Case study Realistic and best uncertainties with humidity instruments

Size: px
Start display at page:

Download "Case study Realistic and best uncertainties with humidity instruments"

Transcription

1 Case study Realistic and best uncertainties with humidity instruments Domen Hudoklin September 2014 Brdo, Slovenia

2 Humidity calibrations Dewpoint primary - against dewpoint generator secondary - against precision dew-point sensor (chilled mirror hygrometer) Relative humidity reference value: dewpoint + temperature reference RH probe Other (mass fraction - upper air, )

3 Traceability scheme LMK case Other NMI pressure meter LMK primary dew-point generator PGR-LMK-01 LT bath LMK primary dew-point generator PGR-LMK-02 HT bath precision resistance bridge precision ref.resistor thermometer SPRT LMK primary realisation of temperature -50 C to 20 C 0,028 C to 0,049 C 20 C to 95 C 0,05 C precision dew-point sensors humidity generator IPRT thermometers -10 C to 70 C 0,03 C -10 C to 70 C 0,12 C unit under cal. (dew-point sensor) unit under cal. 10% to 95% 0,5 % to 1,6 % (rel. humidity sensor)

4 CMC - Calibration and Measurement Capability a CMC is a calibration and measurement capability available to customers under normal conditions: as published in the BIPM key comparison database (KCDB) of the CIPM MRA; or as described in the laboratory s scope of accreditation granted by a signatory to the ILAC Arrangement * BIPM/RMO-ILAC/RAB Working Party, The CMC-BMC Issue: Developments since the Nashville statement, 2007 (

5 BIPM KCDB CMCs - dewpoint

6 BIPM KCDB CMCs - dewpoint

7 Combining different uncertainties In RH calibrations, if the reference is dew-point sensor, then the temperature and dew-point contributions have to be considered separately. u u t t 2 u t d t d 2

8 Temperature dependence of relative humidity D / % t / C / %

9 BIPM KCDB CMCs relative humidity at 95%rh, 98%rh

10 BIPM KCDB CMCs relative humidity at 10%rh (and 0.5%rh)

11 Humidity calibrations Sources of uncertainties: generation of humidity uncertainty of reference value uncertainty of calibration method UUT correlations (generation UUT, )

12 4. UUT Unit under test temporal variation hysteresis resolution curve-fitting loading effect drift (to be added by the end user) reproducibility the influence of UUT to calibration environment (self-heat, humidification psychrometer) uncertainty of electrical devices (bridge, resistor,...)

13 Expanded uncertainty contribution of the UUT LMK case for period of 1 month

Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK code Location code Customers Sites locations: 21-47 High Street, Feltham, Middlesex, TW13 4UN, UK Calibration Centre Bolkiah Garrison BB3510 Negara Brunei Darussalam Contact: Mr Lim Tiong Thai Tel: +673-2-386475

More information

CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES

CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES CRITERIA FOR PERFORMING CALIBRATION AND INTERMEDIATE CHECKS ON EQUIPMENT USED IN ACCREDITED FACILITIES Prepared by: Technical Manager Approved by: Chief Executive Officer Approval Date: 2013-04-24 Effective

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Old Bracknell Lane West Bracknell Berkshire RG12 7AH Contact: Mr M Trotter Tel: +44 (0)1344 459314 Fax: +44 (0)1344 465556 E-Mail: mtrotter@bis.fm

More information

Rense Product Range. Customized Humidity Solutions. Rense Product Range

Rense Product Range. Customized Humidity Solutions. Rense Product Range Customized Humidity Solutions Rense Product Range Background to Michell Instruments Michell Instruments is the international leader in the field of moisture and humidity measurement solutions. With over

More information

Relative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration

Relative Humidity Sensors, Transmitters and Calibration Products Relative Humidity Measurement and Calibration Relative Humidity Measurement and Calibration The widest range of humidity products from a pioneer of RH sensor development The best in Relative Humidity technology to optimize your process Whether you

More information

The BIPM key comparison database

The BIPM key comparison database The BIPM key comparison database Dr Claudine Thomas BIPM KCDB Coordinator The BIPM key comparison database The KCDB is a public-access website http://www.bipm.org/kcdb built up upon a number of databases

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK Unit 7 Contact: Mr S C Sparks Solent Industrial Estate Tel: +44 (0)1489 790296 Hedge End Fax: +44 (0)1489 790294 Southampton E-Mail: info@southcal.co.uk

More information

Accreditation according to ISO/IEC 17025 in Europe. Historic EA Europeing Accreditation Assessment Experience from Technical assessment

Accreditation according to ISO/IEC 17025 in Europe. Historic EA Europeing Accreditation Assessment Experience from Technical assessment Accreditation according to ISO/IEC 17025 in Europe Historic EA Europeing Accreditation Assessment Experience from Technical assessment 1 In year 1803 the foots in Europe had a big differences, depending

More information

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS REFERENCES NIST Special Publication 330 IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS Revised October 2013 International vocabulary of metrology Basic and general concepts and

More information

GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards

GMP 11. Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards March 2003 GMP 11 Good Measurement Practice for Assignment and Adjustment of Calibration Intervals for Laboratory Standards 1. Introduction 1.1. Purpose Measurement processes are dynamic systems and often

More information

Environment-Laboratory Ambient Conditions

Environment-Laboratory Ambient Conditions ISSUE 3.0: MADE MINOR CLARIFICATION CHANGES. FOR CLIENT LABS Purpose To assure laboratory conditions are stable during the performance of testing and data collection. Why this requirement is important?

More information

User's Guide. Digital Psychrometer. Model RH390. F C WBT Dew Point MAX MIN

User's Guide. Digital Psychrometer. Model RH390. F C WBT Dew Point MAX MIN User's Guide Digital Psychrometer Model RH390 MAX MIN F C WBT Dew Point Introduction Congratulations on your purchase of the Extech RH390 Digital Psychrometer. This device measures Humidity, Air Temperature,

More information

Calibration Certificate

Calibration Certificate P.O. Box 87-0087 Phone:856-686-1600 Fax:856-686-1601 www.troemner.com e-mail: troemner@troemner.com SECTION 1: Page 1 of 7 Pages SECTION 2: APPROVED SIGNATORY Joseph Moran SECTION 3: PERSON PERFORMING

More information

DM Series CONTENTS. 1. Features 3. 2. Applications 3. 3. General 4. 4. Calibration 5. 5. Dimensions 6. 6. Specifications 7. 7. Calibration procedure 7

DM Series CONTENTS. 1. Features 3. 2. Applications 3. 3. General 4. 4. Calibration 5. 5. Dimensions 6. 6. Specifications 7. 7. Calibration procedure 7 DM Series CONTENTS 1. Features 3 2. Applications 3 3. General 4 4. Calibration 5 DM-509-(M/T/S)X-(01/02) & DM-509-(H3/H5)-(01/02) 5. Dimensions 6 6. Specifications 7 7. Calibration procedure 7 The DM Series

More information

16. Industrial Platinum Resistance Thermometers

16. Industrial Platinum Resistance Thermometers 129 16. Industrial Platinum Resistance Thermometers Industrial platinum resistance thermometers (IPRTs) are constructed from platinum of a lower quality than is used for SPRTs; typically, values of W(H

More information

Relative Humidity Calibration Kit

Relative Humidity Calibration Kit Revised 6.8.12 Relative Humidity Calibration Kit For Calibrating All RH Measuring Kestrel Meters Model Numbers 3000, 3500, 4000, 4200, 4250, 4300, 4400, 4500 2011 Nielsen-Kellerman Co. While the calibration

More information

How to calibrate an RTD or Platinum Resistance Thermometer (PRT)

How to calibrate an RTD or Platinum Resistance Thermometer (PRT) How to calibrate an RTD or Platinum Resistance Thermometer (PRT) Application Note Introduction There are two types of calibrations applicable to PRTs characterization and tolerance testing. The type of

More information

The Calibration Chain: Role of BIPM,

The Calibration Chain: Role of BIPM, The Calibration Chain: Role of BIPM, PSDLs and ADCLs J. Seuntjens and M. McEwen McGill University & National Research Council CANADA Clinical Dosimetry for Radiotherapy, 2009 AAPM Summer School PSDL ADCL

More information

Professional Thermohygrometer

Professional Thermohygrometer User Manual Professional Thermohygrometer with NIST-Traceable Calibration Model 20250-21 THE STANDARD IN PRECISION MEASUREMENT Introduction The Digi-Sense Professional Thermohygrometer (Model 20250-21)

More information

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT

OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT OMCL Network of the Council of Europe QUALITY MANAGEMENT DOCUMENT PA/PH/OMCL (12) 77 7R QUALIFICATION OF EQUIPMENT ANNEX 8: QUALIFICATION OF BALANCES Full document title and reference Document type Qualification

More information

Uncertainty of Force Measurements

Uncertainty of Force Measurements European Association of National Metrology Institutes Uncertainty of Force Measurements EURAMET cg-4 Version.0 (03/011) Previously EA-10/04 Calibration Guide EURAMET cg-4 Version.0 (03/011) UNCERTAINTY

More information

KALIBRERINGSBEVIS. 25 temperature loggers Ecoguard model TS-020000, identification se table of results.

KALIBRERINGSBEVIS. 25 temperature loggers Ecoguard model TS-020000, identification se table of results. utfärdat av riksmätplats 01 issued by a Swedish National Laboratory Contact person Lars-Erik Josefson 2014-03-27 1 (3) MTv4F006969-K01 Measurement Technology +46 10 516 54 56 larserik.josefson@sp.se EcoGuard

More information

RECOMMENDATION. on the Evaluation of Quality Management System of National Metrology Institutes

RECOMMENDATION. on the Evaluation of Quality Management System of National Metrology Institutes EURO-ASIAN COOPERATION OF NATIONAL METROLOGICAL INSTITUTIONS COOMET ЕВРО-АЗИАТСКОЕ СОТРУДНИЧЕСТВО ГОСУДАРСТВЕННЫХ МЕТРОЛОГИЧЕСКИХ УЧРЕЖДЕНИЙ КООМЕТ Approved on the COOMET Quality Forum 25 th February 2002

More information

Electricity and Magnetism, Georgia, GEOSTM (Georgian National Agency for Standards and Metrology)

Electricity and Magnetism, Georgia, GEOSTM (Georgian National Agency for Standards and Metrology) single references 1 1.018 V Temperature (23 ± 3) C 6 µv 2 95% No 1 single references 10 10 V Temperature (23 ± 3) C 25 µv 2 95% No 2 low ranges DMM > 0.01 10 V Temperature (23 ± 3) C 2E-06U + 3E-06 to

More information

Dew-Point Measurement Solutions

Dew-Point Measurement Solutions Dew-Point Measurement Solutions www.michell.com Dew-Point Measurement Solutions Special Features No downtime Michell Instruments offers a unique sensor exchange programme, which means that you will never

More information

Calibration. How can you get lower costs for the calibration? What is traceability? What is measurement uncertainty? What is calibration?

Calibration. How can you get lower costs for the calibration? What is traceability? What is measurement uncertainty? What is calibration? Calibration Tvinnargatan 25 SE-507 30 Brämhult SWEDEN Phone: +46 33 22 56 30 Fax: +46 33 13 88 71 info@kalibrera.se www.kalibrera.se In May 2011 Elastocon purchased the company Mätcentrum in Eskilstuna,

More information

ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES

ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES ISO/IEC 17025 ACCREDITATION REQUIREMENTS FOR CALIBRATION LABORATORIES AUTHORITY: VICE PRESIDENT EFFECTIVE DATE: 2015/01/01 DOCUMENT NUMBER: MA 2002 TABLE OF CONTENTS Foreword... 3 1. Application... 4 2.

More information

Form and content of COOMET сertificate for reference materials for composition and properties of substances and materials

Form and content of COOMET сertificate for reference materials for composition and properties of substances and materials COOMET Recommendation Form and content of COOMET сertificate for reference materials for composition and properties of substances and materials CООМЕТ R/RM/22:2013 Agreed at the 17 th meeting of COOMET

More information

Basic Steps to Establish ISO/IEC 17025 Compliant Testing Laboratories

Basic Steps to Establish ISO/IEC 17025 Compliant Testing Laboratories Workshop for Caribbean countries to promote the development and implementation of Conformity Assessment programmes St. Augustine (Trinidad and Tobago) 2-4 December Basic Steps to Establish ISO/IEC 17025

More information

Experiences with novel developed secondary conductivity sensors within the German Calibration Service (DKD)

Experiences with novel developed secondary conductivity sensors within the German Calibration Service (DKD) Experiences with novel developed secondary conductivity sensors within the German Calibration Service (DKD) Ulrich Breuel 1, Barbara Werner 2, Petra Spitzer 3, Hans D. Jensen 4 1,2 Zentrum für Messen und

More information

calibrate confidence with Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Abstract Introduction White Paper

calibrate confidence with Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Abstract Introduction White Paper calibrate with confidence White Paper Metrology Concepts: Understanding Test Uncertainty Ratio (TUR) Authored by Keith Bennett Director of Metrology Howard Zion Director of Technical Operations TRANSCAT

More information

RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES

RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES RULES FOR THE PARTICIPATION IN INTER- LABORATORY COMPARISONS AND PROFICIENCY TESTING SCHEMES Reviewed by: Natalija Jovičić Zarić MSc, Quality Manager Approved by: Milica Lukešević, Acting Director Any

More information

How do I measure the amount of water vapor in the air?

How do I measure the amount of water vapor in the air? How do I measure the amount of water vapor in the air? Materials 2 Centigrade Thermometers Gauze Fan Rubber Band Tape Overview Water vapor is a very important gas in the atmosphere and can influence many

More information

Temperature Calibration; Depths of Immersion

Temperature Calibration; Depths of Immersion Temperature Calibration; epths of Immersion Author: John P. Tavener Company Isothermal Technology Limited, Pine Grove, Southport, England Abstract Of all the sources of errors and uncertainties in thermal

More information

Procon Engineering. Technical Document PELR 1002. TERMS and DEFINITIONS

Procon Engineering. Technical Document PELR 1002. TERMS and DEFINITIONS Procon Engineering Technical Document PELR 1002 TERMS and DEFINITIONS The following terms are widely used in the weighing industry. Informal comment on terms is in italics and is not part of the formal

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation Page 1 of 18 Traceability for Life Sciences Testing and (October 2015) 2015 by A2LA. All rights reserved. No part of this document may be reproduced in any form or by any means without the prior written

More information

Infrared Thermometer Calibration A Complete Guide

Infrared Thermometer Calibration A Complete Guide Infrared Thermometer Calibration A Complete Guide Application note With proper setup and planning, infrared calibrations can be accurate. The steps outlined below should be followed to perform accurate

More information

A beginner s guide to humidity measurement

A beginner s guide to humidity measurement A beginner s guide to humidity measurement Stephanie Bell Version for comment October 2011 NPL Management Ltd Internal NPL measurement guides This is one of a series of NPL guides about measurement. These

More information

The Radon Calibration Laboratory at the Federal Office for Radiation Protection (BfS)

The Radon Calibration Laboratory at the Federal Office for Radiation Protection (BfS) The Radon Calibration Laboratory at the Federal Office for Radiation Protection (BfS) T.R. Beck, H. Buchröder, J. Döring, E. Foerster, V. Schmidt Bundesamt für Strahlenschutz (Federal Office for Radiation

More information

Quality Assurance (QA) & Quality Control (QC)

Quality Assurance (QA) & Quality Control (QC) Quality Assurance (QA) & Quality Control (QC) In the laboratory Georges Ruta Water Quality Scientist City West Water 1 City West Water Water Supply & Sewerage Authority Melbourne Pop. >800,000 Area 580

More information

Traceable Moisture Measurements by Means of Residual Water Detection

Traceable Moisture Measurements by Means of Residual Water Detection 1/24 Traceable Moisture Measurements by Means of Residual Water Detection Peter Østergaard Danish Technological Institute peo@dti.dk 2/24 What you will learn from this presentation Why SI traceability

More information

A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS

A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS Presented 2006 May 16-18 2006 ISFFM Queretaro, Mexico A PRIMARY CALIBRATION SYSTEM FOR THE SUPPORT OF HIGH PERFORMANCE GAS FLOW TRANSFER STANDARDS P. Delajoud, M. Bair, C. Rombouts, M. Girard DH Instruments,

More information

ISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters as Reference Standards.

ISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters as Reference Standards. MICRO MOTION WHITE PAPER BY M. LEE, D. STANDIFORD, MICRO MOTION, INC. A. PRUYSEN, EMERSON PROCESS MANAGEMENT ISO 17025 Accreditation for a Standing Start Finish (SSF) Primary Flow Stand with Coriolis Flowmeters

More information

FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)

FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs) FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs) Barriers to Overcoming Interferences and Limitations Alan T. Cariski, MD, JD, and Mike Flis, AdvaMed BGM

More information

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL Issue February 2004 2 TABLE OF CONTENTS SECTION PAGE 1. INTRODUCTION 3 1.1 General 3 1.2 Ceramic Sensing Element 3 1.3 Calibration

More information

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS

LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS XVII IMEKO World Congress Metrology in the 3rd Millennium June 22 27, 2003, Dubrovnik, Croatia LARGE MEASUREMENT RANGE MECHANICAL COMPARATOR FOR CALIBRATION OF LONG GAUGE BLOCKS Tanfer Yandayan TUBITAK-Ulusal

More information

3M CORPORATE METROLOGY SERVICES St. Paul, MN

3M CORPORATE METROLOGY SERVICES St. Paul, MN A2LA has accredited THE AMERICAN ASSOCIATION FOR LABORATORY ACCREDITATION ACCREDITED LABORATORY 3M CORPORATE METROLOGY SERVICES St. Paul, MN for technical competence in the field of Calibration This laboratory

More information

A TRACEABLE CALIBRATION PROCEDURE FOR HYDROCARBON DEW POINT METERS

A TRACEABLE CALIBRATION PROCEDURE FOR HYDROCARBON DEW POINT METERS 23rd World Gas Conference, Amsterdam 2006 A TRACEABLE CALIBRATION PROCEDURE FOR HYDROCARBON DEW POINT METERS Main author H.J. Panneman N.V. Nederlandse Gasunie Gasunie Engineering & Technology Groningen,

More information

Temperature and Humidity

Temperature and Humidity Temperature and Humidity Overview Water vapor is a very important gas in the atmosphere and can influence many things like condensation and the formation of clouds and rain, as well as how hot or cold

More information

Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance

Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance INL/CON-10-18643 PREPRINT Balance Calibration A Method for Assigning a Direct-Reading Uncertainty to an Electronic Balance 2010 NCSL International Workshop and Symposium Michael Stears July 2010 This is

More information

Simplifying Humidity Measurement with Single-Chip Relative Humidity and Temperature Sensor ICs

Simplifying Humidity Measurement with Single-Chip Relative Humidity and Temperature Sensor ICs Simplifying Humidity Measurement with Single-Chip Relative Humidity and Temperature Sensor ICs Introduction Humidity sensing technology is critical to a wide range of applications, including HVAC and refrigeration,

More information

High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors

High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors High Precision Measurement of the Target Mass of the Daya Bay Neutrino Detectors Tom Wise on behalf of Daya Bay collaboration Day Bay is a reactor based experiment dedicated to measurement of the last

More information

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc.

Calibration & Preventative Maintenance. Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration & Preventative Maintenance Sally Wolfgang Manager, Quality Operations Merck & Co., Inc. Calibration: A comparison of two instruments or measuring devices one of which is a standard of known

More information

Primary Standards for Humidity

Primary Standards for Humidity Primary Standards for Humidity Andrea Peruzzi NMi van Swinden Laboratorium Humidity is everywhere You step out of your morning shower, you look at the mirror of your bathroom and its surface is covered

More information

UNIT 6a TEST REVIEW. 1. A weather instrument is shown below.

UNIT 6a TEST REVIEW. 1. A weather instrument is shown below. UNIT 6a TEST REVIEW 1. A weather instrument is shown below. Which weather variable is measured by this instrument? 1) wind speed 3) cloud cover 2) precipitation 4) air pressure 2. Which weather station

More information

Experiment 6 ~ Joule Heating of a Resistor

Experiment 6 ~ Joule Heating of a Resistor Experiment 6 ~ Joule Heating of a Resistor Introduction: The power P absorbed in an electrical resistor of resistance R, current I, and voltage V is given by P = I 2 R = V 2 /R = VI. Despite the fact that

More information

Name Organization Country e-mail Jovan Bojkovski LMK Slovenia jovan.bojkovski@fe.uni-lj.si

Name Organization Country e-mail Jovan Bojkovski LMK Slovenia jovan.bojkovski@fe.uni-lj.si EVITHERM WP8 MEETING 26 March 2003 France, Trappes, LNE Attendees present : Name Organization Country e-mail Thierry Vigneron CAP France thierry.vigneron@chauvin-arnoux.com Eliane Renaot CNAM-INM France

More information

RS DIGITAL THERMOMETER

RS DIGITAL THERMOMETER RS DIGITAL THERMOMETER 1319A INSTRUCTION MANUAL C F MX MN AVG TITLE TABLE OF CONTENTS PAGE 1. INTRODUCTION... 1 2. SPECIFICATIONS... 1 3. NAME OF PARTS AND POSITIONS... 3 4. OPERATION INSTRUCTIONS... 5

More information

Calibration Management Software by CyberMetrics. 1-800-777-7020 +1 (480) 922-7300 www.cybermetrics.com sales@cybermetrics.com

Calibration Management Software by CyberMetrics. 1-800-777-7020 +1 (480) 922-7300 www.cybermetrics.com sales@cybermetrics.com Calibration Management Software by CyberMetrics 1-800-777-7020 +1 (480) 922-7300 www.cybermetrics.com sales@cybermetrics.com Product: SHT1x, SHT2x, SHT3x, SHT7x Series Digital Humidity and Temperature

More information

Hygro-Thermometer + InfraRed Thermometer Model RH101

Hygro-Thermometer + InfraRed Thermometer Model RH101 User's Guide Hygro-Thermometer + InfraRed Thermometer Model RH101 Introduction Congratulations on your purchase of the Extech Hygro-Thermometer plus InfraRed Thermometer. This device measures relative

More information

ENERGY STAR Program Requirements Product Specification for Data Center Storage. Draft 2 Test Method March 2013

ENERGY STAR Program Requirements Product Specification for Data Center Storage. Draft 2 Test Method March 2013 ENERGY STAR Program Requirements Product Specification for Data Center Storage Draft 2 Test Method March 2013 1 2 3 4 1 OVERVIEW The following test method shall be used for determining compliance with

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation Page 1 of 15 R104 GENERAL REQUIREMENTS: ACCREDITATION OF FIELD TESTING AND FIELD CALIBRATION LABORATORIES July 2015 2015 by A2LA All rights reserved. No part of this document may be reproduced in any form

More information

NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING

NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING NIST HANDBOOK 150-5 CHECKLIST CONSTRUCTION MATERIALS TESTING Instructions to the Assessor: This checklist addresses specific accreditation criteria prescribed in NIST Handbook 150-5, Construction Materials

More information

International Laboratory Accreditation Cooperation. Laboratory Accreditation or ISO 9001 Certification? global trust. Testing Calibration Inspection

International Laboratory Accreditation Cooperation. Laboratory Accreditation or ISO 9001 Certification? global trust. Testing Calibration Inspection International Laboratory Accreditation Cooperation Laboratory Accreditation or ISO 9001 Certification? Laboratory Accreditation or ISO 9001 Certification? When selecting a supplier to fulfill your testing,

More information

HTM2500LF Temperature and Relative Humidity Module

HTM2500LF Temperature and Relative Humidity Module Hermetic Housing Humidity calibrated within +/-2% @55%RH Temperature measurement through NTC 10kOhms +/-1% direct output Small size product Typical 1 to 4 Volt DC output for 0 to 100%RH at 5Vdc DESCRIPTION

More information

General and statistical principles for certification of RM ISO Guide 35 and Guide 34

General and statistical principles for certification of RM ISO Guide 35 and Guide 34 General and statistical principles for certification of RM ISO Guide 35 and Guide 34 / REDELAC International Seminar on RM / PT 17 November 2010 Dan Tholen,, M.S. Topics Role of reference materials in

More information

CCM.M-K3.1 comparison / 50 kg Mass

CCM.M-K3.1 comparison / 50 kg Mass CCM.M-K3.1 comparison / 50 kg Mass The 7 th of April 2010 G. MANN - T. MADEC - P.A. MEURY Laboratoire Commun de Métrologie LNE-CNAM Abstract This comparison is a subsequent to the CCM.M-K3 comparison performed

More information

DISSOLVED OXYGEN METER Model : DO-5509

DISSOLVED OXYGEN METER Model : DO-5509 DISSOLVED OXYGEN METER Model : DO-5509 Your purchase of this DISSOLVED OXYGEN METER marks a step forward for you into the field of precision measurement. Although this DISSOLVED OXYGEN METER is a complex

More information

Moisture Control. It s The Dew Point. Stupid! Its not the humidity.

Moisture Control. It s The Dew Point. Stupid! Its not the humidity. Moisture Control Its not the humidity. It s The Dew Point Stupid! Mike Schell EpiphanyTec Inc. Santa Barbara, CA mschell@epiphanytec.com 805 687-3175 Matching great technology to market need! Topics SHR

More information

Determination of Moisture Content

Determination of Moisture Content Determination of Moisture Content 1. Introduction: Drying, storage, marketing, and roasting are four important aspects of coffee handling in which moisture plays an important role: Drying: moisture measurement

More information

S Davidson, M Perkin, M Buckley

S Davidson, M Perkin, M Buckley No. 71 Measurement Good Practice Guide The Measurement of Mass and Weight S Davidson, M Perkin, M Buckley The National Physical Laboratory is operated on behalf of the DTI by NPL Management Limited, a

More information

INSTRUCTION MANUAL DTH31. 1-800-547-5740 Fax: (503) 643-6322 www.ueitest.com email: info@ueitest.com

INSTRUCTION MANUAL DTH31. 1-800-547-5740 Fax: (503) 643-6322 www.ueitest.com email: info@ueitest.com INSTRUCTION MANUAL DTH31 Digital Psychrometer 1-800-547-5740 Fax: (503) 643-6322 www.ueitest.com email: info@ueitest.com Introduction The DTH31 Digital Psychrometer is an advanced handheld humidity and

More information

Installation, Operation and Performance Qualification of a Thermoguard Installation

Installation, Operation and Performance Qualification of a Thermoguard Installation Installation, Operation and Performance Qualification of a Thermoguard Installation Thermoguard Qualification in Accordance with GMP To operate Thermoguard hardware and software within a regulated environment

More information

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document

By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By Authority Of THE UNITED STATES OF AMERICA Legally Binding Document By the Authority Vested By Part 5 of the United States Code 552(a) and Part 1 of the Code of Regulations 51 the attached document has

More information

A Guide for Documenting the Management System for a Testing/Calibration Lab. May 2015

A Guide for Documenting the Management System for a Testing/Calibration Lab. May 2015 A Guide for Documenting the Management System for a Testing/Calibration Lab May 2015 1. Introduction The following guide is to help testing/calibration laboratories create documentation to satisfy the

More information

Torque Key Comparison CCM.T-K1.3

Torque Key Comparison CCM.T-K1.3 Torque Key Comparison CCM.T-K1.3 Measurand Torque: 0 N m, 500 N m, 1000 N m Final Report March 9, 2015 Pilot Institute: Contact Person: E-mail: Co-author: Korea Research Institute of Standards and Science,

More information

Conductivity measurement on pure water according to the recommendations of the USP Pharmacopoeia USP24-NF19

Conductivity measurement on pure water according to the recommendations of the USP Pharmacopoeia USP24-NF19 Conductivity measurement on pure water according to the recommendations of the USP Pharmacopoeia USP24-NF19 By Patricia DALMAS, Product Specialist, Radiometer Analytical Pharmaceutical laboratories working

More information

CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001)

CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001) CALIBRATION OF A THERMISTOR THERMOMETER (version = fall 2001) I. Introduction Calibration experiments or procedures are fairly common in laboratory work which involves any type of instrumentation. Calibration

More information

ROLE OF THE APEC SPECIALIST REGIONAL BODIES

ROLE OF THE APEC SPECIALIST REGIONAL BODIES ROLE OF THE APEC SPECIALIST REGIONAL BODIES ND A S ARD URE D N STA TRUCT E H F T FRAS O S NT NCE IN E M ELE FORMA CON Why is the Standards and Conformance Infrastructure important? ~ Improves everyday

More information

User checks and maintenance of laboratory balances

User checks and maintenance of laboratory balances Technical Note 13 March 2014 Issued: July 1995 Amended and reissued: July 2005, July 2007, October 2007, May 2010, March 2013, March 2014 User checks and maintenance of laboratory balances Copyright National

More information

COMBIMASS. Technical Data COMBIMASS eco-bio +

COMBIMASS. Technical Data COMBIMASS eco-bio + COMBIMASS Technical Data THE SYSTEM COMBIMASS The field transmitters of the COMBIMASS eco series are suitable for gas flow measurement and cover a wide range of different applications. The instruments

More information

Humidity and Dew-Point Calibration

Humidity and Dew-Point Calibration Humidity and Dew-Point Calibration Instruments, Systems & Services Michell Instruments has 40 years' experience providing moisture measurement & calibration solutions to our customers Contents A Background

More information

EQUIPMENT CALIBRATION/MAINTENANCE

EQUIPMENT CALIBRATION/MAINTENANCE EQUIPMENT CALIBRATION/MAINTENANCE The following chart provides a guide for calibration of standard equipment used in the Fish Health Laboratory. All Intervals suggested are subject to individual manufacturer

More information

Gravimetric determination of pipette errors

Gravimetric determination of pipette errors Gravimetric determination of pipette errors In chemical measurements (for instance in titrimetric analysis) it is very important to precisely measure amount of liquid, the measurement is performed with

More information

METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS

METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS METHODS FOR THE CALIBRATION OF ELECTROSTATIC MEASURING INSTRUMENTS Contents Foreword --------------------------------------------------------------------------------------------------------------------

More information

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK code Coalville 6 Way North Cheshire Trading Estate Wirral CH43 3DU Contact: Mr D Williams Tel: +44 (0)151-649 9955 Fax: +44 (0)151-649 9977 E-Mail: sales@paragon-sci.com Website: www.paragon-sci.com Laboratory

More information

Advances in Thermal Dispersion Mass Flow Meter Accuracy

Advances in Thermal Dispersion Mass Flow Meter Accuracy Advances in Thermal Dispersion Mass Flow Meter Accuracy By Dan McQueen President Fluid Components International (FCI) Advances in Thermal Dispersion Mass Flow Meter Accuracy By Dan McQueen President Fluid

More information

NI USB-5681 RF Power Meter Specifications

NI USB-5681 RF Power Meter Specifications NI USB-568 RF Power Meter Specifications General This document lists specifications for the NI USB-568 RF power meter. Minimum or maximum specifications are warranted under the following conditions: hour

More information

Experimental Design 7.2.2013 Doc. Martti Heinonen

Experimental Design 7.2.2013 Doc. Martti Heinonen Introduction to Metrology: SI unit system and measurement standards, traceability, calibration and measurement uncertainty Experimental Design 7.2.2013 Doc. Martti Heinonen Mars Climate Orbiter 2 Impact

More information

HM1500LF Relative Humidity Module

HM1500LF Relative Humidity Module Small size product Product free from Lead, Cr(6+), Cd and Hg Humidity calibrated within +/-2% @55%RH Typical 1 to 3.6 Volt DC output for 0 to 100% RH at 5Vdc supply Ratiometric to voltage supply from 4.75Vdc

More information

TECHNICAL NOTE: SI 01 SPECIFIC REQUIREMENTS FOR THE ACCREDITATION OF INSPECTION BODIES FOR SITE INVESTIGATION

TECHNICAL NOTE: SI 01 SPECIFIC REQUIREMENTS FOR THE ACCREDITATION OF INSPECTION BODIES FOR SITE INVESTIGATION ACCREDITATION SCHEME FOR INSPECTION BODIES TECHNICAL NOTE: SI 01 SPECIFIC REQUIREMENTS FOR THE ACCREDITATION OF INSPECTION BODIES FOR SITE INVESTIGATION Technical Note SI 01: 3 February 2016 The SAC Accreditation

More information

American Association for Laboratory Accreditation

American Association for Laboratory Accreditation Page 1 of 5 INTRODUCTION ISO/IEC 17011 Conformity Assessment General requirements for accreditation bodies accrediting conformity assessment bodies, the standard by which A2LA operates in order to be recognized

More information

GUIDANCE ILAC-G24 INTERNATIONAL OIML D 10 SERIES Edition 2007 (E) DOCUMENT Edition 2007 (E)

GUIDANCE ILAC-G24 INTERNATIONAL OIML D 10 SERIES Edition 2007 (E) DOCUMENT Edition 2007 (E) GUIDANCE ILAC-G24 INTERNATIONAL OIML D 10 SERIES Edition 2007 (E) DOCUMENT Edition 2007 (E) Guidelines for the determination of calibration intervals of measuring instruments Guide pour la détermination

More information

NRC Publications Archive Archives des publications du CNRC

NRC Publications Archive Archives des publications du CNRC NRC Publications Archive Archives des publications du CNRC Digital 3D Imaging and Modeling: A Metrological Approach Beraldin, Jean-Angelo Publisher s version / Version de l'éditeur: Time Compression Technologies

More information

User Guide. Temperature and Humidity Datalogger. Model 42280

User Guide. Temperature and Humidity Datalogger. Model 42280 User Guide Temperature and Humidity Datalogger Model 42280 Introduction Congratulations on your purchase of the Extech 42280 Thermometer and Relative Humidity Datalogger. The 42280 is a wall-mount, tripod

More information

R221 - Specific Requirements: Forensic

R221 - Specific Requirements: Forensic R221 - Specific Requirements: Forensic Page 1 of 6 R221 - Specific Requirements: Forensic (October 2015) 2015 by A2LA. All rights reserved. No part of this document may be reproduced in any form or by

More information

ACL 800 MEGOHMMETER OPERATION MANUAL

ACL 800 MEGOHMMETER OPERATION MANUAL ACL 800 MEGOHMMETER OPERATION MANUAL Meter is warranted for one year from the date of purchase on parts and labor. Calibration is recommended every twelve months. NOTE: THIS MANUAL IS NOT TO BE SEPARATED

More information

Total Heat Versus Sensible Heat Evaporator Selection Methods & Application

Total Heat Versus Sensible Heat Evaporator Selection Methods & Application Total Heat Versus Sensible Heat Evaporator Selection Methods & Application Scope The purpose of this paper is to provide specifying engineers, purchasers and users of evaporators in industrial refrigeration

More information

Figure 1. A typical Laboratory Thermometer graduated in C.

Figure 1. A typical Laboratory Thermometer graduated in C. SIGNIFICANT FIGURES, EXPONENTS, AND SCIENTIFIC NOTATION 2004, 1990 by David A. Katz. All rights reserved. Permission for classroom use as long as the original copyright is included. 1. SIGNIFICANT FIGURES

More information

Sentinel I28 and Blackbelt Mass Flow Calibration and Verification

Sentinel I28 and Blackbelt Mass Flow Calibration and Verification Service Bulletin: #001 Date: August 2012 Sentinel I28 and Blackbelt Mass Flow Calibration and Verification Summary This Application Bulletin will give a short synopsis of the purpose of a calibration and

More information