Total Organic Carbon Analysis for Purified Water and Water for Injection

Similar documents
ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)

Cleaning Validation by TOC Analyzer

HPLC Analysis of Acetaminophen Tablets with Waters Alliance and Agilent Supplies

Phoenix Apollo Laboratory TOC Analyzers and Services Designed To Meet Your Needs

How To Choose A Total Organic Carbon Analyzer

Simultaneous determination of L-ascorbic acid and D-iso-ascorbic acid (erythorbic acid) in wine by HPLC and UV-detection (Resolution Oeno 11/2008)

Reversed Phase High Presssure Liquid Chromatograhphic Technique for Determination of Sodium Alginate from Oral Suspension

EPA Method RSK 175 Dissolved Gasses in Water Matrices using the Teledyne Tekmar HT3 Headspace Analyzer. HT3 Application Note.

Elemental Scientific. Stability. Elemental Scientific

NANYANG TECHNOLOGICAL UNIVERSITY. School of Civil and Environmental Engineering. CV2701: Laboratory 2A. Laboratory Manual. For

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS

Determination of Caffeine in Beverages HPLC-1

Analytical Test Report

LACHAT METHOD NUMBER D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately g CN/L g CN/L. 09e

Determination of the Mass Percentage of Copper in a Penny. Introduction

INTERNATIONAL JOURNAL OF PHARMACEUTICAL RESEARCH AND BIO-SCIENCE

The Determination of Sugars in Molasses by High-Performance Anion Exchange with Pulsed Amperometric Detection

3.4 TEST FOR BACTERIAL ENDOTOXINS. Final text for revision of The International Pharmacopoeia

Standard Operating Procedure for Total Kjeldahl Nitrogen (Lachat Method)

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

EXPERIMENT 4. Determination of Sodium by Flame Atomic-Emission Spectroscopy

Determination of Calcium and Magnesium in Brine

C391-E058K TOC-V. Ser ies. Shimadzu Total Organic Carbon Analyzer

SIMULTANEOUS DETERMINATION OF TELMISARTAN AND HYDROCHLOROTHIAZIDE IN TABLET DOSAGE FORM USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Develop a Quantitative Analytical Method for low (» 1 ppm) levels of Sulfate

CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY

Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.

The Sievers 900 Series TOC Analyzers

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

The introduction of your report should be written on the on the topic of the role of indicators on acid base titrations.

Analysis of Total Organic Carbon

Transfer of a USP method for prednisolone from normal phase HPLC to SFC using the Agilent 1260 Infinity Hybrid SFC/UHPLC System Saving time and costs

How To Test For Contamination In Large Volume Water

Vitamin C quantification using reversed-phase ion-pairing HPLC

ATOMIC ABSORTION SPECTROSCOPY: rev. 4/2011 ANALYSIS OF COPPER IN FOOD AND VITAMINS

Solutions and Dilutions

Anion chromatography using on-line recycled eluents

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Development of validated RP- HPLC method for estimation of rivaroxaban in pharmaceutical formulation

EUROPEAN COMMISSION DIRECTORATE-GENERAL TAXATION AND CUSTOMS UNION TAX POLICY Excise duties and transport, environment and energy taxes

Determination of the Amount of Acid Neutralized by an Antacid Tablet Using Back Titration

Determination of Haloacetic Acids and Dalapon in Drinking Water by SPE and GC/ECD*

BRIEFING Plastic Packaging Systems for Pharmaceutical Use.

Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

High Performance Thin Layer Chromatographic Method for Estimation of Cefprozil in Tablet Dosage Form

Dissolved and precipitated oxalate

STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14

CORESTA RECOMMENDED METHOD Nº 80

Analysis of Free Bromate Ions in Tap Water using an ACQUITY UPLC BEH Amide Column

To determine the equivalence points of two titrations from plots of ph versus ml of titrant added.

Phosphorus Analyzer Procedure Last updated: A. Introduction

Enhanced Organic Precursor Removals Using Aged Filter Media Page 1. Enhanced Organic Precursor Removals Using Aged Filter Media

Acid Base Titrations

Analysis of Vitamin C Using Iodine. Introduction

Coating and Extraction of Honeycomb Denuders

IgM ELISA. For the quantitative determination of IgM in human serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

Online Total Organic Carbon (TOC)

Assay for Citrate and Phosphate in Pharmaceutical Formulations Using Ion Chromatography

Colorimetric Determination of Iron in Vitamin Tablets

Volumetric Analysis. Lecture 5 Experiment 9 in Beran page 109 Prelab = Page 115

Color, True and Apparent

Microbiological Testing of the Sawyer Mini Filter. 16 December Summary

α-cyclodextrin SYNONYMS α-schardinger dextrin, α-dextrin, cyclohexaamylose, cyclomaltohexaose, α- cycloamylase

Color, True and Apparent

Preparation of frequently used solutions

INTERNATIONAL OLIVE COUNCIL

NITRIC OXIDE and NITROGEN DIOXIDE 6014

Sensitive and Rapid Determination of Polycyclic Aromatic Hydrocarbons in Tap Water

SIMULTANEOUS DETERMINATION OF NALTREXONE AND 6- -NALTREXOL IN SERUM BY HPLC

ERDOSTEINE - MONOGRAPH.

Analysis of Various Vitamins in Multivitamin Tablets

DR/4000 PROCEDURE. CHLORINE, Free

USEPA 1 FerroVer Method 2 Method to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls

Simulation of the determination of lead azide content in waste water from explosives manufacture

ph and Acidity in Wine and Fruit Juice

Calibration of Glassware

SPIKE RECOVERY AND DETERMINING A METHOD DETECTION LIMIT Pamela Doolittle, University of Wisconsin Madison, pssemrad@wisc.edu 2014

Micromass LCT User s Guide

Using the AS Automated Sampler in the Simultaneous, Sequential, and Concentrate Modes

With proper calibration, the upper limit may be extended as needed.

Measuring the Conductivity of Pure Water Using USP<645>

Determination of calcium by Standardized EDTA Solution

Upon completion of this lab, the student will be able to: H3O + (aq) + A (aq) BH + (aq) + OH (aq) O + ] [H 3. ph = log[h 3.

Analytical Methods for Cleaning Validation

BY UV SPECTROPHOTOMETRY

Method OIA : Available Cyanide by Ligand Exchange and Flow Injection Analysis (FIA)

Carolina s Solution Preparation Manual

Analytical Test Method Validation Report Template

EXPERIMENT 7. Identifying a Substance by Acid-Base Titration

ph and Acidity in Tomato Sauce

Chemistry 321, Experiment 8: Quantitation of caffeine from a beverage using gas chromatography

Simultaneous determination of aspartame, benzoic acid, caffeine, and saccharin in sugar-free beverages using HPLC

1.1 This test method covers the qualitative and quantitative determination of the content of benzene and toluene in hydrocarbon wax.

HS 1003 Part 2 HS 1003 Heavy Metals Test

FerroVer Method 1 Method to 3.0, 1.0 to 30.0 and 10.0 to mg/l Fe Powder Pillows

Phosphorus, Reactive (Orthophosphate)

Design of Purified Water and Water For Injection Systems. Hugh Hodkinson Engineers Ireland Chemical & Process Division

DETERMINATION OF TOTAL CARBON, TOTAL ORGANIC CARBON AND INORGANIC CARBON IN SEDIMENTS

Transcription:

Total Organic Carbon Analysis for Purified Water and Water for Injection Application Note Abstract In pharmaceuticals, water, specifically Purified Water (PW) and Water for Intravenous Injection (WFI), is vital for drug preparation. Producing and ensuring the cleanliness of this water entails very strict quality control and precise analytical testing methods. The United States Pharmacopeia (USP) and Japanese Pharmacopeia (JP) have promoted Total Organic Carbon (TOC) analysis as the procedure to verify PW and WFI are free of TOCs and therefore able to be used for pharmaceutical purposes. A TOC system is considered suitable, under the USP, if it can recover 90% of a 0.5mgC/L system suitability standard, 1,4- benzoquinone. The JP requirement uses a dodecylbenzenesulfonic acid standard with a 90% recovery criterion. This application note will present data following these strict criteria. In addition, this study will verify the suitability of Teledyne Tekmar s TOC Analyzer (Figure 1) for analysis of TOC at 0.05 mgc/ml, as opposed to the current standard of 0.5 mgc/l. Figure 1. Teledyne Tekmar s Fusion TOC Analyzer Introduction In pharmaceuticals, water, specifically Purified Water (PW) and Water for Intravenous Injection (WFI) is vital for the preparation of intravenous drugs. Since these medications are normally introduced directly into the bloodstream, the reagents used to make them need to be as clean as possible. The United Stated Pharmacopeia (USP) and Japanese Pharmacopeia (JP) lay out strict regulations that mandate that PW and WFI are free of Total Organic Carbon (TOC) to be used for pharmaceutical applications. These methods also demand extremely sensitive instruments to detect TOC at the levels they require. The JP test method for TOC specifies, The instrument should be capable of measuring the amount of organic carbon down to 0.050 mgc/l. 1 Also, the system should be capable of generating not less than 0.450 mgc/l when using a 0.806 mg/l solution of dodecylbenzenesulfonic acid as a sample. This solution has a carbon concentration of 0.500 mgc/l and the JP method specifies that the instrument must be capable of oxidizing 90% (0.450 mgc/l) of this solution as a measured sample.

This application note will show the capability of the Teledyne Tekmar Fusion TOC analyzer of meeting these two requirements set forth in Japanese Pharmacopeia test method for total organic carbon. Experimental-Instrument Conditions For this study, analysis was performed on the Teledyne Tekmar Fusion, a UV/Persulfate TOC Analyzer using the default Japanese Pharmacopeia method (Table 1). General Parameter Value Advanced Parameter Value Sample Volume 9.0 ml Needle Rinse Volume 5.0 ml Dilution 1:1 Vial Prime Volume 2.0 ml Acid Volume 0.1 ml IC Sample Prime Volume 2.0 ml Reagent Volume 0.6 ml IC Sparge Rinse Volume 5.0 ml UV Reactor Pre-rinse Off Baseline Stabilize Time 1.60 min UV Reactor Pre-rinse Volume 0.0mL Detector Pressure Flow 300.0 ml/min Number of UV Reactor Pre-rinse 1 Syringe Speed Waste 10 IC Sparge Time 0.50 min Syringe Speed Acid 7 Detector Sweep Flow 500.0 ml/min Syringe Speed Reagent 7 Pre-Sparge Time 0.0 min Syringe Speed DI Water 7 System Flow 500.0 ml/min NDIR Pressurization 50 psig Sample Preparation Syringe Speed Sample Dispense 7 Syringe Speed Sample Aspirate 4 Syringe Speed UV Dispense 7 Syringe Speed UV Aspirate 5 Syringe Speed IC Dispense 5 Syringe Speed IC Aspirate 5 NDIR Pressure Stabilize Sample Mixing Sample Mixing Cycles 1 1.75 min Off Sample Mixing Volume 10.0 Low Level Filter NDIR Table 1: Fusion Japanese Pharmacopeia Method Parameters KHP (Potassium Hydrogen Phthalate) and dodecylbenzenesulfonic acid were used as sources of TOC. A stock solution of KHP was prepared at 1000 mgc/l in a 1 L volumetric flask. From this KHP stock solution, serial dilutions were made to prepare 0.500 mgc/l and 5.00 mgc/l working solutions. A 50 mgc/l stock solution of dodecylbenzenesulfonic acid was prepared by dissolving 80.6 mg/l of dodecylbenzenesulfonic acid into deionized water. A 0.500 mgc/l working solution was prepared by diluting this stock. Results A. Recovery of Dodecylbenzenesulfonic Acid The Japanese Pharmacopeia mandates that the system should be capable of generating not less than 0.450 mg/l of carbon when using the dodecylbenzenesulfonic acid as the sample. 1 The Fusion TOC analyzer was calibrated using a 5.00 mgc/l KHP stock solution utilizing the auto-dilution feature. The regression factor (r 2 ) for the curve was determined to be 0.9995 over a calibration range of 0.050-2.50 mgc/l. Figure 2 illustrates the calibration curve developed for this study. A sample of dodecylbenzenesulfonic acid with a carbon concentration of 0.500 mgc/l was analyzed as a sample. The results from this analysis can be seen in Table 2. Off

Figure 1: TOC Calibration curve for the Dodecylbenzenesulfonic Recovery Test Sample Name Concentration (mgc/l ) Std. Dev. (mgc/l) %RSD dodecylbenzenesulfonic acid with a carbon concentration of 0.500mg/L* dodecylbenzenesulfonic acid with a carbon concentration of 0.500mg/L* 0.503mg/L 0.009 1.85 0.504mg/L 0.012 2.50 Table 2: TOC data for 0.500 mgc/l dodecylbenzenesulfonic acid. *n=2 The analysis preformed on dodecylbenzenesulfonic acid with a carbon concentration of 0.500mgC/L had a recovery of 100.6%. This satisfies the 90% recovery criteria that is set forth in the total organic carbon test method for the Japanese Pharmacopeia. B. TOC Detection Below 0.050mg/L The Japanese Pharmacopeia also indicates The instrument should be capable of measuring the amount of organic carbon down to 0.050 mgc/l. 1 For this analysis, the Fusion was calibrated using a stock KHP solution of 0.500 mg/l of carbon utilizing the auto-dilution feature. The regression factor (r 2 ) for the calibration curve (Figure 3) was determined to be 0.9995 over a range of 0 to 0.500mgC/L. A set of check standards were analyzed at 0.050 and 0.125 mgc/l to simulate the low level that is required by the JP TOC Method. The results for these check standards are found in Table 3.

Figure 3: Calibration curve for the low level TOC analysis Sample Name Concentration (mgc/l ) Std. Dev. (mgc/l) %RSD Check Sample 0.050mg/L of Carbon 0.0503mg/L 0.0007 1.35 Check Sample 0.050mg/L of Carbon 0.0515mg/L 0.0023 4.39 Check Sample 0.125mg/L of Carbon 0.1248mg/L 0.0018 1.43 Check Sample 0.125mg/L of Carbon 0.1276mg/L 0.0025 1.99 Table 3: Data for the low level TOC analysis The results of this low level analysis meet the method criteria of accurately measuring the amount of organic carbon down to 0.050 mgc/l. Even at such a low level, the samples show excellent reproducibility as evident by the low %RSDs (1.35 to 4.39%). Since these samples were ran as a check standards, the Fusion software automatically subtracts the blank contribution of carbon that is present in the system and deionized water. The data listed in Table 3 satisfies the criteria set forth in the total organic carbon test method for the Japanese Pharmacopeia.

Conclusions When dealing with pharmaceuticals, specifically Purified Water (PW) and Water for Intravenous Injection (WFI) the implications on human health require very strict regulation. One of the methods of determining the cleanliness of these reagents is total organic carbon testing. Both the USP and JP lay out detailed guidelines that must be followed, with the main difference being the system suitability test material. For the JP method, the TOC instrument must be capable of recovering 90% of a 0.500 mg/l carbon sample of dodecylbenzenesulfonic acid. The Fusion met and exceeded these criteria, recovering 100.6% of the 0.500 mgc/l sample of dodecylbenzenesulfonic acid. The second part of the JP TOC system suitability states that the system must be able to measure a carbon sample down to 0.050mgC/L. Data presented using the Fusion also met this criteria, with accurate and reproducible results, even at low levels. This study demonstrates the suitability of Teledyne Tekmar s Fusion TOC Analyzer for this analysis by meeting and surpassing all method performance criteria. References 1. Japanese Pharmacopeia General Test/ Test for Total Organic Carbon pg.93-95 2. U.S. Pharmacopeia General Chapter <643> Total Organic Carbon. www.usp.org