The Cryogenic Refrigeration System for MICE

Similar documents
The Liquefaction of Hydrogen and Helium Using Small Coolers

The Cost of Superconducting Magnets as a Function of Stored Energy and Design Magnetic Induction times the Field Volume

Peter C. Vander Arend Cryogenic Consultants, Inc. Allentown, Pennsylvania

Lecture 7 Thermal Insulation & Cryostat Basics. J. G. Weisend II

How does solar air conditioning work?

Lecture 7 Heat Transfer & Thermal Insulation. J. G. Weisend II

Controlling a 2-phase CO 2 loop using a 2-phase accumulator

THERMAL MANAGEMENT OF COMPUTER SYSTEMS USING ACTIVE COOLING OF PULSE TUBE REFRIGERATIORS

Refrigeration Basics 101. By: Eric Nelson

UNIT 2 REFRIGERATION CYCLE

AIR CONDITIONING TECHNOLOGY

C. starting positive displacement pumps with the discharge valve closed.

COMMERCIAL HVAC CHILLER EQUIPMENT. Air-Cooled Chillers

Refrigerant Changeover Guidelines R-22 to R-407C. Leading the Industry with Environmentally Responsible Refrigerant Solutions

UNDERSTANDING REFRIGERANT TABLES

The Secret of Hydraulic Schematics. BTPHydraulics

Commercial refrigeration has been in the environmental. Refrigerant. as a. Basics Considerations PART 1:

Characteristics of Evaporators

OPTIMAL DESIGN AND OPERATION OF HELIUM REFRIGERATION SYSTEMS *

Refrigerant Changeover Guidelines R-22 to R-407C. Leading the Industry with Environmentally Responsible Refrigerant Solutions

Practice Test. 4) The planet Earth loses heat mainly by A) conduction. B) convection. C) radiation. D) all of these Answer: C

Mechanical Seal Piping Plans

Any Service Technician Can Fix It A Good Service Technician Can Figure Out What s Wrong With It.

Why and How we Use Capacity Control

Automobile Air Conditioning Primer

How Ground/Water Source Heat Pumps Work

Heat. Investigating the function of the expansion valve of the heat pump. LD Physics Leaflets P Thermodynamic cycle Heat pump

Case Study: Innovative Energy Efficiency Approaches in NOAA s Environmental Security Computing Center in Fairmont, West Virginia

Topic Page Contents Page

Control Device Requirements Charts For Oil and Gas Handling and Production Facilities

Open Cycle Refrigeration System

How To Calculate The Performance Of A Refrigerator And Heat Pump

Chapter 7: Oil Vapor Diffusion Pumps

Evaluation Of Hybrid Air- Cooled Flash/Binary Power Cycle

SECTION 5 COMMERCIAL REFRIGERATION UNIT 22 CONDENSERS

Thermal Coupling Of Cooling and Heating Systems

Low GWP Replacements for R404A in Commercial Refrigeration Applications

TABLE OF CONTENT

Defining Quality. Building Comfort. Heating and Cooling Coils

Research on the Air Conditioning Water Heater System

SunMor SM-V30 Solar Water Heater Installation Manual

AN INSTRUMENT FOR GRAVIMETRIC CALIBRATION OF FLOW DEVICES WITH CORROSIVE GASES. J. O. Hylton C. J. Remenyik Oak Ridge National Laboratory *

Heating Water by Direct Steam Injection

ThermoStar Refrigeration Air Dryers

5.2. Vaporizers - Types and Usage

A/C refrigerant system, overview

MECHANICAL AND THERMAL ANALYSES OF THE CABLE/ STRAND STRAIN TEST FIXTURE

Refrigeration and Airconditioning Prof. M. Ramgopal Department of Mechanical Engineering Indian Institute of Technology, Kharagpur

CHAPTER 7 THE SECOND LAW OF THERMODYNAMICS. Blank

New Trends in the Field of Automobile Air Conditioning

BB-18 Black Body High Vacuum System Technical Description

Data Bulletin. Mounting Variable Frequency Drives in Electrical Enclosures Thermal Concerns OVERVIEW WHY VARIABLE FREQUENCY DRIVES THERMAL MANAGEMENT?

Large Systems Commissioning

Data Realty Colocation Data Center Ignition Park, South Bend, IN. Owner: Data Realty Engineer: ESD Architect: BSA LifeStructures

INFRARED MONITORING OF 110 GHz GYROTRON WINDOWS AT DIII D

Steam System Best Practices Condensate System Piping

COURSE TITLE : REFRIGERATION AND AIR CONDITIONING COURSE CODE : 4029 COURSECATEGORY : A PERIODS/WEEK : 5 PERIODS/SEMESTER : 90 CREDITS : 4 OBJECTIVES

Dry-out Design Considerations and Practices for Cryogenic Gas Plants

Liquid helium. Safetygram 22

Select the Right Relief Valve - Part 1 Saeid Rahimi

TEACHER BACKGROUND INFORMATION THERMAL ENERGY

Thermodynamics - Example Problems Problems and Solutions

Max. Length of HTS Cables in the Future

Session 2: Hot Water Supply

Lesson 36 Selection Of Air Conditioning Systems

Vapor Chambers. Figure 1: Example of vapor chamber. Benefits of Using Vapor Chambers

ENERGY CARRIERS AND CONVERSION SYSTEMS Vol. II - Liquid Hydrogen Storage - Takuji Hanada, Kunihiro Takahashi

Thermo Siphon Solar Systems

Laddomat Charging unit

High Pressure Ammonia Systems New Opportunities

Air Eliminators and Combination Air Eliminators Strainers

CHAPTER 9: WATER HEATING

Increasing the evaporation temperature with the help of an internal heat exchanger

Water Volume Calculation in Hydronic Heating & Cooling Systems

LNG Fuel Tank System LNG Tank overview

Water Fired Chiller/Chiller-Heater. WFC-S Series: 10, 20 and 30 RT Cooling

DIRECT STEAM INJECTION HOT WATER SYSTEMS FOR JACKETED HEATING

Contents and Nomenclature

Laddomat 21 Charging Unit

The D-Wave 2X Quantum Computer Technology Overview

National Competency Based Skill Training Refrigeration and Air-conditioning Mechanic Logbook

AIR CONDITIONING EFFICIENCY F8 Energy eco-efficiency opportunities in Queensland Foundries

CONTAMINANT REMOVAL FROM CENTRIFUGAL SYSTEMS

How To Clean Water From An Ammonia Refrigeration System

GTI Small-Scale Liquefaction Technology. March 2013

STEAM HEATING SYSTEM TROUBLESHOOTING GUIDE

RESULTS OF ICARUS 9 EXPERIMENTS RUN AT IMRA EUROPE

The paper addresses the boil-off in the cryogenic industry and details the specifics of it when applied to vehicle LNG tanks.

1932 Lexington Houston, Texas (713) Recommendations for installation of cooling systems

Draining and filling cooling system

LECTURE 28 to 29 ACCUMULATORS FREQUENTLY ASKED QUESTIONS

Chapter 18 Temperature, Heat, and the First Law of Thermodynamics. Problems: 8, 11, 13, 17, 21, 27, 29, 37, 39, 41, 47, 51, 57

Specific Volume of Liquid (Column 7). The volume per unit of mass in cubic feet per pound.

The Challenges of Low Temperature Research. Luke Mauritsen, Brian Smithgall Montana Instruments Corporation, USA

ENERGY SAVING WORT BOILING SISTEM IN BREWING FACTORY

Fully Pumped Systems

INSTALLATION INSTRUCTIONS HOT GAS BYPASS SYSTEM DESIGN MANUAL

ME Heat Transfer Laboratory. Experiment No. 7 ANALYSIS OF ENHANCED CONCENTRIC TUBE AND SHELL AND TUBE HEAT EXCHANGERS

Chapter 5. Components, Symbols, and Circuitry of Air-Conditioning Wiring Diagrams

Transcription:

The Cryogenic Refrigeration System for MICE Michael A. Green Lawrence Berkeley National Laboratory LBNL-51751 SCMAG-794 The proposed cryogenic cooling system for MICE uses a single liquid helium refrigerator to cool the superconducting magnets, the liquid hydrogen absorbers, and the scintillating fiber detectors. It is proposed that the helium refrigerator cold box deliver two-phase helium at 4.4 K at the saturated liquid condition and 14 K helium at a pressure of 14 to 18 bars. The 4.4 K helium will be used to cool the seven superconducting magnet assemblies in the MICE magnet string and the fiber light detectors. The 14 K helium will be used to cool the liquid hydrogen absorbers, the magnet shields, and the gas cooled electrical leads that go from about 50 K to room temperature. Helium returns to the refrigerator from the magnet string and detector box at 4.3 K. The helium gas from the absorbers returns to the refrigerator as a gas at 18 K or lower. The helium that will be used to cool the magnet shields and leads will by pass the refrigerator cold box and return to the refrigerator compressor at about 293 K. Forced Two-phase Helium Cooling for the Magnets It is proposed that the superconducting coils for the MICE magnets be cooled by conduction from the magnet coil support structure. A number of large detector magnets are cooled this way. The magnet support structure will be cooled by two-phase helium in cooling tubes that have an ID about 10 mm. It is proposed that the solenoid magnets be cooled in series using a single cooling circuit that goes from the refrigerator through the magnets and back to the refrigerator. (See Figure 1.) The volume of liquid helium in the solenoids will be very low (of the order of a few liters). The advantages of twophase helium cooling of the magnets are as follows: 1) The cool down of the magnet string is straightforward and well controlled. To first order, the magnet cool down rate is proportional to the magnet mass, proportional to the temperature difference between the flow circuit inlet and the outlet, and inversely proportional to the mass flow through the circuit. 2) The overall mass of a tubular cooled magnet and its cryostat is less than the mass of a magnet with helium bath cooling. 3) The amount of helium in contact with the magnet coil during a quench is limited to the amount of helium in the cooling tube. The safety hazard of liquid helium in tubes during a loss of vacuum or a magnet quench is small compared to a bath cooled cryostat. Tubes always have a higher pressure rating than a cryostat helium vessel. In general, tubes are not covered by the pressure vessel code; the helium vessel in a bath cryostat usually is covered by the pressure vessel code. 4) Helium services to the superconducting magnets can come into the cryostat from any direction. In other words, the helium services and leads do not have to come out of the top of the cryostat. The gas cooled electrical leads can be operated at any orientation including having the cold end up and the warm end down. -1-

Cold Box Helium Compressor 300 K, 1.1 bar He Return 14 K, 17 bar He Supply 4.3 K 1.1 bar He Return 18 K, 1.1 bar He Return 4.3 K 1.2 bar He Supply 500 l Control Dewar D1 F1 A1 Fiber Detector Cooling Coupling Magnet H2 Absorber Focusing Magnet C1 F2 C2 F3 A2 A3 Detector Magnet D2 Figure 1. A Simplified Schematic Diagram of the Helium Cryogenic Cooling System for MICE. Note: The boxes labeled D1 and D2 are detector magnets; the boxes labeled F1, F2 and F3 are focusing magnets; and the boxes labeled C1 and C2 are the coupling solenoids. The absorbers A1, A2 and A3 are shown attached to the focusing magnets. It is assumed that a focusing magnet is in the same cryostat vacuum vessel as the absorber. The physical connection between the absorber and the focusing coil pair is assumed to be through a low thermal conductivity support that limits heat transfer from 18 K to 4 K. -2-

Optional Nitrogen Circuit LN 2 N Gas 2 Helium Cold Box E 1.1 Atm. 300 K He 18 Atm. He Helium Compressor System Helium Recovery and Supply Expansion Engines The number of engines may be greater than shown. E 300 K He from Shields and Leads 14 K He to Absorbers and Shields Cool Down Bypass 18 K He from Absorbers J-T Valve 2-phase He Supply Two-phase He Flow to and from Magnets 500 liter Control Dewar Heat Exchanger Figure 2. A Schematic Representation of the Refrigerator Cold, the Refrigerator Compressors and the Two-phase Cooling Control Dewar. The control dewar and its heat exchanger are key elements for successful two-phase cooling of one or more superconducting magnets. Most of the liquid helium in the cooling circuit is in the control dewar. Note: the refrigerator cold box is shown with two expansion engines. There could me more engines, but typical 100 to 200 W refrigerators have two engines as shown. The J-T valve could be replaced by a cold expander, which will increase the cooling of the refrigerator. The diagram above shows the cool down bypasses for the refrigerator, magnets, and absorbers. -3-

Cooling for the Magnets and the Liquid Hydrogen Absorbers Figure 1 shows how the superconducting magnets and the absorbers are cooled. The superconducting magnets are shown as being cooled in series. The hydrogen absorbers are cooled in parallel. The cooling for the magnet shields and leads off of the 14 K circuit is shown as being in parallel with the absorber cooling. Figure 1 shows the primary control valves for the cooling system. (They are shown as darkened double triangles.) The control valve between the refrigerator and the control dewar is the J-T valve that produces the two-phase helium needed to cool the magnet string. The control valves that go into each of the absorbers will control the helium flow depending on the temperature of the helium stream leaving the absorber. It should be noted that the temperature of the helium entering the control valve is shown as 14 K. This temperature may have to be higher than 14 K because expansion through the control valve will cause cooling of the helium entering the absorber. The helium gas entering the absorber must be above 13 K, which is the freezing temperature for liquid hydrogen in the absorber. The Control Dewar for the Two-phase Helium Distribution System Figure 2 shows the refrigerator cold box, the compressors and the liquid helium control dewar. The use of a control dewar with a heat exchanger (as shown in Figure 2) is essential for reliable operation of a two-phase helium cooling circuit. The role of the control dewar and its heat exchanger is to sub-cool the helium entering the magnet string. The advantages of this approach are: 1) The liquid in the control dewar is colder than the two-phase helium mixture coming off of the J-T valve. The heat flow from the two-phase helium to the helium bath in the dewar condenses the helium in the cooling tube and drops its temperature below the saturation temperature for the helium in the heat exchanger tube. The average density of the helium in the tube increases over a factor of three and as a result, the pressure drop in the flow circuit from magnet to magnet is reduced over a factor of two. 2) The control dewar and its heat exchanger has the effect of damping out the pressure oscillations that often occur in two phase flow circuits. This feature extend even to the so called garden oscillation caused by a twophase mixture going up and down through many turns of a tube. Simply put, two-phase flow circuits that do not work without a control dewar and its heat exchanger will usually work when a control dewar is used. In a well-designed two-phase flow cooling system, the control dewar eliminates or greatly reduces pressure oscillations often seen in flow circuits. 3) The control dewar acts as a buffer vessel that can provided additional cooling at times when the heat load exceeds the capacity of the available refrigeration. Up to 1.5 times the rated capacity of the refrigeration capacity can be delivered to the load as long as there is liquid helium in the control dewar. If one does not allow quench gas to vent back through the control dewar, one can use the liquid helium in the control dewar to provide added cooling for the magnets when they are being re-cooled after a quench. It is proposed that the control dewar be a standard wide neck 500-liter dewar built by a commercial vendor. The neck inside diameter should be at least 100 mm. A 100-mm diameter neck will allow one to install a heat exchanger fabricated from 12.7-mm OD tube. When installed, the heat exchanger would -4-

sit in the lower one third of the dewar liquid space. The use of a standard 500-liter storage dewar means that the control dewar can be relatively inexpensive and it can be certified as being safe. Figure 2 also shows the by-pass that is used when the magnets and the absorbers are being cooled down from room temperature. The by-pass that is shown completely bypasses the cold box. An alternative to this is a by-pass is one that goes into the cold box and allows the return gas to enter the return side of the refrigerator heat exchanger at various points as the return gas is cooled to allow it to enter the refrigerator at that point. The use of by-pass as shown in Figure 2 is a simple way of cooling down any cryogenic system from room temperature to its operating temperature using a refrigerator. 0.03 to 0.12 g/s 14 K He from Cold Box 4 g/s 4.3 K 2-phase He from Previous Magnet 4 g/s 4.3 K 2-phase He to Next Magnet Lead Gas Flow Control F In line Insualtor 0.03 to 0.12 g/s 300 K He Gas to Compressor Suction F Gas Cooled Cu leads T T HTS Leads S/C Magnet at 4.4 K Cold Mass Support Intercept 24 to 40 K Shield 300 K Cryostat Vacuum Wall Figure 3. A Schematic Representation of the Flow of Two-phase Helium and 14 K helium in a Coupling Magnet or Detector Magnet Cryostat. Note: the lead current for the coupling magnet is assumed to be 1000 A. The lead current for the detector solenoids is 200 to 250 A. It should be noted that the detector solenoid has at least four pairs of leads. Figure 3 shows a schematic representation of the cooling of a coupling magnet or a detector magnet. The magnet coil is cooled using two-phase helium in a pipe attached to the coil support structure. Before the two-phase helium leaves the magnet cryostat, the lower part of the high temperature superconductor leads (HTS) must cooled by the two-phase helium in order to intercept any -5-

heat coming down the HTS leads. The use of HTS leads reduces the heat flow into the 4.4 K region. If conventional gas cooled leads were used, helium needed for lead cooling would be drawn off the twophase cooling circuit. In the coupling magnet, 0.1 to 0.12-g s -1 would be drawn off and returned to the compressor at room temperature. As a result, the equivalent 4.4 K refrigeration needed for cooling a single coupling magnet must be increased by about 12 W. The use of HTS leads on the detector magnets will save the equivalent of about 12 W of refrigeration at 4.4 K for each magnet. 5 g/s 14 K He from Cold Box Control Valve H2 Rupture Disc 5 g/s 18 K He to Cold Box to H2 Surge Tank H2 Gas Fill 3 g/s 4.4 K 2-phase He from Previous Magnet 3 g/s 4.2 K 2-phase He to Next Magnet Lead Gas Flow Control 0.03 g/s 300 K He Gas to Compressor Suction F In line Insualtor F Vent Valve T Gas Cooled Cu leads T T HTS Leads LH2 Absorber Gravity S/C Magnet at 4.4 K Cold Mass Support Intercept 300 K Cryostat Vacuum Wall 19 K LH2 Cooled by Natural Convection 24 to 40 K Shield Figure 4. A Schematic Representation of the Cooling of the Focusing Magnet and the Hydrogen Absorber The helium circulating in the walls of the absorber is shown as a heat exchanger that has the helium going up. The hydrogen circulates so that it passes down the absorber walls as it is cooled. The heated hydrogen rises up the center region of the absorber while the cooled hydrogen goes down along the absorber wall. The hydrogen enters the absorber as a gas. It is condensed on the absorber wall to form liquid hydrogen. There should be no need for a liquid fill of the absorber. When the hydrogen is returned to the surge thank, it can return through the same line that is used to fill the absorber. -6-

Cooling of the Liquid Hydrogen Absorber Figure 4 shows a schematic view of how the liquid hydrogen absorber and the focusing solenoid can be cooled from the MICE cryogenic system. It is assumed that the hydrogen in the absorber is cooled by free convection to the walls of the absorber (a separate heat exchanger may be better) that is cooled by helium from the refrigerator that enters the absorber at 14 K. Hydrogen that is in the absorber moves upward because it is less dense. The hydrogen moving upward is replaced by cooler more dense hydrogen that is moving down the walls of the absorber body (or a separate heat exchanger). The absorber body walls forms a counter-flow heat exchanger where heat is transferred from the helium flowing in tubes attached to the wall to the liquid hydrogen flowing down the walls of the absorber. The helium that enters the absorber at 14 K must be split into two streams that flow from the bottom of the absorber to the top of the absorber. If the helium enters the absorber at 14 K and leaves the absorber at 18 K about 0.05-g s -1 of helium flow is needed for each watt of heat removed from the liquid hydrogen. The required helium mass flow in inversely proportional to the temperature difference between the helium inlet and out let. For an absorber with 80 W of heating and helium gas entering at 14 K and exiting at 18 K, the helium flow through the absorber body is about 4 g s -1. Since the liquid hydrogen in the absorber must always be kept sub-cooled, in order to avoid boiling, the maximum allowable absorber temperature is 20 to 22 K depending on maximum allowable pressure on the absorber windows. The temperature difference that drives the free convection and the heat transfer from the liquid hydrogen to the helium from the refrigerator can be no more than 8 K. In reality, this temperature difference is less than half of that value. The design of the free convection flow circuit and the heat exchanger that exchanges the heat from the hydrogen in the absorber to the helium flowing through the absorber is very important. This will be the topic of a different design note. The cooling circuit shown in Figure 4 reflects the need for counter flow heat exchanger from the hydrogen in the absorber to the helium flowing through tubes in the body of the absorber. Figure 4 shows that the helium must enter the absorber body from the bottom; it must exit the absorber body from the top. The hydrogen that is in the absorber is just liquid in a flask. Liquid hydrogen does not have to be transferred to the absorber. The liquid can be condensed from hydrogen gas that enters the absorber through a fill tube. The hydrogen can be drained from absorber rapidly by opening a valve that allows the liquid hydrogen in the absorber to expand into an evacuated tank. This type of safety system has been used for hydrogen bubble chambers with large volumes of liquid hydrogen. A hydrogen absorber with 380-mm diameter windows would have a total liquid hydrogen volume of about 38 liters. The volume of the evacuated tank needed for the hydrogen from three absorbers would be about 100 m 3. The heat leak from the absorber body to the focusing solenoid depends on how the absorber body is supported. If one supports the absorber from room temperature the heat leak to the magnet is small. If one supports the absorber body directly to the magnet coil support structure, the heat leak will be small even when on considers the possible de-centering forces that might occur on the absorber during a focusing magnet quench. From 22 K to 4 K, the thermal conductivity integral of G-10 is about 2 W m -1. Large magnetic forces (say 100 tons) can be carried with heat rates from the absorber body to the magnet support structure that are less than 0.2 W to the 4 K region. -7-

If liquid helium were used in the absorber instead of liquid hydrogen, the absorber body would be cooled using two phase helium at 4.4 K. The absorber would be filled with helium at 2 bar (just below the critical pressure for helium). The driving temperature difference for natural convection cooling would be about 0.6 K instead of the 5 to 6 K for the absorber when it is filled with hydrogen. The maximum allowable heat flow into the absorber fluid drops from 150 W to about 15 W. The heat conducted into the absorber body from the outer windows and the connection to room temperature will be the same as it is for the hydrogen absorber (about 33 W per absorber). This heat must be transferred to the two-phase helium directly without passing into the absorber fluid. The maximum cooling to three absorbers that operate with liquid helium instead of liquid hydrogen is about 140 W at 4.4K. This heat load is about half of the heat flow into the 14 K helium used to cool the hydrogen absorber. Cooling the Scintillating Fiber Detectors The detectors that detect the photons from the scintillating fibers that are in the NICE detectors operate at a temperature between 5.0 and 5.9 K. The D0 experiment at Fermilab used similar detectors, which are cooled by a liquid helium bath. A liquid nitrogen cooling stage intercepts heat that comes down the fibers. The maximum number of channels for MICE is about half the number of channels in the D0 experiment. The D0 experiment at Fermilab uses 200 to 300 W of cooling at 4.4 K plus about 2 kw of cooling at liquid nitrogen temperature. The maximum heat load to the MICE detector cooling is about 150 W at 4.4 K plus 1000 W at 80 K. The actual heat load to the MICE scintillating fiber detectors could be much lower than to 150 W quoted in the previous paragraph. At least two factors will influence the heat load. The first is the number of scintillating fiber channels needed to overcome the background noise due to the dark currents and x-rays from the RF cavities. The number of channels entering the detector is also influenced by how these channels are multiplexed before entering the scintillating fiber light detector. There is at least one scenario that suggests that the number of fiber channel that need to be cooled might be one or two percent of the number of channels entering the D0 scintillating fiber light detector. In this scenario, one or two cryocoolers could produce cooling at 5.5 K and 60 K to cool the detectors and intercept heat coming down the fibers. If a TPG is used as a MICE detector, there should be no need for cryogenic refrigeration. Calculated Heat Loads and the Size of the Helium Refrigerator Table 1 shows the worst case projected heat loads at 4.4 K, the heat loads at 14 to 18 K and the 14 K helium mass flow needed to cool the magnet shields and the upper gas-cooled electrical leads. The table also includes the liquid nitrogen flow needed to intercept heat flowing down the scintillating fibers into the scintillating fiber light detector. From the calculated heat loads, one can determine the equivalent refrigeration at 4.4 K needed to cool MICE. This permits one to determine the size of the refrigerator needed for MICE and the input power and water cooling that is needed to provide the refrigeration for MICE. -8-

Table 1. Heat Loads and Refrigeration Requirements for MICE Parameter Heat Load Heat Load Lead Flow at 4.4 K(W) at 14 K(W) at 14 K (g s -1 ) Seven Super conducting Magnets Two Coupler Magnets 3.2 30.3 0.20 Three Focusing Magnets 5.2 21.9 0.18 Two Detector Magnets 2.8 13.8 0.24 Liquid Helium Transfer Line (includes. Bayonets) 27.4 --- --- Total Heat Load and Mass Flow 38.6 66.6 0.66 Three Liquid Hydrogen Absorbers Radiation Load to Windows 39 Conduction Heat Load to the Absorber Body 60 Beam and Dark Current Heating Allowance 60 Convection Heaters 90 Helium Transfer Lines (includes Bayonets) 41 Heat Load to the 14 K Helium Gas 290 Scintillating Fiber Detectors at 4.4 K 150.0 Liquid Nitrogen Cooling for the Scintillating Fiber Detectors = 1000 W @ 80 K = ~20 liter / hr of LN 2. Equivalent Refrigeration at 4.4 K and Refrigerator Input Power Equivalent 4.4 K Refrigeration Needed Design Size of the Equivalent 4.4 K Refrigerator Total Installed Input Power for the Compressors 307 W ~420 W ~295 kw The calculated refrigeration load for MICE is about 190 W at 4.4 K (including the heat load to the scintillating fiber light detectors), plus 290 W from 14 to 18 K plus 0.66 grams per second of helium gas taken from the cold box at 14 K and returned to the compressor intake at 293 K. This refrigeration is equivalent to about 307 W at 4.4 K. In order to allow for errors and additional margin, a 420 W 4.4 K helium refrigerator should be installed. This means that the installed power needed to power the MICE refrigeration system should be about 300 kw. If liquid helium is used in the absorbers instead of liquid hydrogen, the required helium refrigeration plant capacity will go up from 307 W to about 360 W. One may want to increase the size of the cold box from 420 W at 4.4 K to about 450 W at 4.4 K to allow one to operate the MICE absorbers with liquid helium in place of liquid hydrogen. Since most of the power that goes into the compressor is extracted as heat, the required cooling capacity for the compressors will equal the installed power for the refrigeration system. -9-

Cost Estimate for the MICE Refrigeration The MICE refrigerator is sized between 420 and 450 W at 4.5 K. The costs for a number of different refrigerators is given in Figure 5 below [1]. Using the cost equation given in Figure 5, one comes up with an estimated MICE refrigerator cost of about 1.1-M$. This cost includes, the refrigerator cold box, the compressor station, the helium recovery system, and the refrigerator control system. An addition 450 k$ must be spent to provide the transfer lines and control cryostat for the experiment. 100 Refrigerator Capital Cost (M$) 10 1.1 Cost (M$) = 2.0*Q(kW)^^0.7 Built before 1975 Built after 1975.01.0001.001.01.1 1 Refrigeration Capacity at 4.5 K (kw) 10 100 The Cost of a Variety of 4.5 K Helium Refrigerators Given in 1997 dollars as a Function of Refrigerator Size. To covert to 2002 dollars, multiply the cost obtained from the chart by 1.18. Acknowledgements This work was supported by the Office of Science, United States Department of Energy under DOE contract number DE-AC03-76SF00098 References [1] M. A. Green and R Burns, An Update on Estimating the Cost of Cryogenic Refrigeration, Advances in Cryogenic Engineering 43, p 1661, Plenum Press, (1997) -10-

DISCLAIMER This document was prepared as an account of work sponsored by the United States Government. While this document is believed to contain correct information, neither the United States Government nor any agency thereof, nor The Regents of the University of California, nor any of their employees, makes any warranty, express or implied, or assumes any legal responsibility for the accuracy, completeness, or usefulness of any information, apparatus, product, or process disclosed, or represents that its use would not infringe privately owned rights. Reference herein to any specific commercial product, process, or service by its trade name, trademark, manufacturer, or otherwise, does not necessarily constitute or imply its endorsement, recommendation, or favoring by the United States Government or any agency thereof, or The Regents of the University of California. The views and opinions of authors expressed herein do not necessarily state or reflect those of the United States Government or any agency thereof, or The Regents of the University of California. Ernest Orlando Lawrence Berkeley National Laboratory is an equal opportunity employer. -11-