Choosing the Best Device Sample Size for Verification and Validation. Steven Walfish Staff Statistician, Corporate Quality

Similar documents
Capability Analysis Using Statgraphics Centurion

Quality Risk Management ICH Q9 & ISO Presented by Michael Kerr 11 th November 2011

Confidence Intervals for Cpk

Best Practice In A Change Management System

QUALITY RISK MANAGEMENT (QRM): A REVIEW

Risk Assessment and Management. Allen L. Burgenson Manager, Regulatory Affairs Lonza Walkersville Inc.

Sample Size and Power in Clinical Trials

IVD Regulation Overview. Requirements to Assure Quality & Effectiveness

Confidence Intervals for One Standard Deviation Using Standard Deviation

Improved Utilization of Self-Inspection Programs within the GMP Environment A Quality Risk Management Approach

Design Verification The Case for Verification, Not Validation

Process Validation for Medical Devices

TÜV Rheinland Functional Safety Program Functional Safety Engineer Certification

QA GLP audits. Alain Piton. Antipolis,, France.

Software-based medical devices from defibrillators

Risk Assessment for Medical Devices. Linda Braddon, Ph.D. Bring your medical device to market faster 1

SAMPLE SIZE CONSIDERATIONS

VELUX General Quality Requirements

Confidence Intervals for Cp

Corrective and Preventive Action Background & Examples Presented by:

Quality Risk Management Tools Quality Risk Management Tool Selection When to Select FMEA: QRM Tool Selection Matrix

Risk Management and the Impact of EN ISO 14971:2012 Annex Z

Quality Risk Management Understanding and Control the Risk in Pharmaceutical Manufacturing Industry

Risk assessment is used as a

Quality Risk Management in Pharmaceutical Industry: A Review

INFLUENCE OF MEASUREMENT SYSTEM QUALITY ON THE EVALUATION OF PROCESS CAPABILITY INDICES

University of Paderborn Software Engineering Group II-25. Dr. Holger Giese. University of Paderborn Software Engineering Group. External facilities

FINAL DOCUMENT. Quality Management Systems - Process Validation Guidance. The Global Harmonization Task Force

Medical Device Regulations for Process Validation: Review of FDA, GHTF, and GAMP Requirements

Hypothesis testing - Steps

Statistical Process Control (SPC) Training Guide

3.0 Risk Assessment and Analysis Techniques and Tools

How Far is too Far? Statistical Outlier Detection

Chapter 8 Hypothesis Testing Chapter 8 Hypothesis Testing 8-1 Overview 8-2 Basics of Hypothesis Testing

The use of risk assessment tools for microbiological assessment of cleanroom environments. by Tim Sandle

QUALITY RISK MANAGEMENT

CAPA - the importance of data analysis

ISO 13485:201x What is in the new standard?

Auditing HACCP Programs

Lean Six Sigma Black Belt-EngineRoom

UNICEF s Quality Assurance System for Procurement of Micronutrient Powders (MNP)

Six Sigma Acronyms. 2-1 Do Not Reprint without permission of

ISO Introduction

ISO 14971: Overview of the standard

Design Verification. Introduction

Risk Assessment Tools for Identifying Hazards and Evaluating Risks Associated with IVD Assays

Confidence Intervals for the Difference Between Two Means

THE PROCESS CAPABILITY ANALYSIS - A TOOL FOR PROCESS PERFORMANCE MEASURES AND METRICS - A CASE STUDY

Design of Experiments for Analytical Method Development and Validation

Guidance for Industry

IAS CALIBRATION and TESTING LABORATORY ACCREDITATION PROGRAMS DEFINITIONS

Supplier quality guideline. VOSS Automotive Group Issue Rev00

Failure Analysis Methods What, Why and How. MEEG 466 Special Topics in Design Jim Glancey Spring, 2006

Process Validation: Practical Aspects of the New FDA Guidance

Unit 26 Estimation with Confidence Intervals

What is Process Validation?

Evaluating Current Practices in Shelf Life Estimation

THIS PAGE WAS LEFT BLANK INTENTIONALLY

ORACLE CONSULTING GROUP

Controlling Risks Risk Assessment

STATISTICAL QUALITY CONTROL (SQC)

Hardware safety integrity Guideline

Risk Management in IEC rd Edition. Presented by Alberto Paduanelli Medical Devices Lead Auditor, MHS-UK, TÜV SÜD Product Service

Estimating Software Reliability In the Absence of Data

Overview of Medical Device Design Controls in the US. By Nandini Murthy, MS, RAC

1. What is the critical value for this 95% confidence interval? CV = z.025 = invnorm(0.025) = 1.96

State of Control Over the Lifecycle and Process Validation (New and Legacy Products)

Two-Sample T-Tests Allowing Unequal Variance (Enter Difference)

ICH Q9 Quality Risk Management - an industry view. Peter H. Gough, Eli Lilly and Company

Combination Products. Presented by: Karen S. Ginsbury For: IFF March PCI Pharma

Supplier Quality Management Production Part Approval Process (PPAP) Manual

Chapter 4 and 5 solutions

TÜV FS Engineer Certification Course Being able to demonstrate competency is now an IEC requirement:

Life Sciences Product Development Artifacts Survey Results

THE USE OF STATISTICAL PROCESS CONTROL IN PHARMACEUTICALS INDUSTRY

STANDARD OPERATING PROCEDURES QUALITY ASSURANCE AND QUALITY CONTROL 04

Table of Contents. Preface 1.0 Introduction 2.0 Scope 3.0 Purpose 4.0 Rationale 5.0 References 6.0 Definitions

Some Essential Statistics The Lure of Statistics

Quality System: Design Control Procedure - Appendix

Using DOE with Tolerance Intervals to Verify Specifications

Adverse Impact Ratio for Females (0/ 1) = 0 (5/ 17) = Adverse impact as defined by the 4/5ths rule was not found in the above data.

FINAL DOCUMENT. Implementation of risk management principles and activities within a Quality Management System. The Global Harmonization Task Force

LAB 4 INSTRUCTIONS CONFIDENCE INTERVALS AND HYPOTHESIS TESTING

Non-Inferiority Tests for Two Means using Differences

Keysight Technologies Specifications Guidelines. Technical Overview

IEOR 4404 Homework #2 Intro OR: Deterministic Models February 14, 2011 Prof. Jay Sethuraman Page 1 of 5. Homework #2

Validation and Calibration. Definitions and Terminology

Cartel Electronics. AS 9100 Quality Systems Manual

Biostatistics: DESCRIPTIVE STATISTICS: 2, VARIABILITY

Back Analysis of Material Properties

Final Exam Practice Problem Answers

Quality Risk Management

Criteria for Flight Project Critical Milestone Reviews

ABSORBENCY OF PAPER TOWELS

Data-Base Workflow. From Raw Material Data to Simulation input

Two-Sample T-Tests Assuming Equal Variance (Enter Means)

Building an Effective Supplier Control Program: A review of key program elements & their implementation.

SAFETY MANUAL SIL RELAY MODULE

Transcription:

Choosing the Best Device Sample Size for Verification and Validation Steven Walfish Staff Statistician, Corporate Quality

Agenda Validation scope Statistical tools MSA Design Verification and Validation Other Tools for validation 2

Validation Scope A manufacturer can assure through careful design of the device and packaging, careful design and validation of processes, and process controls, that there is a high probability that all manufactured units will meet specifications and have uniform quality. Where process results cannot be fully verified during routine production by inspection and test, the process must be validated according to established procedures [820.75(a)]. 3

Validation Scope When any of the conditions listed below exist, process validation is the only practical means for assuring that processes will consistently produce devices that meet their predetermined specifications: Routine end-product tests have insufficient sensitivity to verify the desired safety and efficacy of the finished devices; Clinical or destructive testing would be required to show that the manufacturing process has produced the desired result or product. Routine end-product tests do not reveal all variations in safety and efficacy that may occur in the finished devices. 4 The process capability is unknown, or it is suspected that the process is barely capable of meeting the device specifications.

Why Statistics? Manufacturers should: Understand the sources of variation Detect the presence and degree of variation Understand the impact of variation on the process and ultimately on product attributes Control the variation in a manner commensurate with the risk it represents to the process and product. 5

What is commensurate with the risk? Reality in Population Accept Reject Decision Based on Sample Accept Correct Decision Type II Error (b) Consumer Risk Reject Type I Error (a) Producer Risk Correct Decision 6

Identify Critical Quality Attributes Critical quality attributes are usually the responses from the process (i.e. tensile strength) Do not want to select too many or too few. Usually 3-5 is a representative set. Early studies help to determine the best set. 7

Identify Critical Process Factors Critical process factors are usually the inputs of the process (i.e. temperature) Use screening designs to reduce the trivial many to the vital few. Characterize the design space for these critical process parameters. 8

Focus of Characterization Need to know what happens to the process when conditions occur which stress the process Use Worst-case conditions to challenge the process and study how the output changes. This can help to identify measures that will protect the process as process changes occur. 9

Types of Designs Plackett-Burman Screening Design Evaluate n parameters in as little as n+1 batches. Fractional Factorial Design Allow for main-effect only models and some two-way interactions. 10

Screening Design to determine Main Effects Screening Design is based on a Risk Analysis Develop a list of all possible Critical Process Parameters by considering discrete and continuous parameters. Evaluate each parameter for potential impact on the entire process. Perform a detailed review of each parameter to assess relative importance. Assign ranges for parameter evaluation based on process Risk Analysis. 11

How to Select the Best Design Randomization of treatments (runs) helps to distribute the random error over the entire experiment. When doing a fractional factorial, you need to be careful to assign factors to A, B, etc to control for interactions that are under consideration for analysis. 12

How to Select the Best Design When doing a fractional factorial interactions that are not tested directly are assumed to be negligible. This is required since these interactions are confounded with other interactions or main effects. Should you spend more runs identifying main effects or interactions? 13

Why Analyze Measurement Systems? QUANTIFY the amount of observed process variation due to error in the measurement system. EVALUATE: new measurement devices suspected deficient measurement devices impact of repairs on measurement devices new operators involved in the measurement process ASSURE 14 no measurement system error will lead to false acceptance or rejection of product.

Design Controls Design Verification confirms that Design Outputs meet Design Input requirements, Design Verification confirms You Made it Right Design Validation confirms that the product fulfills the user needs and intended use. Design Validation confirms You Made the Right Thing 15

Statistical Building Blocks Utilization of statistics in the design verification and design validation has many different motivations. As a level of confidence that risks are appropriately mitigated as a result of the design and development process. As an input to risk management, statistics can be used to formulate probabilistic models. The confidence level we must asserted through the use statistics are influenced by several factors: Risk Variability 16

Statistical Justification Document if feature is statistical or non-statistical. There should be written criteria for when a test is considered non statistical versus it is considered statistical. Such criteria generally consists of a list of the types of tests considered non statistical. Further, requirements that are statistical but for which recognized standards exist specifying a sample size and acceptance criteria, use of a recognized standard is a valid justification. 17

Statistical Justification Sampling plans consist of both sample size and acceptance criterion. Associated with passing a sampling plan is a confidence statement one can make. A rationale must exist in writing for the sampling plan selected. This rationale should be based on risk to patient. The phrase 95%/90% means 95% confidence (Type I error rate of 5%) of having 90% reliability or conformance to requirements (Type II error rate is 10%). 18

Sample Size for Continuous Data n 2 2 Z Z S 2 You can substitute the t-value for smaller sample sizes. Differs slightly from the method used by software such as JMP. 19

Example Type I error rate is 5% (95% Confidence) Za=1.96 Type II error rate is 1% (99% Power) Zb = 2.326 Unknown Standard Deviation and Delta, but want to detect a 0.5 standard deviation shift. 2-Sided Specification n n 1.96 74 2.326 0.5 2 2 (1) 2 20

Values of Alpha and Beta Confidence Level b 1-Sided Specification a 2-Sided Specification 80% 0.842 0.842 1.282 90% 1.282 1.282 1.645 95% 1.645 1.645 1.960 98% 1.960 1.960 2.241 99% 2.326 2.326 2.576 100% 2.576 2.576 2.807 21

K-Value Sample Size Using tolerance intervals Determine the Confidence Level Determine the percent of population coverage (Reliability) Use a ISO 16269-6 (ISO standard for tolerance intervals) to find the value of K X k * S U X k * S L 22

K Value Example How large a sample do I need to have 95% confidence with 95% reliability for a process with a mean of 10.0 and standard deviation of 0.55? The specification are 8.5 to 12.0 Since the mean is closest to the lower specification From ISO 16269-6, a sample size is approximately 21 1 0 k * 0.55 8.5 23 k 2.727

Attribute C=0 Sample Size (based on all tests being acceptable) At 95% Confidence and 90% Reliability Does not mean 28 devices, but 28 tests (can be on 1 device depending on variability between devices) 24

Setting Validation Acceptance Criteria Confidence Intervals Tolerance Intervals Process Capability 25

Confidence Intervals A Confidence Interval represents a range of values that will most likely include the true population value. The Confidence Interval is derived from the observed sample. The width of the interval is based on: The error of our estimate A multiplier reflecting how sure we want the interval to include the true population value The interpretation of the Confidence Interval is always the same: We are X% confident that the interval includes the true population value 26

Tolerance Intervals Tolerance Intervals are Interval Estimates for individual values in the population. Statement about confidence of coverage. i.e. 95% confident that 95% of the individual values will be within the interval. 27

Process Capability Can set a minimum Cp (Pp) or Cpk (Ppk) for the process. Usually need to use confidence intervals on these estimates to set acceptance criteria and sample size. 28

Example We collected 20 results for tensile strength of a new material used in a medical device. Mean = 8.51N Std Dev. = 0.697 LSL = 5N Min = 7.4 Max=9.5 29

Example Want to test if the MEAN tensile strength is greater than 5N. The 95% Lower Bound on 8.51N is 8.24N. This implies if we sample 20 units from a population whose mean is 8.51N with SD=0.697, the average will be greater than 8.24N 95% of the time. Want to test if the INDIVIDUAL tensile strength is greater than 5N. The 95% confidence with 99% coverage (reliability) is 6.21N. This implies if we sample 20 units from a population whose mean is 8.51N with SD=0.697, the 99% of the individual tensile strength values would be greater than 6.21N. 30

Other Tools for Validation The following is a non-exhaustive list of tools: Failure Mode Effects Analysis (FMEA) Failure Mode, Effects and Criticality Analysis (FMECA) Fault Tree Analysis (FTA) Hazard Analysis and Critical Control Points (HACCP) Hazard Operability Analysis (HAZOP) Preliminary Hazard Analysis (PHA) 31

References GHTF SG3: Quality Management Systems Process Validation Guidance The FDA Quality Systems Regulation (QSR) 32

Conclusions Validation is required when a process cannot be fully verified. Statistical DOE and other methods are used to verify and validate processes. Sample size can be calculated that gives a high probability of meeting the specification. 33

Time for Questions 34