelines (Validation of Analytical Procedures: Test and Methodology Q2). Imp-3 : Fig.1D Imp-5 : Fig.1F Imp-7 : Fig.1H Imp-9 : Fig.1J Imp-4 : Fig.1E Imp-6 : Fig.1G Imp-8 : Fig.1I Imp-10 : Fig.1K MATERIALS AND METHODS Materials and reagents Active pharmaceutical ingredient standards and samples were supplied by Dr. Reddy's Laboratories Limited, IPDO, Hyderabad, India. Commercially available Xarelto in 20- mg tablets was used for the dosage form analysis. HPLCgrade acetonitrile and methanol, analytical-grade Ortho phosphoric acid and 1-Octane sulphonic acid sodium salt were purchased from Merck, Darmstadt, Germany. Water was prepared in-house by using a Millipore Milli-Q Plus water purification system (Millipore Corporate Headquarters, Billerica, MA). Figure 1 (A-K): Chemical structures of Rivaroxaban and its related substances Equipment The LC system used for method development, forced degradation studies and method validation was Acquity UPLC system (Waters, Milford, USA) which consists of a binary solvent manager and a photodiode array (PDA) detector. The output signal was monitored and processed using empower2 software. Cintex digital water bath was used for hydrolysis studies, Photostability studies were carried out in a photostability chamber (Sanyo, Leicestershire, UK).Thermal stability studies were performed in a dry air oven(cintex,mumbai, India). Imp-1 : Fig.1B Rivaroxaban: Fig.1A Imp-2 : Fig.1C Chromatographic conditions The method was developed using Waters Aquity BEH HSS T3 100-mm, 2.1-mm, and 1.8-μm particle sizes. Separation was achieved using a gradient method. Mixture of 1 ml Ortho phosphoric acid and 10 mm Octane 1-sulphonic acid sodium salt finalized as buffer.buffer and acetonitrile in the ratio of 90:10,20:80 was selected as mobile phase A and B respectively and wavelength was 248 nm with a flow rate of 0.45 ml/min at ambient temperature. The injection volume was 3.0 μl.gradient program finalized as (Time/A/B) 0.01/90/10, 4.5/75/25, 7.5/75/25, 9.0/50/50, 10.0/50/50, 10.01/90/10 and 12.0/90/10. 1 ml phosphoric acid in mixture of methanol and acetonitrile in a ratio of 50:50 (v/v) used as diluent. 18
3 LC MS conditions An LC MS system (Agilent 1200 series LC coupled with Applied Biosystems 4000 Q Trap triple quadrupole mass spectrometer with Analyst 1.4 software, MDS SCIEX, USA) was used for to identify the unknown compounds formed during forced degradation studies. An Inertsil C8, 150 mm 4.6 mm, 3.0 μm column (Agilent, USA) was used as the stationary phase. A 1.0 ml Ortho phosphoric acid in 1litre water and 1.0 ml of Ortho phosphoric acid in water and acetonitrile in a ratio of 20:80 (v/v) were used for solvent B. The gradient program (T/B) was set as 0.01/10, 5/10, 20/35, 30/80, 42/80, 42.1/10 and 50/10. 1 ml phosphoric acid in methanol and acetonitrile in a ratio of 50:50 (v/v) used as diluent. The flow rate was 0.8 ml/min. The analysis was performed in positive electro-spray/positive ionization mode, the ion source voltage was 5000 V, and the source temperature was 450 C. GS1 and GS2 were optimized to 30 and 35 psi, respectively. The curtain gas flow was 20 psi. Preparation of stock solutions A stock solution of RVB (2000 μg/ml) was prepared by dissolving an appropriate amount of the drug in dilent. Working solutions containing 500 μg/ml and 100 μg/ml were prepared from this stock solution for the determination of related substances and for the assay determination respectively. A stock solution of impurities (mixture of Imp-1 to 10) at 50 μg/ml was prepared in dilent. Preparation of sample solution Twenty (n = 20) RVB 20-mg Xarelto tablets were weighed, and the pellets were transferred into a clean, dry mortar. Pellets equivalent to 100 mg of the drug were dissolved in 100 ml of diluent to make a 1000-μg/mL solution ml of this solution was diluted to 20 ml with diluent, yielding 500-μg/mL solution.10ml of this solution was diluted to 50 ml with diluent, yielding 100- μg/ml solution. These solutions were filtered through a 0.22-μm nylon membrane filter. The resulting solution was analyzed by UPLC. Stress studies Specificity is the ability of the method to measure the analyte response in the presence of its potential impurities 12. The specificity of the developed LC method for RVB was carried out in the presence of its ten impurities. Stress studies were performed at an initial concentration of 500 μg/ml of RVB in active pharmaceutical ingredients (API) and tablets to provide the stability-indicating property and specificity of the proposed method. Intentional degradation was attempted by the stress conditions of UV light (254 & 365 nm), heat (105 C), acid (0.1-N HCl at 70 C), base (0.1-N NaOH at RT), hydrolysis (70 C) and oxidation (3.0 H2O2 at 70 C) to evaluate the ability of the proposed method to separate RVB from its degradation products. For heat and light studies, the study period was 10 days; whereas for the hydrolysis, base, acid and oxidation studies, the study periods were 48 h, 1 h, 2 h, and 48 h, respectively. The purity of the peaks obtained from the stressed samples was verified using the PDA detector. The purity angle was within the purity threshold limit obtained in all the stressed samples and demonstrated the analyte peak homogeneity. An assay of stressed samples was performed (at 100 µg/ml) by comparison with reference standard and the mass balance ( assay + impurities + degradation products) for each of the stressed samples was calculated. An assay was also calculated for the RVB sample by spiking all ten impurities at the specification level (i.e., 0.15 ). METHOD VALIDATION The described method has been validated for the assay and related substances by UPLC determination as per ICH guidelines 13. Precision The repeatability of the method for the related substances was checked by a sixfold analysis of 500 μg/ml of RVB spiked with 0.75 μg/ml of each of the ten impurities. The RSD was calculated for each impurity peak area. Ruggedness study for related substances of RVB method was demonstrated on different day by varying different analyst, different equipment and different column.. Intermediate precision was determined by a sixfold analysis of 500 μg/ml of RVB spiked with 0.75 μg/ml of each of the ten impurities. The RSD () of the peak area was calculated for each impurity. The precision of the assay was evaluated by performing six (n = 6) independent assays of a test sample of RVB and by comparison with a qualified reference standard. The RSD () of the six results was calculated. The intermediate precision of the assay method was evaluated on different day by varying different analyst, different equipment and different column. Accuracy The accuracy of an analytical procedure expresses the closeness of agreement between the true value and the observed value. The accuracy of the assay method for RVB was evaluated in triplicate (n = 3) at the three concentrations of 50, 100 and 150 μg/ml (50, 100 and 150 ) of drug product, and the recovery was calculated for each added (externally spiked) concentration. For all impurities, the recovery was determined in triplicate for 0.375, 0.75 and μg/ml (50, 100 and 150 ) of the analyte concentration (500 μg/ml) of the drug product, and the recovery of the impurities was calculated. Linearity Linearity test solutions for the related substance method were prepared by diluting stock solutions to the required concentrations. The solutions were prepared at eight concentration levels from LOQ 200 (LOQ, , 0.075, , 0.15, , and 0.30 ) with respect to the normal sample concentration (500 μg/ml). The correlation coefficients, slopes and Y-intercepts of the calibration curve were determined. Linearity test solutions for the assay method prepared from RVB stock solutions at 19
4 five concentration levels from of assay analyte concentration (50, 75, 100, 125 and 150 μg/ml). The peak area versus concentration data was treated as least-squares linear regression analysis. Limit of detection (LOD) and limit of quantification (LOQ) By injecting linear ( with respect to the test concentration) solutions of known concentrations, based on the standard deviation (σ) of the response and the slope (S) of the calibration plot and using the formula LOD = 3.3 σ/s and LOQ = 10 σ/s, the LOD and LOQ for RVB and for the ten impurities were estimated. The precision (n = 6) was also determined at the LOQ level, and the RSD was calculated for the peak area for each impurity and for RVB. Robustness The robustness of an analytical procedure is a measure of its capacity to remain unaffected by small, but deliberate variations in method parameters and provides an indication of its reliability during normal usage. To determine the robustness of the method, the experimental conditions were deliberately changed. The resolution of RVB and the ten impurities was evaluated. The mobile phase flow rate was 0.45 ml/min; to study the effect of the flow rate on resolution, the flow rate was changed to 0.40 and 0.50 ml/min. The effect of the column temperature was studied at 20 C and 30 C (instead of 25 C). The effect of the percent organic strength on resolution was studied by varying acetonitrile by - 10 to + 10 from initial composition. Solution stability and mobile phase stability The stability of RVB in solution was determined by leaving the test solutions of the sample and reference standard in tightly closed volumetric flasks at room temperature for 48 h during which they were assayed at 12 h intervals. The stability of RVB and its impurities in solution for the related substance method was determined by leaving spiked sample (with respect to the specification, i.e., 0.15 level) in a tightly capped closed volumetric flasks at room temperature for 48 h and measured the impurity content of ten impurities at every 12 h. Stability of mobile phase was demonstrated by analysis of freshly prepared sample solution at 12 h intervals for 48 h and comparing the results with those obtained from freshly prepared reference standard solutions. The prepared mobile phase was kept constant during the study period. RESULTS AND DISCUSSION Method development and optimization The main target of the chromatographic method is to achieve the separation of impurities (potential, bi-products and degradation impurities) from the main component of RVB. The maximum absorption wavelength of RVB and its impurities were observed at 248 nm, which is intersecting value obtained from the UV absorption spectra, hence, this particular wave length was selected for LC analysis. The blended solution containing 500 μg/ml of RVB and 0.75 μg/ml of each of the ten impurities was prepared in the diluent. For the initial trail Acquity BEH C8 (100-mm 2.1-mm, 1.7-μm particles) column was chosen with a mobile phase composition of buffer (ortho phosphoric acid and 1-Octane sulphonic acid sodium salt as modifier) and acetonitrile in varies ratios in gradient mode. The separation of imp-4 and imp-5, imp-9 and imp- 10 were found to be merge. Further trails were conducted with Acquity BEH C mm 2.1-mm, 1.7-μm particles) by keeping the mobile phase composition same as above. The separation of imp-9 and imp-10 was found to be good (USP resolution >5). But the separation of imp- 4 and imp-5 was found to be merge and imp-10 and analyte peak was found to be inadequate. Further trails were conducted with Acquity BEH Shield RP 18 (100- mm 2.1-mm, 1.7-μm particles) by keeping the mobile phase composition same as above. Elution order changed for imp-4, imp-5, imp-9 and imp-10.the separation between imp-9, imp-10 and analyte were found to be good (USP resolution >1.5). But imp-4 separates from imp-5and merge with imp-6 and imp-7. Further trails were conducted with Acquity BEH phenyl (100-mm 2.1-mm, 1.7-μm particles) by keeping the mobile phase composition same as above. All impurities separated well (resolution >2) except imp-2 and imp-3 were found to be merge. Separation not achieved by changing the chromatographic conditions like gradient and column temperate Further trails were conducted with Acquity HSS T3 (100- mm 2.1-mm, 1.8-μm particles) by keeping the mobile phase composition same as above. All impurities separated well (resolution >2) but resolution between imp-4 and imp- 5 is less than 1. Further tried with different gradient. All impurities separated well (resolution >2) (Fig.2). At this mobile phase mixture and column temperature, very good separations of all the impurities were seen and the method was finalized at this particular mobile phase and column temperature (Fig.2A). The elution of analyte peak was approximately 6 min. Interference with the excipients (placebo) was also checked and no interference was observed at the impurity peaks and the analyte peak. Trail-1: Column: Acquity BEH C mm, 1.7μm Trail-2: Column: Acquity BEH C mm, 1.7μm 20
5 impurities from pharmaceutical dosage forms ranged from 93.9 to Linearity The linearity calibration plot for the assay method was obtained over the calibration ranges tested, i.e μg/mL and correlation coefficient obtained was greater than The result shows that an excellent correlation existed between the peak area and concentration of the analyte. Linear calibration plot for the related substance method was obtained over the calibration ranges tested, i.e. LOQ to 0.30 for impurities. The correlation coefficient obtained was greater than (Table 2). The above result shows that an excellent correlation existed between the peak area and the concentration of Imp-1, Imp-2, Imp- 3, Imp-4, Imp-5, Imp-6, Imp-7, Imp-8, Imp-9 and Imp-10. Robustness In all the deliberate varied chromatographic conditions (flow rate, column temperature and composition of organic solvent), all analytes were adequately resolved and elution orders remained unchanged. The resolution between critical pairs, i.e. for Imp-4 and Imp-5, for Imp-5 and Imp- 6 and for Imp-9 and Imp-10 was greater than 2.0 and tailing factor for Rivaroxaban and its impurities was less than 1.2. METHOD VALIDATION RESULTS System suitability The system suitability parameters were evaluated for RVB and its ten impurities. The USP tailing factor for all ten impurities and RVB was found to be less than 1.2. The USP resolution (Rs) of RVB and the ten potential impurities was greater than 2.0 between all pairs of compounds. The USP plate count for all ten impurities and RVB was found to more than (Table 1). Precision The RSD of assay of RVB during the assay method repeatability study was 0.4 and the RSD for the area of Imp-1, Imp-2, Imp-3, Imp-4, Imp-5, Imp-6, Imp-7, Imp-8, Imp-9 and Imp-10 in related substance method repeatability study was within 6.5. The RSD of the assay results obtained in the intermediate precision study was within 0.4 and the RSD for the area of Imp-1, Imp-2, Imp-3, Imp-4, Imp-5, Imp-6, Imp-7, Imp- 8, Imp-9 and Imp-10 were well within 5.3, conforming good precision of the method. The RSD values are presented in Table 2. These results confirmed that the method was highly precise. Limits of Detection and Quantification The determined limit of detection, limit of quantification and precision at LOQ values for RVB and its ten impurities are reported in Table 2. Accuracy Recovery of RVB from pharmaceutical dosage forms ranged from 99.4 to Recovery of the ten Stability in Solution and in the Mobile Phase Difference in assay of RVB during solution stability and mobile phase stability experiments was within 1.0. No significant changes (0.02 ) in the amounts of the ten impurities were observed during solution stability and mobile phase stability experiments when performed using the related substances method. The results from solution stability and mobile phase stability experiments confirmed that standard solutions and solutions in the mobile phase were stable for up to 48 h during assay and determination of related substances. All forced degradation samples were analyzed at an initial concentration of 500 μg /ml of RVB with UPLC conditions already mentioned, using PDA detector to evaluate the homogeneity and purity of RVB peak. Degradation was not observed when RVB was subjected to hydrolytic, oxidation, light and heat conditions. Significant degradation was observed in acid (0.1N HCl at 70 C for 2 h) leading to the formation of an unknown impurities. Significant degradation was observed in base (0.1N NaOH at room temperature for 1 h) leading to the formation of an unknown impurity and impurity-10 and which are well separated from the analyte peak and RVB impurities (Fig.3A-B). Homogeneous Peak found for RVB in acid and base hydrolyzed samples (Fig.3C-D). LC-MS analysis was performed as per experimental conditions and mass of the unknown impurity formed in acid was and unknown impurity formed in Base was Known impurity formed in base hydrolysis showing the mass value at and conformed as impurity-10 (Fig.3E-G). Results from force degradation studies have been presented in Table 3. 21
6 Table 1: System suitability parameters. Compound name RT(min) RRT a Resolution b (Rs) Tailing factor (T) Theoretical plates (N) Imp Imp Imp Imp Imp Imp Imp Imp Imp Imp RVB a Relative retention times (RRT) were calculated against the retention time (RT) of Rivaroxaban. b Resolutions were calculated between two adjacent peaks. c Mean±SD. Table 2: Method validation results Method validation results Parameter RVB Imp-1 Imp-2 Imp-3 Imp-4 Imp-5 Imp-6 Imp-7 Imp-8 Imp-9 Imp-10 Precision (RSD) Intermediate precision(rsd) LOQ (μg/ml) LOD (μg/ml) Precision at LOQ Accuracy at LOQ Correlation coefficient 8 7 Slope (b) Intercept (a) Y-intercept at 100 level Accuracy to to to to to to Linearity range is LOQ- 200 with respect to 500 μg/ml RVB for impurities. Linearity range is with respect to 100 μg/ml of RVB for assay Accuracy range is with respect to specification limit to to to to Assay studies were carried out for stress samples (at 100 μg/ml) against qualified RVB reference standard. The mass balance ( assay + sum of all compounds+ sum of all degradants) results were calculated for all stressed samples and found to be more than 99. Thus confirms the stability-indicating power of the developed method. The results from system suitability all related substances well separated from RVB with symmetric peak shapes and shown in Table 1. Precision study evaluated by performing both repeatability and intermediate precision. RSD results for all related substances observed less than 6.5 indicates that the method is highly precise. The correlation coefficient was obtained greater than indicates linearity of RVB method. Accuracy performed at different levels from LOQ to 150 to the specification limit. Average recovery observed between 92.0 to indicates RVB method is highly accurate. During solution and mobile phase stability, variation in impurity content and assay is less than 0.02 and 1.0 respectively when compared to initial results. Hence RVB spiked sample, 22
7 Degradation condition Time Assay (w/w on dried basis) RS by UPLC degradation Mass balance (assay+deg. Products) Remarks/ observation Assay sample and mobile phase were stable for 48 h at room temperature. Robustness study was performed by changing minute variations in column temperature, flow rate and mobile phase composition. Satisfactory results observed w.r.t resolution and tailing. Hence method is highly robust. Results for precision, Linearity and Accuracy results were shown in Table 2. Rao, et al., Int J Res Pharm Sci 2015, 5(2) ; Table 3: Summary results of forced degradation Summary of forced degradation study Figure: 3C: Peak purity plot for RVB in acid hydrolysis HCl, 0.1N 70 C (Acid hydrolysis) NaOH, 0.1N RT (Base hydrolysis) Water hydrolysis 70 C Oxidation by H2O2,3.0,70 C Thermal (105 C) 2hr s hr h rs h rs d ays An unknown impurity was formed 99.1 An known and unknown impurity was formed 99.2 No degradation observed 99.9 No degradation observed 99.4 No degradation observed. UV at 254nm 10d ays No degradation observed. Fig: 3D :Peak purity plot for RVB in base hydrolysis 23
8 CONCLUSION The rapid gradient RP-UPLC method developed for quantitative analysis of RVB and its related substances in pharmaceutical dosage forms is precise, accurate, linear, robust, rugged and specific. Satisfactory results were obtained from validation of the method. The retention time (12 min) enables rapid determination of the related substances and RVB. This method exhibited an excellent performance in terms of sensitivity and speed. Hence this method is cost effective with respect to time and solvent consumption. The method is stability indicating and can be used for routine analysis of production samples and to check the stability of samples of RVB. ACKNOWLEDGEMENTS The authors wish to thank the management of Dr. Reddy s group for supporting this work. Authors wish to acknowledge the Process research group for providing the samples for our research. Special thanks to my colleague Mr. Varaprasad and Purshotham from R&D. science innovations and discoveries.2012; 2(1): Çelebier Mustafa, Reçber Tuba, Engin Koçak et al. RP-HPLC method development an validation for estimation of rivaroxaban in pharmaceutical dosage forms.brazilian Journal of Pharmaceutical Sciences.2013;(49): Chandra sekhar K, Satya vani P, Dhana Lakshmi A et al. A new method development and validation for analysis of rivaroxaban in formulation by RP HPLC; Research Desk. 2012; 1(1): Swarup SP, Vikram D, Nitin S. Enatiomeric separation of Rivaroxaban by a Chiral liquid chromatographic method. International Journal of Pharmacy andpharmaceutical Sciences.2015;7(2). 12. ICH Q1A (R2). Stability Testing of New Drug Substances and Products ICH Q2 (R1). Validation of Analytical Procedures. Text and Methodology REFERENCES 1. Swartz M. UPLC An introduction and review. Journal of Liquid Chromatography & Related Technologies. 2005; 28: Nguyen D TT, Guillarme D, Heinisch S et al. High throughput liquid chromatography with sub-2 micron particles at high pressure and high temperature. Journal of Chromatography A. 2007; 1167: Nov akov a L, Matysov a L, Solich P.Advantages of application of UPLC in pharmaceutical analysis, Talanta. 2006; 68: Susanne R, Alexander S, Jens P et al. Discovery of novel antithrombotic agent BAY Journal of Medicinal Chemistry. 2005; 48(19): E Perzborn, J Strassburger, A Wilmen et al. In vitro and in vivo studies of BAY Journal of Thrombosis and Haemostasis. 2005; 3: Rohde G. Determination of Rivaroxaban-a noval,oral,direct Factor Xa inhibitor in human plasma by high-performance liquid chromatography-tandam mass spectrometry. Journal of Chromatography B. 2008; 872(1-2): Satyanarayana P V V, Siva Madhavi Alavala. New spectrometric methods for the quantitative estimation of Rivaroxaban in formulations. International Journal of Research and Reviews in Pharmacy and Applied science. 2012; 2(4): Satyanarayana P V V, Siva Madhavi Alavala. RP- High-performance liquid chromatographic method for Rivaroxaban in pharmaceutical dosage forms. International journal of 24
Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation
IJPAR Vol.4 Issue 4 Oct Dec - 2015 Journal Home page: ISSN: 2320-2831 Research Article Open Access Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation V.
SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY
CHAPTER 5 SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY CHAPTER 5 Simultaneous determination of Telmisartan
Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
Research Article. Stress study and estimation of a potent anticoagulant drug rivaroxaban by a validated HPLC method: Technology transfer to UPLC
Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2015, 7(10):65-74 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Stress study and estimation of a potent anticoagulant
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
Quality by Design Approach for the Separation of Naproxcinod and its Related Substances by Fused Core Particle Technology Column
Journal of Chromatographic Science Advance Access published October 11, 2012 Journal of Chromatographic Science 2012;00:1 7 doi:10.1093/chromsci/bms162 Article Quality by Design Approach for the Separation
PVV Satyanaryana et al., IJSID, 2012, 2 (1), 226-231. International Journal of Science Innovations and Discoveries
ISSN:2249-5347 IJSID International Journal of Science Innovations and Discoveries An International peer Review Journal for Science Research Article Available online through www.ijsidonline.info RP-HPLC
ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS
VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external
A NEW METHOD DEVELOPMENT AND VALIDATION FOR ANALYSIS OF RIVAROXABAN IN FORMULATION BY RP HPLC
A NEW METHOD DEVELOPMENT AND VALIDATION FOR ANALYSIS OF RIVAROXABAN IN FORMULATION BY RP HPLC K.Chandra sekhar, P. Satya vani, A. Dhana lakshmi, Ch.LL.Devi, Anupama Barik, Narendra Devanaboyina * Department
INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE
Rajgor VM,, 2014; Volume 3(3): 188-197 INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE ANALYTICAL METHOD DEVELOPEMENT AND VALIDATION FOR THE SIMULTANEOUS ESTIMATION OF MEMANTINE HCL AND
Photolytic-Thermal Degradation Study And Method Development Of Rivaroxaban By RP-HPLC
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.5, No.3, pp 1254-1263, July-Sept 2013 Photolytic-Thermal Degradation Study And Method Development Of Rivaroxaban By
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
Journal of Chemical and Pharmaceutical Research
Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research J. Chem. Pharm. Res., 2010, 2(1): 91-99 ISSN No: 0975-7384 A Stability-Indicating RP-LC method for the Determination of Related
High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage Form
ISSN: 0973-4945; CODEN ECJHAO E- Chemistry http://www.e-journals.net Vol. 5, No.3, pp. 427-430, July 2008 High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage
International Journal of Research and Reviews in Pharmacy and Applied science. www.ijrrpas.com
International Journal of Research and Reviews in Pharmacy and Applied science www.ijrrpas.com P.V.V.Satyanarayana*, Alavala Siva Madhavi NEW SPECTROPHOTOMETRIC METHODS FOR THE QUANTITATIVE ESTIMATION OF
# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water
Application Notes # LCMS-35 esquire series Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water An LC-APCI-MS/MS method using an ion trap system
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies
HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies Application Note Small Molecule Pharmaceuticals Authors Jignesh Shah, Tiantian Li, and Anil Sharma Agilent Technologies,
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS
Analysis of the Vitamin B Complex in Infant Formula Samples by LC-MS/MS Stephen Lock 1 and Matthew Noestheden 2 1 AB SCIEX Warrington, Cheshire (UK), 2 AB SCIEX Concord, Ontario (Canada) Overview A rapid,
Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate
Cantaurus, Vol. 7, 5-8, May 1999 McPherson College Division of Science and Technology Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate Janet Bowen ABSTRACT Sulfate is used in
Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis
Application Note AN844 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Page 1 Extraction of, and from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis Introduction Catecholamines
High performance thin layer chromatographic method for estimation of deflazacort in tablet
ISSN: 2231-3354 Received on: 07-09-2011 Revised on: 11-09-2011 Accepted on: 13-09-2011 High performance thin layer chromatographic method for estimation of deflazacort in tablet Patel Satish A and Patel
International Journal of Pharma and Bio Sciences V1(2)2010
1 Central Drugs Laboratory,Kolkata Angshuman Biswas 1 and Arindam Basu 1 *Corresponding author [email protected] ABSTRACT A fast and reliable high performance liquid chromatography method for determination
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract
Using Natural Products Application Solution with UNIFI for the Identification of Chemical Ingredients of Green Tea Extract Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column
Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column Sachiki Shimizu, FUJIFILM Fine Chemicals Co., Ltd., Kanagawa, Japan Kenneth J. Fountain, Kevin Jenkins, and Yoko Tsuda,
Journal of Chemical and Pharmaceutical Research
Available on line www.jocpr.com Journal of Chemical and Pharmaceutical Research ISSN No: 0975-7384 CODEN(USA): JCPRC5 J. Chem. Pharm. Res., 2010, 2(6):62-67 Development and validation of RP- HPLC method
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis
The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis Stephen Lock AB SCIEX Warrington (UK) Overview A rapid, robust, sensitive and specific LC-MS/MS method has been developed for
Overview. Triple quadrupole (MS/MS) systems provide in comparison to single quadrupole (MS) systems: Introduction
Advantages of Using Triple Quadrupole over Single Quadrupole Mass Spectrometry to Quantify and Identify the Presence of Pesticides in Water and Soil Samples André Schreiber AB SCIEX Concord, Ontario (Canada)
HIGH PERFORMANCE THIN LAYER CHROMATOGRAPHIC METHOD FOR DETERMINATION OF TADALAFIL IN TABLET DOSAGE FORM
Volume 1, Issue 3, 2011. ISSN: 2249-3387 Journal home page: http://www.ajptr.com/ HIGH PERFORMANCE THIN LAYER CHROMATOGRAPHIC METHOD FOR DETERMINATION OF TADALAFIL IN TABLET DOSAGE FORM Satish A. Patel
Journal of Chemical and Pharmaceutical Research, 2012, 4(6):3263-3274. Research Article
Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2012, 4(6):3263-3274 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Impurities profiling Method and degradation studies
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates
Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates Paula Hong and Kenneth J. Fountain Waters Corporation, 34 Maple St., Milford, MA, USA APPLICATION
LC-MS/MS for Chromatographers
LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS, with an emphasis on the analysis of drugs in biological matrices LC-MS/MS for Chromatographers An introduction to the use of LC-MS/MS,
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)
Method OIV-MA-AS313-22 Type II method Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008) 1. Introduction
Analytical Test Report
Analytical Test Report Customer: Address (City, State): Purchase Order: Report Number: Project Number: Date Received: Date of Report: Test Location: Boulder, CO. Assay: Part Number: Amino Acids by HPLC
SUCRALOSE. White to off-white, practically odourless crystalline powder
SUCRALOSE Prepared at the 41st JECFA (1993), published in FNP 52 Add 2 (1993). Metals and arsenic specifications revised at the 63rd JECFA (2004). An ADI of 0-15 mg/kg bw was established at the 37th JECFA
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC
SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC Katja SÄRKKÄ, Kari ARINIEMI, Pirjo LILLSUNDE Laboratory of Substance Abuse, National Public Health Institute Manerheimintie,
ERDOSTEINE - MONOGRAPH.
STRUCTURAL FORMULA (±1S-(2-[N-3-(2-oxotetrahydro thienyl)]acetamido)-thioglycolic acid) C 8 H 11 NO 4 S 2 M.W. = 249.307 DESCRIPTION Color : White to ivory white Appearance : Microcrystalline powder SOLUBILITY
LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)
October 2006 RESTRICTED LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006) DRAFT FOR DISCUSSION World Health Organization 2006 All rights reserved. This draft is intended for
Analytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
Transfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System Saving time and costs
Agilent Application Solution Transfer of a USP method for prednisolone from normal phase PLC to SFC using the Agilent 126 Infinity ybrid SFC/UPLC System Saving time and costs Application Note Pharmaceutical
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma
SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from uman Plasma Krishna Rao Dara, Dr. Tushar N. Mehta, Asia Pacific Center of Excellence, Thermo Fisher Scientific, Ahmedabad, India Application
Stability-indicating High Performance Thin Layer Chromatographic Determination of Clozapine in Tablet Dosage Form
Stability-indicating High Performance Thin Layer Chromatographic Determination of Clozapine in Tablet Dosage Form Zahid Zaheer *, Mazhar Farooqui 1, S.R.Dhaneshwar 2 *Y.B.Chavan College of Pharmacy, Dr.
By T.Sudha, T.Raghupathi
Global Journal of Medical research Volume 11 Issue 2 Version 1.0 Type: Double Blind Peer Reviewed International Research Journal Publisher: Global Journals Inc. (USA) Online ISSN: 0975-5888 Reverse Phase
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions
Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions authors Angela (Qi) Shen, Ling Morgan, Marcele L. Barroso, and Xin Zhang; Tandem Labs Tuyen Nguyen; Sepracor Inc.
How To Test For Contamination In Large Volume Water
Automated Solid Phase Extraction (SPE) of EPA Method 1694 for Pharmaceuticals and Personal Care Products in Large Volume Water Samples Keywords Application Note ENV0212 This collaboration study was performed
Development and validation of UV spectrophotometric method for the determination of rivaroxaban
Available online at www.derpharmachemica.com Scholars Research Library Der Pharma Chemica, 013, 5(4):1-5 (http://derpharmachemica.com/archive.html) ISSN 0975-413X CODEN (USA): PCHHAX Development and validation
Received: 19 Jun2010, Revised and Accepted: 20 July 2010
International Journal of Pharmacy and Pharmaceutical Sciences ISSN- 0975-1491 Vol 2, Suppl 4, 2010 Research Article METHOD DEVELOPMENT AND VALIDATION FOR THE ESTIMATION OF OLOPATADINE IN BULK AND PHARMACEUTICAL
Tamsulosin Hydrochloride Capsules
. nal Revision Bulletin Official October 1, 2011 Tamsulosin 1 standard solution, and shake well. Centrifuge at 1500 rpm for 10 min, and use the supernatant, passing it if Tamsulosin Hydrochloride Capsules
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep
Guide to Reverse Phase SpinColumns Chromatography for Sample Prep www.harvardapparatus.com Contents Introduction...2-3 Modes of Separation...4-6 Spin Column Efficiency...7-8 Fast Protein Analysis...9 Specifications...10
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection
Analysis of Polyphenols in Fruit Juices Using ACQUITY UPLC H-Class with UV and MS Detection Evelyn Goh, Antonietta Gledhill Waters Pacific, Singapore, Waters Corporation, Manchester, UK A P P L I C AT
VALIDATION OF HPLC METHOD FOR DETERMINATION OF DOCOSAHEXAENOIC ACID (DHA) IN MULIIVITAMIN CAPSULE DIAH WIDOWATI*, ROS SUMARNY, ESTI MUMPUNI
VALIDATION OF HPLC METHOD FOR DETERMINATION OF DOCOSAHEXAENOIC ACID (DHA) IN MULIIVITAMIN CAPSULE DIAH WIDOWATI*, ROS SUMARNY, ESTI MUMPUNI Fakultas Farmasi Universitas Pancasila Jln. Srengseng Sawah,
Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2
Technical ote Simultaneous Metabolite Identification and Quantitation with UV Data Integration Using LightSight Software Version 2.2 Alek. Dooley, Carmai Seto, esham Ghobarah, and Elliott B. Jones verview:
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. [email protected] Overview GMP requirements for Quality Control laboratories
Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection
52 May/June 2012 Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection by Karen Rome and Allyson McIntyre, AstraZeneca, Macclesfield, SK10 2NA, UK Quantitative analysis
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
UV-VIS SPECTROPHOTOMETRIC METHOD FOR ESTIMATION OF GABAPENTIN AND METHYLCOBALAMIN IN BULK AND TABLET
International Journal of ChemTech Research CODEN( USA): IJCRGG ISSN : 0974-4290 Vol.2, No.1, pp 695-699, Jan-Mar 2010 UV-VIS SPECTROPHOTOMETRIC METHOD FOR ESTIMATION OF GABAPENTIN AND METHYLCOBALAMIN IN
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes
EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes Brussels, 18th May 2005 CED No 494 rev 2 Final TAXUD/3711/2004 POETRY:
ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
How To Use An Acquity Qda Detector
Mass-Directed Isolation of a Synthetic Peptide Using the ACQUITY QDa Detector Jo-Ann M. Jablonski and Andrew J. Aubin Waters Corporation, Milford, MA, USA APPLICATION BENEFITS The ACQUITY QDa Detector
Estimation and Stress Study of a Potent Anticoagulant Drug Rivaroxaban by a Validated HPLC Method
2 ESTIMATION AND STRESS STUDY OF A POTENT ANTICOAGULANT DRUG RIVAROXABAN BY A VALIDATED HPLC METHOD II Estimation and Stress Study of a Potent Anticoagulant Drug Rivaroxaban by a Validated HPLC Method
Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control
Automated Method Development Utilizing Software-Based Optimization and Direct Instrument Control Dr. Frank Steiner, 1 Andreas Brunner, 1 Fraser McLeod, 1 Dr. Sergey Galushko 2 1 Dionex Corporation, Germering,
Applying Statistics Recommended by Regulatory Documents
Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services [email protected] 301-325 325-31293129 About the Speaker Mr. Steven
Liquid Chromatographic Method for the estimation of Donepezil Hydrochloride in a Pharmaceutical Formulation
International Journal of ChemTech Research CODEN( USA): IJCRGG ISSN : 0974-4290 Vol.3, No.1, pp 112-118, Jan-Mar 11 Liquid Chromatographic Method for the estimation of Donepezil Hydrochloride in a Pharmaceutical
DEVELOPMENT AND VALIDATION OF A STABILITY INDICATING HPLC METHOD FOR ANALYSIS OF PEFLOXACIN IN BULK DRUG AND TABLET DOSAGE FORM
Academic Sciences International Journal of Pharmacy and Pharmaceutical Sciences ISSN- 0975-49 Vol 5, Issue 2, 203 Research Article DEVELOPMENT AND VALIDATION OF A STABILITY INDICATING HPLC METHOD FOR ANALYSIS
UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs
Application Note: 439 UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs Guifeng Jiang, Thermo Fisher Scientific, San Jose, CA, USA Key Words Accela UHPLC System MSQ Plus MS Detector Drugs
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS
A High Throughput Automated Sample Preparation and Analysis Workflow for Comprehensive Forensic Toxicology Screening using LC/MS/MS AB SCIEX QTRAP 4500 LC/MS/MS System and Gerstel, Inc. MultiPurpose Sampler
Chemical analysis service, Turner s Green Technology Group
Chemical analysis service, Turner s Green Technology Group Hourly costs, Academic collaborations leading to co-publications: 627 SEK/hr (excl. VAT) Hourly costs, Industry/Academic collaborations, no publications:
Research Article. Development and validation of a HPLC analytical assay method for dapoxetine tablets: A medicine for premature ejaculation
Available online www.jocpr.com Journal of Chemical and Pharmaceutical Research, 2014, 6(7):1613-1618 Research Article ISSN : 0975-7384 CODEN(USA) : JCPRC5 Development and validation of a HPLC analytical
WATERS QUANTITATIVE ANALYSIS solutions
More sensitivity. More speed. What more can you ask for in a complete quantitative analysis solution? WATERS QUANTITATIVE ANALYSIS solutions THE CHALLENGE OF QUANTITATIVE ANALYSIS Tandem mass spectrometry,
Vitamin C quantification using reversed-phase ion-pairing HPLC
Vitamin C quantification using reversed-phase ion-pairing HPLC Thomas Grindberg and Kristy Williams Department of Chemistry, Concordia College, 901 8 th St S, Moorhead, MN 56562 Abstract Vitamin C, an
Analysis of Vegetable Oils by High Performance Liquid Chromatography
Analysis of Vegetable Oils by High Performance Liquid Chromatography Using Evaporative Light Scattering Detection and Normal Phase Eluents Andrew J. Aubin, Cecilia B. Mazza, Donald A. Trinite, and Patricia
Overview. Purpose. Methods. Results
A ovel Approach to Quantify Unbound Cisplatin, Carboplatin, and xaliplatin in Human Plasma Ultrafiltrate by Measuring Platinum-DDTC Complex Using LC/M/M Min Meng, Ryan Kuntz, Al Fontanet, and Patrick K.
Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLC-DAD and HPLC-MS/MS
Standard Analytical Methods of Bioactive Metabolitesfrom Lonicera japonica Flower Buds by HPLCDAD and HPLCMS/MS Jongheon Shin College of Pharmacy Seoul National University Summary The standard analytical
Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna
Conversion of a USP Gas Chromatography Method to Convergence Chromatography: Analysis of Atropa Belladonna Paula Hong and Pat McConville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E
Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis
Thermo Scientific SOLAµ SPE Plates Technical Guide Consistent excellence for bioanalysis SOLAµ - delivering reproducible low volume extractions. Everytime! Thermo Scientific SOLAµ plates are designed for
GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE
ACTA CHROMATOGRAPHICA, NO. 13, 2003 GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE A. Pavlova and R. Ivanova Refining and Petrochemistry Institute, Analytical Department, Lukoil-Neftochim-Bourgas
Center for Drug Evaluation and Research (CDER) Reviewer Guidance' Validation of Chromatographic Methods. November 1994 CMC 3
Center for Drug Evaluation and Research (CDER) Reviewer Guidance' Validation of Chromatographic Methods November 1994 CMC 3 TABLE OF CONTENTS I. INTRODUCTION... 1 II. TYPES OF CHROMATOGRAPHY... 2 A. High
Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 )
Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 ) Andrew Aubin Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S Fast analysis of
α-cyclodextrin SYNONYMS α-schardinger dextrin, α-dextrin, cyclohexaamylose, cyclomaltohexaose, α- cycloamylase
α-cyclodextrin New specifications prepared at the 57th JECFA (2001) and published in FNP 52 Add 9 (2001). An ADI not specified was established at the 57th JECFA (2001). SYNONYMS α-schardinger dextrin,
Validating Methods using Waters Empower TM 2 Method. Validation. Manager
Validating Methods using Waters Empower TM 2 Method Validation Manager (MVM) Anders Janesten Nordic Informatics Sales Specialist 2008 Waters Corporation Decision-centric centric method development information
Development and Validation of RP-HPLC Method for the determination of Methylphenidate Hydrochloride in API
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.6, No.2, pp 462-467, April-June 2014 Development and Validation of RP-HPLC Method for the determination of Methylphenidate
A commitment to quality and continuous improvement
Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous
All the prepared formulations were subjected for following. evaluation parameters and obtained results were showed in Tables 6.3 &
105 6.1 CHARACTERIZATION OF TABLETS All the prepared formulations were subjected for following evaluation parameters and obtained results were showed in Tables 6.3 & 6.4. 6.1.1 Description (Size, Shape,
Jennifer L. Simeone and Paul D. Rainville Waters Corporation, Milford, MA, USA A P P L I C AT ION B E N E F I T S INT RO DU C T ION
A Validated Liquid-Liquid Extraction Method with Direct Injection of Hexane for Clopidogrel in Human Plasma Using UltraPerformance Convergence Chromatography (UPC 2 ) and Xevo TQ-S Jennifer L. Simeone
ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE
ANALYSIS OF FOOD AND NATURAL PRODUCTS LABORATORY EXERCISE Determination of carbohydrates in foodstuff (LC/RID method) Exercise guarantor: Assoc. Prof. Karel Cejpek, Ph.D. CONTENT Required knowledge...
ORIENTAL JOURNAL OF CHEMISTRY An International Open Free Access, Peer Reviewed Research Journal. www.orientjchem.org
ORIENTAL JOURNAL OF CHEMISTRY An International Open Free Access, Peer Reviewed Research Journal www.orientjchem.org ISSN: 0970-020 X CODEN: OJCHEG 2014, Vol. 30, No. (1): Pg. 395-399 High Performance Thin
Implementing New USP Chapters for Analytical Method Validation
Implementing New USP Chapters for Analytical Method Validation March 2010 Ludwig Huber Fax.: +49 7243 602 501 E-mail: [email protected] Today s Agenda Handling Method Changes vs. Adjustments
Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische Chemie, Universitätsspital Zürich, CH-8091 Zürich, Schweiz
Toxichem Krimtech 211;78(Special Issue):324 Online extraction LC-MS n method for the detection of drugs in urine, serum and heparinized plasma Daniel M. Mueller, Katharina M. Rentsch Institut für Klinische
Project 5: Scoville Heat Value of Foods HPLC Analysis of Capsaicinoids
Willamette University Chemistry Department 2013 Project 5: HPLC Analysis of Capsaicinoids LABORATORY REPORT: Formal Writing Exercises PRE-LAB ASSIGNMENT Read the entire laboratory project and section 28C
+++#,# & %!"#$%& '"#()*
+++#,#! "#$#%&"##$!'#&("&!"# $ %& %& & %!"#$%& '"#()* +++#,# Research Article Chemistry International Journal of Pharma and Bio Sciences ISSN 0975-6299 A NOVEL RP-HPLC METHOD FOR THE QUANTIFICATION OF
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification
MultiQuant Software 2.0 for Targeted Protein / Peptide Quantification Gold Standard for Quantitative Data Processing Because of the sensitivity, selectivity, speed and throughput at which MRM assays can
Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana*
Validation and Calibration of Analytical Instruments a D.Gowrisankar, b K.Abbulu, c O.Bala Souri, K.Sujana* a Department of Pharmaceutical Analysis, Andhra University, Visakhapatnam. b Department of Pharmaceutics,