Choice of guiding catheters and guidewires



Similar documents
Coronary & Peripheral Guidewires. Roby Rakhit

TURQUOISE PTCA BALLOON DILATATION CATHETER

Deflectable & Steerable Catheter Handbook

When Procedural Support really matters. Navien TM A+ Intracranial Support Catheter

Introduction to Cardiac Catheterisation The Nurse s Role

One wire. Many possibilities.

Ostial LAD: Single stent approach is the best. Antonio A. Pocoví, MD, FSCAI, MTSAC, Advisory Council Member, CACI

Tracheal Stent System

Hemostasis Solutions Boston Scientific is committed to improving patient care in the management of gastrointestinal bleeding.

The New Standard of Excellence, Redefining Deliverability, Versatility and Control

USE OF SIMULATION TO IMPROVE CARDIOVASCULAR STENT DEVELOPMENT

Ultrasound in Vascular Surgery. Torbjørn Dahl

Experience of Direct Coronary Stenting at National Institute of Cardiovascular Diseases

Biliary and Endovascular Product Catalog

Yukon Choice DES + Translumina - The polymer-free DES solution. Coronary Stent System for Drug Application

Evolu'on of Balloon Angioplasty From POBA to newer technologies NCVH May 26, 2015

Local Coverage Article: Endovascular Repair of Aortic Aneurysms (A53124)

Ch. 138 CARDIAC CATHETERIZATION SERVICES CHAPTER 138. CARDIAC CATHETERIZATION SERVICES GENERAL PROVISIONS

Orsiro Hybrid Drug Eluting Stent Industry's first hybrid DES

REPORTING STENT PLACEMENT FOR NONOCCLUSIVE VASCULAR DISEASE IN LOWER EXTREMITIES

Cardiac Catheterization

Overview of Newer Stent Devices for Aneurysm Treatment

Endovascular Repair of an Axillary Artery Aneurysm: A Novel Approach

RELAY Plus Thoracic Stent-Graft System with Transport Delivery System

Ureteral Pig tail Stent Sets

Clinical Research Intracoronary Stenting with Crushing in Coronary Artery Bifurcation Lesions: Initial Results and Medium-Term Follow Up

Selection & Use of Surgical Needles

California Health and Safety Code, Section

Coronary Stents. What is a Coronary Artery stent?

The Bioresorbable Vascular Stent Dr Albert Ko

SUCTION, ASPIRATING TUBES

Talent Thoracic Stent Graft with THE Xcelerant Delivery System. Expanding the Indications for TEVAR

Is Stenting or Coronary Artery By-pass Grafting the Better Treatment for This Patient?

A Patient s Guide to Minimally Invasive Abdominal Aortic Aneurysm Repair

Coronary Arteries: Number of Vessels/Number of Stents

STONY BROOK UNIVERSITY HOSPITAL VASCULAR CENTER CREDENTIALING POLICY

The left internal mammary artery (LIMA) is the

Right Heart Catheterization from the Arm

Imaging of Thoracic Endovascular Stent-Grafts

INTERVENTION IN OSTIAL CORONARY LESIONS:

SAMPLE. Asia-Pacific Interventional Cardiology Procedures Outlook to Reference Code: GDMECR0061PDB. Publication Date: May 2014

PRACTICE GUIDELINE TITLE: INTRAVENOUS LINE INSERTION: PERIPHERAL AND CENTRAL

Orthopedic Foot Instruments. Dedicated instruments for reconstructive foot surgery.

Wingspan Stent System with Gateway PTA Balloon Catheter

Patient Information Booklet. Endovascular Stent Grafts: A Treatment for Abdominal Aortic Aneurysms

StentViz Enhanced Stent Visualization

The Cardiac Society of Australia and New Zealand

Clinical Training for Visage 7 Cardiac. Visage 7

PERIPROSTHETIC IMPLANTS

GENERAL HEART DISEASE KNOW THE FACTS

WallFlex Biliary RX Stent. Fully, Partially and Uncovered Self-Expanding Metal Stents

An Informative Guide for Heart Catheterization Patients & Families

Coronary Bifurcation Treatment: Update from the European Bifurcation Club. Remo Albiero, MD Ist. Clinico S. Rocco Brescia (Italy)

Majestic Trial 12 Month Results

Different Properties of Coronary Stents György Ring 1, Eszter Bognár 1, Zsófia Bálint-Pataki 1, János Dobránszky 2

An Overview of Transradial Patient Set-up. Susan R. Cooney RN,BSN,CCRN Durham VAMC Cardiac Catheterization Lab Duke University School of Nursing

Development of Simulation Tools Software

Developments in Multi-Layer Medical Tubing Technology

Since 1991, when Parodi et al1 described a minimally

PERIPHERAL VASCULAR DEVICES

Malfunctioning PD Catheters

The Captivator II Snares are the first line of stiff and rounded snares available in multiple sizes with both a hot and cold snaring indication.

Mechanical Properties of Nitinol Stents and Stent-grafts. Comparison of 6 mm Diameter Devices

Instructions for Use. Device Description The AMPLATZER Vascular Plug II is a self-expandable nitinol mesh occlusion device (see Figure 1).

State-of-the-Art Technology in Cardiac CT

Carotid Artery Stenting? Comparison of the outcome of octogenarian patients according to the stent type. Data from a multicentre registry

GRAFTMASTER RX Coronary Stent Graft System PPL (2/1/13)

Captivator II. Single-Use Snares

Vascular Access. Chapter 3

PATIENT INFORMATION BOOKLET

2009/JUL/20 at 9:18 a.m. Doc number: M710950B001 [Uni+] VALIANT THORACIC STENT GRAFT WITH THE CAPTIVIA DELIVERY SYSTEM Instructions for Use

Peripheral Vascular U.S. Product Catalog

Coding Updates for 2013: Cardiology

NOVOSTE BETA-CATH SYSTEM

Purpose Members of the Department of Cardiology will provide cardiology services to patients of McLaren Greater Lansing.

Aneurysm Embolization System for the Treatment of Cerebral Aneurysms. Information for Patients and Families

Aesculap Veterinary Orthopaedics. Targon VET Interlocking Nail

PRECOMBAT Trial. Seung-Whan Lee, MD, PhD On behalf of the PRECOMBAT Investigators

2014 Procedural Reimbursement Guide Select Percutaneous Coronary Interventions

PERCUTANEOUS PD CATHETER IMPLANTATION SYSTEM

CORONARY ARTERY BYPASS GRAFTS, STENTS, AND EXTRACORONARY CARDIAC DZ. Charles White MD

12 Lead ECGs: Ischemia, Injury & Infarction Part 2

P REPLACEMENT SURGERY

Bard LifeStar Biliary Stent System

Overview of the Cardiovascular System

Solid shape molding is not desired in injection molding due to following reasons.

Peripheral Procedures in the Cardiac Cath Lab National AAPC April 2, Agenda

Cardiac Catheterization Lab Procedures

AND SYLLABUS FOR INTERVENTIONAL CARDIOLOGY SUBSPECIALITY TRAINING IN EUROPE

ANTERIOR LUMBAR INTERBODY FUSION (ALIF) Basic Anatomical Landmarks: Anterior Lumbar Spine

The Essential Role of CT Planning and TEE Guidance to Optimize Device Deployment

Knotilus TM. Anchor Instability Repair. Technique Guide

HYFLAM D3: Report of Scenarios 08/02/2012

Adult Forearm Fractures

Full Crown Module: Learner Level 1

What You Should Know About Cerebral Aneurysms

Transcription:

Choice of guiding catheters and guidewires Dr. Ahmad Elsayed Yousef, MD Lecturer of Cardiology, Ain Shams University Werner Frossmann-1929 1

X-ray for angioraphy (1950-1970) Cine frame from the first selective coronary arteriogram recorded by F. Mason Sones, Jr, on October 30, 1958. Cheng T O Circulation. 2003;107:e42 Copyright American Heart Association, Inc. All rights reserved. 2

Andreas Greuntzig-1977 On the shoulder of pioneers, we have witnessed the birth of a sub-speciality. Cardiovascular medicine has been forever transformed. Patient outcomes have never been better!! 3

HISTORY AND ORIGIN OF CATHETER Late Latin, from Greek: KATHETER, came from KATHIENAI, kathe- to send down : kat-, kata-, cata- + hienai- to send. 7 DEFINITION: A catheter is a hollow flexible tube that can be inserted into a body cavity, duct or vessel. Catheters thereby allow injection of fluids. The process of inserting a catheter is catheterization. 8 4

Guide catheter construction Ideal characteristics of catheters Better Torque Control Strength Radiopacity Flexible Atraumatic Tip Low Surface frictional resistance for good trackability over guide wire. 10 5

PARTS OF A CATHETER BODY HUB BODY TIP HUB TIP 11 MEASUREMENT: FRENCH CATHETER SCALE: The French catheter scale (most correctly abbreviated as Fr, but also often abbreviated as FR or F) is commonly used to measure the outer diameter of cylindrical medical instruments including catheters,needles etc. In the French Gauge system as it is also known, the diameter in millimeters of the catheter can be determined by dividing the French size by 3, thus an increasing French size corresponds with a larger diameter catheter. The following equations summarize the relationships: D(mm) = Fr/3 or Fr = D(mm)*3 Most commonly in adult Diagnostic Catheters of 5 7 Fr is used. 12 6

Guiding catheter For each given size of, its ID is either a standard, large or giant lumen Larger sizes o better opacification of the contrast o better guide support o allow pressure monitoring increased risk of ostial trauma, vascular complications and the possibility of kinking of catheter shaft TYPES OF CATHETERS: CLASSIFICATION : Catheters can be classified depending on SIDE HOLES : -: Single Hole -:End Hole with side holes. -:Blocked end with side holes only. SIZES : Abdominal 6-80 cm Thoracic or Carotid Arteries 100-120 cm NOTE: Size depends on : > age of the patient > selective or super selective study > size of the vessels. NOTE: Ideal practice is to use the smallest diameter catheter feasible for any particular study to minimize the risk of arterial damage by the procedure. 14 7

CATHETERS CAN BE BROADLY CLASSIFIED UNDER THESE GROUPS: DIAGNOSTIC CATHETERS Used for Angiographs. GUIDING CATHETERS Used for Angioplasty. Guiding catheters are like angiography catheters only difference is that guiding catheters are more stiffer & firm as it carries Balloon catheters, PTCA wires and stent delivery system. Mild stiffness comes due to the wire braided design. Good Push ability. Good Tractability. 15 8

Guiding catheter configuration Which catheter to choose? 9

Factors for catheter choice Access site: femoral/radial Location of ostium: R/L/IMA/SVG Anatomy: Patient size/ diameter of aorta Equipment required: kissing/bifurcation intervention Back-up Side holes Guiding catheter construction 10

Guiding catheter selection-lad Proper Fit We should have like 45 degrees at the angle while resting against the contralateral wall 11

Shorter curves fits superior orientation Longer curves fits inferior orientation Tip orientation with different curve lengths So what you think about that? Too long What to do? Decrease curve length by 0.5 Too short What to do? Increase curve length by 1 The improper fit 12

RCA osteal take-off Horizontal Inferior Superior Right coronary curves 13

Guiding catheter selection-rca Anterior origin: MP Normal: JR 4.0 Lateral origin: AL 1.0 or 2.0 Guiding catheter selection- SVGs and IMAs 14

Bypass graft curves LIMA LIMA: JR 4.0 OR LIMA catheter 15

SVG- LAD/LCX Superior take off: JR 4.0 OR LCB Horizontal origin: JR 4.0 OR MP SVG to RCA Horizontal origin: JR 4.0 OR MP Inferior take-off : MP 16

LCA- Back up support Extra back up (EBU) curves: Broad secondary curve that braces against the contralateral wall for better back-up support Co-axial support Amplatz coronary family Sometimes help Need extra-cautious manipulation 17

Backup force 3 factors o Catheter size o Area of contact made by cath on Ao o Angle (theta) of cath on the reverse side of Ao The angle (theta) determines the force that can dislodge the guiding catheter. 18

If this angle is 90⁰, it results in a greater backup force. Therefore a lower position is preferable as the point of contact on the reverse side of the aorta because the angle approaches 90º 19

Child in Mother Technique 110cm long 5Fr guide (Child) in 100cm long 6Fr/7Fr guide catheter (Mother) May provide up-to 70% more support Trauma to vessel dissection Air embolism usually occurring during intubation of child catheter. Important issues: Access: Femoral: long sheath is sometimes essential in special situations: iliac tortuosity, AAA. Radial: use slippery wire with visualization of wire tip while advancing catheter. Co-axial positioning of catheter tip is of extreme importance. To disengage Amplatz Left catheter: push in and rotate instead of pulling out. 20

Important issues- pressure tracing from guiding catheter N.B: side holes allow perfusion but don t prevent catheter induced injusry of coronary ostea. Radial approach 21

22

23

Thromboembolism Coronary dissection 24

Take Home messages Ensure Co-Axial positioning at ostium Knowledge of the anatomy is vital Try to use the smallest possible diameter Try to have all sizes of all catheters in your cathlab Be gentle! Pressure is your best friend 25

How to choose between the different guidewires?? Complications of guidewires 26

Complications of guidewires Complications of guidewires 27

Complications of guidewires Historic background In 1977, professor Grüntzig performed the first coronary angioplasty using a blunt, closed-end, inner balloon catheter with a short guidewire attached to its tip. The safe crossing of severe stenoses with this large, nonmanoeuvrable catheter was limited. Moreover with this system it was impossible to perform independent movements of the wire and balloon. In 1982 Simpson et al reported the first experience with a new over-the-wire balloon system. Since then, guidewire technology has rapidly advanced. 28

Guidewires have the following functions: To track through the vessel, To access the lesion, To cross the lesion atraumatically, To provide support for interventional devices. Guidewire performance Characteristics Torque control Trackability Steerability Flexibility Prolapse tendency Radiopacity/visibility Tactile feedback Crossing Support Definition Is an ability to apply rotational force at a proximal end of a guidewire and have that force transmitted efficiently to achieve proper control at the distal end Is an ability of a wire to follow the wire tip around curves and bends without bucking or kinking, to navigate anatomy of vasculature Is an ability of a guidewire tip to be delivered to the desired position in a vessel Is an ability to bend with direct pressure Tendency of the body of a wire not to follow the tip around bends Is an ability to visualise a guidewire or guidewire tip under fluoroscopy. Is tactile sensation on a proximal end of a guidewire that physician has that tells him what the distal end of the guidewire is doing Is an ability of a guidewire to cross lesion with little or no resistance Is an ability of a guidewire to support a passage of another device or system over it 29

Structure of the guidewire CORE The inner part of the guidewire that extends through the shaft of the wire from the proximal to the distal part where it begins to taper. It is the stiffest part of the wire that gives the stability and steerability to the guidewire from its proximal end to the distal tip. Core material Stainless steel: provides excellent support Better transmission of push force and torque characteristics BUT Less flexible More susceptible to kinking. Nitinol: Excellent flexibility More durable than stainless steel BUT Less torqueable. 30

Core diameter Larger diameters improve the support, and allow 1:1 torque response. Smaller diameters enhance the flexibility of the wire. Core taper The core of a guidewire usually is tapered along its length. Shorter taper and smaller numbers of widely spaced gradual tapers enhance the support and transmission of push force Longer tapers and larger numbers of more segmental tapering enhance the flexibility. 31

TIP The tip is the distal end of a guidewire. One-piece core or core-to-tip design: The core extends all the way to the tip. Less flexible These wires have precise steering and tip control. Two-piece or shaping ribbon design: The core stops just short of the distal tip. very flexible less reliable torque control. Coating The purpose of the coating is: To reduce frictions helping the wire negotiate tortuous anatomy. To improve deliverability of interventional equipment over the wire. Two types of coatings are used: Hydrophilic coatings They provides a lubricious, low friction feel inside the vessel and more trackability. They are recommended for operators with some experience as they carry the risk for subintimal movement, dissection and perforation of coronary artery. Hydrophobic coatings These are silicone based coatings which repel water and are applied on the working length of the wire, with the exception of the distal tip. Silicone coating has higher friction, more stable feel inside the vessel. 32

Classification of guidewires There is no uniform clinical classification of guidewires. However, guidewires could be classified by tip flexibility, device support and coating type. BY TIP FLEXIBILITY Floppy ChoICE Floppy - Boston Scientific, Natick, MA, USA; IQ - Boston Scientific; Mailman - Boston Scientific; Asahi Soft - Abbott Vascular, Redwood City, CA, USA; Hi-Torque Balance - Abbott Vascular; Hi-Torque Balance Middleweight - Abbott Vascular; Hi-Torque Transverse - Abbott Vascular; Wizdom ST - Cordis, Johnson & Johnson, Warren, NJ, USA) Intermediate ChoICE Intermediate - Boston Scientific; Hi-Torque Intermediate - Abbott Vascular; Stabilizer - Cordis) Standard ChoICE Standard - Boston Scientific; Shinobi - Cordis) Light ChoICE Floppy - Boston Scientific ChoICE PT Floppy - Boston Scientific Asahi Light - Abbot Vascular Hi-Torque Whisper LS - Abbott Vascular Hi-Torque Balance - Abbott Vascular Wizdom - Cordis BY DEVICE SUPPORT Moderate support IQ - Boston Scientific PT Graphix - Boston Scientific Luge - Boston Scientific Hi-Torque Whisper MS - Abbott Vascular Hi-Torque Balance Middleweight - Abbott Vascular; Stabilizer - Cordis Extra support ChoICE Extra Support - Boston Scientific ChoICE PT Extra Support - Boston Scientific Asahi Grand Slam - Abbott Vascular Hi-Torque Whisper ES - Abbott Vascular Hi-Torque Balance Heavyweight - Abbott Vascular Stabilizer Plus - Cordis 33

BY COATING TYPE Hydrophilic coating ChoICE PT Floppy - Boston Scientific PT Graphix - Boston Scientific Asahi Fielder - Abbott Vascular Hydrophobic coating IQ - Boston Scientific Asahi Soft - Abbott Vascular Guidewire manipulation SHAPING THE WIRE TIP 34

STEERING OF THE WIRE THROUGH THE VESSEL Small alternating rotations to the left and right are used The ideal wire can maintain 1:1 torque response. Torque response is especially important in tortuosity. Free wire tip motions are important. Avoid excessive rotations to prevent the wire tip fracture. Dye injection Usage of different angiographic views can help to cross the lesion safely and successfully Selection of guidewires The selection of a guidewire Guidewire selection is an essential component in a successful coronary intervention. Although the guidewire selection is influenced by criteria related to the vessel anatomy, the lesion morphology and the devices to be used, in real life scenario the guidewire selection is based upon operator's experience and preference. Rather than trying to use all wires that are on the market, it is more common for the operator to become an expert with a limited number of wires. 35

Left main intervention WIRES FOR LEFT MAIN PCI The choice of a guidewire is not of critical importance. Wire selection usually includes spring tip guidewire designed for frontline lesions, for example, ChoICE Floppy (Boston Scientific), Hi-Torque Balance Middleweight (Abbott Vascular) Wires for bifurcation PCI WIRES FOR BIFURCATION PCI The ChoICE Floppy (Boston Scientific), Hi- Torque Balance Middleweight (Abbott Vascular) will be the initial wires to use. In many cases we can perform successful PCI using only floppy wires and adequate other interventional devices 36

Some tips and tricks Before the guidewire insertion into the side branch it is important to choose the most appropriate angiographic view. Accessing the side branch before stent implantation: The ChoICE Floppy (Boston Scientific), Hi-Torque Balance Middleweight (Abbott Vascular) will be the initial wires to use In the presence of difficulties accessing the side branch some hydrophilic wires such as the ChoICE PT Floppy (Boston Scientific), PT Graphix (Boston Scientific) or Asahi Fielder (Abbott Vascular) may become useful. BUT put in mind.. it is important to monitor the distal position of the wire tip. These wires also should not to be jailed because of the risk of wire rupture during pullback. Therefore after accessing the side branch with hydrophilic wire and angulation modification, it is possible to exchange for floppy wire. 37

Tip shaping Tip shaping 38

After stent implantation, we recommend to Looping wire technique with making a loop of the distal segment of the wire is recommended before approaching the main vessel stent and advancing the wire through it. Wires for dissections WIRES FOR DISSECTIONS Wires like a ChoICE Floppy or Asahi Soft might be suitable for crossing a spiral dissection. The parallel wire technique can be recommended if a dissection plane is entered with the first. 39

Wires for calcified lesions Our practice is to start with ChoICE Floppy (Boston Scientific). If it fails to cross the lesion, the next step is to choose floppy hydrophilic wire such as the ChoICE PT Floppy (Boston Scientific) or Asahi Fielder (Abbott Vascular). In order to advance the stent over the less supportive wire we use plaque modification and lesion preparation with cutting balloons. Rotablation can be a good option provided good experience with the procedure. Tortuous anatomy WIRES FOR TORTUOUS ANATOMY In most cases we apply the same strategy as in calcified lesions and start with a very floppy wire. In addition, sometimes supportive wire (buddy wire) for device delivery could be used. 40

The usage of a buddy wire that is advanced in a different vessel may provide additional support. Insertion of guidewires in tortuous coronary arteries frequently induces vessel wall straightening and shortening which can lead to appearance of false lesions or coronary pseudo-stenosis in the angiography. Pullback of the guidewire normally leads to resolution of these "lesions" CTO Dr. Sudhir Rathore MD, MRCP (UK) TCT 2009 41

Please. Go slow Schedule procedure when you have enough time to do the case Take multiple angiographic projections Think about what you are seeing Don t be afraid to stop and bring the patient back later Take home message There are multiple factors that contribute to a successful coronary intervention and appropriate guidewire selection is one of them. The guidewire market is very broad and floating. Variation in structure of guidewire can create significant differences is wire performance that affects the suitability of a particular wire in varying clinical situations. However it is more reasonable to focus on selected guidewires and to understand their properties rather than trying to use all of them. 42

Remember the date!! Thank you for attention 43