Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2
|
|
|
- Annis Crawford
- 9 years ago
- Views:
Transcription
1 Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis, Oregon Prepared by Kathryn Motter And Cameron Jones Revised
2 Page 2 of 11 Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A.2 Table of Contents 1.0 Scope and Application Summary of Method Definitions Interferences Safety Equipment and Supplies Reagents and Standards Preparation of Standards Sample Handling and Storage Quality Control Calibration and Standardization Procedure Beaker Preparation Sample Setup Procedural Notes Data Analysis and Calculations Method Performance Pollution Prevention Waste Management References Tables, Diagrams, Flowcharts, and Validation Data Total and Total Dissolved Solids Summary Results Data Sheet Total and Total Dissolved Solids Weights Record Data Sheet Document Revision History... 11
3 Page 3 of 11 Standard Operating Procedure for the Determination of Total and Total Dissolved Solids CCAL 13A Scope and Application 1.1 Total Dissolved Solids are defined as the material residue left in a vessel, after evaporation of a sample that has passed through a filter. Total Solids are the material residue left after evaporation of an unfiltered sample. The detection limit for this determination is 5 mg/l solid residue. 2.0 Summary of Method 2.1 A thoroughly mixed sample is volumetrically transferred to a tared beaker and evaporated to dryness. The gain in weight represents the total or total dissolved solids per initial volume. CCAL uses Whatman GF/F or GF/C glass microfiber filters; 0.7 and 1.2 um particle retention ratings, respectively. Samples are evaporated at 180ºC for five days. 3.0 Definitions 3.1 DI water: Water that has been through a deionization system to produce water similar to ASTM Type I reagent with 16.7 Mohms resistivity (ASTM) (Reference 16.3). 4.0 Interferences 4.1 Samples with high mineral content may be hygroscopic and require prolonged drying, followed by proper desiccation and rapid weighing. 4.2 Samples high in bicarbonate require prolonged drying to insure complete conversion of bicarbonate to carbonate. 4.3 Avoid excessive residue, as this may result in formation of a watertrapping crust.
4 Page 4 of Safety 5.1 The toxicity or carcinogenicity of each reagent has not been precisely determined; however, each chemical should be regarded as a potential health hazard. Exposure to these chemicals should be reduced to the lowest possible level. Cautions are included for known extremely hazardous materials. 6.0 Equipment and Supplies Note: Brand names, suppliers and part numbers are for illustrative purposes only. No endorsement is implied. Equivalent performance may be achieved using apparatus and materials other than those specified here, but demonstration of equivalent performance that meets the requirements of this method is the responsibility of the laboratory ml Griffin beakers (glass) 6.2 Drying oven equipped with thermostatic control capable of maintaining temperature within 5 o C range 6.3 Desiccator - with moisture indicating desiccant 6.4 Analytical balance - capable of weighing to 0.1 mg 6.5 Filtration system - filter funnel, filter stage, filter barrel, clamps, Erlenmeyer filter flask ml graduated cylinder 6.7 GF/F or GF/C filter papers 6.8 Stainless steel screen with 1 mm 2 mesh. 6.9 Large porcelain Buchner funnel Vacuum system and connecting hoses. 7.0 Reagents and Standards 7.1 Preparation of Standards
5 Page 5 of ppm NaCl Solution: Dissolve g NaCl (oven-dried at 80ºC) into 1000 ml of DI water and mix until dissolution is complete. 8.0 Sample Handling and Storage 8.1 Samples are stored at 4 C in the dark. 9.0 Quality Control 9.1 Preparation of standards is recorded on worksheets and documented by weight of standard added to a given flask before dilution to volume with DI water. All records of certification are kept on file at CCAL Laboratory. 9.2 Blank: DI water run twice each analysis set. 9.3 Quality Control Check Standard: 1000 ppm NaCl check standard run once each analysis set. 9.4 Control: empty beaker run through complete process Calibration and Standardization 10.1 Balances: calibrated yearly by external vendor Acceptable recovery of NaCl check standard is ± 2 % Procedure 11.1 Beaker Preparation Place prenumbered beakers into a 180ºC drying oven and dry for five days to a constant weight Transfer dried beakers to desiccator(s) and allow to stabilize overnight Record beaker numbers to be used on the Data Summary and Weight Record data sheets Zero the balance Weigh clean, dry, empty beakers. Record weight on the Weight Record data sheet.
6 Page 6 of Immediately after all beakers have been weighed, a quality assurance weight check (QA) must be performed on 20% of the beakers by someone other than the person performing the initial measurements After all beakers have been weighed, transcribe the lowest weight for that beaker from the Weights Record data sheet to the tare column of the Summary Results data sheet for that beaker Sample Setup Assign sample numbers to beaker numbers on the analysis data sheet Each analysis batch must have a control beaker to which no sample, blank or standard is added, two DIW blanks, and one 1000 mg/l NaCl standard In a graduated cylinder, measure 100 ml of sample, blank or standard into the appropriate dry, tared beaker Place samples, control, standard and blanks in the 180ºC drying oven and evaporate for at least five days to constant weight Transfer dried beakers to desiccator(s) and allow to stabilize overnight Weigh sample beakers as before ( ), allowing each beaker to stabilize After all beakers have been weighed, transcribe the lowest of consistent weights for that beaker from the Weights Record data sheet to the tare column of the Summary Results data sheet for that beaker The difference between the loaded weight and the tare weight equals the weight of the dissolved solids in 100 ml of sample Procedural Notes Make multiple measurements to assure a constant weight. Repeat weighing procedure until three consecutive measurements match to within 0.2 mg. Hygroscopic matter may result in water retention. If weight continually increases after multiple readings, use the initial weight Dried beakers will gain weight if left in the open air. Minimize time out of the desiccator, and replace the lid of the desiccator after removing/returning a beaker Recheck balance zero after obtaining a final weight for the first beaker, the fifth beaker and the tenth beaker The QA weight must be within ± 0.5 mg of the initial weight. If the QA weight does not fall to within ± 0.5 mg of the final initial
7 Page 7 of 11 weight, then the beakers must be reweighed in reverse order from the out-of-qa-limit beaker until measurements fall to within ± 0.5 mg of the final initial weight. Reweighed beakers must again be QA checked and meet the above criteria or be reweighed until QA criteria are met It is especially important that as little time as possible elapse between initial weighing and QA weighing If multiple balances are being used use the same balance for the tare and loaded measurements to minimize variability If dissolved solids fraction is low (less than l mg) an additional 100 ml of sample can be added to the same beaker and the process repeated Beakers with samples should not be dried more than 15 days to prevent over drying Do not touch beakers with bare hands during the weighing and analysis process as skin oils may leave a detectable residue Data Analysis and Calculations 12.1 Filterable residue (total dissolved solids; TDS) or unfiltered residue (total solids; TS) Results are reported as mg/liter. TS or TDS mg/l = (B - A) x 1000 mg/g L where A = beaker tare weight (g), B = beaker tare weight + dried sediment weight (g) 13.0 Method Performance 13.1 This method was validated through inter-laboratory studies. The CCAL participates in the USGS Standard Reference Water QA program Pollution Prevention 14.1 The chemicals used in this method pose little threat to the environment when properly managed All standards and reagents should be prepared in volumes consistent with laboratory use to minimize the volume of disposable waste.
8 Page 8 of For further information on pollution prevention consult Less is better: Laboratory Chemical Management for Waste Reduction, available from the American Chemical Society s Department of Government Relations and Science Policy, th Street NW, Washington D.C , (202) Waste Management 15.1 It is the laboratory s responsibility to comply with all federal, state and local regulations governing waste management, and to protect the environment by minimizing and controlling all releases from fume hoods and bench operations. Compliance with all sewage discharge permits and regulations is required For further information on waste management, consult The Waste Management Manual for Laboratory Personnel, and Less is Better: Laboratory Chemical Management for Waste Reduction, both available from the American Chemical Society s Department of Government Relations and Science Policy, th Street NW, Washington DC, References 16.1 Standard Methods For The Examination of Water and Wastewater, Method 2540 Total Dissolved Solids Dried at 180ºC. American Public Health Association. 21 st Edition, Code of Federal Regulations. Protection of Environment. Section 40, Appendix B to Part 136. Definition and procedure for the determination of the method detection limit. Revision Revised July 1, Office of the Federal Register, National Archives and Records ASTM. American Society for Testing and Materials. Standard Specifications for Reagent Water. D (Reapproved 1983). Annual Book of ASTM Standards, Vol ASTM: Philadelphia, PA, Tables, Diagrams, Flowcharts, and Validation Data 17.1 Total and Total Dissolved Solids Summary Results Data Sheet
9 Page 9 of 11 Total Dissolved Solids Summary Results Project Log # Analyst Dates: Beakers Weighed Samples Setup Final Weighing Remarks: Sample # Tare (g) Tare + Sediment (g) Volume (ml) Notes Beaker # QDSC1 QDSB1 QDSB2 QDSL1 Control Blank Blank 1000 ppm Std
10 17.2 Total and Total Dissolved Solids Weights Record Data Sheet TS/TDS/CCAL 13A.2 Page 10 of 11 Total Dissolved Solids Weights Record Tare Tare + Sediment (circle one) Log # Analyst Date Beaker # Weighing # Zero / QA Check
11 Page 11 of Document Revision History Original Document: March 2006 Version: 13A.0 Title: Standard Operating Procedure for the Determination of Total Dissolved Solids Edit Date: February 2010 New Version: 13A.1 Address update Add Total Solids procedure and change document title to reflect the addition Edit Date: New Version 13A.2 Section 6.2: update temperature acceptance range Section 11: reorganized for clarity General editing
Total Dissolved Solids:
Total Dissolved Solids: Going From Tedious to TDS Michael Bechtold, Technical Specialist David Smith, Technical Director Total Dissolved Solids (TDS) What is it and why does it matter? How is it currently
Solids, Volatile Dissolved and Fixed Dissolved
, 8277 Solids, Volatile Dissolved and Fixed Dissolved Gravimetric Method 1 Scope and Application: For wastewater. 1 Adapted from Standard Methods for the Examination of Water and Wastewater DOC316.53.001206
With proper calibration, the upper limit may be extended as needed.
While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the
Determination of Insoluble Solids in Pretreated Biomass March 2008 Material
National Renewable Energy Laboratory Innovation for Our Energy Future A national laboratory of the U.S. Department of Energy Office of Energy Efficiency & Renewable Energy Determination of Insoluble Solids
Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls
Nitrate DOC316.53.01066 Cadmium Reduction Method Method 8039 0.3 to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater. Test preparation Instrument-specific
Suspended Solids: Unequal Results from Equal Methods
Suspended Solids: Unequal Results from Equal Methods By John Stone As of November 2003, 40 CFR, Part 136 lists five approved methods for determining Total Suspended Solids (Residue). Major differences
MOISTURE (Karl Fischer, Buffered)
MOIST.03-1 MOISTURE (Karl Fischer, Buffered) PRINCIPLE SCOPE The sample is dissolved in a mixture of methanol and formamide (50:50 v/v) and then titrated with standardized Karl Fischer reagent. The titration
Determination of a Chemical Formula
1 Determination of a Chemical Formula Introduction Molar Ratios Elements combine in fixed ratios to form compounds. For example, consider the compound TiCl 4 (titanium chloride). Each molecule of TiCl
Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters
Determination of Total Suspended Solids (TSS) and Total Volatile Solids (TVS) in Waters of Fresh/Estuarine/Coastal Waters 1. SCOPE and APPLICATION 1.1 Gravimetric analysis is used to determine total suspended
COMMON LABORATORY APPARATUS
COMMON LABORATORY APPARATUS Beakers are useful as a reaction container or to hold liquid or solid samples. They are also used to catch liquids from titrations and filtrates from filtering operations. Bunsen
Standard Methods for the Examination of Water and Wastewater
2540 SOLIDS#(1)* 2540 A. Introduction Solids refer to matter suspended or dissolved in water or wastewater. Solids may affect water or effluent quality adversely in a number of ways. Waters with high dissolved
Color, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units Scope and application: For water, wastewater and seawater; equivalent to NCASI method
Color, True and Apparent
Color, True and Apparent DOC316.53.01037 Platinum-Cobalt Standard Method 1, 2, 3 Method 8025 15 to 500 color units 5 to 500 color units (low range technique) Scope and application: For water, wastewater
4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES
4.0 EXPERIMENT ON DETERMINATION OF CHLORIDES Sl. No. Contents Preamble 4.1 Aim 4.2 Introduction 4.2.1 Environmental Significance 4.3 Principle 4.4 Materials Required 4.4.1 Apparatus Required 4.4.2 Chemicals
SODIUM CARBOXYMETHYL CELLULOSE
SODIUM CARBOXYMETHYL CELLULOSE Prepared at the 28th JECFA (1984), published in FNP 31/2 (1984) and in FNP 52 (1992). Metals and arsenic specifications revised at the 55 th JECFA (2000). An ADI not specified
105 Adopted: 27.07.95
105 Adopted: 27.07.95 OECD GUIDELINE FOR THE TESTING OF CHEMICALS Adopted by the Council on 27 th July 1995 Water Solubility INTRODUCTION 1. This guideline is a revised version of the original Guideline
EDTA Titrations 1: Standardization of EDTA and Analysis of Zinc in a Supplement Tablet. by Professor David Cash. September, 2008
CHEMICAL, ENVIRONMENTAL, AND BIOTECHNOLOGY DEPARTMENT EDTA Titrations 1: Standardization of EDTA and Analysis of Zinc in a Supplement Tablet by Professor David Cash September, 2008 Mohawk College is the
Acid Base Titrations
Acid Base Titrations Introduction A common question chemists have to answer is how much of something is present in a sample or a product. If the product contains an acid or base, this question is usually
The Analytical Balance
Chemistry 119: Experiment 1 The Analytical Balance Operation of the Single-Pan Analytical Balance Receive instruction from your teaching assistant concerning the proper operation of the Sartorius BP 210S
Experiment 12- Classification of Matter Experiment
Experiment 12- Classification of Matter Experiment Matter can be classified into two groups: mixtures and pure substances. Mixtures are the most common form of matter and consist of mixtures of pure substances.
SIEVE ANALYSIS OF FINE AND COARSE AGGREGATES
Test Procedure for SIEVE ANALYSIS OF FINE AND COARSE AGGREGATES TxDOT Designation: Tex-200-F Effective Date: January 2016 1. SCOPE 1.1 Use this test method to determine the particle size distribution of
PREPARATION AND PROPERTIES OF A SOAP
(adapted from Blackburn et al., Laboratory Manual to Accompany World of Chemistry, 2 nd ed., (1996) Saunders College Publishing: Fort Worth) Purpose: To prepare a sample of soap and to examine its properties.
Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow
Phosphorus, colorimetry, phosphomolybdate, automated-segmented flow Parameter and code: Phosphorus, total-in-bottom-material, dry weight, I-6600-88 (mg/kg as P): 00668 1. Application This method is used
Synthesis of Aspirin and Oil of Wintergreen
Austin Peay State University Department of hemistry hem 1121 autions Purpose Introduction Acetic Anhydride corrosive and a lachrymator all transfers should be done in the vented fume hood Methanol, Ethanol
Hot water extractable acidity or alkalinity of paper (Reaffirmation of T 428 om-08) (No changes from Draft 1)
NOTICE: This is a DRAFT of a TAPPI Standard in ballot. Although available for public viewing, it is still under TAPPI s copyright and may not be reproduced or distributed without permission of TAPPI. This
Esterification Method 1 Method 8196 27 to 2800 mg/l (as acetic acid) Reagent Solution
Volatile Acids DOC316.53.01144 Esterification Method 1 Method 8196 27 to 2800 mg/l (as acetic acid) Reagent Solution Scope and application: For digestor sludges. 1 Adapted from The Analyst, 87, 949 (1962).
Apr 17, 2000 LAB MANUAL 1302.0. 1302 PARTICLE SIZE ANALYSIS OF SOILS AASHTO Designation T 88 (Mn/DOT Modified)
Apr 17, 2000 LAB MANUAL 1302.0 1302 PARTICLE SIZE ANALYSIS OF SOILS AASHTO Designation T 88 (Mn/DOT Modified) 1302.1 SCOPE This method describes a procedure for the quantitative determination of the distribution
Mixtures and Pure Substances
Unit 2 Mixtures and Pure Substances Matter can be classified into two groups: mixtures and pure substances. Mixtures are the most common form of matter and consist of mixtures of pure substances. They
Calibration of Glassware
Calibration of Glassware Introduction Glassware is commonly calibrated using a liquid of known, specific density, and an analytical balance. The procedure is to determine the mass of liquid the glassware
Enhanced Organic Precursor Removals Using Aged Filter Media Page 1. Enhanced Organic Precursor Removals Using Aged Filter Media
Enhanced Organic Precursor Removals Using Aged Media Page 1 Enhanced Organic Precursor Removals Using Aged Media O b j e c t i v e s The main goals of this project was to evaluate the dissolved organic
Determination of calcium by Standardized EDTA Solution
Determination of calcium by Standardized EDTA Solution Introduction The classic method of determining calcium and other suitable cations is titration with a standardized solution of ethylenediaminetetraacetic
experiment5 Understanding and applying the concept of limiting reagents. Learning how to perform a vacuum filtration.
81 experiment5 LECTURE AND LAB SKILLS EMPHASIZED Synthesizing an organic substance. Understanding and applying the concept of limiting reagents. Determining percent yield. Learning how to perform a vacuum
In this experiment, we will use three properties to identify a liquid substance: solubility, density and boiling point..
Identification of a Substance by Physical Properties 2009 by David A. Katz. All rights reserved. Permission for academic use provided the original copyright is included Every substance has a unique set
Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material
PM 2.5 Nylon Filter Cleaning Date: August 25,2009 Page I of8 Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material Environmental and Industrial Sciences Division
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850
UNITED STATES CONSUMER PRODUCT SAFETY COMMISSION DIRECTORATE FOR LABORATORY SCIENCES DIVISION OF CHEMISTRY 5 RESEARCH PLACE ROCKVILLE, MD 20850 Test Method: CPSC-CH-E1001-08.2 Standard Operating Procedure
MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1
MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Chemistry 5.310 Laboratory Chemistry EXPERIMENT #5 THE POTENTIOMETRIC TITRATION OF AN ACID MIXTURE 1 I. PURPOSE OF THE EXPERIMENT In this experiment
Stoichiometry Limiting Reagent Laboratory. Chemistry 118 Laboratory University of Massachusetts, Boston
Chemistry 118 Laboratory University of Massachusetts, Boston STOICHIOMETRY - LIMITING REAGENT --------------------------------------------------------------------------------------------------------------------------------------------
USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls
Iron, Total DOC316.53.01053 USEPA 1 FerroVer Method 2 Method 8008 0.02 to 3.00 mg/l Fe Powder Pillows or AccuVac Ampuls Scope and application: For water, wastewater and seawater; digestion is required
Separation by Solvent Extraction
Experiment 3 Separation by Solvent Extraction Objectives To separate a mixture consisting of a carboxylic acid and a neutral compound by using solvent extraction techniques. Introduction Frequently, organic
FAJANS DETERMINATION OF CHLORIDE
EXPERIMENT 3 FAJANS DETERMINATION OF CHLORIDE Silver chloride is very insoluble in water. Addition of AgNO 3 to a solution containing chloride ions results in formation of a finely divided white precipitate
Continuous process of sodium bicarbonate production by Solvay method
Continuous process of sodium bicarbonate production by Solvay method Manual to experiment nr 10 Instructor: Dr Tomasz S. Pawłowski 1 Goal of the experiment The goal of the experiment is introduction of
(a) Method 1 Sample and velocity traverses for stationary sources.
While we have taken steps to ensure the accuracy of this Internet version of the document, it is not the official version. Please refer to the official version in the FR publication, which appears on the
Assessment of Accuracy and Precision
2 chapter Assessment of Accuracy and Precision S.S. Nielsen, Food Analysis Laboratory Manual, Food Science Texts Series, DOI 10.1007/978-1-4419-1463-7_2, Springer Science+Business Media, LLC 2010 9 Chapter
Laboratory Evaluation Checklist - PSP
Laboratory Evaluation Checklist - PSP PUBLIC HEALTH SERVICE U.S. FOOD AND DRUG ADMINISTRATION SHELLFISH PROGRAM IMPLEMENTATION BRANCH SHELLFISH SAFETY TEAM 5100 PAINT BRANCH PARKWAY COLLEGE PARK, MD 20740-3835
How to measure dissolved, suspended & total solids
World Bank & Government of The Netherlands funded Training module # WQ - 10 How to measure dissolved, suspended & total solids New Delhi, May 1999 CSMRS Building, 4th Floor, Olof Palme Marg, Hauz Khas,
Calibration of Volumetric Glassware
CHEM 311L Quantitative Analysis Laboratory Revision 2.3 Calibration of Volumetric Glassware In this laboratory exercise, we will calibrate the three types of glassware typically used by an analytical chemist;
Physical and Chemical Properties and Changes
Physical and Chemical Properties and Changes An understanding of material things requires an understanding of the physical and chemical characteristics of matter. A few planned experiments can help you
Phosphorus, Reactive (Orthophosphate)
Phosphorus, Reactive (Orthophosphate) DOC316.53.01113 Amino Acid Method 1 Method 8178 0.23 to 30.00 mg/l PO 4 Reagent Solution Scope and application: For water, wastewater and seawater. 1 Adapted from
Florida Method of Test for TESTING OF GROUND TIRE RUBBER Designation: FM 5-559
Florida Method of Test for TESTING OF GROUND TIRE RUBBER Designation: FM 5-559 1 SCOPE 1.1 This method is used to determine the physical requirements of ground tire rubber for use in asphalt rubber using
The volume of a penny will be calculated from its mass and density.
Measurement and Density In science a key concern is the quantities involved in chemical processes. These amounts can be directly measured or calculated from other measurements. A measurement consists of
Calibration of Volumetric Glassware
Chemistry 119: Experiment 2 Calibration of Volumetric Glassware For making accurate measurements in analytical procedures, next in importance to the balance is volumetric equipment. In this section volumetric
Analysis of Vitamin C Using Iodine. Introduction
Analysis of Vitamin C Using Iodine Introduction Vitamin C (ascorbic acid) is oxidized to dehydroascorbic acid using a mild oxidizing agent such as iodine. The oxidation is a two- electron process, following
CHEM 2423 Recrystallization of Benzoic Acid EXPERIMENT 4 - Purification - Recrystallization of Benzoic acid
EXPERIMENT 4 - Purification - Recrystallization of Benzoic acid Purpose: a) To purify samples of organic compounds that are solids at room temperature b) To dissociate the impure sample in the minimum
5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS
5.0 EXPERIMENT ON DETERMINATION OF TOTAL HARDNESS Sl. No. Contents Preamble 5.1 Aim 5.2 Introduction 5.2.1 Environmental Significance 5.3 Principle 5.4 Materials Required 5.4.1 Apparatus Required 5.4.2
Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)
Name: Date: Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating) High School Environmental Science AP Module 1 Environmental Lab NGSSS Big Ideas: This module is a laboratory-based
SUPPLEMENTARY MATERIAL
SUPPLEMENTARY MATERIAL (Student Instructions) Determination of the Formula of a Hydrate A Greener Approach Objectives To experimentally determine the formula of a hydrate salt. To learn to think in terms
NITRIC OXIDE and NITROGEN DIOXIDE 6014
NITRIC OXIDE and NITROGEN DIOXIDE 6014 NO MW: 30.01 CAS: 10102-43-9 RTECS: QX0525000 46.01 10102-44-0 QW9800000 METHOD: 6014, Issue 1 EVALUATION: FULL Issue 1: 15 August 1994 OSHA : 25 ppm NO; C 1 ppm
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY
CONFIRMATION OF ZOLPIDEM BY LIQUID CHROMATOGRAPHY MASS SPECTROMETRY 9.1 POLICY This test method may be used to confirm the presence of zolpidem (ZOL), with diazepam-d 5 (DZP-d 5 ) internal standard, in
Dissolving of sodium hydroxide generates heat. Take care in handling the dilution container.
TITRATION: STANDARDIZATION OF A BASE AND ANALYSIS OF STOMACH ANTACID TABLETS 2009, 1996, 1973 by David A. Katz. All rights reserved. Reproduction permitted for education use provided original copyright
QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE
QUALITY ASSURANCE / QUALITY CONTROL POLICY BREG SOIL AND WATER QUALITY LABORATORY UNIVERSITY OF DELAWARE Department of Bioresources Engineering Rm 111 Worrilow Hall 531 S. College Ave. Newark, DE 19716-2140
Where the exp subscripts refer to the experimental temperature and pressure acquired in the laboratory.
Molar Volume of Carbon Dioxide Reading assignment: Julia Burdge, Chemistry 3rd edition, Chapter 10. Goals To determine the molar volume of carbon dioxide gas and the amount of sodium carbonate in a sample.
Solutions and Dilutions
Learning Objectives Students should be able to: Content Design a procedure for making a particular solution and assess the advantages of different approaches. Choose the appropriate glassware to ensure
Loss-on-Ignition Standard Operating Procedure
Page 1 of 5 Loss-on-Ignition Standard Operating Procedure LacCore, National Lacustrine Core Facility 11/11/2013 Contents Purpose... 1 Equipment... 2 Safety... 2 Summary... 2 Procedure... 3 How to use the
The Molar Mass of a Gas
The Molar Mass of a Gas Goals The purpose of this experiment is to determine the number of grams per mole of a gas by measuring the pressure, volume, temperature, and mass of a sample. Terms to Know Molar
Standard Methods for the Examination of Water and Wastewater
2320 ALKALINITY*#(1) 2320 A. Introduction 1. Discussion Alkalinity of a water is its acid-neutralizing capacity. It is the sum of all the titratable bases. The measured value may vary significantly with
PECTINS. SYNONYMS INS No. 440 DEFINITION DESCRIPTION. FUNCTIONAL USES Gelling agent, thickener, stabilizer, emulsifier CHARACTERISTICS
PECTINS SYNONYMS INS No. 440 Prepared at the 71 st JECFA (2009) and published in FAO JECFA Monographs 7 (2009), superseding specifications prepared at the 68 th JECFA (2007) and published in FAO JECFA
Colorimetric Determination of Iron in Vitamin Tablets
Cautions: 6 M hydrochloric acid is corrosive. Purpose: To colorimetrically determine the mass of iron present in commercial vitamin tablets using a prepared calibration curve. Introduction: Iron is considered
Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater
Document: AND Sol Env 08 2013 Environmental Water Testing: Surface Water, Groundwater, Hard Water, Wastewater, & Seawater Matrix specific sample preparation and testing methods for environmental waters
EPA-821-R-98-002 Washington, DC 20460 Agency February 1999
United States Office of Water EPA-821-R-98-002 Environmental Protection Washington, DC 20460 PB99-121949 Agency February 1999 Method 1664, Revision A: N-Hexane Extractable Material (HEM; Oil and Grease)
Determination of the Mass Percentage of Copper in a Penny. Introduction
Determination of the Mass Percentage of Copper in a Penny Introduction This experiment will cost you one penny ($0.01). The penny must be minted after 1983. Any penny will do; for best results the penny
METHOD 353.2 DETERMINATION OF NITRATE-NITRITE NITROGEN BY AUTOMATED COLORIMETRY
METHOD 353.2 DETERMINATION OF NITRATE-NITRITE NITROGEN BY AUTOMATED COLORIMETRY Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division Revision 2.0 August 1993 ENVIRONMENTAL MONITORING
ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND
#3. Acid - Base Titrations 27 EXPERIMENT 3. ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND Carbonate Equilibria In this experiment a solution of hydrochloric
SYNTHESIS AND ANALYSIS OF A COORDINATION COMPOUND OF COPPER
Chemistry 111 Lab: Synthesis of a Copper Complex Page H-1 SYNTHESIS AND ANALYSIS OF A COORDINATION COMPOUND OF COPPER In this experiment you will synthesize a compound by adding NH 3 to a concentrated
PART I: PREPARATION OF SOLUTIONS AND STANDARDIZATION OF A BASE
TITRATION: STANDARDIZATION OF A BASE AND ANALYSIS OF STOMACH ANTACID TABLETS 2009, 1996, 1973 by David A. Katz. All rights reserved. Reproduction permitted for education use provided original copyright
Method 202 Dry Impinger Method for Determining Condensable Particulate Emissions from Stationary Sources. By: Paul Ottenstein
Air Quality Issues in the Rocky Mountain Region PM2.5 Source Testing Method 201A Determination PM 10 and PM 2.5 Emissions from Stationary Sources (Constant Sampling Rate Procedure) Method 202 Dry Impinger
XI. Methods of Analysis DETERMINATION OF POTASSIUM CARBONATE CALCULATIONS REAGENTS PROCEDURE
XI. NOTE: An automatic titrator may be utilized for ease of analysis, especially if this test is performed often. Contact Technical Service for further information concerning automatic titration. Methods
FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows
Iron, Total DOC316.53.01314 FerroVer Method 1 Method 10249 0.1 to 3.0, 1.0 to 30.0 and 10.0 to 300.0 mg/l Fe Powder Pillows Scope and application: For oil and gas field waters; digestion is required for
LACHAT METHOD NUMBER 10-204-00-5-D Rev 1, 21 March 2011 SCOPE AND APPLICATION Approximately 3-500 g CN/L 2-500 g CN/L. 09e
1 SIDE BY SIDE COMPARISON TABLE FOR THE DETERMINATION OF CYANIDE BY UV DIGESTION AND AMPEROMETRIC DETECTION FIA METHOD - 10-204-00-5-D (ASTM D7511-09 e ) TOPIC Method ASTM D7511-09e LACHAT METHOD NUMBER
Experiment 8 Synthesis of Aspirin
Experiment 8 Synthesis of Aspirin Aspirin is an effective analgesic (pain reliever), antipyretic (fever reducer) and anti-inflammatory agent and is one of the most widely used non-prescription drugs. The
DURAPOSIT TM SMT 810 Electroless Nickel
DURAPOSIT TM SMT 810 Electroless Nickel For PWB Metallization Applications Regional Product Availability Asia Japan/Korea North America Europe Description DURAPOSIT TM SMT 810 is an electroless nickel
Stoichiometry Limiting Reagent Laboratory. Chemistry 118 Laboratory University of Massachusetts, Boston
Chemistry 118 Laboratory University of Massachusetts, Boston STOICHIOMETRY - LIMITING REAGENT -----------------------------------------------------------------------------------------------------------------------------
BRIEFING 661.2 Plastic Packaging Systems for Pharmaceutical Use.
BRIEFING 661.2 Plastic Packaging Systems for Pharmaceutical Use. USP proposes the revision and development of a suite of plastic packaging system standards in the current issue of PF. General test chapter
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF)
METHOD 9075 TEST METHOD FOR TOTAL CHLORINE IN NEW AND USED PETROLEUM PRODUCTS BY X-RAY FLUORESCENCE SPECTROMETRY (XRF) 1.0 SCOPE AND APPLICATION 1.1 This test method covers the determination of total chlorine
Halogen Free: What, Why and How. Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc.
Halogen Free: What, Why and How Presented by : Jim Cronin Environmental Monitoring and Technologies, Inc. We will discuss What are Halogens and how we interact with them. Why Halogens, within consumer
STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14
STANDARDIZATION OF A SODIUM HYDROXIDE SOLUTION EXPERIMENT 14 OBJECTIVE The objective of this experiment will be the standardization of sodium hydroxide using potassium hydrogen phthalate by the titration
How to prepare standard solutions
World Bank & Government of The Netherlands funded Training module # WQ -04 How to prepare standard solutions New Delhi, May 1999 CSMRS Building, 4th Floor, Olof Palme Marg, Hauz Khas, New Delhi 11 00 16
Nitrogen, Ammonia. Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N. Ammonia ISE. Test preparation. Instrument-specific table
Nitrogen, Ammonia DOC316.53.01234 Known Addition ISE Method 1 Method 10002 Minimum of 0.8 mg/l NH 3 N Scope and application: For wastewater 2. Ammonia ISE 1 Adapted from the Standard Methods for the Examination
Measurement and Calibration
Adapted from: H. A. Neidig and J. N. Spencer Modular Laboratory Program in Chemistry Thompson Learning;, University of Pittsburgh Chemistry 0110 Laboratory Manual, 1998. Purpose To gain an understanding
Laboratory Exercise: The Precision of Volume Measurements
CHEM 109 Introduction to Chemistry Revision 3.0 Laboratory Exercise: The Precision of Volume Measurements The accurate measurement of the volume of liquids tends to be difficult and chemists have a number
Experiment 1: Measurement and Density
Experiment 1: Measurement and Density Chemistry 140 Learning Objectives Become familiar with laboratory equipment and glassware Begin to see the link between measurement and chemical knowledge Begin to
Coating and Extraction of Honeycomb Denuders
QSM Approval: Coating and Extraction of Honeycomb Denuders 1 INTRODUCTION 1.1 The following procedures are used for the coating of honeycomb denuders with citric acid and sodium carbonate solutions, and
1 Quality Assurance and Quality Control Project Plan
1 Quality Assurance and Quality Control Project Plan The purpose of this section is to describe the quality assurance/quality control program that will be used during the system specific field testing
LIST OF EQUIPMENT IN CHEMISTRY LABORATORY
LIST OF EQUIPMENT IN CHEMISTRY LABORATORY 1 Adana Science and Technology University-Bioengineering Department INSTRUMENTS: Analytical Balance (RADWAG AS220 C/2) 3 Analytical Balance (Denver Instrument)...4
APPENDIX E CHLOROPHYLL-A STANDARD OPERATING PROCEDURES
APPENDIX E CHLOROPHYLL-A STANDARD OPERATING PROCEDURES Title Page No. Standard Operating Procedure For E-1 Chlorophyll-a Determination Used by Sabine River Authority Chlorophyll-a Standard Operating Procedure
METHOD 3510C SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION
METHOD 3510C SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION 1.0 SCOPE AND APPLICATION 1.1 This method describes a procedure for isolating organic compounds from aqueous samples. The method also describes concentration
Odor (Threshold Odor, Consistent Series)
METHOD #: 140.1 (Issued 1971) TITLE: ANALYTE: INSTRUMENTATION: Odor (Threshold Odor, Consistent Series) Odor N/A STORET No. 60 C: 00086 Room Temp: 00085 1.0 Scope and Application 1.1 This method is applicable
ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography)
ACETA.02-1 ACETALDEHYDE and ISOVALERALDEHYDE (Gas Chromatography) PRINCIPLE Isovaleraldehyde (IVA) and acetaldehyde are released from the syrup with the aid of dilute phosphoric acid and heat. The liberated
METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY
METHOD 8020A AROMATIC VOLATILE ORGANICS BY GAS CHROMATOGRAPHY 1.0 SCOPE AND APPLICATION 1.1 Method 8020 is used to determine the concentration of various aromatic volatile organic compounds. The following
Extraction of Caffeine from Energy Drinks
Boston University OpenBU Chemistry http://open.bu.edu Organic Chemistry Laboratory Experiments 2011-07-14 Extraction of Caffeine from Energy Drinks Mulcahy, Seann P. http://hdl.handle.net/2144/1418 Boston