Standardization of hemoglobin A 1c : myth or reality? Andrea Mosca
|
|
|
- Stella Cooper
- 10 years ago
- Views:
Transcription
1 Standardization of hemoglobin A 1c : myth or reality? Andrea Mosca Dept. Pathophysiology and Transplantation Centre for Traceability in Laboratory Medicine (CIRME) University of Milano, Milano (IT) contents IFCC approach to the standardization of HbA 1c - IFCC Network - Manufacturers -Customers Conclusions 2
2 contents IFCC approach to the standardization of HbA1c - IFCC Network - Manufacturers - Customers Conclusions 3 Electrofocusing of hemoglobins HbA3 HbA1c HbAo HbF Intmed methb HbS HbA2 courtesy Prof. J.O. Jeppson 4
3 Primary reference materials (IRMM 466 and 467) Secondary reference materials (blood panels) CRM AD500 (?) IFCC reference measurement procedure (HPLC-CE or HPLC-MS) IFCC Network Manufacturer s working calibrator Manufacturer s product calibrator Patient Sample Manufacturer s internal reference measurement procedure Manufacturer s standing measurement procedure Routine measurement procedure Manufacturer Individual laboratory Mosca et al, Clin Chem Lab Med 2007;45: HbA 1c is the stable adduct of glucose to the N-terminal valine of the β chain of hemoglobin Hexapeptide N-terminal Val β chains α chains 6
4 Flow chart for Reference Methods Hemolysate Enzymatic Cleavage HPLC-Capillary Capillary Electrophoresis HPLC-Mass Spectrometry Approved by IFCC Jeppsson et al, Clin Chem Lab Med 2002;40:78-89 Comparison between the IFCC reference method and three designated comparison methods 12,00 NGSP = 0,92xIFCC + 2,15 10,00 DSM 8,00 6,00 NGSP Mono-S JDS 4,00 Mono S = 0,989xIFCC ,00 JDS = 0,927xIFCC+1,73 0,00 0,00 2,00 4,00 6,00 8,00 10,00 12,00 % HbA1c IFCC method Hoelzel et al Clin Chem 2004;50:
5 Clin Chem 2004; 50:
6 Dϋsseldorf 2 Study Cape Town Study Berlin 2 Study California 2 Study
7 California Berlin Cape Town Dϋsseldorf HbA 1c, mmol/mol (approved) Melbourne Amsterdam Orlando Los Angeles HbA 1c, % (approved) contents IFCC approach to the standardization of HbA 1c - IFCC Network - Manufacturers -Customers Conclusions 14
8 Monitoring Programme Manufacturers 2009 CV in % 6.0% 4.0% 2.0% HPLC Immuno Affinity POCT IFCC-RM mmol/mol NGSP % Best Performance 15 Bias A. Mosca -UniMI 16
9 17 contents IFCC approach to the standardization of HbA 1c - IFCC Network - Manufacturers -Customers Conclusions 18
10 Analytical goals Braga et al, Clin Chim Acta 2011;411: Weykamp et al, Clin Chem 2011;57(8) 20
11 Essential steps for achieving the standardization of HbA 1c Mosca et al, Clin Chem Lab Med 2013, in press 21 Internal Quality Control Level 1 mean: 5.5 % CV = 1.6 % Bio-Rad Variant II, Dual Kit Lyphochek Diabetes Control Lot Level 2 mean: 9.6 % CV = 1.3 %
12 -20 C liofilo IQC CV = 2.4 % -20 C -80 C sangue CV = 2.6 % Cobas Integra Braga et al, Biochimica clinica 2010;34: IQC data from the manufacturers 24
13 EQAS fresh whole blood N = 220 labs 99 % within TE Lenters-Westra 25 et al, Diab Technol Therap 2011; 13: samples in duplicate (blind over one year), every 2 weeks EQAS lyophilized materials
14 EQAS lyophilized materials TE = 6.0 % 27 BC 2008
15 6.5 NGSP units, % int. rif DCCT 1993 SI units, mmol/mol
16 Why using IFCC standardized IFCC? Improved reliability and metrological traceability Abandoning tests with poor performance Measurement units aligned to the S.I. system Avoiding the complications of using two different measurement units (different reference intervals, different decisional limitis, different analytical goals) Simpler report Expanding the physiopathological range Greater attention to the result Better relationship to the physiopathological meaning Rohlfing et al, Diabetes Care 2002;25:275-8
17 Survey on HbA 1c Italy (1/2) (data from 237 laboratories, ) Survey on HbA 1c Italy (2/2) (data from 237 laboratories, )
18 Conclusions A complete reference system is available for HbA 1c Laboratory professionals have to carefully evaluate the methods (IFCC certificates), to calibrate them and to run periodical IQC and EQAS exercises EQAS results are difficult to be compared each other (matrix effect, target values, programmes, criteria) Manufacturers should help into developing more adequate control materials Scientific Societies should constantly stimulate and monitor the process 35 Standardization of hemoglobin A 1c : myth or reality? AND/OR 36
19 Aknowledgments RenataPaleari(University of Milano, Milano, IT) (HPLC analysis, data elaboration) Cas Weykamp (Qeen Beatrix H, Winterswijk, NL) (IFCC Network HbA 1c ) Patricia Kaiser, Hans Reinauer(INSTAND ev, Duesseldorf, DE) (ID-MS reference measurement procedure) 37
The IFCC HbA 2 standardization
The IFCC HbA 2 standardization program Andrea Mosca Centro Interdipartimentale per la Riferibilità Metrologica in Medicina di Laboratorio (CIRME) Dip. di Scienze e Tecnologie Biomediche Università degli
POCT in diagnosing and monitoring of Diabetes Mellitus. 13th EFLM Continuous Postgraduate Course, Sverre Sandberg, Noklus / EFLM
POCT in diagnosing and monitoring of Diabetes Mellitus - 13th EFLM Continuous Postgraduate Course, Sverre Sandberg, Noklus / EFLM The most important constituents - Glucose monitoring and diagnosing - U-albumin
Il processo di standardizzazione dell emoglobina A2. Andrea Mosca
Il processo di standardizzazione dell emoglobina A2 Andrea Mosca Centro Interdipartimentale per la Riferibilità Metrologica in Medicina di Laboratorio (CIRME) Dip. di Scienze e Tecnologie Biomediche Università
Harmonisatie. Cas Weykamp SKML Sectie Humorale Immunologie Congres De Waarde van de Expert 9 juni 2015
Harmonisatie Cas Weykamp SKML Sectie Humorale Immunologie Congres De Waarde van de Expert 9 juni 2015 Patient Patient Optimale Diagnose Behandeling Patient Optimale Diagnose Behandeling Patient Optimale
Screening for hemoglobin disorders: Tosoh G7 thalassemia mode
Screening for hemoglobin disorders: Tosoh G7 thalassemia mode Henri Brinkman, Frans Verheul General Clinical Laboratory (AKL) Ysselland Hospital Capelle aan den Yssel The Netherlands Stockholm, October
The Effect of Insulin Lispro on Glycemic Control in a Large Patient Cohort. Eliasson, M.D., Ph.D. 5
DIABETES TECHNOLOGY & THERAPEUTICS Volume 10, Number?, 2008 Mary Ann Liebert, Inc. DOI: 10.1089/dia.2007.0297 LIND ET AL. GLYCEMIC EFFECTS OF INSULIN LISPRO The Effect of Insulin Lispro on Glycemic Control
Symmetry in Diagnostics HPLC ANALYZER. Gold Standard Accuracy by Ion-Exchange HbA1c. TOSOH BIOSCIENCE www.tosohbioscience.us
Symmetry in Diagnostics HPLC ANALYZER Gold Standard Accuracy by Ion-Exchange HbA1c TOSOH BIOSCIENCE www.tosohbioscience.us HPLC ANALYZER HbA1c analysis in 1.6 minutes Direct determination of stable HbA1c
Evaluation of the Menarini Arkray HA 8140 hemoglobin A 1c analyzer
Clinical Chemistry 43:6 968 975 (1997) Endocrinology and Metabolism Evaluation of the Menarini Arkray HA 8140 hemoglobin A 1c analyzer W. Garry John, 1 * Francois Braconnier, 2 Kor Miedema, 3 Carlos Aulesa,
Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT
Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT Alexander Belenky, PhD Director -Product Development Samsung - Nexus-Dx Edward Brennan, PhD
Contact. Laboratory for Clinical Trials Laboratoire des Essays Cliniques Laboratorio de Ensayos Clinicos
Contact LKF - Laboratorium für Klinische Forschung GmbH Lise-Meitner-Straße 25-29 D-24223 Raisdorf, Germany Phone: +49(0)43 07-82 76 0 Fax: +49(0)43 07-82 76 79 Internet: www.lkf-kiel.de E-mail: [email protected]
Graham Beastall. [email protected]. Outline of Talk
Harmonization of Methods in Laboratory Medicine: A Means to Improve Patient Safety Graham Beastall [email protected] Outline of Talk Introduction Why Standardize or Harmonize Methods? Traceability
Hb A distribution in cord blood
2 ND European Hemoglobinopathy Forum: Insights on the Diagnosis of Hemoglobin disorders November 29th, 2011 Madrid Hb A distribution in cord blood (normal vs β + or β o thalassemia carriers) Giovanni Ivaldi
CAPILLARYS Hb A1c USING THE CAPILLARYS 2 FLEX-PIERCING INSTRUMENT
CAPILLARYS Hb A1c USING THE CAPILLARYS 2 FLEX-PIERCING INSTRUMENT 2013/06 INTENDED USE The CAPILLARYS Hb A1c kit is designed for separation and quantification of the HbA 1c glycated fraction of hemoglobin
Scottish Diabetes Survey 2013. Scottish Diabetes Survey Monitoring Group
Scottish Diabetes Survey 2013 Scottish Diabetes Survey Monitoring Group Contents Contents... 2 Foreword... 4 Executive Summary... 6 Prevalence... 8 Undiagnosed diabetes... 18 Duration of Diabetes... 18
A comparison of blood glucose meters in Australia
Diabetes Research and Clinical Practice 71 (2006) 113 118 www.elsevier.com/locate/diabres A comparison of blood glucose meters in Australia Matthew Cohen *, Erin Boyle, Carol Delaney, Jonathan Shaw International
CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES
CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES Prepared by: Technical Manager Approved by: Chief Executive Officer Approval Date: 2013-04-24 Effective
DIABETES: Applying Evidence- Based Medicine in Telehealth George E. Dafoulas MD, MBA in HSM, PhDc e- trikala SA, Greece
DIABETES: Applying Evidence- Based Medicine in Telehealth George E. Dafoulas MD, MBA in HSM, PhDc e- trikala SA, Greece www.united4health.eu Table of Contents Overview of current status and need for Evidence
Buyers guide. Point of care devices for the measurement of HbA 1c and low concentration albumin in urine CEP08057
Buyers guide Point of care devices for the measurement of HbA 1c and low concentration albumin in urine CEP08057 June 2009 Contents 2 Introduction... 3 Technical considerations... 6 Operational considerations...
CRM s and the Dilemma of the 2nd Source. Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company
CRM s and the Dilemma of the 2nd Source Shawn Kassner, Sr Product Manager Tim Miller, Sr Organic Chemist Phenova, A Phenomenex Company What are we talking about? ISO 17025 requirements for labs and CRM
& Microalbuminuria. Glycosilated Hemoglobin. Two key parameters to prevent late diabetic complicatons. Dr. Antuña de Alaíz
& Microalbuminuria Glycosilated Hemoglobin Two key parameters to prevent late diabetic complicatons Dr. Antuña de Alaíz Index Microalbuminuria Introduction... Importance of the diabetic nephropathy...
PRINCIPLE. REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE
SYNCHRON System(s) Chemistry Information Sheet Copyright 2010 Beckman Coulter, Inc. Creatine Kinase REF 442635 (200 tests/cartridge) REF 476836 (400 tests/cartridge) For In Vitro Diagnostic Use ANNUAL
There is a need for a coordinating role by the Commission supported by the Framework programme, to facilitate and harmonise those developments.
0(752/2*
Self-Monitoring of Blood Glucose in Noninsulin- Treated Diabetes
Summary and Conclusions Self-Monitoring of Blood Glucose in Noninsulin- Treated Diabetes A Systematic Review Swedish Council on Health Technology Assessment SBU Board of Directors and Scientific Advisory
Procedure for Equipment Calibration and Maintenance
Procedure for Equipment Calibration and Maintenance 1.0 Purpose This procedure specifies the schedule and requirements for calibration, performance verification, and maintenance of State Crime Laboratory
An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and
An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and CoaguChek systems Cardiovascular disease: the #1 killer in
Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians
Unaccredited Point-of-Care Laboratory Testing Guideline for Physicians Prepared by the Advisory Committee on Laboratory Medicine College of Physicians & Surgeons of Alberta Serving the Public by guiding
Validation of measurement procedures
Validation of measurement procedures R. Haeckel and I.Püntmann Zentralkrankenhaus Bremen The new ISO standard 15189 which has already been accepted by most nations will soon become the basis for accreditation
Calcium. Table 1: Difference between method means in percent
Calcium Measurement of total calcium is widely used for both the diagnosis and the monitoring of a range of conditions related to the bones, heart, nerves, and kidneys. Total calcium measurements include
Six Sigma Quality New Metrics vs Old Myths
Six Sigma Quality New Metrics vs Old Myths James O. Westgard, PhD Professor, Dept Path & Lab Medicine University of Wisconsin Medical School President, Westgard QC, Inc. Madison, WI http://www.westgard.com/sixsigprimer.htm
NABL accreditation is a formal recognition of the technical competence of a medical testing laboratory
NABL accreditation is a formal recognition of the technical competence of a medical testing laboratory Third party assessment NABL has Mutual Recognition Arrangement with APLAC and ILAC NABL accredited
A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories
A Practical Guide to Internal Quality Control (IQC) for Quantitative Tests in Medical Laboratories () Edited by Richard Pang, PhD, FACB Senior Consultant Pro Q Consulting Services Hong Kong For Hong Kong
GUIDELINES ON MEDICAL DEVICES. IVD GUIDANCES : Borderline issues A GUIDE FOR MANUFACTURERS AND NOTIFIED BODIES
EUROPEAN COMMISSION ENTERPRISE DIRECTORATE-GENERAL Single Market : regulatory environment, standardisation and New Approach Pressure equipment, medical devices, metrology MEDDEV. 2.14/1 rev. 1 January
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services. ISPRA 25 June 2009
Reference Materials for Environmental Performance Testing Dr Steve Wood Head of Regulatory and Legislative Services ISPRA 25 June 2009 Outline Background to LGC UK MCERTS scheme Reference materials production
Quality Requirements of a Point of Care Testing Service
Quality Requirements of a Point of Care Testing Service Adil I. Khan MSc, PhD Assistant Professor of Pathology Director of Point of Care Testing & Clinical Chemistry Laboratory, Temple University Episcopal
Your diabetes: Understanding your blood glucose test results. Information for patients Diabetes Service. HbA1c. Large Print. What is the HbA1c test?
Your diabetes: Understanding your blood glucose test results This leaflet explains the different ways your diabetes blood glucose control is measured. Knowing the level of glucose in your blood is important
Using a Flow Sheet to Improve Performance in Treatment of Elderly Patients With Type 2 Diabetes
Special Series: AAFP Award-winning Research Papers Vol. 31, No. 5 331 Using a Flow Sheet to Improve Performance in Treatment of Elderly Patients With Type 2 Diabetes Gary Ruoff, MD; Lynn S. Gray, MD, MPH
For In Vitro Diagnostic Use
SYNCHRON System(s) Chemistry Information Sheet Creatine Kinase REF (200 tests/cartridge) 442635 REF (400 tests/cartridge) 476836 For In Vitro Diagnostic Use ANNUAL REVIEW Reviewed by: Reviewed by: Date
Cord Blood Screening for Haemoglobin E (Hb E) Syndrome by Capillary Electrophoresis
Cord Blood Screening for Haemoglobin E (Hb E) Syndrome by Capillary Electrophoresis Rosline H., W. Asmuni W. Mohd Saman, Imilia Ismail, Che Anuar CY, Shafini Mohd Yusoff, Rosnah B Department of Hematology,
The Need to Harmonize Clinical Laboratory Test Results
The Need to Harmonize Clinical Laboratory Test Results A White Paper of the American Association for Clinical Chemistry July 2015 American Association for Clinical Chemistry Table of Contents Executive
Scottish Diabetes Survey 2014. Scottish Diabetes Survey Monitoring Group
Scottish Diabetes Survey 2014 Scottish Diabetes Survey Monitoring Group Contents Table of Contents Contents... 2 Foreword... 4 Executive Summary... 6 Prevalence... 8 Undiagnosed diabetes... 21 Duration
International Consortium for Harmonization of Clinical Laboratory Results. Operating Procedures
International Consortium for Harmonization of Clinical Laboratory Results Operating Procedures Approved: February 11, 2014 Background Results from clinical laboratory measurement procedures should be comparable
EN ISO 22870 accreditation for blood gases
EN ISO 22870 accreditation for blood gases Michel Vaubourdolle SFBC Accreditation WG chairman EFLM ISO/A WG chairman Head of Department of Laboratory Medicine* and Pathology University Hospitals Paris-East,
Contractor Information. LCD Information. Local Coverage Determination (LCD): HbA1c (L32939) Contract Number 11202
Local Coverage Determination (LCD): HbA1c (L32939) Contractor Information Contractor Name Palmetto GBA opens in new window Contract Number 11202 Contract Type MAC - Part B LCD Information Document Information
MS in Transfusion and Transplantation Sciences Assessment Plan I. Program Outcomes
MS in Transfusion and Transplantation Sciences Assessment Plan I. Program Outcomes Program Outcomes (POs) are defined as: the knowledge, skills, or behaviors that a program s students should be able to
Your diabetes: Understanding your blood glucose test results. Information for patients Diabetes Service. HbA1c. Large Print. What is the HbA1c test?
Your diabetes: Understanding your blood glucose test results This leaflet explains about the different ways your diabetes blood glucose control is measured. Knowing the level of glucose in your blood is
Laboratory Procedure Manual
Laboratory Procedure Manual Analyte: Matrix: Glycohemoglobin Whole Blood Method Tosoh A1c 2.2 Plus Glycohemoglobin Analyzer as performed by: Fairview-University Medical Center University Campus Collaborative
Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1
Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification
James O. Westgard, PhD, 1,2 and Sten A. Westgard, MS 2. Abstract
Clinical Chemistry / ANALYTIC QUALITY TODAY The Quality of Laboratory Testing Today An Assessment of σ Metrics for Analytic Quality Using Performance Data From Proficiency Testing Surveys and the CLIA
Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)
Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs) James H. Nichols, PhD, DABCC, FACB Professor of Pathology, Microbiology, and Immunology Medical Director, Clinical
Scottish Diabetes Survey
Scottish Diabetes Survey 2011 Scottish Diabetes Survey Monitoring Group Foreword The Scottish Diabetes Survey 2011 data reflects many aspects of the quality of diabetes care across the whole of Scotland.
FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)
FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs) Barriers to Overcoming Interferences and Limitations Alan T. Cariski, MD, JD, and Mike Flis, AdvaMed BGM
Content Sheet 10-1: Overview of External Quality Assessment (EQA)
Content Sheet 10-1: Overview of External Quality Assessment (EQA) Role in quality management system Assessment is a critical aspect of laboratory quality management, and it can be conducted in several
Laboratory Quality Metrics
Laboratory Quality Metrics Laila O. Abdel-Wareth. MBBCh, FCAP, FRCPC, FRCPath, EMHCA Chief Medical Officer- Pathology & Laboratory Medicine Services SEHA Chair of Pathology & Laboratory Medicine Institute
Comparison of Three Methods for Measurement of HbA1c
Türk Biyokimya Dergisi [Turkish Journal of Biochemistry Turk J Biochem] 2010; 35 (4) ; 344 349. Technical Report [Yöntem Bildirisi] Yayın tarihi 15 Aralık, 2010 TurkJBiochem.com [Published online 15 December,
Quality Assurance of Vitamins
Quality Assurance of Vitamins Ronda Greaves Medical Scientist: The Royal Children s Hospital Melbourne Chair: AACB Vitamins working party [email protected] APCCMS 14th Jan 2010 Outline Definition
ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi
ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi The Innovation Hub, Pretoria Presentation layout 1. Introduction ISO/IEC 17025 2. Requirements
Certificate of Analysis NFKK Reference Serum X Components in Human Serum, Lot Number: NFKK2002a
Scandinavian Society of Clinical Chemistry Certificate of Analysis NFKK Reference Serum X Components in Human Serum, Lot Number: NFKK2002a Version 7 Description of the material The NFKK Reference Serum
BEIPH Final Report. QCMD 2010 Hepatitis B Virus DNA (HBVDNA10A) EQA Programme. William G MacKay on behalf of QCMD and its Scientific Council July 2010
QUALITY CONTROL for MOLECULAR DIAGNOSTICS The Altum Building, Todd Campus, West of Scotland Science Park, Glasgow, G20 0XA Scotland Tel: +44 (0) 141 945 6474 Fax: +44 (0) 141 945 5795 www.qcmd.org [email protected]
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant 2 USP Stimuli Article Lifecycle Management of Analytical Procedures: Method Development, Procedure
Laboratory Procedure Manual
Laboratory Procedure Manual Analyte: Matrix: Glycohemoglobin Whole Blood Method Tosoh G 7 HPLC Glycohemoglobin Analyzer as performed by: Fairview-University Medical Center University Campus Collaborative
The Problem of the Use of different Units for the same Analyte
The Problem of the Use of different Units for the same Analyte EQALM Symposium 2009 Ulla Tiikkainen ulla.tiikkainen @labquality.fi Comparability of results Measurement traceability to the reference material
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (11) 204 3R HANDLING AND USE OF REFERENCE STANDARDS IN THE OMCL NETWORK Full document title and reference Document type Legislative
*FASTING BLOOD GLUCOSE AND GLYCOSYLATED HAEMOGLOBIN LEVELS IN RANDOMLY SELECTED GHANAIAN DIABETIC PATIENTS THE CLINICAL IMPLICATIONS
*FASTING BLOOD GLUCOSE AND GLYCOSYLATED HAEMOGLOBIN LEVELS IN RANDOMLY SELECTED GHANAIAN DIABETIC PATIENTS THE CLINICAL IMPLICATIONS F.K.N. Arthur, 1 F.A. Yeboah, 2 K. Nsiah, 1 P.K.N. Nkrumah, 1 K.A. Afreh
SMF Awareness Seminar 2014
SMF Awareness Seminar 2014 Clinical Evaluation for In Vitro Diagnostic Medical Devices Dr Jiang Naxin Health Sciences Authority Medical Device Branch 1 In vitro diagnostic product means Definition of IVD
Electronic Raw Data and the Use of Electronic Laboratory Notebooks
7th WRIB- M-2 Workshop Electronic Raw Data and the Use of Electronic Laboratory Notebooks Ron Shoup Outline for Today Raw Data- defining it, creating it, and keeping it for posterity Hybrid Data- paper,
Optimizing the Post-implementation Monitoring of a LC-MS/MS Method
Optimizing the Post-implementation Monitoring of a LC-MS/MS Method Julianne Cook Botelho, PhD Centers for Disease Control and Prevention Atlanta, GA AACC- Mass Spectrometry in the Clinical Lab: Best Practice
Laboratory medicine as one of the pharmacists competencies
2014 EAFP Annual Conference Laboratory medicine as one of the pharmacists competencies Borut Božič Examinations of the body fluids in order to recognize the state of health of the individuals Greco-Roman
Quality Management Systems for Seed Testing Laboratories: Presented to the 2010 CSAAC Meeting. Valerie Martz Senior Laboratory Accreditation Officer
Quality Management Systems for Seed Testing Laboratories: Presented to the 2010 CSAAC Meeting Valerie Martz Senior Laboratory Accreditation Officer Outline 1. Quality Management Systems (QMS): Generic
A Sensitive Assay Method of Creatine Kinase-MB Isoenzyme in Human Serum Using Anti-CK-M Antiserum and Firefly Luciferase*
Vol. 27. No. 3. 1983 (151) A Sensitive Assay Method of Creatine Kinase-MB Isoenzyme in Human Serum Using Anti-CK-M Antiserum and Firefly Luciferase* Toshio Imai**, Mitsutaka Yoshida**, Kohichi Suzuki***
ADMINISTRATIVE MANUAL Policy and Procedure
ADMINISTRATIVE MANUAL Policy and Procedure TITLE: Point of Care Testing NUMBER: CH 30-111 (Laboratory Diagnostic Bedside Testing) Effective Date: January 2014 Page 1 of 6 Applies To: Holders of Administrative
Natalia Olchanski, MS, Paige Lin, PhD, Aaron Winn, MPP. Center for Evaluation of Value and Risk in Health, Tufts Medical Center.
ISPOR 2013, New Orleans, LA Using EMR data for conducting retrospective studies: Opportunities and Pitfalls Natalia Olchanski, MS, Paige Lin, PhD, Aaron Winn, MPP Center for Evaluation of Value and Risk
Workshop A Tara Kadis
Workshop A Tara Kadis Considerations/barriers in decision making about insulin verses GLP-1 use in people with type 2 diabetes Which Insulin regimes should we consider? Diabetes is a progressive multi-system
Application of Sigma Metrics for the Assessment of Quality Assurance in Clinical Biochemistry Laboratory in India: A Pilot Study
Ind J Clin Biochem (Apr-June 2011) 26(2):131 135 DOI 10.1007/s12291-010-0083-1 ORIGINAL ARTICLE Application of Sigma Metrics for the Assessment of Quality Assurance in Clinical Biochemistry Laboratory
Sigma Metrics in the Clinical Laboratory: From Theory to Practice
Sigma Metrics in the Clinical Laboratory: From Theory to Practice D R. D A N A B A I L E Y U N I V E R S I T Y O F T O R O N T O G A M M A - D Y N A C A R E Quality indicators are everywhere 1 Objectives
Insulin therapy in various type 1 diabetes patients workshop
Insulin therapy in various type 1 diabetes patients workshop Bruce H.R. Wolffenbuttel, MD PhD Dept of Endocrinology, UMC Groningen website: www.umcg.net & www.gmed.nl Twitter: @bhrw Case no. 1 Male of
David Shu, MD, FRCPC Endocrinology, Royal Columbian Hospital October 8 th, 2010
David Shu, MD, FRCPC Endocrinology, Royal Columbian Hospital October 8 th, 2010 Objectives At the end of the talk, the participants will be able to: 1. Identify the increasing prevalence of type 2 diabetes
Overview of ISO 17511 for 'Biologicals'
JCTLM Symposium on Reference Measurement Systems for Biologicals International Bureau of Weights and Measures Pavillon de Breteuil, Sèvres, FR 2004-12-15 Overview of ISO 17511 for 'Biologicals' R. Dybkaer
Investigation of the effect of isomaltulose (PalatinoseTM) on metabolic parameters in subjects with Type 2 Diabetes.
PLEASE NOTE: This trial has been registered retrospectively. Trial Description Title Investigation of the effect of isomaltulose (PalatinoseTM) on metabolic parameters in subjects with Type 2 Diabetes.
Prevalence of Diabetic Retinopathy: A Pilot Study from the SEARCH for Diabetes in Youth Study
Prevalence of Diabetic Retinopathy: A Pilot Study from the SEARCH for Diabetes in Youth Study Elizabeth J Mayer-Davis, PhD University of North Carolina at Chapel Hill For the SEARCH Study Group, May 2014
EffiValidation 3.0 software basic training module
EffiValidation 3.0 software basic training module Oct 3, 2012 www.effichem.com Agenda EffiChem company EffiValidation 3.0 software : overview International standards and guidelines Software features Structure
Global Database on Blood Safety
Global Database on Blood Safety Summary Report 2011 1 Key facts Global Blood Collection: Around 92 million blood donations are collected annually from all types of blood donors (voluntary unpaid, family/replacement