I. Methodology for Onda's Hydrophone Calibrations



Similar documents
A Robust Hydrophone for HIFU Metrology

APPLICATION NOTE AP050830

APPLICATION NOTE ULTRASONIC CERAMIC TRANSDUCERS

0HDVXULQJWKHHOHFWULFDOSHUIRUPDQFH FKDUDFWHULVWLFVRI5),)DQGPLFURZDYHVLJQDO SURFHVVLQJFRPSRQHQWV

SIGNAL GENERATORS and OSCILLOSCOPE CALIBRATION

Using Current Transformers with the 78M661x

WHY DIFFERENTIAL? instruments connected to the circuit under test and results in V COMMON.

Current and Temperature Ratings

Agilent AN 1316 Optimizing Spectrum Analyzer Amplitude Accuracy

Accurate Loss-of-Signal Detection in 10Gbps Optical Receivers using the MAX3991

Lock - in Amplifier and Applications

MICROPHONE SPECIFICATIONS EXPLAINED

ε: Voltage output of Signal Generator (also called the Source voltage or Applied

LogicLab s.r.l Via della valle Carate Brianza (MI) Tel Fax

Op-Amp Simulation EE/CS 5720/6720. Read Chapter 5 in Johns & Martin before you begin this assignment.

Agilent AN 1315 Optimizing RF and Microwave Spectrum Analyzer Dynamic Range. Application Note

Current Probes, More Useful Than You Think

Basic Op Amp Circuits

Lab Exercise 1: Acoustic Waves

Chapter 12 Driven RLC Circuits

Positive Feedback and Oscillators

Reading: HH Sections , (pgs , )

MODEL 1211 CURRENT PREAMPLEFIER

Laboratory 4: Feedback and Compensation

Description. Output Stage. 5k (10k) - + 5k (10k)

Acousto-optic modulator

Output Ripple and Noise Measurement Methods for Ericsson Power Modules

MATRIX TECHNICAL NOTES

1. Oscilloscope is basically a graph-displaying device-it draws a graph of an electrical signal.

Amplifier for Small Magnetic and Electric Wideband Receiving Antennas (model AAA-1B)

S-Parameters and Related Quantities Sam Wetterlin 10/20/09

DVT913 TV CHANNEL CONVERTER

Ultrasound Distance Measurement

A 1 to 2 GHz, 50 Watt Push-Pull Power Amplifier Using SiC MESFETs. high RF power. densities and cor- capacitances per watt.

High Pressure Measurements with the High Pressure Microphone Calibrator Type Introduction. The High Pressure Microphone Calibrator Type 4221

Diode Applications. by Kenneth A. Kuhn Sept. 1, This note illustrates some common applications of diodes.

On-site Measurement of Load and No-load Losses of GSU Transformer

FREQUENCY RESPONSE ANALYZERS

Keysight Technologies Understanding the Fundamental Principles of Vector Network Analysis. Application Note

BJT Amplifier Circuits

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

LS RS. Figure 1: Assumed inductor model

E190Q Lecture 5 Autonomous Robot Navigation

DRAFT. University of Pennsylvania Moore School of Electrical Engineering ESE319 Electronic Circuits - Modeling and Measurement Techniques

Gage Applied Technologies

CONDUCTED EMISSION MEASUREMENT OF A CELL PHONE PROCESSOR MODULE

DSA800 Series Spectrum Analyzer

PIPELINE INSPECTION UTILIZING ULTRASOUND TECHNOLOGY: ON THE ISSUE OF RESOLUTION By, M. Beller, NDT Systems & Services AG, Stutensee, Germany

Experiment 7: Familiarization with the Network Analyzer

Use Data Budgets to Manage Large Acoustic Datasets

RF Network Analyzer Basics

EXPERIMENT NUMBER 5 BASIC OSCILLOSCOPE OPERATIONS

QUICK START GUIDE FOR DEMONSTRATION CIRCUIT BIT DIFFERENTIAL ADC WITH I2C LTC2485 DESCRIPTION

AVR127: Understanding ADC Parameters. Introduction. Features. Atmel 8-bit and 32-bit Microcontrollers APPLICATION NOTE

Taking the Mystery out of the Infamous Formula, "SNR = 6.02N dB," and Why You Should Care. by Walt Kester

RF Measurements Using a Modular Digitizer

DUAL%CHANNEL BROADBAND%LINEAR%AMPLIFIER Model&A800D

CS4525 Power Calculator

BJT Amplifier Circuits

Alternating-Current Circuits

Engineering Sciences 151. Electromagnetic Communication Laboratory Assignment 3 Fall Term

TESTS OF 1 MHZ SIGNAL SOURCE FOR SPECTRUM ANALYZER CALIBRATION 7/8/08 Sam Wetterlin

Kit Watt Audio Amplifier

Optimizing IP3 and ACPR Measurements

Linear Parameter Measurement (LPM)

Introduction to acoustic imaging

Enhancing Second Harmonic Suppression in an Ultra-Broadband RF Push-Pull Amplifier

Physics 120 Lab 6: Field Effect Transistors - Ohmic region

Laboratory Manual and Supplementary Notes. CoE 494: Communication Laboratory. Version 1.2

PowerAmp Design. PowerAmp Design PAD135 COMPACT HIGH VOLATGE OP AMP

Using Simple Calibration Load Models to Improve Accuracy of Vector Network Analyzer Measurements

Impedance 50 (75 connectors via adapters)

DRTS 33. The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters

Eliminator 15-inch two-way high-output stage system 350-watts continuous Ring-Mode Decoupling (RMD ) for improved sound quality and dynamic integrity

How To Sell A Talan

Constructing a precision SWR meter and antenna analyzer. Mike Brink HNF, Design Technologist.

UNDERSTANDING POWER FACTOR AND INPUT CURRENT HARMONICS IN SWITCHED MODE POWER SUPPLIES

TETRIS High Impedance Active Probe. Features:

10EaZy SW White Paper Choosing the right hardware

Lab 9: The Acousto-Optic Effect

FREJA 300 Relay Testing System

Keysight Technologies 8 Hints for Better Spectrum Analysis. Application Note

AVX EMI SOLUTIONS Ron Demcko, Fellow of AVX Corporation Chris Mello, Principal Engineer, AVX Corporation Brian Ward, Business Manager, AVX Corporation

F = S i /N i S o /N o

Electrical tests on PCB insulation materials and investigation of influence of solder fillets geometry on partial discharge

Keysight N9320B RF Spectrum Analyzer

XR-2206 Monolithic Function Generator

Robot Perception Continued

HIGH RELIABILITY POWER SUPPLY TESTING

Op amp DC error characteristics and the effect on high-precision applications

Precision, Unity-Gain Differential Amplifier AMP03

Op Amp Circuit Collection

Sound absorption and acoustic surface impedance

AN-691 APPLICATION NOTE One Technology Way P.O. Box 9106 Norwood, MA Tel: 781/ Fax: 781/

T = 1 f. Phase. Measure of relative position in time within a single period of a signal For a periodic signal f(t), phase is fractional part t p

INTERFERENCE OF SOUND WAVES

Frequency Response of Filters

Current Probes. User Manual

Yaesu FT MHz PA output and filter simulations Marc Vlemmings, PA1O - Eindhoven, The Netherlands

TETRIS 1000 / High Impedance Active Probe Order-No: Features:

Transcription:

190 ammerwood venue Sunnyvale, 94089 Phone (408) 745-0383 Fax (408) 745-0956 www.ondacorp.com I. Methodology for Onda's ydrophone alibrations alibration is obtained by comparison techniques to reference hydrophones calibrated by the National Physical aboratory (NP) in the UK. For calibration below 0 Mz, the stepped single frequency comparison technique is used, and for calibration above 0 Mz, the harmonic comparison technique is employed. The procedures are as follows: Stepped single frequency comparison technique: 1) The reference hydrophone is placed at the focus of a broadband source [1], which is driven by a sinusoidal tone burst. The reference hydrophone is aligned by maximizing the hydrophone signal with the source driven at the maximum end of its frequency range (0 Mz for a 1-0 Mz calibration, 5 Mz for a 0.5-1 Mz calibration). The voltage driving the source is then swept in across the frequency range (i.e., 1 to 0 Mz, or 0.5 to 1 Mz). t each frequency, the output of the reference hydrophone is measured and digitally filtered to separate the primary frequency component from any harmonics that may be present in the signal. The reference hydrophone has previously been calibrated by NP, so that the voltage it detects can be translated into a pressure value, hence calibrating the pressure output of the source as a function of frequency, at the source s focal position. ) The reference hydrophone is then replaced by the hydrophone to be calibrated, and re-aligned. The voltage driving the source is again swept across the same frequency settings, and the output of the hydrophone at the fundamental of the driving signal is determined and recorded. This voltage is then divided by the pressure amplitude (measured in step 1) to determine the output of the hydrophone in Volts per Pascal (V/Pa). armonic comparison technique: 1) The reference hydrophone is placed at a position in the non-linear acoustic field produced by a plane circular source transducer driven by a.5mz sinusoidal tone-burst signal. The position was chosen where many strong harmonics of the fundamental frequency are generated in the distorted pressure waveform due to non-linear propagation. Using a 50 Mz high-pass filter, the reference hydrophone is aligned by maximizing the hydrophone s high frequency components in the plane at that position to an optimal orientation. The pressure waveform at that position is recorded after the high-pass filter was removed. n Onda_ydroalMethod_0131018.doc Page 1 of 6

FFT of the waveform is calculated, and the voltage amplitude at each harmonic is determined. The reference hydrophone has previously been calibrated by NP, so that the voltage it detects can be translated into a pressure value, hence calibrating the pressure output of the source at each harmonic. ) The reference hydrophone is replaced by the hydrophone to be calibrated, which is then aligned the same way at the same position as the reference hydrophone. The pressure waveform is recorded with the absence of the high-pass filter. n FFT of the pressure waveform is then calculated, and the voltage at each harmonic is determined. This voltage is then divided by the pressure amplitude (measured in step 1) to determine the output of the hydrophone in Volts per Pascal (V/Pa). II. alibration Uncertainty The uncertainties in the calibration are predominantly systematic in origin, the random error is negligible in comparison. The dominant source of uncertainty is the uncertainty in the calibration of the reference hydrophone (as provided by NP). The other sources of uncertainty are temperature variation (± 1 º, resulting in calibration variations of ± 1%) source stability (± %) digitizer error (± %) and positional repeatability (this effect varies with frequency, from ±.5% at 1 Mhz to ± 5% at 0 Mz). The combined effect of these error sources is calculated according to [-3], with a coverage factor of k = to calculate the uncertainty with a 95% confidence level. This yields uncertainties that are typically within ± 1.5 db from 0.5 to 1 Mz, ± 1 db from 1 to 15 Mz, ± 1.5 db from 15 to 0 Mz, ±. db from 0 to 40 Mz and ± 3 db from 40 to 60 Mz. Onda_ydroalMethod_0131018.doc Page of 6

III. Important onsiderations Regarding Electrical oading onditions M and MB hydrophones are sold with a standard 1-0 Mz calibration into a 50-ohm load. G, NP (formerly NV), N (formerly NZ), and NR hydrophones are modular i.e., they allow the user to choose different amplifiers or no amplifier at all. Because a hydrophone's sensitivity depends on the electrical impedance of the detector or amplifier it is attached to, modular hydrophones are sold with an open-circuit calibration, called formally the "End-of-cable Open ircuit Sensitivity", or EO sensitivity for short, and designated as M c ( f ) [4-5]. M c ( f ) is determined by measuring the sensitivity of the hydrophone with a laboratory amplifier, and then subtracting the effect of that amplifier's gain, as well as that of its finite (i.e., non-infinite load). Note that the G, NP, and N hydrophones may be connected directly to an amplifier without any cable, so the EO sensitivity is defined as the sensitivity that would be measured at the hydrophone's connector into an infinite load. Of course, actual hydrophone usage is made with the customer's choice of amplifier and/or other form of detector. In such cases, the customer may calculate the effect of the amplifier/detector configuration on the hydrophone sensitivity, by mathematically combining hydrophone calibration data and amplifier data, as described below: Obtaining amplifier-loaded sensitivity from EO data The sensitivity of the hydrophone/preamp combination ( M ( f ) ) can be determined using the following formula [4]: where 1/ Re( Z ) Im( ) + Z ( f ) = G( f ) M c ( f ) {Re( Z ) + Re( Z )} + {Im( Z ) + Im( Z )} M (1) G ( f ) = amplifier gain (as a function of frequency) Z = input impedance of amplifier Z = impedance of hydrophone Re( Z) indicates the real part of a complex number Z Im( Z) indicates the imaginary part of a complex number Z If the hydrophone and amplifier impedances are primarily capacitive (as is the case with all Onda hydrophones and amplifiers), Eq. (1) can be approximated: Onda_ydroalMethod_0131018.doc Page 3 of 6

M ( f ) = G( f ) M c ( f ) (a) + where and and are the capacitance of the amplifier and hydrophone, respectively. is provided with the EO calibration data shipped with each modular hydrophone. Onda provides with the calibration sheets for all 010 and 00 amplifiers, and can be purchased for an additional price for other Onda amplifiers. Note that in some cases, adaptors or short cables may be used between the hydrophone connector and the amplifier. This additional connection length may have an effect which can be approximated by a connection capacitance, designated, which can be accounted for by the following modification to Eq. (): M ( f ) = G( f ) M c ( f ) (b) + + common example of such a connector is the right angle SM connector (Onda part number R- MF) which is used to connect an Onda G hydrophone to an 010 or 00 in a right-angle configuration; = 1.6 pf for this case. Note: ustomers are advised to perform their own assessment of the accuracy of Eqs. (a-b) for their configuration. In particular, customers should be aware that extension cables connecting the hydrophone to the preamplifier or detector may not be accurately modeled as purely capacitive at higher frequencies; in such cases, customers are advised to consider purchasing a separate calibration which includes the use of their extension cables. Onda_ydroalMethod_0131018.doc Page 4 of 6

IV. Other ommon alibration Units M ( f ) is in units of voltage divided by pressure and is typically expressed in V/Pa. lternatively, it can be expressed in units of db re 1V/µPa using Eq. (3): dbre 1V / µ Pa[ M ( f )] = 0*log10[ M ( f )] 10 (3) It is also common to express the calibration in terms of acoustic intensity. Strictly speaking, a simple calibration factor exists only under conditions of a sinusoidal signal and under the assumption that intensity is equal to the time-averaged value of the pressure squared divided by the acoustic impedance of the medium. Under these conditions the following relation applies: I = VRMS / K (4) where I is the acoustic intensity, and V RMS is the root-mean-square voltage of the sinusoidal signal. The calibration factor K is given by K = z a[ M ( f )] (5) where za is the acoustic impedance of the medium (1.5 MRayls for water). Because it is common to express I in units of W/cm^, K 10 { volts cm / Watt} = 1.5x10 [ M ( f ){ volts / Pa}] for water (6) V. Examples Example I. n NV-0400 is supplied by Onda with an EO calibration: M c ( f ) = 50 nv/pa at 5 Mz, and = 80 pf We now wish to determine the sensitivity for this hydrophone connected to an -010 preamplifier. s determined from the datasheet for the -010, the amplifier impedance is capacitive with = 7 pf, and the gain is 0 db ( G ( f ) = 10). pplication of Eqs. (a) and 3 shows that M ( 5Mz) = 460 nv/pa, or 46.7. db re 1V/µPa. pplication of Eq. (6) yields K (5 Mz) = 3.17 x 10-3 V cm /W. Onda_ydroalMethod_0131018.doc Page 5 of 6

Example II. n G-000 is supplied by Onda with an EO calibration: M c ( f ) = 45 nv/pa at 5 Mz, and = 13 pf We now wish to determine the sensitivity for this hydrophone connected to an -010 preamplifier through an R-MF connector: s determined from the datasheet for the -010, the amplifier impedance is capacitive with = 7 pf, and the gain is 0 db ( G ( f ) = 10). pplication of Eqs. (b) and 3 shows that M ( 5Mz) = 71 nv/pa, or 51.3dB re 1V/µPa. pplication of Eq. (6) yields K (5 Mz) = 1.10 x 10-3 V cm /W. References [1]. Selfridge and P.. ewin, Wideband Spherically Focused PVDF coustic Source for alibration of Ultrasound ydrophone Probes, IEEE Trans. UFF, 47(6) 137-1376, 000. [] R.. Preston, D.R. Bacon, and R.. Smith, alibration of Medical Ultrasound Equipment: Procedures and ccuracy ssessment, IEEE Transactions on Ultrasonics, ferroelectrics, and Frequency ontrol, Vo. UFF-5(), pp. 110-11, 1988. [3] M.. Ziskin, Measurement Uncertainty in Ultrasonic Exposimetry, in Ultrasonic Exposimetry, M.Ziskin and P. ewin, eds. (R Press), hapter 14, 1993. [4] UM/NEM: coustic Output Measurement Standard for Diagnostic Ultrasound Equipment. merican Institute of Ultrasound in Medicine, 14750 Sweitzer ane, Suite 100, aurel MD 0707-5906; National Electrical Manufacturers ssociation, 1300 North 17 th Street, Suite 1847, Rosslyn V 09. [5] IE 617-: Ultrasonics ydrophones Part : alibration for Ultrasonic Fields Up to 40 Mz. Onda_ydroalMethod_0131018.doc Page 6 of 6