Evaluation of Several Columns and Solvents for Post-Extraction Gel Permeation Chromatography (GPC) Clean-up of Fish Tissue Prior to PCB Analysis

Similar documents
Fipronil Analysis By GC/XSD following Post-Extraction Gel Permeation Chromatography Cleanup

This study was performed in collaboration with CT Laboratories of Baraboo, WI, USA.

This study was performed by Patrick Sullivan and staff at TestAmerica Analytical Testing Corp., Westfield, MA, USA

How To Test For Contamination In Large Volume Water

Automation of Solid Phase Extraction and Column Chromatographic Cleanup

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.

[ Care and Use Manual ]

SUCRALOSE. White to off-white, practically odourless crystalline powder

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

Caribbean Coastal Pollution Project (CCPP)

Thermo Scientific HyperSep Solid Phase Extraction Method Development Guide

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Analysis of Vegetable Oils by High Performance Liquid Chromatography

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum

Determination of Tricyclazole in Water Using Solid Phase Extraction and Liquid Chromatography

Experiment 5: Column Chromatography

Purification of reaction mixtures using flash chromatography.

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

LUMEFANTRINE Draft proposal for The International Pharmacopoeia (October 2006)

Aliphatic/Aromatic Separation of Extractable Petroleum Hydrocarbons in Solids or Water by Silica gel Fractionation

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

PCBs DETERMINATION IN TRANSFORMER OIL BY SPE AND GAS CHROMATOGRAPHY ANALYSIS

ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

ORGANIC SAMPLE PREPARATION

Guide to Reverse Phase SpinColumns Chromatography for Sample Prep

An Automated Method for the Selective Solid Phase Extraction of Zearalenone from Wheat Using Molecularly Imprinted Polymers

APPLICATIONS MANUAL. Ifosfamide in blood serum Ingredients in blood serum Organophosphorus pesticides in tea leaf... 5

Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Thermo Scientific SOLA SPE cartridges and plates Technical Guide. Join the revolution. unparalleled performance

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

A Complete Solution for Method Linearity in HPLC and UHPLC

METHOD 8082A. POLYCHLORINATED BIPHENYLS (PCBs) BY GAS CHROMATOGRAPHY

METHOD 8310 POLYNUCLEAR AROMATIC HYDROCARBONS

Pesticide Analysis by Mass Spectrometry

Gas Chromatography Liner Selection Guide

High-Throughput 3-D Chromatography Through Ion Exchange SPE

Oasis HLB Cartridges and 96-Well Plates

Separation of Amino Acids by Paper Chromatography

Gel Filtration Standard

Fast and Accurate Analysis of Vitamin D Metabolites Using Novel Chromatographic Selectivity and Sample Preparation

HORIZONTAL concerning the layout, editorial and technical content, this document needs revising. In order to harmonize all documents within

Thermo Scientific SOLAµ SPE Plates Technical Guide. Consistent excellence. for bioanalysis

METHOD POLYCHLORINATED BIPHENYLS (PCBs) BY GAS CHROMATOGRAPHY

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Data File. Sephadex G-25 media and pre-packed columns. Introduction. Sephadex G-25 Bead structure. Desalting/buffer exchange and gel filtration

The Use of Hydrogen Carrier Gas for GC/MS

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Introduction. The following definitions may help you better understand the components of the data report.

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Analysis of Fat-Soluble Vitamin Capsules using UltraPerformance Convergence Chromatography (UPC 2 )

SPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0)

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

Overview. Purpose. Methods. Results

Research for leachables of nuclear grade cation exchange resin

HEXANES. Insoluble in water, soluble in ether, alcohol, and acetone. Neutral to methyl orange (ph indicator) Not more than 0.

Agilent Enhanced Matrix Removal Lipid LOSE THE LIPIDS, FIND YOUR ANALYTES

Determination of Caffeine in Beverages HPLC-1

Paper Chromatography: Separation and Identification of Five Metal Cations

The basic division explosives for forensic purposes is shown in Fig. 1

CORESTA RECOMMENDED METHOD Nº 80

III Analytical Methods

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

GC/MS Analysis of Vanillin and Ethyl Vanillin in Distilled Spirits

Extraction and Properties of the Polyphenol, Catechin, as an Antioxidant

Analytical Test Report

Determination of Fat in Dried Milk Products Using Accelerated Solvent Extraction (ASE)

AUTOMATED SAMPLE PREPARATION INTELLIGENT SYSTEM SOLUTION

Innovative vs Traditional

NWTPH-HCID. Hydrocarbon Identification Method for Soil and Water

METHOD 3510C SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION

Organic Lab 1 Make-up Experiment. Extraction of Caffeine from Beverages. Introduction

Which analytical tools/instruments/techniques are used for POPs. A general walk through the analysis

α-cyclodextrin SYNONYMS α-schardinger dextrin, α-dextrin, cyclohexaamylose, cyclomaltohexaose, α- cycloamylase

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

GCxGC COUPLED TO FAST SCANNING QUADRUPOLE MS FOR TRACE ANALYSIS OF POPs

Fractional Distillation and Gas Chromatography

PURIFICATION. Preparative Chromatography Modular Solutions from Manual Systems to Full Automation. Scale Up. Reverse. Phase. Impurity Isolation

Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes

Key Words. Introduction

United States Department of Agriculture Food Safety and Inspection Service, Office of Public Health Science

A New QuEChERS Dispersive SPE Material and Method for Analysis of Veterinary Drug Residue by LC-MS-MS

For Chromatography, you must remember Polar Dissolves More, not like dissolves like.

TANNIC ACID. SYNONYMS Tannins (food grade), gallotannic acid, INS No. 181 DEFINITION DESCRIPTION

Method Development for Size-Exclusion Chromatography of Monoclonal Antibodies and Higher Order Aggregates

Transcription:

Evaluation of Several Columns and Solvents for Post-Extraction Gel Permeation Chromatography (GPC) Clean-up of Fish Tissue Prior to PCB Analysis Keywords: Automated GX-271 GPC Clean-up System, Contaminants, EnviroSep-ABC GPC Sample Clean-up Column, Fish, GC/ECD, Gel Permeation Chromatography (GPC), PCBs This study was performed in collaboration with the Wisconsin State Lab of Hygiene in Madison, WI and presented at the 2009 SETAC North America Meeting in New Orleans, LA. Introduction The determination of PCBs, as well as other environmental contaminants, in fish tissue requires extensive sample cleanup prior to analysis by gas chromatography with an electron capture detector (GC-ECD). The high fat content of fish tissue can cause buildup of nonvolatile materials on the GC injection port and the analytical column resulting in poor analytical results and high instrument maintenance costs. Gel Permeation Chromatography (GPC) is a size-exclusion clean-up procedure that uses organic solvents and a hydrophobic gel (primarily a cross-linked divinylbenzenestyrene copolymer) to separate macromolecules. It is a highly effective method for the removal of high molecular interferences such as lipids, proteins, pigments and cellular components from animal tissue. GPC clean-up is often used for the clean-up of fish tissue extracts prior to analysis for halogenated compounds such as PCBs and chlorinated pesticides. The most common GPC clean-up procedures for fish extract cleanup utilize a 700 mm x 25 mm glass column containing 70 g of Bio-Rad Envirobeads SX-3 resin with a mobile phase of 100% dichloromethane or 60 g of Envirobeads SX-3 resin with a mobile phase of 1:1 dichloromethane/cyclohexane at a flow rate of 5 ml/min. Although both columns are highly efficient in removing lipids from fish extracts, sample throughput is limited to approximately one hour per sample extract and both clean-up procedures utilize large amounts of solvent per sample.

The purpose of this study was to use a Gilson GX-271 Automated GPC Clean-up System (Figure 1) to evaluate several GPC clean-up columns and mobile phases that would give faster throughput and use less solvent for the post-extraction clean-up of fish tissue prior to PCB analysis via GC-ECD. Figure 1. Gilson Automated GX-271 GPC Clean-up System with UV Detector (part number 21110000) Experimental Conditions Materials All solvents were distilled in glass suitable for GC, HPLC, pesticide residues analysis and spectrophotometry. All reagents were ACS grade quality or better. GPC standards were prepared according to USEPA Method 3640A and contained corn oil, bis(2-ethylhexyl)phthalate, methoxychlor, perylene and sulfur. Stock solutions of PCBs included Congeners BZ #14, BZ #65 and BZ #166 obtained from Ultra Scientific. Extraction Weigh 10 g of ground fish tissue in a beaker. Fortify with 1g corn oil and add 60 g anhydrous sodium sulphate and 230 ml dichloromethane. Spike with appropriate surrogate PCB standards. Pour the mixture through a column containing Florisil topped with a 1 ml layer of anhydrous sodium sulphate. Collect eluent and evaporate to near dryness using a gentle stream of nitrogen. Reconstitute in 5 ml dichloromethane.

GPC Clean-up The following columns and conditions were employed for GPC clean-up: Table 1. GPC Clean-up Protocols Column Mobile Phase Flow Rate Injection Volume OI Analytical glass column with 60g Envirobeads SX-3 Phenomenex EnviroSep-ABC Column (21 x 350mm) with guard column 1:1 dichloromethane/cyclohexane 1:1 dichloromethane/cyclohexane 5 ml/min 5 ml 5 ml/min 5 ml Phenomenex EnviroSep-ABC Column (21 x 350mm) with guard column Jordi Flash Gel Fluorinated DVB (10 X 250mm) with guard column OI Analytical Optima Column packed with 21g Envirobeads SX-3 100% dichloromethane 5 ml/min 5 ml 100% dichloromethane 2 ml/min 1 ml 1:1 ethyl acetate/cyclohexane 5 ml/min 1 ml Column calibration used a GPC calibration standard (as described above), a Gilson 112 UV Detector set at 254 nm and Gilson TRILUTION LC software with GPC Methods (See Figure 2). Based on the UV trace, column eluate was collected just after bis(2-ethlhexyl)phthalate elution and stopped after perylene elution. After GPC clean-up, collected fractions were further cleaned up with silica gel to remove any pesticides and then concentrated with a gentle stream of nitrogen and reconstituted in appropriate solvent for GC analysis.

Figure 2. An example of a Sample List using GPC methods in TRILUTION LC v2.1 GC Analysis PCBs were analyzed with an Agilent HP5890-II GC with a DB5 column (60 m x 0.25 mm ID, 0.1uM phase) configured with ECD.

Results Table 2. GPC Column Parameters for Fish Extract Clean-up Column Dump Volume (ml) Collect Volume (ml) Total Run Time Column Lipid Loading Capacity OI Analytical Glass, 1:1 DCM/CYX 100 110 60 min 1 g EnviroSep-ABC, 1:1 DCM/CYX 75 48 32 min 0.5 g EnviroSep-ABC, 100% DCM 75 43 30 min 0.5 g Jordi Flash Fluorinated DVB, 100% DCM 18 22 30 min 0.05 g Optima Column, 1:1 CYX:ethyl acetate 45 60 28 min 0.2 g Figure 3. Chromatogram of a USEPA Method 3640A calibration standard using an EnviroSep-ABC column with a mobile phase of 100% dichloromethane. The flow rate is 5 ml/min with UV detection at 254 nm.

Figure 4. Chromatogram of a USEPA Method 3640A calibration standard using an EnviroSep-ABC column with a mobile phase of 1:1 dichloromethane/cyclohexane. The flow rate is 5 ml/min with UV detection at 254 nm. Figure 5. Chromatogram of a USEPA Method 3640A calibration standard using an OI Analytical Optima column with a mobile phase of 1:1 ethyl acetate/cyclohexane. The flow rate is 5 ml/min with UV detection at 254nm. Figure 6. Chromatogram of a fish tissue extract during GPC clean-up using an EnviroSep-ABC column with a mobile phase of 100% dichloromethane at 5 ml/min. Fraction collection time marks are noted.

Table 3. PCB Recovery in Fish (n=3) Column Mobile Phase PCB BZ #14 PCB BZ #65 PCB BZ #166 OI Analytical Glass, 60g Envirobeads SX-3 1:1 dichloromethane/cyclohexane 74.7 81.6 81.5 EnviroSep-ABC 1:1 dichloromethane/cyclohexane 80.5 86.8 92.9 EnviroSep-ABC 100% dichloromethane 87.9 86.7 91 Jordi Flash Fluorinated DVB 100% dichloromethane 61.1 58.8 62.9 OI Optima Column, 1:1 CYX/EA 1:1 ethyl acetate/cyclohexane 90 89 117.5 Figure 7. GC/ECD chromatogram of fish tissue extract after GPC clean-up showing PCB congeners.

Conclusion GPC clean-up is a highly effective tool for removal of lipids and other high-molecular weight interferences prior to analysis for PCBs by GC/ECD. All tested columns were able to separate the lipids in fish extracts from the PCBs. The traditional glass column packed with 60 g Envirobeads SX-3, 1:1 dichloromethane:cyclohexane achieved excellent separation and had the highest lipid loading capacity. However, it utilized the most solvent and had the lowest throughput. The Phenomenex stainless steel EnviroSep-ABC column, using either of the mobile phases tested, had twice the throughput of the traditional column and used half the amount of solvent. It provided very good recoveries (82.5 to 92.9%) of the PCBs tested. This column has the advantage of easily switching to a different mobile phase, if necessary. Lipid loading capacity is approximately one half that of the traditional column. The OI Analytical Optima column with a 1:1 ethyl acetate/cyclohexane mobile phase was also effective, but demonstrated more recovery variability between extracts and had a lipid loading capacity of approximately 20% of the traditional glass column. The Jordi Flash Gel Fluorinated DVB column utilized the least amount of solvent but had the lowest lipid loading capacities and the poorest recoveries (58.8 to 62.9%) compared to all the other columns tested. Throughput was the same as the Phenomenex column. GPC clean-up readily lends itself to automation and helps reduce GC maintenance costs by preventing buildup on GC injection ports and columns. Envirobeads is a registered trademark of Bio-Rad Laboratories, Inc. EnviroSep is a trademark of Phenomenex Corporation. Florisil is a registered trademark of U.S. Silica Company. Optima is a trademark of OI Analytical Corporation. Application Note ENV1201, December 2009