White Paper: Dario Blood Glucose Monitoring System - Performance Evaluation

Similar documents
Evaluation of Accuracy and User Performance of the TRUE METRIX Self-Monitoring Blood Glucose System *

Accuracy Test Omnitest 3 Blood Glucose Monitoring System. B. Braun Diabetes Care

Blood Glucose Monitoring: The Facts about Accuracy

FDA Public Hearing on Clinical Accuracy Requirements for Point of Care Blood Glucose Meters (BGMs)

A comparison of blood glucose meters in Australia

USING CLSI GUIDELINES TO PERFORM METHOD EVALUATION STUDIES IN YOUR LABORATORY

510(K) SUMMARY FEB Submitter: Name of the device: * Common Name: Clinical Chemistry Analyzer (with optional ISE Module) Attachment #2

Method Validation/Verification. CAP/CLIA regulated methods at Texas Department of State Health Services Laboratory

SMF Awareness Seminar 2014

Content Sheet 7-1: Overview of Quality Control for Quantitative Tests

Validation of measurement procedures

SureStep Pro Linearity Test Kit and Analytical Measurement Range Verification (AMR)

Inform II. Quick Reference Guide. for Healthcare Professionals BLOOD GLUCOSE MONITORING SYSTEM

Title: Glucose Testing: StatStrip Glucose Meter Page 1 of 6

US Diabetic Devices Industry Research Report: KenResearch

Analytical Test Method Validation Report Template

I know my value. CoaguChek XS Plus system Smart INR monitoring at your practice. 1 of 6

Self-Monitoring of Blood Glucose

Preface. Preface. A Healthcare professional should be contacted when Customer Service is not available.

Quantifying measurement error from digital instruments

To avoid potential inspection

Practice Advisory Blood Glucose Meter Accuracy Issued September 16, 2013

SKUP Scandinavian evaluation of laboratory equipment for primary health care

Onboard Quality Controls for the CoaguChek XS System

Unit 31 A Hypothesis Test about Correlation and Slope in a Simple Linear Regression

A guidebook for people with diabetes

Blood Glucose Monitoring Basics

STANDARD OPERATING PROCEDURE FOR NOVA STAT STRIP CONNECTIVITY BLOOD GLUCOSE METERS FOR USE IN LEEDS NHS TRUST

Management of Diabetes

Performance Variability of Seven Commonly Used Self-Monitoring of Blood Glucose Systems: Clinical Considerations for Patients and Providers

Self-Monitoring of Blood Glucose (SMBG)

General. What is a laboratory test? What is a waived test?

Textbook: Chapter 12 pages , Chapter 13 pages

Blood Sugar Monitoring

1. What is the critical value for this 95% confidence interval? CV = z.025 = invnorm(0.025) = 1.96

Title: Fingerstick Glucose by Accu-Chek Inform

HA Convention 2007 Service Priorities & Programmes 8

Validation and Calibration. Definitions and Terminology

Accu-Chek Inform II system. Professional glucose testing for the wireless age

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

Quality Control of the Future: Risk Management and Individual Quality Control Plans (IQCPs)

Quality Assurance/Quality Control in Acid Deposition Monitoring

2. What is the general linear model to be used to model linear trend? (Write out the model) = or

Who gets you to the next level?

Using R for Linear Regression

Control Charts and Trend Analysis for ISO Speakers: New York State Food Laboratory s Quality Assurance Team

A Self-Study Packet for The Johns Hopkins Hospital Point-of-Care Testing Program copyright 2011 Johns Hopkins Hospital All rights reserved

Blood Glucose Monitoring Test Systems for Prescription Point-of- Care Use Draft Guidance for Industry and Food and Drug Administration Staff

Telserve Licensed Healthcare Professional User Manual

An important first step in identifying those at risk for Cardiovascular disease The Accutrend Plus system: from the makers of the ACCU-CHEK and

PD-UserHbk-01. Pages printed from this document are uncontrolled - For latest versions please refer to electronic laboratory handbook

BLOOD GLUCOSE METER. Standard Owner s Booklet for Single Patient Use Only

Caution! Please read complete Owner s Booklet and all product Instructions for Use.

HEMATOCRIT/HCT AND CALCULATED HEMOGLOBIN/HB

Calibration and Linear Regression Analysis: A Self-Guided Tutorial

OneStep Fecal Occult Blood RapiDip InstaTest. Cat #

For people with diabetes. A blood sugar diary

Prognostic impact of uric acid in patients with stable coronary artery disease

Blood Sugar Testing. Bayer Health Facts

Quality Requirements of a Point of Care Testing Service

Report. European Co-ordinated Post Market Surveillance Operation under COEN authority. Blood Glucose Meters

Lab 2. Spectrophotometric Measurement of Glucose

after meals Monitoring: Control your the Power to bedtime before meals ActiveCare Basics ActiveCare Basics

Test strips and meter designed for glucose self-measurement manufactured by Abbott. Report from an evaluation organised by SKUP

EQUATIONS and INEQUALITIES

Samsung POINT OF CARE Systems - Cardiac/Acute Care Biomarkers LABGEO IB Clinical Chemistry Analytes LABGEO PT

Comparison of Glucose Determinations on Blood Samples Collected in Three Types of Tubes

BLOOD GLUCOSE METER. Getting Started Guide for Single Patient Use Only

Energy Performance Indicators (EnPI) Tim Dantoin Focus on Energy

TEST REPORT: SIEVERS M-SERIES PERFORMANCE SPECIFICATIONS

SD CodeFree BLOOD GLUCOSE MONITORING SYSTEM

Blood Glucose Monitoring System USER GUIDE

QUESTION 3. HOW SHOULD THE DATA AND REPORTING BE INTERPRETED?

Using Microsoft Excel to Plot and Analyze Kinetic Data

Key Features in the AgaMatrix Jazz

Understanding Analytical Chemistry (Weighing, Mixing, Measuring and Evaluating)

Tutorial for the TI-89 Titanium Calculator

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

OAML Program for the Clinical Placement of Medical Laboratory Assistants

Risk Calculation Software Requirements for Down's Syndrome Screening

Basic Statistics and Data Analysis for Health Researchers from Foreign Countries

The Accuracy of Commercial Blood Uric Acid Meters on Blood Uric Acid Level Measurement

CALCULATIONS & STATISTICS

HIV Rapid Test Kits USAID Approval and Technical Guidance June 17, 2013 Revision

Nova Max Blood Glucose Monitor Owner s Guide

Chapter 10. Key Ideas Correlation, Correlation Coefficient (r),

Correlation and Regression

How to Verify Performance Specifications

Hypertension and Diabetes Status Bangladesh Demographic and Health Survey

Urinalysis Compliance Tools. POCC Webinar January 19, 2011 Dr. Susan Selgren

CoaguChek System. XS Plus. User Manual

AP Physics 1 and 2 Lab Investigations

Point-of-care (POC) versus central laboratory instrumentation for monitoring oral anticoagulation

The Effect of Temperature on Conductivity Measurement

Solving Systems of Linear Equations Graphing

GLUCOSE TOLERANCE (FOR DIABETESMELLITUS)- NON-PREGNANT ADULT/CHILDREN (Under 18 years old) (OGTT)

Blood Glucose - How to test

Transcutaneous Bilirubinometer

table to see that the probability is (b) What is the probability that x is between 16 and 60? The z-scores for 16 and 60 are: = 1.

Introduction to Biosensors. Professor Brian Birch LIRANS University of Luton UK

Transcription:

White Paper: Dario Blood Glucose Monitoring System - Performance Evaluation Date: Feb 10, 2014 LabStyle Innovation

TABLE OF CONTENTS 1. PURPOSE... 3 2. SCOPE... 3 3. REFERENCE... 4 4. DEFINITIONS... 4 5. STRUCTURE AND ORGANIZATION OF STUDY... 5 6. OVERVIEW OF EVALUATION... 5 7. RESULT ANALYSIS OF EVALUATION... 6 8. CONCLUSIONS... 9 2

1. PURPOSE This Performance Evaluation study was conducted to evaluate the accuracy of blood glucose level results obtained from fingertip using Labstyle Innovations Dario Blood Glucose Monitoring System (BGMS) compared to reference equipment (YSI 2300 STATPLUS) and to evaluate the ease of use of the Dario BGMS. The performance evaluation was conducted following an independent study performed by Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbh an der Universität Ulm in January 2014. The study showed excellent correlation between SMBG results obtained by OEM based technology and the results obtained by the reference YSI (gold standard). The results where within the acceptance criteria of EN ISO15197:2013 for system accuracy. 2. SCOPE 2.1 Device The device used in this study was the Dario Blood Glucose Monitoring System (BGMS). The device comprises of the following components: Dario Test Strip Dario Lancing Device Dario Blood Glucose Meter (Dongle) Smart Mobile Device Software 2.2 Intended Use Dario Blood Glucose Monitoring System (BGMS) is indicated for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips. The system is intended to be used by a single patient and should not be shared. The Dario BGMS is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in monitoring the effectiveness of diabetes control. The Dario BGMS is not to be used for the diagnosis of or screening of diabetes or for neonatal use. 3

Dario Blood Glucose meter (BGM) must be used with Dario blood glucose test strip, Dario control solutions, Dario lancing device, Dario lancets, and Dario application software 2.3 Test Principle The test is based on electrochemical biosensor technology and the principle of capillary action. The electrical current generated by the reaction of glucose with the reagent of the strip is measured by the meter and is displayed on an ios-based smart mobile device (e.g., iphone, ipod, ipad) as the corresponding blood glucose level. The strength of the current produced by the reaction depends on the amount of glucose in the blood sample. 2.4 Technical Background Dario technology is based on OEM GOD strip technology which was recently evaluated by an independent laboratory. The OEM BGMS accuracy study was performed by Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbh an der Universität Ulm in January 2014. The study results confirmed that the system meets the accuracy requirements of the ISO 15197:2013. Three test strip lots were tested demonstrating accuracy of 95.5%, 98%, and 97%. These results meet the system accuracy requirements of the ISO 15197:2013 and reinforces the notion that the Dario strip technology complies with the same requirements. 3. REFERENCE 1. EN ISO 15197:2013 In vitro diagnostic test systems Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus, Clause 8. User performance evaluation 8.3 User evaluation / 8.4 Evaluation of instructions for use. 2. BS EN 13612:2002 Performance evaluation of in vitro diagnostic medical devices. 4. DEFINITIONS Reference value: YSI (mg/dl) X 1.06 plasma conversion value: 4

The capillary whole blood results obtained by the YSI were multiplied by 1.06 since the YSI is calibrated for venous blood as determined by the below chart 1 : 5. STRUCTURE AND ORGANIZATION OF STUDY The study was conducted in internal laboratory. The following individuals were involved in the performance of this study: Operators - Mr. SH Ha (Medical Technologist, SD Biosensor, Inc.) - Mr. MK Kin (QC, SD Biosensor, Inc.) - Ms. YS Ahn (Nurse, YONSEI Medical Clinic) Study condition: Overall studies by SMBG meter were followed in accordance with Dario BGM system User Instruction Guide, i.e. same intended use condition. 6. OVERVIEW OF EVALUATION The study was conducted over a day period, Sept. 25 th, 2013. A total of 87 subjects were recruited. Fresh capillary blood was obtained from each subject. A range of glucose concentration was tested 28-458 mg/dl. Capillary blood samples were modified - raised or lowered to achieve the required distribution. Methods for adjusting the glucose concentration are described in IST 15197:2013(E) 6.1.2. The samples were tested by using 3 Dario Blood Glucose Meters from different lots and Dario Test Strip (three different test strip lots: Lot No.: SPS1013001, SPS1013002, SPS1013003) and by YSI. 1 Davidson & Parker: Clinical Diabetes Mellitus: A Problem Oriented Approach.2000, p.461 5

7. RESULT ANALYSIS OF EVALUATION The results and accuracy tables are represented in accordance with ISO 15197:2013 Standard for system accuracy comparing the results obtained with Dario BGMS to YSI. Investigator presented regression equations with confidence intervals and plots showing all data points according to ISO 15197:2013. Sample distribution: Glucose Concentration (mg/dl) Percentage of the sample <50 5% 50-80 8% 7% 81-120 20% 121-200 30% 201-300 15% 301-400 5% 5% 401> 5% Evaluation results Individual results of the blood-glucose monitoring system were plotted as the dependent variable and the average of the duplicate reference value (= YSI values * 1.06) as the independent variable. Identical scales and intervals were used for the x- and y-axes. Slope and y-intercept were calculated by regression analysis procedure (EXCEL/TOOL/DATA ANALYSIS/ Regression Analysis). Dario (SPS1013001 Dario(SPS1013002) Dario(SPS1013003) Slope 1.0021 1.0029 1.0012 Y-Intercept 1.2465-0.3243 1.6083 R 0.9972 0.9968 0.9969 6

Figure 1: Consensus Error Grid (mg/dl) Dario SPS 1013001 Figure 2: Consensus Error Grid (mg/dl) Dario SPS 1013002 7

Figure 3: Consensus Error Grid (mg/dl) Dario SPS 1013003 System Accuracy according to ISO15197:2013 Reference: YSI (YSI value * 1.06) Dario BGMS Test Results SPS 1013001 System accuracy results: Dario BGM system (SPS 1013001) concentrations <100 mg/dl concentrations 100 mg/dl Within ± 5 Within ± 10 Within ± 15 Within ± 5 % Within ± 10 % Within ± 15 % mg/dl mg/dl mg/dl 60/72 83% 94/128 73% 122/128 95% 128/128 Dario BGMS Test Results SPS 1013002 System accuracy results: Dario BGM system (SPS 1013002) concentrations <100 mg/dl concentrations 100 mg/dl Within ± 5 Within ± 10 Within ± 15 Within ± 5 % Within ± 10 % Within ± 15 % mg/dl mg/dl mg/dl 62/72 86% 71/72 99% 90/128 70% 121/128 95% 127/128 99% Dario BGMS Test Results SPS 1013003 8

System accuracy results: Dario BGM system (SPS 1013003) concentrations <100 mg/dl concentrations 100 mg/dl Within ± 5 Within ± 10 Within ± 15 Within ± 5 % Within ± 10 % Within ± 15 % mg/dl mg/dl mg/dl 61/72 85% 93/128 73% 119/128 93% 125/128 98% 8. CONCLUSIONS The Performance Evaluation was performed according to the ISO 15197:2013, 6.3 System accuracy. According to the result analysis of this evaluation, the Dario BGMS showed excellent correlation coefficient (R= 0.997, 0.997, 0.997) with YSI reference method (venous plasma - calibrated). The results fell within the acceptance criteria of ISO15197:2013 for system accuracy requirements: Ninety-five percent (95%) of the measured glucose values shall fall within either ±15mg/dL of the average measured values of the reference measurement procedure at glucose concentrations < 100mg/dL or within ±15% at glucose concentrations 100 mg/dl. - 99% of individual glucose measured values shall fall within zones A and B of the Consensus Error Grid(CEG) for type 1 diabetes. Additionally, of individual glucose measured values were within zones A of the Consensus Error Grid(CEG) for type 1 diabetes. The results of this Performance Evaluation study together with the results of the independent study performed in Germany support the conclusion that the Dario BGMS meets the accuracy requirements of EN ISO15197:2013. 9