Purity Determinations By Differential Scanning Calorimetry 1



Similar documents
Experiment 1: Colligative Properties

Thermochemistry. r2 d:\files\courses\ \99heat&thermorans.doc. Ron Robertson

Transfer of heat energy often occurs during chemical reactions. A reaction

Bomb Calorimetry. Example 4. Energy and Enthalpy

Experiment 12E LIQUID-VAPOR EQUILIBRIUM OF WATER 1

MOLECULAR WEIGHT BY BOILING POINT ELEVATION

CHEM 105 HOUR EXAM III 28-OCT-99. = -163 kj/mole determine H f 0 for Ni(CO) 4 (g) = -260 kj/mole determine H f 0 for Cr(CO) 6 (g)

Figure 1. A typical Laboratory Thermometer graduated in C.

TA INSTRUMENTS DIFFERENTIAL SCANNING CALORIMETER (DSC) Insert Nickname Here. Operating Instructions

FULL PAPER Standardization of PCM Characterization via DSC

DETERMINING THE ENTHALPY OF FORMATION OF CaCO 3

Experiment 2 Kinetics II Concentration-Time Relationships and Activation Energy

SOLUBILITY, IONIC STRENGTH AND ACTIVITY COEFFICIENTS

vap H = RT 1T 2 = kj mol kpa = 341 K

Thermochemistry: Calorimetry and Hess s Law

Experiment 13: Determination of Molecular Weight by Freezing Point Depression

Graphite Furnace AA, Page 1 DETERMINATION OF METALS IN FOOD SAMPLES BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROSCOPY (VERSION 1.

States of Matter CHAPTER 10 REVIEW SECTION 1. Name Date Class. Answer the following questions in the space provided.

In order to solve this problem it is first necessary to use Equation 5.5: x 2 Dt. = 1 erf. = 1.30, and x = 2 mm = m. Thus,

IDEAL AND NON-IDEAL GASES

Physical Properties of a Pure Substance, Water

Phase Transitions and Differential Scanning Calorimetry

Freezing Point Depression: Why Don t Oceans Freeze? Teacher Advanced Version

Bomb Calorimetry. Electrical leads. Stirrer

DSC Differential Scanning Calorimeter

KINETIC DETERMINATION OF SELENIUM BY VISIBLE SPECTROSCOPY (VERSION 1.8)

The first law: transformation of energy into heat and work. Chemical reactions can be used to provide heat and for doing work.

Thermodynamics of Mixing

Organic Chemistry Calculations

Performing Calculatons

CHEM 36 General Chemistry EXAM #1 February 13, 2002

EXPERIMENT 12: Empirical Formula of a Compound

Chapter 2 Chemical and Physical Properties of Sulphur Dioxide and Sulphur Trioxide

= atm. 760 mm Hg. = atm. d. 767 torr = 767 mm Hg. = 1.01 atm

Gibbs Free Energy and Chemical Potential. NC State University

The Analytical Balance

Chapter 11 Properties of Solutions

AP CHEMISTRY 2007 SCORING GUIDELINES. Question 2

AAHS-CHEMISTRY FINAL EXAM PREP-REVIEW GUIDE MAY-JUNE 2014 DR. GRAY CLASS OF 2016

Determination of the enthalpy of combustion using a bomb calorimeter TEC

Experiment 6 Coffee-cup Calorimetry

Apparatus error for each piece of equipment = 100 x margin of error quantity measured

To calculate the value of the boiling point constant for water. To use colligative properties to determine the molecular weight of a substance.

Fractions and Linear Equations

Molar Mass of Butane

Evaluation of Quantitative Data (errors/statistical analysis/propagation of error)

Experiment 6 ~ Joule Heating of a Resistor

Experiment 1: Adiabatic Bomb Calorimeter (Dated: September 27, 2010)

PURIFICATION TECHNIQUES

RESULTS OF ICARUS 9 EXPERIMENTS RUN AT IMRA EUROPE

Thermodynamics. Thermodynamics 1

Distillation vaporization sublimation. vapor pressure normal boiling point.

Gas Laws. The kinetic theory of matter states that particles which make up all types of matter are in constant motion.

Thermochemical equations allow stoichiometric calculations.

Chapter Test B. Chapter: Measurements and Calculations

DETERMINATION OF THE HEAT STORAGE CAPACITY OF PCM AND PCM-OBJECTS AS A FUNCTION OF TEMPERATURE. E. Günther, S. Hiebler, H. Mehling

Differential Scanning Calorimetry DSC

Determination of the enthalpy of combustion using a bomb calorimeter TEC. Safety precautions

The Determination of an Equilibrium Constant

11 Thermodynamics and Thermochemistry

Reaction Rates and Chemical Kinetics. Factors Affecting Reaction Rate [O 2. CHAPTER 13 Page 1

Using Excel (Microsoft Office 2007 Version) for Graphical Analysis of Data

Experiment 5: Phase diagram for a three-component system (Dated: April 12, 2010)

Experiment 3 Limiting Reactants

Combustion Calorimetry

EXPERIMENT 5. Molecular Absorption Spectroscopy: Determination of Iron With 1,10-Phenanthroline

ORGANIC LABORATORY TECHNIQUES NEVER distill the distillation flask to dryness as there is a risk of explosion and fire.

The Empirical Formula of a Compound

ACID-BASE TITRATIONS: DETERMINATION OF CARBONATE BY TITRATION WITH HYDROCHLORIC ACID BACKGROUND

Pre-Lab Notebook Content: Your notebook should include the title, date, purpose, procedure; data tables.

Using Freezing-Point Depression to Find Molecular Weight

48 Practice Problems for Ch Chem 1C - Joseph

Austin Peay State University Department of Chemistry CHEM Empirical Formula of a Compound

CHEMISTRY STANDARDS BASED RUBRIC ATOMIC STRUCTURE AND BONDING

a) Use the following equation from the lecture notes: = ( J K 1 mol 1) ( ) 10 L

4.2 Bias, Standards and Standardization

Adiabatic Bomb Calorimetry

Phase diagram of water. Note: for H 2 O melting point decreases with increasing pressure, for CO 2 melting point increases with increasing pressure.

Thermodynamics and Equilibrium

Chemistry 13: States of Matter

Method To Solve Linear, Polynomial, or Absolute Value Inequalities:

Exp 13 Volumetric Analysis: Acid-Base titration

Prentice Hall. Chemistry (Wilbraham) 2008, National Student Edition - South Carolina Teacher s Edition. High School. High School

The Point-Slope Form

Fractional Distillation and Gas Chromatography

POURING THE MOLTEN METAL

Chemistry 212 VAPOR PRESSURE OF WATER LEARNING OBJECTIVES

Determination of Molar Mass by Boiling Point Elevation of Urea Solution

Experiment #4, Ohmic Heat

Everest. Leaders in Vacuum Booster Technology

Heterogeneous Catalysis and Catalytic Processes Prof. K. K. Pant Department of Chemical Engineering Indian Institute of Technology, Delhi

Melting Point Determination Application Note #1

EXPERIMENT 7 Reaction Stoichiometry and Percent Yield

We will study the temperature-pressure diagram of nitrogen, in particular the triple point.

Table of content. L81/RITA high speed Thermo Balance. Quattro Dilatometer. L75/1250/B/S Macro Dilatometer. New air cooled furnace program

The Determination of an Equilibrium Constant

Name Class Date. In the space provided, write the letter of the term or phrase that best completes each statement or best answers each question.

Standards Development For Differential Scanning Calorimetry

Chem 420/523 Chemical Thermodynamics Homework Assignment # 6

(S&G 5th ed. Expt 27, 6th, 7th & 8th eds. Expt 26)

Transcription:

Purity Determinations By Differential Scanning Calorimetry 1 Purpose: Determine the purity of a compound using freezing point depression measurements with a differential scanning calorimeter. Prelab: The instructions for the Perkin Elmer DSC 4 are given at the end of this manual. Introduction A common problem in analytical chemistry is the determination of the purity of substances. This laboratory uses differential scanning calorimetry to determine the purity of organic compounds. This method depends only on the physical-chemical behavior of the compound, and no reference standard is necessary. The pure melting point of the sample does not even need to be known. The method is accurate for samples over 98% pure, but it does not measure impurities which are soluble in the solid phase or insoluble in the melt. It is also inapplicable to chemicals which decompose at their melting points or have inordinately high vapor pressures. This technique is particularly valuable in the pharmaceutical and basic chemicals industry. Differential Scanning Calorimetry Differential scanning calorimetry (DSC) monitors heat effects associated with phase transitions and chemical reactions as a function of temperature. In a DSC the difference in heat flow to the sample and a reference at the same temperature, is recorded as a function of temperature. The reference is an inert material such as alumina, or just an empty aluminum pan. The temperature of both the sample and reference are increased at a constant rate. The heat flow difference dh dt = ( dh dt ) sample - (dh dt ) reference 1 can be either positive or negative. Here dh/dt is the heat flow measured in mcal sec -1. In an endothermic process, such as most phase transitions, heat is absorbed and, therefore, heat flow to the sample is higher than that to the reference. Hence dh/dt is positive. N 2 Inlet Sample Resistance Heater Temperature Sensors Reference Resistance Heater Sample Reference Sample Reference Side View (without furnace block) Furnace Block Top View (cover off) Figure 1. Differential Scanning Calorimeter.

Purity by DSC 2 The calorimeter consists of a sample holder and a reference holder as shown in Figure 1. Both are constructed of platinum to allow high temperature operation. Under each holder is a resistance heater and a temperature sensor. Currents are applied to the two heaters to increase the temperature at the selected rate. The difference in the current to the two holders, necessary to maintain the holders at the same temperature, is used to calculate dh/dt. A flow of nitrogen gas is maintained over the samples to create a reproducible and dry atmosphere. The nitrogen atmosphere also eliminates air oxidation of the samples at high temperatures. The sample is sealed into a small aluminum pan. The reference is usually an empty pan and cover. The pans hold up to about 10 mg of material. Melting produces a positive peak in a DSC thermogram. The integral under the DSC peak, above the baseline, gives the total enthalpy change for the phase transition: ( dh dt ) dt = H sample 2 sample This integral can easily be done using computer systems. The resulting value is usually presented in arbitrary units, which can be called planimeter units. (A planimeter is an old-fashioned mechanical device for determining areas.) Since the integral is in arbitrary units, a calibration must be performed with a known weight of a substance with a well-known enthalpy of melting. Indium metal is used for this purpose. Then the enthalpy of the process under study is determined from: H sample = H indium sample area in planimeter units indium area in planimeter units 3 Theory The purity of a sample for which only the molecular weight and sample weight are known can be ascertained by a simple mathematical treatment of the data from a DSC scan. The computation is based on the freezing point depression colligative property equation 2 : T = RT*2 fus H x B and K f = RT*2 4 fus H where T is the melting point depression; R is the gas constant, 1.9872 cal mole -1 deg -1 ; T* is the melting point of the pure solvent; fus H is the enthalpy of fusion per mole; x B is the molality of the solute; and K f is the the molal freezing point depression constant. Equation 4 can be rearranged to give the form most convenient for use in DSC work: fush Mol % impurity = 100 RT* 2 T = 100 1 K T 5 f where T* is the melting point of a sample with zero impurity. With this equation, the amount of impurity (or solute) in solution can be determined by multiplying T by the constant, fus H/RT* 2. In molecular weight determinations, K f must be determined for the solvent. Extensive tables of K f values are available. In purity measurements, the pure compound is the

Purity by DSC 3 solvent, and the constant, which is different for each substance to be determined, must be computed from the DSC curve. The area under the DSC melting curve gives the enthalpy of fusion for the sample and the enthalpy of fusion per mole is readily computed from the weight of the sample. T* and T are obtained from the DSC data. Analysis of a melting curve is best understood by looking at the reverse process, freezing. In Figure 2, the left-hand box represents a sample with zero impurity with the melting point of T*. The next box shows a real sample with X concentration of impurity, and its melting point, T 1. The difference between T* and T 1 is the melting point depression, T, for the sample. 0 X 2X 3X T* T1 T2 T3 T* T1 T = slope T2 T3 1 2 3 1/Fraction Melted Figure 2. Freezing point depression for liquid-soluble, solid-insoluble impurities. Two assumptions inherent in colligative property calculations are now used: the impurity is insoluble in the solid phase and soluble in the melt. Under these conditions, it is evident that when the sample is half frozen (melted), all the impurity is in the liquid phase, which is now reduced to half its former volume, and the impurity concentration is increased to 2X. Because concentration and melting point depression are linearly related, the depression at this point is twice that of the original sample (T* -T 2 = 2 T). In the last box in Figure 2, one third of the sample is liquid, the concentration of impurity in the liquid phase is 3X, and the melting point depression is 3 T. The increase in impurity concentration is inversely proportional to the fraction melted. The inverse of the fraction melted is plotted with the corresponding temperature as the dependent variable. A straight line results whose slope is the melting point depression, T, and whose zero intercept is the extrapolated melting point, T*, of the theoretically pure substance. As a solid melts, the impurities dissolve in the initial liquid phase and the same relationship between fraction melted and temperature depression is obtained. The melting curve is better suited for analysis than the freezing curve, as the former is easily done under equilibrium

Purity by DSC 4 conditions, whereas freezing introduces the problem of supercooling, a nonequilibrium condition unsuitable for calculation. Example Calculation An example of a DSC melting curve is shown in Figure 3. The sample size was 2.2850 mg with a molecular weight of 263.12 g mol -1. Because the amount of heat absorbed is directly proportional to the amount of sample melted and to the area under the melting curve, fractions of these values may be used interchangeably. The DSC melting curve is divided into a series of partial areas, Figure 3. Each partial area extends from the onset of melting to a chosen temperature. These temperatures are chosen at roughly equally spaced intervals up to the point where the sample is half melted. Beyond this point, about 2/3 up the peak, melting is too rapid for equilibrium conditions to prevail and equation 5 does not hold. The fraction melted at any temperature is equal to the fraction of the total area occurring below that temperature. The partial areas extracted from Figure 3 are given in Table I. The inverse of the fraction melted can be obtained directly by dividing the total area by the partial area. These inverses are plotted against their corresponding temperatures (Tb,i), Figure 4 (a). 3 2. 5 2 f u s H mcal/sec 1. 5 1 0. 5 3 0 8. 5 9 0 K 0 3 0 6 3 0 7 3 0 8 3 0 9 T ( K ) Figure 3. Example of a DSC Melting Curve; Methyl-4-(2,4-dichlorophenoxy)butyrate. The partial areas are bounded at the high temperature side with a line matching the slope of the low temperature side of the indium melting curve. Table I. Example Data from Figure 3. Temperature, K (At baseline) Area Sum 1/F Area sum + 15% correction 1/F (corrected) 308.119 52 55.2 483 6.84 308.256 146 19.7 577 5.72 308.422 371 7.74 802 4.12 308.590 781 3.68 1212 2.72 2871 3302

Purity by DSC 5 308.9 308.8 308.7 308.6 308.5 308.4 308.3 308.2 308.1 308 0 20 40 60 1/F 308.9 308.8 308.7 308.6 308.5 308.4 308.3 308.2 308.1 a. b. 0 2 4 6 8 Figure 4. (a.) Data from Table I before correction. (b.) Data after 15% correction, with curve fit giving the slope, 0.112 K, and intercept (=T*), 308.89 K. 1/F In the case of very pure samples (over 99.99 mol %), the plot is linear. Curvature is observed in the case of less pure compounds. As the sample becomes less pure, its melting range increases and the point where melting actually begins is difficult to observe. A correction must be made for this pre-melting 3. For convenience, this correction is expressed as a percentage of the total area measured, as other samples from the same lot will have a similar correction. The data are tabulated in Table I, where the first column records the temperature of each base line intercept, and the second records the total area up to the corresponding temperature. The next column lists the reciprocal of the fraction melted (l/f) at each temperature. These values are plotted against temperature in Figure 4(a). Because the line thus obtained is curved, a correction for pre-melting is applied, which was obtained by trial-and-error to minimize the uncertainty of the fit parameters. In this case a 15% correction was found necessary to linearize the plot, Figure 4(b). When this 15% (431) of the total measured area (2871) is added to each area, the corrected areas tabulated in Table I are obtained. The new total area (3302) divided by the corrected partial areas gives a series of corrected l/f values. The linear plot results when the corrected l/f values are plotted with corresponding temperatures as the dependent variable. The zero intercept, T*, is 308.89 K, and the slope, T, is 0.112 K. To calibrate the instrument a 99.999% pure indium metal standard was run. The enthalpy of fusion of indium is 6.80 cal g -1. In this example the area under the indium melting peak was 793 area units (mcal sec -1 C) for a sample weighing 2.3218 mg. The number of mcal per area units is then 6.80 mcal mg 2.3218 mg 793 area units = 0.0199 mcal area units The enthalpy of fusion of the sample is obtained from the corrected area as follows: 0.0199 mcal area units 3302 area units 2.2850 mg 263.12 mg millimole = 7567 mcal millimole = fus H

Purity by DSC 6 Insertion into Equation 5 gives: 100 7567 cal mol -1 0.112 K 1.9872 cal K -1 mol -1 (308.89 K) 2 = 0.447 mole % impurity Data Analysis Using Logger Pro The traditional way to analyze the data in this experiment is to integrate the partial areas by hand using a planimeter. However, computerized data acquisition allows many of the calculations to be automated. The calculation of the partial areas can be greatly simplified using the built in spreadsheet facility in Logger Pro. Consider the baseline subtracted thermogram as a series of data points at temperature Ti with power Pi, Figure 5. 3 2. 5 2 P i mcal/sec 1. 5 1 0. 5 0 3 0 6 3 0 7 3 0 8 T b,i T i T Figure 5. Baseline subtracted thermogram. The desired corrected area is the shaded portion. To find the partial area up to a temperature Ti, the area of the triangle whose side has a slope equal to the slope of the melting transition for indium is subtracted. The indium slope is used to delineate portions of the melting curve integral because it corrects for thermal lag inherent in the system 4. The area of the triangle is ½ height x base: corrected areai = integrali - Pi(Tb,i - Ti ) 2 (1) Let m be the slope of the melting transition for indium in units of mcal/sec/ C. The formula for the line with slope m that passes through the data point (Ti,Pi) is: P = m T + (Pi m Ti) (2) (Verify, using Eq. 2, that when T = Ti that P = Pi.) The temperature Tb,i is the intersection of this line with the baseline at P = 0 :

Purity by DSC 7 0 = m Tb,i + (Pi m Ti) (3) Solving for Tb,i gives: Tb,i Ti = -Pi m or Tb,i = Ti - Pi m (4) Substitution back into Eq. 1 gives corrected areai = integrali - Pi2 2m This simple result makes the calculations much easier to implement. The purity.cmbl method for Logger Pro has columns to do these calculations. However, the value of the slope m needs to be input. The default value of 300 mcal/sec/ C must be replaced in the Iadj and Tb columns by your slope, as given below in the Procedure section. Then you can copy and paste the Iadj and Tb columns into Excel to complete your calculations. The calculations of the partial integrals by hand as given above results in just a few data points (four in the above example). Using computer based integration gives a partial integral at every data point. There fore your plot will have many data points instead of just four, giving correspondingly better curve fitting results. Procedure: The instructions for the Perkin Elmer DSC 4 are given at the end of this manual. 1. Indium Calibration: Indium, 99.999%, is used for both temperature and heat calibration. The melting point is 156.60 C and the enthalpy of fusion is 6.80 cal g -1. Area calibration is accomplished by measuring the area corresponding to the melting of a weighed sample of indium and indium's enthalpy of fusion as discussed above. Temperature calibration is accomplished by noting the maximum in the indium melting peak. The difference in the literature and measured melting temperature is used to correct the final pure melting point determined below. Weigh a pan and lid to the nearest 0.01 mg. Add 5-7 mg of indium metal and reweigh. The samples are weighed into the pan before sealing, as metal from the pan is occasionally lost by stripping during sealing. Determine the melting curve from 154-156 o C, at a rate of 0.2 C min -1. Use the DSCautostart.cmbl as usual. Remember to set the starting temperature and scan rate into the temperature column of Logger Pro. Start with a range of 2 mcal sec -1, however, the appropriate setting will depend on the size of the sample. Subtract the baseline from your thermogram before doing the integral. Integrate the curve plotted as a function of temperature. Normally you would integrate the curve plotted as a function of time, but the integrals for the purity determination are done on the plot of power verses temperature. We need to do both integrals in the same way. The units of the integral will be (mcal sec -1 C) and the units will cancel in the final results. Determine the slope of the low temperature side of the indium melting peak. In Logger Pro, pull down the Analyze menu and choose Tangent. Moving the mouse on the potential vs. temperature curve will list out the slope. Record this value.

Purity by DSC 8 2. Purity Determination: Weigh a pan and lid and add 2-3 mg of the sample to be determined. Reweigh, and remember to record the weights to the nearest 0.01 mg. Determine the melting curve at a 1 C min -1 program rate to determine the approximate melting point of the sample. Then determine the melting curve from about 1-2 C below the melting point to just above the melting point at a scan rate of 0.2 C min -1. For this second run use the purity.cmbl method. Remember to set the starting temperature and scan rate into the temperature column of Logger Pro. Also enter the slope of the indium calibration curve into the Iadj and Tb columns. For example, if the slope is 34.3 mcal sec -1 C then for Iadj: and for Tb: integral("subtracted Potential", "Temperature")-0.5*"Subtracted Potential"^2/34.3 "Temperature"-"Subtracted Potential"/34.3 Use the same range setting as you used for the indium calibration, 2 mcal sec -1, at the slower scan rate. After the run is complete, subtract the baseline in the normal way by entering the slope and intercept into the appropriate spot in the P column. For example if your curve fit gives a slope of 3.456e-4 and an intercept of 0.312 you would enter for the P column: "Potential"-0.312-3.456e-4*"Temperature" The bottom plot should now show the corrected area as a function of temperature. Integrate the full melting peak as you did for indium in the middle P vs. temperature plot. Determine the temperature where melting begins. Determine the temperature where the power, P, is about 2/3 of the maximum P value. Highlight the data between these two temperatures in the Logger Pro spreadsheet for Iadj and Tb, and record six data pairs evenly spaced in this region. Note that Iadj is the Area Sum in Table I before the pre-melting correction and Tb is the temperature extrapolated to the baseline, which is also the temperature needed for the calculations in Table I. Calculations Determine the fraction of the sample melted as shown in Table I. You will have six partial areas. The plot in Figure 4 is constructed. Adjust the correction for pre-melting to construct the most linear plot and give the smallest standard deviations for the slope and intercept. Calculate the enthalpy of fusion, the pure melting point (corrected), and the mole % impurity. Report Include a data table similar to Table I, and include the six data points that you used for determining partial areas. Report the enthalpy of fusion, the pure melting point (corrected using the indium melting temperature calibration), the freezing point depression ( T), and the mole % impurity. Include your final plot. Use significant figure rules to estimate the uncertainty of your enthalpy of fusion. Use the uncertainties from least squares curve fitting to estimate the standard deviation in your freezing point depression, pure melting point, and the mole % impurity. Discuss the chemical significance of this experiment: what other techniques are used by organic chemists to determine purity, and what is the advantage of the DSC technique compared to these methods?

Purity by DSC 9 References 1. This write up is excerpted almost totally from C. Plato and S. R. Glasgow, Jr., Anal. Chem., 1969, 41 (2), 330. 2. P. W. Atkins, Physical Chemistry, 7th ed., Freeman, New York, 2002, Section 7.5. 3. Thermal Analysis Newsletter, Perkin-Elmer Corp., Norwalk, CT. 4. A. P. Gray, Instrument News, 1966, 16 (3), 9.