Case study Realistic and best uncertainties with humidity instruments
|
|
- Earl Hampton
- 8 years ago
- Views:
Transcription
1 Case study Realistic and best uncertainties with humidity instruments Domen Hudoklin September 2014 Brdo, Slovenia
2 Humidity calibrations Dewpoint primary - against dewpoint generator secondary - against precision dew-point sensor (chilled mirror hygrometer) Relative humidity reference value: dewpoint + temperature reference RH probe Other (mass fraction - upper air, )
3 Traceability scheme LMK case Other NMI pressure meter LMK primary dew-point generator PGR-LMK-01 LT bath LMK primary dew-point generator PGR-LMK-02 HT bath precision resistance bridge precision ref.resistor thermometer SPRT LMK primary realisation of temperature -50 C to 20 C 0,028 C to 0,049 C 20 C to 95 C 0,05 C precision dew-point sensors humidity generator IPRT thermometers -10 C to 70 C 0,03 C -10 C to 70 C 0,12 C unit under cal. (dew-point sensor) unit under cal. 10% to 95% 0,5 % to 1,6 % (rel. humidity sensor)
4 CMC - Calibration and Measurement Capability a CMC is a calibration and measurement capability available to customers under normal conditions: as published in the BIPM key comparison database (KCDB) of the CIPM MRA; or as described in the laboratory s scope of accreditation granted by a signatory to the ILAC Arrangement * BIPM/RMO-ILAC/RAB Working Party, The CMC-BMC Issue: Developments since the Nashville statement, 2007 (
5 BIPM KCDB CMCs - dewpoint
6 BIPM KCDB CMCs - dewpoint
7 Combining different uncertainties In RH calibrations, if the reference is dew-point sensor, then the temperature and dew-point contributions have to be considered separately. u u t t 2 u t d t d 2
8 Temperature dependence of relative humidity D / % t / C / %
9 BIPM KCDB CMCs relative humidity at 95%rh, 98%rh
10 BIPM KCDB CMCs relative humidity at 10%rh (and 0.5%rh)
11 Humidity calibrations Sources of uncertainties: generation of humidity uncertainty of reference value uncertainty of calibration method UUT correlations (generation UUT, )
12 4. UUT Unit under test temporal variation hysteresis resolution curve-fitting loading effect drift (to be added by the end user) reproducibility the influence of UUT to calibration environment (self-heat, humidification psychrometer) uncertainty of electrical devices (bridge, resistor,...)
13 Expanded uncertainty contribution of the UUT LMK case for period of 1 month
Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK
code Location code Customers Sites locations: 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK Calibration Centre Bolkiah Garrison BB3510 Negara Brunei Darussalam Contact: Mr Lim Tiong Thai Tel: +673-2-386475
More informationCRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES
CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES Prepared by: Technical Manager Approved by: Chief Executive Officer Approval Date: 2013-04-24 Effective
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Old Bracknell Lane West Bracknell Berkshire RG12 7AH Contact: Mr M Trotter Tel: +44 (0)1344 459314 Fax: +44 (0)1344 465556 E-Mail: mtrotter@bis.fm
More informationRense Product Range. Customized Humidity Solutions. Rense Product Range
Customized Humidity Solutions Rense Product Range Background to Michell Instruments Michell Instruments is the international leader in the field of moisture and humidity measurement solutions. With over
More informationRelative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration
Relative Humidity Measurement and Calibration The widest range of humidity products from a pioneer of RH sensor development The best in Relative Humidity technology to optimize your process Whether you
More informationThe BIPM key comparison database
The BIPM key comparison database Dr Claudine Thomas BIPM KCDB Coordinator The BIPM key comparison database The KCDB is a public-access website http://www.bipm.org/kcdb built up upon a number of databases
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Unit 7 Contact: Mr S C Sparks Solent Industrial Estate Tel: +44 (0)1489 790296 Hedge End Fax: +44 (0)1489 790294 Southampton E-Mail: info@southcal.co.uk
More informationAccreditation according to ISO/IEC 17025 in Europe. Historic EA Europeing Accreditation Assessment Experience from Technical assessment
Accreditation according to ISO/IEC 17025 in Europe Historic EA Europeing Accreditation Assessment Experience from Technical assessment 1 In year 1803 the foots in Europe had a big differences, depending
More informationIAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS
REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and
More informationGMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards
March 2003 GMP 11 Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards 1. Introduction 1.1. Purpose Measurement processes are dynamic systems and often
More informationEnvironment-Laboratory Ambient Conditions
ISSUE 3.0: MADE MINOR CLARIFICATION CHANGES. FOR CLIENT LABS Purpose To assure laboratory conditions are stable during the performance of testing and data collection. Why this requirement is important?
More informationUser's Guide. Digital Psychrometer. Model RH390. F C WBT Dew Point MAX MIN
User's Guide Digital Psychrometer Model RH390 MAX MIN F C WBT Dew Point Introduction Congratulations on your purchase of the Extech RH390 Digital Psychrometer. This device measures Humidity, Air Temperature,
More informationCalibration Certificate
P.O. Box 87-0087 Phone:856-686-1600 Fax:856-686-1601 www.troemner.com e-mail: troemner@troemner.com SECTION 1: Page 1 of 7 Pages SECTION 2: APPROVED SIGNATORY Joseph Moran SECTION 3: PERSON PERFORMING
More informationDM Series CONTENTS. 1. Features 3. 2. Applications 3. 3. General 4. 4. Calibration 5. 5. Dimensions 6. 6. Specifications 7. 7. Calibration procedure 7
DM Series CONTENTS 1. Features 3 2. Applications 3 3. General 4 4. Calibration 5 DM-509-(M/T/S)X-(01/02) & DM-509-(H3/H5)-(01/02) 5. Dimensions 6 6. Specifications 7 7. Calibration procedure 7 The DM Series
More information16. Industrial Platinum Resistance Thermometers
129 16. Industrial Platinum Resistance Thermometers Industrial platinum resistance thermometers (IPRTs) are constructed from platinum of a lower quality than is used for SPRTs; typically, values of W(H
More informationRelative Humidity Calibration Kit
Revised 6.8.12 Relative Humidity Calibration Kit For Calibrating All RH Measuring Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500 2011 Nielsen-Kellerman Co. While the calibration
More informationHow to calibrate an RTD or Platinum Resistance Thermometer (PRT)
How to calibrate an RTD or Platinum Resistance Thermometer (PRT) Application Note Introduction There are two types of calibrations applicable to PRTs characterization and tolerance testing. The type of
More informationThe Calibration Chain: Role of BIPM,
The Calibration Chain: Role of BIPM, PSDLs and ADCLs J. Seuntjens and M. McEwen McGill University & National Research Council CANADA Clinical Dosimetry for Radiotherapy, 2009 AAPM Summer School PSDL ADCL
More informationProfessional Thermohygrometer
User Manual Professional Thermohygrometer with NIST-Traceable Calibration Model 20250-21 THE STANDARD IN PRECISION MEASUREMENT Introduction The Digi-Sense Professional Thermohygrometer (Model 20250-21)
More informationOMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT
OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification
More informationUncertainty of Force Measurements
European Association of National Metrology Institutes Uncertainty of Force Measurements EURAMET cg-4 Version.0 (03/011) Previously EA-10/04 Calibration Guide EURAMET cg-4 Version.0 (03/011) UNCERTAINTY
More informationKALIBRERINGSBEVIS. 25 temperature loggers Ecoguard model TS-020000, identification se table of results.
utfärdat av riksmätplats 01 issued by a Swedish National Laboratory Contact person Lars-Erik Josefson 2014-03-27 1 (3) MTv4F006969-K01 Measurement Technology +46 10 516 54 56 larserik.josefson@sp.se EcoGuard
More informationRECOMMENDATION. on the Evaluation of Quality Management System of National Metrology Institutes
EURO-ASIAN COOPERATION OF NATIONAL METROLOGICAL INSTITUTIONS COOMET ЕВРО-АЗИАТСКОЕ СОТРУДНИЧЕСТВО ГОСУДАРСТВЕННЫХ МЕТРОЛОГИЧЕСКИХ УЧРЕЖДЕНИЙ КООМЕТ Approved on the COOMET Quality Forum 25 th February 2002
More informationElectricity and Magnetism, Georgia, GEOSTM (Georgian National Agency for Standards and Metrology)
single references 1 1.018 V Temperature (23 ± 3) C 6 µv 2 95% No 1 single references 10 10 V Temperature (23 ± 3) C 25 µv 2 95% No 2 low ranges DMM > 0.01 10 V Temperature (23 ± 3) C 2E-06U + 3E-06 to
More informationDew-Point Measurement Solutions
Dew-Point Measurement Solutions www.michell.com Dew-Point Measurement Solutions Special Features No downtime Michell Instruments offers a unique sensor exchange programme, which means that you will never
More informationCalibration. How can you get lower costs for the calibration? What is traceability? What is measurement uncertainty? What is calibration?
Calibration Tvinnargatan 25 SE-507 30 Brämhult SWEDEN Phone: +46 33 22 56 30 Fax: +46 33 13 88 71 info@kalibrera.se www.kalibrera.se In May 2011 Elastocon purchased the company Mätcentrum in Eskilstuna,
More informationISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES
ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES AUTHORITY: VICE PRESIDENT EFFECTIVE DATE: 2015/01/01 DOCUMENT NUMBER: MA 2002 TABLE OF CONTENTS Foreword... 3 1. Application... 4 2.
More informationForm and content of COOMET сertificate for reference materials for composition and properties of substances and materials
COOMET Recommendation Form and content of COOMET сertificate for reference materials for composition and properties of substances and materials CООМЕТ R/RM/22:2013 Agreed at the 17 th meeting of COOMET
More informationBasic Steps to Establish ISO/IEC 17025 Compliant Testing Laboratories
Workshop for Caribbean countries to promote the development and implementation of Conformity Assessment programmes St. Augustine (Trinidad and Tobago) 2-4 December Basic Steps to Establish ISO/IEC 17025
More informationExperiences with novel developed secondary conductivity sensors within the German Calibration Service (DKD)
Experiences with novel developed secondary conductivity sensors within the German Calibration Service (DKD) Ulrich Breuel 1, Barbara Werner 2, Petra Spitzer 3, Hans D. Jensen 4 1,2 Zentrum für Messen und
More informationcalibrate confidence with Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Abstract Introduction White Paper
calibrate with confidence White Paper Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Authored by Keith Bennett Director of Metrology Howard Zion Director of Technical Operations TRANSCAT
More informationRULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES
RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES Reviewed by: Natalija Jovičić Zarić MSc, Quality Manager Approved by: Milica Lukešević, Acting Director Any
More informationHow do I measure the amount of water vapor in the air?
How do I measure the amount of water vapor in the air? Materials 2 Centigrade Thermometers Gauze Fan Rubber Band Tape Overview Water vapor is a very important gas in the atmosphere and can influence many
More informationTemperature Calibration; Depths of Immersion
Temperature Calibration; epths of Immersion Author: John P. Tavener Company Isothermal Technology Limited, Pine Grove, Southport, England Abstract Of all the sources of errors and uncertainties in thermal
More informationProcon Engineering. Technical Document PELR 1002. TERMS and DEFINITIONS
Procon Engineering Technical Document PELR 1002 TERMS and DEFINITIONS The following terms are widely used in the weighing industry. Informal comment on terms is in italics and is not part of the formal
More informationAmerican Association for Laboratory Accreditation
Page 1 of 18 Traceability for Life Sciences Testing and (October 2015) 2015 by A2LA. All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written
More informationInfrared Thermometer Calibration A Complete Guide
Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate
More informationA beginner s guide to humidity measurement
A beginner s guide to humidity measurement Stephanie Bell Version for comment October 2011 NPL Management Ltd Internal NPL measurement guides This is one of a series of NPL guides about measurement. These
More informationThe Radon Calibration Laboratory at the Federal Office for Radiation Protection (BfS)
The Radon Calibration Laboratory at the Federal Office for Radiation Protection (BfS) T.R. Beck, H. Buchröder, J. Döring, E. Foerster, V. Schmidt Bundesamt für Strahlenschutz (Federal Office for Radiation
More informationQuality Assurance (QA) & Quality Control (QC)
Quality Assurance (QA) & Quality Control (QC) In the laboratory Georges Ruta Water Quality Scientist City West Water 1 City West Water Water Supply & Sewerage Authority Melbourne Pop. >800,000 Area 580
More informationTraceable Moisture Measurements by Means of Residual Water Detection
1/24 Traceable Moisture Measurements by Means of Residual Water Detection Peter Østergaard Danish Technological Institute peo@dti.dk 2/24 What you will learn from this presentation Why SI traceability
More informationA PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS
Presented 2006 May 16-18 2006 ISFFM Queretaro, Mexico A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS P. Delajoud, M. Bair, C. Rombouts, M. Girard DH Instruments,
More informationISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters as Reference Standards.
MICRO MOTION WHITE PAPER BY M. LEE, D. STANDIFORD, MICRO MOTION, INC. A. PRUYSEN, EMERSON PROCESS MANAGEMENT ISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters
More informationFDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)
FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs) Barriers to Overcoming Interferences and Limitations Alan T. Cariski, MD, JD, and Mike Flis, AdvaMed BGM
More informationEASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL
EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration
More informationLARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS
XVII IMEKO World Congress Metrology in the 3rd Millennium June 22 27, 2003, Dubrovnik, Croatia LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS Tanfer Yandayan TUBITAK-Ulusal
More information3M CORPORATE METROLOGY SERVICES St. Paul, MN
A2LA has accredited THE AMERICAN ASSOCIATION FOR LABORATORY ACCREDITATION ACCREDITED LABORATORY 3M CORPORATE METROLOGY SERVICES St. Paul, MN for technical competence in the field of Calibration This laboratory
More informationA TRACEABLE CALIBRATION PROCEDURE FOR HYDROCARBON DEW POINT METERS
23rd World Gas Conference, Amsterdam 2006 A TRACEABLE CALIBRATION PROCEDURE FOR HYDROCARBON DEW POINT METERS Main author H.J. Panneman N.V. Nederlandse Gasunie Gasunie Engineering & Technology Groningen,
More informationTemperature and Humidity
Temperature and Humidity Overview Water vapor is a very important gas in the atmosphere and can influence many things like condensation and the formation of clouds and rain, as well as how hot or cold
More informationBalance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance
INL/CON-10-18643 PREPRINT Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance 2010 NCSL International Workshop and Symposium Michael Stears July 2010 This is
More informationSimplifying Humidity Measurement with Single-Chip Relative Humidity and Temperature Sensor ICs
Simplifying Humidity Measurement with Single-Chip Relative Humidity and Temperature Sensor ICs Introduction Humidity sensing technology is critical to a wide range of applications, including HVAC and refrigeration,
More informationHigh Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors
High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors Tom Wise on behalf of Daya Bay collaboration Day Bay is a reactor based experiment dedicated to measurement of the last
More informationCalibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.
Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known
More informationPrimary Standards for Humidity
Primary Standards for Humidity Andrea Peruzzi NMi van Swinden Laboratorium Humidity is everywhere You step out of your morning shower, you look at the mirror of your bathroom and its surface is covered
More informationUNIT 6a TEST REVIEW. 1. A weather instrument is shown below.
UNIT 6a TEST REVIEW 1. A weather instrument is shown below. Which weather variable is measured by this instrument? 1) wind speed 3) cloud cover 2) precipitation 4) air pressure 2. Which weather station
More informationExperiment 6 ~ Joule Heating of a Resistor
Experiment 6 ~ Joule Heating of a Resistor Introduction: The power P absorbed in an electrical resistor of resistance R, current I, and voltage V is given by P = I 2 R = V 2 /R = VI. Despite the fact that
More informationName Organization Country e-mail Jovan Bojkovski LMK Slovenia jovan.bojkovski@fe.uni-lj.si
EVITHERM WP8 MEETING 26 March 2003 France, Trappes, LNE Attendees present : Name Organization Country e-mail Thierry Vigneron CAP France thierry.vigneron@chauvin-arnoux.com Eliane Renaot CNAM-INM France
More informationRS DIGITAL THERMOMETER
RS DIGITAL THERMOMETER 1319A INSTRUCTION MANUAL C F MX MN AVG TITLE TABLE OF CONTENTS PAGE 1. INTRODUCTION... 1 2. SPECIFICATIONS... 1 3. NAME OF PARTS AND POSITIONS... 3 4. OPERATION INSTRUCTIONS... 5
More informationCalibration Management Software by CyberMetrics. 1-800-777-7020 +1 (480) 922-7300 www.cybermetrics.com sales@cybermetrics.com
Calibration Management Software by CyberMetrics 1-800-777-7020 +1 (480) 922-7300 www.cybermetrics.com sales@cybermetrics.com Product: SHT1x, SHT2x, SHT3x, SHT7x Series Digital Humidity and Temperature
More informationHygro-Thermometer + InfraRed Thermometer Model RH101
User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative
More informationENERGY STAR Program Requirements Product Specification for Data Center Storage. Draft 2 Test Method March 2013
ENERGY STAR Program Requirements Product Specification for Data Center Storage Draft 2 Test Method March 2013 1 2 3 4 1 OVERVIEW The following test method shall be used for determining compliance with
More informationAmerican Association for Laboratory Accreditation
Page 1 of 15 R104 GENERAL REQUIREMENTS: ACCREDITATION OF FIELD TESTING AND FIELD CALIBRATION LABORATORIES July 2015 2015 by A2LA All rights reserved. No part of this document may be reproduced in any form
More informationNIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING
NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING Instructions to the Assessor: This checklist addresses specific accreditation criteria prescribed in NIST Handbook 150-5, Construction Materials
More informationInternational Laboratory Accreditation Cooperation. Laboratory Accreditation or ISO 9001 Certification? global trust. Testing Calibration Inspection
International Laboratory Accreditation Cooperation Laboratory Accreditation or ISO 9001 Certification? Laboratory Accreditation or ISO 9001 Certification? When selecting a supplier to fulfill your testing,
More informationHTM2500LF Temperature and Relative Humidity Module
Hermetic Housing Humidity calibrated within +/-2% @55%RH Temperature measurement through NTC 10kOhms +/-1% direct output Small size product Typical 1 to 4 Volt DC output for 0 to 100%RH at 5Vdc DESCRIPTION
More informationGeneral and statistical principles for certification of RM ISO Guide 35 and Guide 34
General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in
More informationCCM.M-K3.1 comparison / 50 kg Mass
CCM.M-K3.1 comparison / 50 kg Mass The 7 th of April 2010 G. MANN - T. MADEC - P.A. MEURY Laboratoire Commun de Métrologie LNE-CNAM Abstract This comparison is a subsequent to the CCM.M-K3 comparison performed
More informationDISSOLVED OXYGEN METER Model : DO-5509
DISSOLVED OXYGEN METER Model : DO-5509 Your purchase of this DISSOLVED OXYGEN METER marks a step forward for you into the field of precision measurement. Although this DISSOLVED OXYGEN METER is a complex
More informationMoisture Control. It s The Dew Point. Stupid! Its not the humidity.
Moisture Control Its not the humidity. It s The Dew Point Stupid! Mike Schell EpiphanyTec Inc. Santa Barbara, CA mschell@epiphanytec.com 805 687-3175 Matching great technology to market need! Topics SHR
More informationDetermination of Moisture Content
Determination of Moisture Content 1. Introduction: Drying, storage, marketing, and roasting are four important aspects of coffee handling in which moisture plays an important role: Drying: moisture measurement
More informationS Davidson, M Perkin, M Buckley
No. 71 Measurement Good Practice Guide The Measurement of Mass and Weight S Davidson, M Perkin, M Buckley The National Physical Laboratory is operated on behalf of the DTI by NPL Management Limited, a
More informationINSTRUCTION MANUAL DTH31. 1-800-547-5740 Fax: (503) 643-6322 www.ueitest.com email: info@ueitest.com
INSTRUCTION MANUAL DTH31 Digital Psychrometer 1-800-547-5740 Fax: (503) 643-6322 www.ueitest.com email: info@ueitest.com Introduction The DTH31 Digital Psychrometer is an advanced handheld humidity and
More informationInstallation, Operation and Performance Qualification of a Thermoguard Installation
Installation, Operation and Performance Qualification of a Thermoguard Installation Thermoguard Qualification in Accordance with GMP To operate Thermoguard hardware and software within a regulated environment
More informationBy Authority Of THE UNITED STATES OF AMERICA Legally Binding Document
By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has
More informationA Guide for Documenting the Management System for a Testing/Calibration Lab. May 2015
A Guide for Documenting the Management System for a Testing/Calibration Lab May 2015 1. Introduction The following guide is to help testing/calibration laboratories create documentation to satisfy the
More informationTorque Key Comparison CCM.T-K1.3
Torque Key Comparison CCM.T-K1.3 Measurand Torque: 0 N m, 500 N m, 1000 N m Final Report March 9, 2015 Pilot Institute: Contact Person: E-mail: Co-author: Korea Research Institute of Standards and Science,
More informationConductivity measurement on pure water according to the recommendations of the USP Pharmacopoeia USP24-NF19
Conductivity measurement on pure water according to the recommendations of the USP Pharmacopoeia USP24-NF19 By Patricia DALMAS, Product Specialist, Radiometer Analytical Pharmaceutical laboratories working
More informationCALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001)
CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001) I. Introduction Calibration experiments or procedures are fairly common in laboratory work which involves any type of instrumentation. Calibration
More informationROLE OF THE APEC SPECIALIST REGIONAL BODIES
ROLE OF THE APEC SPECIALIST REGIONAL BODIES ND A S ARD URE D N STA TRUCT E H F T FRAS O S NT NCE IN E M ELE FORMA CON Why is the Standards and Conformance Infrastructure important? ~ Improves everyday
More informationUser checks and maintenance of laboratory balances
Technical Note 13 March 2014 Issued: July 1995 Amended and reissued: July 2005, July 2007, October 2007, May 2010, March 2013, March 2014 User checks and maintenance of laboratory balances Copyright National
More informationCOMBIMASS. Technical Data COMBIMASS eco-bio +
COMBIMASS Technical Data THE SYSTEM COMBIMASS The field transmitters of the COMBIMASS eco series are suitable for gas flow measurement and cover a wide range of different applications. The instruments
More informationHumidity and Dew-Point Calibration
Humidity and Dew-Point Calibration Instruments, Systems & Services Michell Instruments has 40 years' experience providing moisture measurement & calibration solutions to our customers Contents A Background
More informationEQUIPMENT CALIBRATION/MAINTENANCE
EQUIPMENT CALIBRATION/MAINTENANCE The following chart provides a guide for calibration of standard equipment used in the Fish Health Laboratory. All Intervals suggested are subject to individual manufacturer
More informationGravimetric determination of pipette errors
Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with
More informationMETHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS
METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS Contents Foreword --------------------------------------------------------------------------------------------------------------------
More informationSchedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK
code Coalville 6 Way North Cheshire Trading Estate Wirral CH43 3DU Contact: Mr D Williams Tel: +44 (0)151-649 9955 Fax: +44 (0)151-649 9977 E-Mail: sales@paragon-sci.com Website: www.paragon-sci.com Laboratory
More informationAdvances in Thermal Dispersion Mass Flow Meter Accuracy
Advances in Thermal Dispersion Mass Flow Meter Accuracy By Dan McQueen President Fluid Components International (FCI) Advances in Thermal Dispersion Mass Flow Meter Accuracy By Dan McQueen President Fluid
More informationNI USB-5681 RF Power Meter Specifications
NI USB-568 RF Power Meter Specifications General This document lists specifications for the NI USB-568 RF power meter. Minimum or maximum specifications are warranted under the following conditions: hour
More informationExperimental Design 7.2.2013 Doc. Martti Heinonen
Introduction to Metrology: SI unit system and measurement standards, traceability, calibration and measurement uncertainty Experimental Design 7.2.2013 Doc. Martti Heinonen Mars Climate Orbiter 2 Impact
More informationHM1500LF Relative Humidity Module
Small size product Product free from Lead, Cr(6+), Cd and Hg Humidity calibrated within +/-2% @55%RH Typical 1 to 3.6 Volt DC output for 0 to 100% RH at 5Vdc supply Ratiometric to voltage supply from 4.75Vdc
More informationTECHNICAL NOTE: SI 01 SPECIFIC REQUIREMENTS FOR THE ACCREDITATION OF INSPECTION BODIES FOR SITE INVESTIGATION
ACCREDITATION SCHEME FOR INSPECTION BODIES TECHNICAL NOTE: SI 01 SPECIFIC REQUIREMENTS FOR THE ACCREDITATION OF INSPECTION BODIES FOR SITE INVESTIGATION Technical Note SI 01: 3 February 2016 The SAC Accreditation
More informationAmerican Association for Laboratory Accreditation
Page 1 of 5 INTRODUCTION ISO/IEC 17011 Conformity Assessment General requirements for accreditation bodies accrediting conformity assessment bodies, the standard by which A2LA operates in order to be recognized
More informationGUIDANCE ILAC-G24 INTERNATIONAL OIML D 10 SERIES Edition 2007 (E) DOCUMENT Edition 2007 (E)
GUIDANCE ILAC-G24 INTERNATIONAL OIML D 10 SERIES Edition 2007 (E) DOCUMENT Edition 2007 (E) Guidelines for the determination of calibration intervals of measuring instruments Guide pour la détermination
More informationNRC Publications Archive Archives des publications du CNRC
NRC Publications Archive Archives des publications du CNRC Digital 3D Imaging and Modeling: A Metrological Approach Beraldin, Jean-Angelo Publisher s version / Version de l'éditeur: Time Compression Technologies
More informationUser Guide. Temperature and Humidity Datalogger. Model 42280
User Guide Temperature and Humidity Datalogger Model 42280 Introduction Congratulations on your purchase of the Extech 42280 Thermometer and Relative Humidity Datalogger. The 42280 is a wall-mount, tripod
More informationR221 - Specific Requirements: Forensic
R221 - Specific Requirements: Forensic Page 1 of 6 R221 - Specific Requirements: Forensic (October 2015) 2015 by A2LA. All rights reserved. No part of this document may be reproduced in any form or by
More informationACL 800 MEGOHMMETER OPERATION MANUAL
ACL 800 MEGOHMMETER OPERATION MANUAL Meter is warranted for one year from the date of purchase on parts and labor. Calibration is recommended every twelve months. NOTE: THIS MANUAL IS NOT TO BE SEPARATED
More informationTotal Heat Versus Sensible Heat Evaporator Selection Methods & Application
Total Heat Versus Sensible Heat Evaporator Selection Methods & Application Scope The purpose of this paper is to provide specifying engineers, purchasers and users of evaporators in industrial refrigeration
More informationFigure 1. A typical Laboratory Thermometer graduated in C.
SIGNIFICANT FIGURES, EXPONENTS, AND SCIENTIFIC NOTATION 2004, 1990 by David A. Katz. All rights reserved. Permission for classroom use as long as the original copyright is included. 1. SIGNIFICANT FIGURES
More informationSentinel I28 and Blackbelt Mass Flow Calibration and Verification
Service Bulletin: #001 Date: August 2012 Sentinel I28 and Blackbelt Mass Flow Calibration and Verification Summary This Application Bulletin will give a short synopsis of the purpose of a calibration and
More information