Total Chlorine Sensor

Similar documents
Ozone Sensor. Model 499A OZ CAUTION CAUTION. Instruction Sheet PN 51A-499AOZ/rev.F January 2007

For additional information, please visit our website at

Trace Dissolved Oxygen Sensor

Dissolved Oxygen Sensors

ph Sensor Model 397 CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY WARNING ATEX DIRECTIVE Instruction Sheet PN 51A-397/rev.

Submersion/Insertion Toroidal Sensor

Sanitary and CIP Toroidal Sensor

ENDURANCE Conductivity Sensors

General Purpose ph/orp Sensors

Submersion/Insertion Toroidal Sensor

Toroidal Conductivity Sensor

General Purpose ph/orp Sensor

DISSOLVED OXYGEN PROBE

OxyGuard. Dissolved Oxygen Probe

Milwaukee USER MANUAL. Milwaukee. Smart DO Meter PORTABLE DISSOLVED OXYGEN METER MODEL: SM600. Authorized Dealer: ISMIL600 11/01

Stainless Steel Single and Dual Circulation Kits

6 H2O + 6 CO 2 (g) + energy

Operator Quick Guide EC SENSOR

Trillium 40 Axis Spring Tensioner Wire Replacement Instructions

Milwaukee USER MANUAL

REPAIR AND MAINTENANCE MANUAL

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

Oceanscience Cable Chimp II Cableway ROV System User Guide and Warranty

INSTALLATION INSTRUCTIONS BL-8704

INSTALLATION INSTRUCTIONS MK-8704

TraceTek TTDM Series Leak Detection and Location Modules Replacement Parts Installation Instructions

Your safety and the safety of others are very important.

Scotchgard Multi-Layer Protective Film 1004

KUSTER ELECTRONIC GAUGES

Munchkin Heat Exchanger Maintenance

INGROUND, ROPED, AND HOLELESS JACK SEAL REPLACEMENT

Multi-Parameter Electrochemical / Optical Water Quality System

Instruction Manual HI Portable Waterproof Microprocessor Dissolved Oxygen Meter

Reliance DuraStar ORDINARY LOCATION INSTALLATION, OPERATING & MAINTENANCE INSTRUCTIONS

IntelliChlor Flow-Temperature Switch Replacement Kit Installation Guide

Waters Corporation. Waters 2690/5 USER & TROUBLESHOOTING GUIDE

AMERBRITE UNDERWATER WHITE LED POOL LAMP

123 Industrial Loop Road Paynesville, MN Phone:

Lenntech psi End port pressure vessels 2.5. User s Guide to: Phoenix Vessel Technology Limited. Model number: 1503

5800 Temperature Sensor Cable Assembly

SCREAMIN' EAGLE "HEAVY BREATHER" PERFORMANCE AIR CLEANER KIT

with installation dynafact boost GAUGE this manual is for use with systems

Professional Plus. Calibration Tips

Care, Maintenance, and Storage of YSI 6-Series Probes and Sondes

MODEL G300 BRAKE BLEEDER

SureSite Magnetic Liquid Level Indicator

THERMOMETER INSTALLATION, OPERATION AND MAINTENANCE

Thermostatic Valve Replacement Cartridge and Rebuild Kit

SMD Rework Station TABLE OF CONTENTS

3. SEISCO PARTS & SERVICE REMOVAL AND REPAIR GUIDE

300 SERIES 331, 332, 333, 344, 356 AND 367 MODELS

ABV20 Air Actuated Boiler Blowdown Valve Installation and Maintenance Instructions

Chill 4 Hole Bath Shower Mixer

ATS Overhead Table Shelf System INSTRUCTION MANUAL

3M Water Filtration Products. CFS01 Series, CFS02 Series, CFS11 Series, and CFS12 Series Housings. Installation and Instruction Manual For:

Fisher 249 Caged Displacer Sensors

Coolflo Safe Touch Shower Mixing Valve ST20015CP

USER MANUAL. 10 GPM Small Single Line Cryogenic Coupling

Electrical Connection. Provide Proper Electrical Supply. Electrical Requirements. Connecting using the supplied cord and plug

Rosemount 5300 and 5400 Series Terminal Compartment Spare Part Instruction

Chloride and Salinity

User's Guide. ExStik DO600. Dissolved Oxygen Meter RECALL. Dissolved Oxygen DO600

ETL listed for installations within 5 ft. (1.5M) of outer edge of water Rev E Page 1 of 10

Universal Installation and Maintenance Instructions 71CD-1212 Spill Containment Bucket

Carpet Washer. vax.co.uk VRS5W. Vax Careline: (UK) (ROI) Vax model number: Version 1.0

PART 1 - INTRODUCTION...

Wiper Motor Marinco 2.5. Installation Instructions

Wind Direction Smart Sensor (S-WDA-M003)

DTM04 TANK MONITOR DTM08 TANK MONITOR Dtm12 TANK MONITOR. Installation and Operation Manual

Manual for GlobePharma Mini-Press II Rotary Tablet Press

Total solder points: 129 Difficulty level: beginner advanced LIQUID LEVEL CONTROLLER K2639 ILLUSTRATED ASSEMBLY MANUAL H2639IP-1

High Purity ph Sensor

4-600, 1000 psi. User s Guide to: Phoenix Vessel Technology Limited. End port pressure vessels Model numbers: 1386, 1384.

Installing a Probe. Typical Installation

Pipe Valves, Fittings and Nipples Pressures to 15,200 psi

Instruction Manual. Image of SP-3015 & SP Important Safeguards. Automatic Dispensing Hot Water Pot with Reboil Function

Models SFP33 & SFP32 Pressure Transmitters

Weatherproof Extension Case (Part # S-EXT-CASE)

INSTALLATION & OPERATING INSTRUCTIONS

Installation Guidelines: Resistance Thermometers and Temptran Thermometers for Building Automation Systems

STANDARD OPERATING PROCEDURE Dissolved Oxygen Measurement, Meter Calibration and Maintenance

Fleck Service Manual INSTALLATION AND START-UP PROCEDURE TABLE OF CONTENTS JOB SPECIFICATION SHEET

User Guide Dissolved Oxygen Probes

CappController Pipette Controller. Operation Manual Version:

For exploded diagram and part number information, refer to the Spare Parts Catalog available on our website at

ENGINE FUEL FUEL FILTER... FUEL HEATER... INJECTOR... SUPPLY PUMP... COMMON RAIL... FUEL PRESSURE LIMITTER...

Installation and Operating Instructions Installation Instructions for SS EPE-316L Series

AquaMaster. Installation Manual

Support Cleaning Apparatus. Operation, Cleaning, & Safety Manual

Outlines: Original Artwork: LeMaitre. Light Post Adapter. Light Post Adapter. Fits Olympus, Storz, and Stryker Equipment* English Instructions for Use

2006 JUDY SERVICE GUIDE

CoolWave r 2 Phase Control Board and Cable Replacement Kits

OWNER S MANUAL. FOOTWORKS tm Faucet Controller. Foot Operated Faucet Control Valves for Commercial Sinks

GeniJet Pipeless Motor Spa Installation

Fluid Level Gauge Fluid Level Sensor Temperature Switch FSA / FSK / TS

AMX300 DirectConnect Mixing Valve Kits

2100 AD Mirror Elevator Ball Nut Replacement Procedure

AU680 Every Other Week or Every 3,000 Samples (ISE) Maintenance. AU680 Weekly (Analyzer and ISE) Maintenance

Powers Controls TH 192 S Single Temperature Room Thermostat

Unique DV-ST UltraPure - Manuel: for valve sizes DN8 to DN80 (¼ to 3 )

Transcription:

Instruction Sheet PN 51A-499ACL02/rev.G May 2007 Model 499A CL-02 Total Chlorine Sensor For additional information, please refer to the Instruction Manuals CD shipped with this product, or visit our website at www.emersonprocess.com/raihome/liquid/. SPECIFICATIONS - SENSOR Pressure: 0 to 65 psig (101 to 549 kpa abs) Temperature: 32 to 122 F (0 to 50 C) Process Connection: 1 inch MNPT Wetted Parts: Noryl 1, Viton 2, silicone Cathode: gold 1 Noryl is a registered trademark of General Electric. 2 Viton is a registered trademark of DuPont Performance Elastomers. CAUTION SENSOR/PROCESS APPLICATION COMPATIBILITY The wetted sensor materials may not be compatible with process com position and operating conditions. Application compat ibility is entirely the responsibility of the user. CAUTION Do not exceed pressure and temperature specifications. Pressure: 65 psig max (549 kpa abs max) Temperature: 32 to 122 F (0 to 50 C) INSTALLATION The sensor is intended for use only in the SCS921, SCS921A, or TCL Sample Conditioning System for total chlorine. Consult instruction manual for the sample conditioning system for more information.

MODEL 499A CL-02 WIIRNG & MAINTENANCE WIRING When making connections through a junction box (PN 22719-02), wire point-to-point. NOTE: Use a wire nut and pigtail (included) when connecting several wires to the same terminal. FIGURE 1. Pin Out Diagram for Model 499ACL-01- VP Sensor (top view of connector end of sensor) FIGURE 2. Wiring Sensor with Standard Cable to Model 1055 Analyzer FIGURE 3. Wiring Sensor with Optimum EMI/RFI or Variopol Cable to Model 1055 Analyzer FIGURE 4. Wiring Sensor with Standard Cable to Model 1056 Analyzer FIGURE 5. Wiring Sensor with Optimum EMI/RFI or Variopol Cable to Model 1056 Analyzer 2

MODEL 499A CL-02 MAINTENANCE FIGURE 6. Wiring Sensor with Standard Cable to Model 54eA Analyzer FIGURE 7. Wiring Sensor with Optimum EMI/RFI or Variopol Cable to Model 54eA Analyzer FIGURE 8. Wiring Sensor with Standard Cable to 5081-A Transmitter. FIGURE 9. Wiring Sensor with Optimum EMI/RFI or Variopol Cable to 5081-A Transmitter. FIGURE 10. Wiring Sensor with Standard Cable to Xmt-A Transmitter. FIGURE 11. Wiring Sensor with Optimum EMI/RFI or Variopol Cable to Xmt-A Transmitter. 3

MAINTENANCE GENERAL When used in clean water, the total chlorine sensor requires little maintenance. Generally, the sensor needs maintenance when the response becomes sluggish or noisy or when readings drift follow calibration. Maintenance frequency is best determined by experience. If the sensor is a used in potable water, expect to clean the membrane every month and replace the membrane and electrolyte solution every three months. Sensors used in dirty water require more frequent maintenance and calibration. However, if experience shows the sensor is holding calibration and not drifting appreciably between calibration intervals, the maintenance interval can be extended. CAUTION: PRESSURIZED SPRAY INJURY Before removing the sensor from the process stream for maintenance, be sure the process pressure is reduced to 0 psig and the process temperature is at a safe level! CAUTION Fill solution may cause irritation. May be harmful if swallowed. Read and follow manual. CLEANING THE MEMBRANE. Keep the membrane clean. Clean the membrane with water sprayed from a wash bottle. Use a soft tissue to gently wipe the membrane. REPLACING THE MEMBRANE. 1. Hold the sensor with the membrane facing up. 2. Unscrew membrane retainer. Remove the membrane assembly and O-ring. See Figure 12. 3. Inspect the cathode. If it is tarnished, clean it by gently rubbing in the direction of the existing scratches (do not use a circular motion) with 400-600 grit silicon carbide finishing paper. Rinse the cathode thoroughly with water. 4. Put a new O-ring in the groove. Hold the membrane assembly with the cup formed by the membrane and membrane holder pointing up. Fill the cup with electrolyte solution. 5 Next, place a drop of electrolyte solution on the cathode. Invert the membrane assembly and place it over the cathode stem. 6. Screw the membrane retainer back in place. 7. Hold the sensor with the membrane pointing down. Shake the sensor a few times, as though shaking down a clinical thermometer. REPLACING THE ELECTROLYTE SOLUTION AND MEMBRANE. 1. Unscrew the membrane retainer and remove the membrane assembly and O-ring. See Figure 12. 2. Hold the sensor over a container with the cathode pointing down. 3. Remove the fill plug and allow the electrolyte solution to drain out. 4. Wrap the plug with several turns of pipe tape and set aside. 4 5. Prepare a new membrane. Hold the membrane assembly with the cup formed by the membrane and membrane holder pointing up. Fill the cup with electrolyte solution. See PREPARATION OF FILL SOLUTION. 6. Hold the sensor at about a 45-degree angle with the cathode end pointing up. Add electrolyte solution through the fill hole until the liquid overflows. Tap the sensor near the threads to release trapped air bubbles. Add more electrolyte solution if necessary. 7. Place the fill plug in the electrolyte port and begin screwing it in. After several threads have engaged, rotate the sensor so that the cathode is pointing up and continue tightening the fill plug. Do not overtighten. 8. Place a new O-ring in the groove around the cathode post. Cover the holes at the base of the cathode stem with several drops of electrolyte solution. 9. Insert a small blunt probe, like a toothpick with the end cut off, through the pressure equalizing port. See Figure 12. NOTE Do not use a sharp probe. It will puncture the bladder and destroy the sensor. Gently press the probe against the bladder several times to force liquid through the holes at the base of the cathode stem. Keep pressing the bladder until no air bubbles can be seen leaving the holes. Be sure the holes remain covered with electrolyte solution. 10. Place a drop of electrolyte solution on the cathode, then place the membrane assembly over the cathode. Screw the membrane retainer in place. 11. The sensor may require several hours operating at the polarizing voltage to equilibrate after the electrolyte solution has been replenished.

FIGURE 12. Sensor Parts SPARE PARTS 33523-00 Electrolyte Fill Plug 9550094 O-Ring, Viton 2-014 33521-00 Membrane Retainer 23501-02 Total Chlorine Membrane Assembly: includes one membrane assembly and one O-ring 23502-02 Total Chlorine Membrane Kit: includes 3 membrane assemblies and 3 O-rings 9210438 Total Chlorine Sensor Fill Solution, 4 oz (120 ml) 1 Viton is a registered trademark of DuPont Performance Elastomers. 5

6 NOTES:

NOTES: 7

The right people, the right answers, right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE http://www.raihome.com Specifications subject to change without notice. Credit Cards for U.S. Purchases Only. Emerson Process Management Liquid Division 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250 http://www.raihome.com Rosemount Analytical Inc. 2007