Metrological traceability a concept for standardization in laboratory medicine

Similar documents
CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES

Overview of ISO for 'Biologicals'

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

OIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

The BIPM key comparison database

International Consortium for Harmonization of Clinical Laboratory Results. Operating Procedures

American Association for Laboratory Accreditation

ISO and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

ISO/IEC ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES

CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company

Case study Realistic and best uncertainties with humidity instruments

RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES

There is a need for a coordinating role by the Commission supported by the Framework programme, to facilitate and harmonise those developments.

Graham Beastall. Outline of Talk

Validation of measurement procedures

Regulatory Requirements for Medical Device Calibration Programs

Procedures and General Requirements

GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION

Terminology in Analytical Measurement Introduction to VIM 3

Standardization of hemoglobin A 1c : myth or reality? Andrea Mosca

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

American Association for Laboratory Accreditation

Schedule of Accreditation issued by United Kingdom Accreditation Service High Street, Feltham, Middlesex, TW13 4UN, UK

Experis Gas Mixtures. When accuracy, stability and reliability are a must

Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1

accreditation will it still be needed or will other schemes show up?

ROLE OF THE APEC SPECIALIST REGIONAL BODIES

QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES

RECOMMENDATION. on the Evaluation of Quality Management System of National Metrology Institutes

The age of competence: an update on the international laboratory accreditation scene for veterinary testing laboratories

Accreditation according to ISO/IEC in Europe. Historic EA Europeing Accreditation Assessment Experience from Technical assessment

Validation and Calibration. Definitions and Terminology

How do I gain confidence in an Inspection Body? Do they need ISO 9001 certification or ISO/IEC accreditation?

EUROLAB 25 years. The role of metrology and testing to characterize materials and products. EUROLAB 25th Anniversary Seminar

Form and content of COOMET сertificate for reference materials for composition and properties of substances and materials

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

Quantifying Uncertainty in Analytical Measurement

International Laboratory Accreditation Cooperation. Laboratory Accreditation or ISO 9001 Certification? global trust. Testing Calibration Inspection

Procedure for Equipment Calibration and Maintenance

European cooperation for EAL-G23 THE EXPRESSION OF UNCERTAINTY IN QUANTITATIVE TESTING. The Expression of Uncertainty in Quantitative Testing

Guidelines on the Reporting of Compliance with Specification

The Calibration Chain: Role of BIPM,

EffiValidation 3.0 software basic training module

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

ISO/IEC ACCREDITATION REQUIREMENTS FOR FORENSIC INSPECTION BODIES

NABL accreditation is a formal recognition of the technical competence of a medical testing laboratory

Certifying Reference Materials to ISO Guide 34

Content Sheet 10-1: Overview of External Quality Assessment (EQA)

Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009

Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories

Adding Value to Laboratory Medicine

List of EA Publications. And International. Documents

American Association for Laboratory Accreditation

PROCESS FOR REVIEW AND APPROVAL OF NOMINATED CERTIFIED REFERENCE MATERIALS AND REFERENCE MEASUREMENT METHODS/PROCEDURES JCTLM WG1-P-03

American Association for Laboratory Accreditation

SAMPLE. Verification of Comparability of Patient Results Within One Health Care System; Approved Guideline (Interim Revision)

IAEA-TECDOC Development and use of reference materials and quality control materials

MANAGEMENT REVIEW FOR LABORATORIES AND INSPECTION BODIES

TECHNICAL NOTE: SI 01 SPECIFIC REQUIREMENTS FOR THE ACCREDITATION OF INSPECTION BODIES FOR SITE INVESTIGATION

NABL NATIONAL ACCREDITATION

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

Maria Belli. (former APAT) IX. Congress of clinical chemistry & laboratory medicine Prague September 2009

Torque Key Comparison CCM.T-K1.3

(Uncertainty) 2. How uncertain is your uncertainty budget?

State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist

QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES. October 2008 Revision 08

The IFCC Global Campaign of Diabetes Mellitus. IFCC General Conference, Antalya, April 13-15th 2008 Sverre Sandberg, Howard Morris, Matthew McQueen

Calibration of Electronic Test Equipment: An Introduction to ISO & Accreditation

ISO How to Meet. Requirements for Method Verification. Prepared by:

Mise en pratique of the definition of the kilogram

THE INTERNATIONAL HARMONIZED PROTOCOL FOR THE PROFICIENCY TESTING OF ANALYTICAL CHEMISTRY LABORATORIES

Harmonisatie. Cas Weykamp SKML Sectie Humorale Immunologie Congres De Waarde van de Expert 9 juni 2015

Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland

R103 - GENERAL REQUIREMENTS: PROFICIENCY TESTING FOR ISO/IEC LABORATORIES

Analytical Laboratories, Inc.

Selection, Use and Interpretation of Proficiency Testing (PT) Schemes

Definition of the Measurand: CRP

American Association for Laboratory Accreditation

Non-Controlled Copy. Quality Manual Revision Log. Rev # Date Description

How To Test A Titrette And Digital Burette

P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014

JCGM 106:2012. Evaluation of measurement data The role of measurement uncertainty in conformity assessment

High Level Importance: ISO Microbiology compliance

Accreditation and Competence?

The National Quality Infrastructure

Gravimetric determination of pipette errors

GUIDANCE ILAC-G24 INTERNATIONAL OIML D 10 SERIES Edition 2007 (E) DOCUMENT Edition 2007 (E)

Info 15:2 TRAINING 2015/2016. Info 13:19

Stimuli to the Revision Process

STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION

POCT in diagnosing and monitoring of Diabetes Mellitus. 13th EFLM Continuous Postgraduate Course, Sverre Sandberg, Noklus / EFLM

EU Regulation on in-vitro Diagnostic Medical Devices call for urgent action.

LAB 37 Edition 3 June 2013

The IFCC HbA 2 standardization

MEDICAL DEVICE GUIDANCE

Recognition of my Product (Testing Laboratory) Why I need recognition, what is the effect?

The Laboratory Quality Assurance System. A Manual of Quality Procedures and Forms. 3rd Edition

Transcription:

DOI 10.1515/cclm-2012-0710 Clin Chem Lab Med 2013; 51(5): 953 957 Opinion Paper Lothar Siekmanna, * Metrological traceability a concept for standardization in laboratory medicine Abstract: The concept of measurement traceability provides the most important strategy in achieving standardization in laboratory medicine aimed at equivalent measurement results regardless of the principle of measurement, the method, the actual (test kit) and the laboratory where analyses are carried out. Keywords: measurement traceability; reference materials; reference s; reference measurement services; standardization. a Previous chair of the IFCC Committee on Traceability in Laboratory Medicine (C-TLM). *Corresponding author: Prof. Dr. Lothar Siekmann, Institute for Clinical Chemistry and Clinical Pharmacology, Sigmund Freud Strasse 25, 53127 Bonn, Germany, E-mail: Lothar.Siekmann@ukb.uni-bonn.de Hierarchical scheme of measurement traceability Measurement traceability (defined as property of a measurement result whereby the result can be related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty according to VIM 2008) [ 1 ] forms the basis for standardization in laboratory medicine which is described in detail in ISO 17511 [ 2 ]. The strategy of traceability in laboratory medicine is described in detail in ISO 17511 [ 2 ]. As demonstrated in Figure 1, a result of measurement in a patient sample should be traceable along a cascade of mea surement procedures and calibrators/reference materials of increasing metrological order up to the highest level represented by the definition of the measurand in SI units. This hierarchical scheme represents a sequence of s and materials of increasing metrological order from the routine sample obtained from a patient applying the end-user s routine (test kit) to the highest level: the definition of the measurand in SI units. The increasing uncertainty μ c (y) from the top level down to the result in a patient sample is symbolized by the horizontal bars. The responsibilities for the procedures and materials at different levels lie, on the top, with the International Bureau for Weights and Measures (BIPM), which serves as a functional unit for the General Conference on Weights and Measures (CGPM); this is followed by the National Metrology Institutes (NMI), the Accredited Reference (Calibration) Laboratories (AR), the Manufacturer s Laboratories () down to the basic level of the enduser s routine (testing) laboratories. In many situations, it is acceptable to skip pairs of materials and procedures. However, it is important to maintain the highest levels in the scheme, e.g., the primary reference of outstanding specificity and trueness to assign values to secondary matrix reference materials. The hierarchical scheme for the measurement of catalytic concentrations of enzymes shows some differences as, in this case, the top level cannot be described by a primary calibrator and its certification. Instead, the enzyme measurand has to be defined by an exactly described conventional reference. This concept is described in ISO 18153 [ 3 ] and the respective primary reference s have been published by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) for a series of enzymes [ 4 ]. Requirements for establishing the concept of traceability The implementation of the traceability concept obviously requires the establishment of internationally accepted: reference s; reference materials; and reference measurement laboratory services.

954 Siekmann: Metrological traceability in laboratory medicine SI-Unit Definition of the measurand Certification of a reference material (purity) μ c (y) Uncertainty BIPM, CGPM NMI ISO 17511 (modified) Primary calibrator NMI, AR Secondary calibrator Primary reference NMI, AR AR, BIPM: International Bureau of Weights and Measures Traceability Manufacturer s working (master) calibrator Mf. s product calibrator Secondary reference Manufacturer s standing End-user s routine, AR Manufacturer and/or end-user CGPM: General Conference on Weights and Measures NMI: National Metrology Institute AR: Accredited Reference Measurement Laboratory Routine sample End-user : Manufacturer s Laboratory Result End-user Figure 1 Hierarchical scheme of traceability in laboratory medicine (modified according to ISO 17511). With respect to these requirements, the Joint Committee for Traceability in Laboratory Medicine (JCTLM) [ 5 ], established by the BIPM, the IFCC and the International Laboratory Accreditation Cooperation (ILAC), has launched two projects in its working groups (WG). WG-1 lists tables of reference materials and reference procedures on the JCTLM website [ 5 ]. The materials and procedures are first inspected by review teams dedicated to different groups of measurands on the basis of criteria defined in the International Standards ISO 15193 [ 6 ] and 15194 [ 7 ]. The criteria for listing reference measurement materials and s are described in detail in the respective procedure manuals on the JCTLM website [5 ]. The list of materials and methods is updated annually according to the decisions made by the JCTLM Executive Committee. The identification of reference measurement laboratories is the objective of JCTLM WG-2. Establishing networks of reference laboratories is in addition to reference s and reference materials one of the biggest challenges in implementing the concept of measurement traceability. Each individual proposal of a reference laboratory for a particular measurement service (measurand) is first inspected by a review team dedicated to the appropriate group of measurands and then approved by the JCTLM Executive Committee according to the rules of the corresponding WG-2 procedure manual [ 5 ]. There is general agreement now that reference laboratories should be identified: according to the metrological level of the procedures applied; on the basis of accreditation as calibration laboratory according to ISO 15195 [ 8 ] and ISO 17025 [ 9 ]; and on the basis of their ability to demonstrate performance in regular inter-laboratory comparisons (ring trials). With respect to the metrological level the measurement principle [e.g., isotope dilution-mass spectrometry (IDMS)] is probably one of the most important criteria to accept a method as reference in view of its outstanding specificity and trueness. In addition, the use of a primary reference material as calibrator is essential. Furthermore, all gravimetric and volumetric steps must be carefully calibrated gravimetrically using certified test weights. Each certified measurement result achieved after a campaign of at least three independent measurement occasions must be accompanied by an individual uncertainty statement. Also, extensive quality assessment of each analytical campaign is obligatory. Matrix control materials with certified target values must be included in each series of measurements.

Siekmann: Metrological traceability in laboratory medicine 955 Laboratories must show their accreditation as a reference (calibration) laboratory to demonstrate adherence to the requirements of ISO standards 15195 and 17025. In order to support the JCTLM activities regarding inter-laboratory comparisons, the IFCC has launched a ring trial program for reference laboratories (RELA) particularly dedicated to reference measurement laboratories for some 30 different measurands [ 10 ]. Ring trial results not only demonstrate the competence of individual laboratories, they also reveal the equivalence of individual laboratory results or, in some cases, the bias of different reference procedures. In this special issue of Clinical Chemistry and Laboratory Medicine, the article External Quality Assessment Scheme for reference laboratories review of eight years experience [ 11 ] describes the details. Hierarchical scheme of laboratories In summary, a hierarchical scheme of laboratories has been established as demonstrated in Figure 2. At the top, we find the NMIs; they must demonstrate their competence in key comparisons and are listed in the CIPM-MRA-Key-Comparison-Data-Base (International Committee on Weights and Measures Mutual Recognition Arrangement Key Comparison Data Base). Their services include certification of primary and secondary Outline of the Calibration and Measurement Hierarchy in Laboratory Medicine CIPM MRA Key Comparisons Data Base (KCDB) CIPM National Metrology Institutes Listed: CIPM KCDB list (Appendix C) of National Metrology Institutes providing Calibration and Measurement Capabilities (CMCs) Realizations of the SI units providing: a) Certified values for reference materials, b) Calibration and measurement capabilities on the highest metrological level for comparative measurements e.g. in Ring Trials for Reference Laboratories Reference (Calibration) Laboratories JCTLM a) Accreditation as Calibration Laboratories according to ISO 17025 and 15195, b) Use of JCTLM listed reference methods, c) Participation in Ring Trials for Reference Laboratories Listed: JCTLM list of reference laboratories Reference measurement service for a) Diagnostica manufacturers, b) Regulatory bodies c) Proficiency testing organisations, certifying reference method values for ring trials for routine laboratories Routine (Testing) Laboratories National Regulatory Bodies a) Accreditation according to ISO 15189, b) Participation in ring trials of proficiency testing organisations, Reference method certified by reference laboratories Measurement service for patient samples Figure 2 Hierarchical scheme of laboratories for calibration and measurement in laboratory medicine.

956 Siekmann: Metrological traceability in laboratory medicine reference materials and provision of calibration and measurement capabilities on the highest metrological level, e.g., by participating in Ring Trials for Reference (Calibration) Laboratories. Thereby, they provide the link to the second level of laboratories in the hierarchical order. The Reference (Calibration) Laboratories have to demonstrate their competence: by accreditation as calibration laboratories according to ISO 15195 [8 ]; by using JCTLM accepted reference measurement procedures; and by regular participation in ring trials dedicated to reference/calibration laboratories. Once the requirements for particular services (measurands) are fulfilled, these laboratories will be listed by the JCTLM on the respective website. The service of the reference/calibration laboratories is to support diagnostic kit manufacturers and regulatory bodies in the establishment and control of traceability of diagnostic test kits. Furthermore, they provide certified reference method target values for External Quality Assessment Schemes (EQAS), thus linking to the basic group of routine/testing laboratories. These may demonstrate their competence by accreditation according to ISO 15189 [ 12 ] and by participation in ring trials where target values are certified by reference/ calibration laboratories. Obviously, their service is the analysis of samples from patients. Implementation of the concept of measurement traceability The implementation of the concept of traceability improves standardization in laboratory medicine as evident from the results of EQAS for routine laboratories using reference method target values. The concept of traceability as described here has to date become available for only a limited number of measurands for which certified primary reference materials and/or reference s are available. Up to now, the JCTLM has listed 241 reference materials, 156 reference s and 63 reference measurement services from eight reference laboratories. Extension of the concept of traceability to include further measurands by developing new reference materials and reference s and establishing reference measurement services is scheduled for the future. Nevertheless, it can be estimated that in the current situation, about 65% of all measurements in a typical clinical chemical laboratory can be covered by the concept of traceability. For measurements where the concept is not applicable another approach has been proposed based on harmonization. This has been discussed extensively in recent years [ 13 ]. However, it must be emphasized that harmonization should not be applied when the metrological concept of traceability can be implemented. Conflict of interest statement Author s conflict of interest disclosure: The author stated that there is no conflict of interest regarding the publication of this article. Research funding: None declared. Employment or leadership: None declared. Honorarium: None declared. Received June 27, 2012; accepted October 9, 2012; previously published online November 12, 2012 References 1. International vocabulary of metrology Basic and general concepts and associated terms (VIM). 3rd ed. 2008 version with minor corrections. JCGM 200, 2012. 2. In vitro diagnostic medical devices Measurement of quantities in biological samples Metrological traceability of values assigned to calibrators and control materials, ISO 17511, 2003. 3. In vitro diagnostic medical devices Measurement of quantities in biological samples Metrological traceability of values for catalytic concentration of enzymes assigned calibrators and control materials, ISO 18153, 2003. 4. Siekmann L, Bonora R, Burtis CA, Ceriotti F, Clerc-Renaud P, Ferard G, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 C, Part 1. The concept of reference procedures for the measurement of catalytic concentrations of enzymes. Clin Chem Lab Med 2002;40:631 4. 5. Joint Committee for Traceability in Laboratory Medicine (JCTLM). Available from: http://www.bipm.org./jctlm/. Accessed 15 October 2012. 6. In vitro diagnostic medical devices Measurement of quantities in samples of biological origin Requirements for content and presentation of reference s, ISO 15193, 2009.

Siekmann: Metrological traceability in laboratory medicine 957 7. In vitro diagnostic medical devices Measurement of quantities in samples of biological origin Requirements for certified reference materials and the content of supporting documentation, ISO 15194, 2009. 8. Laboratory medicine Requirements for reference measurement laboratories, ISO 15195, 2003. 9. General requirements for the competence of testing and calibration laboratories, ISO 17025, 2005. 10. RELA surveys. Available from: http://www.dgkl-rfb.de:81/index. shtml. Accessed 15 October 2012. 11. Kessler A, Siekmann L, Weykamp C, Geilenkeuser WJ, Dreazen O, Middle J, et al. External Quality Assessment Scheme for reference laboratories review of eight years experience. Clin Chem Lab Med 2013;51:997 1005. 12. Medical laboratories Particular requirements for quality and competence, ISO 15189, 2007. 13. Miller WG, Myers GL, Gantzer, Kahn SE, Sch ö nbrunner ER, Thienpont LM, et al. Roadmap for harmonization of clinical laboratory s. Clin Chem 2011;57: 1108 17.