Nuclear Magnetic Resonance and the Measurement of Relaxation Times of Acetone with Gadolinium



Similar documents
Nuclear Magnetic Resonance

A look at the utility of pulsed NMR

E. K. A. ADVANCED PHYSICS LABORATORY PHYSICS 3081, 4051 NUCLEAR MAGNETIC RESONANCE

NMR Techniques Applied to Mineral Oil, Water, and Ethanol

Nuclear Magnetic Resonance and Its Application in Condensed Matter Physics

Pulsed Nuclear Magnetic Resonance An Experiment for UCSB s Advanced Laboratory

Basic Principles of Magnetic Resonance

Nuclear Magnetic Resonance (NMR) Spectroscopy

Nuclear Magnetic Resonance

NMR SPECTROSCOPY. Basic Principles, Concepts, and Applications in Chemistry. Harald Günther University of Siegen, Siegen, Germany.

How To Understand The Measurement Process

NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706

The Hydrogen Atom Is a Magnet.

Understanding Precessional Frequency, Spin-Lattice and Spin-Spin Interactions in Pulsed Nuclear Magnetic Resonance Spectroscopy Introduction Theory

Nuclear Magnetic Resonance (NMR) Spectroscopy cont... Recommended Reading:

Nuclear Magnetic Resonance Spectroscopy

Nuclear Magnetic Resonance Spectroscopy

How To Understand Electron Spin Resonance

Trans Fats. What is a trans fat? Trans fatty acids, or trans fats as they are known, are certain

Atomic and Nuclear Physics

Experiment 5. Lasers and laser mode structure

Prof.M.Perucca CORSO DI APPROFONDIMENTO DI FISICA ATOMICA: (III-INCONTRO) RISONANZA MAGNETICA NUCLEARE

Proton Nuclear Magnetic Resonance Spectroscopy

Generation and Detection of NMR Signals

NMR for Physical and Biological Scientists Thomas C. Pochapsky and Susan Sondej Pochapsky Table of Contents

Advanced Physics Labs SEPT Pulsed NMR

4. It is possible to excite, or flip the nuclear magnetic vector from the α-state to the β-state by bridging the energy gap between the two. This is a

Spin-lattice and spin-spin relaxation

ELECTRON SPIN RESONANCE Last Revised: July 2007

NMR - Basic principles

13C NMR Spectroscopy

Introduction to Nuclear Magnetic Resonance Spectroscopy

NMR SPECTROSCOPY A N I N T R O D U C T I O N T O... Self-study booklet NUCLEAR MAGNETIC RESONANCE δ PUBLISHING

Spin-Lattice Relaxation Times

Pulsed Fourier Transform NMR The rotating frame of reference. The NMR Experiment. The Rotating Frame of Reference.

Ph 3504 Nuclear Magnetic Resonance and Electron Spin Resonance

The Experiment Some nuclei have nuclear magnetic moments; just as importantly, some do not

Nuclear Magnetic Resonance

Massachusetts Institute of Technology Department of Chemistry 5.33 Advanced Chemical Instrumentation FALL SEMESTER 2005

Chapter 19 Nuclear Magnetic Resonance Spectroscopy (NMR)

Determination of Molecular Structure by MOLECULAR SPECTROSCOPY

NMR Pulse Spectrometer PS 15. experimental manual

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

Nuclear Magnetic Resonance Spectroscopy

Organic Chemistry Tenth Edition

Nuclear Magnetic Resonance Spectroscopy

Used to determine relative location of atoms within a molecule Most helpful spectroscopic technique in organic chemistry Related to MRI in medicine

Step Response of RC Circuits

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS

Tetramethylsilane (TMS) Trimethylsilyl d 4. -propionic acid (TMSP) Dioxane. O - Na + Dimethylfura n. Potassium Hydrogen Phthalate. Sodium Maleate CH 3

Pulsed Nuclear Magnetic Resonance and Spin Echo

Pulsed/CW NMR Spectrometer PS2-A/B/C INSTRUCTOR S MANUAL

NMR Spectroscopy in Notre Dame

Nuclear Magnetic Resonance (NMR)

Physics of Imaging Systems Basic Principles of Magnetic Resonance Imaging II

Proton Nuclear Magnetic Resonance ( 1 H-NMR) Spectroscopy

What is NMR? Innovation with Integrity. Nuclear Magnetic Resonance NMR

Name Date Class ELECTRONS IN ATOMS. Standard Curriculum Core content Extension topics

Proton Nuclear Magnetic Resonance Spectroscopy

Lab 1: The Digital Oscilloscope

Chapter 13 Spectroscopy NMR, IR, MS, UV-Vis

NMR and IR spectra & vibrational analysis

Proton NMR. One Dimensional H-NMR. Cl S. Common types of NMR experiments: 1-H NMR

GE Medical Systems Training in Partnership. Module 8: IQ: Acquisition Time

Nuclear Magnetic Resonance Spectroscopy Notes adapted by Audrey Dell Hammerich, October 3, 2013

EDEXCEL NATIONAL CERTIFICATE/DIPLOMA UNIT 5 - ELECTRICAL AND ELECTRONIC PRINCIPLES NQF LEVEL 3 OUTCOME 4 - ALTERNATING CURRENT

Upon completion of this lab, the student will be able to:

2. Spin Chemistry and the Vector Model

PROTON NUCLEAR MAGNETIC RESONANCE SPECTROSCOPY (H-NMR)

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

Signal to Noise Instrumental Excel Assignment

Experiment 8: Undriven & Driven RLC Circuits

GenTech Practice Questions

Current Probes, More Useful Than You Think

Electronic Supplementary Information

8.1 Relaxation in NMR Spectroscopy

Lab 9: The Acousto-Optic Effect

RF Measurements Using a Modular Digitizer

NMR Signal Properties & Data Processing

An Adaptive Signal Processing Approach to Dynamic Magnetic Resonance Imaging

Preface. (modified from Alastair G. Smith, Surgeons Hall, Edinburgh, October 1939) H. H. Schild

Practical guide for quantitative 1D NMR integration Eugenio Alvarado, University of Michigan, 05/10/10

Application of Nuclear Magnetic Resonance in Petroleum Exploration

INTERFERENCE OF SOUND WAVES

AP PHYSICS 2012 SCORING GUIDELINES

INFRARED SPECTROSCOPY (IR)

5. Measurement of a magnetic field

NMR Nuclear Magnetic Resonance

Wireless Medical Telemetry Laboratory

= δx x + δy y. df ds = dx. ds y + xdy ds. Now multiply by ds to get the form of the equation in terms of differentials: df = y dx + x dy.

Ultrasonic Wave Propagation Review

Atomic Structure Ron Robertson

E/M Experiment: Electrons in a Magnetic Field.

MRI. Chapter M. Contents M.1

Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist.

Determining the Structure of an Organic Compound

Signal Manipulation. time domain NMR signal in MHz range is converted to khz (audio) range by mixing with the reference ( carrier ) frequency

Primer on Magentic Resonance Imaging Tomography c 2004

Pulsed Nuclear Magnetic Resonance (version February 2008)

Relaxation Can T, Be Longer Than T,?

Transcription:

Nuclear Magnetic Resonance and the Measurement of Relaxation Times of Acetone with Gadolinium Xia Lee and Albert Tsai June 15, 2006 1

1 Introduction Nuclear magnetic resonance (NMR) is a spectroscopic technique frequently employed by chemists to elicit the structures of chemicals, especially organic molecules, in a liquid or solid form. In the field of medicine, where a 2 dimensional version of NMR termed magnetic resonance imaging (MRI) is employed, this technique has also had significant impacts on the development of non-invasive diagnosis of diseases. First introduced in the 1940 s by Felix Bloch and Edward Purcell to measure the magnetic moment of nuclei in liquids and solids, NMR is based on protons in a nucleus having an intrinsic spin angular momentum, thus a magnetic moment. When a constant magnetic field is applied, a nucleus will resonate like a mechanical oscillator when driven in to an excited energy state by an EM wave (in the radio frequency for NMR) at the correct frequency, which is determined by the strength of the magnetic field and the magnetic moment of the nucleus. Traditionally, NMR was done by sweeping the magnetic field strength while applying a continuous EM wave and measuring where the signal from the nuclei emitted peaked (continuous wave NMR, or CW NMR). However, modern NMR is usually done by applying a pulse of RF waves (pulsed NMR), which contains a broad spectrum of frequency components, and then measuring the resulting radio signal emitted from the nuclei after the pulse. This signal is termed the free induction decay (FID) as it decays similar a damped oscillator. However, modern NMR is usually done by applying a pulse of RF wave (pulsed NMR), which contains a broad spectrum of frequency components, and measuring the radio signal, termed the free induction decay (FID) as the signal decays like a damped oscillator, emitted from the nuclei after the pulse. Using a Fourier transform of the FID signal, the resonance frequency of the nuclei can be determined. Moreover, pulsed NMR can provide information concerning the physical properties of the nuclei measured. One of the properties is the relaxation times, or the time required for the nuclei in the sample to return to their ground state after being saturated by a strong EM pulse, of the sample. In this experiment, we attempt to measure the relaxation times of acetone, using proton NMR, with various concentrations of gadolinium added as well as glycerol, which will serve as a standard to compare our results to measurements from other experiments. In this experiment, we attempt to measure, using proton NMR, the relaxation times of acetone with various concentrations of gadolinium added, as well as glycerol, which will serve as a standard by which to compare our results to measurements from other experiments. 2 Theory of Experiment For a formal and comprehensive treatment on the theory of NMR and relaxation times, please refer to a textbook on the subject of NMR spectroscopy [1]. This section will instead focus primarily on the theory directly relevant to our experiment. 2

2.1 Measuring T 1 To measure T 1, the spin-lattice relaxation time, a 90 o pulse is first applied to the sample so that the z-component of the total magnetization vector is rotated onto the xy-plane, which induces a FID. Then another 90 o pulse is applied after a certain time τ to again rotate the magnetization that decayed back to the z-axis into the xy-plane, where the precession of the magnetic moment becomes measurable. The 90 o pulse can be repeated in series to obtain a clearer signal through averaging. Mathematically, this process can be described by a simplified version of the Bloch equations in the z-component [2]: dm z = M 0 M z dt T 1 This first-order differential equation has a solution of: Linearizing this equation gives: M z (τ) = M 0 ( 1 e τ/t 1 ) ( ln 1 M ) z (τ) = τ M 0 T 1 M z is directly related to the signal strength of the FID, and M 0 can be measured from the FID by applying ( a 90 o pulse ) after allowing the sample to settle for several seconds. Therefore, by plotting ln 1 Mz(τ) M 0 versus τ, T 1 can be derived from the slope of the linear fitting function. 2.2 Measuring T 2 In theory, T 2 can be measured by examining the envelope of the FID directly. However, inhomogeneity in the magnetic field applied causes the individual magnetic moments of the nuclei to precess at slightly different frequencies and de-phase, where they will begin to partially cancel each other, as the signal decays. This will reduce the component of the total magnetization vector in the xy-plane faster than by T 2 alone. This effective decay time is termed T 2 inhomogeneous and is always less than the pure T 2. To solve this problem, a 180 o pulse is applied after a time τ after the initial 90 o pulse. This reverses the phase of the individual magnetic moments of the nuclei so that the higher frequency moments now lag the lower frequency ones and vice versa. After a time τ following the 180 o pulse, the magnetic moments will be in-phase again and generate a signal, a spin echo, whose strength depends solely on the pure T 2 and the time that has elapsed since the 90 o pulse (2τ). Mathematically, this process can also be described using a simplified version of the Bloch equations in the transverse direction [2]: dm xy dt = M xy T 2 3

This equation has a solution of: M xy (τ) = M xy0 ( e τ/t 2 ) Linearizing this equation gives: ( ) Mxy (τ) ln = ln (M xy (τ)) ln (M xy0 ) = τ M xy0 T 2 By plotting ln (M xy (2τ)) versus 2τ, T 2 can be obtained from the slope of the linear fitting function. 2.3 Effects of Gadolinium on Relaxation Times The longitudinal relaxation process (T 1 ), termed the spin-lattice relaxation, is determined by the time required for an excited proton in a molecule to transfer its spin angular momentum to other materials in its environment. Gadolinium, having 7 unpaired electrons in its highest f orbital, is a paramagnetic atom that provides a route to quickly allow the protons to transfer their spin angular momentum. Thus, adding a gadolinium compound into a solution is expected to decrease the T 1 of the solution. On the other hand, the transverse relaxation process (T 2 ), termed the spin-spin relaxation, is determined by the time needed for excited protons in a molecule to transfer their spin angular momentum to each other. Since this entirely an intra-molecular process, addition of gadolinium into the solution should have little or no effect on T 2. 3 Results 3.1 Experimental Setup The sample is placed in a glass test tube and secured in a sample holder between two Helmholtz coils where the magnetic field generated by the coils is relatively homogeneous. The Helmholtz coils are powered by a DC power supply with a voltage of approximately 35V and a current of 2.6A (the current was adjusted to obtain the optimal signal when the samples were pulsed). Inside the sample holder, there is a set of coil for transmitting the pulsed RF wave used to excite the sample and another set of coil to receive the signal emanating from the sample. The two signal coils are arranged perpendicular to the magnetic field generated by the Helmholtz coils and to each other. Figure 1 illustrates how the different components of the experimental apparatus are connected. A computer running LabView generates an electric pulse (5V) used to gate the signal from an RF signal generator set at 8MHz (8MHz is approximately the resonance frequency of protons in a 1T field). The gate is opened during the pulse and sends the 8MHz RF pulse through an amplifier to the transmitting coil sample. After the pulse, the signal from the sample is received through the receiver coil and sent to a demixer. The demixer filters out the artifact signals generated by the strong RF pulse and prevents those 4

signals from saturating out the signals from the samples. The phase shifter corrects the for the phase difference caused by the two signal paths from the RF generator having different path lengths. The final signal is then viewed on an oscilloscope. The gating signal is also connected to the oscilloscope for triggering purposes. 3.2 Measurements for T 1 To determine T 1, M z, which is directly proportional to signal strength in volts, at several different τ s were measured. Figure 2, using data for pure acetone recorded directly from the oscilloscope, demonstrates how these measurements were done on the oscilloscope. To recover a clean signal, signals from 32 90 o pulses with the same τ were averaged together. The difference in voltage between the first two extrema clearly identifiable as parts of the FID signal was recorded (between the two horizontal red lines under the arrows in Figure 2). After a set of measurements (usually ten data points) for a sample had been obtained, the M 0 for that sample was measured using a similar technique, although the signal in this case was not averaged. The points were then plotted and a least-square linear fitting was done to them, as demonstrated in Figure 3. The values for T 1 derived from repeating the above steps for all the samples are: 12.8±9ms for glycerol, 247±6ms for pure acetone, 69.0±6.2ms for 5ml acetone with 0.01g Gd added, 52.4±4.2ms for 5ml acetone with 0.05g Gd added, and 10.5±0.6ms for 5ml acetone with 0.1g Gd added. 3.3 Measurements for T 2 To obtain T 2, the screen of the oscilloscope was centered to a time 2τ after the initial 90 o pulse (or τ after the 180 o pulse) for each measurement. The strength of the spin echo was then measured as illustrated in Figure 4. The voltage difference between the central peak and the peak that gave the largest voltage difference of the two that surrounded it was recorded as M xy (2τ) (between the two horizontal red lines under the arrows in Figure4). After such a set of M xy at different 2τ s was obtained from a sample, the points were then plotted and a least-square linear fitting was done as in Figure 5. The values for T 2 obtained from repeating the above steps for all the samples are: 10.7±0.4ms for glycerol, 13.2±1.4ms for pure acetone, 10.6±0.3ms for 5ml acetone with 0.01g Gd added, 9.40±0.43ms for 5ml acetone with 0.05g Gd added, and 9.68±0.73ms for 5ml acetone with 0.1g Gd added. 4 Discussions The T 1 and T 2 values obtained during this experiment are summarized in Table 1. Compared to two other reports that reported relaxation times for glycerol [3, 6], our T 1 and T 2 appear to be in general agreement (within a factor of 2) with the values reported 5

in those reports. Unfortunately, no published materials containing the relaxation time of glycerol were found. Therefore, it may be difficult to determine the accuracy of our calculated relaxation times. For acetone, T 1 decreased as more Gd was added; this result agrees with the expectation that adding a paramagnetic material will shorten the spin-lattice relaxation time. Between 0.05g and 0.1g of Gd, there is a sharp drop (by a factor of 5) in T 1 when Gd concentration is doubled, suggesting there may be a critical concentration of Gd when a dramatic drop in T 1 occurs. Nevertheless, until the T 1 of more concentrations can be determined, the preceding proposal remains highly uncertain. T 2 for acetone remained relatively constant throughout the entire Gd concentration range tested. This agrees well with the prediction that spin-spin relaxation is not significantly affected by the proton s external environment. In conclusion, the results obtained in this experiment appear to be in good agreement with the expected outcomes. Nevertheless, as we are unable to locate published relaxation times for acetone and glycerol, our experiment should be regarded as only a moderate success. 5 Materials and Methods 5.1 Sample Preparation The glycerol and acetone used were taken from the general chemical stock in the Physics 173 lab at the University of California, San Diego. For each sample, 5ml of the desired sample was pipetted from the stock solution into a 15ml glass test tube. The gadolinium compound used was purchased from Sigma-Aldrich and the name of the product is Resolve- Al TM Gd (chemical formula Gd(OCC(CH 3 ) 3 CHCOC(CH 3 ) 3 ) 3, CAS number 14768-15-1). The Gd compound was weighted using an electronic balance and added to the acetone in the sample test tube. The solution was then lightly vortexed to dissolve the Gd compound. The prepared samples were then placed into the sample holder in the NMR machine and measurements were taken. 5.2 Oscilloscope Data The oscilloscope used, a Tekronix TDS3034B, allows waveforms displayed on its screen to be stored to a disk. The format chosen to store the waveforms is a standard CSV (comma separated values) file format. Microsoft Excel was used to plot the stored data. 5.3 LabView Configuration and Pulse Generation For measuring T 1, a single counter under the control of LabView, with adjustable pulse width and delay, was used to generate the pulse. To generate a spin echo, signals from two independent counters, where the first counter gated the second one, were combined through an XOR gate. 6

The proper duration for a 90 o pulse was found by searching for the shortest pulse width where the FID signal from a sample was at a maximum. 180 o is two times the width of a 90 o pulse. 5.4 Error Propagation In general, if a value is derived from a function as such: y = f (x 1, x 2,...,x n ) and each of the variable of the function, x i, has an error, assumed to be Gaussian, δx i, then the error in the final value is [4]: [ δy = n ( ) ] 2 f δx i x i so In the case for T 1, i=1 ( y = f (M z, M z0 ) = ln 1 M ) z M z0 For T 2, ( ) 2 ( ) 2 f f δy = δm z + δm 0 M z M 0 ( = M ) 2 ( z M z δm z + M 0 M z M 0 (M 0 M z ) δm 0 ) 2 so y = f (M xy ) = ln(m xy ) δy = ( ) 2 f δm xy = δm xy M xy M xy Since the slope of the linear function, a, used to fit the data for both T 1 and T 2 is the negative inverse of the actual T values, the error in the T values become: 5.5 Linear Fitting δt = δa a 2 Mathematica 5.2 from Wolfram Research was used to fit the data sets to their best fitting linear functions using the least-squared method. This is done using the Regress function in the Statistics standard add-on package in Mathematica, and the data points were weighted according to the inverse of their variance when calculating the least-square of the fit. Please 7

refer to the documentation in Mathematica for information on how to use the Regress function [5]. 8

Tables and Figures Sample T 1 (ms) T 2 (ms) Glycerol 12.8±0.9 10.7±0.4 Acetone 247±6 13.2±1.4 5ml Acetone + 0.01g Gd 69.0±6.2 10.6±0.3 5ml Acetone + 0.05g Gd 52.4±4.2 9.40±0.43 5ml Acetone + 0.1g Gd 10.5±0.6 9.68±0.73 Table 1: 1: T 1 and T 2 for the samples measured. Figure 1: Diagram of experimental setup. 9

Figure 2: Measuring M z from a FID. The blue line is the signal from the receiver coil, and the yellow line (not to scale, it is actually 5V in height) is the gating signal from the computer. 10

T1 5ml Acetone 0 10 20 30 40 50 60 70 80 90 100-0.1-0.2 ln(1-v/m0) -0.3-0.4-0.5-0.6 y = ax + b a = 0.00405 ± 0.00010 b = 0.190 ± 0.005 R = 0.996 T = 247 ± 6ms 1 2 M = 71.4± 1mV 0-0.7 Time (ms) Figure 3: Linear fitting for T1. Figure 4: Measuring M xy using a spin echo. 11

T2 0.05g Gd in 5ml Acetone 4 3.5 3 2.5 ln(v) 2 1.5 1 0.5 y = ax + b a = 0.106 ± 0.005 b = 3.95± 0.03 R = 0.983 T = 9.40± 0.43ms 2 2 0 2 4 6 8 10 12 14 16 18 20 2*Time (ms) Figure 5: Linear fitting for T 2. 12

References [1] Harald Günther. (1995) NMR Spectroscopy, 2nd ed. West Sussex, United Kingdom: John Wiley & Sons. [2] Hornak JP. The Basics of NMR. http://www.cis.rit.edu/htbooks/nmr/ June 14, 2006. [3] Frazier R. The Measurement of Relaxation Times in Mineral Oil and Glycerin Using Pulsed Magnetic Nuclear Resonance. http://frazier.home.cern.ch/frazier June 8, 2006. [4] Taylor JR. (1997) An Introduction to Error Analysis, 2nd ed. Sausalito, CA: University Science Books. [5] Wolfram Research, Inc. (2005) Mathematica, Version 5.2. Champaign, IL: Wolfram Research, Inc. [6] Magat K & Mandyam V. (2004) A look at the utility of pulsed NMR. http://physics.ucsd.edu/neurophysics/courses/physics_173_273/ June 15, 2006. 13