Evaluation of Targets for Color Calibrating Digital Color Images from an Optical Bright-Field Transmission Microscope

Similar documents
Calibration Slide for Histopathology task force Teleconference 20 February :00 (UK) / 10:00 (EST)

Color Management Terms

Realization of a UV fisheye hyperspectral camera

Multispectral stereo acquisition using 2 RGB cameras and color filters: color and disparity accuracy

AxioCam HR The Camera that Challenges your Microscope

ZEISS Axiocam 506 color Your Microscope Camera for Imaging of Large Sample Areas Fast, in True Color, and High Resolution

Basler. Area Scan Cameras

How To Use An Edge 3.1 Scientific Cmmos Camera

Overview Image Acquisition of Microscopic Slides via Web Camera

AxioCam MR The All-round Camera for Biology, Medicine and Materials Analysis Digital Documentation in Microscopy

Thermo Scientific ArrayScan XTI High Content Analysis Reader. revolutionizing cell biology with the power of high content

High-resolution Imaging System for Omnidirectional Illuminant Estimation

Technical Considerations Detecting Transparent Materials in Particle Analysis. Michael Horgan

Virtual Reality Techniques for the Visualization of Biomedical Imaging Data

NyONE - Cell imaging in a bird s eye view 4. NyONE...resolution matters! 8. Features & benefits 10. Fluorescence excitation channels 12

pco.edge 4.2 LT 0.8 electrons 2048 x 2048 pixel 40 fps :1 > 70 % pco. low noise high resolution high speed high dynamic range

Basler pilot AREA SCAN CAMERAS

Understanding HD: Frame Rates, Color & Compression

USB 3.0 Camera User s Guide

UV-NIR LASER BEAM PROFILER

Overview. Proven Image Quality and Easy to Use Without a Frame Grabber. Your benefits include:

Basler scout AREA SCAN CAMERAS

Spyder 5EXPRESS Hobbyist photographers seeking a simple monitor color calibration solution.

MIRAX SCAN The new way of looking at pathology

Principles of Microscopy and Confocal and Fluorescence Microscopy

ICC Recommendations for Color Measurement

Choosing a digital camera for your microscope John C. Russ, Materials Science and Engineering Dept., North Carolina State Univ.

Datasheet Preliminary

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

APPLICATION NOTE. Basler racer Migration Guide. Mechanics. Flexible Mount Concept. Housing

Implementing ISO12646 standards for soft proofing in a standardized printing workflow

Fast Z-stacking 3D Microscopy Extended Depth of Field Autofocus Z Depth Measurement 3D Surface Analysis

D/N VARI-FOCAL IR CAMERA

Road-Map for Interference Filter based Spectral Sensors

T-REDSPEED White paper

CBERS Program Update Jacie Frederico dos Santos Liporace AMS Kepler

Lecture 20: Scanning Confocal Microscopy (SCM) Rationale for SCM. Principles and major components of SCM. Advantages and major applications of SCM.

Digital Imaging and Communications in Medicine (DICOM) Supplement 145: Whole Slide Microscopic Image IOD and SOP Classes

Are Digital Workflows Difficult?

High speed 3D capture for Configuration Management DOE SBIR Phase II Paul Banks

The camera shall be based upon standardised components and proven technology generally using open and published protocols.

CULTURAL HERITAGE USER GUIDE

Medical Imaging Display Colour Space (mrgb) Teleconference 17 July :00 (UK) / 10:00 (EST)

Basler. Line Scan Cameras

RGB Color Managed Workflow Example

Dome Camera with IR Night Vision

EIZO Monitor CalMAN Setup Guide

product overview pco.edge family the most versatile scmos camera portfolio on the market pioneer in scmos image sensor technology

Inspection and Illumination Systems for Visual Quality Assurance. Large selection Attractive prices Tailored solutions. optometron.

KIP 720 CIS SCANNING SYSTEM WITH ADVANCED KIP REAL TIME THRESHOLDING TECHNOLOGY

A Comprehensive Set of Image Quality Metrics

LED red (-): Measuring value < lower tolerance threshold LED red (+): Measuring value > upper tolerance threshold. Page 1/6

RealScan-S. Fingerprint Scanner RealScan-S

Basler beat AREA SCAN CAMERAS. High-resolution 12 MP cameras with global shutter

LED Fresnel Super Series...1. LED Softlight Series...9. Super LED F Super LED F Super LED Piccolo F Super LED F 10...

Wisconsin Heritage Online Digital Imaging Guidelines Version 2.0 September 2009

KIP 2300 HIGH PRODUCTION CCD SCAN SYSTEM

Color Accurate Digital Photography of Artworks

TIL311 HEXADECIMAL DISPLAY WITH LOGIC

From Pixel to Info-Cloud News at Leica Geosystems JACIE Denver, 31 March 2011 Ruedi Wagner Hexagon Geosystems, Geospatial Solutions Division.

Otis Photo Lab Inkjet Printing Demo

Calibration of a High Dynamic Range, Low Light Level Visible Source

Séries U3-CMOS. Nouvelles. Caméras Numériques USB3. Ultra rapides

Calibrating your monitor

Self-Calibrated Structured Light 3D Scanner Using Color Edge Pattern

D/N VANDAL RES. VARI-FOCAL IR DOME CAMERA

CMOS OV7660 Camera Module 1/5-Inch 0.3-Megapixel Module Datasheet

Network Camera SNC-P1

Revised Minutes Calibration slide for histopathology task force Teleconference 29 August :00 (UK) / 10:00 (EDT)

Chapter 12 Filters for FISH Imaging

Wide format scanning reformatted Small Light Easy to use Portable Self-contained

KS732 Family. Rev KS732. 1/3 CCD Color USB Camera

Untangling the megapixel lens myth! Which is the best lens to buy? And how to make that decision!

Why use ColorGauge Micro Analyzer with the Micro and Nano Targets?

KIP 720 CIS SCANNING SYSTEM WITH ADVANCED KIP REAL TIME THRESHOLDING TECHNOLOGY

The camera shall be based upon standardised components and proven technology generally using open and published protocols.

A pretty picture, or a measurement? Retinal Imaging

MT-EL-H1709M PV Module/Solar String Electro-luminescence (EL) Tester

DSLR Astrophotography

Advances in scmos Camera Technology Benefit Bio Research

SCAMAX 402 I 412 I 422 WE ARE PIONEERING! P R E C I S I O N E N G I N E E R I N G P R O V I D E S SECURITY AND EFFECTIVENESS. SCAMAX powered by InoTec

The Universal DAQ Device. Connect and measure immediately!

FPD IN-LINE SPECTRAL TESTING SOLUTIONS

User's Guide. Pinless Moisture Meter. Model MO257

Color Management Handbook

Scanners and How to Use Them

AS-M5630U Sony 32X Optical Zoom HD 2MP Network Security Camera Module

1. Introduction to image processing

Medical Imaging Display Colour Space (mrgb) Teleconference 16 January :00 (UK) / 10:00 (EST)

White Paper. "See" what is important

Measuring the Point Spread Function of a Fluorescence Microscope

MRC High Resolution. MR-compatible digital HD video camera. User manual

Planetary Imaging Workshop Larry Owens

ENVI Classic Tutorial: Atmospherically Correcting Multispectral Data Using FLAASH 2

D/N VANDAL RES. VARI-FOCAL DOME CAMERA

Improved predictive modeling of white LEDs with accurate luminescence simulation and practical inputs

Color Management Handbook

QUARTZ FA-LIMS. Laboratory Information Management System for Failure Analysis THE ONLY LIMS SYSTEM DESIGNED SPECIFICALLY FOR FAILURE ANALYSIS

Color holographic 3D display unit with aperture field division

Transcription:

Evaluation of Targets for Color Calibrating Digital Color Images from an Optical Bright-Field Transmission Microscope Hong Wei, Michael H Brill, Tae Park DATACOLOR 06/18/2014 Presentation to ISCC Camera

Sources of Color Variability Quality of optics Light source Koehler/aperture Microscope set up Camera set up File type Alterations Quality of sensor Controls Acquisition software Image Acquisition Image Adjustment Color correction Gamma Communication Color Management Image Calibration Monitor Calibration 2

ChromaCal Calibration Slide Calibration slide 3 @ www.chromacal.com

ChromaCal Example of Use Ground truth color from multispectral camera on microscope 2 Camera 1 on Microscope 1 Camera 1 on Microscope 2 Camera 2 on Microscope 2 After CHROMACAL Calibration Before Calibration 4

Scope Develop methodology to assess ChromaCal technology Create reference slide for testing. Evaluate ChromaCal using the reference slide following ICC MIWG guidelines. Provide the microscopy community with potential tools to test their own technologies. 5

Challenges No commercially available test color chart (slide) that is designed specifically for digital microscope images. Insufficient published data on the stain colors. No standard color management solution for digital microscope images. 6

Reference Slide & Development Algorithm 7

Reference Slide Development 4 steps 1) Collect transmission spectra of biological samples with certain stain(s)* by multispectral imaging. 2) Compute the color coordinates of the sample with the stain(s) using a preferred illuminant, such as CIE D50. 3) Group all the found colors together in the color space and delimit from them the color gamut of the stain(s). 4) Identify test color filters that match or close to the stain colors. * Currently, the activity has been limited to Hematoxylin & Eosin stains only. But the same concept can be applied to other stains. 8

Step 1: Transmission Spectra Collection with Multi-spectral Camera Band: 31 bands, 400-700nm, 7nm bandwidth Dynamic Range: 12 bits Pixel format: 1392 x 1040 (1,447,680 transmission spectra) Color Accuracy Samples: 22 H&E slides Total of transmission spectra collected: 31,848,960 9

Step 2&3: Color Coordinates Calculation & H&E Color Gamut H&E Color Gamut L CIE XYZ @D50 a CIE Lab b 31,848,960 a L 10 b

Step 4: Identify Reference Color Filters H&E Stain Reference Slide L Reference a Reference L b b 11 Reference a

Features against ICC Requirements ICC Proposed Requirements* the colors of stains; measurement should be repeatable; Easy scanning of the slide on digital microscope systems. Features of H&E Reference Slide All the 20 colors are within or very close to the H&E stain color gamut; Large patch size (~3.8mm), allowing easy measurement with typical measurement setup; between glass slide and cover glass with optical cement; Large space for barcode; Glass slide, same 3 x1 dimension * Adapted from ICC MIWG (Medical Imaging Working Group) white paper at work http://www.color.org/groups/medical/digital_microscope_test_materials_and_test_methods-v3.pdf 12

ChromaCal Performance Testing 13

Color Performance Test Work Flow test colors, D50 Raw_ E, 2000 Rendered Labs of 20 test colors Multispectral Imaging CIE XYZs (srgb to XYZ, D50) http://www.color.org/groups/medical/digital_microscope_test_materials_and_test_methods-v3.pdf 14

Color Performance Test Work Flow test colors, D50 Cal_ E, 2000 CIE Labs of 20 test colors Multispectral Imaging CIE XYZs after ChromaCal Calibration (D50) http://www.color.org/groups/medical/digital_microscope_test_materials_and_test_methods-v3.pdf 15

ChromaCal Color Accuracy with Tungsten light @ 3000K Microscope: Zeiss Axio Objective: Zeiss Achroplan 20x/0.45 Digital Camera: Lumenera Infinity 2-1R 1.4M pixels CCD @3000K Raw_ E Cal_ E Improved Accuracy Mean 11.46 2.87 75% Max 14.49 9.44 35% 16

ChromaCal Color Accuracy Tungsten with Daylight Filter Added a daylight filter (Kodak 80A) to increase the CCT of the light source from 3200K to 4830K Adapted from http://www.olympusmicro.com/primer/photomicrography/colortemperature.html Without Daylight Filter With Daylight Filter Mean 2.87 2.05 Max 9.44 4.75 17

ChromaCal Color Accuracy Tungsten vs. LED Tungsten w/daylight filter LED Mean 2.05 2.01 Max 4.75 4.64 18

ChromaCal Color Accuracy is sensitive to light source CCT LED Daylight Filter 19

Light Source Color Temperature Image courtesy of Jerry Sedgewick, Imaging & Analysis, LLC 20

ChromaCal Power of Auto White Balance Cal_DE Test color @4800K ChromaCal @4800K Test color @2600K ChromaCal @4800K Test color @2600K ChromaCal @4800K, AWB Mean 2.05 6.77 2.35 Max 4.75 10.23 6.60 21

Summary Developed a H&E stains reference slide containing 20 large color patches. Meets the requirements from ICC MIWG. Evaluated the color Performance of the ChromaCal Image Calibration using this reference slide. Future Work complicity of the H&E color gamut Conduct the similar work on other stains. 22