Medium External Fixator Tibial Shaft Box Frame. For small-statured adults and pediatric patients, using outrigger posts.

Similar documents
Technique Guide. Screw Removal Set. Instruments for removing Synthes screws.

Battery Power Line. Battery-driven power tool system for orthopaedics and traumatology.

Orthopedic Foot Instruments. Dedicated instruments for reconstructive foot surgery.

Modular Sternal Cable System. Flexibility and strength in sternal closure and repair.

Technique Guide. Reamer/Irrigator/Aspirator (RIA). For intramedullary reaming and bone harvesting.

PERIPROSTHETIC IMPLANTS

LCP Compact Foot/Compact Hand.

Small Battery Drive. For orthopaedic trauma applications.

Technique Guide. Large Fragment LCP Instrument and Implant Set. Part of the Synthes locking compression plate (LCP) system.

TransFx External Fixation System Large and Intermediate Surgical Technique

MINI FRAGMENT SYSTEM. Instruments and implants for 1.5 mm, 2.0 mm, and 2.4 mm plate fixation PRODUCT OVERVIEW

Titanium Wire with Barb and Needle. For canthal tendon procedures.

Technique Guide. 4.5 mm LCP Proximal Femur Plates. Part of the Synthes Periarticular LCP Plating System.

Technique Guide. DHS/DCS Dynamic Hip and Condylar Screw System. Designed to provide stable internal fixation.

Scout Vessel Guard. A cover for vessels during anterior lumbar spine surgery.

MatrixNEURO. The next generation cranial plating system.

External Fixation Systems IMPORTANT MEDICAL INFORMATION

Technique Guide. Norian SRS Fast Set Putty. Calcium phosphate bone void filler.

Cable System. For Orthopaedic Trauma Surgery.

Technique Guide. Orthopaedic Cable System. Cerclage solutions for general surgery.

1.5 Tesla and 3 Tesla Magnetic Resonance Imaging (MRI) Guidelines for the Senza System

Technique Guide. 3.5 mm and 4.5 mm Curved Locking Compression Plates (LCP). For minimally invasive osteosynthesis.

MatrixNEURO CMF Matrix

LCP Superior Clavicle Plate. The anatomically precontoured fixation system with angular stability for clavicle shaft and lateral clavicle.

ImageReady MRI Full Body Guidelines for Precision Montage MRI Spinal Cord Stimulator System

LCP Olecranon Plate. The anatomical fixation system with angular stability for olecranon and proximal ulnar fractures.

LifeStent XL Biliary Stent System

Posterior Referencing. Surgical Technique

Aesculap Veterinary Orthopaedics. Targon VET Interlocking Nail

REVEAL LINQ LNQ11. Insertable Cardiac Monitor MRI procedural information. MRI Technical Manual

Anterior Lumbar Interbody Fusion (ALIF). Instrument set supports placement of ALIF spacers using anterior or anterolateral approach.

V-TEK IVP System 2.7 System 4.0

Technique Guide. Titanium Trochanteric Fixation Nail System. For intramedullary fixation of proximal femur fractures.

Zimmer Small Fragment Universal Locking System. Surgical Technique

Zimmer Periarticular Proximal Tibial Locking Plate

Expert HAN. Expert Hindfoot Arthrodesis Nail.

FDA Guidelines for Magnetic Resonance Equipment Safety

Instructions for Use. Device Description The AMPLATZER Vascular Plug II is a self-expandable nitinol mesh occlusion device (see Figure 1).

MR Information Date: March 11, 2014

NCB Distal Femur System. Surgical Technique

Expert LFN. Lateral Femoral Nail.

DHS/DCS System. Including LCP DHS and DHS Blade.

Surgical technique. End Cap for TEN. For axial stabilization and simultaneous protection of soft tissue.

Anatomic Percutaneous Ankle Reconstruction of Lateral Ligaments (A Percutaneous Anti ROLL)

OptiLock Periarticular Plating System For Proximal Tibial Fractures. Pre-Launch Surgical Technique

Subject: Magnetic Resonance Imaging (MRI) Information on Sorin Group Heart Valve Prostheses and Annuloplasty Devices

How To Fix A Radial Head Plate

Expert TN. Tibial Nail.

Technique Guide. 2.4 mm LCP Distal Radius System. A comprehensive plating system to address a variety of fracture patterns.

LISS PLT. The less invasive stabilization system for proximal tibia fractures.

Technique Guide. 2.7 mm/3.5 mm Variable Angle LCP Ankle Trauma System. Part of the Synthes variable angle locking compression plate (VA-LCP) system.

PSI Patient Specific Implants. Derived from CT data for excellent reconstructive results.

Expert ALFN. Adolescent Lateral Femoral Nail.

M1A/M14 5/8-24 thread adapter Installation Instructions

M A N U A L

Technique Guide. VersiTomic. Michael A. Rauh, MD. Anterior Cruciate Ligament Reconstruction

DRAFT. Triathlon TS Knee System. Surgical Protocol. Version 1

Guidance for implant removal. Straumann Dental Implant System

Achilles Tendon Repair, Operative Technique

The information contained in this document is intended for healthcare professionals only.

Scout Vessel Guard. A hydrogel membrane with exceptional handling characteristics.

TFN Titanium Trochanteric Fixation Nail System. For intramedullary fixation of proximal femur fractures.

ARCS Pre-Implantology

ACTIV ANKLE DISTAL AND DIAPHYSEAL FIBULA INNOVATION MEANS MOTION

Zimmer Natural Nail System. Cephalomedullary Nail Surgical Technique STANDARD

PATIENT SPECIFIC PLATES FOR MANDIBLE: CUSTOM MADE PLATES FOR TRAUMA AND RECONSTRUCTION

How To Use A Phoenix Retrograde Femoral Nail

AZEK Rail Install Guide

Lathe Milling Attachment

ZERO-P VA. Variable angle zero-profile anterior cervical interbody fusion (ACIF) device SURGICAL TECHNIQUE

TABLE OF CONTENTS. Surgical Technique 2. Indications 4. Product Information Patient Positioning and Approach 2

TriLock 1.5 Implants for the Phalanges

TwinFix Cannulated Compression Screw

VariAx Distal Radius Locking Plate System. Anatomical & Universal Volar Plates Dorsal Plates Fragment Specific Plates

esims Web Overview. Synthes Information Management Suite.

Lentur Cable System. Surgical Technique

HORIZONTAL INSTALLATION

Introduction to JIGS AND FIXTURES

Implant Extraction Guide Module One & Two

The information contained in this document is intended for healthcare professionals only.

C O N V E Y O R C O M P O N E N T S C H A I N S B E L T S B E A R I N G S

6 inch A-Arm Lift Kit WARNING: / installation instructions. will fit CLUB CAR DS. included:

PS-SEAL PTFE based high performance seal

Resharpening Companion

Elo Touch Solutions Wall-mounting Kit for the 5501L IDS Touchmonitors

Ecoflex Wall Penetration Systems: An efficient, sustainable solution for penetrating concrete or block walls

Stainless Steel Single and Dual Circulation Kits

Structural Holding System

6900PTFX bioprosthesis Carpentier-Edwards PERIMOUNT Magna mitral pericardial

Belt Tensioning Methods for Small Package Conveyors What s the best solution?

NCB Proximal Humerus System. Surgical Technique

Zimmer Periarticular Elbow Locking Plate System

Installation instruction CASAFLEX - Connecting piece DN 100

W I T H I N J O I ST S I N STA L L AT I ON

CLAW II. Polyaxial Compression Plating System SURGIC A L T ECHNIQUE

STEADYfast Stabilizer Installation Notes Fifth Wheel and Travel Trailers 11/23/13

NCB Periprosthetic Femur Plate System. Surgical Technique

Super Sounds Soccer. Model B8507

Working Drawing and Assemblies. Chapter 10

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

Transcription:

Medium External Fixator Tibial Shaft Box Frame. For small-statured adults and pediatric patients, using outrigger posts. Technique Guide Part of the Medium External Fixation System

MRI Information Synthes Medium External Fixation devices are labeled MR Conditional according to the terminology specified in ASTM F2503-05, Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment. Nonclinical testing demonstrated that, when used in the specific configurations stated in Synthes labeling, Synthes Medium External Fixation devices are MR Conditional. Representative Synthes Medium External Fixation devices used in a typical construct include clamps, rods and various attachments. A patient with a Synthes Medium External Fixation frame may be scanned safely after placement of the frame under the following conditions. Static magnetic field of 1.5 Tesla when the fixation frame is positioned: 7 cm or less from within the outside edge of the bore of the MRI at Normal Operating Mode or; Completely outside of the MRI bore in First Level Controlled Mode Static magnetic field of 3.0 Tesla when the fixation frame is positioned: 7 cm or less from within the outside edge of the bore of the MRI at Normal Operating Mode or; Completely outside of the MRI bore in First Level Controlled Mode Highest spatial gradient magnetic field of 900 Gauss/cm or less Maximum MR system reported whole body averaged specific absorption rate (SAR) of 2 W/kg for the Normal Operatinig Mode and 4 W/kg for the First Level Controlled Mode for 15 minutes of scanning Use only whole body RF transmit coil, no other transmit coils are allowed, local receive only coils are allowed. Note: In nonclinical testing, the Synthes external fixation frame was tested in several different configurations. This testing was conducted with the construct positioned 7 cm from within the outside edge of the MRI bore. The results showed a maximum observed heating for a wrist fixation frame of 6ºC for 1.5 T and less than 1ºC for 3.0 T with a machine reported whole body averaged SAR of 2 W/kg. Patients may be safely scanned in the MRI chamber at the above conditions. Under such conditions, the maximal expected temperature rise is less than 6ºC. Because higher in vivo heating cannot be excluded, close patient monitoring and communication with the patient during the scan is required. Immediately abort the scan if the patient reports burning sensation or pain. To minimize heating, the scan time should be as short as possible, the SAR as low as possible, and the device should be as far as possible from the edge of the bore. Temperature rise values obtained were based upon a scan time of 15 minutes. The above field conditions should be compared with those of the user s MR system, to determine if the item can safely be brought into the user s MR environment. If placed in the bore of the MR scanner during scanning, Synthes MR Conditional external fixation devices may have the potential to cause artifact in the diagnostic imaging. All components of Synthes external fixation frames must be identified as MR Conditional prior to being placed in or near an MR environment. Artifact information MR image quality may be compromised if the area of interest is in the same area or relatively close to the position of the Synthes Medium External Fixation construct, and it may be necessary to optimize MR imaging parameters, to compensate for the presence of the fixation frame. Representative devices used to assemble a typical Synthes Medium External Fixation frame have been evaluated in the MRI chamber and worst-case artifact information is provided below. Overall, artifacts created by Synthes Medium External Fixation devices may present issues if the MR imaging area of interest is in or near the area where the fixation frame is located. For FFE sequence: Scan duration: 3 min, TR 100 ms, TE 15 ms, flip angle 15º and SE sequence: Scan duration: 4 min, TR 500 ms, TE 20 ms, flip angle 70º radio echo sequence, worst-case artifact will extend approximately 10 cm from the device. Warning Do not place any radio frequency (RF) transmit coils over the external fixation frame. Synthes Medium External Fixator Tibial Shaft Box Frame Technique Guide

Medium External Fixator Tibial Shaft Box Frame When to use This frame is ideal for the fixation of tibia shaft fractures with open or closed soft tissue injury. The 6-position pin clamp allows full adjustability for fracture reduction. 1 2 Relevant anatomy Tibial Schanz screws should be placed in the AP plane (as shown in the illustrated frame) for maximum stability. Alternatively, they may be placed anteromedially to avoid drilling along the crest. For pediatric patients, use fluoroscopic imaging, placing the most distal and the most proximal Schanz screws at least 2 cm from the physes. While the physes are 4 mm 6 mm wide, they have undulating shapes. Therefore, to be in the safe zone and avoid injury to a growth plate, allow 1 cm for its total width. 1 Schanz screw is perpendicular to bone 20 140 90 Clamp is parallel to bone 30 Legend Optimal zones for Schanz screw insertion Safe zones for Schanz screw insertion 40 120 60 20 Tibial Safe Zones 2 1. J. Alonso and M. Horowitz. Use of the AO External Fixator in Children. Journal of Pediatric Orthopaedics. 1987;7. 594 600. 2. A. Fernández. External Fixation. AO Principles of Fracture Management. T. Rüedi and W. Murphy, ed. Dübendorf, Switzerland; AO Publishing. 2000. 239. Illustration modified and used with permission. Reference Behrens, F. and K. Searls. External Fixation of the Tibia. Journal of Bone and Joint Surgery. 1986;68-B. 246 254. Synthes 1

Recommended Components for Basic Frame Product Item Quantity Number Needed 294.78x 5.0 mm Self-Drilling 4 Schanz Screw 390.027 Medium Pin Clamp, 6 position 2 390.031 Medium Combination Clamp 2 390.029 30 Outrigger Post, 8 mm 2 395.7xx 8.0 mm Carbon Fiber Rod 1 394.993 Protective Cap, 4 for 5.0 mm Fixation Pins 395.781 Protective Cap, for 2 8.0 mm Carbon Fiber Rods 2 Synthes Medium External Fixator Tibial Shaft Box Frame Technique Guide

Technique Overview 1 Insert Schanz screws Use the 6-Position Drill Guide Handle (392.963) or pin clamp technique to ensure proper pin spacing. 2 2 Attach pin clamp Tighten the vise plates. 1 3 3 Attach outrigger posts Thread the posts into the vise plates to a hard stop. For angled posts, turn the post counterclockwise to the desired orientation. Lock in position by turning the lock nut clockwise until tight. 5 2 1 4 4 Attach carbon fiber rods Attach carbon fiber rods to the outrigger posts with combination clamps. 1 3 1 5 Reduce fracture Reduce the fracture and tighten all clamps. Note: To increase stiffness, add a second rod to the frame by repeating Steps 3 and 4 on the opposite side of the pin clamps. Synthes 3

Medium Pin Clamp Technique 1 Insert first Schanz screw Insert a Schanz screw through the drill sleeve and end position of the Medium Pin Clamp (390.026 or 390.027), using the clamp as an insertion guide. Note: The clamp should be parallel, and the Schanz screws perpendicular, to the bone. 2 Insert second Schanz screw Insert a second Schanz screw through the opposite end of the clamp. Tighten the vise plates. Pin Clamp Medium Pin Clamp, 6 position (390.027) Stargrind cover: remove to add outrigger posts Vise plate tightening point Note: Each side of the pin clamp can accept either an outrigger post or a rod attachment. Note: Additional Schanz screws may be inserted as needed. Outrigger Posts Lock nut Straight (390.028) Thread into pin clamp 30 Post (390.029) 90 Post (390.030) 4 Synthes Medium External Fixator Tibial Shaft Box Frame Technique Guide

Optional Frame Configurations Box frame with straight outrigger posts Box frame with 90 outrigger posts Frame with diagonal crossbar for additional stiffness Ankle spanning frame with 30 outrigger posts Synthes 5

Medium External Fixator Set with Self-Drilling Schanz Screws Stainless Steel (01.302.602) or Titanium (01.302.604) Graphic Case 690.450 Graphic Case, for Medium External Fixator Implants in Set 01.302.602 293.74 5.0 mm Steinmann Pin with Central Thread, 200 mm, 2 ea. Self-Drilling Schanz Screws, 4 ea 294.777 4.0 mm diameter, 125 mm. 294.778 4.0 mm diameter, 150 mm 294.785 5.0 mm diameter, 175 mm 294.786 5.0 mm diameter, 200 mm Implants in Set 01.302.604 293.74 5.0 mm Steinmann Pin with Central Thread, 200 mm, 2 ea. Titanium Self-Drilling Schanz Screws, 4 ea. 494.777 4.0 mm diameter, 125 mm 494.778 4.0 mm diameter, 150 mm 494.785 5.0 mm diameter, 175 mm 494.786 5.0 mm diameter, 200 mm Instruments (for both sets) 310.19 2.0 mm Drill Bit, quick coupling, 100 mm, 2 ea. 310.37 3.5 mm Drill Bit, quick coupling, 195 mm, 2 ea. 321.158 Combination Wrench, 8 mm width across flats 392.955 4.0 mm/2.5 mm Drill Sleeve 392.969 Combination T-Wrench, 8 mm 393.101 Drive Adaptor with quick coupling, for 4.0 mm Schanz Screws 393.103 Drive Adaptor with quick coupling, for 5.0 mm Schanz Screws 393.105 Small Universal Chuck with T-Handle 394.181 3.5 mm Trocar, short 394.182 3.5 mm Trocar, long 394.183 2.5 mm Trocar Note: For additional information, please refer to package insert. For detailed cleaning and sterilization instructions, please refer to http://us.synthes.com/medical+community/cleaning+and+sterilization.htm or to the below listed inserts, which will be included in the shipping container: Processing Synthes Reusable Medical Devices Instruments, Instrument Trays and Graphic Cases DJ1305 Processing Non-sterile Synthes Implants DJ1304 6 Synthes Medium External Fixator Tibial Shaft Box Frame Technique Guide

395.911 Drill Sleeve Handle 395.912 5.0 mm/ 3.5 mm Drill Sleeve, short 395.913 5.0 mm/3.5 mm Drill Sleeve, long 395.921 6.0 mm/ 5.0 mm Threaded Drill Sleeve, short 395.922 4.0 mm Threaded Drill Sleeve 395.923 6.0 mm/ 5.0 mm Threaded Drill Sleeve, long Fixation Material (for both sets) 390.031 Medium Combination Clamp, 8 ea. 390.032 Dynamization Clip, for Medium Combination Clamp, 4 ea. 390.033 Medium Multi-Pin Clamp, 4 position, 2 ea. 390.034 Rod Attachment, for Medium Multi-Pin Clamp, 4 ea. 390.035 Medium Open Adjustable Clamp, 4 ea. 390.036 Medium Multi-Pin Clamp, 6 position, 2 ea. 390.037 8.0 mm/11.0 mm Combination Clamp, 2 ea. 394.991 Protective Caps, for 4.0 mm Fixation Pins, 1 pkg. of 10 394.993 Protective Caps, for 5.0 mm Fixation Pins, 1 pkg. of 10 395.781 Protective Caps, for 8.0 mm Carbon Fiber Rods, 4 pkgs. of 2 8.0 mm Carbon Fiber Rods 395.779 160 mm, 2 ea. 395.782 200 mm 395.784 220 mm 395.786 240 mm, 2 ea. 395.788 280 mm, 2 ea. 395.792 320 mm, 2 ea. 395.796 360 mm, 2 ea. 395.797 400 mm Also Available Implants Schanz Screws 294.43.48 4.0 mm, spade point, 60 mm 150 mm 294.52.57 5.0 mm, blunted trocar point, 100 mm 250 mm 294.71.76 4.5 mm, blunted trocar point, 80 mm 200 mm Self-Drilling Schanz Screws 294.774.779 4.0 mm, 60 mm 175 mm 294.782.788 5.0 mm, 100 mm 250 mm Titanium Self-Drilling Schanz Screws 494.774.779 4.0 mm, 60 mm 175 mm 494.782.788 5.0 mm, 100 mm 250 mm Steinmann Pins with Central Thread 293.64 5.0 mm diameter, 150 mm 293.69 5.0 mm diameter, 175 mm Also Available Instruments 03.302.001 Medium Open Compressor 392.963 6-Position Drill Guide Handle Also Available Fixation Material 390.026 Medium Pin Clamp, 4 position 390.027 Medium Pin Clamp, 6 position 390.028 Straight Outrigger Post, 8 mm 390.029 30 Outrigger Post, 8 mm 390.030 90 Outrigger Post, 8 mm Also Available Sets 105.957 Power Drive Set 150.16 ComPact Air Drive II Set Also Available for Graphic Case 690.350.13 Label Sheet Pack, for Schanz Screws and Carbon Fiber Rods Synthes 7

Synthes (USA) 1302 Wrights Lane East West Chester, PA 19380 Telephone: (610) 719-5000 To order: (800) 523-0322 Fax: (610) 251-9056 Synthes (Canada) Ltd. 2566 Meadowpine Boulevard Mississauga, Ontario L5N 6P9 Telephone: (905) 567-0440 To order: (800) 668-1119 Fax: (905) 567-3185 www.synthes.com 2005 Synthes, Inc. or its affiliates. All rights reserved. Synthes is a trademark of Synthes, Inc. or its affiliates. Printed in U.S.A. 5/11 J5839-D