Arterial blood pressure

Similar documents
Hemodynamic Monitoring: Principles to Practice M. L. Cheatham, MD, FACS, FCCM

Direct Arterial Blood Pressure Monitoring Angel M. Rivera CVT, VTS (ECC) Animal Emergency Center Glendale, WI March 2003

Section Two: Arterial Pressure Monitoring

Normal & Abnormal Intracardiac. Lancashire & South Cumbria Cardiac Network

Section Four: Pulmonary Artery Waveform Interpretation

Arterial pressure monitoring Direct arterial pressure monitoring permits continuous measurement of systolic, diastolic, and mean pressures and allows

Normal Intracardiac Pressures. Lancashire & South Cumbria Cardiac Network

Paediatric Intensive Care Unit Nursing Procedure: Care of Arterial Lines.

Measuring central venous pressure

Traumatic Cardiac Tamponade. Shane KF Seal 19 November 2003 POS

Edwards FloTrac Sensor & Edwards Vigileo Monitor. Understanding Stroke Volume Variation and Its Clinical Application

Physics of Invasive Blood Pressure Monitoring

Vascular System The heart can be thought of 2 separate pumps from the right ventricle, blood is pumped at a low pressure to the lungs and then back

Factors Affecting Blood Pressure. Vessel Elasticity Blood Volume Cardiac Output

INTERDISCIPLINARY CLINICAL MANUAL Policy and Procedure

Acute heart failure may be de novo or it may be a decompensation of chronic heart failure.

The EasySense unit can detect that the Smart Q Heart Rate Sensor is connected and the range it is set to.

Note: The left and right sides of the heart must pump exactly the same volume of blood when averaged over a period of time

Haemodynamic Monitoring Learning Package. Name:. HORNSBY KU-RING-GAI HOSPITAL INTENSIVE CARE UNIT

Doppler. Doppler. Doppler shift. Doppler Frequency. Doppler shift. Doppler shift. Chapter 19

Functions of Blood System. Blood Cells

HEART HEALTH WEEK 3 SUPPLEMENT. A Beginner s Guide to Cardiovascular Disease HEART FAILURE. Relatively mild, symptoms with intense exercise

Circulatory System Review

Fellow TEE Review Workshop Hemodynamic Calculations Director, Intraoperative TEE Program. Johns Hopkins School of Medicine

CRITICAL CARE MONITORING: WHAT, WHEN, AND HOW? Susan G. Hackner, BVSc.MRCVS.DACVIM.DACVECC.

Exchange solutes and water with cells of the body

3 rd Russian-Bavarian Conference on Bio-Medical Engineering

Edwards FloTrac Sensor & Edwards Vigileo Monitor. Measuring Continuous Cardiac Output with the FloTrac Sensor and Vigileo Monitor

Vtial sign #1: PULSE. Vital Signs: Assessment and Interpretation. Factors that influence pulse rate: Importance of Vital Signs

SARASOTA MEMORIAL HOSPITAL NURSING PROCEDURE

The pulse Tamás Fenyvesi

Spurious systolic hypertension 143

Evaluation copy. Blood Pressure. Project PROJECT DESIGN REQUIREMENTS

Page 1. Introduction The blood vessels of the body form a closed delivery system that begins and ends at the heart.

SECTION III: HEMODYNAMIC PRINCIPLES

Heart and Vascular System Practice Questions

Blood Vessels and Circulation

Heart Sounds & Murmurs

Cardiovascular Physiology

Nursing Education and Research

INTRODUCTION TO EECP THERAPY

Anatomy and Physiology: Understanding the Importance of CPR

Chapter 20: The Cardiovascular System: The Heart

Blood Pressure. Blood Pressure (mm Hg) pressure exerted by blood against arterial walls. Blood Pressure. Blood Pressure

Introduction to CV Pathophysiology. Introduction to Cardiovascular Pathophysiology

Management of the Patient with Aortic Stenosis undergoing Non-cardiac Surgery

Heart Failure EXERCISES. Ⅰ. True or false questions (mark for true question, mark for false question. If it is false, correct it.

Milwaukee School of Engineering Case Study: Factors that Affect Blood Pressure Instructor Version

MEASURING AND RECORDING BLOOD PRESSURE

Guideline Title: ICU Management of Intra-Aortic Balloon Pump

MECHINICAL VENTILATION S. Kache, MD

Low-gradient severe aortic stenosis with normal LVEF: A disturbing clinical entity

Cardiovascular diseases. pathology

REMOVAL OF A PICC. Possible Cause Nursing Actions Prevention

CHEST TUBES AND CHEST DRAINAGE SYSTEMS

V: Infusion Therapy. Alberta Licensed Practical Nurses Competency Profile 217

Samba Preclin Samba 200. Product Presentation

Monitoring of Cerebral Blood Flow. Transcranial Doppler Laser Doppler Flowmetry Thermal dilution method (Hemedex)

UW MEDICINE PATIENT EDUCATION. Aortic Stenosis. What is heart valve disease? What is aortic stenosis?

Workshop B: Essentials of Neonatal Cardiology and CHD Anthony C. Chang, MD, MBA, MPH CARDIAC INTENSIVE CARE

OEM MAXNIBP Frequently Asked Questions

Update on Small Animal Cardiopulmonary Resuscitation (CPR)- is anything new?

Clinical Training for Visage 7 Cardiac. Visage 7

Heart Murmurs. Outline. Basic Pathophysiology

Electrocardiography Review and the Normal EKG Response to Exercise

GENERAL HEART DISEASE KNOW THE FACTS

Chapter 16. Learning Objectives. Learning Objectives 9/11/2012. Shock. Explain difference between compensated and uncompensated shock

How To Understand What You Know

Common types of congenital heart defects

Dr Bashir Alaour, Dr William English, Royal Cornwall Hospitals NHS Trust, UK Correspondence:

What to Know About. Atrial Fibrillation

Measurement of Blood Pressure

Manual Blood Pressure Measurement

Fundamentals of Critical Care: Hemodynamics, Monitoring, Shock

Potential Causes of Sudden Cardiac Arrest in Children

Catheter insertion of a new aortic valve to treat aortic stenosis

Magnetic Resonance Quantitative Analysis. MRV MR Flow. Reliable analysis of heart and peripheral arteries in the clinical workflow

ELECTRON SPIN RESONANCE Last Revised: July 2007

Diagnostic and Therapeutic Procedures

Overview of the Cardiovascular System

PATIENT INFORMATION GUIDE TO ATRIAL FIBRILLATION

GUIDE TO ATRIAL FIBRILLATION

A PATIENT S GUIDE TO CARDIAC CATHETERIZATION

Chest Pain. Acute Myocardial Infarction: Differential Diagnosis and Patient Management. Common complaint in ED. Wide range of etiologies

NORTH WALES CRITICAL CARE NETWORK

Crash Cart Drugs Drugs used in CPR. Dr. Layla Borham Professor of Clinical Pharmacology Umm Al Qura University

Overdamped system response

Cardiovascular Responses to Exercise

Chapter 16: Circulation

Minimally Invasive Mitral Valve Surgery

Troubleshooting Beyond the Basics. Intra-Aortic Balloon Pump

Anatomi & Fysiologi The cardiovascular system (chapter 20) The circulation system transports; What the heart can do;

Mr GH: Pericardial Window. Anaesthetic Management of Cardiac Tamponade

TRANSPORT OF CRITICALLY ILL PATIENTS

How To Treat A Heart Attack

Starling s Law Regulation of Myocardial Performance Intrinsic Regulation of Myocardial Performance

Transcription:

Arterial blood pressure Just because something is invasive doesn t mean it improves survival. The Mean Arterial Pressure - MAP = (Systemic vascular resistance) x (cardiac output) - Crudely, MAP = (diastolic pressure) + (pulse pressure divided by 3) Why is the MAP more relevant than systolic and diastolic blood pressure? - Least dependent on measurement site or technique (whether invasive or not) - Least altered by damping - Determines tissue blood flow via autoregulation Non-invasive (oscillometric) cuff blood pressure monitoring: - Wildly inaccurate. o Underestimates high blood pressures o Overestimates low blood pressure o For the normal range, the 95% confidence interval is +/- 15mmHg! o In case of arrhythmia, the accuracy is even worse.

Arterial line mechanics Which artery should I puncture? - There does not appear to be any difference in complications, irrespective of site. - Femoral artery is better in low output states (better reflects the aortic pressure). The monitoring device Pressurized bag of saline Screwable Slow Flush Valve Squeezable Fast Flush Valve Phlebostatic axis Midaxillary line, 4 th intercostal space Clamp Fluid filled non-compliant tubing: no more than 1.2 metres Sampling port - The arterial pressure wave travels at 6-10 metres/sec - The cannula in the artery is connected to the transducer via some non-compliant fluid-filled tubing Priming the non-compliant pressure tubing - The idea is that the fluid in the tubing transmits the pressure wave to the transducer - The whole principle rests on a continuous cylinder of saline connecting the artery to the pressure transducer Damping : anything that has a shock absorber effect on the art line - Air bubbles, long tubing, or compliant tubing all of these absorb some of the force of the pulse wave decreasing the amplitude of the oscillations. - Damping results in a slurred waveform with overestimation of the diastolic and underestimation of systolic; however the MAP value is usually preserved. - In contrast, a kinked or clogged art line will see MAP systolic and diastolic all trending towards zero. Zeroing the art line - The system is zered at the phlebostatic axis (otherwise it would read diastolic blood pressures of ~ 760mmHg) - The device is zeroed when the air-fluid interface is opened to atmospheric pressure Leveling the art line - The transducer has to be zeroed at the phlebostatic axis. - If it is zeroed at the level of the cannula, the hydrostatic pressure of the column of blood will be added to the measured blood pressure. For every 10cm below the phlebostatic axis, the art line will add 7.4mmHg of pressure. Which fluid should I flush with? - Heparinized saline increases accuracy, but does not prolong patency. - Constant infusion 3ml/hr; when flushed, flush rate is 30-60ml/hr Pressure Transducer

Normal arterial line waveforms The normal art line waveform mmhg 120 Anacrotic limb Dicrotic limb Dicrotic notch: caused by the closing of the aortic valve and retrograde flow against it. Systolic Pulse pressure 80 Diastolic 0 Area under the graph = MAP Difference in arterial waveforms according to site of insertion Aorta: big elastic vessel; the trace is slurred because it damps Brachial artery Radial artery Femoral artery Dorsalis pedis The further you get from the aorta, - The taller the systolic peak - The further the dicrotic notch The systolic peak is steeper the further down the arterial tree you travel because of reflected waves. That is to say, the narrowing and bifurcation of blood vessels reflects some of the pulse back at the aortic valve. In young people, this is a positive feature, as their relatively elastic vessels recoil slowly and the reflection wave is delayed, arriving to the aortic valve after it closes, and nourishing the coronary arteries.

Respiratory variation of arterial line waveforms Spontaneous respiration Intrathoracic Pressure Zero As intrathoracic pressure decreases on inspiration, the negative pressure is transmitted to the aorta (the thorax sucks the blood back in) and the baseline of the art line trace is displaced down. Art line trace Variation in systolic blood pressure is typically less than 10mmHg. There is a small stroke volume variation (on inspiration, stroke volume is increased because of increased venous return+ In a spontaneously breathing patient, it is normal to observe a downward displacement of the arterial pressure baseline with each inspiration. This phenomenon is exaggerated in patients with cardiac tamponade or restrictive pericarditis. Positive Pressure Ventilation Pressure support Intrathoracic Pressure PEEP Zero Art line trace Variation in systolic blood pressure is greater with mechanical ventilation; Furthermore, there is a variation in stoke volume because of reduced venous return during a positive pressure breath. The more hypovolemic the patient is, the greater the inspiratory decrease of systolic blood pressure and troke volume. In a mechanically ventilated patient, it is normal to observe a decrease in arterial pressure baseline, systolic blood pressure and mean arterial pressure. This phenomenon is exaggerated in patients who are hypovolemic, as it is the influence of the tug-of-war between positive pressure of the ventilator and the diastolic pressure of the venous system. It s a fight: PEEP vs central venous pressure. If the PEEP wins, the right heart doesn t fill enough, consequently the left heart doesn t fill enough, and the stroke volume is decreased as per the Frank-Starling mechanism. If the patient has low central venous pressure (eg. dehydrated or hemorrhaging), PEEP wins more often. Of course all of this goes out the window if the ventricles are abnormal, eg. in LV dysfunction- then, the positive pressure might actually INCREASE the stroke volume, because it augments the transmural pressure of the barely coping ventricle, and assist in ejecting the blood from the thorax.

Abnormal arterial line waveforms Pathological arterial line waveforms Atherosclerosis and hypertension Reflected wave Dicrotic notch Reflected wave - Steep sharp systolic peak - Reflected waves are visible In the atherosclerotic elderly the vessels have poor compliance, and the reflection wave returns early, before the aortic valve closes. Thus, it contributes to afterload. You can sometimes see it on the art line trace. This contribution disappears in vasodilation, and appears with vasoconstriction. The non-compliant vessels do not stretch in response to the systolic pressure, and thus the pressure rises rapidly at the beginning of systole. Aortic stenosis - Slurred gradual systolic peak Abnormally slow systolic ejection gradient In presence of resistance to outflow, the systolic peak will be slow to arrive ( as the left ventricle struggles to squeeze blood past the aortic valve). The dicrotic limb should remain relatively normal. Aortic regurgitation - Abnormally widened pulse pressure Abnormally widened pulse pressure The diastolic will be well below the systolic, with abnormally widened pulse pressure, because the blood regurgitates easily back into the left ventricle as it fills. The compliance of the left ventricle causes the arterial pressure to dip in diastole as some of the pressure is absorbed by the act of left ventricular filling.

Arterial line dynamic response testing Dynamic Response is a function of Natural Resonant Frequency and Damping Coefficient The Natural Resonant Frequency: How fast the system vibrates in response to a pressure signal The Damping coefficient: How quickly those vibrations come to rest in the system The Square Wave Test When you squeeze the fast flush valve, you let the transducer taste some of the 300mmHg in the pressurized saline bag. This produces a waveform that rises sharply, plateaus, and drops off sharply when the flush valve is released again. This is the square wave. The accurate, responsive, adequately damped art line trace The dicrotic notch is clear and distinct The natural resonant frequency is the time between oscillations. It should be around 30 Hz (i.e. 30 times per second) to give an accurate signal at heart rates of around 180 (3 Hz). Systolic Fast flush valve is activated A good art line trace has a distinct dicrotic notch, and after the fast flush test there are two oscillations only. The over-damped art line trace The dicrotic notch is lost, the waveform appears abnormally smooth and curved. MAP Diastolic Two oscillations occur after the flush valve is released ; each oscillation s amplitude must be no more than 1/3 rd of the previous oscillation. Fast flush valve is activated Actual systolic False, underestimated systolic MAP remains the same Diastolic remains the same Just one oscillation after the flush, and then they die off, absorbed in the air bubble or clot or whatever else is overdamping the system The over-damped trace will lose its dicrotic notch, and there wont be more than one oscillation. This happens when there is clot in the catheter tip, or an air bubble in the tubing. The under-damped art line trace False, overestimated systolic Actual systolic Fast flush valve is activated MAP remains the same Diastolic remains the same Ringing, repeated oscillations; the line oscillates like a gong. There is so little damping that these oscillations just wont die, and continue to reverberate through the circuit. The under-damped trace will overestimate the systolic, and there will be many post-flush oscillations.