A New PC Based Semi-Automatic TLD Badge Reader System For Personnel Monitoring

Similar documents
Submitted to Radiation Measurements INVESTIGATION OF OSL SIGNAL OF RESISTORS FROM MOBILE PHONES FOR ACCIDENTAL DOSIMETRY

Use of the VALIDATOR Dosimetry System for Quality Assurance and Quality Control of Blood Irradiators

THERMAL ANEMOMETRY ELECTRONICS, SOFTWARE AND ACCESSORIES

FREQUENCY RESPONSE ANALYZERS

BB-18 Black Body High Vacuum System Technical Description

The Ultimate British Valve Tester

Process Control Primer

Radiation Strip Thickness Measurement Systems

PC BASED PID TEMPERATURE CONTROLLER

Transmitter Interface Program

Temperature Controller CD401/CD901/CD701/CD501/CD100/CD101 Instruction Manual

ABB i-bus KNX Fan Coil Controller FC/S 1.1

Photovoltaic Solar Energy Unit EESFB

Electro-Pneumatic Bus Door Control Mechanism

TECHNICAL DATASHEET #TD1404AX PWM CONTROLLED SOLENOID DRIVER

AS A Low Dropout Voltage Regulator Adjustable & Fixed Output, Fast Response

A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION

Measuring Laser Power and Energy Output

IEC ESD Immunity and Transient Current Capability for the SP72X Series Protection Arrays

HIGH RELIABILITY POWER SUPPLY TESTING

Loop Calibration and Maintenance

Physical Specifications (Custom design available.)

Series AMLDL-Z Up to 1000mA LED Driver

Designed for multi-injector endurance testing, the

Measuring, Controlling and Regulating with labworldsoft

Four/Five Axis TB6560 CNC Driver Users Manual

Diagram of components 2. Reducer..3

TIG INVERTER INSTRUCTION MANUAL

Three Axis TB6560 CNC Driver Users Manual

Basic Technology Series

RI-215A Operator s Manual. Part Number: RK Revision 0 Released: 10/3/05

4-Channel Thermometer / Datalogger

Dyeing Programmer DP - 01

Power Amplifier Gain Compression Measurements

ABB i-bus EIB / KNX Analogue Input AE/S 4.2

Measuring Temperature withthermistors a Tutorial David Potter

Failure code manual. content

handbook, 2 columns handbook, halfpage 085 CS

Variable Capacity Compressors, a new dimension for refrigeration engineers to explore

Wahl C50 Calibrator FEATURES

Development of High-Speed High-Precision Cooling Plate

Technical Manual. FAN COIL CONTROLLER COOLING or HEATING ANALOG or PWM Art A

PCM Encoding and Decoding:

PORTER MASS FLOW MEASUREMENT AND CONTROL SYSTEMS

Adding Heart to Your Technology

Nuclear Physics Lab I: Geiger-Müller Counter and Nuclear Counting Statistics

NUCLEAR MAGNETIC RESONANCE. Advanced Laboratory, Physics 407, University of Wisconsin Madison, Wisconsin 53706

Tadahiro Yasuda. Introduction. Overview of Criterion D200. Feature Article

JUMPFLEX 857 Series ma V ma. Transducers / Relay and Optocoupler Modules

Impedance 50 (75 connectors via adapters)

Application Note AN1

Application Information Fully Integrated Hall Effect Motor Driver for Brushless DC Vibration Motor Applications

Evaluating AC Current Sensor Options for Power Delivery Systems

Technical Datasheet Scalar Network Analyzer Model MHz to 40 GHz

Cooking at the Speed of light!

Latest Power Plant Control System

S101D01/S101D02 S201D01/S201D02

PROFORMA ACCEPTANCE / QUALITY ASSURANCE TESTS OF REMOTE AFTERLOADING BRACHYTHERAPY UNIT

Basic RTD Measurements. Basics of Resistance Temperature Detectors

Melting Point. Electrothermal PRODUCT SPECIFICATIONS OPERATION

Technical Support Bulletin Nr. 20 Special AC Functions

Multi-Range Programmable DC Power Supplies 9115 Series

Split Controller. User Manual SM-SS-F 2009/01/04

A Study of Speed Control of PMDC Motor Using Auto-tuning of PID Controller through LabVIEW

notes MERRTT IS-302 Radiological Survey Instruments and Dosimetry Devices

NEBB STANDARDS SECTION-8 AIR SYSTEM TAB PROCEDURES

Microcontroller-based experiments for a control systems course in electrical engineering technology

Micro-Step Driving for Stepper Motors: A Case Study

THERMO KING TRUCK & TRAILER UNIT ALARM CODES THIS DOCUMENT SHOWS ALL CURRENT ALARM CODES FOR THERMO KING TRUCK AND TRAILER UNITS.

User Guide. Heavy Duty Dissolved Oxygen Meter. Model

RISH EM 3490 DS Dual Source Energy Meter RISH EM 3490 DS. Application : Product Features:

14.5GHZ 2.2KW CW GENERATOR. GKP 22KP 14.5GHz WR62 3x400V

OMNIYIG. .5 TO 2 GHz, 1 TO 4 GHz, 6 to 18 GHz Thin Film YIG-TUNED OSCILLATORS OUTLINE DIMESIONS MOUNTING SURFACE GND +15V + TUNE - HTR HTR

Online Infrared Flue Gas Analyzer Gasboard 3000

Production of X-rays. Radiation Safety Training for Analytical X-Ray Devices Module 9

EFC Frequency converters

Using Thermocouple Sensors Connecting Grounded and Floating Thermocouples

High Power Programmable DC Power Supplies PVS Series

Adding Pneumatic Preset Counter. Type 497. Continuously visible preset Integrated pneumatic reset 3 or 5-digit display Convenient button setting

Automation System TROVIS 6400 TROVIS 6493 Compact Controller

Study on the Auto Air Conditioning Controller Testing System Based on Lab Windows/CVI

In the previous presentation, we discussed how x-rays were discovered and how they are generated at the atomic level. Today we will begin the

Electric Coolant Pumps. Always at the Correct Temperature

Applications: X-ray Microtomography, Streak Tube and CRT Readout, Industrial & Medical Imaging X-RAY GROUP

LED DIMMING A HELPFUL GUIDE...

Programming Logic controllers

EVAL-UFDC-1/UFDC-1M-16

Tube Control Measurement, Sorting Modular System for Glass Tube

EASIDEW PORTABLE HYGROMETER INSTALLATION, OPERATION AND MAINTENANCE MANUAL

LS-1 Series Tungsten Halogen Light Sources Installation and Operation Instructions

Multimeter measurements on variable frequency drives using the new Fluke 289 DMM

A PIC16F628 controlled FLL (Frequency Locked Loop) VFO for HF

Ordering Part Numbers:

Automotive Sensor Simulator. Automotive sensor simulator. Operating manual. AutoSim

7-41 POWER FACTOR CORRECTION

INTEGRATED CIRCUITS DATA SHEET. TDA7000 FM radio circuit. Product specification File under Integrated Circuits, IC01

M 140i. Multifunction Calibrator. Operation Manual

Application Note, Rev.1.0, September 2008 TLE8366. Application Information. Automotive Power

Transcription:

A New PC Based Semi-Automatic TLD Badge Reader System For Personnel Monitoring M.S.Kulkarni, Ratna Pradeep & S.Kannan Radiation Protection & Instrumentation Section Radiation Safety Systems Division Bhabha Atomic Research Centre Mumbai 400 085 INDIA INTRODUCTION Personnel monitoring of more than 40,000 radiation workers in India is being carried out routinely using a Bhabha Atomic Research Centre (BARC) made thermoluminescent dosimeter (TLD) badge based on CaSO 4 :Dy phosphor embedded in Teflon discs [1], and a photographic film badge. Till recently the TLD badges were read using the manual type TLD badge reader system with an external interface unit for connection to a PC for glow curve storage and dose evaluation [2]. The film badge is being replaced by the TLD due to the operational convenience and the low cost of the latter resulting in a very significant increase in the number of persons being monitored using TLD. A new PC based semiautomatic TLD badge reader system with a number of design improvements for faster processing of the TLD badges with reduced man power has been developed. The new reader has features such as non contact N 2 gas heating, reproducible temperature profile of the gas, built-in safeguards against gas flow, heater and EHT failures, dark current compensation, background subtraction, auto ranging, automatic switching of EHT to a lower value when high TL output is encountered to enable a wide range of measurement with high sensitivity to lower levels of TL, etc. An elaborate software is written in C language for storage and recall of glow curves, evaluation of doses and printout of dose reports. INSTRUMENT DESCRIPTION The semiautomatic TLD badge reader system is based on a Personnel Computer with all the front end functions in the reader controlled by an Atmel 89C51 microcontroller which is connected to PC through an RS-232 serial interface. Fig. 1 shows the functional block diagram of the reader. The reader comprises of a microcontroller based electronic control circuits, TLD card transport mechanism for moving the TLD badges to the reading position, PMT housing, gas heater & an on-off proportional temperature control circuit and a solenoid for switching the gas flow on & off. The entire operation of the reader is controlled by a menu driven, user friendly, software residing in the PC. An assembly language software in the 89C51 microcontroller controls the various motorized movements, monitoring of various circuits, generation of EHT for the photo-multiplier tube (PMT), measurement of PMT anode current, etc., on command from the PC. The movement and the positioning of TLD cards inside the reader is achieved using three compact DC motors and the positioning is accurately sensed using opto-electronic sensors. The circuits are opto-isolated from the rest of the circuit to prevent motor driver noise from affecting the rest of the circuit. The motors are driven in both forward and reverse direction by a software controlled pulse width modulated outputs from the microcontroller for achieving precise positioning. A anodized aluminum magazine (maximum capacity of 50 TLD cards) loaded with the TLD cards is moved in a U channel into the reader by a Y motor which also raises the card for pickup and placement. A motor X is used to move the TLD badge to the reading position. Sensing of the different motor positions is done using opto-interrupters and the presence / absence of the dosimeter card in the magazine is sensed by using an opto-reflector. Another DC motor Z is used for moving a shutter to prevent infra-red and stray light from reaching the photo cathode of the PMT, when the TLD card is not in the reading position. This dc motor is also used for bringing a built-in check light source (in 63 Ni incorporated in a plastic scintillator) to the reading position for checking the PMT sensitivity. A high stability temperature compensated photo multiplier tube type EMI 9125B is used for light detection. A software controlled high voltage power supply (EHT, -200V to 1200V) with a load regulation and stability better than 0.05% is designed for the PMT. A control circuit is used to switch this EHT from a high to a low value automatically, under the software control, when the dosimeter reading exceeds 200,000 counts (equal to 0.2Sv of Cs 137 gamma dose) such that the gain of the PMT is reduced by a factor of 100. This is done to safeguard the PMT from exceeding the recommended maximum anode current of 50 µa under high radiation dose condition and to cover a range of 1 µsv to 10 Sv. A current to frequency (I-F) converter is used to convert the PMT anode

current into a train of pulses whose frequency is proportional to the current. The I-F converter has a linearity of ±1% from 1nA to 10 µa (10Hz to 100 KHz). A dual on/off proportional temperature controller is designed for uniform and rapid heating of the TLD badges. Cold N 2 gas is forced through a Nichrome coil heater at a flow rate of 5LPM to get the hot gas at the outlet of the heater. The heater is maintained at 300 o C using feedback from a thermocouple located inside the heater during the idling period when readout of TLD cards is not in progress. The temperature of the hot gas is raised to 280 o C within 5 seconds and clamped at 280 o C till the end of the readout time, using feedback from a second thermocouple located in the hot gas stream below the exit nozzle. A solenoid valve is used to switch the gas supply On & Off at the cold end of the heater. The glow curves obtained are uniform. A typical glow curve recorded for a TLD badge along with temperature time profile is shown in fig.2. As can be seen from the glow curve a read out time of 30 seconds is adequate. The left over TL is found to be less than 10% by the second readout of the dosimeter cards. The read out time per badge (containing three 0.8mm thick TL dosimeters) is 90 secs and 50 badges are read in less than 90 minutes. SOFTWARE A user-friendly menu driven software package has been developed for the semiautomatic TLD badge reader system. The package has features like creating data files for storing glow curve data, generation of the personnel & institution data files, data analysis, recall of glow curves, dose evaluation, printout of dose reports, record-keeping, etc. A 16 digit identification code is assigned for each user. This code contains entire information of the person wearing the TLD badge, which includes the institution number to which the user belongs and his personal number, service year and service month, frequency of service, nature of radiation environment in which the user is working (eg. gamma, beta/gamma, X/gamma, X-ray, etc), and the location code (indicating where the TLD badge is worn). The dose evaluation is carried out using empirical relations depending upon the code entered and a printout of dose report can be obtained. Once all the TLD cards are loaded in the magazine (maximum 50 cards), the start reading cycle can be invoked. This prompts the entry of the dose data file name and the 16 digit identification numbers serially in the same order in which the badges are loaded in the magazine. After this code is entered the reading cycle begins and the solenoid valve opens at the inlet of the gas and the heating process starts. The gas temperature is continuously monitored and displayed on-line on the screen along with the glow curve and temperature profile. The entire glow curve data is stored in the hard disc at the end of the reading cycle. The reader is calibrated either by changing the EHT through the software (which in turn varies the gain of the PMT), or by entering the calibration factor for fine adjustment. The calibration factor is obtained using a set of dosimeter cards exposed to a known dose and reading them in the reader. An optimum EHT (900 1200V) is usually selected for the PMT to obtain the best signal to noise ratio. After the readout is over, the readings are normalized and mean dose and standard deviation is calculated using the software option. The calibration factor is then the ratio of the actual dose given to the TLD badges to the mean dose. DETECTION THRESHOLD & LINEARITY The detection threshold the lowest dose that can be measured with the reader system (also defined as 2.26 times the standard deviation (σ ) of the annealed dosimeter readings as per the latest International Electro technical Commission draft [4] is improved in the reader by increasing the readout resolution to 1µSv, minimizing the infrared background using a infrared filter, reducing non-radiation induced background by using N 2 gas for heating. A high stability, temperature compensated, low dark current (0.1nA) PMT tube is used in the reader system. Before the readout of each dosimeter badge the dark current is measured and stored in the PC and subtracted from subsequent dosimeter readings. The EHT is switched from a high to a low value automatically when the dosimeter reading exceeds 200,000 counts (approximately 0.2Sv of 137 Cs gamma dose) so as to reduce the gain of the PMT by a factor of 100. This is done to safeguard the PMT from the fatigue when dosimeters of high radiation dose are read. A range of 1 µsv to 10 Sv for the reader system is thus achieved. To determine the detection threshold, a set of 30 randomly selected TL dosimeter badges (used regularly for personnel monitoring) were annealed and read in the reader for three times. A mean background reading of 63.17 µsv ± 22.07 µsv (mean σ) equivalent was obtained from these readings. The observations are shown in the table1. However, when the same experiment was done with the fresh TLD badges having minimum sensitivity variation, a mean background of 17.5 µsv with a std. deviation of 10 µsv was found, indicating the minimum detection threshold can be improved to 1mR using the TLD badge

reader system. Table 1 also shows the reader response for a test dose of 5 msv and the standard deviation and residual TL which is well within 10%. Cycle Number Average dose(µsv) for annealed badges Standard Deviation Test Dose Average dose(µsv) of test badges Standard Deviation Residue 1 63.26 22.13 5mSv 5505 405.12 8.9% 2 63.34 21.90 5mSv 5506 404.91 8.8% 3 62.92 22.19 5mSv 5511 405.01 8.7% Table 1: Background reading and standard deviation of the routinely used TLD badges indicating the minimum measurable dose and the reader response for a test dose of 5 msv along with the std. deviation and residual TL. A stable and reproducible built-in light source made up of 63 Ni coupled to a plastic scintillator [5] is used to check the performance of the light detection system. To check the linearity of response of the TL dosimeter output with respect to dose, the TLD badges were exposed to the different levels of gamma dose from 60 Co radiation field. The dose (in Sv) verses the TL output is found to be linear from 50 µsv to 10Sv. Table 2. shows the response of TL dosimeters read in the reader system for various doses. The results were found to be reproducible in the various personnel monitoring centers using the above PC based semiautomatic TLD badge reader system. A study was made to find the reusability of TLD badges with the new reader system. For this, a fresh set of TLD badges were subjected to more than 50 cycles of anneal, exposure and readout in the reader system. It was found that there is no significant change in the sensitivity and shape of the TLD badges even after 50 cycles (table 3). These studies were repeated by the other groups working on these TLD reader system, and the results were found to be similar. Dose Level 50 µsv 500 µsv 1 msv 100 msv 1 Sv 10 Sv Response 49.22 518.28 965.45 97885 998664 9648780 Table 2. Response of TL dosimeters read in the reader system for various doses. Cycle 1 10 20 30 40 50 Number Sensitivity 1.01 1.05 1.055 1.04 1.042 1.041 Residue 8.1% 7.98% 7.79% 7.70% 8.32% 8.01% Table 3. Sensitivity of TLD badges used with the reader system RESULTS & CONCLUSION The new PC based semiautomatic TLD badge reader system facilitates a fast readout of TLD badges with a resolution of 1µSv and covers a large dose range of 50µSv to 10 Sv without any range switching. The detection threshold is reduced to 50 µsv for the routine personnel monitoring badges. The genuineness of the dose can be readily verified using the glow curve recall facility. The hot N2 heating of the TLD badges results in the reduction in readout time for 0.8 mm thick TL dosimeter to 30 sec and 0.4 mm thick dosimeters to less than 20 secs and increases the life span of the dosimeters. Operator related errors are completely eliminated due to direct transfer of data to the PC. Glow curve with the temperature profile for any dosimeter can be readily recalled when required for analysis.

REFERENCES 1) K.G.Vohra, R.C.Bhatt, Bhuwan Chandra, A.S.Pradhan, A.R.Lakshmanan & S.S.Shastry, Health Phys. 38 (Feb) (1980), 193 2) M.S.Kulkarni, P.Ratna, S.Kannan & T.V.Venkateswaran, Nuclear Instruments and Methods in Physics Research A 334 (1993) 512-515 3) P.Christensen & R.V.Griffith, Radiat. Prot. Dosim., 54, 279-285 4) International Electrotechnical Commission (IEC) (1997) Draft/IEC 45B 77 5) J.S.Nagpal, J. Udaykumar, A.G.Page & G.Venkatraman, Radiat. Prot. Dosim., 60, 181-184 6) S.Kannan, M.S.Kulkarni, P. Ratna & M.D.Lalsare, Proc. 23 rd IARP Conference, Feb. 19-21, 1997. 7) G.Varadharajan, A.K.Bakshi & A.S.Pradhan, Proc. 24 th IARP Conference, January 1999

magazine TC Heater temp. TEMPERATURE CONTROLLER 89C51 Microcontroller SYSTEM Y Motor TC for gas temp. RS-232 X motor Pno EHT TEMP D1:3547 Z motor I-F Converter EHT Fig.1 BLOCK DIAGRAM OF THE PC BASED TLD BADGE READER SYSTEM Fig.2 A typical glow curve recorded for a TLD badge along with temperature time profile