Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter



Similar documents
CHAPTER 5 QC Test For Radiographic Equipment. Prepared by:- Kamarul Amin bin Abu Bakar School of Medical Imaging KLMUC

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

5.2 ASSESSMENT OF X-RAY TUBE LEAKAGE RADIATION AND X-RAY TUBE OUTPUT TOTAL FILTRATION

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

Image Quality and Radiation Dose for Intraoral Radiography: Hand-Held Held (Nomad), Battery Powered

Role of the Medical Physicist in Clinical Implementation of Breast Tomosynthesis

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION SOURCES

X-Mind. Instinct for perfection

IAC Ch 41, p.1. Procedure means a stereotactically guided breast biopsy performed on a patient for diagnostic purposes.

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

Irradiation Field Size: 5cmX5cm 10cmX10cm 15cmX15cm 20cmX20cm. Focus-Surface Distance: 100cm. 20cm Volume of Ion Chamber : 1cmX1cmX1cm

SECTION 1: REQUIREMENTS FOR CERTIFICATES OF COMPLIANCE FOR CLASSES OF RADIATION APPARATUS

QUALITY ASSURANCE PROGRAMS FOR DIAGNOSTIC RADIOLOGY FACILITIES. (a) Applicability. (b) Definitions

Recognition. Radiation Survey Objectives. Objectives. Part 1 Documentation Radiation Source Survey Objectives Radiation Detectors Techniques

X-ray Production. Target Interactions. Principles of Imaging Science I (RAD119) X-ray Production & Emission

REGULATION: QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS N.J.A.C. 7:28-22

PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT

Quality Control and Maintenance Programs

SUBCHAPTER 22 QUALITY ASSURANCE PROGRAMS FOR MEDICAL DIAGNOSTIC X-RAY INSTALLATIONS

Radiation Dose Measurements in Routine X ray Examinations

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

INTRODUCTION. A. Purpose

Digital Breast Tomosynthesis QC Requirements

IAEA-TECDOC-1447 Optimization of the radiological protection of patients: Image quality and dose in mammography (coordinated research in Europe)

EPA Protocols: Compliance testing of dental X-ray apparatus used for plain radiography

Production of X-rays and Interactions of X-rays with Matter

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

CT: Size Specific Dose Estimate (SSDE): Why We Need Another CT Dose Index. Acknowledgements

X-ray Imaging Systems

Manual for Accreditation of Agencies for Quality Assurance of Diagnostic X-ray Equipment

Veraview IC5 HD High definition, digital imaging excellence. Thinking ahead. Focused on life.

Radiation Safety Characteristics of the NOMAD Portable X-ray System

Diagnostic X-ray facilities as per quality control performances in Tanzania

I.7. SOFT TECHNIQUE STILL IN USE IN CHEST RADIOGRAPHY PROS AND CONS

Quality Assurance. The selection of the equipment. Equipment Specifications. Medical Exposure Directive 97/43 Euratom. Quality Assurance Programme

TA3 Dosimetry and Instrumentation. Evaluation of the Field Size in Dental Diagnostic Radiology System ANDRADE PS 1, POTIENS MP 1

HIGH PERFORMANCE MOBILE SURGICAL C-ARM KMC-950

Radiation Strip Thickness Measurement Systems

Dose Measurement in Mammography; What are we measuring? David E. Hintenlang, Ph.D. DABR University of Florida

SUMMARY OF CURRENT UK LEGISLATION AND GUIDELINES

kv-& MV-CBCT Imaging for Daily Localization: Commissioning, QA, Clinical Use, & Limitations

Name: Date: Team: Lab Experiment # 3. Focused Grid Positioning Errors. Computed Radiography and Direct Digital Radiography

The Exposure Index and Its Standardization

Clinical Physics. Dr/Aida Radwan Assistant Professor of Medical Physics Umm El-Qura University

Q.A. Collectible. Sponsored by CRCPD s Committee on Quality Assurance in Diagnostic X-Ray (H-7)

Update on ACR Digital Mammography QC Manual

DS 600. A contact free flux gate based current measurement sensor 600A rms

WEATHER STATION FOR SOLAR FARM MONITORING

Quality Assurance of Radiotherapy Equipment

TISSUE MIMICKING GEL QUALITY LE PHANTOM SERIES DESIGN. performance the ultrasound labs ofand. icking material has the same attenuation mim-

Portuguese Health Physics Society: Calibrating Pen Dosimeters with and without a Phantom

X-ray Imaging Systems

Atomic and Nuclear Physics Laboratory (Physics 4780)

Scan Time Reduction and X-ray Scatter Rejection in Dual Modality Breast Tomosynthesis. Tushita Patel 4/2/13

Page: 1 of 6 Page: 1 of 6

Purchasing a cardiac CT scanner: What the radiologist needs to know

Cathode Ray Tube. Introduction. Functional principle

Worldwide Quality Assurance networks for radiotherapy dosimetry

Owner s Manual

Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer Patented

Vacuum Evaporation Recap

Infrared / K Thermometer Remote Temperature Measurement ( Non-Contact ) 1327K INSTRUCTION MANUAL

MAVO-MONITOR / MAVO-SPOT Instrument Set for Contact or Distant Measurements of Luminances

Copyright 2007 Casa Software Ltd. ToF Mass Calibration

Measurement of Charge-to-Mass (e/m) Ratio for the Electron

PHYS 222 Spring 2012 Final Exam. Closed books, notes, etc. No electronic device except a calculator.

Radiographic Grid. Principles of Imaging Science II (RAD 120) Image-Forming X-Rays. Radiographic Grids

Patient Exposure Doses During Diagnostic Radiography

RADIATION THERAPY. Dosimetry Pioneers since 1922 NEW

Application of Digital Radiography for the Detection and Classification of Pneumoconiosis

Chapter XIII. PDD TAR TMR Dose Free Space

How To Use A Ds340 (Dsp)

In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the

Compliance Guidance for FLUOROSCOPIC QUALITY CONTROL

Hygro-Thermometer + InfraRed Thermometer Model RH101

Ch. 227 ANALYTICAL X-RAY EQUIPMENT 25

7.0, 4 Wire Touch Panel

PERFORM-X DIGITAL X-RAY SYSTEM PERFORM-X. with ceiling mounted tube support PERFORM-X

Quality control tests of diagnostic radiology equipment in Hungary, and its radiation protection aspects

DICOM Correction Item

Dupline Car Park System Type GP x 724-US 3-Colour Ultrasonic Sensor

Lab 4: Magnetic Force on Electrons

E420 SHAFT ALIGNMENT. Entry level measurement and alignment system for rotating machines BUILT-IN WIRELESS UNITS LARGE COLOUR DISPLAY QUICK START

C-10 Radiation Monitoring Network

Dual Laser InfraRed (IR) Thermometer

Easy Quality Control with - PTW Equipment. Code of Practice - Quality control of X-ray equipment in diagnostic radiology

Annual MQSA Inspection Questions NMQAAC April 19, 2004

RADIATION SAFETY STANDARD

Dual Laser InfraRed (IR) Thermometer with Color Alert

Physics testing of image detectors

Using Current Transformers with the 78M661x

MQSA Quality Control Manual for Monochrome Displays for Mammography

I-Max Touch Range. PAN / CEPH / 3D digital panoramic unit. Evolutive 3 in 1 panoramic unit

We bring quality to light. MAS 40 Mini-Array Spectrometer. light measurement

Introduction. Digital Mammography Detector QC FFDM. What is FFDM QC and why is it. important? What to know before you start. Eric A. Berns, Ph.D.

Quality Control of Full Field Digital Mammography Units

DOSES TO EYES AND EXTREMITIES OF MEDICAL STAFF DURING INTERVENTIONAL RADIOLOGY PROCEDURES

Radiation Detection and Measurement

Transcription:

Study the Quality Assurance of Conventional X-ray Machines Using Non Invasive KV meter T.M.Taha Radiation Protection Department, Nuclear Research Center, Atomic Energy Authority, Cairo.P.O.13759 Egypt. ABSTRACT The aim of the present paper is to investigate some factors affecting on quality assurance of conventional x-ray such as reproducibility of tube voltage, dose out put, time, and x-ray tube efficiency, Accuracy of kvp, ma, time, and focal spot size. and half value layer.examinations of these factors are studied using non invasive kv meter NeroMax 8000 which connected with suitable ionization chambers. The quality assurance tests of X-ray diagnostic examination are measured and compared with the international tolerance. Key words. Quality Assurance, NeroMAX8000, X-ray INTRODUCTION The principle goal of quality assurance of x-ray machine is obtain accurate and timely diagnosis. The secondary goals are minimization of radiation exposure and obtain high image quality. This can be assess by performance the x-ray machine by optimum operating parameters such as reproducibility of tube voltage, dose out put, time, x-ray tube efficiency, Accuracy of kvp, ma, time, focal spot size. and half value layer Many authors publish work concerning quality assurance of x-ray machine, AAPM, (1) describes quality assurance protocol for diagnostic x-ray equipment at the radiologic technologist level. Some important parameters in diagnostic x-ray such as film processor monitor, exposure time, mas reciprocity, peak tube potential, and x-ray out put and beam quality, and focal spot size were described. J.Bosnjak et.al 2008 (2) has been systematically performed quality control surveys of diagnostics imaging equipment in Republic of Srpska. Harold Elford Johns 1961 (3) has been expressed the quality of x-ray in terms of the half value layer. The quality assurances of diagnostic x-ray are based on the Basic Safety Standard BSS (4) and International Commission of Radiological Protection, the use of diagnostic reference levels (DRL for patients, ICRP- Report No., 1966. (5) The aim of the present paper is to investigate some factors affecting on quality assurance of conventional x-ray such as reproducibility of tube voltage, dose out put, time, x-ray tube efficiency, Accuracy of kvp, ma, time, and check the focal spot size and half value layer. 105

Material and Method Reproducibility of dose, time and high voltage setting- parameters were measured with Nero Max 8000, which were placed on the couch inside the selected field size and six exposures were made. The reproducibility Pz was calculated using the formula. Pz = SD /Zav. 100% Where : SD is the estimator of standard deviation of a series of measurements dose[mgy], time [ms] or voltage [KV], Zav is the means value of the parameters measured [dose[mgy], time[ms] or voltage [KV]. Accuracy of tube voltage and time setting were examined for each machine. Three exposures were recorded for KV and time accuracy. The accuracy Rx was calculated using the formula. Rx= X m -X n /X n. % Where: Xm is the measured value of time [ms] or voltage [KV], and Xn is the nominal value of time [ms] or voltage [KV] Extensive measurements c made to assess of changes in KVp, ma, s and mas on reproducibility and linearity of radiation output. It formed over a range of clinical settings. Calibrated ionization chamber used to measure output expressed as ugy per mas, at a set distance, without backscatter. The linearity was checked using the following equation. X1-X2 /X1+X2 <0.1 Where X1 and X2 are two successive readings. KV, ma, Time accuracy were performed by made the operating conditions constant and change these parameters over a range of clinical settings. This work has been done in nuclear medicine department in Alkhair hospital, Mansora City, Aldaqhlia Country. Beam Quality Filtration of x-ray machine was measured at 70 KVp, 32 mas, 200 ma, 0.16 sec and source to detector distance 66 cm. The dose output was measured using Nero max -8000.The dose output was measured with different thickness. The thickness which gave half of zero doses is considered as the half value thickness. It is the thickness of some standard material required to reduce the intensity of a beam to one half its original value, Harold Elford Johns (1961) 3 106

Focal spot size using determination using STAR resolution pattern Focal spot is defined as the area on the target of the x-ray tube which the electron stream strikes and from which x-rays are emitted. called also focus. The imaging diameter of star resolution pattern was measured via fixation the star resolution pattern in outrance of x-ray beam and record its image on the x-ray film. From the image the focal spot size for x-ray machine under study was calculated using the following equation. f = Π θ D / 180 (M-1) M = Imaging diameter /Original diameter D: the internal diameter between starting and ending the resolution θ = 2 0 F = focal spot size, mm X-ray Beam Alignment The x-ray beam alignment was checked using perpendicularity of reference axis with table/bucky with field size of 20 x 20 cm, 5 mas, 70 KVp and 100 cm source to detector distance. The quality assurance procedures of x-ray equipment were carried out for six machines of the same type and manufacture..neromax 8000 has been used to check the quality assurance of x- ray machine. The NERO Max consists of the control console, detector cable, two filter cards, ma's leads, Excel Add-in, AC adapter, and HVL plates; the compact control console houses the rechargeable battery, supper bright easy to read backlit display, eight control buttons, and the sophisticated electronic necessary for accurate, reproducible measurements. Connectors or power input, RS-232, printer, scope output and the NERO mix detector contains sensors for simultaneously measuring KV, exposure or rate and am or mass. Solid state detectors are used to measure KV, An ion chamber, located in the top of the detector, is used for exposure / rate measurements. The filter cards contain the various filters needed to accurately measure kilo voltage. Each filter card is coded so that the NERO max knows which filter is in use and its position.. Reproducibility RESULTS AND DISCUSSION X-ray tube out put and reproducibility (difference between the measured and nominal value) for high voltage, time,, Dose output for six x-ray machines were carried out six times for machines numbered from M1-M2 and presented as shown in table (1) 107

Table 1 the results of Quality Assurance control: efficiency of X-ray tube, reproducibility of dose, time and high voltage. Code FFD (m) Efficiency (µgy/mas) FFD [cm] Reproducibility [%] Dose Time High Voltage M1 100 62 100 0.1 0.20 0.70 M2 100 80 100 0.2 - - M3 100 76 100 0.4 0.01 0.10 M4 100 67 100 0.1 0.3 0.60 M5 100 78 100 0.1 3.00 0.60 M6 100 62 100 0.7 0.20 0.10 Reproducibility of dose output was ranged from 0.1 to 0.7 %, of time was ranged from 0.01 to 3 % and of high voltage was ranged from 0.1% to 0.7% which is lower than the tolerance ;limit The efficiency of the x-ray tubes were ranged from 62% to 80% which were greater than 25 µgy/mas for 1 m FDD, Focus to Detector Distance and at 80 kv tube voltage (6). KV and Time Accuracy KV accuracy for different settings of six x-ay machines voltage was examined by setting the source to detector distance at 66 cm of exposure, time at 0.1 sec for different KV intervals from 50-120 KV and average of KV accuracy was presented as shown in table 2. In addition Time accuracy for x-ray machines was checked by variation the time interval from 50-1000 second as shown in table 2. Table. 2 KV and time accuracy for different settings of six x-ray machines Machine No. Mean kvp Accuracy % Error Mean time Accuracy % Error M1 1.50 3.00 M2 20.00, Recalibrated and adjusted to 5.00 4.21 M3 5.00 4.20 M4 1.74 3.70 M5 4.00 4.10 M6 1.30 3.80 KV accuracy is good at all KVp stations for six machine except one of the examined machines gave accuracy of 20 % which is higher than the tolerance limit.(± 5%). That mean this machine needs calibration. It recalibrated and adjusted to 5% KV accuracy. Time accuracy is good at all time settings stations for all examined machine which is lower than the tolerance limit. (± 10%). (7) 108

Beam Quality Half value layers were calculated as shown un table 3. that thickness has an ability to prevent the hazard of soft x-ray, by reducing the surface doses during x-ray imaging. Table. 3. Measurement of the Half value thickness Machine No. Half Value Layer, mm M1 2.31 M2 2.40 M3 2.70 M4 2.32 M5 2.60 M6 2.40 HVL is exceed the minimum value, passed above 2.3 mm Al at 70 KeV. This is within the accepted value of IAEA.. Focal spot size using STAR resolution pattern The results of focal spot size estimation are presented as shown in table.4. The estimated focal spot size was compared with assigned value stated by the manufacture and was within the tolerance level ±0.3 of nominal focal spot (NNL). Table. 4 Focal spot sizes for some conventional x-ray machines Code "Nominal size" mm Acceptable size, mm Estimated size, mm M1 1.3 1.6 1.52 M2 1.0 1.3 1.088 M3 0.9 1.2 0.870 M4 1.0 1.3 1.07 M5 1.0 1.12 M6 1.2 1.5 0.87 The estimated sizes of focal spot are lower than the acceptable values X-ray Beam Alignment The beam alignment was calculated as shown in table 5. 109

Table 5. X-Ray Beam Allignment % Machine No. Perpendicularity, cm M1 0.7 M2 0.4 M3 0.6 M4 0.3 M5 0.5 M6 0.6 The alignment of x-ray machines was investigated using perpendicularity test tool The alignment was ranged from 0.3 to 0.7 cm which is lower than the tolerance level (± 1 cm at 100 cm at FDD) (7) CONCLUSION The quality assurance tests of x-ray machines under study obtained accurate and timely diagnosis. The measured output doses were within the international reference doses. The quality assurance program should generalize for all hospitals to ensure the quality of the x-ray machines under services. REFERENCES (1) A merican Association of Physicists Medicine " Basic Quality Diagnostic Radiology " Diagnostic Radiology Committee Task Force On Quality Assurance Protocol, AAPM Report No.4, 1981. (2) J. Bosnjak, O. Ciraj-Bjelac and Herzegovina (Republic of Srpska) (3) Harold Elford Johns" The Physics of Radiology" second edition. Charles C Thomas, Publisher, Publications number 419, 1961. (4) International Commission Of Radiological Protection" Basic Safety Standard for ionizing radiation " Safety Series: BSS115-1994. (5) International Commission Of Radiological Protection, ICRP-No-46 1966 (6) Polish Health Minster executive order dated 24. December 2002 (in Polish) Dz U.2002. Nr 241, poz.2098. (7) John R.Cameron" Medical Physics" John Wiley and Sons, Inc. 1978, p.390. 110