Reprocessing. Critical & Semi-Critical Equipment
|
|
|
- Thomasine Morgan
- 9 years ago
- Views:
Transcription
1 Reprocessing (Cleaning, Disinfection & Sterilization) Critical & Semi-Critical Equipment - A Physician Toolkit - Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit
2 Table of Contents Preface...1 Basic Physician Office Supplies for Reprocessing and Semi-Critical Medical Equipment...2 Sterilization Monitoring...3 Policies, Procedures & Guidelines...4 Reprocessing Medical Equipment Policy, Procedures & Guidelines...5 Policy Template for Reprocessing of Medical Equipment...9 Procedure Template for Steam Sterilization...10 Procedure Template for High Level Disinfection...11 Control Sheet Template for Cleaning and Disinfection of Flexible Endoscopes Manually...13 Control Sheet Template for Cleaning and Disinfection of Flexible Endoscopes Using Automated Endoscope Reprocessor...14 Sterilization Log Template...15 Common Deficiencies...16 Cleaning, Disinfection & Sterilization Products...17 Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit
3 Preface This document, A Physician Toolkit for Reprocessing Critical and Semi-Critical Medical Equipment, has been developed to assist office physicians and their staff to meet best practice standards on reprocessing medical equipment. Additional references and standards that physicians and their staff should be aware of and compliant with are: Provincial Infectious Diseases Committee (PIDAC) Best Practices for Cleaning, Disinfection and Sterilization, 2006; Canadian Committee on Antibiotic Resistance (CCAR) Infection Prevention & Control Best Practices, 2007; and A Physician s Guide to Occupational Health and Safety Responsibilities. Links to all of the above noted references are available on the College s website at Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 1
4 Basic Physician Office Supplies for Reprocessing Critical & Semi-Critical Medical Equipment Personal Protective Equipment (PPE) Face Protection: mask, protective eye ware (safety glasses or goggles) and/or full face shield). Gloves: gloves must be of sufficient weight to be highly tear resistant (regular exam gloves are not appropriate protection for reprocessing). Gowns: gowns must be of a quality to provide an adequate barrier to moisture (isolation gowns are not appropriate). Cleaning Brushes (reusable or disposable) or cloths for physical cleaning. Detergent or enzymatic cleaning solutions (e.g. Medzyme). High Level Disinfection High level disinfectant solution (e.g. Glutaraldehyde). Products must have a Drug Identification Number (DIN) from Health Canada. High level disinfectant solution chemical test strips. Log and documentation record forms. Sterilization (manufacturers may have starter kits which include all of the products listed here) External chemical indicator/autoclave tape (maybe built-in with peel pouches) placed on packages with each load. Internal chemical indicator: placed inside each package for sterilization. Biological indicator (spore-laden strips or vials) (usually Geobacillus stearothermophilus for steam sterilizers). Used once each day the sterilizer is in use, or with every load if sterilizing implantable devices. Logs books/forms for documentation of sterilization parameters and autoclave maintenance. Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 2
5 Sterilization Monitoring Sterilization of medical equipment is a two-step process involving: (i) (ii) Decontamination (aka cleaning) to remove >80% of microbes; and Sterilization to kill the remaining microbes. Steam autoclaves are the most efficient and non-toxic method of sterilization in a physician office setting. Because of the difficulty in proving the sterility of a medical instrument, the effectiveness of the sterilization process must be monitored. Three complementary types of monitoring are required: Mechanical/Digital indicators to monitor the autoclave s physical parameters (time, temperature and pressure) for each cycle/load; Chemical indicators - external indicators on the outside of each wrapped package and internal chemical indicators inside each package, which change color when exposed to the right conditions for sterilization; and Biological Indicators that confirm the actual annihilation of microbial spores. All three types of monitors are important to confirming that the conditions necessary to achieve sterility have been met. Among these, biological indicators may be the most unfamiliar to physicians. Options for biological monitoring include: Purchasing a self-contained testing kit with incubator for on-site monitoring; or Purchasing only the spore strips, and then transporting them to a microbiology laboratory that offers environmental testing. (Unfortunately, few laboratories in the province offer this service.) Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 3
6 Policies, Procedures & Guidelines The following policies, procedures, guidelines and templates are not official documents of the College of Physicians & Surgeons of Alberta. They are provided for information only. We hope these documents will assist you in creating your own specific documents for your office setting. Comments and questions can be forwarded to Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 4
7 Reprocessing Medical Equipment Policy, Procedures & Guidelines The following information is provided to assist you in developing policies and procedures relating to reprocessing of medical equipment in the office setting. Templates for developing office/clinic based policy and procedures are included. A. General Requirements: 1. Medical equipment that is to be reprocessed must be labeled reusable by the manufacturer and must be accompanied by the manufacturer s written instructions for reprocessing (cleaning, disinfection and sterilization). 2. Critical or semi-critical medical equipment labeled as single use or disposable shall not be reprocessed. 3. In order to determine the level of reprocessing that is required for a piece of medical equipment, the following risk classification is used: Critical equipment - contacts sterile tissue or the vascular system and requires sterilization. (surgical instruments, biopsy forceps, suture scissors, devices entering sterile body cavities, etc.) Semi-critical equipment - comes in contact with intact mucous membranes or non-intact skin and requires a minimum of high level disinfection (vaginal specula, nasal specula, vaginal/anal ultrasound probes, gastrointestinal/nasopharyngeal endoscopes, laryngoscopes, etc.) 4. Medical equipment reprocessing must take place in a designated area that is separate from patient care areas, or is not to occur when patients are present. 5. Personnel responsible for reprocessing in the office setting must have documented training. 6. Office-specific procedures are developed for medical equipment reprocessing that are written, current, and incorporate existing recognized standards of practice ( PIDAC, CCAR and AH&W). Manufacturer s instructions must be accessible and incorporated into office specific procedures. 7. Personnel responsible for reprocessing must wear the appropriate personal protective equipment (PPE) (e.g. gowns, gloves and facial protection) while reprocessing medical devices. 8. At least one dedicated sink or hand hygiene station must be present in the reprocessing area. Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 5
8 9. Medical equipment is to be stored in clean areas that are protected from contamination, vermin, excessive handling and crushing. They are NEVER stored beneath the sink, in the soiled equipment cleaning area or in the immediate area where examinations/procedures are performed (e.g. beneath the end of the table where pelvic exams are done.) B. Cleaning: 1. All medical equipment must be cleaned first, using water and detergents or enzymatic solutions that are appropriate for use on instruments, prior to any subsequent disinfection or sterilization. 2. Detergents or enzymatic solutions are used according to manufacturer s instructions that are written on the label. 3. Detergents or enzymatic cleaning solutions are discarded following each use. 4. There are written office-specific procedures for cleaning that include protocols for containment of contaminated equipment at the point of use, transport to disassembly, sorting, soaking, physical removal of soil, rinsing, drying, inspection and wrapping (if necessary). 5. Cleaning may be done manually (using detergent or enzymatic solution, water and friction) or mechanically in an automated washer decontaminator/disinfector. 6. Automated cleaning equipment, if used, must be installed and operated according to the manufacturer s instructions, and a preventive maintenance program for the equipment must be established and documented. 7. If ultrasonic cleaning equipment is used: Pre-cleaning of devices is necessary to remove gross matter; Rinsing of devices with clean, fresh tap water must follow the ultrasonic cycle; Efficacy of the equipment must be tested using a method recommended by the manufacturer; Solution must be changed at least daily and whenever it appears soiled; and The unit must be covered when in operation. 8. Devices must be dried following cleaning. Devices must be dry before storage, immersion in disinfectant or wrapping and sterilization. 9. Devices are visually inspected to ensure that no visible soil remains. Devices that do not pass inspection are returned for re-cleaning. 10. Devices with lumens or channels must be cleaned using an appropriately sized brush or stylet as recommended by the manufacturer. Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 6
9 11. Cleaning accessories (e.g. brushes, sponges) are disposable, or must be thoroughly cleaned and high level disinfected or sterilized between uses. 12. Sinks used for cleaning medical devices are cleaned at least once daily. C. High Level Disinfection 1. Semi-critical medical devices require a minimum of high level disinfection. 2. The disinfectant has a Drug Identification Number (DIN) from Health Canada. 3. The disinfectant label indicates that the product is a chemo sterilant or a high level disinfectant (HLD). 4. The HLD is prepared and used according to the manufacturer s instructions specified on the label, MSDS or accompanying product literature. 5. An appropriate chemical test strip specified by the disinfectant manufacturer is purchased and used to test disinfectant minimum effective concentration (MEC) at least daily. 6. Results of all disinfectant MEC testing are recoded in a log. 7. When opened, each container of chemical test strips is checked using a quality control procedure recommended by the manufacturer to verify accuracy. 8. Results of all quality control testing of test strips are recorded in a log. 9. Containers of test strips are dated when opened and not used beyond the shelf life indicated by the manufacturer. 10. There is documentation that the correct HLD solution is used when solution is changed. 11. Devices are completely immersed in HLD for the recommended time. D. Sterilization 1. Critical equipment is sterilized using an approved process. (Unacceptable sterilization methods include boiling, glass bead sterilizers, microwaves and ultraviolet light.) 2. The sterilizers are installed and have a documented preventive maintenance program according to the manufacturer s recommendations. 3. The type of sterilizer used has cycles capable of sterilizing the instruments that are used in the office setting. Check with sterilizer manufacturer s written claims. 4. There are written procedures for sterilization of all medical equipment. Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 7
10 5. Chemical indicators are placed on the outside of each load and inside each package to be sterilized. 6. Each sterilization cycle is monitored with physical parameters (time, temperature and pressure) that are recorded on a print out or otherwise logged. 7. Sterilizers are monitored with an appropriate biological indicator each day the sterilizer is used. 8. A log of biological indicator monitoring is maintained. 9. There is a written recall procedure that is followed in the event of a failed biological indicator. 10. Each package that is sterilized is labeled with a date and load number. 11. A log is kept of each load and items in the load. 12. Instruments/devices are sterilized in the open position and according to the manufacturer s instructions if disassembly is required for sterilization. 13. All reprocessed instrument sets and devices are stored in a manner to keep them clean and dry. Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 8
11 Policy Template for Reprocessing of Medical Equipment Name/Address of Office/Clinic: Date Issued: Date Reviewed: (Suggest annually): Signature of Owner/Operator of Clinic/Office: Policy - Reprocessing of Medical Equipment/Instruments: Purpose: Reprocessing is the cleaning, disinfection and sterilization of reusable medical equipment between each patient use. The safe use and reprocessing of all reusable medical equipment as well as proper care and maintenance of reprocessing equipment is required to reprocess safely. The following shall be adhered to when reprocessing medical equipment: 1. Items labeled for single use only shall not be reused. 2. The following risk classification shall be used to determine the minimum level of reprocessing to be performed on items that are labeled reusable : Critical: Item that enters sterile body site or vascular system and requires cleaning followed by sterilization. Semi-Critical: Item that comes in contact with intact mucous membranes or non-intact skin and requires cleaning followed by a minimum of high level disinfection. Non-Critical: Item that comes in contact with intact skin that requires cleaning followed by low level disinfection. 3. The manufacturer s instructions for cleaning, disinfection and sterilization of medical equipment shall be utilized and followed in office procedures. 4. Written office/clinic procedures for reprocessing shall be followed. 5. The manufacturer s instructions for installation and preventive maintenance of equipment used in reprocessing medical devices and instruments shall be followed and documented. 6. All processes related to cleaning, disinfection and sterilization of medical equipment shall be monitored and documented. 7. Staff responsible for reprocessing of medical equipment shall be knowledgeable and aware of current standards of practice ( PIDAC, CCAR, AH&W). All education and competencies shall be documented. Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 9
12 Procedure Template for Steam Sterilization Name of Equipment/Instrument: Performed by: Personal Protective Equipment: Gloves Gown Face Shield/Mask Safety glasses/goggles Disassembly instructions, if applicable: (Include pictures or diagrams for complex or unusual equipment.) After each use: 1. Prepare (detergent/enzymatic) solution by adding ml to ml water. (Follow instructions on label for correct concentration.) 2. Immediately immerse instrument into freshly prepared solution of (indicate name of detergent/enzymatic product used) for minutes to prevent drying of soil. (Follow detergent/enzymatic manufacturer s instructions for soak time.) 3. Clean with detergent/enzymatic solution. Requires: Brush Cloth Ultrasonic 4. Rinse with tap water to remove detergent/enzymatic and soil residue. 5. Dry with lint free cloth. 6. Reassemble, as necessary, according to manufacturer s instructions. 7. Wrap in (indicate type of wrapping material: peel pouch) 8. If not part of the wrap, apply internal and external chemical indicators. 9. Label with date of sterilization and load number. 10. Place in sterilizer. 11. Select cycle (appropriate cycle e.g., wrapped, porous, nonporous). 12. Check printout/display to ensure all cycle parameters have been met and sign on print out or in log book. 13. Remove when cycle is finished and packs are completely dry. 14. Storage location:. References (i) (ii) (name of manufacturer) instructions for reprocessing PIDAC or CCAR Date Issued: Date Reviewed: Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 10
13 Procedure Template for High Level Disinfection Name of Instrument: Performed by: Personal Protective Equipment: Gloves Face Shield/Mask Gown Safety glasses/goggles Disassembly instructions, if applicable: (Include pictures or diagrams for complex or unusual equipment.) Disinfectant used: Check the following: Label Concentration After each use of equipment: Date product activated/opened Results of minimum effective concentration (MEC) testing of disinfectant (done daily at a minimum using appropriate test strip) 1. Prepare (detergent/enzymatic) solution by adding ml to ml water. (Follow instructions on label for correct concentration.) 2. Immediately immerse instrument into freshly prepared solution of (indicate name of detergent/enzymatic product used) for minutes to prevent drying of soil. (Follow detergent/enzymatic manufacturer s instructions for soak time.) 3. Clean with detergent/enzymatic solution. Requires: Brush Cloth Ultrasonic 4. Rinse with clean tap water to remove detergent and soil residue. 5. Dry with lint free cloth to avoid diluting the disinfectant chemical. 6. Remove cover from container of high level disinfectant. 7. Completely immerse device or instrument in high-level disinfectant. 8. Cover container of high level disinfectant and begin timing (set timer). 9. Soak for minutes at room temperature (according to manufacturer s instructions). 10. Rinse with clean tap water using three separate rinses to ensure removal of residual disinfectant. 11. If instrument has channels or lumens, purge with alcohol to facilitate drying. 12. Dry. 13. Storage location: Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 11
14 References (i) (ii) (iii) (name of manufacturer) instructions for reprocessing (name of device/instrument). Date and source of instructions. Instructions for preparation and use of (name of high level disinfectant) PIDAC/ CCAR Date Issued: Date Reviewed: Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 12
15 Control Sheet Template for Cleaning and Disinfection of Flexible Endoscopes Manually* Date: Instrument code number or identifier: Endoscope processed (time): I.D. Number of soaking container if more than one container is used for disinfection: Comments: Signature of person reprocessing scope: Ensure water-resistant cap is on Perform leak test Wash exterior of scope Brush valves and piston Brush all channels Suction biopsy port with cleaning adapter Aspirate enzymatic solution through biopsy channel Connect appropriate adaptors for instrument Inject enzymatic product in all channels Soak in enzymatic solution for a minimum of minutes (see enzymatic instructions) Rinse scope and flush all channels with fresh tap water Dry exterior of scope and flush remaining water from channels Completely immerse scope, valves and cleaning brushes in HLD Inject HLD into all channels of the scope using appropriate clean channel adaptors Ensure all channels are filled with HLD and no air pockets remain Ensure temperature of room is not less than degrees (see HLD instructions Set timer for (see HLD instructions) When disinfection time is complete, purge all channels with air to remove HLD Rinse scope and flush channels with three separate rinses of clean tap water (with sterile or submicron filtered water) Dry channels with compressed medical grade psi. (check manufacturer pressure limit) Flush all channels with 70% isopropyl alcohol Repeat dry channels with air Hang/lay scope with all valves and water-resistant cap removed * Adapted from CSA Z Decontamination of reusable medical devices. Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 13
16 Control Sheet Template for Disinfection of Flexible Endoscopes Using Automated Endoscope Reprocessor (AER)* Date: Instrument code number or identifier: Endoscope processed (time): AER used: Comments: Signature of person reprocessing scope: Ensure water-resistant cap is on Perform leak test Wash exterior of scope Brush valves and piston Brush all channels Suction biopsy port with cleaning adapter Aspirate enzymatic solution through biopsy channel Connect appropriate adaptors for instrument Inject enzymatic product in all channels Flush additional ports Soak in enzymatic solution for a minimum of minutes (see enzymatic instructions) Rinse scope and flush all channels with fresh tap water Place in washer using appropriate connectors for final disinfection Place valves, separate from scope, into washer Place cleaning brush(es) into washer Dry with compressed medical grade air at psi (check manufacturer pressure limit) Flush all channels with 70% isopropyl alcohol Repeat dry with compressed medical grade air Hang scope with all valves and water-resistant cap removed Clip initial print out from AER to control sheet * Adapted from CSA Z Decontamination of reusable medical devices. Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 14
17 Sterilization Log Template Sterilization Parameter and Cycle Documentation Autoclave # Date and Time Load Number Contents of the load Temperature achieved ( C or F) Time length of cycle (minutes Internal Chemical Indicator (Pass or Fail) Daily Biological Indicator (Pass or Fail) Daily Biological Indicator Control (Pass or Fail) Internal Chemical indicator: Type in use: Lot number: Expiry: Biological indicator: Type in use: Lot number: Expiry: Reprocessing staff name Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 15
18 Common Deficiencies Identified from Recent On-Site Audits 1. Biological Monitoring not being performed. 2. That all required Personal Protective Equipment is either not available or not worn by staff when reprocessing. 3. That the manufacturer s instructions for reprocessing equipment are not available. 4. That sterilization monitoring parameters (digital, mechanical chemical and biological) are not documented. 5. That high level disinfectant chemical test strip results are not documented. 6. That quality procedures on new containers of chemical test strips, and lots of biological indicators, are not performed and/or documented. 7. That reprocessing is occurring in patient care areas with patients present. 8. That re-useable cleaning accessories are not cleaned and minimally high level disinfected or sterilized between uses. 9. That documented policies and procedures are not developed for all steps of reprocessing and training procedures for reprocessing staff. 10. Inappropriate wrapping materials and tape used for packaging equipment for sterilization. Use self sealing pouches. 11. Internal, chemical indicators not being placed in each package for sterilization. Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 16
19 Cleaning, Disinfection & Sterilization Products The following products are examples of appropriate cleaning, disinfection and sterilization products. These products are available and can be ordered by contacting a medical supply vendor. Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 17
20 Instrument Detergents and Enzymatic Cleaners NPH Klenz Neutral Detergent Enzymatic Cleaners - foam or sponge Medzyme Enzymatic Cleaner Instrument Lubricants Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 18
21 Instrument Cleaning Accessories (Must be disposable, or cleaned and minimally high level disinfected or sterilized between uses.) Glutaraldehyde High Level Disinfectant & Test Strips Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 19
22 Cidex OPA (orthophtalaldehyde) and test strips Biological Indicators Self-contained BI Process Challenge Device PCD or Test Pack containing BI Log for BI monitoring Incubator Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 20
23 Chemical Indicators for Sterilization Steam Peracetic Acid (Steris System 1 Steam/Ethylene oxide (EO) Bowie-Dick Air Removal Test Packs Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 21
24 Post HLD Rinsing (Sterile or sub-micron filtered water is recommended) Sterile Water for Irrigation - 500ml 1000ml, 3000ml volumes Pall-Aquasafe Faucet Water Filter Read labels on all cleaning/ disinfection products. Ensure directions are followed. (Sample enzymatic cleaner label) Reprocessing Critical & Semi-Critical Equipment - A Physician Tool Kit - 22
Infection Prevention + Control
Physiotherapy Alberta regulates and leads the practice of physiotherapy in Alberta. Contact us for more information on this or other position statements. 780.438.0338 1.800.291.2782 [email protected]
position statement INFECTION CONTROL INTRODUCTION POSITION STATEMENT PERFORMANCE EXPECTATIONS
INFECTION CONTROL INTRODUCTION Appropriate infection control is an essential element of clinical practice management based on its importance to the health and safety of patients, practitioners and the
ON TRACK Reprocessing Inservice / Competency for GI Endoscopes (EVIS, EXERA, EXERA II)
THIS CHECKLIST IS DESIGNED FOR USE SOLELY AS A CUSTOMER EDUCATIONAL TOOL, AND IS NOT INTENDED TO REPLACE OR ANY WAY MODIFY THE OLYMPUS INSTRUCTION MANUAL/REPROCESSING MANUAL. BE SURE TO FOLLOW THE DETAILED
InfectIon PreventIon checklist for outpatient settings:
InfectIon PreventIon checklist for outpatient settings: Minimum Expectations for Safe Care National Center for Emerging and Zoonotic Infectious Diseases Division of Healthcare Quality Promotion CS224818
University of Kentucky / UK HealthCare Policy and Procedure. Policy # A03-030
University of Kentucky / UK HealthCare Policy and Procedure Policy # A03-030 Title/Description: High-Level Disinfection with Cidex OPA Purpose: To provide instructions for high level disinfection with
Cleaning and High Level Disinfection of Semi-Critical Medical Devices
I. Purpose: A. To provide guidance for cleaning, processing and high level disinfection (HLD) of semi-critical devices using cylinders, bins/pans, or automated endoscopic reprocessors. This process is
REPROCESSING OF FLEXIBLE CYSTOSCOPES
ALERT Joint AUA/SUNA White Paper on the REPROCESSING OF FLEXIBLE CYSTOSCOPES (2/8/14): The STERIS System 1E (SS1E) liquid chemical sterilant processing system has become available for reuseable processing
Alberta Health. Standards for Cleaning, Disinfection and Sterilization of Reusable Medical Devices for Health Care Facilities and Settings
Alberta Health Standards for Cleaning, Disinfection and Sterilization of Reusable Medical Devices for Health Care Facilities and Settings June 2012 Medical Devices for Health Care Facilities and Settings
National Endoscopy Programme. Decontamination Standards for Flexible Endoscopes
National Endoscopy Programme Decontamination Standards for Flexible Endoscopes Updated March 2009 Introduction Providing an effective endoscope decontamination service within a safe environment is an essential
Cidex OPA Solution Training Checklist and Competency for Semi-Critical Medical Devices
Name Method of Validation: Direct observation Today s Date: Unit/Department Please have user do the following: 1. Locate Cidex OPA Safety Data Sheet. 2. Verbalize which semi-critical devices are cleaned/disinfected
1/22/2012. When is it a good time to learn about sterile processing in your facility? Continuing Education Contact Hours. Learning Objectives
Continuing Education Contact Hours Participants must complete the entire presentation/seminar to achieve successful completion and receive contact hour credit. Partial credit will not be given. All of
Brock University Facilities Management Operating Procedures
Subject: Bodily Fluid Clean-Up Number: FMOP 2-3 Approval: Executive Director Issue Date: 22 Sep 08 Responsibility: Manager Custodial Services Review Period: 2 Years PROCEDURES FOR BODILY FLUID CLEAN-UP
Body Art Facility Infection Prevention And Control Plan Guideline
Body Art Facility Infection Prevention And Control Plan Guideline In accordance with the California Health and Safety Code, Section 119313, a body art facility shall maintain and follow a written Infection
Biosafety Spill Response Guide
Yale University Office of Environmental Health & Safety Biosafety Spill Response Guide Office of Environmental Health & Safety 135 College Street, 1 st Floor, New Haven, CT 06510 Telephone: 203-785-3550
INFECTION CONTROL SURVEYOR WORKSHEET
Exhibit 351 Ambulatory Surgical Center (ASC) INFECTION CONTROL SURVEYOR WORKSHEET (Rev: 142, Issued: 07-17-15, Effective: 07-17-15, Implementation: 07-17-15) Name of State Agency or AO (please specify)
BC Children s Hospital Mistake-Proofing Equipment Reprocessing. Laurence Bayzand Janice Penner Sue Fuller Blamey
BC Children s Hospital Mistake-Proofing Equipment Reprocessing Laurence Bayzand Janice Penner Sue Fuller Blamey Case for Collaboration Patient Safety Learning System (PSLS) trending identified an increased
Cleaning, Disinfection and Sterilization
Cleaning, Disinfection and Sterilization Paula Bullock, M.Ed., MT(ASCP), CIC Objectives Determine how to select and use the best agents/systems for cleaning List important components in cleaning, disinfection,
Standard laryngoscope blade assemblies Directions for use
1 2014 Welch Allyn, Inc. MM 721358 Ver. C Standard laryngoscope blade assemblies Directions for use Intended use About this document The rigid laryngoscope is a device used to examine and visualize a patient
Arizona State University Institutional Animal Care and Use Committee STANDARD INSTITUTIONAL GUIDELINE RODENT SURGERY
Arizona State University Institutional Animal Care and Use Committee STANDARD INSTITUTIONAL GUIDELINE RODENT SURGERY It is the policy of the IACUC that all survival surgeries involving rodents be conducted
EAST CAROLINA UNIVERSITY INFECTION CONTROL POLICY
EAST CAROLINA UNIVERSITY INFECTION CONTROL POLICY Department: Adult and Pediatric HealthCare/ECU-Doctor s Park, Building #2 Date Originated: October 28, 1992 Date Reviewed: 10.28.92, 12.14.94, Date Approved:
Inherent in any infection control strategy are two significant concepts: (i) Routine Practices, and (ii) Risk Assessment.
Infection Control Standards of Practice for Chiropodists and Podiatrists I. INTRODUCTION Infection control is considered an integral part of patient care. Concerns regarding the possible spread of blood-borne
33 Infection Control Techniques
CHAPTER 33 Infection Control Techniques Learning Outcomes 33.1 Describe the medical assistant s role in infection control. 33.2 Describe methods of infection control. 33.3 Compare and contrast medical
Cleaning, Disinfection, and Sterilization. Basics of Infection Prevention 2-Day Mini-Course 2013
Cleaning, Disinfection, and Sterilization Basics of Infection Prevention 2-Day Mini-Course 2013 2 Objectives Describe basic principles of cleaning, disinfection, sterilization Identify when to use cleaning,
THE BASICS OF DECONTAMINATION
THE BASICS OF DECONTAMINATION Objectives: Review the basic factors that impact cleaning and decontamination of surgical devices. Present the fundamentals of manual and automatic cleaning Review various
01.05 Cleaning, Sterilization, High-Level Disinfection and Storage of Patient Care Devices and Other Items
01.05 Cleaning, Sterilization, High-Level Disinfection and Storage of Patient Care Devices and Other Items Page 1 of 36 INDEX Page Purpose, Audience, Policy, General Recommendations & Electrical Safety...
Are You Always Ready for The Joint Commission or CMS Survey?
Are You Always Ready for The Joint Commission or CMS Survey? Sandy Garcia, RN, BSN, MBA Lead Consultant Certified Lean/Six Sigma Black Belt 1 Are You? Were You? Do You? Have You? 2 1 Infection Prevention
Christina Bradley Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK
Christina Bradley Hospital Infection Research Laboratory Queen Elizabeth Hospital Birmingham, UK Inadequate decontamination procedures and equipment malfunction were two leading causes of post endoscopic
Appendix J IBC Biohazard Spill Management Plan
OVERVIEW Prevention is the most important part of any spill management plan. Be sure to read and understand standard operating procedures (SOP s) and protocols for safe manipulation of biohazards before
Laboratory Biosafty In Molecular Biology and its levels
Laboratory Biosafty In Molecular Biology and its levels Workshop 16-17 Oct..2012 Guidelines Does not mean optional Laboratory Biosafety The Laboratory Biosafety Manual is an important WHO publication
Infection Control in the Dental Practice through Proper Sterilization
Midmark Corporation 60 Vista Drive P.O. Box 286 Versailles, Ohio 45380-0286 937-526-3662 midmark.com Infection Control in the Dental Practice through Proper Sterilization In today's world, the need for
THE BASICS OF STERILIZATION
THE BASICS OF STERILIZATION Objectives: State the importance of sterilization to patient care Review three key essentials of the sterilization process Describe sterilization methods used, application,
Disinfection and Sterilization of reusable medical devices in General Practice
Disinfection and Sterilization of reusable medical devices in General Practice Authors: - Donald MacCulloch, Pat Mead, Arthur Morris - AUGUST 2002 new zealand accredited laboratory ISO 9002 Registered
Response to Biological Spills in the Laboratory (Intentional or Accidental)
Response to Biological Spills in the Laboratory (Intentional or Accidental) Exposure Management For splash to eyes, mucous membranes, or broken area of the skin Irrigate eyes with clean water, saline or
DNA Analyst Training Laboratory Training Manual Protocol 2.02 Clean Technique
DNA Analyst Training Laboratory Training Manual Protocol 2.02 Clean Technique This laboratory protocol (or part thereof) has been provided as an example of a laboratory SOP, courtesy of the Illinois State
Cleaning. By the end of this chapter, you will be able to: Introduction. Definitions. Chapter 9
Chapter 9 By the end of this chapter, you will be able to: l Define the terms cleaning, decontamination, disinfection l List the factors that affect disinfection l Describe the aims of a cleaning schedule
Private Water Supplies Sampling Manual. A Field Guide
Private Water Supplies Sampling Manual A Field Guide Foreword This sampling manual details standard procedures for the collection, storage and transportation of samples at private water supplies in accordance
Bloodborne Pathogens Program Revised July, 5 2012
Bloodborne Pathogens Program Revised July, 5 2012 Page 1 of 16 Table of Contents 1.0 INTRODUCTION...3 1.1 Purpose...3 1.2 Policy.3 2.0 EXPOSURE CONTROL METHODS 4 2.1 Universal Precautions.4 2.2 Engineering
POLICY 08:18:00 BLOODBORNE PATHOGENS CONTROL PLAN
POLICY 08:18:00 BLOODBORNE PATHOGENS CONTROL PLAN I. Purpose and Scope The purpose of this plan is to establish guidelines and precautions for the handling of materials which are likely to contain infectious
Training on Standard Operating Procedures for Health Care Waste Management Swaziland 12 May, 2011
Training on Standard Operating Procedures for Health Care Waste Management Swaziland 12 May, 2011 Safe Infectious Waste Handling and Transport Objective Waste Overview Roles and Responsibilities of Waste
Mercyhurst University Athletic Training Program Bloodborne Pathogens Exposure Control Plan
Mercyhurst University Athletic Training Program Bloodborne Pathogens Exposure Control Plan In accordance with the Occupational Safety Health Administration (OSHA) Bloodborne Pathogens Standard, 29 CFR
4. Infection control measures
4. Infection control measures Apart from general hygienic practices and vaccination, staff of institutions should also adopt specific infection control measures against communicable diseases. The measures
Steam Sterilization Cycles for Lab Applications
Steam Sterilization Cycles for Lab Applications Presented by Gary Butler STERIS Life Sciences August 2009 Early Steam Sterilizers Koch Upright Sterilizer First Pressurized Sterilizer First built by Pasteur-Chamberlain
Best Practices for Cleaning, Disinfection and Sterilization in Health Authorities March 2007
Best Practice Guidelines for the Cleaning, Disinfection and Sterilization of Medical Devices in Health Authorities March 2007 Patient Safety Branch Ministry of Health 1 Foreword This document was developed
Infectious Waste Management Plan
Infectious Waste Management Plan Infectious Waste Management Plan USC Health & Safety Programs Unit 777-5269 POLICY: A. In keeping with the University of South Carolina's policy of providing protection
STERILE PROCESSING MANUAL Policy & Procedure
STERILE PROCESSING MANUAL Policy & Procedure TITLE: Dress Code and Personal Protective NUMBER: Equipment (PPE) Sterile Processing Department (SPD) Effective Date: March 2014 Page 1 of 6 Applies To: Holders
EAST CAROLINA UNIVERSITY INFECTION CONTROL POLICY. Methicillin-resistant Staph aureus: Management in the Outpatient Setting
EAST CAROLINA UNIVERSITY INFECTION CONTROL POLICY Methicillin-resistant Staph aureus: Management in the Outpatient Setting Date Originated: Date Reviewed: Date Approved: Page 1 of Approved by: Department
INFECTION CONTROL POLICY MANUAL
Page 1 of 7 POLICY MANUAL Key Words: personal protective equipment, PPE, safety equipment, infection control, standard precautions Policy Applies to: All staff employed by Mercy Hospital. Credentialed
Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes
Standards of Infection Prevention in Reprocessing Flexible Gastrointestinal Endoscopes Acknowledgements Copyright 2016 (SGNA). First published in 1996, revised 2000, 2005, 2007, 2008, 2011, 2012, 2015
INFECTION PREVENTION CHECKLIST FOR OUTPATIENT SETTINGS: MINIMUM EXPECTATIONS FOR SAFE CARE
INFECTION PREVENTION CHECKLIST FOR OUTPATIENT SETTINGS: MINIMUM EXPECTATIONS FOR SAFE CARE National Center for Emerging and Zoonotic Infectious Diseases Division of Healthcare Quality Promotion Version
Biosafety Guidelines. 5. Pipetting Mechanical pipetting devices must be available and used. Mouth pipetting is prohibited.
Biosafety Guidelines Biosafety Level 1 Laboratory Biosafety Level 1 is suitable for work involving well-characterized agents not known to consistently cause disease in immunocompetent adult humans, and
Reprocessing Flexible Endoscopes Avoiding Reprocessing Errors Critical for Infection Prevention and Control
Reprocessing Flexible Endoscopes Avoiding Reprocessing Errors Critical for Infection Prevention and Control By Bradley Catalone, Ph.D., and George Koos Flexible endoscopy procedures are now a routine part
Federal Wage System Job Grading Standard For Laboratory Working, 3511. Table of Contents
Federal Wage System Job Grading Standard For Laboratory Working, 3511 Table of Contents WORK COVERED... 2 WORK NOT COVERED...2 TITLES... 2 GRADE LEVELS... 2 NOTE TO USERS... 3 LABORATORY WORKER, GRADE
Expert Opinion. on the efficacy of. EndoDet and EndoDis + EndoAct. for cleaning and disinfection of. Olympus Gastroscope Type GIF-2T200.
Expert Opinion on the efficacy of EndoDet and EndoDis + EndoAct for cleaning and disinfection of Olympus Gastroscope Type GIF-2T200 with the Endo Thermo Disinfector ETD3 tested according to the recommendations
Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes
Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes Acknowledgements Copyright 2012, Society of Gastroenterology Nurses and Associates, Inc. (SGNA). First published in
Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes
Standards of Infection Control in Reprocessing of Flexible Gastrointestinal Endoscopes Acknowledgements Copyright 2012, Society of Gastroenterology Nurses and Associates, Inc. (SGNA). First published in
MEDICAL WASTE MANAGEMENT
MEDICAL WASTE MANAGEMENT Biological Safety INTRODUCTION PURPOSE Regulated medical waste is a designation for wastes that may contain pathogenic microorganisms which was previously termed infectious waste.
THE BASICS OF PACKAGING
THE BASICS OF PACKAGING Objectives: Review the regulatory requirements for packaging Present the purpose, function and essential characteristics of packaging Describe packaging options, their use, and
KGI MEDICAL WASTE MANAGEMENT PLAN
1revised 2012 I. Facility Information KGI MEDICAL WASTE MANAGEMENT PLAN a. Contact Person: Barbara Erwin Director Research Operations/Chemical and Biological Laboratory Safety/Chair. 909-607-0160 i. Facility:
Best Practices for Loaner Surgical Instruments
Best Practices for Loaner Surgical Instruments Copyright 2015, Sterile Processing University, LLC, All Rights Reserved. This material may not be copied or used without permission of the author. Nancy Chobin,
Water should be removed as soon as possible once the safety of the structure has been verified.
FACT SHEET Posted October 18 2005 Check List for Infection Control Concerns when Reopening Healthcare Facilities Closed due to Extensive Water and Wind Damage Prior to opening a healthcare facility that
Safe Handling of Cytotoxic Materials
Safe Handling of Cytotoxic Materials Kara Henman RN, MN, CON(C) Oncology Practice Consultant Cancer Care Nova Scotia What are hazardous drugs? Chemotherapy Immunosuppressive agents Biological agents Antiviral
Clinical Guideline for: Aseptic Technique
Clinical Guideline for: Technique Summary This guideline provides the principles of, Non Touch, and Clean Techniques to be implemented in the hospital environment. Key Points The essential elements of
BLOODBORNE PATHOGENS EXPOSURE CONTROL PLAN
BLOODBORNE PATHOGENS EXPOSURE CONTROL PLAN Facility Name: University of Arkansas at Little Rock Date of Preparation: 01-17-06 In accordance with the OSHA Bloodborne Pathogens Standard, 1910.1030, the following
Your Guide to Peritoneal Dialysis Module 3: Doing Peritoneal Dialysis at Home
Your Guide to Peritoneal Dialysis Module 3: 6.0959 in Preparing to do PD One of the most important things about PD is to keep the dialysis area and anything that comes in contact with the PD equipment
BESTPRACTICE. HTM01-05 Q&A: You can now follow us on Twitter! We have launched our HTM01-05 questions and answers service on Twitter.
ISSUE 4 BESTPRACTICE For Essential Requirements, hand washing is still a viable method of cleaning your instruments, but how do you validate for compliance with HTM01-05? Ensuring your instruments are
3039878000or8009926372
3039878000or8009926372 bi s a l e s @me s a l a bs. c om Regulatory officials and sterilization experts have voiced concerns regarding the appropriateness of using a Biological Indicator (BI) Ampoule interchangeably
Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material
PM 2.5 Nylon Filter Cleaning Date: August 25,2009 Page I of8 Standard Operating Procedure for Cleaning Nylon Filters Used for the Collection of PM 2. 5 Material Environmental and Industrial Sciences Division
GUIDELINES: Aseptic Technique for Rodent Survival Surgeries Surgical Area Animal Preparation 1.
GUIDELINES: Aseptic Technique for Rodent Survival Surgeries The following Guidelines are provided to ensure correct aseptic technique for all rodent survival surgeries. Surgical Area 1. The surgical area
Hand Hygiene and Infection Control
C Hand Hygiene and Infection Control Sirius Business Services Ltd www.siriusbusinessservices.co.uk Tel 01305 769969 [email protected] Whatever your First Aid, Fire Safety or Health & Safety
Chemotherapy Spill Response:
Chemotherapy Spill Response: Antineoplastic Spills Outside Of A Fume Hood Lisa Hudley, RN Training Coordinator Safety & Environmental Compliance William Guess Director Safety & Environmental Compliance
Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy!
Flexible Endoscopes: The A,B,C s of Monitoring Manual Cleaning Efficacy! Dr. Michelle J. Alfa, Ph.D., FCCM Medical Director, Clinical Microbiology, Diagnostic Services of Manitoba, Winnipeg, Canada Disclosure:
1.40 Prevention of Nosocomial Pneumonia
1.40 Purpose Audience Policy Statement: The guideline is designed to reduce the incidence of pneumonia and other acute lower respiratory tract infections. All UTMB healthcare workers and contract healthcare
Transducer and System Care and Cleaning. 2013 Koninklijke Philips N.V. All rights reserved. Published in USA.
l Transducer and System Care and Cleaning 4535 617 39631 Rev B December 2013 2013 Koninklijke Philips N.V. All rights reserved. Published in USA. Philips Ultrasound 22100 Bothell-Everett Highway Bothell,
ResMed Masks. Clinical Use Only. Disinfection and Sterilization Guide. NASAL MASKS Mirage Activa. Mirage Kidsta. Mirage Vista.
ResMed Masks and Sterilization Guide S Mirage Activa Mirage Kidsta Mirage Vista Ultra Mirage II Ultra Mirage NON-VENTED NASAL PILLOWS SYSTEMS Mirage Swift II NEW NASAL PILLOWS SYSTEM Mirage Swift NASAL
Biosafety Level 2 Criteria
Biosafety Level 2 Criteria Biosafety in Microbiological and Biomedical Laboratories (BMBL) 5th Edition Biosafety Level 2 (BSL-2): Biosafety Level 2 builds upon BSL-1. BSL-2 is suitable for work involving
Laboratory Safety Inspection Checklist (Biology and Biochemistry)
EOSMS 201-1A Date: 11/15/14 Page 1 of 9 Building Room Inspection Date Department Principal Investigator(s) Inspector(s) Instructions: This form is to be completed for laboratory self-inspection. Complete
Ambulance Service. Patient Care. and. Transportation Standards
Ambulance Service Patient Care and Transportation Standards Ministry of Health and Long-Term Care Emergency Health Services Branch Patient Care A. General Each operator and each emergency medical attendant
4. Cryogenic gloves are generally designed to protect the hands from intense cold or heat.
PROTECTIVE GLOVES In many University laboratories, exposure to chemicals, infectious agents, sharp objects, extreme temperatures and other hazards can create a potential for injury to the hand. Wherever
STERILIZATION د STERILIZATION
STERILIZATION. د STERILIZATION Is a term referring to any process that eliminates (removes) or kills all forms of microbial life, including transmissible agents (such as fungi, bacteria, viruses, spores
INFECTION CONTROL PRECAUTIONS
INFECTION CONTROL PRECAUTIONS Outline Standard Precautions Droplet Precautions Contact Precautions Airborne Precautions References STANDARD PRECAUTIONS Use Standard Precautions, or the equivalent, for
Health Service Executive Code of Practice for Decontamination of Reusable Invasive Medical Devices
PART 5a: RECOMMENDED PRACTICES FOR DENTAL SERVICES IN A CENTRAL DECONTAMINATION UNIT Health Service Executive Code of Practice for Decontamination of Reusable Invasive Medical Devices Reader Information
BSL 1 Laboratory Biosafety Manual
BSL 1 Laboratory Biosafety Manual Version 1.0 Idaho State University, Office for Research Institutional Biosafety Committee (IBC) 1651 Alvin Ricken Drive, Pocatello, ID 83201-8046 Phone: 208-282-2179 Fax:
OCCUPATIONAL EXPOSURE TO BLOODBORNE PATHOGENS (29 CFR 1910.1030)
I. Introduction OCCUPATIONAL EXPOSURE TO BLOODBORNE PATHOGENS (29 CFR 1910.1030) It is estimated that approximately 5.6 million workers in health care and other fields are exposed to bloodborne pathogens.
Bed Cleaning Procedure
This is an official Northern Trust policy and should not be edited in any way Bed Cleaning Procedure Reference Number: NHSCT/10/308 Target audience: Nursing and Midwifery Staff Sources of advice in relation
Checklist for a Physical Containment Level 2 (PC2) Laboratory Facility
Checklist for a Physical Containment Level 2 (PC2) Laboratory Facility (The Gene Technology Regulator s Guidelines for Certification of Facilities/Physical Containment Requirements applies) Facility Name:.
OSHA s Bloodborne Pathogens Standard 1910.1030
OSHA s Bloodborne Pathogens Standard 1910.1030 Jens Nissen & Kennan Arp Iowa OSHA Enforcement 515-281-3122 [email protected] or [email protected] Bloodborne Pathogens Standard Federal Law 29 CFR 1910.1030
INSTITUTIONAL POLICY AND PROCEDURE (IPP) Department: Manual: Section:
HOSPITAL NAME INSTITUTIONAL POLICY AND PROCEDURE (IPP) Department: Manual: Section: TITLE/DESCRIPTION POLICY NUMBER HANDLING OF H AZARDOUS DRUGS EFFECTIVE DATE REVIEW DUE REPLACES NUMBER NO. OF PAGES APPROVED
Oxivir Tb and Accelerated Hydrogen Peroxide (AHP) Frequently Asked Questions
Oxivir Tb and Accelerated Hydrogen Peroxide (AHP) Frequently Asked Questions What is Accelerated Hydrogen Peroxide (AHP)? AHP is a patented synergistic blend of commonly used, safe ingredients that when
CENTRAL SERVICE REGULATIONS FOR AMBULATORY SURGERY CENTERS
CENTRAL SERVICE REGULATIONS FOR AMBULATORY SURGERY CENTERS Last updated July 2014 Legal disclaimer: The International Association of Healthcare Central Service Materiel Management (IAHCSMM) provides the
Mock OSHA Inspection:
Mock OSHA Inspection: PASS or FAIL? Infection Control OSHA Dental Practice Act HIPAA Presented by Leslie Canham, CDA, RDA, CSP (Certified Speaking Professional) In the dental field since 1972, Leslie helps
OCCUPATIONAL SAFETY AND HEALTH ADMINISTRATION (OSHA)
OCCUPATIONAL SAFETY AND HEALTH ADMINISTRATION (OSHA) The OSHA/VOSH 1910.1030 Blood borne Pathogens Standard was issued to reduce the occupational transmission of infections caused by microorganisms sometimes
In use test of a Washer-disinfector reprocessing of flexible endoscopes
of a Washer-disinfector reprocessing of flexible endoscopes Olympus ETD 3 with PAA Clean Endoscope Copenhagen Hospitals Denmark In-use test of Olympus ETD 3 Washer-disinfector with PAA Cleaner and disinfectant
Decontamination and Waste Management www.biosecurity.sandia.gov
Decontamination and Waste Management www.biosecurity.sandia.gov SAND No. 2006-3684C Sandia is a multiprogram laboratory operated by Sandia Corporation, a Lockheed Martin Company, for the United States
IX. Decontamination and Spills
IX. Decontamination and Spills IX.1. Definitions Sterilization: the act or process, physical or chemical, which destroys or eliminates all forms of life, especially microorganisms. Decontamination: reduction
Revised 4-3-2012 EHS Biosafety. 1 Select appropriate containers/bags for autoclaving.
Requirements for Decontamination by Autoclaving Summary: Steam sterilization (autoclaving) is one of the most effective methods for decontaminating biohazardous material. Autoclaves use saturated steam
Biomedical Waste Management Plan
Biomedical Waste Management Plan USF Biomedical Waste Management Plan for: Facility Address: Facility Phone Number: Facility Contact: Division of Environmental Health and Safety 4202 E. Fowler Ave, CRS
To provide direction for the safe handling, administration and disposal of hazardous drugs.
Subsection: MEDICATION Related terms: Cytotoxic Drugs, Antineoplastic Drugs Authorized by: Clinical Directors CS-04-02-01 Page 1 of 9 Date Established: October 2006 Date For Review: September 2014 Dates
Giving safe injections
Giving safe injections A guide for nurses and others who give injections World Health Organization International Council of Nurses The World Health Organization (WHO) defines a safe injection to be one
Outlines: Original Artwork: LeMaitre. Light Post Adapter. Light Post Adapter. Fits Olympus, Storz, and Stryker Equipment* English Instructions for Use
Outlines: Original Artwork: LeMaitre VASCULAR Fits Olympus, Storz, and Stryker Equipment* English Instructions for Use Fits Olympus, Storz, and Stryker Equipment* Instructions for Use - English Model Numbers