Measuring. User Manual



Similar documents
Care and Use of the Compound Microscope

To measure an object length, note the number of divisions spanned by the object then multiply by the conversion factor for the magnification used.

Name Class Date Laboratory Investigation 4B Chapter 4: Cell Structure

Leica MZ125. Modular high-performance stereomicroscope with 12.5:1 zoom

MICROSCOPY. To demonstrate skill in the proper utilization of a light microscope.

FirstView 3 Reflector Telescope Owner s Manual

MT-30 & MT-90 Series. Advanced Academic Microscopes/ Advanced Academic Polarizing Microscope INSTRUCTION MANUAL

MITOSIS IN ONION ROOT TIP CELLS: AN INTRODUCTION TO LIGHT MICROSCOPY

30 mm Diameter Main Tube 1-4x24 Models

National Optical & Scientific Instruments Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

CrossFireII. 1-Inch Diameter Main Tube RIFLESCOPE MANUAL RIFLESCOPE RIFLESCOPE

WAVELENGTH OF LIGHT - DIFFRACTION GRATING

Microscope Lab Introduction to the Microscope Lab Activity

PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES PORTABLE MICROSCOPES

EXPERIMENT #1: MICROSCOPY

VISM Evolution Scope Series

Installation and use of Millett Tactical TRS-1 and TRS-2 scopes TRS-1 TRS-2. Before starting, make sure firearm is UNLOADED!!

TS-E24mm f/3.5l TS-E45mm f/2.8 TS-E90mm f/2.8 Instructions

National Optical & Scientific Instrument Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

Compound microscope (Hund)

RIFLESCOPE RIFLESCOPE MANUAL

How To Use An Asbestos Microscope

18-270mm F/ Di II VC PZD for Canon, Nikon (Model B008) mm F/ Di II PZD for Sony (Model B008)

GRID AND PRISM SPECTROMETERS

VISM CQB Scope Series

Geometric Optics Converging Lenses and Mirrors Physics Lab IV

R/T TACTICAL MONOCULAR

AF 70~300 mm F/4-5.6 Di LD Macro 1:2 (Model A17)

Living up to Life. Leica LAS Live Image Builder Dynamic Software for Fast Widefield Imaging and Extended Depth of Focus

Infrared Viewers. Manual

The Vortex Spitfire 1x Prism Scope

National Optical & Scientific Instruments Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

THE COMPOUND MICROSCOPE

Solution Derivations for Capa #14

Leica A60 S / Leica A60 F

Operator s Maintenance Manual STEINER BINOCULARS. Military ruggedness and precision 07/03

2) A convex lens is known as a diverging lens and a concave lens is known as a converging lens. Answer: FALSE Diff: 1 Var: 1 Page Ref: Sec.

SP AF 17~50 mm F/2.8 XR Di-II LD Aspherical [IF] (Model A16)

Leica Stereo Explorer. Visualizing and Analyzing Microstructures in 3D. Living up to Life

Product Range Overview

Shining advantages. The key to the inside of living cells: Leica MZ FLIII fluorescence stereomicroscope

Protocol for Microscope Calibration

Be careful not to scratch or hit front edge of the side viewing micro prisms onto hard objects!

Leica DM300 User Manual

CALIBRATION FOR LAL20X & LAL24X

Understanding astigmatism Spring 2003

A Fishy Tale. Observing the Circulatory System of a Goldfish with a Compound Light Microscope

9/16 Optics 1 /11 GEOMETRIC OPTICS

Vision Correction in Camera Viewfinders

WSD130 SOLDERING STATION

Chapter 1 Parts C. Robert Bagnell, Jr., Ph.D., 2012

The Kaps SOM -Series. Stereo Examination Microscopes for ENT Specialists. The variable space-saving system for every ENT practice

SP AF 300mm F/2.8 LD [IF] for Canon (Model 360EE)

OPERATING INSTRUCTIONS XJL 101/101A REFLECTED LIGHT METALLURGICAL MICROSCOPE

MACRO PHOTO LENS. MP-E 65mm f/ ENG. Instruction

Lenses and Telescopes

Under 20 of rotation the accuracy of data collected by 2D software is maintained?

Lecture 17. Image formation Ray tracing Calculation. Lenses Convex Concave. Mirrors Convex Concave. Optical instruments

The Krasnogorsk-3. Operating Instructions. Sold by NCS Products, New York

Chapter 27 Optical Instruments The Human Eye and the Camera 27.2 Lenses in Combination and Corrective Optics 27.3 The Magnifying Glass

PowerLine Magnetic-Laser Sheave Alignment System Instructions

First let us consider microscopes. Human eyes are sensitive to radiation having wavelengths between

1. Examine the metric ruler. This ruler is 1 meter long. The distance between two of the lines with numbers on this ruler is 1 centimeter.

Rodenstock Photo Optics

VALLIAMMAI ENGINEERING COLLEGE SRM NAGAR KATTANKULATHUR DEPARTMENT OF MECHANICAL ENGINEERING ME2304 ENGINEERING METROLOGY AND MEASUREMENT

National Optical & Scientific Instruments Inc Landmark 35 Drive San Antonio, Texas Phone (210) Fax (210)

Features: Mounting the Optic

AP Physics B Ch. 23 and Ch. 24 Geometric Optics and Wave Nature of Light

DEPENDABLE ACCURACY. RIGHT TO THE POINT.

Scanners and How to Use Them

RESOLUTION CHARTS AND GRATINGS RESOLUTION CHARTS AND GRATINGS RESOLUTION CHARTS AND GRATINGS RESOLUTION CHARTS AND GRATINGS RESOLUTION CHARTS AND

A new advance in routine inspections INVERTED MICROSCOPE CKX41/CKX31

EXPERIMENT O-6. Michelson Interferometer. Abstract. References. Pre-Lab

MICROSCOPE & OPTICAL CATALOG

Living up to Life. Leica KL300 LED. Compact, modular cold light source for routine stereomicroscopes

PlaneWave CDK Telescope Instructions CDK12.5, 17, 20 and 24

74 Chapter CO 2 Cleaning

Related topics: Application Note 27 Data Analysis of Tube Formation Assays.

EyeMag Medical Loupes from ZEISS Seeing is believing NEW! LED illumination

CS Digital Microscope User Manual

Spring Term. Lecturer: Assoc. Prof. Dr. M. Zeki COŞKUN. Department of Geomatics Engineering coskun@itu.edu.tr

Using an Abney Level to Measure Relative Heights

Exercise 2. The Compound Light Microscope

RAY OPTICS II 7.1 INTRODUCTION

Module Connectors and Cable Specifications

Physics 116. Nov 4, Session 22 Review: ray optics. R. J. Wilkes

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Leica M651. Manual surgical microscopes for microsurgical disciplines. Living up to Life

EF-S18-135mm f/ IS

HS-T RIFLESCOPE. riflescope manual. Second Focal Plane MOA

Handheld USB Digital Endoscope/Microscope

Instruction Manual Genesis 200 EQ

SLIT LAMP BQ 900 Sophisticated microscopy

Endoscope Optics. Chapter Introduction

EF-S18-55mm f/ IS II COPY ENG. Instruction

ENGINEERING METROLOGY

Bullets, documents, tools: Compare for yourself - with optimum precision

In this project, you will be observing at least three objects with a telescope or binoculars, and drawing what you see.

NIGHT VISION MONOCULAR PHANTOM

Fast Z-stacking 3D Microscopy Extended Depth of Field Autofocus Z Depth Measurement 3D Surface Analysis

Transcription:

0 1 2 3 4 5 6 7 8 9 10 11 Measuring User Manual

Accessories for measuring tasks Stage micrometer (1) for calibration Graticules with various measuring pitches (2) in mm and inches Graticule with mesh (3) Graticule with co-ordinate grid Lengths Counting Calibrating Angle measurements Eyepieces Software for measuring tasks are measured using the graticules with measuring pitches, e.g. 12mm : 120 or 5mm : 100. cells, nuclei and the suchlike within one particular area can be undertaken using the grid plate with mesh. When calibrating with the stage micrometer, the value of an interval is defined on the eyepiece grid plate in relation to the specimen magnification. can be taken with the aid of the graticule with co-ordinate grid and the angular division on the rotary polarisation table. Please refer to the instrument s directions for use with regard to using the graticule in eyepieces. Two adjustable eyepieces are required for equipment with measuring graticules so that the sharpness can be set precisely. If you take digital photographs with your stereomicroscope or macroscope, we would recommend the Leica IM image archiving and image management software with measuring module for interactive measuring, inscribing and marking. 2 Measuring User Manual

0 1 2 3 4 5 6 7 8 9 10 11 12 0.01 mm 12 mm : 120 2 0.1 mm 0.5 mm 1 mm 0 1 0 20 40 60 80 100 5 mm : 100 2 100 x 1 mm 2 3 50 mm 1 Care, cleaning Avoid corrosive cleaning agents and materials which will scratch the surfaces. Observe the information relating to care in the operating manual for your instrument. Do not keep unused graticules and stage micrometers in the original containers in which they were supplied. You will not cause any damage by undertaking cleaning with soft, dry hairbrushes bellows optical cleaning cloths cotton buds fine, soft, non-fluffing cloths: if removing fingerprints, dampen cloth; if dealing with stubborn dirt, soak cloth in alcohol. Measuring User Manual 3

Setting diopters and grid plate Co-ordinate the eyepieces precisely to your eyes as described until you can clearly see the graticule and specimen at the same time (free from parallax). Preparations Set distance from eye, 0 eyepieces, illumination etc. as described in the instrument s directions for use. Set «0» diopters on both eyepieces. Illuminate field of vision without specimen. 0 Use graticule to regulate eyepiece Do not look into the eyepieces. Turn eyepiece with graticule anticlockwise until it reaches the stop. Look into the eyepieces Slowly turn eyepiece with graticule clockwise until you can clearly see the measurement lines. No longer adjust eyepiece with grid plate. Focusing test specimen Place test specimen (e.g. stage micrometer) under the lens. 4 Measuring User Manual

Select the lowest magnification. Observe specimen through the eyepiece with graticule and focus. Select the highest magnification. Use focusing drive to optimise sharpness. Regulate eyepiece without graticule Do not look into the eyepieces Turn eyepiece without grid plate anticlockwise until it reaches the stop. Select the lowest magnification. Observe specimen through the eyepiece without graticule and slowly turn eyepiece clockwise until the eye clearly sees the specimen. Select the highest magnification. Observe specimen through both eyepieces. If necessary, use the focusing drive to re-focus. Checking eyepiece is free of parallax Look into the eyepieces and slightly move your head. Graticule and specimen must not move towards one another, i.e. they must be clearly focused in one plane (free of parallax). If this is not the case, please repeat the process. Measuring User Manual 5

Calibrating graticules Why calibrate? The graticule is only magnified by the eyepiece. The total magnification of the specimen does however depend on the lens, magnification changer, eyepiece and additional tube possibly used (e.g. coaxial lighting sources) and changes whenever these factors are altered. When calibrating with the stage micrometer, the value of an interval on the graticule is defined in relation to the specimen magnification. The true dimensions of a specimen can only be established once the calibrated value has been defined. You do not have to calibrate You have to calibrate Only calibrate once Engaging zoom settings if you only undertake relative measurements or comparisons. if you need absolute measurement values. The calibration value of each of your optics/ magnification combinations used only has to be established once. Keep a table of the appropriate calibration values (refer to example on page 12). When working with instruments with engageable zoom settings (Leica MZ6, MZ75, MZ95 etc.) or adjustable stops (Leica S6 models etc.), the specific magnifications can always be re-produced exactly (refer to directions for usage for the stereomicroscope). When using instruments with zoom magnification changers without engageable setting stages, inaccuracies could arise when again undertaking settings. If a high degree of measurement precision is required, you should therefore undertake calibration whenever changing magnification. 6 Measuring User Manual

Calibrating Place stage micrometer on the table insert and set sharpness. Select the magnification with which you will later want to take measurements. Set eyepiece graticule and lens micrometer to be free of parallax (see page 4). y 0 1 2 3 4 5 6 7 8 9 10 11 12 10 1 2 3 4 5 6 7 8 x Align stage micrometer (X) close to and in parallel with the eyepiece graticule (Y). Focus both pitches down to a single point. In the example: 0 and 10 Count: The number of mm (inches) of the stage micrometer (X) corresponds to a certain number of intervals on the eyepiece graticule (Y). In the example: 7,8mm on the stage micrometer corresponds to 120 intervals on the eyepiece graticule Calculate calibration value using calibration formula and note down. Calibration formula X Y stage micrometer: number of mm (inches) eyepiece graticule: number of intervals X Y mm (inch) = calibration value mm (inch) Measuring User Manual 7

Measuring, counting Measuring length 0 1 2 3 4 5 6 7 8 9 10 11 12 Remove the stage micrometer and focus on a specimen. Count how many intervals on the eyepiece graticule cover the measurement distance required. Multiply the number of intervals by the calibration value. The result is the absolute length of the measurement distance in mm (inch). Counting Remove the stage micrometer and focus on the specimen. Count particles (cells) within a known area. The area may comprise several squares. 8 Measuring User Manual

Examples 1. Measuring pitch 12mm : 120 y 0 1 2 3 4 5 6 7 8 9 10 11 12 10 1 2 3 4 5 6 7 8 x Stage micrometer with 0.1mm graduation Calibrating 7.8mm on the stage micrometer (X) 120 intervals on the eyepiece graticule (Y) 7.8 120 mm = 0.065mm calibration value 0 1 2 3 4 5 6 7 8 9 10 11 12 Measuring Intervals on the eyepiece graticule: 112 Result: 112 intervals 0.065mm = 7.3mm length of measurement distance Measuring User Manual 9

2. Measuring pitch 5mm : 100 y 0 20 40 60 80 100 0 1 x Stage micrometer with 0.1mm graduation Calibrating 1.37mm on the stage micrometer (X) 100 intervals on the eyepiece graticule (Y) 1.37 100 mm = 0.014mm calibration value 0 20 40 60 80 100 Measuring Intervals on the eyepiece graticule: 34 Result: 34 intervals 0.014mm = 0.47mm length of measurement distance 10 Measuring User Manual

3. Mesh pitch 100 1mm 2 y 9 x 20 1 2 3 4 5 6 7 Calibrating 6.2mm on the stage micrometer (X) 10 intervals on the eyepiece graticule (Y) 6.2 10 mm = 0.62mm calibration value A B Counting Area of boxed in square A: 0.38mm 2 Result: 4 cells per 0.38mm 2 Area of boxed in square B: 9.5mm 2 Result: 16 cells per 9.5mm 2 Measuring User Manual 11

Table Calibrating 1. Focus stage micrometer. 2. Count: number of mm (inches) on the stage micrometer (X) number of intervals on the eyepiece graticule (Y) 3. Calculate calibration value (one interval of the eyepiece graticule): X Y Instrument = mm (inch) calibration value Measuring 4. Focus on specimen rather than stage micrometer. 5. Count and calculate: number of intervals on the eyepiece graticule mm (inch) calibration value = Magnification stage Lens Tube factor mm (inch) length of measurement distance Eyepieces Graticule Calibration value Illustrations, descriptions and technical data are not binding and may be changed without notice. M2-275-0en Leica Microsystems (Switzerland) Ltd CH-9435 Heerbrugg, 2001 Printed in Switzerland V.2004 RDV Leica Microsystems (Switzerland) Ltd Stereo & Macroscope Systems CH-9435 Heerbrugg Telephone +41 71 726 33 33 Fax +41 71 726 33 99 www.leica-microsystems.com www.stereomicroscopy.com