IFC 070 Technical Datasheet



Similar documents
OPTIMASS 6000 Technical Datasheet

OPTIMASS 1400 Technical Datasheet

HYDRUS ULTRASONIC METER

IFC 300 Technical Datasheet

Electro-Magnetic Flowmeters COPA-XT 2-Wire DT43

1 WATERFLUX - Comparisonon COO Rectangular Magmeter in water applications

Sentinel LCT. Panametrics Ultrasonic Flowmeter for Liquid Custody Transfer Measurement. GE Measurement & Control Solutions. Benefits.

DMH. Magnetic Inductive Flowmeter. for conductive fluids. Measuring range: m 3 /h. Accuracy: ± 0.3% of reading ± 0.

Measurement. SITRANS F US Inline. Transmitter FUS080/FUE080 4/219

Certified according to DIN EN ISO 9001 Technical Datasheet KVO Series Vortex Flow Meter

CORONA E / FLYPPER ELECTRONIC METER MULTI-JET

SHARKY 775 COMPACT ENERGY METER ULTRASONIC

Electromagnetic Flow Meter

How To Use A Flowmeter

RHE12. Hazardous Area Coriolis Mass Flow Transmitter. Features. Applications. Benefits

Duct Humidity Transmitter

Open channel & partially filled pipe Ultrasonic flow meters

Certified according to DIN EN ISO 9001 Technical Datasheet C-Flow Coriolis Mass Flow Meter

Digital inductive conductivity transmitter

LLB-OCMFM 2000 PORTABLE DOPPLER AREA-VELOCITY FLOWMETER

Non-invasive CLAMP-ON ultrasonic flowmeter for liquids

3DLevelScanner II TECHNICAL DATA SHEET

AMTRON E-30 Compact heat and cooling meter

SHARKY 775 COMPACT ENERGY METER ULTRASONIC

Technical Information RIA14

713 Open Channel Flowmeter and Transmitter. All-in-one package for open channel flow measurement.

Technical Data Sheet. 3DLevelScanner

SITRANS F M MAG 8000 Precise Metering with battery powered water meters. sitrans f

How To Read A Pipe From A Sensor

Electronic Control Devices The European Product Catalogue 2007

For Multi-Parameter Meters see mvx

WaterMAG. Electromagnetic Flowmeter for Water Applications. Model MCB10A OVERVIEW APPLICATIONS FEATURES. No. SS2-MCB

DREXELBROOK. USonic-R Series. Ultrasonic Level Measurement System. User Friendly

DMP 331. Industrial Pressure Transmitter for Low Pressure. Stainless Steel Sensor

Instructions Manual. Electromagnetic sensor Series FLOMID FX. instrumentation for fluids. R-MI-FlomidFX Rev.: 0 English version

PORTABLE FLOW METER KITS TDS-100H

MOUNTING AND OPERATING INSTRUCTIONS

Technical data. General specifications. Indicators/operating means. Electrical specifications Operating voltage U B Power consumption P 0 Interface

Ferrostat Speed Sensor Series DSF Explosion Proof Versions EEx

R4D0-FB-IA* FieldBarrier. Features. Assembly. Function. Connection IN + - S OUT Spur 2. Spur 1. Zone 1. Zone 0

Grant Agreement No SFERA. Solar Facilities for the European Research Area SEVENTH FRAMEWORK PROGRAMME. Capacities Specific Programme

Technical data. General specifications. Indicators/operating means. 30 Hz Multiplex operation 30 Hz / n, n = number of sensors, n 5

2.24. Pump Controller 704. Data Sheet

DiwiTherm, Resistance Thermometers Model TR75, with digital Display Battery or Solar Powered

Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters

BM SERIES WAVE COMPACT 6.3 GHZ CW SERIES PULSE RADAR LEVEL INSTRUMENTS (6.3 GHz)

Centro-Matic Automatic Lubrication Systems System Controls

SITRANS F flowmeters. SITRANS F M System information SITRANS F M electromagnetic flowmeters. 4/12 Siemens FI

Inductive sensor NI10-M18-Y1X-H1141

COMBIMASS. Technical Data COMBIMASS eco-bio +

FAQ s on Siemens Magmeters

Heating, air-conditioning, cooling

How To Control A Power Supply With A Mini Transmitter And Switch (Power Supply)

Electronic Control Devices The European Product Catalogue 2007

Power consumption (kw)

Instruction Manual. T3111 T3111L T3111P T3111Ex

SITRANS F flowmeters SITRANS F M. Battery operated water meter MAG /93

Material: Weight: Bearing Life: Shaft Loads: Operating Temp.: Storage Temp.: Shock: Vibration: Bump: Humidity: Enclosure Rating: Cable: Cable Glands:

SITRANS F flowmeters SITRANS F M

DSD 70 Dual Channel Hall Effect Speed Sensor for Railway Applications, compliant with EN 50155

T U R B I N E G A S M E T E R

Electronic pressure switch with display Model PSD-30, standard version Model PSD-31, with flush diaphragm

Material: Weight: Bearing Life: Shaft Speed: Starting Torque: Mass Moment of Inertia: Shaft Loads:

THE THERMAL FLOW METER, A GAS METER FOR ENERGY MEASUREMENT

Typical Aluminum GFM Mass Flow Meter

Physical Specifications (Custom design available.)

MTI10 Insertion and MTL10 In-line Thermal Mass Flowmeter and Temperature Transmitter

SITRANS F flowmeters SITRANS FUS080. Flowmeter FUE380 with approval. 4/202 Siemens FI Overview

Dimensions. LUC-M20 with F12 housing and process connection 2" approx. 86 AF60 G2 2 NPT Ø50. Additional dimensions see section dimensions.

Build your own flowmeter

OLI. Oil level indicator

Programmable set for Ethernet Modbus/TCP in IP67 TI-BL67-PG-EN-2

Analog signal converters CC-E I/I Current / current isolators

Flow switches Series AD/VH

Plastic tube flowmeters Series PT/PS

SITRANS F flowmeters SITRANS FUS080. SITRANS FUS380/FUE380/SONOCAL /179. flow sensor pipe, 4 transducers/transducer cables and a transmitter

T-SERIES INDUSTRIAL INCLINOMETER ANALOG INTERFACE

O C M III. Open Channel Flow Monitor

Magnetic field sensor for pneumatic cylinders BIM-INT-Y1X-H1141

OEM Pressure Transmitters for heavy-duty applications, Type MBS 1200 and MBS 1250

Mobrey Ultrasonic. MSP900SH Level and MSP900FH Flow Transmitters. Product Data Sheet February 2015 IP2032, Rev BA

Outdoor 33.6W Dual Port Passive Power-over-Ethernet Midspan For External Security Cameras and Wireless Access Points

How To Power A Schen

Take total control of your HVAC system with cost-saving flow and thermal energy meters

Clear Glass Sensor. Ordering Information E3S-CR67/62. Optimum Sensor for Detecting Transparent Glass and Plastic Bottles

Liquid Level Limit Switch liquiphant T FTL 20

Fiber Optic High Density ACC Sensor System Faser Optisches Beschleunigungs-Messsystem

Isolating Barriers for our sensors A5S1 in Ex-areas Series D461

Temperature sensor with integrated transmitter for industrial applications, MBT 3560

Mounting Instructions. Torque Transducer. A en

Low Frequency Velocity Sensor

Technical Data. Dimensions

Tubular Analog model resolution: 0,5 mm (SLOW mode), 1 mm (FAST mode) Right angle Repeatibility: 0,7 mm Vdc

System 6000 Electropneumatic Converter for Direct Current Signals Type 6116 i/p Converter

SITRANS F flowmeters SITRANS F M. Transmitter MAG 5000/ /32 Siemens FI Overview

Ron 3025 Crane Scale with 1"/25mm Display

Series P215PR/RM Direct- and Remote-Mount Pressure-Actuated Condenser Fan Speed Controllers for Single-Phase Motors

Smart Electromagnetic Flowmeter Open channel Flowmeter Detector

4 non-safe digital I/O channels 2 IO-Link Master V1.1 slots. Figure 1. Figure 2. Type code. TBPN-L1-FDIO1-2IOL Ident no

RISH CON - Hz. Salient Features : Application : Product Features: FREQUENCY TRANSDUCER

Transcription:

IFC 070 Technical Datasheet Electromagnetic signal converter Battery driven with low energy consumption Quick and easy to install and operate External data logger and GSM module for remote data transfer KROHNE

CONTENTS IFC 070 1 Product features 3 1.1 Solution for the water and wastewater industry... 3 1.2 Features... 5 1.3 Measuring principle... 7 2 Technical data 8 2.1 Technical data... 8 2.2 Measuring accuracy... 11 2.3 Dimensions and weights... 12 3 Installation 13 3.1 Notes on installation...13 3.2 Mounting position and flange deviation... 13 3.3 Vibration... 14 3.4 Magnetic field... 14 4 Electrical connections 15 4.1 Safety instructions... 15 4.2 Installation of converter... 15 4.3 Grounding... 15 4.4 Connection of signal cable... 16 4.5 Terminal assignment of converter... 17 5 Notes 18 2

IFC 070 PRODUCT FEATURES 1 1.1 Solution for the water and wastewater industry The IFC 070 electromagnetic signal converter is designed for use in water industry such as potable water distribution networks, for revenue billing and for water abstraction. It is battery powered, making it especially suitable for applications where no power connection is available and provides certainty in case of power failure. Optional a GSM and data logger module is available for the remote transmission of measurement data and status information. 1 Battery-powered totaliser 2 Rigid flow sensor 3

1 PRODUCT FEATURES IFC 070 Highlights Battery operated, stand alone water meter, battery life up to 15 years Excellent performance in low flow conditions and over a wide flow range Easy installation, short inlet and outlet runs, no filters and bi-directional flow Maintenance free operation, no moving parts, no wear and obstruction free Optional, battery powered data logger and GSM module for remote data transfer Industries Water abstraction Distribution of potable water Irrigation Applications Raw water, potable water, irrigation water Well chambers Monitoring of distribution networks Pipeline leakage detection Measuring water consumption 4

IFC 070 PRODUCT FEATURES 1 1.2 Features GSM and data logger module There is an increased demand by utilities for remote reading. Often water meters are installed at remote locations in the distribution network or below the ground for example in busy city centers. The IFC 070 can be supplied with a state-of-the-art data logger and GSM modem. The stored data is tranmitted (for example once a day) by SMS and can be forwarded to the customer's management system. Quick to install and easy to operate - Remote or Compact The IFC 070 signal converter can be supplied in either compact or remote design. In case of a remote design, the converter can be installed on the wall or on a pipe. The functionality of the compact and remote versions is identical. 5

1 PRODUCT FEATURES IFC 070 Low energy consumption The IFC 070 signal converter has an extreme low power consumption. It delivers precise and reliable measurements for many years operating on batteries. With a sampling rate of 1/15 Hz, the water meter can operate for more than fifteen years. Long term reliability In addition to a long battery lifetime of up to 15 years, the IFC 070 provides diagnostic information. The IFC 070 has two status outputs for self checking, battery warnings and counter overrun. 6

IFC 070 PRODUCT FEATURES 1 1.3 Measuring principle An electrically conductive fluid flows inside an electrically insulating pipe through a magnetic field. This magnetic field is generated by a current, flowing through a pair of field coils. Inside of the fluid, a voltage U is generated: U = v * k * B * D in which: v = mean flow velocity k = factor correcting for geometry B = magnetic field strength D = inner diameter of flow meter The signal voltage U is picked off by electrodes and is proportional to the mean flow velocity v and thus the flow rate q. The signal voltage is quite small (typically 1 mv at v = 3 m/s / 10 ft/s and field coil power of 1 W). Finally, a signal converter is used to amplify the signal voltage, filter it (separate from noise) and convert it into signals for totalising, recording and output processing. 1 Voltage (induced voltage proportional to flow velocity) 2 Electrodes 3 Magnetic field 4 Field coils 7

2 TECHNICAL DATA IFC 070 2.1 Technical data The following data is provided for general applications. If you require data that is more relevant to your specific application, please contact us or your local representative. Additional information (certificates, special tools, software,...) and complete product documentation can be downloaded free of charge from the website (Downloadcenter). Measuring system Measuring principle Application range Measured value Primary measured value Secondary measured value Faraday's law Electrically conductive fluids Flow velocity Volume flow Design Modular construction Compact version Remote version User interface Display Units Cable connections The measurement system consists of a flow sensor and a signal converter. It is available as compact and as remote version. More information about the sensor can be found in the relevant documentation. With WATERFLUX 3000 sensor: WATERFLUX 3070 C With OPTIFLUX 2000 sensor: OPTIFLUX 2070 C With WATERFLUX 3000 sensor: WATERFLUX 3070 F With OPTIFLUX 2000 sensor: OPTIFLUX 2070 F Maximum cable length 25 m / 75 ft 8 digits LCD Display of positive and negative counter, sum counter, flow rate Status indication for battery, flow / counter direction, empty pipe Volume in m 3, US Gallons Flow rate in m 3 /h, USGPM Standard: 2x M20x1.5 Optional: ½" NPT, PF½ Measuring accuracy Reference conditions Maximum measuring error Repeatability Medium: water Temperature: 20 C / 68 F Inlet section: 5 DN Operating pressure: 1 bar / 14.5 psig ±0.2% of measured value ±1.5 mm/s / 0.06 inch/s For detailed information on the measuring accuracy, see chapter "Measuring accuracy". ±0.1% (v > 0.5 m/s / 1.5 ft/s) 8

IFC 070 TECHNICAL DATA 2 Operating conditions Temperature Process temperature -5 +70 C / 23 +158 F Ambient temperature -40 +65 C / -40 +149 F Storage temperature -50 +70 C / -58 +158 F Chemical properties Physical condition Liquids Electrical conductivity 20 μs/cm Recommended flow velocity in pipeline: -9...9 m/s / -30...30 ft/s in sensor: -18...18 m/s / -59...59 ft/s Process conditions Raw water, ground- and surface water Potable water Irrigation water Installation conditions Dimensions and weights Materials Housing Connection box (remote versions only) Electrical connections Power supply Battery Typical lifetime Alarm Battery replacement In- and output Outputs Communication For detailed information see chapter "Dimensions and weights". Die-cast aluminium, polyurethane coated Die-cast aluminium, polyurethane coated Standard 1 Lithium battery (D-cell) Optional 2 Lithium batteries (D-cell) External battery pack with 4 lithium batteries (D-cell) See graph at the end of this table. Pre-alarm 1 year before battery depletion and final alarm Possible without loss of totalizer data 2 Passive pulse outputs for remote totalising: f 500 Hz; I 10 ma; U: 5...24 VDC (P 100 mw) 2 Passive status outputs: I 10 ma; U: 5...24 VDC (P 100 mw) External data logger / GSM module, SMS protocol to: SCADA system (at customer's site) OPC server (to be connected to customer's OPC client) PCWin (mini-scada can be supplied by KROHNE) 9

2 TECHNICAL DATA IFC 070 Approvals and certifications CE Sign Hazardous areas Non-Ex ATEX Other approvals and standards This device fulfills the statutory requirements of the EC directives. The manufacturer certifies successful testing of the product by applying the CE mark. Standard Not available Protection category acc. to IEC IP 66/67 (NEMA 4/4X/6) 529 / EN 60529 Shock- and vibration resistance IEC 68-2-3 Electromagnetic compatibility Directive: 89/336/EEC Harmonized standard: EN 61326-1 : 2006 Low voltage directive Directive: 2006/95/EC Harmonized standard: EN 61010 : 2001 Custody transfer Standard: without Europe: MI-001 pending Worldwide: OIML R-49 pending Typical lifetime of batteries (at 25 C) Figure 2-1: X = Sampling interval in seconds, Y = typical lifetime in years 1 dual battery pack 2 single battery 10

IFC 070 TECHNICAL DATA 2 2.2 Measuring accuracy Figure 2-2: X [m/s]: flow velocity Y [%]: deviation from the actual measured value 1 with WATERFLUX 3000 water meter 2 with OPTIFLUX 2000 flow sensor 11

2 TECHNICAL DATA IFC 070 2.3 Dimensions and weights Compact version a = 170 mm / 6.7" b = 122 mm / 4.8" c = 140 mm / 5.5" Sizes of sensor can be found in the relevant datasheet. Remote version, signal converter b = 122 mm / 4.8" c = 335 mm / 13.2" H = 310 mm / 12.2" 12

IFC 070 INSTALLATION 3 3.1 Notes on installation Inspect the cartons carefully for damage or signs of rough handling. Report damage to the carrier and to the local office of the manufacturer. Check the packing list to check if you received completely all that you ordered. Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate. 3.2 Mounting position and flange deviation Figure 3-1: Mounting position and flange deviation 1 L max 2 L min Mount flowmeter either with converter aligned upwards or downwards. Install flowmeter in line with the pipe axis. Pipe flange faces must be parallel to each other. Max. permissible deviation of pipe flange faces: L max - L min 0.5 mm 13

3 INSTALLATION IFC 070 3.3 Vibration Figure 3-2: Avoid vibrations 3.4 Magnetic field Figure 3-3: Avoid magnetic fields 14

IFC 070 ELECTRICAL CONNECTIONS 4 4.1 Safety instructions 4.2 Installation of converter 4.3 Grounding All work on the electrical connections may only be carried out with the power disconnected. Take note of the voltage data on the nameplate! Observe the national regulations for electrical installations! Observe without fail the local occupational health and safety regulations. Any work done on the electrical components of the measuring device may only be carried out by properly trained specialists. Look at the device nameplate to ensure that the device is delivered according to your order. Check for the correct supply voltage printed on the nameplate. Only applicable for remote versions. Mount converter with mounting plate on wall or standpipe. Keep distance between sensor and signal converter as short as possible. Observe length of the delivered signal cable. Figure 4-1: Grounding 15

4 ELECTRICAL CONNECTIONS IFC 070 4.4 Connection of signal cable The signal cable is only used with remote versions. The standard cable includes both electrode and field current leads, the optional type A / B cable is only used for the electrodes. In that case, the field current cable is no part of the supply. Figure 4-2: Cable connection at converter side, standard cable 1 cable length: 13 cm / 5" 2 cable length: 5 cm / 2" 3 brown + white cable, used for field current 4 purple and blue cable, used for electrode signals 5 cable length: 8 cm / 3" 6 Shield (terminal 1 of connector X2 + U-clamp Prepare appropiate cable lengths (1...3) Connect the shield to the U-clamp, the brown cable to terminal 7 and the white to terminal 8. Connect the shield to terminal 1, the purple cable (red in case of type A or B cable) to terminal 2 and the blue (white in case of type A or B cable) to terminal 3. 16

IFC 070 ELECTRICAL CONNECTIONS 4 4.5 Terminal assignment of converter Figure 4-3: Removing side cap Figure 4-4: Terminal assignment 1 Status output 1 2 Status output 2 3 Not connected 4 Ground 5 Pulse output A 6 Pulse output B Electrical values Pulse output passive: f 500 Hz; I 10 ma; U: 5...24 VDC (P 100 mw) Status output passive: I 10 ma; U: 5...24 VDC (P 100 mw) 17

5 NOTES IFC 070 18

IFC 070 NOTES 5 19

K K K KROHNE product overview KROHNE - Subject to change without notice. Electromagnetic flowmeters Variable area flowmeters Ultrasonic flowmeters Mass flowmeters Vortex flowmeters Flow controllers Level meters Temperature meters Pressure meters Analysis products Measuring systems for the oil and gas industry Measuring systems for sea-going tankers Head Office KROHNE Messtechnik GmbH & Co. KG Ludwig-Krohne-Str. 5 D-47058 Duisburg Tel.:+49 (0)203 301 0 Fax:+49 (0)203 301 10389 info@krohne.de The current list of all KROHNE contacts and addresses can be found at: