The TSL 6.4 GHz ECR ion source status, improvements and measurements



Similar documents
Status of High Current Ion Sources. Daniela Leitner Lawrence Berkeley National Laboratory

Development of Radiation Resistant Quadrupoles Based on High Temperature Superconductors for the Fragment Separator

HIGH CURRENT OPERATION OF THE ACSI TR30 CYCLOTRON

Coating Technology: Evaporation Vs Sputtering

Cathode Ray Tube. Introduction. Functional principle

A new extraction system for the upgraded AIC-144 cyclotron

High Power Neutral Beam Injection in LHD

Principles of Ion Implant

An example: helium isotopes. An example: helium isotopes. Limits to Detection/Measurement. Lecture 14 Measurements II: Mass spectrometry

The ithemba LABS Radioactive Beam Project. R A Bark

Activities at the University of Frankfurt (IAP)

FIVE PROJECTS OF HIGH RF POWER INPUT COUPLERS & WINDOWS FOR SRF ACCELERATORS*

13C NMR Spectroscopy

Nuclear Structure. particle relative charge relative mass proton +1 1 atomic mass unit neutron 0 1 atomic mass unit electron -1 negligible mass

DEMONSTRATION ACCELERATOR DRIVEN COMPLEX FOR EFFECTIVE INCINERATION OF 99 Tc AND 129 I

Modification of Pd-H 2 and Pd-D 2 thin films processed by He-Ne laser

INFRARED MONITORING OF 110 GHz GYROTRON WINDOWS AT DIII D

PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.

STUDY OF THE TRANSVERSE BEAM EMITTANCE OF THE BERN MEDICAL CYCLOTRON

How To Improve The Ganil Control System

Current Probes. User Manual

Multi-elemental determination of gasoline using Agilent 5100 ICP-OES with oxygen injection and a temperature controlled spray chamber

World-first Proton Pencil Beam Scanning System with FDA Clearance

DENSITY MEASURING SYSTEMS

TOF FUNDAMENTALS TUTORIAL

THE ALIGNMENT STRATEGY FOR THE SPIRAL2 SUPERCONDUCTING LINAC CRYOMODULES

The Periodic Table: Periodic trends

A Remote Plasma Sputter Process for High Rate Web Coating of Low Temperature Plastic Film with High Quality Thin Film Metals and Insulators

2 ATOMIC SYSTEMATICS AND NUCLEAR STRUCTURE

C.-K. Ng. Stanford Linear Accelerator Center. and. T. Weiland. University oftechnology. FB18, Schlossgartenstr. 8. D64289, Darmstadt, Germany.

RF SYSTEM FOR VEPP-5 DAMPING RING

Product Specification

Magnetic Field and Magnetic Forces

State of the art in reactive magnetron sputtering

EXPERIMENTAL CONDITIONS FOR CROSS SECTION MEASUREMENTS FOR ANALYTICAL PURPOSES. L. Csedreki 1. Abstract. I. Introduction

Reliability and Availability Aspects of. the IPHI Project

Laser drilling up to15,000 holes/sec in silicon wafer for PV solar cells

GA A25827 EFFECTS OF ELECTRON CYCLOTRON CURRENT DRIVE, COUNTER-NBI, AND ROTATIONAL ENTRAINMENT ON NEOCLASSICAL TEARING MODE CONTROL IN DIII-D

How To Analyze Plasma With An Inductively Coupled Plasma Mass Spectrometer

ION ENERGY DISTRIBUTION FUNCTION MEASURED BY RETARDING FIELD ENERGY ANALYZERS

14.5GHZ 2.2KW CW GENERATOR. GKP 22KP 14.5GHz WR62 3x400V

4 SENSORS. Example. A force of 1 N is exerted on a PZT5A disc of diameter 10 mm and thickness 1 mm. The resulting mechanical stress is:

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

Electron Beam and Sputter Deposition Choosing Process Parameters

arxiv:astro-ph/ v1 15 Sep 2005

Shielding Effectiveness Test Method. Harbour s LL, SB, and SS Coaxial Cables. Designs for Improved Shielding Effectiveness

Section 21. Telecommunication Hardware. BuyLog Catalog 21-1

AORC Technical meeting 2014

Spherical Aberration Corrections for an Electrostatic Gridded Lens

3. What would you predict for the intensity and binding energy for the 3p orbital for that of sulfur?

Lasers Design and Laser Systems

Overview of Proton Beam Cancer Therapy with Basic Economic Considerations

High Rate Oxide Deposition onto Web by Reactive Sputtering from Rotatable Magnetrons

High Power Tests of Beryllium Oxide Windows to the Lower Hybrid Current Drive Launcher in JET

Quantum Computing for Beginners: Building Qubits

K O M A C. Beam Commissioning of 100-MeV KOMAC Linac. Korea Multi-purpose Accelerator Complex 양 성 자 가 속 기 연 구 센 터

Series 6000 Torque measured metal bellow coupling

Information about the T9 beam line and experimental facilities

ITER - business in fusion

Alternative Linear Motion Systems. Iron Core Linear Motors

Neuere Entwicklungen zur Herstellung optischer Schichten durch reaktive. Wolfgang Hentsch, Dr. Reinhard Fendler. FHR Anlagenbau GmbH

MEASUREMENTS OF THRUST LOAD IN CABLE SYSTEMS SEMI-RIGIDLY INSTALLED IN DUCT / MANHOLE STRUCTURES

Atoms and Elements. Outline Atoms Orbitals and Energy Levels Periodic Properties Homework

High Power VASIMR Experiments using Deuterium, Neon and Argon

Chapter 1: Moles and equations. Learning outcomes. you should be able to:

Active noise control in practice: transformer station

B) atomic number C) both the solid and the liquid phase D) Au C) Sn, Si, C A) metal C) O, S, Se C) In D) tin D) methane D) bismuth B) Group 2 metal

source at CERN G.McMonagle AB/RF

ROTATING MACHINES. Alignment & Positioning of. Fast, easy and accurate alignment of rotating machines, pumps, drives, foundations, etc.

Facilities Perspectives

CHAPTER 4 DESIGN OF INTEGRAL SLOT AND FRACTIONAL SLOT BRUSHLESS DC MOTOR

Linac RF Commissioning with the SNS HPRF Systems

Projects and R&D activities

AUTOMATION OF OPERATIONS ON THE VEPP-4 CONTROL SYSTEM

COLLATED QUESTIONS: ELECTROMAGNETIC RADIATION

Enhanced Stripline Scanning Array B.M. Cahill and J.C. Batchelor

AMD Analysis & Technology AG

Atomic Theory: History of the Atom

& Tunnel Cross Section

Ferrostat Speed Sensor Series DSF Explosion Proof Versions EEx

RI-80 SMD Series Dry Reed Switch

Experimental results for the focal waveform and beam width in the focusing lens with a 100 ps filter

User checks and maintenance of laboratory balances

8 Speed control of Induction Machines

Mars Atmosphere and Volatile EvolutioN (MAVEN) Mission

Vacuum Pumping of Large Vessels and Modelling of Extended UHV Systems

First Measurements with U-probe on the COMPASS Tokamak

Selected Radio Frequency Exposure Limits

Understanding Ignition Waveforms

SIMULATIONS OF ELECTRON CLOUD BUILD-UP AND SATURATION IN THE APS *

Coaxial Cable Products Guide. Connectivity for Business-Critical Continuity

METHOD OF STATEMENT FOR STATIC LOADING TEST

BM SERIES WAVE COMPACT 6.3 GHZ CW SERIES PULSE RADAR LEVEL INSTRUMENTS (6.3 GHz)

Tom Wilson Product Marketing Manager Delivery Systems Varian Medical Systems International AG. CERN Accelerator School, May 2015

Transcription:

NUKLEONIKA 2003;48(Supplement 2):S99 S104 PROCEEDINGS The TSL 6.4 GHz ECR ion source status, improvements and measurements Daniel van Rooyen, Dan Wessman Abstract The TSL 6.4 GHz ECR ion source performs reliably and is well optimized for the various ion species that are routinely provided for experiments. Beam intensities are comparable with other similar sources but at the lower end of the spectrum. We are thus investigating a number of methods of improvement. The development of a micro-oven for low melting point materials was successfully concluded. Further development is needed to improve the consumption rate in order to enable more effective use of expensive isotopes. Measurements with operation of the source in the after-glow mode were successful but the pulse to pulse reproducibility should be further improved. Although the maximum gain compared to the CW mode is satisfactory, accomplishing a higher factor would be even more advantageous for beams delivered to CELSIUS. A systematic study of various parameters was started in order to find optimal operating conditions running in the after-glow mode. Key words ECR ion source after-glow mode Introduction The TSL 6.4 GHz ECR ion source was constructed in a collaboration between TSL and JYFL (Jyväskylä) as a copy of the Michigan RT 6.4 GHz. The first beams were extracted in November 1990. Subsequently, a similar ECRIS was constructed for installation at JYFL. ECR beams are vertically injected into the cyclotron, which is a single pole machine with three sectors for vertical focussing. The maximum average field is 1.75 T at a maximum useful radius of 1.2 m. The bending limit is given by K = 192 Q 2 /A MeV; the focussing limit (protons) of 100 MeV is circumvented by using frequency modulation up to 180 MeV. Refer to Refs. [3, 7] for more detailed descriptions of the characteristics of the cyclotron. The ECRIS is routinely used for a number of periodically recurring experiments in various fields: physics (fusion reactions, CHICSI experiments in the CELSIUS storage ring); bio-medical research (DNA double-stranded break repair etc.); ion track micro-technology (nano-wires and GMR based magnetic field sensors). Beams most often extracted are: B, N, O, Ne, Mg, Ar, Xe. D. van Rooyen, D. Wessman The Svedberg Laboratory, Uppsala University, P.O. Box 533, S-751 21 Uppsala, Sweden, Tel.: +46 18/ 471 3445, Fax: +46 18/ 471 3833, e-mail: daniel.vanrooyen@tsl.uu.se Received: 26 November 2002, Accepted: 20 May 2003 Status and performance of the TSL ECR ion source Considering the long and extensive use of the TSL ECR ion source, near optimal tuning conditions were established for the different ion species. It should be noted that the source operation proved to be very reliable and beam-time is very seldom lost due to breakdowns or insufficient intensity. In a comparison of similar sources, it is clear that the per-

S100 D. van Rooyen, D. Wessman Fig. 1. Comparison of the performance of the TSL 6.4 GHz ECR with various other similar ECR, s. (The data for the JYFL 6.4 GHz reflects the status before the magnetic field upgrade.) formance of our source is comparable, but at the lower end of the spectrum (Fig. 1). In the recent two years it was again attempted to verify the nature of the difference between the JYFL and TSL ECR ion sources. Due to the many factors playing a role in the performance of an ECR of which is not the least the previous ion history, even repeating previous performances are not always trivial. Nevertheless we are seriously embarking on (1) a study of possible explanations of these differences and (2) various actions that can be taken to improve its performance. New beams User interest was expressed for a sulphur [13] beam, which was satisfactorily developed but not accelerated. The developed (and accelerated) silicon beam was satisfactory with the intensity comparing reasonably well with reports from other laboratories. Beam currents were as in the Table 1. Fe beam development using the MIVOC (Metal Ions from Volatile Organic Compounds) [1] method resulted in stable beams but with beam currents only on the verge of usable intensities. Beam intensities of up to a factor 2 higher could not be kept stable for reasonably long periods. Instead of implementing the micro-oven it is thus as yet not an option to rather use the MIVOC method for metallic beams of (relative) expensive isotopes. Development of micro-oven A micro-oven for low melting point solids was developed during 2000, with a first run with Mg 25 and Mg 26 in September 2000, and the oven further improved for a next run in February 2001. Details of the oven are shown in Figs. 2 and 3. A primary condition in the design was the essentiality to be able to refill the oven without disturbing the ECRIS vacuum. Although the oven is now functioning well, further optimization is necessary. Due to the limited space available with access via the radial ports (Fig. 4), the volume of the oven is severely restricted with the implication that the maximum mass is limited to ca. 230 mg Mg thus limiting runs to ca. 90 hours before refilling is needed. Consumption rate and ionization efficiency measurements were routinely done. Ionization efficiency is defined as, for example, the summed beam current for all Mg ions relative to the mass of Mg metal consumed. Depending on the beam intensity needed, and thus the degree of heating, the consumption rate (Mg) is varying between 2.5 and 4.5 mg/h. The ionization efficiency of ca. 0.7% compares well with figures reported [2]. For the moment we are content with the oven performance as it is as yet not foreseen that beams from solids will become a strong component of our beam spectrum. Even though Mg 25,26 is not so extremely expensive, it would be advantageous to prevent loosing large percentages of the material on colder inner surfaces. A heated inner structure Table 1. Analyzed silicon, sulphur and iron beam intensities from the TSL 6.4 GHz ECR ion source. ECRIS Laboratory Data source Beam currents (eµa) Silicon Sulphur Iron Q 5+ 7+ 3+ 5+ 7+ 11+ 13+ JYFL, 6.4 GHz Jyväskylä, Finland Web (before upgrade) 26.5 10.2 36.5 2.5 5.6 LBL, 9.2&6.4 GHz Berkeley, USA [5] 72 10 20 63 5 NAC, 10.1 GHz Faure, South Africa NAC beam records 10.4 0.9 3.2 TSL, 6.4 GHz Uppsala, Sweden TSL beam records 14 6 22.8 1.5 1.6 * HIL UW, 10 GHz Warsaw, Poland [10] 50 60 * Fe intensities of 2.5 to 3 µa could not be kept stable for longer periods than several hours.

The TSL 6.4 GHz ECR ion source status, improvements and measurements S101 Fig. 2. Close-up of oven tip (26 mm outside diameter). Fig. 5. Ar 11+ beam current measurement with the ECRIS operated in the after-glow mode. concentric with the plasma chamber inner surface would be worthwhile to be developed. Increasing the volume in a new micro-oven design will also be a high priority. With even better optimization of the various source parameters it is believed that both the consumption rate and ionization efficiency could be further improved. As time allows we will thus proceed with these mentioned improvements. Operation in pulsed mode Fig. 3. Detail of micro-oven. Fig. 4. TSL ECR ion source indicating radial port where microoven is positioned. More often, at TSL as elsewhere, it is essential to have the maximum possible average beam available to users which thus necessitates the use of DC ECR beams. This is not the case, for example, when heavy ions are produced for acceleration in CELSIUS. At various laboratories it has been established that pulsing of the microwave power leads to a factor of 2 to 3 increase in the extracted beam intensity measured immediately after cut-off of the RF pulse. For high charge states increase factors of up to 10 were also reported [6]. Previous measurements with the TSL ECRIS operated in the so-called after-glow mode extracting lighter ions were not very encouraging due to plasma instabilities. Ar 11+ is of specific interest for acceleration in CELSIUS and running the ECRIS in pulsed mode optimized for this (heavier) ion was further investigated. With re-tuning of several of the source parameters, a stable condition of the plasma could be accomplished. The after-glow peak of 10 µa was repetitive with time and is thus an increase almost by a factor of 2 compared to the DC beam (Fig. 5). For this relative low charge state this result was very satisfactory. For this test measurement the relevant parameters are pre-sented in Table 2. Conclusions from a preliminary systematic study are: increasing the pulsing frequency does not lead to a significant increase in beam intensity and leads to the plasma being more susceptible to instabilities; the RF on pulse should be sufficiently long for the plasma to stabilize; tuning is simpler if the RF off pulse is sufficiently long for the plasma to completely die away but there seems to be an advantage in building up the plasma before it completely quenches.

S102 D. van Rooyen, D. Wessman Table 2. Typical parameters for operation in the pulsed mode (after-glow mode). Klystron off Klystron on Pulsing Length of after-glow (ms) (ms) frequency (Hz) beam pulse (ms) 128 273 3 10 Improvements and measurements Biased disk A small change to enable increased voltage deliverable to the biased disk led to an intensity increase of ca. 20% for higher charge states, i.e. Xe 27+, which made acceleration of these beams more comfortable. History effects of Si and B It could not statistically be indicated sufficiently well enough that a run with Si, and thus creating a SiO 2 layer on the plasma chamber inner surface, has a positive effect on subsequent beam intensities. One may conclude that the cold electron effect of the SiO 2 is not noticeable when superimposed on the effect of the biased disk. A positive effect that was indeed clear is that optimal tuned conditions for subsequent beams are reached with more ease. Contrary to measurements performed at other laboratories (private communication J. Ärje, JYFL, Jyväskylä), a run with B seems to have a negative effect on subsequent beams. A possible explanation is that our observance applies only to high Q beams (for example: 129 Xe 27+ ) where very high vacuum is essential and that B, using the MIVOC method with decaborane H 14 B 10, is still outgassing from the inlet lines. Fig. 6. Comparison of axial magnetic field before and after upgrade of coil configuration with reference to the SC-ECRIS magnetic field. (Figure courtesy of Hannu Koivisto, JYFL.) The SC-ECRIS at NSCL/MSU is a very well performing 6.4 GHz ECR ion source. From a study of the magnetic field characteristics of the SC-ECRIS at NSCL/MSU it was concluded [9] that the magnetic field of the RT 6.4 GHz ECR s can be adapted and the magnitude increased to such an extent to resemble the configuration of the SC-ECRIS (Fig. 6). During 2001 the magnetic field configuration of the JYFL 6.4 GHz ECRIS was adapted to incorporate these calculations with very positive results as shown in Fig. 7. In collaboration with JYFL, a similar reconstruction of the TSL ECRIS is scheduled to start in November 2002. Currently the 9 axial coils are symmetrically positioned in 3 groups of 3 each at injection, centre and extraction. A new yoke will be manufactured and the 9 coils regrouped in only an injection set (5 coils) and extraction group (4 coils). The new yoke will completely surround each set of coils enabling a further enhancement of the magnetic field. Additionally, the fields at the injection and extraction region will be strengthened by respectively adding a plug of high quality magnetic steel at injection and two rings in the vicinity of the plasma electrode. The strength of the hexapole magnetic field will also be increased. This will be accomplished by adding a cylinder of magnetic steel closely surrounding the plasma chamber. With completion of the last scheduled run with heavy ions at the end of December 2002, we start with the disassembly of the ECR ion source. Towards the end of January 2003 the coil modifications should be completed and all yoke components on site and ready for assembly. During February the new yoke and coils will be assembled and all services connected. The final stage of positioning the plasma chamber and the injection assembly (extra iron plug, biased disk, gasfeeds) will be completed such that pumping vacuum can start at the beginning of March. We allow for ca. two Magnetic field upgrade Fig. 7. Improvement in beam intensities for the Jyväskylä 6.4 GHz ECR ion source after adaptation of the magnetic field configuration. (Figure courtesy of Hannu Koivisto, JYFL.)

The TSL 6.4 GHz ECR ion source status, improvements and measurements S103 weeks of pumping and outgassing which then provides a final 3 weeks for testing before the first experiments with heavy ions from the ECR ion source are scheduled for the beginning of April 2003. All drawings are now finalized and we have received the first quotations for those yoke components whose dimensions are too large to be manufactured internally. The first of the necessary new components involved in the upgrade have also been manufactured and machining will continue dependent on available workshop time. Future prospects Apart from the magnetic field upgrade, further future prospects are: Improving plasma characteristics VECC, Calcutta reports [12] the very positive effect of plating the inner surface of the plasma chamber with Al and thereby increasing beam intensities with up to a factor of 2. This confirms the established belief that provision of secondary cold electrons (in this case from the oxidised Al) to the ECR plasma has a significant effect stabilizing the plasma, resulting also in significantly higher beam intensities extracted. Using a biased disk positioned axially at injection is an alternative source of cold electrons, although new results seems to indicate that the effect of the biased disk is rather related to optimisation of the plasma potential or improved extraction conditions. Similar to the detailed study of various effects and operational modes of biased disks at RIKEN (Japan) [8] and ATOMKI (Hungary) [4], it should thus also be experimentally established what the optimal operational mode for the TSL ECR ion source is. Improving MIVOC beams The MIVOC method is known as a very convenient technique to produce metallic ions, especially as the use of an oven or any other heating or sputtering is avoided. The method does, on the other hand, require exact procedures and careful tuning of the source. Understanding of the MIVOC method and the related tuning of the ECRIS have been improved and valuable experience has been gained. It is believed that we can still further improve on reached intensities. Furthermore, this knowledge can now be extended to other elements (Ni, Co, Cu, Ge and also I 2 ). Optimization of the extraction system On a longer term it would be worthwhile to study the extraction system and verify whether possibilities exist to improve the current system. Additionally, the injection beam-line should be re-evaluated and the possibility of an updated emittance system investigated. New inflector To improve the acceptance and transmission of beams from the ECRIS into/through the cyclotron, a new inflector was designed [11]. The relevant parameters (e.g. k which is related to the direction of the electric field) are chosen to minimize the inflector fringe field. The electrode spacing at entrance, d 0, is 8 mm and the width of the electrode is 16 mm, thus improving the aspect ratio from 1 to 2. The off-centre distance between the centre of the cyclotron and the centre of the trajectory at the inflector exit was chosen as 11.7 mm. Calculations for the ion trajectories for 3 reference ions were done. These calculations indicated that the inflector exit parameters and the energy gain is sufficient to clear all the central region components. Construction of the inflector is nearly completed and the first tests are scheduled for October 2002. Conclusion The source is performing well and in a stable and dependable fashion. Extracted beam intensities can be improved. First tests for the operation of the ECRIS in the after-glow mode gave positive results. We have, with reference to various aspects, embarked on a programme to improve our ECRIS. In addition, we expect that the development and further optimization of the new inflector will improve intensities available from the cyclotron. References 1. Ärje J, Leinonen H, Seppälä R et al. (1999) Status report of the JYFL-ECR ion sources. In: Proc of the 14th ECRIS Workshop, 3 6 May 1999, CERN, Geneva, Switzerland, pp 116 119 2. Barué C, Been T, Dupuis M et al. (1999) Metallic ion developments at GANIL. In: Proc of the 14th ECRIS Workshop, 3 6 May 1999, CERN, Geneva, Switzerland, pp 111 115 3. Berglund M (1993) Studies of axial injection in the Gustav Werner Cyclotron. UPTEC 93 104E. Institute of Technology, Uppsala University 4. Biri S, Nakagawa T, Kidera M et al. (1999) Highly charged ion production using an electrode in biased and floating modes. In: Proc of the 14th ECRIS Workshop, 3 6 May 1999, CERN, Geneva, Switzerland, pp 81 85 5. Brown IG (ed.) (1989) The physics of ion sources. Wiley, New York 6. Geller R (1996) Electron cyclotron resonance ion sources and ECR plasmas. IOP, Bristol 7. Hedblom K, Berglund M (1993) Beam optics calculations for the axial hole in the Gustav Werner Cyclotron. TSL-Note 93-04. Uppsala University 8. Kidera M, Higurashi Y, Nakagawa T et al. (1999) Effect of biased electrode to high energetic electrons in plasma of RIKEN 18 GHz ECRIS. In: Proc of the 14th ECRIS Workshop, 3 6 May 1999, CERN, Geneva, Switzerland, pp 86 89 9. Koivisto H, Heikkinen P, Ärje J et al. (2001) Development work with the JYFL ECR ion sources. In: Proc of the 16th Int Conf on Cyclotrons and their Applications held at NSCL/MSU, 13 17 May 2001, East Lansing, USA. AIP 600:255 257

S104 D. van Rooyen, D. Wessman 10. Sudlitz K, Kulczycka E (1999) 10 GHz ECRIS for Warsaw Cyclotron. In: Proc of the 14th ECRIS Workshop, 3 6 May 1999, CERN, Geneva, Switzerland, pp 205 207 11. Toprek D, Reistad D, Lundstrȯ. m B et al. (2002) Design of the central region in the Gustav Werner Cyclotron at the Uppsala University. Nucl Instrum Meth Phys Res A 486:539 544 12. VECC Progress report (2000) VECC, Calcutta, India 13. Zyromski KE, Loveland W, Souliotis GA et al. (2001) Fusion enhancement in the 32,38 S + 181 Ta reaction. Phys Rev C 63:024615