AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

Similar documents
Thermal Desorption Unit TDU

AppNote 6/2003. Analysis of Flavors using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

TechNote 3/2002. The Use of Different PTV Inlet Liner Types for Trapping Alkanes, Aromatics and Oxygenated Compounds During Thermal Desorption

MEPS - Micro Extraction by Packed Sorbent Online SPE for GC and LC sample preparation - Extraction to injection in a single process

A VISION to On-Line SPE PTV GC MS Determination of Organic Micro Pollutants in Surface Water

Analysis of Organophosphorus Pesticides in Milk Using SPME and GC-MS/MS. No. GCMS No. SSI-GCMS Shilpi Chopra, Ph.D.

How To Analyze Pesticide Levels In Lemon Oil

AppNote 2/2012. Determining Phenolic Compounds in Whisky using Direct Large Volume Injection and Stir Bar Sorptive Extraction KEYWORDS ABSTRACT

This study was performed by Patrick Sullivan and staff at TestAmerica Analytical Testing Corp., Westfield, MA, USA

THERMAL DESORPTION. Introduction and Principles. Focusing on Volatiles

LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research

Sorbents. [ A] [ A] aq. Strong affinity = large K D. Digital chromatography: all-or. or-nothing LC mechanism to extremes K =

Fast Continuous Online Analysis of VOCs in Ambient Air using Agilent 5975T LTM GC/MSD and Markes TD

Pesticide Analysis by Mass Spectrometry

AMD Analysis & Technology AG

# LCMS-35 esquire series. Application of LC/APCI Ion Trap Tandem Mass Spectrometry for the Multiresidue Analysis of Pesticides in Water

The Characterization of Perfume Fragrances Using GC/MS, Headspace Trap and Olfactory Port

How To Test For Contamination In Large Volume Water

Expectations for GC-MS Lab

Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

Determination of Pesticide Residues in Drinking Water Using Automated Solid-Phase Extraction and Gas Chromatography with Nitrogen Phosphorus Detection

Gas Chromatography. Let s begin with an example problem: SPME head space analysis of pesticides in tea and follow-up analysis by high speed GC.

Sensitive and Rapid Determination of Polycyclic Aromatic Hydrocarbons in Tap Water

Determination of Polycyclic Aromatic Hydrocarbons (PAH) in Food with GC-MS/MS

GC METHODS FOR QUANTITATIVE DETERMINATION OF BENZENE IN GASOLINE

Gas Chromatography Liner Selection Guide

METHOD 8310 POLYNUCLEAR AROMATIC HYDROCARBONS

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Extraction of Cannabinoids in Marijuana and Edibles by QuEChERS

AppNote 9/2013. Automated Derivatization, SPE Cleanup and LC/MS/MS Determination of Glyphosate and Other Polar Pesticides KEYWORDS ABSTRACT

Automation of Solid Phase Extraction and Column Chromatographic Cleanup

A Comparison between EPA Compendium Method TO- 15 and EPA Method 8260B for VOC Determination in Soil Gas

Enhancing GCMS analysis of trace compounds using a new dynamic baseline compensation algorithm to reduce background interference

This study was performed in collaboration with CT Laboratories of Baraboo, WI, USA.

Fractional Distillation and Gas Chromatography

AppNote 10/2012. Large Volume Injection with GC/MS Multi Column Switching for Direct Determination of Ethyl Carbamate in Alcoholic Beverages KEYWORDS

Theory and Instrumentation of GC Introduction

Determination of N-cyclohexyl-diazeniumdioxide (HDO) containing compounds in treated wood using GC-MS

Indoor Air Comfort Test Report

The Use of Hydrogen Carrier Gas for GC/MS

CCME. Reference Method for the Canada- Wide Standard for Petroleum Hydrocarbons in Soil - Tier 1 Method

APPENDIX L META ENVIRONMENTAL FORENSIC REPORTS

APPENDIX C. DATA VALIDATION REPORT

PCBs DETERMINATION IN TRANSFORMER OIL BY SPE AND GAS CHROMATOGRAPHY ANALYSIS

AN INTRODUCTION TO HEADSPACE SAMPLING IN GAS CHROMATOGRAPHY FUNDAMENTALS AND THEORY. Andrew Tipler

METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY

Application. Gas Chromatography September 1997

Name Lab #3: Solubility of Organic Compounds Objectives: Introduction: soluble insoluble partially soluble miscible immiscible

Bologna Maggio 2011; 1 MS Environmental Day

The First Quantitative Analysis of Alkylated PAH and PASH by GCxGC/MS and its Implications on Weathering Studies

Technical Report No.: B. Rev. 00 Dated Parts of Ceiling Lamp and Parts of Spot Lamp

Extraction to Injection in a Single Process

Series 2. Continuous monitoring of trace chemical vapours

Application Note. Determination of Nitrite and Nitrate in Fruit Juices by UV Detection. Summary. Introduction. Experimental Sample Preparation

Analysis of Pesticides in Vegetables Using the Agilent 1260 Infinity Analytical SFC System with Triple Quadrupole MS Detection

III Analytical Methods

Technical Report. Automatic Identification and Semi-quantitative Analysis of Psychotropic Drugs in Serum Using GC/MS Forensic Toxicological Database

ß-CYCLODEXTRIN SYNONYMS

Application Note # MS-14 Fast On-site Identification of Drugs with the mobile GC/MS system E²M

Test Report No. PAL e/sac Altenberge, Test of glassware. Sample Entry: (re-test material)

Multi-Shot Pyrolyzer EGA/PY-3030D. Flexible Versatile Reproducible

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

AppNote 3/1994. Trace Analysis of PAH s and PCB s in Soil through On-Line Direct Thermal Desorption

SCION SQ Series GC-MS

Application of Aroma Release Monitoring in Real Time as a Powerful Tool to Understand Consumer Perception

The Use of Micro Flow LC Coupled to MS/MS in Veterinary Drug Residue Analysis

Aromatic plants: from extraction to analysis

Enhancing Utility and Throughput of SPME using Advanced Automation Strategies

Online SPE-LC-APCI-MS/MS for the Determination of Steroidal Hormones in Drinking Water

Intelligent use of Relative Response Factors in Gas Chromatography-Flame Ionisation Detection

for SolEx TM HRP Cartridges SolEx TM HRP RSLC Columns

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

METHOD 550. July J.W. Hodgeson. W.J. Bashe (Technology Applications Inc.) T.V. Baker (Technology Applications Inc.)

Analyzing Small Molecules by EI and GC-MS. July 2014

A Complete Solution for Method Linearity in HPLC and UHPLC

Solid-phase microextraction: a powerful sample preparation tool prior to mass spectrometric analysis

Determination of Anabolic Steroids in Horse Urine by SPE and LC-MS/MS

ApplicationNOTE Introduction

Opiates in Urine by SAMHSA GC/MS

Analysis of Drugs by using the Ultra Inert Inlet Liners with Wool

Detection of Estrogens in Aqueous and Solid Environmental Matrices by Direct Injection LC-MS/MS

SPE and HPLC. Dr Iva Chianella Lecturer in Analytical Chemistry Cranfield Health +44 (0)

Air Sampling for Gases and Vapors. Patrick N. Breysse, PhD, CIH Peter S.J. Lees, PhD, CIH Johns Hopkins University

Quantitative analysis of anabolic steroids in control samples from food-producing animals using a column-switching LC-HESI-MS/MS assay

UHPLC/MS: An Efficient Tool for Determination of Illicit Drugs

Analysis of Liquid Samples on the Agilent GC-MS

Extraction, Derivatization and Ultra-Sensitive GC/Triple Quadrupole Analysis of Estrone and Estradiol in Human Serum

AppNote 1/2012. Rapid Cleanup and Comprehensive Screening of Pain Management Drugs in Urine using Automated Disposable Pipette Extraction and LC-MS/MS

The Automated SPE Assay of Fipronil and Its Metabolites from Bee Pollen.

Distillation Experiment

Extraction of Epinephrine, Norepinephrine and Dopamine from Human Plasma Using EVOLUTE EXPRESS WCX Prior to LC-MS/MS Analysis

Series 2. Air Server. Options for UNITY 2. A robust, cryogen-free solution for round-the-clock on-line air monitoring

AppNote 1/2014. Automated Hydrolysis, DPX Extraction and LC/MS/MS Analysis of Pain Management Drugs from Urine KEYWORDS ABSTRACT

SPE, LC-MS/MS Method for the Determination of Ethinyl Estradiol from Human Plasma

Method Development of LC-MS/MS Analysis of Aminoglycoside Drugs: Challenges and Solutions

Transcription:

AppNote 2/2 Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples Pat Sandra Department of Organic Chemistry, University of Gent, Krijgslaan 281 S4, B-9 Gent, Belgium Erik Baltussen Laboratory of Instrumental Analysis, Eindhoven University of Technology, P.O. Box 513, NL-56 MB Eindhoven, The Netherlands Frank David Research Institute for Chromatography, Kennedypark 2, B-85 Kortrijk Belgium Andreas Hoffmann Gerstel GmbH & Co.KG, Aktienstrasse 232-234, D-45473 Mülheim an der Ruhr, Germany INTRODUCTION For the analysis of organic compounds in aqueous environmental matrices several sample preparation methods are used. Volatile organic compounds can be analysed using (static) headspace, purge and trap or liquid-liquid extraction. Semivolatiles are analysed after liquid-liquid extraction or solid phase extraction (SPE). During the past years, miniaturisation has become a dominant trend in analytical chemistry. Typical examples of miniaturisation in sample preparation techniques are micro liquid-liquid extraction (in-vial extraction), ambient static headspace and disk cartridge SPE. In combination with stateof-the-art analytical instrumentation, this resulted in faster

analysis, higher sample throughput, lower solvent consumption, less manpower in sample preparation while maintaining or even improving sensitivity. Some 1 years ago, a new micro-extraction method was developed by Arthur and Pawliszyn [1], namely solid-phase micro-extraction (SPME). SPME can be used for the extraction of organic compounds from aqueous samples or from a gas (headspace) phase. Mostly used are fibers coated with 7-1 µm polydimethylsiloxane (PDMS) as extraction medium. SPME is by nature an equilibrium technique, based on the partitioning of the solutes between the silicone phase and the aqueous (and/or gas) matrix. Recent studies [2-4] have correlated this equilibrium with octanol/water distribution coefficients (k (o/w) ). These studies demonstrated that for solutes with low k (o/w) (k<1) low recoveries are obtained. This is main1y due to the phase ratio between the aqueous and PDMS phase. The amount of PDMS used in SPME is typically in the order of.5 µl or less, thereby limiting the enrichment on the PDMS fiber. Based on these observations, a new approach using stir bars coated with PDMS was recently developed [5]. In this approach, 5-3 µl PDMS coatings are used. Consequently, the sensitivity is increased by a factor of 1 to 1. Complete recovery is possible for solutes with k (o/w) larger than 5. Solutes with k (o/w) from 1 to 5 can also be analysed using calibration as is done in SPME. This technique is called stir bar sorptive extraction (SBSE). A theoretical comparison between SPME and SBSE is given in the graph below showing the recovery (complete extraction of solutes into the PDMS phase) in function of the octanol - water distribution coefficient. While high recoveries (> 5%) are only obtained for solutes with k (o/w) > 1 using SPME, the recovery obtained by stir bar sorptive extraction is already higher than 5 % for solutes with k (o/w) > 1. Recovery (%) 1 9 8 7 6 5 4 3 2 1 SBSE SPME 1 1 1 1 1 1 PDMS coated stir bars are now commercially available (Twister TM, Gerstel GmbH). Magnetic stirring rods are incorporated in a glass jacket and coated with a 1 mm layer of PDMS. Two Twisters are available : 1 mm L x 3.2 mm o.d. and 4 mm L x 3.2 mm o.d. PDMS coated stir bars. Typically the 1 mm stir bars are used for 1-5 ml sample volumes and the 4 mm stir bars are used for 1-25 ml sample volumes. PDMS Glass Magnet EXPERIMENTAL Sample preparation. Sample extraction is performed by placing a suitable sample amount (typically 1-25 ml) in a vial, adding a stir bar and stirring during 3-12 min. After extraction, the stir bar is introduced in a glass thermal desorption tube (4 mm i.d. x 187 mm L), placed in a thermal desorption unit and thermally desorbed. Desorption temperatures are application dependant and are between 15-3 C during 5-15 min. Alternatively liquid desorption can be used. Instrumentation. For thermal desorption a TDS-2 system (Gerstel GmbH) is used. The thermal desorption unit is mounted on a HP 689 GC (Agilent Technologies, Wilmington, DE, USA) equipped with a CIS 4 PTV inlet. The desorbed solutes are cryofocussed in the PTV inlet. After desorption, the PTV is programmed to inject the solutes and analyse them on a capillary column. RESULTS AND DISCUSSION As predicted by theory, the absolute recovery obtained by stir bar sorptive extraction is higher than the recovery obtained by SPME. This is clearly illustrated by the analysis of a water sample spiked with polyaromatic hydrocarbons. For the SBSE experiment, a 6 ml water sample (5% methanol added to prevent wall adsorption) was spiked with 3 ng/l PAHs and extracted using a 1 mm stir bar. For SPME, a 6 ml sample spiked at 3 ng/l was extracted using a 1 µm PDMS fiber. Extraction time was 3 min and stirring speed was 14 rpm for both techniques. Desorption was done at respectively 25 C (SBSE) AN/2/2-2

and 3 C (SPME). The chromatograms are shown below. With SBSE, all compounds are extracted to a similar extend, while in SPME the more apolar (higher boiling) solutes are extracted in higher amounts than the more polar (more volatile) PAHs. The overall sensitivity using SBSE is also significantly higher (see Y-axis scale and the sample is 1 times more diluted for SBSE). Abundance 1 8 6 4 2 5. 7.5 1. 12.5 14. 24 2 16 12 8 4 1 1 2 2 3 5 5. 7.5 1. 12.5 14. Figure1. Comparison of relative recovery obtained for polycyclic aromatic compounds using SBSE versus SPME (Peaks: 1. Naphthalene, 2. Acenaphthylene, 3. Acenaphthene, 4. Fluorene, 5. Phenanthrene, 6. Anthracene, 7. Fluoranthene, 8. Pyrene). 4 6 3 5 6 4 7 7 8 8 SBSE SPME 1 fold concentration Stir bar sorptive extraction can be used for both volatile and for semivolatile solutes. In this way, the same technique can cover EPA methods 524 and 624 (volatiles by GC-MS) AND methods 525 and 625 (semivolatiles by GC-MS). A typical analysis of a water sample spiked at 1 ppb level with a mixture of semivolatile priority pollutants is given in the chromatogram below. A 1 ml sample was extracted using a 1 mm stir bar during 4 min with stirring at 14 rpm. The stir bar was then thermally desorbed at 25 C during 1 min in splitless mode. The desorbed solutes were cryofocussed in a CIS 4 PTV inlet at -5 C. After desorption the PTV was programmed to 3 C and the compounds were analysed on a 3 m x.25 mm i.d. x.25 µm HP-5 MS column. Detection was done using MS in scan mode. Compounds ranging from N-nitrosodimethylamine (peak1) to chrysene (peak 35) are easily detected with LODs from 5 ng/l (for most polar nitrosamines) to 1 ng/l (for PAHs, phthalates, organochlorine pesticides, etc.). AN/2/2-3

Figure 2. Analysis of water sample spiked with semivolatiles at 1 ppb level. The analysis of volatiles by SBSE is given in the following chromatograms. A 1 ml sample was spiked at 2 ppb level and extracted using a 1 mm stir bar during 4 min with stirring at 14 rpm. The stir bar was then thermally desorbed at 25 C during 1 min in splitless mode. The desorbed solutes were cryofocussed in a CIS 4 PTV inlet at -15 C. After desorption the PTV was programmed to 25 C and the compounds were analysed on a 3 m x.25 mm i.d. x 1 µm HP-1 column. Detection was done using MS in scan mode. Compounds ranging from 1,1,1-trichloroethane (peak 1) up to 1,2,3-trichlorobenzene (peak 44) are easily detected with LODs from 1-1 ng/l. Figure 3. Analysis of water sample spiked with volatile priority pollutants at 2 ppb level using SBSE - thermal desorption - GC-MS. AN/2/2-4

The limits of the technique are illustrated by the analysis of a surface water sample. Stir bar sorptive extraction was done with a 4 mm stir bar in a 2 ml sample during 6 min at a stirring speed of 1 rpm. The analysis was performed with the MS in scan mode. Volatile aromatic hydrocarbons are detected at 5 ppt level and dichlorobenzenes are detected at 1 ppt level. Abundance 5 T, 4 ppt 4 B, 5 ppt 1,3DCB, 2.1 ppt 3 2 D 3 Ion 78 Ion 91 D 3 1,4DCB, 1 ppt Ion 146 1 A6 A7 A8 TIC 5. 7.5 1. 12.5 15. 17.5 2. D4 A9 A1 D 5 Figure 4. Analysis of surface sample by SBSE - thermal desorption - GC-MS (Peaks: B: benzene, T: toluene, 1,3-DCB: 1,3-dichlorobenzene, 1,4-DCB: 1,4-dichlorobenzene). CONCLUSIONS Stir bar sorptive extraction is a powerful technique for the extraction and analysis of both volatile and semivolatile organic compounds in aqueous environmental samples. In comparison with SPME, higher recoveries and higher sensitivity are obtained. REFERENCES [1] Arthur, C.L. and Pawliszyn, J., Anal. Chem. 199, 62, 2145. [2] Dugay, J.; Miege, C; Hennion, M.-C., J Chromatogr., A 1998, 795, 27. [3] De Bruin, L.S.; Josephy, P.D.; Pawliszyn, J.B., Anal Chem. 1998, 7, 1986. [4] Beltran, J.; Lopez, F.J.; Cepria,.; Hernandez, F., J. Chromatogr., A 1998, 88, 257. [5] Baltussen, E., Sandra, P., David F., and Cramers, C., J. Microcolumn Separations, 1999, 11, 737. AN/2/2-5

Aktienstraße 232-234 D-45473 Mülheim an der Ruhr +49 () 28-7 65 3- +49 () 28-7 65 3-33 gerstel@gerstel.de www.gerstel.de www.gerstel.com GERSTEL Inc. Caton Research Center 151 Caton Center Drive Suite H Baltimore, MD 21227 United States of America +1 (41) 247 5885 +1 (41) 247 5887 info@gerstelus.com www.gerstelus.com Technisches Büro Berlin Marburger Straße 2 1789 Berlin Telefon ( 3) 21 9 98 28 Fax ( 3) 21 9 98 27 e-mail: tb_berlin@gerstel.de Technisches Büro Bremen Parkallee 117 2829 Bremen Telefon (4 21) 3 47 56 24 Fax (4 21) 3 47 56 42 e-mail: tb_bremen@gerstel.de Technisches Büro Karlsruhe Greschbachstraße 6a 76229 Karlsruhe Telefon (7 21) 9 63 92 1 Fax (7 21) 9 6392 22 e-mail: tb_karlsruhe@gerstel.de Technisches Büro München Stefan-George-Ring 29 81929 München Telefon ( 89) 93 8 65 14 Fax ( 89) 93 8 61 9 e-mail: tb_muenchen@gerstel.de GERSTEL, GRAPHPACK und TWISTER sind eingetragene Warenzeichen der. GERSTEL, GRAPHPACK and TWISTER are registered tradenames of Printed in the Fed. Rep. of Germany Copyright by