ISO How to Meet. Requirements for Method Verification. Prepared by:
|
|
|
- Elfreda Allen
- 9 years ago
- Views:
Transcription
1 How to Meet ISO Requirements for Method Verification Prepared by: AOAC INTERNATIONAL 481 N. Frederick Ave, Suite 500 Gaithersburg, MD 20877, USA The Scientific Association Dedicated to Analytical Excellence
2 Acknowledgments At the request of the AOAC Technical Division for Laboratory Management (TDLM), the Analytical Laboratory Accreditation Criteria Committee (ALACC) prepared this guide in an effort to provide guidance to laboratory managers by clarifying the activities needed to verify that a laboratory can perform a method successfully. Experts from various groups representing different sectors of the analytical community drafted and reviewed the guide. The guide shows the consensus of a diverse group of stakeholders, including accreditation bodies, laboratories, auditors, statisticians, and pharmaceutical and feed industries. European viewpoints are also represented. The guide was written under the leadership of: M.L. JANE WEITZEL, Watson Pharmaceuticals, USA, Chair of ALACC ALACC sub-chairs: SUZANNE M. LEE, General Mills, USA, Chair of the Chemistry Subcommittee MICHELE SMOOT, Silliker Laboratories Group, USA, Chair of the Microbiology Subcommittee NUBIA VIAFARA, Cangene Corp., Canada, Chair of the Pharmaceutical Subcommittee In addition, MICHAEL BRODSKY, Brodsky Consultants, Canada, contributed the microbiology section.
3 PURPOSE The purpose of the guide is to define the activities that are required to fulfill method verification based on analytical method performance characteristics. ISO 17025:2005 section states: The laboratory shall confirm that it can properly operate standard methods before introducing the tests or calibrations. If the standard method changes, the confirmation shall be repeated. In this guide, to confirm is the same as to verify. Verification that a laboratory can adequately operate a standard method requires that the laboratory provide objective evidence the performance parameters specified in the test method have been met with the matrices to which the method is being applied. Most often, the critical requirements are the accuracy and the precision (generally accepted as repeatability and reproducibility) which are reflected in the measurement uncertainty. The objective evidence is the accuracy and precision obtained from actual lab data. SCOPE AND APPROACH The scope of this guide encompasses AOAC Official Methods SM, EPA, FDA and FSIS official methods, and NADA methods and methods used in Microbiological, Food and Pharmaceutical labs. Different industries may have differing terminology when describing categories of analytical methods and analytical parameters. This guide attempts to use the terminology commonly used by AOAC. Different industries may have specific requirements. A particular federal agency or client may have very specific criteria for method verification. In this case the client s or agency s requirements would override those in this guide. The analytical test methods are grouped according to the category of method based on its purpose. The lab can identify the category of test method it is verifying and find the corresponding parameters that need to be verified. When a method is verified, the laboratory is required to demonstrate that it can achieve certain specific performance characteristics/parameters established during the validation study. The validation study must contain all pertinent performance characteristics. Certain performance characteristics, such as linearity, will not vary from lab to lab and do not need to be verified. Other parameters, such as repeatability, are specific to the lab performing the method and need to be verified. Thus, the performance characteristics that need to be verified are a subset of the performance characteristics included in a method validation. This guide treats chemical test methods and microbiology methods separately. CHEMICAL METHODS Categories of Chemical Methods Chemical analytical methods fulfill many different purposes, from quantifying an analyte at a low concentration to identifying a material. With such a variety of methods, it is logical that different test methods require varying verification. For ease of discussion, the test methods can be divided into six different categories 2007 ALACC Guide 3
4 based on their purpose. The categories are listed below. For each of the categories of test methods only relevant performance characteristics need to be included in a method verification. The approach of this guide is to list all performance characteristics needed for verification, and explain the reason for verifying the performance characteristic. The six categories of chemical analytical methods are: 1. Confirmation of Identity, a method that ensures a material is what it purports to be or confirms the detection of the target analyte. 2. Quantifying an analyte at a low concentration. 3. Determining if an analyte is present above or below a specified, low concentration (often called a Limit Test). The specified concentration is close to the LOQ. 4. Quantifying an analyte at a high concentration. 5. Determining if an analyte is present above or below a specified, high concentration (often called a Limit Test). The specified concentration is substantially above the LOQ. 6. Qualitative test. Since the activities needed for method verification are a subset of those needed for validation, the required performance characteristics for validation will be presented first. The performance characteristics needed for the validation of each of six main categories of chemical test methods are identified in Table 1. If a performance characteristic is not needed for validation, it is not needed for verification. In Table 1, Yes means the performance characteristic must be included for validation and No means the performance characteristic does not need to be included for validation. Requirements of Method Verification for the Six Categories of Chemical Test Methods (Tables 2 6) In Tables 2 5, Yes means the performance characteristic must be included for verification and No means the performance characteristic does not need to be included for verification ALACC Guide 4
5 Table 1. Categories of Chemical Test Methods: Since the activities needed for method verification are a subset of those needed for validation, the required performance characteristics for validation are presented in this table Performance Characteristic Identification 1 Performance Characteristics Included in a Validation Analyte at Low Concentration Quantitative 2 Analyte at Low Concentration Limit Test 3 Analyte at High Concentration Quantitative 4 Analyte at High Concentration Limit Test 5 Qualitative 6 Accuracy No Yes No Yes Yes No Precision No Yes No Yes Yes No Specificity Yes Yes Yes Yes Yes Yes LOD No Yes Yes Yes/No No No LOQ No Yes No Yes/No No No Ruggedness No Yes No Yes No No Linearity/Range No Yes No Yes No No Table 2. Category 1: Confirmation of Identity A method that ensures a material is what it purports to be or confirms the detection of the target analyte Performance Characteristic Verification Verification Activities Reason for Verification Specificity No if the lab s samples are identical to those in the standard method and if any differences in instrumentation do not impact specificity. Yes if the lab s samples differ from those in the standard method. Yes if differences between instruments could affect specificity. NA Same as those required for validation. The activity need only deal with the unique aspect s of the lab s samples or instrumentation. If the samples have the same matrix, the specificity which is based on basic principles, will not be impacted. Basic principles are chemical reactions, e.g. reaction of Ag with Cl to create a precipitate. Specificity can be impacted by differences in instrumentation ALACC Guide 5
6 Table 3. Category 2: Analyte at Low Concentration, Quantitative Performance Characteristic Verification Verification Activity Reason for Verification Accuracy Yes If the concentration range for which the method is validated is narrow (<1 order of magnitude), analyze one reference material/standard/spike at one concentration. Otherwise, demonstrate accuracy at each concentration level (low, middle and high) by analyzing one reference material/standard/spike at each level. Precision Yes Perform the repeatability test once. If the method covers a concentration range >1 order of magnitude, then the repeatability test must include low, middle and high concentrations. Over a narrow concentration range, the accuracy and precision should not vary, therefore, the demonstration at one concentration is sufficient. Over a wide concentration range, the accuracy and precision can vary, thus they need to be verified at the different concentration levels. Over a narrow concentration range, the accuracy and precision should not vary, therefore, the demonstration at one concentration is sufficient. Over a wide concentration range, the accuracy and precision can vary, thus they need to be verified at the different concentration levels. Intermediate precision, between analysts, is handled by making sure the analysts are trained and can adequately perform the method. Specificity No/Yes See Specificity in General Requirements See Specificity in General Requirements LOD Yes Run a sample close to LOD LOD is very likely to be matrix and instrument specific LOQ Yes Run a sample close to LOQ LOQ is very likely to be matrix and instrument specific Table 4. Category 3: Analyte is present above or below a specified, low concentration (Limit Test) Performance Characteristic Verification Verification Activity Reason for Verification LOD Yes Run a sample close to LOD LOD is very likely to be matrix and instrument specific LOQ Yes Run a sample close to LOQ LOQ is very likely to be matrix and instrument specific Specificity No/Yes See Specificity in General Requirements See Specificity in General Requirements 2007 ALACC Guide 6
7 Table 5. Category 4: Quantifying an analyte at high concentration and Category 5: Analyte above or below a specified, high concentration (often called a Limit Test) Performance Characteristic Verification Verification Activity Reason for Verification Accuracy Yes If the method is a limit test or if the concentration range for which the method is validated is narrow (<1 order of magnitude), analyze one reference material/standard/spike at one concentration. Otherwise, demonstrate accuracy at each concentration level, low middle and high by analyzing one reference material/standard/spike at each level. Precision Yes Perform the repeatability test once. If the method covers a concentration range >1 order of magnitude, then the repeatability test must include low, middle and high concentrations. Over a narrow concentration range, the accuracy and precision should not vary, therefore, the demonstration at one concentration is sufficient. Over a wide concentration range, the accuracy and precision can vary, thus they need to be verified at the different concentration levels. Over a narrow concentration range, the accuracy and precision should not vary, therefore, the demonstration at one concentration is sufficient. Over a wide concentration range, the accuracy and precision can, thus they need to be verified at the different concentration levels. Intermediate precision, between analyst, is handled by making sure the analysts are trained and can adequately perform the method. Specificity No/Yes See Specificity in General Requirements See Specificity in General Requirements Table 6. Category 6: Qualitative Tests Qualitative tests are used to identify a specific element or compound (analyte) based on the response of a material to the test. The most important characteristic of a qualitative test is its ability to reliably identify the analyte in the presence of other substances. This is referred to as the specificity. Method validation includes determining any cross reactivity with other known entities. The lack of cross reactivity demonstrates the specificity of the method. If samples are identical to those for which the method is intended, no verification of specificity is required. If any matrix components are unique to the lab s samples, the lab will need to demonstrate there is no impact on specificity. The method precision of qualitative tests is generally expressed as false-positive/false-negative rates and is determined at several concentration levels. Verification of a lab s ability to properly operate a qualitative method can be demonstrated by analyzing populations of negative and positive fortified samples. For example, for each different sample matrix, duplicate samples are analyzed at three levels. Suggested levels are blanks (no analyte), low level (near the lower range of the method) and high level (near the high end of the range). Standard additions can be used to obtain the correct concentration levels. Rates comparable to those stated in the validated method demonstrate the labs ability to operate the method ALACC Guide 7
8 USING MEASUREMENT UNCERTAINTY IN METHOD VERIFICATION OF CHEMICAL TESTS The estimate of measurement uncertainty (MU) for a measurand is the indicator of precision, and the MU is one of the components that can be used to verify a lab can perform the method satisfactorily. When there is no validation data available, the MU may be the only parameter that can be compared to the specification to ensure the method works. The comparison of a lab s performance (bias and precision) can be compared to those from the collaborative study using the approach as described in ISO Technical Specification ISO/TS 21748, Guidance for the use of repeatability, reproducibility and trueness estimates in measurement uncertainty estimation. Additional guidance, including examples of using the approach suggested in ISO/TS will be posted on the AOAC Website. MICROBIOLOGICAL METHODS The performance characteristics needed for the validation and verification of each of three main categories of microbiological test methods are identified in Table 7. Verification Guidelines Verification of microbiological methods also requires that the following parameters are addressed: 1. Laboratory competency of achieving method performance characteristics on an on-going basis. 2. Analyst performance: Can your analysts perform the method with the equivalent degree of precision and accuracy? Table 7. Categories of Microbiological Test Methods: Performance Characteristics Included in a Validation Study Performance Characteristic Identification Quantitative Qualitative (P/A) Verification (where applicable) Relative Accuracy Yes Yes No No Matrix Effects No Yes Yes Yes Precision No Yes No Yes Selectivity No Yes Yes No Specificity Yes Yes Yes No Inclusivity Yes Yes Yes No Exclusivity Yes Yes Yes No False-Positive Rate No Yes Yes No False-Negative Rate No Yes Yes No LOD No Yes Yes No LOQ No Yes No No Ruggedness Yes Yes Yes No Linearity/Range No Yes No No 2007 ALACC Guide 8
9 Measurement Uncertainty for Microbiology Measurement uncertainty is a parameter associated with the result of a measurement that characterizes the dispersion of the values that could reasonably be attributed to the measurand, in other words, the variance or standard deviation of the result or the degree of confidence in an analytical result. Calculation of MU is more reliable with a minimum of 30 observations in order to obtain the 95% confidence limits of the overall precision of the method. Therefore it is not unreasonable to expect that each of the required performance characteristics would also be based on 30 observations. MU usually requires the inclusion of repeatability and reproducibility of all factors that contribute to >10% of the variability and uses the root sum of squares in the calculation, e.g. where there are 2 key factors (Uc) = RSD1 2 + RSD2 2. In its simplest determination, however relativemu or RelativeUe = 2 RSDR (Relative Standard Deviation of Reproducibility) when all factors that could impact on the reliability of a result are taken into consideration. In this regard, Validation data can be used comparatively for method verification since the validation study provides reference values for RSDR and RSDr (Relative Standard Deviation of Reproducibility and Relative Standard Deviation of Repeatability). For example a validation study of Pour Plate counting obtained the following data: Calculation of combined uncertainty for counting: RSDr 7.7% (0.077) (within analysts) RSDR 18.2% (0.182) (between analysts) Sum of squares: (0.077) 2 + (0.182) 2 = Combined relative uncertainty = (0.0371) = = 19.3% Expanded relative uncertainty for counting: (Use coverage factor k = 2 for 95% confidence) = % = 38.6% The MU for this technique is 38.6% and becomes the reference value for the MU determined by in-house method verification. Experimental data for calculating MU can be obtained from: Proficiency testing (PT) programs Reference samples Spike recovery Method verification replicates 2007 ALACC Guide 9
10 Sample duplicates Additional information on estimating Measurement Uncertainty for microbiological methods can be found in the American Association for Laboratory Accreditation in the Guideline G108 Guidelines for Estimating Uncertainty for Microbiological Counting Methods. Analyst Performance in Microbiology Only analysts who have been adequately trained to perform a method should participate in method verification. Training and qualification of analysts require a written protocol of activities with documented results of achievement. Trained analysts need to be assessed on an on-going basis. Assessment tools may include: Blind samples (known positive and known negative samples) Observation of performance Written test; e.g. on Quality Control Practices, calculations, interpretation of results, knowledge of quality policies and procedures Daily checks on precision of duplicate counts* Proficiency testing** *Precision of Duplicate Counts: 1. All analysts perform duplicate counts on samples for each type of matrix, record as D1 and D2 2. Convert data to Log10 3. Determine the absolute value of the range and its log (Rlog) for each pair (D1 D2) 4. Calculate the mean for each set of data = 5. Analysts can now run daily duplicates 6. 99% Precision criteria: R Plot results on a control chart to monitor variability R n log 2007 ALACC Guide 10
11 Ref. SMEWW th ed. (9 11) **Proficiency Testing 1. Intralaboratory Proficiency Testing: Internally controlled check sample or spiked samples used as: An ongoing assessment of analytical performance and competency by individual analysts within a laboratory A means to demonstrate competence for tests not covered by external proficiency programs To detect training needs 2. Interlaboratory Proficiency Testing Proficiency panels from outside source; e.g. AOAC, FDA (a) Provides a measure of precision and relative accuracy of analytical methods performed by different laboratories (b) Provides an estimate of the relative accuracy and precision of results between laboratories (c) Can be used for intralaboratory proficiency testing, but by itself is not designed for assessment of the competency of individual analysts GENERAL REQUIREMENTS (for chemical and microbiological methods) The lab must have a management system consistent with that described in ISO/IEC 17025:2005. The analysts must be adequately trained: The analyst must have the appropriate knowledge, experience, and training to perform the procedure. The analyst must be competent in performing the given functions of the lab: operating the instrumentation performing the analysis as specified understanding the analytical technique The training must be documented ALACC Guide 11
12 The verification exercise must be planned, approved, and documented and a final report kept on file to demonstrate the lab can perform the method adequately. Verification documentation should be created that: Describes the procedure to be verified Details the analytical performance characteristics to be evaluated Establishes acceptance criteria used to determine that the procedure performs suitably Justifies deviations from Reference Method General Requirements Specificity For specificity in all categories of methods, if samples are identical to those for which the method is intended and validated, and the method is based on basic principles then no verification is needed. If the samples have the same matrix, the specificity which is based on basic principles, will not be impacted. Basic principles are chemical reactions, e.g. reaction of Ag with Cl to create a precipitate. For some methods, the specificity can be affected by the instrument used. In these cases the lab should assess if the instrument differences could affect the specificity, and if so, include specificity in the verification, e.g. the different resolution and/or detection systems in inductively coupled plasma optical emission spectrophotometers may result in different interferences. METHOD SELECTION FIT FOR USE Analytical methods are evaluated based on attributes such as accuracy, precision, specificity, sensitivity, detectability and practicality. Compromise between attributes is inherent in the selection of methods. However, any method selected for use must be appropriate to the requirements of the regulatory function and must be within the capabilities of the laboratory staff. Depending on the documentation available on a method, varying degrees of method verification or validation are recommended before it is adopted for routine use. The lab must first ensure the chosen method is applicable to the samples the lab will be analyzing with the method. If there is any difference between the lab s samples and those for which the method is validated, the extent of the difference and its impact must be assessed. The impact of the difference could require a partial validation to include the performance characteristics that are impacted or the method may require complete validation. CIRCUMSTANTIAL CHANGES/DEVIATIONS FROM THE METHOD AS VALIDATED For a number of reasons, a lab may not be able to perform the method exactly as validated. In these circumstances, the method verification may have to deal with these differences. A number of unforeseen scenarios are shown in Table 8. If a scenario does occur the listed equivalency study shall be performed. There may be other scenarios; this list is not all inclusive ALACC Guide 12
13 Table 8. Parameter in Validated Method Difference from Validated Method Required Equivalence Study a. Apparatus and equipment used a. Either some of the apparatus and equipment are outdated or the lab does not have them, and the lab wants to use available apparatus and equipment b. Specific detector b. The lab may not have the specific detector and the lab would like to use one that is available. c. Sample preparation: Occasionally samples are prepared in liquid nitrogen to prevent loss of analyte(s) d. Other possible scenarios: i. Columns and cartridges ii. Mobile phase iii. Mobile phase gradient iv. Throughput c. If the lab has production type environment, it can process sample quickly at room temperature. i. The lab wishes to use available columns and cartridges of different vendors. ii. Occasionally mobile phase used in the validated method is found to be corrosive and detrimental to pump and seals. In this case a compatible substitute can be used. iii. The lab may not have the required 3-pump gradient system; solvents are pre-mixed for two-pump gradient. iv. Subtle modifications in extraction and purification steps may be needed such as exchange of extraction solvent of less health hazard solvent, shaking vs. vortexing, use of compatible solvent, use of SPE cartridges of different vendor(s), etc. An instrument equivalence bridging study is needed to ensure that the numbers generated by alternate apparatus and equipment relate in some reproducible way to the concentrations determined in the validated method. b. An equivalence study is needed comparing the numbers obtained between the two detectors to ensure that the numbers generated relate in some reproducible way to the concentrations determined in the validated method. c. A bridging study comparing the two sample preparations is needed to ensure that the numbers generated relate in some reproducible way to the concentrations determined in the validated method. d. Any of the scenarios mentioned in column 2 require bridging to ensure that the numbers generated relate in some reproducible way to the concentrations in the validated method ALACC Guide 13
14 PHARMACEUTICAL To prevent confusion, the use of the term Identity in the pharmaceutical industry is explained here. In the pharmaceutical industry, identity tests are intended to ensure the identity of an analyte in a sample. Identity is the material what it is supposed to be? In other analytical fields Identity has a different meaning, e.g. in the Eurachem definition. It is necessary to establish that the signal produced at the measurement stage, or other measured property, which has been attributed to the analyte, is only due to the analyte and not from the presence of something chemically or physically similar or arising as a coincidence. This is confirmation of identity. Analytical Methods A Laboratory Guide to Method Validation and Related Topics, p. 14 Thus in the pharmaceutical industry the term Identity describes a category of analytical test while in the Eurachem guide Identity describes an analytical parameter. FOOD Chemical Testing Category 3: Method determining if an analyte is present above or below a specified, low concentration (often called a Limit Test). This category is pretty delicate as it is located at the interface of quantitative qualitative determinations. However, for food safety it represents a very important category, i.e. residues of banned vet drug, which must not be present in food of animal origin. In the EU this is covered by Decision 657/2002/EC. In the EU much more emphasis is placed on the validation of such methods in comparison to those included in this guide. DEFINITIONS Accuracy For the purposes of this document, the term accuracy is defined as: The closeness of agreement between the measured value and the accepted, true, or reference value. Accuracy is indicative of the bias of the measurement process. Accuracy is often evaluated by repetitively spiking the matrix or placebo with known levels of analyte standards at or near target values. The fraction or percentage of added analyte recovered from a blank matrix is often used as the index of accuracy. Added analyte, however, may not always reflect the condition of the natural analyte in the materials submitted for analysis. (Official Methods of Analysis of AOAC INTERNATIONAL, 18th Ed., Appendix E). This guide acknowledges that the Official Methods of Analysis (OMA) definition of accuracy is not fully in line with the current version of the VIM-3rd edition which defines in clause 2.13 (3.5) measurement accuracy accuracy of measurement accuracy closeness of agreement between a measured quantity value and a true quantity value of a measurand NOTES 2007 ALACC Guide 14
15 1. The concept measurement accuracy is not a quantity and is not given a numerical quantity value. A measurement is said to be more accurate when it offers a smaller measurement error. 2. The term measurement accuracy should not be used for measurement trueness and the term measurement precision should not be used for measurement accuracy, which, however, is related to both these concepts. 3. Measurement accuracy is sometimes understood as closeness of agreement between measured quantity values that are being attributed to the measurand. Intermediate Precision Intermediate precision: The precision of an analytical procedure expresses the closeness of agreement between a series of measurements obtained from multiple sampling of the same homogeneous sample under the prescribed conditions. Intermediate precision expresses within-laboratories variations: different days, different analysts, different equipment, etc. (ICH Validation of Analytical Procedures: Text and Methodology, Q2R). Intermediate precision: Measurement precision under a set of conditions of measurement. (ISO/DGuide , International vocabulary of basic and general terms in metrology (VIM). Third edition, 2.24 without notes.) Repeatability Repeatability (of results of measurements): Closeness of the agreement between the results of successive measurements of the same measurand carried out subject to all of the following conditions: same measurement procedure same observer same measuring instrument, used under the same conditions same location repetition over a short period of time (ISO Guide 30: 1992 A7) Repeatability Measurement precision under the set of repeatability conditions of measurement. (ISO/DGuide , International vocabulary of basic and general terms in metrology (VIM). Third edition, 2.22) 2007 ALACC Guide 15
16 Reproducibility Reproducibility (of results of measurements): Closeness of the agreement between the results of measurements of the same measurand, where the measurements are carried out under changed conditions such as: principle or method measurement observer measuring instrument location conditions of use time (ISO Guide 30:1992 A.8) Note: For the purposes of this ALACC guide a change to the principle of measurement would create a new method that would have to be validated. Reproducibility Measurement precision under reproducibility conditions of measurement. (ISO/DGuide , International vocabulary of basic and general terms in metrology (VIM). Third edition, 2.26 without notes.) Standard Method A method that has been validated by an authoritative body. Most food and pharmaceutical methods are found in AOAC, USDA, FDA, EPA, AOCS, AACC, ISO, IUPAC, USP, and FCC method manuals. Many trade associations publish their own methods and provide useful resources. A few examples include the Corn Refiners Association, National Food Processors Association, Association for Dressings and Sauces, and the American Spice Trade Association. Validated Method The planned and documented procedure to establish the method s performance characteristics. The performance characteristics or the validation parameters of the method determine the suitability for its intended use. They define what the method can do under optimized conditions of matrix solution, analyte isolation, instrumental settings, and other experimental features. The inclusion of particular validation parameters in a validation protocol depends on the application, the test samples, the goal of the method, 2007 ALACC Guide 16
17 and domestic or international guidelines or regulations, as applicable. (Official Methods of Analysis of AOAC INTERNATIONAL, 18th Ed., Appendix E) Comments: Validation of a method establishes, by systematic laboratory studies that the method is fit-for-purpose, i.e., its performance characteristics are capable of producing results in line with the needs of the analytical problem. The important performance characteristics include: selectivity and specificity (description of the measurand), measurement range, calibration and traceability, bias linearity, limit of detection/limit of quantitation, ruggedness, and precision. The above characteristics are interrelated. Many of these contribute to the overall measurement uncertainty and the data generated may be used to evaluate the measurement uncertainty. The extent of validation must be clearly stated in the documented method so that users can assess the suitability of the method for their particular needs. Examples of validated methods can be obtained from specific organizations such as AOAC INTERNATIONAL. Verification Provision of objective evidence that a given item fulfils specified requirements. EXAMPLES (a) Confirmation that a given reference material as claimed is homogeneous for the quantity value and measurement procedure concerned, down to a measurement portion having a mass of 10 mg. (b) Confirmation that performance properties or legal requirements of a measuring system are achieved. (c) Confirmation that a target measurement uncertainty can be met. NOTES 1. When applicable, measurement uncertainty should be taken into consideration. 2. The item may be, e.g., a process, measurement procedure, material, compound, or measuring system. 3. The specified requirements may be, e.g., that a manufacturer s specifications are met. 4. Verification in legal metrology, as defined in VIML [10], and in conformity assessment in general, pertains to the examination and marking and/or issuing of a verification certificate for a measuring system. 5. Verification should not be confused with calibration. Not any verification is a validation. 6. In chemistry, verification of identity of entity involved, or of activity, requires a description of the structure or properties of that entity or activity ALACC Guide 17
18 VIM 3, 2.44 Verification Confirmation, through the provision of objective evidence, that specified requirements have been fulfilled (ISO 9000: without notes). Comment: In connection with the management of measuring equipment, verification provides a means for checking that the deviations between values indicated by a measuring instrument and corresponding known values of a measured quantity are consistently smaller than the maximum allowable error defined in a standard, regulation, or specification peculiar to the management of the measuring equipment. The result of verification leads to a decision either to restore in service, perform adjustments, or repair, or to downgrade, or to declare obsolete. In all cases it is required that a written trace of the verification performed shall be kept on the measuring instrument s individual record ALACC Guide 18
Validation and Calibration. Definitions and Terminology
Validation and Calibration Definitions and Terminology ACCEPTANCE CRITERIA: The specifications and acceptance/rejection criteria, such as acceptable quality level and unacceptable quality level, with an
Methods verification. Transfer of validated methods into laboratories working routine. Dr. Manuela Schulze 1
Methods verification Transfer of validated methods into laboratories working routine Dr. Manuela Schulze 1 1. Introduction 2. Definitions and differences validation verification 3. How to perform verification
ALACC Frequently Asked Questions (FAQs)
Updated October 17, 2012 ALACC Contents Web Introduction to FAQ... 2 FAQ #6... 3 ALACC Section: How to Meet ISO 17025 Requirements for Method Verification (Page 9)... 3 FAQ #13... 4 ALACC Section ALACC
American Association for Laboratory Accreditation
Page 1 of 12 The examples provided are intended to demonstrate ways to implement the A2LA policies for the estimation of measurement uncertainty for methods that use counting for determining the number
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant
Lifecycle Management of Analytical Procedures; What is it all about? Jane Weitzel Independent Consultant 2 USP Stimuli Article Lifecycle Management of Analytical Procedures: Method Development, Procedure
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS.
GUIDELINES FOR THE VALIDATION OF ANALYTICAL METHODS FOR ACTIVE CONSTITUENT, AGRICULTURAL AND VETERINARY CHEMICAL PRODUCTS October 2004 APVMA PO Box E240 KINGSTON 2604 AUSTRALIA http://www.apvma.gov.au
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry
Step-by-Step Analytical Methods Validation and Protocol in the Quality System Compliance Industry BY GHULAM A. SHABIR Introduction Methods Validation: Establishing documented evidence that provides a high
1. PURPOSE To provide a written procedure for laboratory proficiency testing requirements and reporting.
Document #: FDPD-QMS.024.003 Page 1 of 12 Table of Contents 1. Purpose 2. Scope 3. Responsibility 4. References 5. Related Documents 6. Definitions 7. Safety 8. Equipment/Materials Needed 9. Process Description
IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS
REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and
DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES
DECISION LIMITS FOR THE CONFIRMATORY QUANTIFICATION OF THRESHOLD SUBSTANCES Introduction This Technical Document shall be applied to the quantitative determination of a Threshold Substance in a Sample
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY
USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY Breakout Session 3B Tuesday, May 1 8:30 10 am James Blackwood, MS, CLSI David D. Koch, PhD, FACB, DABCC, Pathology & Laboratory
VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY Q2(R1)
INTERNATIONAL CONFERENCE ON HARMONISATION OF TECHNICAL REQUIREMENTS FOR REGISTRATION OF PHARMACEUTICALS FOR HUMAN USE ICH HARMONISED TRIPARTITE GUIDELINE VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND METHODOLOGY
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA
FOOD FOR THOUGHT Topical Insights from our Subject Matter Experts UNDERSTANDING WHAT IS NEEDED TO PRODUCE QUALITY DATA The NFL White Paper Series Volume 7, January 2013 Overview and a Scenario With so
Implementing New USP Chapters for Analytical Method Validation
Implementing New USP Chapters for Analytical Method Validation March 2010 Ludwig Huber Fax.: +49 7243 602 501 E-mail: [email protected] Today s Agenda Handling Method Changes vs. Adjustments
Contents. Abstract...i. Committee Membership... iii. Foreword... vii. 1 Scope...1
ISBN 1-56238-584-4 Volume 25 Number 27 ISSN 0273-3099 Interference Testing in Clinical Chemistry; Approved Guideline Second Edition Robert J. McEnroe, PhD Mary F. Burritt, PhD Donald M. Powers, PhD Douglas
ASSURING THE QUALITY OF TEST RESULTS
Page 1 of 12 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6. B changed Division of Field Science and DFS to Office
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories
Guide to Method Validation for Quantitative Analysis in Chemical Testing Laboratories Wilton Park House, Wilton Place, Dublin 2, Ireland Tel +353 1 607 3003 Fax +353 1 607 3109 E-mail [email protected] Web
ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology. Step 5
European Medicines Agency June 1995 CPMP/ICH/381/95 ICH Topic Q 2 (R1) Validation of Analytical Procedures: Text and Methodology Step 5 NOTE FOR GUIDANCE ON VALIDATION OF ANALYTICAL PROCEDURES: TEXT AND
Ecology Quality Assurance Glossary
Ecology Quality Assurance Glossary Edited by William Kammin, Ecology Quality Assurance Officer Accreditation - A certification process for laboratories, designed to evaluate and document a lab s ability
Guidance for Industry
Guidance for Industry Q2B Validation of Analytical Procedures: Methodology November 1996 ICH Guidance for Industry Q2B Validation of Analytical Procedures: Methodology Additional copies are available from:
Control Charts and Trend Analysis for ISO 17025. Speakers: New York State Food Laboratory s Quality Assurance Team
Control Charts and Trend Analysis for ISO 17025 Speakers: New York State Food Laboratory s Quality Assurance Team 1 ISO 17025 Requirements, 5.9.1: The laboratory shall have quality control procedures for
1 Quality Assurance and Quality Control Project Plan
1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing
To avoid potential inspection
Good Analytical Method Validation Practice Deriving Acceptance for the AMV Protocol: Part II To avoid potential inspection observations for test method validations by the regulatory agencies, it has become
APPENDIX N. Data Validation Using Data Descriptors
APPENDIX N Data Validation Using Data Descriptors Data validation is often defined by six data descriptors: 1) reports to decision maker 2) documentation 3) data sources 4) analytical method and detection
Applying Statistics Recommended by Regulatory Documents
Applying Statistics Recommended by Regulatory Documents Steven Walfish President, Statistical Outsourcing Services [email protected] 301-325 325-31293129 About the Speaker Mr. Steven
OIML D 18 DOCUMENT. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION
INTERNATIONAL DOCUMENT OIML D 18 Edition 2008 (E) The use of certified reference materials in fields covered by metrological control exercised by national services of legal metrology. Basic principles
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
Appendix F: Guidelines for Standard Method Performance Requirements
Appendix F: Guidelines for Standard Method Performance Requirements Contents Introduction to Standard Method Performance Requirements 1 Annex A: Format of a Standard Method Performance Requirement 5 Annex
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES. October 2008 Revision 08
QUALITY MANUAL GREAT LAKES INSTITUTE FOR ENVIRONMENTAL RESEARCH ANALYTICAL LABORATORIES October 2008 Revision 08 Great Lakes Institute for Environmental Research University of Windsor Windsor, Ontario
QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES
NB: Version adopted by the World Assembly of Delegates of the OIE in May 2012 CHAPTER 1.1.4. QUALITY MANAGEMENT IN VETERINARY TESTING LABORATORIES SUMMARY Valid laboratory results are essential for diagnosis,
USE OF REFERENCE MATERIALS IN THE LABORATORY
USE OF REFERENCE MATERIALS IN THE LABORATORY What is a reference material? A general definition of a reference material is a material or substance one or more of whose property values are sufficiently
Overview of ISO 17511 for 'Biologicals'
JCTLM Symposium on Reference Measurement Systems for Biologicals International Bureau of Weights and Measures Pavillon de Breteuil, Sèvres, FR 2004-12-15 Overview of ISO 17511 for 'Biologicals' R. Dybkaer
Assay Qualification Template for Host Cell Protein ELISA
Assay Qualification Template for Host Cell Protein ELISA Introduction: With ever increasing importance being placed on host cell protein (HCP) removal during the purification process, it is critical to
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation
International GMP Requirements for Quality Control Laboratories and Recomendations for Implementation Ludwig Huber, Ph.D. [email protected] Overview GMP requirements for Quality Control laboratories
Medical Cannabis Laboratory Approval Program
Medical Cannabis Laboratory Approval Program Application Process and Required Documentation After the publication of the Medical Cannabis Laboratory Application, currently expected by February 16, 2015,
Quantifying Uncertainty in Analytical Measurement
EURACHEM / CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition QUAM:000.P1 EURACHEM/CITAC Guide Quantifying Uncertainty in Analytical Measurement Second Edition Editors S L R Ellison
Design of Experiments for Analytical Method Development and Validation
Design of Experiments for Analytical Method Development and Validation Thomas A. Little Ph.D. 2/12/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
General and statistical principles for certification of RM ISO Guide 35 and Guide 34
General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in
European cooperation for EAL-G23 THE EXPRESSION OF UNCERTAINTY IN QUANTITATIVE TESTING. The Expression of Uncertainty in Quantitative Testing
European cooperation for EAL-G23 THE EXPRESSION OF UNCERTAINTY IN QUANTITATIVE TESTING Accreditation of Laboratories Publication Reference EAL-G23 The Expression of Uncertainty in Quantitative Testing
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL)
Definition of Minimum Performance Requirements for Analytical Methods of GMO Testing European Network of GMO Laboratories (ENGL) 13 October 2008 Date of application: 13 April 2009 INTRODUCTION The scope
Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.
Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known
How to Verify Performance Specifications
How to Verify Performance Specifications VERIFICATION OF PERFORMANCE SPECIFICATIONS In 2003, the Centers for Medicare and Medicaid Services (CMS) updated the CLIA 88 regulations. As a result of the updated
State Meat and Poultry Inspection (MPI) Program Laboratory Quality Management System Checklist
This checklist is intended to help laboratories improve the quality of their testing programs with regard to meat and poultry testing samples, in order to ensure those laboratories are at least equal to
TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS
TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS. PURPOSE The purpose of this report is to verify the performance characteristics of the Sievers M-Series TOC Analyzers. Performance was quantified
American Association for Laboratory Accreditation
Page 1 of 6 1.0 INTRODUCTION These accreditation requirements are applicable to the special competition animal drug testing laboratory accreditation program. This program is designed to meet the requirements
A commitment to quality and continuous improvement
Guidance for the Validation of Analytical Methodology and Calibration of Equipment used for Testing of Illicit Drugs in Seized Materials and Biological Specimens A commitment to quality and continuous
Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory
Method Validation/Verification CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory References Westgard J. O.: Basic Method Validation, Westgard Quality Corporation Sarewitz
Gap Analysis of ISO 15189:2012 and ISO 15189:2007 in the field of Medical Testing
Gap Analysis May 2013 Issued: May 2013 Gap Analysis of and in the field of Medical Testing Copyright National Association of Testing Authorities, Australia 2013 This publication is protected by copyright
ANALYTICAL METHODS INTERNATIONAL QUALITY SYSTEMS
VALIDATION OF ANALYTICAL METHODS 1 GERT BEUVING INTERNATIONAL PHARMACEUTICAL OPERATIONS TASKS: - Internal auditing - Auditing of suppliers and contract manufacturers - Preparing for and guiding of external
Validating Methods using Waters Empower TM 2 Method. Validation. Manager
Validating Methods using Waters Empower TM 2 Method Validation Manager (MVM) Anders Janesten Nordic Informatics Sales Specialist 2008 Waters Corporation Decision-centric centric method development information
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research
LC-MS/MS Method for the Determination of Docetaxel in Human Serum for Clinical Research J. Jones, J. Denbigh, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 20581 Key Words SPE, SOLA,
Analytical Test Method Validation Report Template
Analytical Test Method Validation Report Template 1. Purpose The purpose of this Validation Summary Report is to summarize the finding of the validation of test method Determination of, following Validation
ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi
ISO 17025 and SANAS guidelines for use of reference materials and PT scheme participation Shadrack Phophi The Innovation Hub, Pretoria Presentation layout 1. Introduction ISO/IEC 17025 2. Requirements
The Fitness for Purpose of Analytical Methods. A Laboratory Guide to Method Validation and Related Topics
The Fitness for Purpose of Analytical Methods A Laboratory Guide to Method Validation and Related Topics Second Edition 2014 The Fitness for Purpose of Analytical Methods A Laboratory Guide to Method
(Uncertainty) 2. How uncertain is your uncertainty budget?
(Uncertainty) 2 How uncertain is your uncertainty budget? Paper Author and Presenter: Dr. Henrik S. Nielsen HN Metrology Consulting, Inc 10219 Coral Reef Way, Indianapolis, IN 46256 Phone: (317) 849 9577,
The Determination of Uncertainties in Charpy Impact Testing
Manual of Codes of Practice for the Determination of Uncertainties in Mechanical Tests on Metallic Materials Code of Practice No. 06 The Determination of Uncertainties in Charpy Impact Testing M.A. Lont
GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION
International Document OIML 1 st CD GENERAL REQUIREMENTS FOR THE PROGRAMME OF REFERENCE MATERIAL CERTIFICATION IN SERIAL PRODUCTION International Organization of Legal Metrology Contents 01 Scope 2 12
P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014
P07 CALA Application of Requirements in ISO/IEC 17025:2005 Revision 3.4 April 9, 2014 1 TABLE OF CONTENTS FOREWORD... 1 1.0 SCOPE... 1 1.5 Safety Requirements... 1 4.0 MANAGEMENT REQUIREMENTS... 1 4.1
ENVIRONMENTAL LABORATORY SECTOR VOLUME 1 MANAGEMENT AND TECHNICAL REQUIREMENTS FOR LABORATORIES PERFORMING ENVIRONMENTAL ANALYSIS
EL-VM-ISO-00 Formatted ENVIRONMENTAL LABORATORY SECTOR VOLUME MANAGEMENT AND TECHNICAL REQUIREMENTS FOR LABORATORIES PERFORMING ENVIRONMENTAL ANALYSIS Module : Quality Systems General Requirements Working
QUALITY ASSURANCE and QUALITY CONTROL DATA STANDARD
QUALITY ASSURANCE and QUALITY CONTROL DATA STANDARD Standard No.: EX000012.1 January 6, 2006 This standard has been produced through the Environmental Data Standards Council (EDSC). The Environmental Data
Method Validation. Introduction. Página Web 1 de 16
Página Web 1 de 16 Method Validation Introduction Strategy for Validation of Methods Validation of Standard Methods Revalidation Parameters for Method Validation Selectivity/specificity Precision and Reproducibility
Quality Assurance/Quality Control in Acid Deposition Monitoring
Quality Assurance/Quality Control in Acid Deposition Monitoring Acid Deposition and Oxidant Research Center (Network Center of EANET) Introduction Wet deposition data are used for -assessments of spatial
Interlaboratory studies
Interlaboratory studies Department of Food Chemistry and Analysis, ICT Prague Vladimír Kocourek Prague, 2012 Interlaboratory studies Various titles: Interlaboratory proficiency test or studies Interlaboratory
Signal, Noise, and Detection Limits in Mass Spectrometry
Signal, Noise, and Detection Limits in Mass Spectrometry Technical Note Chemical Analysis Group Authors Greg Wells, Harry Prest, and Charles William Russ IV, Agilent Technologies, Inc. 2850 Centerville
Assay Development and Method Validation Essentials
Assay Development and Method Validation Essentials Thomas A. Little Ph.D. 10/13/2012 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
NABL NATIONAL ACCREDITATION
NABL 162 NABL NATIONAL ACCREDITATION BOARD FOR TESTING AND CALIBRATION LABORATORIES GUIDELINES FOR PROFICIENCY TESTING PROGRAM for TESTING AND CALIBRATION LABORATORIES ISSUE NO : 04 AMENDMENT NO : 00 ISSUE
LAB ID and/or LABORATORY NAME: ASSESSOR NAME: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting. Method Number: ELAP method code 7037
LAB ID and/or LABORATORY NAME: ASSESSOR NAME: DATE: CONTINUOUS RADON MONITOR (CRM) by Proportional Counting Method Number: ELAP method code 7037 SOP Number: Revision Number: SOP Date: Personnel / Data
rtech laboratories began working toward ISO 9002:1994 in ISO 17025: Practical Benefits of Implementing a Quality System STATISTICAL ANALYSIS
1038 HONSA & MCINTYRE : JOURNAL OF AOAC INTERNATIONAL VOL. 86, NO. 5, 2003 STATISTICAL ANALYSIS ISO 17025: Practical Benefits of Implementing a Quality System JULIE D. HONSA and DEBORAH A. MCINTYRE rtech
18.6.1 Terms concerned with internal quality control procedures
18.6.1 Terms concerned with internal quality control procedures Quality assurance in analytical laboratories Quality assurance is the essential organisational infrastructure that underlies all reliable
Gravimetric determination of pipette errors
Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with
Fitness For Purpose Methods. Dan Schmitz Director Analytical Research and Services Abbott Nutrition
Fitness For Purpose Methods Dan Schmitz Director Analytical Research and Services Abbott Nutrition Agenda Describe AOAC process for Fitness for Purpose Review method validation principles and common protocols
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine
Procedure for Writing Technical Procedures
Procedure for Writing Technical Procedures 1.0 Purpose This procedure specifies when a technical procedure is needed and outlines the required elements for writing the technical procedure. 2.0 Scope This
Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods
Page 1 of 6 Analytical Methods: A Statistical Perspective on the ICH Q2A and Q2B Guidelines for Validation of Analytical Methods Dec 1, 2006 By: Steven Walfish BioPharm International ABSTRACT Vagueness
OPERATIONAL STANDARD
1 of 11 1. Introduction The International Safe Transit Association (ISTA), a non-profit association whose objective is to prevent product damage and excess packaging usage within the distribution environment.
Analytical Procedures and Method Validation
IPPC Discharges Monitoring Workshop Analytical Procedures and Method Validation Peter Webster Regional Chemist (EPA Cork) Contents... Overview of data quality Selection of suitable Analytical Methods Choosing
ISO/IEC 17025 QUALITY MANUAL
1800 NW 169 th Pl, Beaverton, OR 97006 Revision F Date: 9/18/06 PAGE 1 OF 18 TABLE OF CONTENTS Quality Manual Section Applicable ISO/IEC 17025:2005 clause(s) Page Quality Policy 4.2.2 3 Introduction 4
QbD Approach to Assay Development and Method Validation
QbD Approach to Assay Development and Method Validation Thomas A. Little Ph.D. 11/105/2014 President Thomas A. Little Consulting 12401 N Wildflower Lane Highland, UT 84003 1-925-285-1847 [email protected]
and Drug Administration/Office of Regulatory Affairs (FDA/ORA) laboratories.
Page 1 of 13 Sections Included in this Document and Change History 1. Purpose 2. Scope 3. Responsibilities 4. Background 5. References 6. Procedure/(6.C.4. added or electronically ) 7. Definitions 8. Records
EffiValidation 3.0 software basic training module
EffiValidation 3.0 software basic training module Oct 3, 2012 www.effichem.com Agenda EffiChem company EffiValidation 3.0 software : overview International standards and guidelines Software features Structure
Sample Analysis Design Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors:
Step 2 Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples Step 2 Calibration/Standard
Quality Management System Policy Manual
Quality Management System Burns Engineering, Inc. 10201 Bren Road East Minnetonka, MN 55343 4. 4.1 General Requirements The Quality Manual consists of the quality system policy documents and the quality
NIST HANDBOOK 150 CHECKLIST
NIST HANDBOOK 150 CHECKLIST Instructions to the Assessor: This checklist addresses the general accreditation criteria prescribed in NIST Handbook 150, NVLAP Procedures and General Requirements (2006 edition).
There is a need for a coordinating role by the Commission supported by the Framework programme, to facilitate and harmonise those developments.
0(752/2*
Measuring equipment in a laboratory.
Measuring equipment in a laboratory. Laboratory and metrology are two words that practically do not function separately. They are integrally connected to measuring equipment for tests and measurements.
STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION
Directive from the Scientific Director Page 1 of 16 STANDARD OPERATING GUIDELINES: EVIDENTIAL BREATH ALCOHOL INSTRUMENT CALIBRATION 1 Scope To describe the procedure for breath alcohol instrument calibration
G104 - Guide for Estimation of Measurement Uncertainty In Testing. December 2014
Page 1 of 31 G104 - Guide for Estimation of Measurement Uncertainty In Testing December 2014 2014 by A2LA All rights reserved. No part of this document may be reproduced in any form or by any means without
Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland
NUKLEONIKA 2010;55(2):149 154 ORIGINAL PAPER Proficiency testing schemes on determination of radioactivity in food and environmental samples organized by the NAEA, Poland Halina Polkowska-Motrenko, Leon
Proficiency Testing In A Global Analytical Environment Benefits & Challenges. AOAC September 2007
Proficiency Testing In A Global Analytical Environment Benefits & Challenges AOAC September 2007 1 The Coca-Cola Company The Coca-Cola Company is the largest manufacturer, distributor and marketer of nonalcoholic
CALIBRATION PRINCIPLES
1 CALIBRATION PRINCIPLES After completing this chapter, you should be able to: Define key terms relating to calibration and interpret the meaning of each. Understand traceability requirements and how they
Process Validation: Practical Aspects of the New FDA Guidance
Process Validation: Practical Aspects of the New FDA Guidance ISPE Boston Chapter Meeting April 18, 2013 Rusty Morrison Commissioning Agents, Inc. Objectives / Summary What is Process Validation? Regulatory
Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension
International Journal of PharmTech Research CODEN (USA): IJPRIF ISSN : 0974-4304 Vol.2, No.2, pp 1634-1638, April-June 2010 Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination
AOAC Guidelines for Single Laboratory Validation of Chemical Methods for Dietary Supplements and Botanicals
AOAC Guidelines for Single Laboratory Validation of Chemical Methods for Dietary Supplements and Botanicals CONTENTS 1.0 Introduction 1.1 Definitions 1.1.1 Validation 1.1.2 Method of analysis 1.1.3 Performance
Fast, Reproducible LC-MS/MS Analysis of Dextromethorphan and Dextrorphan
Fast, Reproducible LC-MS/MS Analysis of and Kimberly Phipps, Thermo Fisher Scientific, Runcorn, Cheshire, UK Application Note 685 Key Words Accucore C18, dextromethorphan, dextrorphan, SOLA CX Abstract
The Fitness for Purpose of Analytical Methods
The Fitness for Purpose of Analytical Methods EURACHEM Guide The Fitness for Purpose of Analytical Methods A Laboratory Guide to Method Validation and Related Topics This document has been developed by
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.
Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY I. BACKGROUND (VERSION 1.0) Atomic absorption spectroscopy (AAS) is a widely used
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests
Content Sheet 7-1: Overview of Quality Control for Quantitative Tests Role in quality management system Quality Control (QC) is a component of process control, and is a major element of the quality management
Cleaning Validation in Active pharmaceutical Ingredient manufacturing plants
Cleaning Validation in Active pharmaceutical Ingredient manufacturing plants September 1999 Table of contents 1. Foreword...... 2 2. Objective...... 3 3. Scope........ 4 4. Potential residues... 5 5. Current
UNCERTAINTY AND CONFIDENCE IN MEASUREMENT
Ελληνικό Στατιστικό Ινστιτούτο Πρακτικά 8 ου Πανελληνίου Συνεδρίου Στατιστικής (2005) σελ.44-449 UNCERTAINTY AND CONFIDENCE IN MEASUREMENT Ioannis Kechagioglou Dept. of Applied Informatics, University