In use test of a Washer-disinfector reprocessing of flexible endoscopes



Similar documents
ETD3 the new standard in endoscope reprocessing.

Expert Opinion. on the efficacy of. EndoDet and EndoDis + EndoAct. for cleaning and disinfection of. Olympus Gastroscope Type GIF-2T200.

Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!

Guidelines 85. Contents. 1. Scope of this guideline. 1. Scope of this guideline

ON TRACK Reprocessing Inservice / Competency for GI Endoscopes (EVIS, EXERA, EXERA II)

National Endoscopy Programme. Decontamination Standards for Flexible Endoscopes

Detergents in CSSD. Wim Renders

Christina Bradley Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK

Residual Organic Soils on Patient used Instruments

Reprocessing Flexible Endoscopes Avoiding Reprocessing Errors Critical for Infection Prevention and Control

Medical Device Reprocessing: How to Ban the Biofilm!

3M Microbiology. 3M Clean-Trace Hygiene Monitoring Systems. Effective. Hygiene. Monitoring

Back to Basics: Flexible Endoscope Processing LISA SPRUCE, DNP, RN, CNS-CP, CNOR, ACNS, ACNP, FAAN

Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes

Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes

THE BASICS OF DECONTAMINATION

Reprocessing. Critical & Semi-Critical Equipment

Infection Prevention + Control

Methods of Verification of Cleaning in CSSD. Dr. Winfried Michels Mechelen Oct.2012

Infection Control in the Dental Practice through Proper Sterilization

Microbiological Testing of the Sawyer Mini Filter. 16 December Summary

A new wave in instrument washing and disinfection Introducing the new HYDRIM G4 WD series

position statement INFECTION CONTROL INTRODUCTION POSITION STATEMENT PERFORMANCE EXPECTATIONS

BESTPRACTICE. HTM01-05 Q&A: You can now follow us on Twitter! We have launched our HTM01-05 questions and answers service on Twitter.

Environmental Monitoring of Clean Rooms

Environmental Monitoring

Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes

REPROCESSING OF FLEXIBLE CYSTOSCOPES

InfectIon PreventIon checklist for outpatient settings:

Cleaning Guidelines for Care Homes. Includes cleaning standards for the general environment and equipment

URGENT FIELD CORRECTION. REVISED ED 530XT OPERATION MANUALS: Cleaning, Disinfection and Storage, CODE 202B D Operation

Contents. Introduction 2 Components What is ATP? Measuring ATP with Bioluminescence Technology Additional Uses of SystemSURE Plus 6

BACTERIAL ENUMERATION

Requirements for transport sequence, cleaning and disinfection

Health Technical Memorandum 01-06: Decontamination of flexible endoscopes. Part C: Operational management

Arrangements for the Maintenance and Testing of Laboratory Autoclaves

Alberta Health. Standards for Cleaning, Disinfection and Sterilization of Reusable Medical Devices for Health Care Facilities and Settings

Bench Autoclave. Standard Operating Procedure. For Installation, Use and Maintenance

Updates on CFPP and Standards

EMERGENCY ACTION PLAN CONTAMINATED WATER SUPPLY (BIOLOGICAL)

[[ ;/LON/P&M/896/8/09 Date: Notice inviting quotation for Repairing heating system

BACTERIA COUNTS IN RAW MILK

Ozone in Cooling Towers : Hygiene for us

Bacteria Control Troubleshooting Guide for Automated Watering Systems

Chapter 7 Equipment and Utensil Cleaning and Sanitization

Precautionary Boil Water Notices: Frequently Asked Questions

A Guide to ATP Hygiene Monitoring

Company background and. portfolio overview

Cleaning, Disinfection, and Sterilization. Basics of Infection Prevention 2-Day Mini-Course 2013

Raw Milk Quality Tests Do They Predict Fluid Milk Shelf-life or Is it time for new tests?

Gastroenterology Nursing

Are You Always Ready for The Joint Commission or CMS Survey?

Current methods: late results (2 to 3 days) contamination spread in the distribution system water consumed

Body Art Facility Infection Prevention And Control Plan Guideline

CLEANING AND SANITIZING

Ethylene Oxide (EtO or EO) The Future Of Gas Sterilization

Clinic Washer-Disinfectors. Washer-Disinfector Line

Heat Transfer. Condensation. Revaporization

Creatine Kinase (CK) Enzymatic Assay Kit Manual Catalog #:

Cleaning and High Level Disinfection of Semi-Critical Medical Devices

Where do the germs live in your environment? April 2010 Shelly Padgett, BSN, RN Nurse Educator, 7N, Central Monitoring, and Cardiac Access Center

City of Houston Health Department Laboratory Water & Dairy Section 2250 Holcombe Blvd. Houston, TX DRINKING WATER INFORMATION & TESTING

Introduction. Introduction. Why do we need microbiological diagnostics of udder infections? Microbiological diagnostics How is it done?

Standard Operating Procedure

Cleaning, Disinfection and Sterilization

Laboratory Biosafty In Molecular Biology and its levels

ESGE ESGENA guideline: Cleaning and disinfection in gastrointestinal endoscopy Update 2008

Private Water Supplies Sampling Manual. A Field Guide

1.40 Prevention of Nosocomial Pneumonia

MagExtractor -Genome-

ENVIRONMENTAL MONITORING

Disc Diffusion Susceptibility Methods

Emergency Guideline for Food Facilities during Boil Water Advisory or Other Public Notices

ACO Cleaning principles for drainage

Biological Indicators and

UTILIZATION of PLASMA ACTIVATED WATER in Biotechnology, Pharmacology and Medicine. JSC TECHNOSYSTEM-ECO Moscow, Russia April, 2009

Best Practices for Loaner Surgical Instruments

INFECTION CONTROL PRECAUTIONS

IX. Decontamination and Spills

Plant Genomic DNA Extraction using CTAB

Environmental Monitoring

MTT Cell Proliferation Assay

Environmental Monitoring in Isolators and Clean Rooms ICR settle and contact plates ICRplus contact plates

Water should be removed as soon as possible once the safety of the structure has been verified.

Follicle Dermal Papilla Cell

Decontamination and Waste Management

UltraClean Soil DNA Isolation Kit

Biosafety Spill Response Guide

Transducer and System Care and Cleaning Koninklijke Philips N.V. All rights reserved. Published in USA.

ArtisanLink Staining System is an automated special stains slide

CHAPTER V: DISPOSAL OF WASTES CONTAMINATED WITH INFECTIOUS AGENTS

LAB 4. Cultivation of Bacteria INTRODUCTION

Domestic Chemical Water Treatment Manufacturers Association

Healthcare Water System Repair Following Disruption of Water Supply

absolute dental - cross-infection control

Welcome to Implementing Inquirybased Microbial Project. Veronica Ardi, PhD

Cal Water 1961 Petra Lane, Placentia, CA (800) CAL-WATER (800) FAX: (714)

- WATER DAMAGE RESTORATION -

Biological Sciences Initiative

COLLECTION OF DRINKING WATER SAMPLES AND TREATMENT OF WELLS

Transcription:

of a Washer-disinfector reprocessing of flexible endoscopes Olympus ETD 3 with PAA Clean Endoscope Copenhagen Hospitals Denmark

In-use test of Olympus ETD 3 Washer-disinfector with PAA Cleaner and disinfectant from Ecolab, Germany By Dr. T. Slotsbjerg, Hvidovre Hospital, Denmark 31 flexible endoscopes were tested before 416 endoscopies in less than 8 weeks. Introduction In a quality control program for cleaning and disinfection of flexible gastrointestinal endoscopes (FE) at hospitals in Copenhagen flush water from the water channel is used to evaluate the whole procedure for reprocessing FE. A continuous quality control is established with sampling one day a month before all endoscopies (before 5-6% of the endoscopies per year). The reprocessing of the biopsy channels of FE are controlled by audit. A report forwards to the endoscopy units every month and control charts 1 are worked out. Outliers from defect endoscopes and washer-disinfectors (WD s ) are excluded and the percentage of clean endoscopes is determined for every year. A database contains results from more than 14.000 control samples. The results and methods from the quality control are also used to evaluate new products for cleaning and disinfection of FE. Order from Olympus Medical Systems Europa GMBH Wendenstrasse 14-18 20097 Hamburg Germany Washer disinfector(s) Two Olympus ETD 3 with per acetic acid (PAA) Both ETD3 PAA machines did have the following main features: Automatic Leak Test. Pre-Cleaning stage with water (no heating). Cleaning stage with 0,6% EndoDet at 35 C for 3 minutes. Disinfection stage with 1,2% EndoDis (PAA) and 1,2% EndoAct at 35 C for 5 minutes. Rinse 1 with water (no heating). Final Rinse with Water (no heating). Optional selectable Drying function (15 minutes) at 60 C. ETD3 PAA 1/13

Each stage of the process runs with fresh water (no reuse of water) and all chemical products are single use. All incoming fresh water is treated with UV-light radiation and with small amount of PAA (10-15ppm). The ETD3-PAA is doing an automatic System Check every morning after first Switch-On. During the System Check all water feeding tubes and the integrated Water Softener are disinfected with 0.5% PAA (EndoDis) solution. The machines at Bispebjerg hospital are serial products. They are equipped with Automatic Endoscope Identification (EndoID) and Flow Control. Chemical products For manual cleaning: 3-enzyme, Olympus Denmark Cleaner in WD: EndoDet 0.6%, Ecolab Disinfectant in WD: EndoDis 1.2% + EndoAct 1.2%, Ecolab Endoscopy department The Endoscopy Unit Surgical Gastrointestinal Department K Bispebjerg University Hospital Copenhagen DK Flexible endoscopes included Gastroscopes, sigmoidenoscopes and colonoscopes (Olympus) Test laboratory Clinical Microbiology Department, Hvidovre Hospital, has carried out the laboratory work. Definitions A clean endoscope: CFU < 5 per 0.2 ml flush water corresponding to 0-20 CFU per ml (range for acceptable drink water). A critical endoscope: CFU are between 5 and 50 CFU per 0.2 ml flush water. Critical FE s with 11-50 CFU per 0.2 ml flush water is correlated to fixation of organic materials to the inner surface of the water channel (Data from the quality control program) A high-risk endoscope: CFU >50 per 0.2 ml flush water. High-risk FE with more than 50 CFU per o.2 ml flush water is correlated to proliferation of microorganisms during storage. ETD3 PAA 2/13

A stable procedure At least 95% of the endoscopes should be clean No occurrence of high-risk endoscopes. An unstable procedure Less than 95% of the endoscopes are clean and/or occurrence of high-risk endoscopes. (The reprocessing is out of control in g-control chart with a centre line = 20 clean FE between a not clean FE s ) 1, 2 The mean of the previous year 98.6 % of the endoscopes were clean in 2006. Acceptable lower limit of clean endoscopes The acceptable lower limit of numbers of clean endoscope is 98.6 % - (3 X SD) (Joint Commission International). SD = Standard deviation Questions before the in use test 1. Is the reprocessing of FE a stable procedure? 2. Is the percentage of clean FE > 98.6% - 3 X SD? 3. Are significant residuals of organic material demonstrable in FE water channels after reprocessing? 4. Prevents reprocessing in ETD 3 proliferation of micro organisms (and development of biofilm) in FE channels during storage? 4 5. Are significant residuals of organic material demonstrable in biopsy port valves after reprocessing? 3 6. Are the process temperatures and times constantly? Condition for the investigation The endoscopy unit follows the Danish recommendations for cleaning and disinfection of flexible endoscopes. o When the endoscope is still connected to the power source we flush the channels of with water. o Manual cleaning in the reprocessing area of all channels and the outer of the endoscope in a detergent solution with brushing. o Before storage we flush the channels with 70% alcohol, and the endoscopes can be used within 3 days. Duodenoscopes are decontaminated and flushed with alcohol before use. Sampling from the endoscopes Samplings are carried out immediately before an endoscopy. The endoscopes are connected with a sterile water bottle with sterile water. The water channels are sampled by depressing the air/water feed button and irrigating the channel with water from the water bottle. Approximately 5 ml is collected into a sterile test tube at the distal end. These samples contain material from the water bottle and the water and the distal jointed air/water channel. At the end of the day control samples are obtained from the water bottles. The test tubes are stored in a cold store and forward to the Clinical Microbiologic Department the same day. ETD3 PAA 3/13

Laboratory procedure In the laboratory ten droplets of 20 µl are put down a 5% blood agar the same day. After incubation for 48 hours at 35 C in CO 2 the number of CFU was counted. Identification of bacteria and fungi Samples with <5 CFU per 0.2 ml are identified as species. Other samples are identified after the routine of the laboratory. Staphylococci are identified as coagulase positive (S. aureus) and coagulase negative (CNS) Isolates from FE s are split up in environmental/staff related or patient related microorganisms. Sample size 400 420 samples from endoscopes are obtained immediately before an endoscopy. Excluded samples Samples from endoscopes are excluded if A contaminated endoscope has a defect demonstrable on service. A contaminated FE is decontaminated in a WD with a defect demonstrable on service. The endoscopes are contaminated during alcohol flushing of FE channels (Bacillus spp.) Samples with > 4 CFU per 0.2 ml related to inadequate compliance with the reprocessing protocol. Time/temperature curves Time/temperature curves of the reprocessing program in ETD 3 are recorded before and after the in use test on an EBI-125A temperature logger (± 0.1 C) from ebro- Germany. Cleaning tests ATP-test Residual Adenosine Tri-Phosphate (ATP), which is found in large quantities in human blood and other tissue fluids, are used as cleanness indicator for FE water channels and biopsy port valves. An UNI-LITE Xcel portable luminometer and total ATP Aqua-Trace and Clean-Trace test kits are used (Biotrace International). The relation between Relative Light Units (RLU) values from UNI-LITE Xcel and ATP in fento-mol (10-15 mol) is previous detected. Log RLU = log ATP +1, log RLU = log ATP-unit, 3 log ATP-units corresponds to 10-3 μl blood. Water channels ATP measures (Aqua-Trace test kits) on more than 25 random samples from the water channels of FE. Biopsy port valves 3 ATP measures (Clean-Trace swab test kit) on 25 random biopsy port valves after manual cleaning and decontamination in ETD 3. ETD3 PAA 4/13

Data analyses The number of clean and not clean endoscopes follows a binominal distribution. Outliners from defects endoscopes are excluded and the percent of clean endoscopes are determined. The acceptable lower limit of numbers of clean endoscope is 98.6% - (3 X SD) (Joint Commission International). If an endoscope is not clean in more than one sampling the critical numbers of not clean results will be determinate. If more than 8 endoscopes are not clean a Number-Between g-type Statistical Quality Control (1, 2) Charts will be made-up ATP residual values are imported in cleaning control chats Water channels: Centre line (CL) and Upper Control Limit (UCL) are calculated. The total cleaning process is reported as in or out of statistical control. In samples from the water channels in new FE s CL and UCL are 1,4 and 1,9 log ATP-units respectively. Biopsy port valves: CL and UCL are calculated. The total cleaning process is reported as in or out of statistical control. Results 416 samples were obtained and lower acceptable limit for clean FEs is (98.6% - 1.7%) = 96.9% Table 1 and figure 1 show that 413 of 416 endoscopes (99.3%) were clean before an endoscopy (0-4 CFU per 0.2 ml). Table 1 shows furthermore that 3 of 416 (0.2%) endoscopes were critical FE s with 5, 7 and 8 CFU per 0.2 ml. water respectively. No endoscopes were excluded. The isolates were coagulase-negative staphylococci (CNS) and gram-negative coliform bacteria. As standard of reference table 1 shows results of quality control samples from FE s at the Endoscopy department after reprocessing in Olympus ETD2+ WD s before the in use test. After reprocessing in ETD3 PAA and ETD 2+ the percent of critical FE with 10-50 CFU per 0.2 ml flush water was 0% and 1.4% respectively and the difference is significant (p=0.017). Table 2 and figure 2 show that 99.6% and 98.2% of the FE was clean after reprocessing the same day and 1-3 days before respectively. The difference is not significant. Figure 3 shows a control chart with ATP determinations in 48 of samples from the water channel of the endoscopes. The control chart shows a total cleaning process in statically control. The centreline of the control cart (CL = 1.4 log ATP-units) indicates that PAA don t fix organic materials at the inner surfaces of the water channels, but a high upper control limit (2.9 log ATP-units) indicates that the manual cleaning can be improved. Figure 4 shows a control chart with ATP residuals in 26 biopsy port valves after manual cleaning and reprocessing in ETD3 PAA. The reprocessing of the valves is out of statistical control (four outliners). ATP-residuals in the biopsy port valves outliner corresponds to < 10-4 μl blood. ETD3 PAA 5/13

Table 3 and figure 5 and 6 show that process temperatures and times during the in use test were constantly in the ETD3 PAA WD s. Conclusion Chemical reprocessing at 35 C of flexible endoscopes using ETD3 PAA is a stable procedure with 99.3% clean endoscopes (0-4 CFU per 0.2 ml flush water). Three samples (0.7%) presented 5 8 CFU per 0.2 ml and were all detected within 7 days. 99.3% clean FE s are above the lover acceptable limit for clean FE s at 96.9%. The reprocessing of 416 FE including the ETD3 PAA process, using cold-water final rinse during 8 weeks has prevented significant proliferation of microorganisms with biofilm formation and clusters of High risk FE during storage (0-3 days). Therefore the cold-water final rinse doesn't contaminate the endoscopes significantly. Removal of organic material from the water channel of FE s was a statistical stable process, but the manual cleaning could be improved with use of an alternative detergent. The cleaning process of biopsy port valves was out of statistical control, but the ATP residuals in the biopsy port valves outliners were not critically. An alternative detergent for manual cleaning is under consideration 4. The process temperatures and times were constantly during the in-use test in the two ETD3 PAA WD s. ETD3 PAA can consequently be recommended for chemical reprocessing of flexible endoscopes on the following conditions Detergents used for manual cleaning should be approved of Olympus FE channels should be flushed with 70% alcohol before storage in accordance with the Danish Guideline for Cleaning and Disinfecting of Flexible Endoscopes (CDC category 1A 5 ). Flexible endoscopes (except duodenoscopes, bronchoscopes and cystoscopes) can be used within 72 ours after a reprocessing. Torsten Slotsbjerg MD, Consultant Clean Endoscope Hvidovre Hospital Denmark ETD3 PAA 6/13

Table 1: 416 samples from the water channel of 31 flexible gastrointestnal endoscopes obtaind before an endoscopy after manual cleaning with 3-enzyme and reprocessing in ETD3 PAA WD s. To comparation are results of 433 quality control samples from the endoscopy department (manual cleaning with 3-enzyme and reprocessing in ETD2+ WD s ) Table 1 Clean endoscopes 0-4 CFU per 0.2 ml Critical endoscopes 5-49 CFU per 0.2 ml. 5-10 CFU (low critical) 11-49 CFU High-risk endoscopes >=50 CFU per 0.2 ml Number of FE ETD3 PAA 413 (99.3%) 3 (0.7%) 0 * 0 ETD 2+ (0.24% glutaraldehyde) 424 (97.9%) 3 (0.7%) 6 (1.4%) * 0 *) p=0.017 400 ETD3 PAA Washer-disinfectors Reprocessing of endoscopes 416 samples from the water channel of flexible endoscopes before an endoscopy 350 Number of samples 300 250 200 150 100 50 0 0 1 2 3 4 5 6 7 8 9 10 >10 CFU per 0.2 ml water Figure 1: Results of 416 samples from the water channel of the 31 flexible GI endoscopes ETD3 PAA 7/13

obtained before an endoscopy after reprocessing of the endoscopes in ETD3 PAA Table 2: Distribution of CFU per 0.2 ml flush water and isolates in samples from Flexible endoscopes reprocessed the same day and after 1-3 days respectively ETD3 PAA Sampling day after reprocessing Same day 1 2 3 Total Number (%) of samples 271 (65.1%) 110 (26.5%) 10 (2.4%) 25 (6.0%) 416 Number of clean FE 270 (99.6%) 108 (98.2%) 10 25 413 (99.3%) Number of low critical FE 1 (0.4%) 2 (1.8%) 3 (0.7%) Bacteria Coliforms + CNS ++ Endoscopes CF Q160 gastroscopes Two CF Q160 gastroscopes Washerdisinfector WD 1 WD1 + WD2 + Patient related, ++ environmental/staff related Frequency 14% 12% 10% 8% 6% 4% 2% 0% ETD3 PAA Washer-disinfectors Samples with growth 1 2 3 4 5 6 7 8 9 10 >10 CFU per 0.2 ml water Reprossessing same day 1-3 days after reprocessing Figure 2: Distribution of CFU per 0.2 ml flush water in samples with growth reprocessed the same day and after 1-3 days respectively ETD3 PAA 8/13

Residuals of organic materials in the water channel of flexible endoscopes after reprocessing 3,5 3,0 Log ATP units 2,5 2,0 1,5 1,0 0,5 0,0 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 45 47 Sample number Sample Centre line (CL) Upper control limit (UCL) Figure 3: Residuals of organic material in the water channel of flexible endoscope after manual cleaning and reprocessing in an ETD3 PAA Washer-disinfector. 2 Biopsy Port Valves Reciduals of organic materials Log ATP units 1,5 1 0,5 1 3 5 7 9 11 13 15 17 19 21 23 25 Sample number Sample CL UCL LCL Outliners Figure 4: Residuals of organic material in biopsy port valves of flexible endoscope after manual cleaning and reprocessing in an ETD3 PAA Washer-disinfector (LCL = lower control limit) ETD3 PAA 9/13

Table 3: Results of time/temperature measuring of during reprocessing of flexible endoscopes in the ETD3 PAA washer-disinfectors. First part of the in-use test Second part of the in-use test WD 1 WD 2 WD 1 WD 2 Time Temp. Time Temp. Time Temp. Time Temp. Pre-rinsing 2:32 29.2 2:29 29.5 2:56 30.0 2:54 30.5 Washing 7:30 35.9 7:15 36.2 7:34 36.2 7:19 36.2 Part of the ETD3 PAA reprocessing Disinfection 8:47 36.0 8:52 36.5 8:18 37.1 8:31 37.0 Rinsing 1 3:18 34.7 3:00 35.8 3:29 34.4 3:26 35.5 Rinsing 2 4:05 (* 34.3 6:04 34.3 5:33 33.6 5:56 35.3 * ) Rinsing2 time was 6:00 minutes in the following reprocessing. ETD3 PAA 10/13

WD1 WD2 Figure 5: Temperature/time curves from ETD3 PAA Washer-disinfectors from first part of the inuse test. ETD3 PAA 11/13

WD1 WD 2 Figure 6: Temperature/time curves from ETD3 PAA Washer-disinfectors from second part of the in-use test. ETD3 PAA 12/13

References 1. JC Benneyan: Statistical Quality Control Methods in Infection Control and Hospital Epidemiology, Part II: Chart use, Statistical Properties, And Research Issues; Infect Control Hosp Epidemiol 1998; 19: 265-283. 2. www.clean-endoscope.com : Posters from SHEA 2004 and 2005, ECCMID 2004 3. Parente, David M., BS., BA., MBA; NAMSA Advisory Services: Consulting and Testing Services Group Atlanta, Georgia: Could biopsy port valves be a source for potential flexible endoscope contamination?; NAMSA 2006. 4. K. Marion et al.: Using an efficient biofilm detaching agent: an essential step for the improvement of endoscope reprocessing protocols: J Hosp Infect 2006 64, 136-142. 5. Multi-society guideline for reprocessing flexible gastrointestinal endoscopes; Am J Infect Control 2003; 31: 309-15 ETD3 PAA 13/13