Instruction Manual. Model FU20 ph/orp Combination sensor. IM 12B6J3-02E-E 7th edition



Similar documents
General Purpose ph/orp Sensor

General Purpose ph/orp Sensors

Evolving your plant with Field Digital Solutions

Industrial Wireless Solutions - Advantages and Disadvantages

PART 1 - INTRODUCTION...

ph Measurement and Control

JUMO tecline HD ph and JUMO tecline HD Rd

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

User Guide. ExStik TM ph (PH100 & PH110) and ORP (RE300) Waterproof Series Pens. Patent Pending

Combination ph/redox (ORP) Sensors

Trace Dissolved Oxygen Sensor

ExStik TM ph Waterproof Meters

SenTix 950 SenTix 980

Milwaukee USER MANUAL

Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes

JUMO BlackLine CR-GT/-EC/-GS Conductive 2-Electrode Conductivity Sensor

Measurement by Ion Selective Electrodes

Installation Instructions / Warranty

OxyGuard. Dissolved Oxygen Probe

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

CHEMICAL CLEANING UNIT Product number:

Global Water Instrumentation, Inc.

BLUE RIBBON CORP. BC001 Birdcage Installation Manual

Submersion/Insertion Toroidal Sensor

Care and use guide. Contents. a handy solution made easy. page

Common Problems with Online Water Quality Analyzers. Greg Macy Director AquatiPro LLC

TB2CS Conductivity sensor 2-Electrode. Superior accuracy in low level conductivity measurement ranges

CONDUCTIVITY SENSOR. Description D0382. Figure 1. The Conductivity Sensor

Network-based Control System STARDOM. Network-based Control System. Bulletin 34P02A00-11E. Versatile Data Server Software

Eutech 2700 Series. Bench Meters. ph ORP ION Conductivity TDS Salinity Resistivity Dissolved Oxygen Temperature

Ozone Sensor. Model 499A OZ CAUTION CAUTION. Instruction Sheet PN 51A-499AOZ/rev.F January 2007

OXIGENO DISUELTO. Dissolved Oxygen. Install & forget. Sistema autolimpiable. Autoclean system

Features, Limitations & Potential Errors

User's Manual. Heavy Duty Dissolved Oxygen Meter Model

Dissolved Oxygen Sensors

INSPECTION, MAINTENANCE AND RECHARGE SERVICE MANUAL P/N 16303

Plastic tube flowmeters Series PT/PS

SINGLE RANGE DS METERS Operation Instructions

Select Radiators Installation Guide

Instruction Manual Temp 4/5/6 Temperature Meter Series

REDOX AND REFERENCE ELECTRODE USER MANUAL

REPAIR AND MAINTENANCE MANUAL

Conductive Level Switch FK 2 Technical Data:

How To Control A Power Supply With A Mini Transmitter And Switch (Power Supply)

Arecont Vision H.264 Color or Day/Night SurroundVideo Series Installation Manual

Cables. Accessories. For Force, Torque and Strain Sensors

User Guide IM/AP200 Rev. R. AP200 series Rugged ph/redox (ORP) sensor systems with rapid temperature response for critical processes

Toroidal Conductivity Sensor

Global Water Instrumentation, Inc.

Subminiature Load Cell Model 8417

Precision Miniature Load Cell. Models 8431, 8432 with Overload Protection

ENDURANCE Conductivity Sensors

ph Measurements of Common Substances

Models SFP33 & SFP32 Pressure Transmitters

BASYS INSTALLATION INSTRUCTIONS FOR ELECTRONIC CAPACITANCE SENSING LAVATORY FAUCETS EFX-3XX Pedestal, Sensor Activated Lavatory Faucets

Product Data. Model 575 Series Submersible Level Transmitters ELECTRONIC PRESSURE MEASUREMENT PRODUCTS DESCRIPTION FEATURES APPLICATIONS

AKRAPOVIC SLIP-ON EXHAUST SYSTEM for the HONDA CB1000R

Convection Ovens. BX Classic. Models BX4 / FG 189C, BX / FG 158C, BX10 / FG 180C. Operator's Manual

Digital inductive conductivity transmitter

CSF16 and CSF16T Stainless Steel Steam Filters

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

SITRANS F flowmeters SITRANS F M

Instruction Manual HI Portable Waterproof Microprocessor Dissolved Oxygen Meter

STC 2P050 Series Solenoid Valves NSF Compliant Materials, 2 Way, Direct Acting

81000 Series Aluminum Gate Valve Installation and Repair Manual

General Specifications

Pipe Valves, Fittings and Nipples Pressures to 15,200 psi

Leak Detection System

Installation Instructions / Warranty

Introducing an All New Line of Cleaning Chemicals

Thermo Scientific Rotor Care Guide. Protect your rotor investment

For additional information, please visit our website at

Water Leak Detection System

Data Sheet PS1B Pressure Switch

Retractable Housings. InTrac 781 / 784 Versatile and Rugged For the Toughest Process Conditions INGOLD

Full Technical Catalog For Closed-loop Sampling System

XONYX Solid Surface Owner's Manual

1805 Series Relief Valves

Boiler Control Solution. Boiler Control Solution. Instruments and Solution for Automatic Boiler Control LF3BUSS02-00EN.

Technical data sheet TDS0070

Pro1030 USER MANUAL. English

Stainless Steel Single and Dual Circulation Kits

Professional Plus. Calibration Tips

The Control of ph and Oxidation Reduction Potential (ORP) in Cooling Tower Applications By Charles T. Johnson, Walchem Corporation

PRETREATMENT TECHNICAL DATA SHEET WATER EMULSIFIABLE RUST PREVENTATIVE PRODUCT DESCRIPTION

Instructions Manual. Electromagnetic sensor Series FLOMID FX. instrumentation for fluids. R-MI-FlomidFX Rev.: 0 English version

Direct ISE Method Method to 1000 mg/l Na + Sodium ISE

Series 510 Submersible Level Transmitters

Omega Pipettor. Operating Instructions

RUBBER SLEEVE KNIFE GATE VALVE

TABLE 1. Rugged cast aluminum alloy casing,celcon waterways with nichrome parts White (unless stainless steel housing)

LOSB9595 1/36 th Brushless ESC and 10250Kv Motor Combination

Software Maintenance Services

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

The Theory of ph Measurement

ITE. Indirect calorifier ITE - 400/500/600/750/1000. Installation, User and Service Manual. Innovation has a name.

Signal Manager Installation and Operation Manual

Fluid Level Gauge Fluid Level Sensor Temperature Switch FSA / FSK / TS

Thermostatic valve Type AVTA

Transcription:

Instruction Manual Model FU20 ph/orp Combination sensor 7th edition

TABLE OF CONTENTS 1. General............................................................ 1 1-1. Introduction........................................................ 1 1-2. Unpacking and Checking.............................................. 1 1-3. Warranty and Service................................................. 1 2. General Specifications FU20........................................... 2 2-1. Measuring elements.................................................. 2 2-2. Construction materials................................................ 2 2-3. Functional specifications (at 25 C)....................................... 2 2-4. Dynamic specifications................................................ 2 2-5. Operating range..................................................... 2 3. Installation Site FU20................................................. 3 3-1. Typical Installation................................................... 3 3-2. Preparing the Sensor for Use........................................... 3 3-3. Mounting the Sensor into the Process.................................... 3 4. Wiring............................................................. 7 4-1. Conventional ph (& ORP) Wiring........................................ 7 4-2. Wiring for ORP measurement with normal reference......................... 7 4-3. Wiring for ORP measurement with ph reference............................ 7 5. General Calibration & Maintenance Procedure........................... 8 5-1. Calibration for ph measurement........................................ 8 5-2. Calibration for ORP measurement....................................... 8 5-3. Maintenance....................................................... 8 6. Model Codes........................................................ 9 7. Spare Parts........................................................ 10

1 1. GENERAL 1-1. Introduction This instruction manual provides information for the installation and use of the FU20, four-in-one wide body ph sensors. The FU20 is the choice for the majority of typical waste water and process applications. The Model FU20 offers a simple and cost effective solution for a wide variety of waste water and process applications. This all-in-one sensor provides simultaneous measurement of ph, redox (ORP) and temperature if desired. The rugged Ryton body is designed for easy installation into on-line and immersion applications via the 3/4 NPT threaded connections provided on both ends of the sensor. Optional quick-removal adapters in both stainless steel and titanium are available to make calibration and maintenance even easier. The FU20 comes with an integral cable of optional lengths designed to minimize the adverse affects noise can have on the measurement. This cable is pre-finished to make connection to any industrial ph instrument as simple as possible. The FU20 can also include a multipole VP connector and then requires seperate cable type WU10-V-S that is available in 5 fixed lengths. 1-2. Unpacking and Checking Upon delivery, unpack the sensor carefully and inspect it to ensure it was not damaged during shipment. If damage is found, retain the original packing materials (including outer box) and then immediately notify the carrier and the relevant Yokogawa sales office. Make sure the Model Code on the sensor is the same as on the packing list. Also, check any option(s) that were ordered are included and correct. The sensor has an identification number on the cable with the full Model Code and a Serial Number included. 1-3. Warranty and Service Yokogawa products and parts are guaranteed free from defects in workmanship and material under normal use and service for a period of (typically) 12 months from the date of shipment from the manufacturer. Individual sales organizations can deviate from the typical warranty period, and the conditions of sale relating to the original purchase order should be consulted. Damage caused by wear and tear, inadequate maintenance, corrosion, or by the effects of chemical processes are excluded from this warranty coverage. In the event of warranty claim, the defective goods should be sent (freight paid) to the Service Department of the relevant sales Organization for repair or replacement (at Yokogawa s discretion). The following information must be included in the letter accompanying the returned goods: 1. Model Code and Serial Number. 2. Original Purchase Order and Date. 3. Length of time in service and description of the process. 4. Description of the fault and circumstances of the failure. 5. Process/environmental conditions that may be related to the failure of the sensor 6. Statement as to whether warranty or non-warranty service is requested. 7. Complete shipping and billing instructions for return of material, plus the name and phone number of a contact person that can be reached for further information. Returned goods that have been in contact with process fluids must be decontaminated and disinfected prior to shipment. Goods should carry a certificate to this effect, for the health and safety of our employees. Material Safety Data sheets must be included for all components of the process to which the sensors/fittings have been exposed.

2 2. GENERAL SPECIFICATIONS FU20 2-1. Measuring elements : ph glass electrode : Silver chloride reference : Solid platinum electrode : Pt1000 temperature sensor. 2-2. Construction materials Body Earthing pin O-ring Reference junction Cable : PPS 40GF (Ryton with glass filling) : Solid Platinum : Viton : Porous PTFE : Coaxial with 4 extra leads 2-3. Functional specifications (at 25 C) Isothermal point : ph 7 Glass impedance - Dome shape : 250 to 500 MΩ - Flat Surface : 400 to 1000 MΩ Reference system : Ag/AgCl with saturated KCI Junction resistance : 1.0 to 10 kω Liguid outled : non-flow double junction Temperature element : Pt1000 to IEC 751 Asymmetry potential : 8 ± 15 mv Slope : > 96 % (of theoretical value) NOTE: The temperature sensor included in the FU20 is intended to provide indication and cell compensation. The construction has not been tested to the pressure vessel standards required for plant temperature control. 2-4. Dynamic specifications Response time ph step (7 to 4) : < 15 sec for 90% Response time temp step (10 C) - Dome shape : < 3 min for 90% - Flat Surface : < 6 min for 90% Stabilisation time (0.02 ph unit/ 10 s) : < 2 minutes 2-5. Operating range ph : 0 to 14 Temperature : -10 to 105 C (14 to 221 F) Pressure : 0 to 10 bar ( 0 to 145 PSIG) Conductivity : > 50 µs/cm Storage temperature : -10 to 50 C (14 to 122 F) NOTE: Specifications should not be considered in isolation. For example the ph range can be 2-12 ph, where the measurement is at elevated temperatures. For advice about specific applications please contact you local sales office.

3 3. INSTALLATION OF FU20 For optimum measurement results, the FU20 should be installed in a location that offers an acceptable representation of the process composition and DOES NOT exceed the specifications of the sensor. The FU20 is designed with threads on either end of the body 3 / 4 NPT to allow installation in a wide variety of applications. 3-1. Typical Installation The FU20 sensor is designed for in-line installation, via a bypass loop as an immersion assembly. For best results the FU20 should be mounted with the process coming towards the sensor (Figure 2) and positioned at least 15 above the horizontal plane to eliminate air bubbles in the glass bulb. Next, slide the mating part of the adapter from the cable end of the FU20 (step 4) sensor and mount into the adapter making sure that the O-ring seats properly (step 5 & 6). Hand tightened the adapter nut (step 7). NOTE: DO NOT over tighten the sensor, as damage may result. See Section 4 for instructions to wire the sensor to the instrument. 3-2. Preparing the Sensor for Use Remove the sensor from its shipping box and slide off the wet pocket (the tube filled with solution to keep the measuring elements from drying out in storage or shipment). FU20 IMPORTANT NOTICE Before mounting the sensor, it should be calibrated. The calibration procedure is normally described in the Instruction Manual of the ph transmitter, but a general calibration is given in section 5 of this manual. TeflonTape 3.3 Mounting the Sensor The FU20 can be mounted using the threads on the body of the sensor (Figure 1). Apply Teflon tape to the appropriate threaded end, then install in the process. The FU20 can also be mounted using the threaded end of the quick release adapter. Apply Teflon (PTFE) tape to the threaded end of the quick release adapter (step 2). Then install the adapter (/NSS, /NTI, /BSP or /BTI) in the process, tightening it by applying a suitable torque using a spanner on the adapter flats, steps 1 & 3. FU20 Figure 1. Sensor Installation

4 Process Flow Wrench Flats Do NOT overtighten Teflon Tape Teflon Tape Figure 2. Sensor & adapter Installation

5 FU20 + FD40 FU20 + ADAPTER /SN3 /SR3 /FN4 /FR4 Fitting Process Flow O-ring Adapter ( /FPS ) /HCNF, SPRAY CLEANER Cleaning Solution (water, acid, etc.) FU20 O-ring Figure 3. Installation examples for the FU20

Ø 22.0 (0.86) Ø 26.0 (1.02 ) 6 3/4 TAPERED THREAD 3/4 TAPERED THREAD -NPT 19.7(0.78 ) 1 NPT or ISO 7/1-R1 19.7 (0.78 ) 22.5 (0.89 ) 51.0 (2.3 ) /NSS, /NTI, /BSS, /BTI 150.0 (5.9 ) Ø 35.5 (1.4 ) 10.0 (0.39 ) 99.0 (3.9 ) -FSM Figure 4. Dimensions FU20 /NSS, /NTI, /BSP, /BTI 1 NPT or ISO 7/1-R1 25.0 (1 ) Ø 39.0 (1.54 ) Figure 5. Dimensions 1 FU20 adapter Stainless Steel & Titanium /FPS Figure 6. Dimensions FU20 adapter for FF40, F S40 and FD40 fittings (K1523DD)

7 4. WIRING FU20 Red (11), Temp-1 Blue (12), Temp-2 Black (13), Reference Clear (15), ph Brown (16), ph shield Yellow (14), ORP/LE ø6.5 MODEL CODE Serie nummer VP-connector WU10 (ph) A (Ref.) B (Temp)F (ph shield) C (Temp) E (ORP/LE) D (Temp-1) Red E (Temp-2 ) Blue F (ph shield) Black C (ph) Clear A (Ref.) Brown B (ORP/LE) Yellow D ø6.5 MODEL CODE Serie nummer 4.1. Cable marking and pin allocation. Signal description FU20 cable Wire color Cable wires (&EXA terminal) VP connector number Pin WU10-V cable Wire color ph Clear 15 A Clear ph-shield Brown 16 C Black Reference Black 13 B Brown ORP/Liquid earth Yellow 14 D Yellow Temp-1 Red 11 E Red Temp-2 Blue 12 F Blue 4.2. Instrument connection of sensor as function of measurement Function ph ph&orp ph&rh ORP ph compen- Signal descr. sated ORP ph 15 15 15-13 Shield of ph 16 16 16-17 REF 13 13 13 13 - ORP/Liquid Earth 14 14 14 15 15 TEMP 11 11 11 11 11 TEMP(2) 12 12 12 12 12 Jumper - - - 13-14 14-15 The settings for the ph converters must be set according to the instructions in the Instruction Manual of the converter. In the following table these settings are summarized.

8 4.2.1. Settings for PH402G and PH202G(S) Function ph ph compensated Setting (default) ph&orp ph&rh ORP ORP code 01 0(pH) 0(pH) 0(pH) 1(ORP) 1(ORP) code 02 0(off) 1(ORP) 2(rH) 0(off) 0(off) code 03 1.1.1 1.1.1 1.1.1 0.0.1 0.0.1 code 04 0.0.1 0.0.1 0.0.1 0.0.1 1.1.1 code 31 0 2 2 jumper input 1 no no no yes yes jumper input 2 yes yes yes yes no 4.2.2. Settings for PH150 and PH450G Function ph ph compensated Setting (default) ph&orp ph&rh ORP ORP sensor setup ph ph+orp ph+orp ORP ORP measurement setup ph+orp ph+rh Impedance setting input1 High High High Low Low input 2 Low Low Low Low High jumper input 1 no no no yes yes jumper input 2 yes yes yes yes no 5. GENERAL CALIBRATION & MAINTENANCE PROCEDURE 5-1. Calibration for ph measurement To calibrate a ph sensor, two buffer solutions with known ph values are required. It is recommended that one buffer solution has a value near to ph 7.00. Depending on the process value to be measured, the second buffer solution should be either acidic (below 7.00) or alkaline (above 7.00). Normally, the IEC buffers (4.01, 6.87 and 9.18) are used. The following is a very general 2-point calibration procedure. 1. Clean the sensor (deposits may be removed using a 5% (approximate) solution of HCl). 2. Rinse sensor thoroughly with clean water. 3. Immerse the sensor in the first buffer (6.87 ph is recommended). 4. Refer to appropriate Instrument Instruction Manual for Calibration procedures (Auto; Manual; Sample, etc.), 5. Rinse sensor thoroughly with clean water. 6. Immerse the sensor in the second buffer (4.01 or 9.18 recommended). 7. Repeat Step 4. During calibration the temperature compensation should be active. The instrument automatically compensates for the sensitivity change of the ph sensor at different temperatures. After the calibration is complete replace or re-install the sensor into the process. 5-2. Calibration of ORP and rh measurements For calibration of ORP and rh the procedures for MANUAL CALIBRATION can be used as described in the instruction manual of the analyser. The commonly used standards for ORP and rh calibration are made from Chinhydron (Quinhydrone) powder dissolved in ph buffer solutions. (1 gram/200 ml) The ORP value is a function of the Reference cell of the measurement and the ph value of the buffer solution.

9 The FU20 has a reference system of Saturated Silver/Silver Chloride and the values are given in the following table as function of the measurement function ( 4.2) Function ORP ph comp. ORP rh ph= 1,68 403 88 23,6 ph= 4,01 265 88 23,6 ph= 6,87 96 88 23,6 ph= 7,00 88 88 23,6 The accuracy of the standard values is approximately +/- 10 mv. 5-3. Maintenance All ph sensors require routine maintenance to keep the measuring elements clean and functioning. Depending on the process, different cleaning solutions may be required. EXAMPLES: 1. Deposits of limes, hydroxides or carbonates can be removed by immersing the sensor in a solution containing dilute hydrochloric acid (5% is recommended). Afterwards rinse with water. 2. Deposits of oil and fat can be removed with hot water with a detergent. When the results are unsatisfactory, a mild (carbonate based) abrasive can be used. 3. Protein deposits should be removed with a protein enzymatic solution. For instance a solution containing 8.5ml concentrated hydrochloric acid and 10 grams of pepsin in 1 liter of water will do. After cleaning the probe is reinserted into the process by following the reverse procedure. NOTE: Avoid using non-polar solvent like tri-chloro ethylene, toluene or hexane. Cleaning with iso-propanol or methanol is acceptable. The non-polar solvents will break up the gel-layer on the glass bulb and requires that the electrode to be soaked in water for at least 12 hours before functioning normally again. The Teflon (PTFE) diaphragm of the sensor can sometimes be regenerated by putting it in hot (60 to 80 C) 3 molar Potassium Chlorine (KCl) solution and letting it cool to room temperature. This procedure clears the junction of process material and then draws the conductive (KCl) back into the junction. 6. MODEL CODES Model Code Suffix Code Option Description FU20 Wide body sensor Cable length -03 3 mtr -05 5 mtr -10 10 mtr -20 20 mtr -VP Vario Pinconnector (no cable) Temp. element -T1 Pt1000 Model -NPT Dome shape model -FSM Flat surface model Options /HCNF Hastelloy cleaning system /FPS Adapter F*40 from noryl /NSS 1 NPT adapter, SS (316L) /NTI 1 NPT adapter, Titanium /BSS 1 BSP adapter, SS (316L) /BTI 1 BSP adapter, Titanium

7. SPARE PARTS Spare part Description FU20 K1523DD /FPS Adapter for FF40, FS40 and FD40 fittings (PPO) K1547PK /NSS 1 NPT, Stainless Steel adapter (Viton O-ring) K1547PL /BSS ISO 7/1-R1, Stainless Steel adapter (Viton O-ring) K1547PM /NTI 1 NPT, Titanium adapter (Viton O-ring) K1547PN /BTI ISO 7/1-R1, Titanium adapter (Viton O-ring) K1500 FR Viton O-rings 29.82*2.62 (5 pcs) for 1 adapter K1500FS EPDM O-rings 29.82*2.62 (5 pcs) for 1 adapter K1500FT Silicone O-rings 29.82*2.62 (5 pcs) for 1 adapter Cleaning system for FU20 K1547PJ Hastelloy cleaning system (HCNF) K1547PG Hastelloy nozzle and mounting set (HCNF) K1547PH Nylon tube (10 metre) and tube mounting set for chemical cleaning system Buffer solutions K1520BA Starters Kit: (3x 500 ml) Solution ph 4.01, ph 6.87, ph 9.18 K1520BB Buffer Solution (500 ml) ph 1.68 K1520BC Buffer Solution (500 ml) ph 4.01 K1520BD Buffer Solution (500 ml) ph 6.87 K1520BE Buffer Solution (500 ml) ph 9.18 Connection equipment BA10 Junction box for ph extension cables WF10-XXX-F ph signal cable with terminated ends. Available lenghts (XXX = 005, 010, 025 & 050 m) WU10-V-S-XX Variopin cable (XX = 02, 05, 10, 15 and 20m) YOKOGAWA ELECTRIC CORPORATION World Headquarters 9-32, Nakacho 2-chome, Musashino-shi Tokyo 180-8750 Japan www.yokogawa.com YOKOGAWA CORPORATION OF AMERICA 2 Dart Road Newnan GA 30265 USA www.yokogawa.com/us YOKOGAWA EUROPE BV Euroweg 2 3825 HD AMERSFOORT The Netherlands www.yokogawa.com/eu YOKOGAWA ELECTRIC ASIA Pte. LTD. 5 Bedok South Road Singapore 469270 Singapore www.yokogawa.com/sg YOKOGAWA CHINA CO. LTD. 3F Tower D Cartelo Crocodile Building No.568 West Tianshan Road Changing District Shanghai, China www.yokogawa.com/cn YOKOGAWA MIDDLE EAST B.S.C.(c) P.O. Box 10070, Manama Building 577, Road 2516, Busaiteen 225 Muharraq, Bahrain www.yokogawa.com/bh Yokogawa has an extensive sales and distribution network. Please refer to the European website (www.yokogawa.com/eu) to contact your nearest representative. IM 12X0X0-E-E Subject to change without notice Copyright Subject to change without notice Copyright Printed in The Netherlands, 00-000 (A) I Printed in The Netherlands, 07-1004 (A) I